Suction
Suction
needed
daily Wash jar and patient tubing with sterilizing Detergent. user
solution Disinfectant.
Clean filters
Clean air vents
Sterilize jars, tubing, other components that
come into contact w/ patient fluids between
each use in solution of water, detergent,
and disinfectant
weekly Ensure electrical plug and cord are in good Bacteria filter. User
condition BME/T
Check collection bottle/jar for cracks,
chips, and other damage
Change bacteria filter if wet or discolored
Semi Clean or replace air intake filter Air intake filter BME/T
annually Clean brushes on motors as necessary Electrical safety
Inspect power cord and plug analyzer.
Ensure vacuum works over full range of
suction pressures if there is a control/knob
Verify that overflow valve (float valve)
works properly when container is filled
with water
Grounding resistance between chassis and
ground pin should not exceed 0.5 ohms
PPM for Suction machine.