Sensymaster Fmt430, Fmt450: Thermal Mass Flowmeter
Sensymaster Fmt430, Fmt450: Thermal Mass Flowmeter
—
Measurement made easy
Precise and dynamic direct mass flow
measurement of gas
—
Efficient top-quality thermal sensor elements
• Single-chip design on ceramic carrier material for top long-
term stability
• Effective protective frames of the sensor elements with flow-
forming characteristics for top repeat precision
—
Powerful electronic unit on a common ABB platform
• Quick commissioning with Easy Setup
• Easy operation thanks to common ABB operation concept
• Modular inputs and outputs with plug-in cards
• ‘Plug-and-play’ electronic unit swap with
SensorApplicationMemory
• Top precision with dynamic temperature compensation
—
ApplicationSelector
• Up to 8 configurable applications for top flexibility
—
Integrated diagnosis and device verification
• Low costs thanks to longer maintenance cycles
• Higher system availability with preventative maintenance
• Process security thanks to sensor element verification
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 3
—
Overview – models
Sensor
—
… Overview – models
Process connections
FMT091 – Wafer type design FMT094 – Weld-on adapter FMT094 – Weld-on adapter with clamp ring threading
Pipe components
FMT091 – Wafer type design In accordance with EN 1092-1: DN 40 to 200, PN 40
In accordance with ASME B16.5: 1½ to 8 in, CL 150 to 300
FMT092 – Partial measuring section Flange in accordance with EN 1092-1: DN 40 to 100 (larger nominal diameters on request), PN 10 to 40
Flange in accordance with ASME B16.5: 1½ to 8 in, CL 150 to 300
Male thread DN 25 to 80, R1 to 3 in
FMT094 – Weld-on adapter For rectangular ducts or pipe diameters ≥ DN 100 (4 in.), PN 16 to 40
Wetted materials
Stainless steel, galvanized
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 5
Transmitter
—
… Overview – models
Device description Measuring principle
The SensyMaster FMT430, FMT450 works in accordance with Thermal flow metering procedures use different ways to
the measuring principle of a hot-film anemometer. This evaluate the flow dependent cooling of a heated resistor as
measurement method allows for direct measurement of the measuring signal.
gas mass flow. In a hotfilm anemometer with constant temperature
Taking into account the standard density, the norm volume difference control, the heated platinum resistor is maintained
flow can be displayed without the need for additional at a constant overtemperature in relation to an unheated
pressure and temperature compensation. platinum sensor inside the gas flow.
The transmitter is equipped with an analog / HART® output
(4 to 20 mA) and two fast digital outputs that can be The heating power required for maintaining the
configured as pulse, frequency or binary outputs. overtemperature depends directly on the flow rate and the
Optionally, the transmitter can be extended using plug-in material properties of the gas. With a known (and constant)
cards with further inputs and outputs as well as a Modbus® gas composition the mass-flow can be determined by
or PROFIBUS DP® interface. electronically evaluating the heater current / mass-flow curve
The SensyMaster FMT430, FMT450is used in the process without additional pressure and temperature compensation.
industry for the flow measurement of gases and gas Together with the standard density of the gas this results
mixtures. directly in the standard volume flow.
Considering the high measuring range dynamics up to 1:100,
an accuracy smaller than 1 % of the measuring value is
achieved.
2 Sensor connection
The transmitter has three signals available. In addition to the
3 Thermal sensor elements
heating power, the temperatures of the measuring medium
Figure 4: Sensor (example, wafer type design) and the heater resistance are included herein, which can be
used to compensate the temperature dependency of gas
The SensyMaster FMT430, FMT450 is composed of the parameters.
components sensor and pipe component (process
connection). By storing the gas data in the transmitter the optimal
The pipe component can be delivered in various designs. In tailoring can be calculated and performed at any operating
addition, a weld-on adapter makes it possible to install the point.
flowmeter sensor in rectangular ducts or pipelines with any
diameter.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 7
optimum measuring dynamic can always be achieved even Biogas Type 1* Helium – He
• The SensyMaster measuring principle can offer an even Argon – Ar Hydrogen sulphide– H2S
larger measuring range. Natural gas Type 1* Methanol – CH3OH
Acetone – C3H6O Neon – Ne
Typical applications Acetylene – C2H2 Nitrogen oxide – NO
• Gas volume measurement in the chemical industry and Ammonia – NH3 Ozone – O3
process technology 1.2-Butadiene – C4H6 Pentane – C5H12
• Pressurized air balancing 1.3-Butadiene – C4H6 Propadiene – C3H8
• Gas burner controls Butane – C4H10 Propylene – C3H6
• Digester gas and activation air measurements in sewage 1 Butane – C4H8 Water vapor – H2O
plants Carbon monoxide – CO
• Gas measurement in air separators * In the case of natural gas and biogas, the exact composition must be
• Hydrogen measurements in the process provided.
Note
ApplicationSelector – integrated data bank In the case of hydrogen and helium gas types as pure gas or
for gases as a component of a gas mixture with a concentration of over
10 %, an optional process gas calibration should always be
The thermal mass flowmeter SensyMaster FMT430, FMT450
additionally ordered.
has an integrated gas data bank (see table Available gas
That way, additional measuring errors due to the special
types).
characteristics of gases are avoided.
—
… Overview – models
Diagnosis and self-monitoring FillMass batch function
The thermal mass flowmeter SensyMaster FMT430, FMT450 Only for FMT450
also includes the internal monitoring of the transmitter and
the sensor.
Amongst other things, the following functions and
components are monitored:
• Monitoring of the power supply
• Limit value monitoring of the process values,
temperature monitoring of the measuring medium
• Monitoring of the sensor element for line break and
short-circuit
• Monitoring of the SensorApplicationMemory
Diagram legend
VO Valve open (filling started)
VC Valve closed (fill quantity reached)
t1 Valve closing time
t2 Overrun time
—
Flowmeter sensor
Installation conditions
Installation location and assembly Inlet and outlet sections
Note the following points when selecting the installation The figures below show the recommended inlet and outlet
location and when mounting the sensor: sections for various installations.
• The ambient conditions (IP rating, ambient
temperature range Tambient) of the device must be
adhered to at the installation location.
• Sensors and transmitters must not be exposed to
direct sunlight. If necessary, provide a suitable means
of sun protection on site. The limit values for ambient
temperature Tambient must be adhered to.
• On flange devices, ensure that the counterflanges of
the piping are aligned plane parallel. Only install flange
devices with suitable gaskets.
• Prevent the sensor from coming into contact with
other objects.
• The device is designed for industrial applications.
No special EMC protective measures are required if
the electromagnetic fields and interference at the
installation location of the device comply with ‘Best
Practice’ (in accordance with the standards listed in
the declaration of conformity).
Maintain a suitable distance from electromagnetic
fields and interference that extend beyond the usual
dimensions.
