SINESON RF

(SNE-7100)
- User manual -

Thank you very much for buying our products. This manual
was prepared to help you use this machine safely. It is simple
to handle this machine and you can have enough effect by the
directions. But if you use this machine wrongly, you may
cause risks. So please make sure that you read all items and
acquire fully before using this machine. The guarantee is
attached to this manual. Please be careful not to lose it as it is
necessary when dealing with failure or requesting A/S.

1
1. To use safely ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

2. Warnings ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

3. Precautions ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

4. Precautions for use ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙
C
o 5. Contents ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

n 6. Name and specifications of each part ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

t 7. How to use ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙


e
8. Handling and storing after use ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙
n
9. How to replace fuse ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙
t
s 10. FAQ (Q & A) ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

◉ 11. Statement subject∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

12. About A/S ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

◯ Regulations on free repair ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

◯ Letter of guarantee ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

◯ Reference ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙

2
1. To use safely
Während der Behandlung niemals Elektrode und Touchscreen gleichzeitig berühren
(Gefahr von Überspannung). Das gilt bei der Selbstbehandlung wie auch bei der
Patientenbehandlung

RET-Elektroden (Resistive RF in der Tiefe) ist immer Chromstahl und mit 1


angeschnittener Rundaufnahme (geht nicht auf CET-Elektrode)

CET-Elektrode (kapazitive RF an der Oberfläche) ist immer in schwarzer


Beschichtung und mit 2 angeschnittener Rundaufnahme (geht nicht auf RET-
Elektrode)

Die beiden RET-Elektroden können nicht miteinander parallel benutzt werden


(Systemüberlastung). Die beiden CET-Elektroden können parallel miteinander
benutzt werden.

● Please read the precautions for safety carefully before use and use it rightfully

● The marks or precautions stated here are for helping you use the product safely and for
preventing users from being damaged or harmed.

● Precautions divided expected results into [warnings] and [cautions] when dealing with
the product wrongly according to the severity of harms or damages.

Understand the followings and comply with the statement subjects.

You should comply with the statement subjects as they are very important for safety.

Types and details of signs

If you ignore this sign and treat the product wrongly, it may
cause serious injury or even death of a user before use.
Warning Serious injury means loss of eyesight, wounds, burns, low-temperature
damages, electric shock, fracture and addiction which may lead to
aftereffects or cause the need of hospitalization and long-term outpatient
treatment.

If you ignore this sign and use the product wrongly, it may cause
Caution damages or injuries of a user or property damages.
Injuries mean wounds, burns or electric shock which needs not
hospitalization or long-term outpatient treatment.

Use example <Example of marked details>


Be careful of high temperature.

This sign means warning or caution.

3
Use example <Example of marked details>
Do not disassemble, modify or repair the
product.

This sign means prohibition of a behavior.

Use example <Example of marked details>


Unplug the power of the product.

This sign means forcible instruction for


behaviors.

2. Warnings

Do not repair, disassemble or modify the Do not install the main body where strong
main body. It may cause injuries due to frequency or magnet may occur. It may cause
ignition or malfunction. malfunction of the body.
Contact the assembler for repair.
It may cause fire or accidents.

Don’t disassemble. Prohibition

Do not use the product to infants. Do not use Do not use the product where the rated
the product to children or the disabled. Keep voltage is not 230V (50Hz). The use may cause
the main body, probes or accessories out of heat, which causes fire, electric shock or
the reach of children. malfunction.

It may cause electric shock or injury. It may cause electric shock or injury.

Prohibition Matters to be observed

4
Do not use too many wires in one socket. Do not plug the power cord or touch the main
Plug the power code tightly into the socket, body with wet hands. It may cause electric
and do not connect it to the socket which shock.
moves when tightly inserted. It may cause
fire due to heat.
It may cause heating, ignition or rupture.

It may cause fire or electric shock.


Be aware of fire

Be aware of electric shock

Do not use broken probes. It may cause skin Do not pull the probe cord. Also, you should
damages. grab a plug, not a cord when unplugging the
If a head or a cord is broken or disconnected, power cord from the socket.
stop using it and request for repair service to
your assembler.

Prohibition Essential matters.

Do not install the product in a wet or dusty Do not impair, process, bend, pull, turn, bind
place, or places with oils or moisture such as or heat the plug cord or power cord. It may
places near kitchen tables or humidifiers. cause fire or electric shock.

