GSI CORTI QUICK GUIDE

This Quick Guide is provided as a reference for initiating testing with the GSI Corti™. It is recommended that the user refer to the Corti User Manual for full instructions and information on the operation of the Corti.

Corti Quick Guide: Screening Corti Quick Guide: Diagnostic Test 7. To print results, press the up button on the navigation panel. The
While the Corti is in the off position, connect the HDMI connector of While the Corti is in the off position, connect the HDMI connector of display will indicate the wireless printer that is paired with the
the Probe to the Corti hand-held unit. Make sure the probe cable is the Probe to the Corti hand-held unit. Make sure the probe cable is Corti. Press the right or left button on the front panel button to
fully inserted. fully inserted. connect and initiate printing.
NOTE: All results that are stored in the Corti will be printed.
Probe Tip Placement Probe Tip Placement NOTE: The Corti will automatically power down after printing.
1. Place a clean probe tube into the micro probe. 1. Place a clean probe tube into the micro probe. 8. To proceed to the next evaluation, refer to Number 1 of Performing
Press the tip into the probe to ensure the tube is Press the tip into the probe to ensure the tube is a Diagnostic Test.
completely seated. completely seated.
2. Place the ear tip over the probe 2. Place the ear tip over the probe Quick Guide to Button Func-
tube. The ear tip must be flush with tube. The ear tip must be flush with tions
the probe. the probe.
Navigation Buttons
Performing a Screening Test Selecting the Protocol Down Button
1. Insert the probe with ear tip into the ear canal. 1. If the unit is off, power on the unit by pressing the down button • Single press – turns the Corti power on
2. Power on the Corti by pressing the down button on the front on the navigation panel. • Single press – enters the protocol screen
panel. 2. Press the down button (Change) one time to enter the protocol and the menu screens
3. Press the right button to select the right ear test, or press the left menu. Right Button
button to select the left ear test. 3. Use the right or left button to scroll through the Corti protocols. • Main screen – starts a test on the right
The protocol will be listed in the center of the display screen. ear
4. The Corti performs an automatic probe and calibration test.
• All screens – scrolls through options in
5. The screening automatically starts. The results are viewed on the 4. When you reach the desired protocol, press the up button on the
the selected menu item
OLED screen. navigation panel to return to the testing screen.
Left Button
6. When the test is completed, the screen will display a “Pass, Refer, 5. Note the selected protocol is listed in the center of the display
• Main screen – starts a test on the left ear
Noisy, No Seal or Fit Err”. If a “Noisy, No Seal or Fit Err” is screen.
• All screens – scrolls through options in
displayed, re-adjust the probe placement and retest. Performing a Diagnostic Test the selected menu item
7. To review results, press the down button on the navigation panel. 1. If the unit is off, power on the unit by pressing the down button Up Button
NOTE: To review test results, press down button before moving on the navigation panel. • Long Press – turns the Corti power off
to the next test (opposite ear). If you do not review the results, • Single Press – moves up one level in the menu structure
2. Press the right button to test the right ear; press the left button to
they will be unavailable for viewing until the results are either • Single Press – moves to the print menu at the completion of a test
test the left ear.
printed or transferred to the database.
8. To print results, press the up button on the navigation panel. The 3. The Corti performs an automatic probe and calibration test. Menu Structure
display will indicate the wireless printer that is paired with the 4. The testing automatically starts. The results are viewed on the 1. Power Up Screen – displays software and firmware version of the
Corti. Press the right or left button on the front to connect and OLED screen. Corti.
initiate printing. 2. Main Screen – displays the selected test protocol and R and L for
5. When the test is completed, the Corti will return to the main test
NOTE: The Corti will automatically power down after printing. screen. The results are ready for interpretation. If a “Noisy, No Seal testing each ear; allows scrolling and selection of the stored
9. To proceed to the next screening test, insert the probe into the or Fit Err” is displayed, re-adjust the probe placement and retest. protocols.
opposite ear. Press the left button to test the left ear or the right 3. Date/Time Screen – displays the date and time and allows
6. To review results, press the down button on the navigation panel.
button to test the right ear. changes via the left and right arrows.
NOTE: To review test results, press review (down button) before
4. System Screen – displays system operation settings and allows
moving to the next test (other ear). If you do not review the results,

8104290 Rev 5
changes via the left and right arrows.
they will be unavailable for viewing until the results are either
printed or transferred to the database. 5. Protocol Screens – displays the protocol settings and allows
changes via the left and right arrows. Changes to settings are
available on the diagnostic Corti unit.

Setting The Clinical Standard www.grason-stadler.com l Customer Service: 800-700-2282