Figure 7: Inlet and outlet sections
Seals
Users are responsible for selecting and mounting suitable Installation Inlet section Outlet section
gaskets (material, shape). A Pipe extension min. 15 × DN min. 5 x DN
Note the following points when selecting and mounting
B Pipe reduction min. 15 × DN
gaskets:
C 90° Pipe elbow min. 20 × DN
• Use gaskets made from a material that is compatible
D 2 × 90° pipe elbow in one min. 25 × DN
with the measuring medium and measuring medium
level
temperature.
E 2 × 90° pipe elbow in two min. 40 × DN
• Gaskets must not extend into the flow area, since
levels
possible turbulence may influence the accuracy of the
F Turn-off device min. 50 × DN
device.
10 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Flowmeter sensor
To achieve the specified measuring accuracy, the indicated Installation at high ambient temperatures
inlet and outlet sections are required.
In case of combinations of several inlet-side errors, e.g. valve
and reduction, a longer inlet section must always be taken
into account.
1 Insulation
Measuring accuracy
Measured error Reference conditions
The stated measured error only applies under the reference
Calibration with air
conditions in the stated measuring range. Special calibration
Calibration gas Air
on request.
Temperature 21 °C, ±2°C
Designation of gas Atmospheric pressure
connections
Relative humidity 40 to 60 %
Test laboratory In accordance with ISO / IEC 17025
A Measured error in %
The exact reference conditions are noted in the respective
1 FMT430
calibration certificate.
2 FMT450
—
… Flowmeter sensor
Influence of the pipe cross-section Process conditions
If the inside diameter configured in the device does not
Measured medium temperature
correspond with the real diameter of the piping, measuring
Devices with ceramic element and flange connection
errors in the flow measurement occur.
• Standard and explosion-proof design:
−20 to 150 °C (−4 to 302 °F)
• High temperature design:
−20 to 300 °C (−4 to 572 °F)
• DVGW design:
0 to 70°C (32 to 158 °F)
Relative humidity
Maximum 85 % RH, annual average ≤ 65 % RH
IP rating
In accordance with EN 60529: IP 65 / IP 67
NEMA IP rating
NEMA 4X
—
… Flowmeter sensor
Measuring range table
The recommended value for applications with air or nitrogen (other gases on request) under atmospheric conditions. For
hydrogen and helium, the measuring range lower limit is typically approx. 10 % of the upper limit.
Note
For additional information on dependencies and restrictions, and for help on product selection, please refer to the Online ABB
Product Selection Assistant (PSA) for flow rate at abb.com/flow-selector.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 15
—
Transmitter
Optional plug-in cards
The transmitter has two slots (OC1, OC2) into which plug-in
cards can be inserted to provide additional inputs and
outputs.
The slots are located on the transmitter motherboard and
can be accessed after removing the front housing cover.
1 Single-compartment housing
2 Dual-compartment housing
Features
• 4 to 20 mA current / HART 7 output.
• In the event of an alarm, current output can be adjusted
to 21 to 23 mA (NAMUR NE43).
• Programmable digital output. Can be configured as a
frequency, pulse or binary output. Figure 14: Plug-in cards
• Two slots for optional plug-in cards for retrofitting
additional current outputs, digital outputs and digital Plug-in card Quantity
inputs. *
• Parameterization by means of HART communication. 1 Current output, 4 to 20 mA passive (red) 2
• Damping: can be adjusted 0.2 to 100 s (1 τ). Order no.: 3KQZ400029U0100
• Low flow cut-off: 0 to 10 % for current and pulse output.
2 Passive digital output (green) 1
• Measuring medium parameters can be changed at any
Order no.: 3KQZ400030U0100
time (pressure and temperature influence, units, etc.).
3 Passive digital input (yellow) 1
• Simulation of current and binary output (manual process
Order no.: 3KQZ400032U0100
execution).
4 24 V DC voltage supply (blue) 1
Order no.: 3KQZ400031U0100
LCD indicator (option)
5 Modbus RTU® RS485 (white) 1
• Indicator of all measured values of the SensyMaster
Order no.: 3KQZ400028U0100
(e.g. mass flow, standard volume flow, temperature).
6 PROFIBUS DP® (white) 1
• Application-specific visualizations which the user can
Order no.: 3KQZ400027U0100
select. Four operator pages can be configured to display
* The ‘Number’ column indicates the maximum number of plug-in cards
multiple values in parallel.
of the same type that can be used.
• Plain text fault diagnostics
• Menu-guided parameterization with four buttons.
Note
• Easy Set-up function for fast commissioning.
For an overview of the possible plug-in card combinations,
• Operation through the front glass via capacitive buttons.
refer to Ordering Information on page 51.
16 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Transmitter
IP rating Signal cables
In accordance with EN60529: IP 65 / IP 67, NEMA 4X The signal cable used for the connection of the transmitter
and sensor must fulfill at least the following technical
specifications.
Vibration
In accordance with EN 60068-2 Cable specification
Temperature Data
Maximum signal cable length
0.25 mm2 (AWG 24) 25 m (82 ft)
Standard Optional* 0.34 mm2 (AWG 22) 40 m (131 ft)
Ambient temperature −20 to 70 °C −40 to 70 °C 0.5 mm2 (AWG 20) 65 m (213 ft)
(−4 to 158 °F) (−40 to 158 °F) 0.75 mm2 (AWG 19) 100 m (328 ft)
Storage temperature −40 to 70 °C —
(−40 to 158 °F) Recommended cables
* In preparation It is recommended to use an ABB signal cable with the order
number 3KQZ407123U0100 for standard applications.
Note The ABB signal cable fulfills the above-mentioned cable
When operating below -20 °C (-4 °F), the LCD display can no specification and can be utilized unrestrictedly up to an
longer be read and the electronics should be operated with as ambient temperature of Tamb. = 80 °C (176 °F).
few vibrations as possible.
Full functionality is assured at temperatures above -20 °C (-
4 °F).
Housing design
Electrical connections
Electrical connection (HART protocol)
A Transmitter B Sensor
Connections for the power supply Connections for inputs and outputs
DC voltage
V1 / V2 Plug-in card, slot OC1
Terminal Function/comments
V3 / V4 Plug-in card, slot OC2
1+ +
For details, see Optional plug-in cards on page 15.
2- −
PE / Protective earth (PE)
Potential equalization
—
… Transmitter
Electrical data for inputs and outputs
Note Current output 32 / Uco, 31 / 32 (basic device)
When using the device in potentially explosive atmospheres, Can be configured for outputting mass flow, volume flow,
note the additional temperature data in Use in potentially density and temperature via on-site software.
explosive atmospheres on page 40!