(Near baths, kitchen tables and humidifiers)

Do not use at shower room. Prohibition

5
Do not use the probe near eyes. Do not use the probe except on skins. Also, do
not use it where there are volatile or
Don't use it near eyes.
inflammable materials (benzene or thinner).

It may cause fire or accidents.

Prohibition
Prohibition

When you don't use the main body for long, Do not use this product with other
make sure to plug the power cord from the electronics. Also, do not wrap the plug with
socket for safety. cloths or paper. It may cause malfunction.

It may cause fire. It may cause malfunction.

Essential matters. Prohibition

6
3. Precautions

Do not use or leave the product near fire or in Do not put the main body or cords near the
places with very high temperatures such as a heating apparatus.
place under direct sun light. Also, do not heat
(It may cause damages to the main body, probe
the product in the fire. accessories or cords, which leads to fire and electric
(It may cause heating, fire or rupture.) shock. Also, it may cause heating, fire and rupture.)

Caution: Flammables.
Be aware of fire.

Do not spill water or coffee on the plug, the Be careful not to drop or hit the main body
main body or a front panel. or probes.
If you spill water or coffee, it may cause heating. So, It may cause damages to or breakdown in the
turn the main power off promptly, unplug the power product.
cord from the socket, and contact to the assembler.

Prohibition
Prohibition

If abnormal sounds are detected in the main Place the main body on the horizontal place.
body or probes, stop using the product Especially, place the main body in a safe
immediately, unplug the power cord, and place in order not to be dropped when use.
request for repairing work.
It may cause damages to or breakdown in the
product.

Matters to be observed
Matters to be observed

Do not use the product when lightning strikes.


It may cause an over-current due to the
lightning.

Prohibition

7
4. Precaution for use
Those who are in poor health, who take sleeping pills,
who are very tired, and who are treated by doctor should
not use this product. Also, if you feel pain or abnormality,
stop using the product.
Those who use the following electrical devices for medical care
shall not use the product.
1. Implantable medical devices such as Health Mecca
2. Electrical life-support equipment such as a heart-lung machine
3. Device-type medical equipment such as an electrocardiograph
※ This product may cause malfunction of electrical devices for medical care.

Those who have the following symptoms shall not use this product.
● Those with an acute disease
● Those with abnormal blood pressure, infectious diseases or tubercular diseases
● Those with a malignant tumors or heart diseases. Pregnant women
● Those with allergic constitution. Those with bleeder's disease
● Those who take steroids for a long time or those with hemotelangiosis
due to dyshepatia
● Those with atomic damage or officially designated infectious diseases
● Those with facial paralysis, or those who are receiving outpatient treatment
● Those who take medicines. Children under 12
● Pregnant or nursing women
● Those with infectious diseases
● Those who are not fully recovered after surgery
● Those with heart diseases, cholelithiasis or diabetes
● Those with heart, brain diseases, ganglion, hemophthalmia or hepatosis
● Those who are implanted with implants in the body
● Women on the moon cycle. those who had incontinence surgery
● Those with febrile diseases
● Those with atopic dermatitis
※ It may cause accidents or skin troubles.
※ Those who receive outpatient treatment shall consult with the doctor.
※ Do not use this product on children.
※ If you feel bad or the product causes abnormality on your skins when use, stop using the product
immediately.
※ Repeatedly use on the same area or the wrong use may cause damages to skins. If it causes
abnormality to your skin, immediately cool the relevant area and consult with a medical specialist.

8
(But, regular use may cause sensitive skins to be turned into red color. Please read the order of use
and precautions, and implement the right aftercare.)

** Precautions during use


Very important
Please, read all with care.

1. Be careful that the skins of both people controlling and controlled will not go brushed.

2. Do not use it together with an electric floor mat or blanket, other products using coil,

and the conductor with large volume.

3. Be careful that the electrode and Ground plate will not be bumped.

4. When you stop using it for a short time (less than one minute), set the

machine PAUSE

When you stop it for more than one minute, you have to turn it off.

5. When you have to take care of your customers consecutively, do not turn the

main switch off, for operating the cooling fan.

6. When you keep using it for one hours, stop it for about ten minutes.

7. If the generation of heat is not normal during its use, do not raise the power

to the extremely.

If the power indicator doesn’t work, it can result from the damages of the internal parts.

Therefore, you have to turn it off and call your dealer.

