Period Task activity Materials personnel

needed
daily  Wash jar and patient tubing with sterilizing Detergent. user
solution Disinfectant.
 Clean filters
 Clean air vents
 Sterilize jars, tubing, other components that
come into contact w/ patient fluids between
each use in solution of water, detergent,
and disinfectant

weekly  Ensure electrical plug and cord are in good Bacteria filter. User
condition BME/T
 Check collection bottle/jar for cracks,
chips, and other damage
 Change bacteria filter if wet or discolored

 Make sure there is a sufficient supply of


bacterial filters
 Check that float valve moves freely
 Ensure anti-static tubing is used

Semi  Clean or replace air intake filter Air intake filter BME/T
annually  Clean brushes on motors as necessary Electrical safety
 Inspect power cord and plug analyzer.
 Ensure vacuum works over full range of
suction pressures if there is a control/knob
 Verify that overflow valve (float valve)
works properly when container is filled
with water
 Grounding resistance between chassis and
ground pin should not exceed 0.5 ohms
PPM for Suction machine.