Power supply L / N, 1+ / 2−
AC voltage
Terminals L/N
Operating voltage 100 to 240 V AC, 50 / 60 Hz
Power consumption < 20 VA
A Current output 31 / Uco, active B Current output 31 / 32, passive
—
… Transmitter
Digital output 41 / 42, 51 / 52 (basic device) Modbus® / PROFIBUS DP® interface V1 / V2 (plug-in card)
Can be configured as pulse, frequency or binary output via A Modbus or PROFIBUS DP interface can be implemented by
on-site software. using the ‘Modbus RTU, RS485 (white)’ or ‘PROFIBUS DP,
RS485 (white)’ plug-in cards.
Note
• Terminals 42 / 52 have common grounding. Digital
outputs 41 / 42 and 51 / 52 are not electrically isolated
from each other. An electrically isolated digital output can
be made using a plug-in module.
• If using a mechanical counter, it is advisable to set a pulse
width of ≥ 30 ms and a maximum frequency of
fmax ≤ 3 kHz.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 21
Current output V1 / V2, V3 / V4 (plug-in module) Digital output V1 / V2, V3 / V4 (plug-in module)
Up to two additional plug-in modules can be implemented via An additional binary output can be implemented via the
the ‘Passive current output (red)’ option module. ‘Passive digital output (green)’ plug-in module.
Can be configured for outputting mass flow, volume flow, Can be configured as an output for flow direction signaling,
density and temperature via on-site software. alarm output etc. via on-site software.
—
… Transmitter
Digital input V1 / V2, V3 / V4 (plug-in module) 24 V DC loop power supply (plug-in module)
A digital input can be implemented via the ‘Passive digital Use of the ‘loop power supply (blue)’ plug-in card allows a
input (yellow)’ plug-in module. passive output on the transmitter to be used as an active
Can be configured as an input for external counter reset, output. See also Connection examples on page 23.
external output deactivation etc. via on-site software.
Connection examples
Input and output functions are configured via the device Digital output 41 / 42, 51 / 52 passive on distributed
software in accordance with the desired application. control system
Note
Each ‘loop power supply (blue)’ plug-in card must only power
one output.
It must not be connected to two outputs (for example digital
A Transmitter Ex. 2 Input 2
output 41 / 42 and 51 / 52)!
B Distributed control system / RX Resistor for current limitation
Memory programmable RI Distributed control system
controller internal resistance
Ex. 1 Input 1
—
… Transmitter
Current output V3 / V4 active Connection versions digital output 41 / 42, 51 / 52
When the ‘loop power supply 24 V DC, blue’ plug-in card is Depending on the wiring of digital outputs DO 41 / 42 and
used, the current output on the plug-in card can also be wired 51 / 52, they can be used parallel or only individually. The
as the active current output. electrical isolation between the digital outputs also depends
on the wiring.
Digital communication
HART® Communication Modbus® communication
Note Note
The HART® protocol is an unsecured protocol, as such the The Modbus® protocol is an unsecured protocol, as such the
intended application should be assessed to ensure that these intended application should be assessed to ensure that these
protocols are suitable before implementation. protocols are suitable before implementation.
In connection with the DTM (Device Type Manager) available Modbus is an open standard owned and administrated by an
to the device, communication (configuration, independent group of device manufacturers styled the
parameterization) can be carried out FDT 0.98 or 1.2 Modbus Organization (www.modbus.org).
(DSV401 R2). Using the Modbus protocol allows devices made by different
Other tool or system integrations (e.g. Emerson AMS / manufacturers to exchange information via the same
Siemens PCS7) on request. communication bus, without the need for any special
The necessary DTMs and other files can be downloaded from interface devices to be used.
www.abb.com/flow.
Modbus protocol
HART output Terminals V1 / V2
Terminals Active: Uco / 32 Configuration Via the Modbus interface or via the local
Passive: 31 / 32 operating interface in connection with Asset
Protocol HART 7.1 Vision Basic (DAT200) and a corresponding
Transmission FSK modulation on current output 4 to 20 mA in Device Type Manager (DTM)
accordance with the Bell 202 standard Transmission Modbus RTU - RS485 serial connection
Baud rate 1200 baud Baud rate 2400, 4800, 9600, 19200, 38400, 56000, 57600,
Signal amplitude Maximum 1.2 mAss 115200 baud
Factory setting: 9600 baud
Parity None, even, odd
Factory setting of the HART® process variables
Factory setting: odd
Process variable Process value
Stop bit One, two
Primary Value (PV) Mass flow
Factory setting: One
Secondary Value (SV) Temperature
IEEE format Little endian, big endian
Tertiary Value (TV) Mass (counter)
Factory setting: Little endian
Quaternary Value (QV) Standard flow rate
Typical response time < 100 ms
Response delay time 0 to 200 milliseconds
The process values of the HART® variables can be set in the
Factory setting: 10 milliseconds
device menu.
26 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Transmitter
PROFIBUS DP® communication
Note
The PROFIBUS DP® protocol is an unsecured protocol, as such
the intended application should be assessed to ensure that
these protocols are suitable before implementation.
PROFIBUS DP interface
Terminals V1 / V2
Configuration Via the PROFIBUS DP interface or via the local
operating interface in connection with Asset Vision
Basic (DAT200) and a corresponding Device Type
Manager (DTM)
Transmission In accordance with IEC 61158-2
1 Modbus master 3 Modbus slave 1
Baud rate 9.6 kbps, 19.2 kbps, 45.45 kbps, 93.75 kbps, 187.5
2 Terminating resistor 4 Modbus slave n to 32 kbps, 500 kbps, 1.5 Mbps
Figure 33: Communication with the Modbus protocol The baud rate is automatically detected and does
not need to be configured manually
Cable specification Device profile PA Profile 3.02
The maximum permissible length is dependent on the baud Bus address Address range 0 to 126
rate, the cable (diameter, capacity and surge impedance), the Factory setting: 126
number of loads in the device chain, and the network
configuration (2-core or 4-core). Only one of the three different GSD files provided by ABB is
• At a baud rate of 9600 and with a conductor cross-section needed for commissioning.
of at least 0.14 mm2 (AWG 26), the maximum length is
1000 m (3280 ft). Parameterization of the device can be performed via the
• When using a 4-core cable as a 2-wire wiring system, the display, or through a device driver in the form of an EDD
maximum length must be halved. (Electronic Device Description) or DTM (Device Type
• The spur lines must be short, a maximum of 20 m (66 ft). Manager).
• When using a distributor with ‘n’ connections, each
branch must have a maximum length of 40 m (131 ft) You can download EDD, DTM and GSD from
divided by ‘n.’ www.abb.com/flow.
The maximum cable length depends on the type of cable The files required for operation can also be downloaded from
used. The following standard values apply: www.profibus.com.