9
5. Contents (Basic)
■ Main body ■ RET / CET an Bipolar Handle (3EA)

■ RET Electrodes (3 EA) / CET Electrode (5EA) ■ Acrylic holder

■ Ground plate and bar (2 EA)

10
6. Name and specifications of each part

■ Upper part of the main body

■ Rear side of the main body

④ ②

⑤ ①

① : Power switch & Power cord socket

② : Bipolar connective socket

③ : Ground plate / bar connective socket

④ : CET handle connective socket

11
⑤ : RET handle connective socket

■ Control Panel – Main screen


④ ⑤ ⑥ ⑦ ⑧ ⑩



⑬ 
③ ⑬

No. NAME FUNCTION

RET function selection button


1 RET Selection Button
*The button color will be changed to orange color under activation.

CET function selection button


2 CET Selection Button
*The button color will be changed to orange color under activation.

BIPOLAR function selection button


3 BIPOLAR Selection Button
*The button color will be changed to orange color under activation.

Decrease the ‘Intensity’. (10~100)


4 Intensity Decreasing Button
*Interval 10.

Decrease the ‘Intensity’. (1~100)


5 Intensity Decreasing Button
*Interval 1.

6 Intensity Display Window This window shows selected ‘Intensity’

Increase the ‘Intensity’ (1~100)


7 Intensity Increasing Button
*Interval 1.

12
Increase the ‘Intensity’. (10~100)
8 Intensity Increasing Button
*Interval 10.

9 Sound On/Off Device working sound on/off button

No. NAME FUNCTION

Pause the device


10 Pause Button
*This button is only available under ‘Run’ condition.

Output Power display


11 This window shows current output power of device.
Window

12 Run/Stop Button Run / Stop toggle button (Device working only under ‘Run’)

13 Time Control Button Increase (∧) / Decrease (∨) device working time

14 Time Display Window This window shows time remaining from established time.

■ Control Panel – BIPOLAR Screen


③ ⑦
① ② ④ ⑤

⑪ ⑩

⑨ ⑨

No. NAME FUNCTION

Decrease the ‘Intensity’. (10~100)


1 Intensity Decreasing Button
*Interval 10.

13
Decrease the ‘Intensity’. (1~100)
2 Intensity Decreasing Button
*Interval 1.

3 Intensity Display Window This window shows selected ‘Intensity’

No. NAME FUNCTION

Increase the ‘Intensity’ (1~100)


4 Intensity Increasing Button
*Interval 1.

Increase the ‘Intensity’. (10~100)


5 Intensity Increasing Button
*Interval 10.

6 Sound On/Off Device working sound on/off button

Pause the device


7 Pause Button
*This button is only available under ‘Run’ condition.

8 Run/Stop Button Run / Stop toggle button (Device working only under ‘Run’)

9 Time Control Button Increase (∧) / Decrease (∨) device working time

Save custom preset value which should be selected through number


10 SAVE Button
11.

Select desired value which will be used frequently.


How to save value,
- Select desired button among P1, P2, P3
11 Custom Preset
- Select desired Intensity
- Press ‘10 SAVE Button’.
3 values can be saved in P1, P2, P3

14
7. How to use
1. Connect Power cable
Insert the power cord into the socket on the rear side of the main body.

2. Connect the Ground Plate (Bar) / RET, CET and Bipolar handles into each socket on the rear side of the
main body.

3. Insert the desired electrode into the CET / RET Handle.

Type of
Description Application
Electrode

20mm dia conical shaped active


Feet, nose & hair for Care
electrode

30mm dia round active electrode Face, neck, arm & knee for skin care

Arm, leg, shoulder & knee for body


40mm dia round active electrode
shaping

60mm dia round active electrode Back, thigh & abdomen for body shaping

80mm dia round active electrode Back, thigh & abdomen for body shaping

Face, cheek, jaw & forehead for skin


30mm dia round active electrode
care

Arm, leg, shoulder & knee for body


60mm dia round active electrode
shaping

Back, thigh & abdomen for body


80mm dia round active electrode
shaping

Always make visual check of the surface of the active electrode before use.
Do not use damaged electrode in any case, it may cause sparkle.

15
※WHAT IS CET , RET AND BIPOLAR?

C.E.T(Capacitive Electric Transfer).


Attaching(holding) un- insulated patient plate (bar) on the patient body(hand), then moving insulated
electrode on the patient skin. Electromagnetic energy generated by vigorous electrode activities, heats
the dermis. It is very effective for epidermis treatment.

R.E.T(Resistive Electric Transfer).


