Introduction
This series of monitors includes the Thermo Scientific TEOM 1405, a standard style TEOM
monitor, the Thermo Scientific TEOM 1405-D, a dichotomous sampler without FDMS
technology, the Thermo Scientific TEOM 1405-F, a single channel TEOM monitor with FDMS
technology, and the Thermo Scientific TEOM 1405-DF, a continuous dichotomous ambient air
monitor with two FDMS. This technical note provides a comparison between the standard
TEOM 1405 monitor with a similarly equipped TEOM 1400ab monitor.
To verify the performance of the TEOM 1405, three 1405 and two 1400 monitors were
collocated and operated in a sampling trailer at the Thermo Fisher Scientific facilities in
Franklin, MA. The inlets of all monitors were separated at one meter spacing and all inlets were
approximately 1.5 meters above the roof of the trailer. The results from all samplers were
compared over a two week time period when sampling PM2.5. Both continuous data and daily
averages were compared for the two types of instruments.
The Thermo Scientific sampling trailer with the Series 1405 and Series 1400
TEOM monitors installed is pictured above. The trailer is located at the Thermo
Fisher Scientific facility in Franklin, MA.
The results are presented in graphs below. The first graph shows a representative plot of the
rolling 1-hour average mass concentrations of the three 1405 TEOM monitors and the two 1400
TEOM monitors over a 12-hour period. The results show that all three TEOM 1405 monitors and
the two TEOM 1400 monitors provide similar results over this representative test period. This
segment is presented as the ambient mass concentrations range from less than 5 µg/m3 to over 30
µg/m3.
50
TEOM 1405 Unit 009
45 TEOM 1405 Unit 123
40 TEOM 1405 Unit 124
Mass Concentration (ug/m^3)
25
20
15
10
0
30.July.08 30.July.08 30.July.08 30.July.08 31.July.08 31.July.08 31.July.08 31.July.08 31.July.08
12:00 15:00 18:00 21:00 00:00 03:00 06:00 09:00 12:00
Date/Time
Comparison of hour average mass concentration values for three Series 1405
TEOM monitors and two Series 1400 TEOM monitors when collecting PM2.5.
The monitors were installed and tested at the Thermo Fisher Scientific facility in
Franklin, MA.
20
15
10
y = 1.0214x - 0.2179
R2 = 0.988
5
0
0 5 10 15 20 25
Series 1400 TEOM Mass Concentration (ug/m^3)
Comparison of daily average values from three Series 1405 TEOM monitors and two Series
1400 TEOM monitors during July and August 2008. The monitors were installed and operated at
the Thermo Fisher Scientific facility in Franklin, MA.