• Up to 6 m (20 ft):
cable with standard shielding or twisted-pair cable. ABB provides three different GSD files which can be
• Up to 300 m (984 ft): integrated in the system.
double twisted-pair cable with overall foil shielding and
integrated earth cable. ID number GSD file name
• Up to 1200 m (3937 ft): 0x9740 PA139740.gsd 1xAI, 1xTOT
double twisted-pair cable with individual foil shielding 0x3435 ABB_3435.gsd 6xAI, 2xTOT, 1xDI, 2xDO
and integrated earth cables. Example: Belden 9729 or 0x9700 PA139700.gsd 1xAI
equivalent cable.
Users decide at system integration whether to install the full
A category 5 cable can be used for Modbus RS485 up to a
range of functions or only part. Switching is made using the
maximum length of 600 m (1968 ft). For the symmetrical
‘Ident Nr. Selector’ parameter.
pairs in RS485 systems, a surge impedance of more than
100 Ω is preferred, especially at a baud rate of 19200 and
Refer to .
above.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 27
—
Pipe components
The pipe components are available with the following process
connections:
—
… Pipe components
Material loads for process connections Integrated hot tap fitting
The integrated hot tap fitting is used instead of the
previously described pipe components and weld-on adapters
if taking out the sensor should be practically possible without
gas escaping during running operation.
Figure 37: Maximum pressure/temperature values for integrated hot tap fitting
Note
Connection flanges PN 16 with four screw holes must be used
in the integrated hot tap fitting in wafer type design DN 65.
Wafer type designs 2 to 8 in only for connection flanges ASME
B16.5, Cl 150.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 29
—
Dimensions
Sensor
Integral mount design
All dimensions and weights are specified in mm (inch) or kg (lb).
Sensor connection For nominal piping diameter L h (installation length) Approximate weight
mm (in) mm (in) Kg (lb)
Flange DN 25 DN 25 to 350 (1 to 14 in) 271 (10.64) 263 (10.35) 6.5 (14.3)
> DN 350 to 700 (> 14 to 28 in) 433 (17.05) 425 (16.73) 7 (15.4)
> DN 700 (> 28 in) 783 (30.83) 775 (30.51) 7.5 (16.5)
Clamp ring fitting DN 25 to 350 (1 to 14 in) 326 (12.83) 318 (12.52) 5.5 (12.1)
> DN 350 to 700 (> 14 to 28 in) 488 (19.21) 480 (18.90) 6 (13.2)
> DN 700 (> 28 in) 838 (32.99) 830 (32.68) 7 (15.4)
Threaded connection DIN 11851 DN 25 to 80 (1 to 3 in) 136 (5.53) 120 (4.72) 4.7 (10.4)
Note
The specified nominal piping diameters apply for the use of the sensor with pipe components without ball valves or hot tap
fittings.
30 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Dimensions
Remote mount design
All dimensions and weights are specified in mm (in) or kg (lb).
Sensor connection For nominal piping diameter L h (installation length) Approximate weight
mm (in) mm (in) Kg (lb)
Flange DN 25 DN 25 to 350 (1 to 14 in) 271 (10.64) 263 (10.35) 5 (11)
> DN 350 to 700 (> 14 to 28 in) 433 (17.05) 425 (16.73) 5.5 (12)
> DN 700 (> 28 in) 783 (30.83) 775 (30.51) 6 (13)
Clamp ring fitting DN 25 to 350 (1 to 14 in) 326 (12.83) 318 (12.52) 4 (8.8)
> DN 350 to 700 (> 14 to 28 in) 488 (19.21) 480 (18.90) 4.5 (9.9)
> DN 700 (> 28 in) 838 (32.99) 830 (32.68) 5.5 (12)
Threaded connection DIN 11851 DN 25 to 80 (1 to 3 in) 136 (5.53) 120 (4.72) 3.2 (7)
Note
The specified nominal piping diameters apply for the use of the sensor with pipe components without ball valves or hot tap
fittings.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 31
Transmitter
—
… Dimensions
Pipe components
All dimensions and weights are specified in mm (in) or kg (lb).
FMT091 – Wafer type design in accordance with EN 1092-1, PN 40 – Sensor connection: flange DN 25
Nominal h D1 D2 D3 Weight
diameter
DN 40 263 (10.35) 43.1 (1.70) 88 (3.46) 94 (3.70) 4.5 (10)
DN 50 54.5 (2.15) 102 (4.02) 109 (4.29) 5.0 (11)
DN 65 70.3 (2.77) 122 (4.80) 129 (5.08) —
DN 80 82.5 (3.25) 138 (5.43) 144 (5.67) 7.0 (15.5)
DN 100 107.1 (4.22) 162 (6.38) 170 (6.69) 8.5 (18.7)
DN 125 131.7 (5.19) 188 (7.40) 196 (7.72) —
DN 150 159.3 (6.27) 218 (8.58) 226 (8.90) 11.5 (25.5)
DN 200 206.5 (8.13) 285 (11.22) 293 (11.54) —
FMT091 – Wafer type design in accordance with ASME B 16.5, CL 150 – Sensor connection: flange DN 25
Nominal h D1 D2 D3 Weight
diameter
1½ in 263 (10.35) 40.9 (1.61) 73 (2.87) 85 (3.35) —
2 in 52.6 (2.07) 92 (3.62) 103 (4.06) —
3 in 78.0 (3.07) 127 (5.00) 135 (5.31) —
4 in 102.4 (4.03) 157 (6.18) 173 (6.81) —
6 in 154.2 (6.07) 216 (8.50) 221 (8.70) —
8 in 202.7 (7.98) 270 (10.63) 278 (10.94) —
FMT091 – Wafer type design in accordance with ASME B 16.5, CL 300 – Sensor connection: flange DN 25
Nominal h D1 D2 D3 Weight
diameter
1½ in 263 (10.35) 40.9 (1.61) 73 (2.87) 94 (3.70) —
2 in 52.6 (2.07) 92 (3.62) 110 (4.33) —
3 in 78.0 (3.07) 127 (5.00) 148 (5.83) —
4 in 102.4 (4.03) 157 (6.18) 180 (7.09) —
6 in 154.2 (6.07) 216 (8.50) 249 (9.80) —
8 in 202.7 (7.98) 270 (10.63) 307 (12.09) —
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 33
Figure 43:
FMT092 – Partial measuring section with flange in accordance with EN 1092-1, Form B1, PN 40 – Sensor connection: flange DN 25
Nominal h D1 D4 L1 L2 Weight
diameter
DN 25 263 (10.35) 28.5 (1.12) 115 (4.53) 486 (19.13) 600 (23.62) 5.5 (12.0)
DN 40 43.1 (1.70) 150 (5.91) 731 (28.78) 860 (33.86) 8.0 (17.5)
DN 50 54.5 (2.15) 165 (6.50) 837 (32.95) 1000 (39.37) 11 (24.3)
DN 65 70.3 (2.77) 185 (7.28) 1190 (46.85) 1400 (55.12) –
DN 80 82.5 (3.25) 200 (7.87) 1450 (57.09) 1700 (66.93) –
DN 100 107.1 (4.22) 235 (9.25) 1870 (73.62) 2200 (86.61) –
FMT092 – Partial measuring section with flange in accordance with EN 1092-1, Form B1, PN 10 – Sensor connection: threaded connection DIN 11851
Nominal ØD inside D5 L3 L4 Weight
diameter
DN 25 27.3 (1.07) 115 (4.53) 410 (16.14) 550 (21.65) –
DN 40 41.9 (1.65) 150 (5.91) 615 (24.21) 820 (32.28) –
DN 50 53.9 (2.12) 165 (6.50) 810 (31.89) 1080 (42.52) –
DN 80 79.9 (3.15) 200 (7.87) 1200 (47.24) 1600 (62.99) –
FMT092 – Partial measuring section with male thread, PN 10 – sensor connection: threaded connection DIN 11851
Nominal ØD inside R male thread L3 L4 Weight
diameter
DN 25 27.3 (1.07) R1” – 33.7 x 1.2 410 (16.14) 550 (21.65) –
DN 40 41.9 (1.65) R1 1/2” – 48.3 x 3.2 615 (24.21) 820 (32.28) –
DN 50 53.9 (2.12) R2” – 60.3 x 3.2 810 (31.89) 1080 (42.52) –
DN 80 79.9 (3.15) R3” – 88.9 x 4.5 1200 (47.24) 1600 (62.99) –
34 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Dimensions
All dimensions and weights are specified in mm (in) or kg (lb).