Accelerated electrode activities where central tissues between electrode and patient plate generate
deeper heat. It is very effective to produce deeper internal thermal effect on the deep local areas like
visceral fat where is hard to see prominent effectiveness by CET treatment. The deeper internal thermal
conduction enables high effective treatment on abdomen and lumbus

BIPOLAR
Doesn’t need ground plate (bar) for operation. Mainly used for face and scalp treatment.
Not likely the mono-polar system which uses the ground plate (bar) at long opposite party of body, the
supplied bipolar handle has the 3+, 3- poles at a probe part, so it is more sensitive than mono-polar and
that's why the patient could feel that kind of tingle sensation.
Therefore it should uses with less intensity below 50% in case of sensitive part like cheek bone, direct
eye line, chin, and 60% over for normal part of cheek with skin flash as per the reaction of patient.

4. Turn the power on the switch which is in the rear side of the main body.

5. Attach (Grab) Ground Plate (Bar).


Ground Plate : Ground plate is attached to patient’s low back according to treatment wide areas like
abdomen, Hip or thigh before starting the operation.
Ground Bar : Patient grabs ground bar by hand according to treatment small areas like face,
forehead, surroundings of eyes or arm before starting the operation.

6. Set up treatment time.


Set up desired treatment time.
- Increase/Decrease time by adjusting ‘Time Control Button’.
- The adjustable range is 1 to 30 minutes.

7. Start treatment

16
Start the device. ‘Run/ Stop‘ toggle button will be changed from RUN to STOP or STOP to RUN by
pressing button per once.

8. Set up level
- Increase/ Decrease the power level by adjusting ‘Intensity Increasing/Decreasing Button’
- The adjustable range is 1 to 100.

9. Initiates treatment
Apply enough RF cream on the desired area and then, attaching the Handle on the target area.
Adjust each level while handle is attached on the skin.
※ RF energy will be emitted after you press ‘Run/Stop’ button
- You can check the remained working time on the ‘Time Display Window’
* Caution : Handle and Ground plate(bar) should be fully attached (grabbed)
on the patient body while treatment
*Adjusting appropriate level based on individual's skin condition, select proper size of electrodes for
the purpose of treatment.
※The treatment with too high level could be caused of burn. So, don’t apply too high level.

*TIME & NUMBER OF USES

SESSION AT LEAST 8 SESSIONS PER PROBLEM AREA

INTERVAL 2~3 DAYS BETWEEN SESSION

AREA TIME ELECTRODE INTENSITY

FACE 15 min 20, 30 mm & 40mm Appropriate

BODY 15 – 30min 60 & 80mm Normal

17
▪ Other Conditions during the period of treatment:
Taking enough water before and after treatment
Between sessions additional lymphatic treatment will be recommended to help activate the drainage
of the dissolved fat cell thru urine or lymphatic system.
Diets are recommended to maintain fat reduction.

*Cautions for the proper treatment


-The adequate treatment times for a patient per day : 1~2sessions on a one or two parts.
-It needs 3months for a new session program re-starting after the 1’st session.
-The average treatment session for a program is 4weeks and before starting more session, the patient
needs to discuss with doctor for decision of more 1 session, totally 5weeks.

*APPLICABLE AREAS FOR TREATMENT


Bariatrics Clinics - Obesity control by each area of human body
- Combustion of body fatness
- Correction of human form
Mesotherapy - Increase Mesotherapy Treatment Effectiveness
Skin Clinic - Lifting Effectiveness / - Whitening Effectiveness
- Control of Acne and Flecks / - Control of Wrinkles
Pain Clinic - Considerable reduction of joint stiffness
- Pain relief
Alopecia Clinic - Promote hair growth by giving an impulsion to the
hair follicle when the hair roots exist
- Strengthen the weak and thin hair
Obstetrics - Control of the Sagging abdomen after child birth

9. Treatment completion
When set up time is passed, the device will be stopped automatically.
You can stop the machine by pressing ‘Run/Stop’ button when you want to stop treatment. Or you can
pause the device by pressing ‘Pause’ button. You can resume treatment by pressing ‘Pause’ button,
again.
After completing the treatment, disconnect the power cord.

10. Clean after treatment.


Clean the treated area with cloths or tissues.
Clean RF Cream off on the used electrodes and the main body with cloths or tissues, and place the
handle and electrodes on the right position in the acrylic holder.

