Chemical Fact File ®

Fluorides (as F) — OSHA ID 110


OSHA Method ID 110 is applicable to the collection and analysis of fluoride containing particulates or hydrogen fluoride (HF) in the
workplace. This method is suitable for eight-hour TWA or 15-minute STEL determinations. HF and fluoride containing particulates are
used in aluminum production, stainless steel pickling, glass etching, oil well acidizing, gasoline production, and uranium production
processes.

Required Equipment: 2. For calibrating the pump, use the sampling train described
1. An air sampling pump capable of sampling at the recom- above remove the inlet plug of the filter cassette, and connect the
mended flow rate with the sampling medium in line, such as: inlet of the filter cassette to the calibrator with flexible tubing.
Calibrate the pump flow rate to the rate specified in the method.
• SKC 224-XR Series Universal Sampler When calibration has been completed, remove the filter cassette,
• SKC AirChek® 2000 Sampler reinsert the plugs, and save the filter cassette for recalibration at the
• SKC AirChek 52 Sampler end of sampling. Request SKC Application Guide #1366 or #1163
for more information on calibrating a pump.
2. An air flow calibrator, such as:
3. For sampling, set up a sampling train as above except use a
• SKC UltraFlo® Calibrator Cat. No. 709
new filter cassette and do not remove the inlet plug on the filter
• DC-Lite Flowmeter Cat. No. 717-01
cassette inlet until ready to sample. The inlet of the filter cassette
should be facing down. Attach the cassette holder clip to a worker's
3. SKC Coated Filter Cassette Cat. No. 225-9001
collar and the pump to the worker's belt. Remove the filter cassette
inlet plug and turn the pump on.
4. SKC Filter Cassette Holder Cat. No. 225-1
4. Sample at an accurately known flow rate for the recommended
Optional Equipment: period of time.
1. SKC Cassette Sealing Bands Cat. No. 225-25
5. At the end of the sampling period, turn the pump off and note
SKC Application Guides: the ending time. Cap the inlet and the outlet of the cassette with the
1. Sampling Train—Filters, #1166 plugs provided. Record any pertinent sampling information.
2. Calibrating a Pump Using a Film Flowmeter, #1163
3. Calibrating a Pump Using an Electronic Calibrator, #1366 6. Calibrate the pump with the representative sampling media in
line to verify that the flow has not changed by more than 5%.
TWA STEL
7. Submit one blank with every ten samples from the same lot
Flow Rate 1500 ml/min 1500 ml/min
number as the sample filters. Blanks should be subjected to exactly
Sample Time 1 hour 15 minutes
the same handling as the samples (load, seal, and transport) except
Air Volume 90 L 22.5 L
that no air is drawn through them.
OSHA PEL 2.5 mg/m3 None
— Hydrofluoric acid 3 ppm None
7. Pack sample filters, field blanks, and all pertinent information
(per 29 CFR 1910-1000)
securely for shipment to a laboratory for analysis.

Sampling and Analysis: Analyzing Method:


1. To set up a filter cassette sampling train, prepare a coated filter Ion-specific electrode (ISE)
cassette, seal it with a sealing band (optional), insert the filter
cassette into the filter cassette holder with the cassette inlet facing Copyright 1992 - 2002
down, and remove the outlet plug. Connect the fitting on the end of
the rubber tubing of the filter cassette holder to the filter cassette
outlet. With the flexible tubing provided, connect the filter cassette
holder to the pump inlet. Request SKC Application Guide #1166 for
more information on preparing filter sampling trains.

Notice: This publication is intended for general information only and should not be used as a substitute for reviewing applicable government regulations, equipment operating instructions, or legal standards. The information
contained in this document should not be construed as legal advice or opinion nor as a final authority on legal or regulatory procedures.

Publication 1227 Rev 0208