FMT092 – Partial measuring section with flange in accordance with ASME B 16.5, CL 150 – Sensor connection: flange DN 25
Nominal h D1 D4 L1 L2 Weight
diameter
1 in 263 (10.35) 26.6 (1.05) 108 (4.25) 454 (17.87) 560 (22.05) –
1½ in 40.9 (1.61) 127 (5.00) 741 (29.17) 864 (34.02) –
2 in 52.6 (2.07) 154 (6.06) 846 (33.31) 1003 (39.49) –
3 in 78.0 (3.07) – – – –
4 in 102.4 (4.03) – – – –
FMT092 – Partial measuring section with flange in accordance with ASME B 16.5, CL 300 – Sensor connection: flange DN 25
Nominal h d1 D4 L4 L3 Weight
diameter
1 in 263 (10.35) 26.6 (1.05) 123.9 (4.88) 454 (17.87) 560 (22.05) –
1½ in 40.9 (1.61) 155.4 (6.12) 741 (29.17) 864 (34.02) –
2 in 52.6 (2.07) 165.1 (6.50) 846 (33.31) 1003 (39.49) –
3 in 78.0 (3.07) – – – –
4 in 102.4 (4.03) – – – –
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 35
Welding adapter
Dimensions in mm (in)
Note
When mounting the weld-on adapter, observe the following points:
• The welding adapters must be shortened to the dimension L before installation, in accordance with: L = h —(1/2 x Ø D).
• The distance h from the upper edge of the flange to the pipe central axis must be within a tolerance of ± 2 mm (± 0.08“).
• Maintain the right angle to the pipe axis (max. tolerance ± 2°).
• The adapter centering pin must be aligned with the pipe axis in the flow direction (outflow side, behind the measuring
point).
36 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Dimensions
Weld-on adapter with ball valve
Dimensions in mm (in)
Note
When mounting the weld-on adapter, observe the following points:
• The welding adapters must be shortened to the dimension L before installation, in accordance with: L = h —(1/2 x Ø D).
• The distance h from the upper edge of the flange to the pipe central axis must be within a tolerance of ± 2 mm (± 0.08“).
• Maintain the right angle to the pipe axis (max. tolerance ± 2°).
• The adapter centering pin must be aligned with the pipe axis in the flow direction (outflow side, behind the measuring
point).
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 37
Note
When mounting the weld-on adapter, observe the following points:
• Always mount the weld-on adapter together with the union nut on the piping. Mounting it at a later time is not possible.
• The welding adapters must be shortened to the dimension L before installation, in accordance with: L = h —(1/2 x Ø D).
• The distance h from the upper edge of the adapter to the pipe central axis must be within a tolerance of ± 2 mm
(± 0.08 inch).
• Maintain the right angle to the pipe axis (max. tolerance ± 2°).
• Observe the thickness of pipeline wall and the degree of shrinkage when welding on.
• The adapter centering pin must be aligned with the pipe axis in the flow direction (outflow side, behind the measuring
point).
• Once welding is complete, there must be a passage of at least 28 mm (1.10 inch) free for the purpose of mounting the
sensor (drill to create if necessary)
38 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Dimensions
Welding adapter with compression fitting
All dimensions in mm (in)
h – sensor length h3 – installation length L = h3 −(½ × ØD) ∅ D – outer pipe diameter*(min. / max.)
263 (10.35) 244 (9.61) to be calculated > 100 to 350 (> 3.94 to 13.78)
425 (16.73) 406 (15.98) > 350 to 700 (> 13.78 to 27.56)
775 (30.51) 756 (29.76) > 700 to 1400 (> 27.56 to 55.12)
Note
When installing the adapter tubes for compression fittings, observe the following points:
• The welding adapter tubes should be shortened to dimension L before installation, in accordance with: L = h3 −(½ × ØD)
• It is imperative that you maintain a right angle to the longitudinal and transversal axes of the meter tube (maximum
tolerance ±2°).
• Make sure that after welding the free opening of the adapter tube if at least Ø28 mm (1.10 in).
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 39
—
Use in potentially explosive atmospheres
Note
Further information on the Ex-Approval of devices can be found in the type examination certificates or the relevant certificates
at www.abb.com/flow .
Device overview
ATEX / IECEx
• Zone 1, 21
• Zone 0
1 Single-compartment housing
2 Dual-compartment housing
3 Zone 0 within the meter tube
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 41
cFMus
Standard / No explosion protection Class I Div. 2 / Zone 2 Class I Div. 1 / Zone 1 (Zone 0)
Model number FMT4x0 Y0 FMT4x0 F2 FMT4x0 F1
Integral mount design
• Standard
• Div. 2 / Zone 2
• Div. 1 / Zone 1
• Div. 1 / Zone 1
1 Single-compartment housing
2 Dual-compartment housing
42 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Use in potentially explosive atmospheres
Ex marking
ATEX / IECEx
Note
• A specific marking applies, depending on the design.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.
cFMus
Note
• A specific marking applies, depending on the design.
• ABB reserves the right to modify the Ex-marking. Refer to the name plate for the exact marking.
—
… Use in potentially explosive atmospheres
Temperature data
Temperature resistance for the connecting cable
The temperature at the cable entries of the device is dependent on the measuring medium temperature Tmedium and the
ambient temperature Tamb..
For the electrical connection of the device, use only cables with sufficient temperature resistance in accordance with the
following table.
From an ambient temperature of Tamb. ≥ 60 °C (≥ 140 °F), the wires in the connection boxes with the enclosed silicone hoses
need to be additionally insulated.
