Professional Documents
Culture Documents
Bilirubin-D Mindray bs-300
Bilirubin-D Mindray bs-300
Bilirubin-D Mindray bs-300
Package Size
[REF] 10741 375 ml auto-BILIRUBIN-D liquicolor Reagent kit
13160 4 x 5 ml AUTOCAL
Recommended control serum
13511 6 x 5 ml HumaTrol N
13512 6 x 5 ml HumaTrol P
13951 6 x 5 ml SERODOS
13151 6 x 5 ml SERODOSplus
[IVD]
R1 = [RGT1], ready for use
R2 = [RGT2], ready for use
¼ For detailed information kindly refer to the most recent package insert.
Calibration
Calibration frequency
User defined; re-calibration is recommended
• after reagent lot change
• as required following quality control procedures.
Control frequency
It is recommended that 2 levels of control be assayed daily.
Parameters
No # Test DBil Method Endpoint Direction Increase
Unit mg/dl Decimals 2 Prim. Wave 546 Sec. Wave 0
Sample vol 20 R1 Vol 200 R2 Vol 50 Line Limit 20
Calibration
Request
Rule Two point linear
Replicates 2 Difference 0
Interval 0 Black Response 0/0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0