Note
The signal cable supplied by ABB can be used without restrictions up to an ambient temperature of ≤ 80 °C (≤ 176 °F).
Measuring medium temperature (Ex data) for model FMT4x0-A1… in Zone 1, Zone 21
The table shows the maximum permissible measuring medium temperature as a function of ambient temperature and
temperature class. The permissible measuring medium temperature specified in Environmental and process conditions for
model FMT4xx… on page 44 must not be up-scaled!
Temperature class
Ambient temperature Tamb. T1 T2 T3 T4 T5 T6
−40 °C to 50 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 122 °F)
−40 °C to 60 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 140 °F)
−40 °C to 70 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 158 °F)
Measuring medium temperature (Ex data) for model FMT4x0-A2… in Zone 2, Zone 22
The table shows the maximum permissible measuring medium temperature as a function of ambient temperature and
temperature class. The permissible measuring medium temperature specified in Environmental and process conditions for
model FMT4xx… on page 44 must not be up-scaled!
Temperature class
Ambient temperature Tamb. T1 T2 T3 T4 T5 T6
−40 °C to 40 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) 95 °C (203 °F) 80 °C (176 °F)
(−40 °F to 104 °F)
−40 °C to 50 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) 95 °C (203 °F) —
(−40 °F to 122 °F)
−40 °C to 60 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) — —
(−40 °F to 140 °F)
−40 °C to 70 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) — —
(−40 °F to 158 °F)
46 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Use in potentially explosive atmospheres
Measuring medium temperature (Ex data) for model FMT4x0-F1… in Class I Division 1 and Class II Division 1
The table shows the maximum permissible measuring medium temperature as a function of ambient temperature and
temperature class. The permissible measuring medium temperature specified in Environmental and process conditions for
model FMT4xx… on page 44 must not be up-scaled!
Temperature class
Ambient temperature Tamb. T1 T2 T3 T4 T5 T6
−40 °C to 50 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 122 °F)
−40 °C to 60 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 140 °F)
−40 °C to 70 °C
280 °C (536 °F) 185 °C (365 °F) 90 °C (194 °F) 90 °C (194 °F) — —
(−40 °F to 158 °F)
Measuring medium temperature (Ex data) for model FMT4x0-F1… in Class I Division 2 and Class II Division 2
The table shows the maximum permissible measuring medium temperature as a function of ambient temperature and
temperature class. The permissible measuring medium temperature specified in Environmental and process conditions for
model FMT4xx… on page 44 must not be up-scaled!
Temperature class
Ambient temperature Tamb. T1 T2 T3 T4 T5 T6
−40 °C to 40 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) 95 °C (203 °F) 80 °C (176 °F)
(−40 °F to 104 °F)
−40 °C to 50 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) 95 °C (203 °F) —
(−40 °F to 122 °F)
−40 °C to 60 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) — —
(−40 °F to 140 °F)
−40 °C to 70 °C
300 °C (572 °F) 290 °C (554 °F) 195 °C (383 °F) 130 °C (266 °F) — —
(−40 °F to 158 °F)
Notice on dust-ignition protection for USA and Canada in accordance with NEC
The surface temperature of the device must not under any circumstances up-scale 85 °C (185 °F) if there is there carbonaceous
dust or dust which can carbonate.
Electrical data
Overview
A Power supply B Inputs / outputs, communication C Signal cable (remote mount design only)
• Type of protection ATEX / IECEx: • Type of protection ATEX / IECEx: • Type of protection ATEX / IECEx:
Non-sparking ‘Ex ec’ non-sparking ‘Ex ec’ Non-sparking ‘Ex ec’
• Type of protection USA / Canada: • Type of protection USA / Canada: • Type of protection USA / Canada:
‘non IS’ Non-Incendive ‘NI’ Non-Incendive ‘NI’
• Maximum 250 Vrms • Terminals: 31, 32, Uco, V1, V2, V3, V4, 41, 42, 51, • Terminals: A, B, UFE, GRN
• Terminals: 1+, 2-, L, N, 52
A Power supply B Inputs / outputs, communication C Signal cable (remote mount design only)
• Type of protection ATEX / IECEx: • Type of protection ATEX / IECEx: • Type of protection ATEX / IECEx:
Non-sparking ‘Ex eb’ non-sparking ‘Ex ec’ Non-sparking ‘Ex eb’
• Type of protection USA / Canada: • Type of protection USA / Canada: • Type of protection USA / Canada:
‘non IS’ Non-Incendive ‘NI’ explosionproof ‘XP’
• Maximum 250 Vrms • When installing in ‘Ex ia’, suited intrinsically safe • Terminals: A, B, UFE, GRN
• Terminals: 1+, 2-, L, N, isolation amplifiers must be used for the
connection.
• Terminals: 31, 32, Uco, V1, V2, V3, V4, 41, 42, 51,
52
Note
When installing in ‘Ex ia’ or ‘IS’ type of protection, the type of protection is determined by the type of electrical connection. The
information in Changing the type of protection in the operating instruction must be observed when changing the type of
protection!
48 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Use in potentially explosive atmospheres
Zone 0, 1, 21 and Division 1 – Model: FMT4xx-A1, FMT4xx-F1
Outputs on basic device Operating values (general) Type of protection ‘ec’ / ‘NI’
UN IN UN IN
Current / HART output 31 / UCO, active 30 V 30 mA 30 V 30 mA
Terminals 31 / UCO
Current / HART output 31 / 32, passive 30 V 30 mA 30 V 30 mA
Terminals 31 / 32
Digital output 41 / 42, active* 30 V 30 mA 30 V 30 mA
Terminals 41 / 42 and V1 / V2*
Digital output 41 / 42, active** 30 V 30 mA 30 V 30 mA
Terminals 41 / 42 and UCO / 32**
Digital output 41 / 42, passive 30 V 30 mA 30 V 30 mA
Terminals 41 / 42
Digital output 51 / 52, active* 30 V 30 mA 30 V 30 mA
Terminals 51 / 52 and V1 / V2*
Digital output 51 / 52, passive 30 V 30 mA 30 V 30 mA
Terminals 51 / 52
All outputs are electrically isolated from each other and from the power supply.
Digital outputs 41 / 42 and 51 / 52 are not electrically isolated from each other. Terminals 42 / 52 have the same potential.
* Only in conjunction with additional ‘24 V DC loop power supply (blue)’ plug-in card in slot OC1.
** Only in conjunction with current output UCO / 32 in ‘Powermode’, see Current output Uco / 32 as loop power supply for digital output 41 / 42 or
51 / 52 on page 19.
Inputs and outputs with optional plug-in cards Operating values (general) Type of protection ‘ec’ / ‘NI’
UN IN UN IN
Current output V3 / V4, active* 30 V 30 mA 30 V 30 mA
Terminals V3 / V4 and V1 / V2*
Current output V1 / V2, passive** 30 V 30 mA 30 V 30 mA
Current output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
Digital output V3 / V4, active* 30 V 30 mA 30 V 30 mA
Terminals V3 / V4 and V1 / V2*
Digital output V1 / V2, passive** 30 V 30 mA 30 V 30 mA
Digital output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
Digital input V3 / V4, active* 30 V 3,45 mA 30 V 3,45 mA
Terminals V3 / V4 and V1 / V2
Digital input V1 / V2, passive* 30 V 3,45 mA 30 V 3,45 mA
Digital input V3 / V4, passive*
Terminals V1 / V2** or V3 / V4**
Modbus® / PROFIBUS DP® 30 V 30 mA 30 V 30 mA
Terminals V1 / V2
* Only in conjunction with additional ‘24 V DC loop power supply (blue)’ plug-in card in slot OC1.
** The terminal assignment depends on the model number or the slot assignments. For connection examples, see Connection examples on page 23.
50 SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C
—
… Use in potentially explosive atmospheres
Special connection conditions
Note
The AS plug-in card (24 V DC loop power supply) may only be Devices connected to the relevant equipment must not be
used to power the internal inputs and outputs on the device. operated at over 250 Vrms AC or 250 V DC to ground.
It must not be used to power external circuits!
Installation in accordance with ATEX or IECEx must comply
Note with the applicable national and international standards and
If the protective earth (PE) is connected in the flowmeter's directives.
terminal box, you must ensure that no dangerous potential
difference can arise between the protective earth (PE) and Installation in the USA or Canada must comply with ANSI /
the potential equalization (PA) in areas with explosion risk. ISA RP 12.6, ‘Installation of intrinsically safe systems for
hazardous (classified) locations’, the ‘National Electrical Code
Note (ANSI / NFPA 70), sections 504, 505’ and the ‘Canadian
• For devices with a power supply of 11 to 30 V DC, on- electrical code (C22.1-02)’.
site external overvoltage protection must be provided.
• You must make sure that the overvoltage is limited to Apparatus connected to the flowmeter must have
140 % (= 42 V DC) of the maximum operating voltage. appropriate explosion protection approval in accordance
with the Entity concept.
The output circuits are designed so that they can be The apparatus must have intrinsically safe circuits.
connected to both intrinsically-safe and non-intrinsically-safe The apparatus must be installed and connected in
circuits. accordance with the relevant manufacturer documentation.
• Combining intrinsically safe and non-intrinsically safe The electrical specifications in Electrical data on page 47
circuits is not permitted. must be observed.
• On intrinsically safe circuits, potential equalization should
be established along the entire length of the cable used
for the signal outputs.
• The rated voltage of the non-intrinsically safe circuits is
UM = 30 V.
• Intrinsic safety is preserved If the rated voltage UM = 30 V
is not up-scaled when connections are established to non-
intrinsically safe external circuits.
• The information in Changing the type of protection in the
operating instruction must be observed when changing
the type of protection.
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 51
—
Ordering Information
Note
For additional information on dependencies and restrictions, and for help on product selection, please refer to the Online
Product Selection Assistant (PSA) at www.abb.us/flow-selector.
The following table provides an overview of the possible combinations of plug-in card combinations that can be selected when
ordering the device.
—
… Ordering Information
SensyMaster FMT430
Thermal Mass Flowmeter, for standard applications, compact and clever
Base model FMT430 XX XX X X XX XX XX XX X
SensyMaster FMT430 Thermal Mass Flowmeter
Explosion Protection Certification
Without Y0
ATEX / IECEx (Zone 2 / 21) A2
ATEX / IECEx (Zone 1 / 21) A1
ATEX / IECEx (Zone (0) 1 / 21) A3
cFMus (Class 1 Div. 2 / Zone 2) F2
cFMus (Class 1 Div. 1 / Zone 1) F1
Measuring Medium
Air or other clean gas (One gas component only) C1
Gas mixtures with max. 23.5 Vol% O2 (eg. Natural gas or Biogas) C2
Oxygen / gas mixtures > 23.5 Vol% O2, oil and grease-free,
with O2 certificate (max. 150 °C / 302 °F) P1
Ammonia H3
Sensor Element Type / Temperature Range of Measuring Medium
Ceramic sensor elements / Standard temperature range −20 to 150 °C (−4 to 302 °F) A
Ceramic sensor elements / High temperature range −20 to 300 °C (−4 to 572 °F) B
Mounting Length / Flowmeter Sensor Material
120 mm (4.7 in) / AISI 316Ti SST (1.4571) (DN 25 to DN 125 [1 to 5 in]) 1*
263 mm (10.4 in) / AISI 316Ti SST (1.4571) (DN 25 to DN 350 [1 to 14 in]) 2*
425 mm (17 in) / AISI 316Ti SST (1.4571) (> DN 350 to DN 700 [> 14 to 28 in]) 3*
775 mm (31 in) / AISI 316Ti SST (1.4571) (> DN 700 [> 28in.]) 4*
Sensor Connection
Flange DN 25, nominal pressure 4 MPa (40 bar, 580 psi) D3
Compression fitting, stainless steel, nominal pressure 2 MPa (20 bar, 290 psi)
(−20... 140 °C (−4 to 284 °F)) (−20... 140 °C (−4 to 284 °F)) (> DN80 (> 3 inch)) G2
Thread DIN 11851, nominal pressure 1.6 MPa (16 bar, 232 psi)
(−20 to 140 °C (−4 to 284 °F)) (−20 to 140 °C (−4 to 284 °F)) F1
Connection Design / Transmitter Housing Type / Transmitter Housing Material / Cable Glands
Integral / Single compartment / Aluminium / 3 × M20 × 1.5 S1
Integral / Single compartment / Aluminium / 3 × NPT ½ in S2
Integral / Dual compartment / Aluminium / 3 × M20 × 1.5 D1
Integral / Dual compartment / Aluminium / 3 × NPT ½ in D2
Integral / Dual compartment / Aluminium / 3 × NPT ½ in (Exd, XP) D5
Integral / Dual compartment / Aluminium / 3 × M20 × 1.5 (Exd, XP) D6
Remote (Selection for remote transmitter) Y0
Without Transmitter Y5
Connection Design / Sensor Housing Type / Sensor Housing Material / Cable Glands
Remote / Single compartment / Aluminium / 1 × M20 × 1.5 A1
Remote / Single compartment / Aluminium / 1 × NPT ½ in A2
Remote / Single compartment / Stainless Steel / 1 × M20 × 1.5 U1
Remote / Single compartment / Stainless Steel / 1 × NPT ½ in U2
Without Y0
* Nominal size ranges when using flanged pipe components or weld-on adapters without ball valve
—
… Ordering Information
Additional ordering information XX XX XXX XX XX XXX XX
SensyMaster FMT430 Thermal Mass Flowmeter
Integrated Digital Display (LCD)
No Display, with Blind Cover L0
With Push Buttons and Display (TTG) and Glass Cover L2
Documentation Language
German M1
English M5
Western Europe Scandinavia MW
Eastern Europe ME
Configuration Type
Parameters set to factory default NC1
Parameters set customer specific (e.g. function of the outputs, etc.) NCC
Calibration Type
5-point calibration, standard measuring range, incl. factory certificate R3
7-point calibration, extended measuring range, incl. factory certificate (non Ex version only) R4
Device Identification Plate
Stainless steel plate with TAG no. T1
Adhesive label with TAG no. TC
Ambient Temperature Range
Extended −40 to 70 °C (−40 to 158 °F) (in preparation) TA9
Extended Diagnostic Options
VeriMass SensorCheck function V2
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 55
SensyMaster FMT450
Thermal Mass Flowmeter for advanced applications
—
… Ordering Information
—
… Ordering Information
SensyMaster FMT432 / FMT452
Remote Transmitter for Thermal Mass Flowmeter FMT430
Base model
SensyMaster FMT432 Thermal Mass Flowmeter Transmitter FMT432 XX XX XX XX
SensyMaster FMT452 Thermal Mass Flowmeter Transmitter FMT452 XX XX XX XX
Without
Without Y0
ATEX / IECEx (Zone 2 / 21) A2
ATEX / IECEx (Zone 1 / 21) A1
ATEX / IECEx (Zone (0) 1 / 21) A3
cFMus (Class 1 Div. 2 / Zone 2) F2
cFMus (Class 1 Div. 1 / Zone 1) F1
Connection Design / Transmitter Housing Type / Transmitter Housing Material / Cable Glands
Remote / Single compartment, wall mounted / Aluminium / 4 × M20 × 1.5 W1
Remote / Single compartment, wall mounted / Aluminium / 4 × NPT ½ in W2
Remote / Dual compartment, wall mounted / Aluminium / 4 × M20 × 1.5 R1
Remote / Dual compartment, wall mounted / Aluminium / 4 × NPT ½ in R2
Remote / Dual compartment / Aluminium / 3 × M20 × 1.5 (Exd, XP) R5
Remote / Dual compartment / Aluminium / 3 × NPT ½ in (Exd, XP) R6
Outputs
Current output 1 (active or passive), digitial output 1 & 2 (passive), HART G0
Current output 1 (active or passive), digitial output 1 & 2 (passive), current output 2 (passive), HART G2
Current output 1 (active or passive), digitial output 1 & 2 (passive), digital Input (passive), HART G8
Current output 1 (active or passive), digitial output 1 & 2 (passive), HART, MODBUS M1
Current output 1 (active or passive), digitial output 1 & 2 (passive), HART, PROFIBUS DP D1
Power Supply
100 to 230 V AC, 50 / 60 Hz A
24 V AC / DC B
—
… Ordering Information
SensyMaster FMT091 Pipe component / Wafer Design (Type 1)
—
… Ordering Information
SensyMaster FMT092 Pipe component, partial measuring section (type 2)
Max. 1,6 MPa (16 bar, 232 psi)
Base model FMT092 X XXX XX XX XX XX XX
SensyMaster FMT092 Pipe component, partial measuring section (type 2)
Design
Standard S
Integrated flow straighteners F
Nominal Diameter
DN 25 (1 in) 025
DN 40 (1½ in) 040
DN 50 (2 in) 050
DN 65 (2½ in) 065
DN 80 (3 in) 080
DN 100 (4 in) 100
DN 125 (5 in) 125
DN 150 (6 in) 150
DN 200 (8 in) 200
Process Connection
Flanges DIN PN 40 D4
Flanges ANSI / ASME B16.5 Class 150, Schedule 40 S A1
Flanges ANSI / ASME B16.5 Class 300, Schedule 40 S A3
R-External thread 1.6 MPa (16 bar, 232 psi) N6
Sensor Connection
Flange DN 25, nominal pressure 4 MPa (40 bar, 580 psi) with centering pin D3
Thread DIN 11851, nominal pressure 1.6 MPa (16 bar, 232 psi) with centering pin F1
Measuring Medium
Air or other clean gas (only one type of gas) C1
Gas mixtures with max. 23.5 Vol% O2 (e.g. natural gas or biogas) C2
Oxygen / gas mixtures > 23.5 Vol% O2, oil and grease-free, with O2 certificate P1
(max. 150 °C / 302 °F)
Hydrogen (max. 8 bar / 0.8 MPa / 116 psi, including process gas calibration) P2*
Helium (max. 8 bar / 0.8 MPa / 116 psi, including process gas calibration) P3*
Ammonia H3
Pipe Material
Stainless steel AISI 316Ti (1.4571) S2
Steel S 235 (1.0037) galvanized S3
Mounting Length of the Sensor
120 mm (4.7 in) L1
263 mm (10.4 in) L2
* For measuring medium H2 or He in nominal size DN 25 to DN 50 or 1 to 2 in please use pipe component 2 with flow straightener.
—
… Ordering Information
SensyMaster FMT094 Pipe component, weld-on adapter
—
Questionnaire
Customer: Date:
Telephone: Email:
Model:
FMT230 FMT430 Not determined
FMT250 FMT450
Application data:
Operating pressure min. / norm. / max. [bar abs, psi, other] Temperature min. / norm. / max. [°C, °F]
________________________________________________________________ ________________________________________________________________
Flow rate min. / norm. / max. [kg/h, lbs/h, Nm3/h, other] Normal conditions (in volume flow)
0°C, 1013mbar other
________________________________________________________________ 20°C, 1013mbar _______________________________
Gas data:
Gas type (pure gas):
___________________________________________________________________________________________________________________________________
Gas mixture (name, vol. %)1) Component 1 Component 2 Component 3 Component 4 Component 5
__________________ __________________ __________________ __________________ __________________
Transmitter design
Design: Signal cable length (remote mount design) Communication:
Integral mount design Single-compartment housing 5m 25 m current output / HART
Remote mount design Dual-compartment housing 15 m Modbus RTU
Note
The order can only be confirmed and a delivery date specified once full technical clearance has been obtained!
SENSYMASTER FMT430, FMT450 THERMAL MASS FLOWMETER | DS/FMT430/450-EN REV. C 67
—
Trademarks
HART is a registered trademark of FieldComm Group, Austin, Texas, USA
Modbus is a registered trademark of Schneider Automation Inc.
Kalrez and Kalrez Spectrum are registered trademarks of DuPont
Performance Elastomers.
Viton is a DuPont de Nemours trademark
Changing the type of protection in the operating instruction
Sales Service
—
ABB Measurement & Analytics
—
We reserve the right to make technical changes or modify the contents of this document without
DS/FMT430/450-EN Rev. C
prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained therein.
Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is
forbidden without prior written consent of ABB.