18
FACE CARE (front)

19
FACE CARE (sides)

20
21
ABDOMEN CARE

Treat the stomach in the clockwise direction

22
Take great care when treating the parts of bones and flank as they are very sensitive.

23
BACK CARE

Take great care when treating the parts of bones and flank as they are very sensitive.

24
25
LEGS (FRONT)

26
27
LEGS (BACK)

28
29
8. Handling and storing after use

1. Regular handling
◎ Every time you use the electrode, wipe it with a dry cloth.
◎ Make sure to keep the surface of the electrode clean.
◎ Every time you use it, wipe it with a dry cloth and treat it with an infrared sterilizer.
◎ Keep it with its power off. Wipe its surface out with a dry cloth. Keep the used parts clean.
◎ Keep it away from direct sunlight, sharp changes of temperature or high temperature and humidity.
◎ Do not leave the measurement, repair or alter of the device to non-professional staff.
◎ Keep it away from shock and vibration.

2. Regular inspection

◎ Make sure to inspect the device and its accessories before use.
◎When you use the device which has not been used for a long time, make sure to check if it works
properly with safety.

30
9. How to replace fuse

Unplug the power cord from the As the fuse is made of glass, you should not
socket. drop it or exercise excessive force on it.

As a spare fuse is set in the power inserting part on the back side of the main body, replace the fuse in the
following method when the fuse is disconnected.

1. Unplug the power cord from the power inserting part and the socket.

2. Pull the fuse holder on the lower part of the power


inserting part using a flat-head screwdriver as shown
in the figure.
1) Insert the end of the screwdriver into the hole
on the upper part of the fuse holder.
2) Pull the fuse holder to the power inserting part
using the principle of the lever. You can pull the fuse holder
simply by pressing the fuse holder with one hand and by
exercising light force on it. Check that the line inside the fuse is
disconnected.
3. Pull the disconnected fuse after pulling the fuse holder out.
4. Pull the spare fuse using a pen and replace it.

5. Complete to replace the fuse by tightly inserting the fuse holder to the power socket.

If you exercise excessive force when pulling the fuse holder out using a flat-head
screwdriver, it may cause injury to the body. Do not exercise too high force.

As the fuse is made of glass, be careful not to drop or break the fuse when handling.

If there is no spare fuse, please contact us or purchase a commercial one.


The standard fuse is 5A, 250V.

31
10. FAQ (Q & A)

1 Q : The power is not turned on even when pressing the power switch.

A1 : Is the power cord plugged surely?


A2 : Isn't the fuse disconnected?
(Please check whether the fuse is disconnected or not by referring to 'How to replace the fuse-25P' in the
manual. When the fuse is disconnected, please replace the fuse according to the method.)

2 Q : Cannot feel the shot even by contacting the probe on the body.

A1 : Is the Standby button changed into a ready mode?


A2 : Please check the sound and operation display lamp. If it is not operated, please contact the AS center.

3 Q : The handle is excessively hot.

A1 : It is thought that the reason is abnormal load just after the use or due to continuous use.
A2 :After stopping operation and disconnecting the power supply, you should get inspection by contacting
the A/S center.

4 Q : The main body is excessively hot.


1) Some parts of a plug or the power cord are abnormally hot.
2) It does not normally operate even after conducting 1~3 checks.

A1 : It is thought to be a breakdown of parts or the machine.


Stop using the machine and contact to the customer service center.

32
11. Statement subject
1. Name of product : Diathermy treatment system

2. Model number : SNE-7100

3. Purpose of use : Equipment to relieve pain by giving more than 2MHz RF energy to

human body

4. Manufacturing no. : Marked on the main body

5. Rated voltage : 100~ 250 VAC, 50/60 Hz

6. Power consumption : 370VA

7. Output Frequency : 470 Khz.

8. Output Power :

TYPE OUTPUT POWER


RET 170W (100Ω load)
CET 300W (500Ω load)
BIPOLAR 45W (100Ω load)

9. Level of protection from electric shock: 1st grade, BF type equipment

10. Weight : 5.3kg (Main body only)

11. Size of the main body (mm) : 415*350*200mm

12. Avaiable Electrode size:

Basic Accessary:

CET: 20mm, 30mm, 40mm, 60mm, 80mm RET: 30mm, 60mm, 80mm

Multipolar: 45mm

33
H/P: RET H/P 1 pcs, CET H/P: 1 pcs, Bi-polar cable, Patient Plate, Patient bar

Option:

Y shape connector(for both hands): for CET, for RET

H/P: CET H/P, RET H/P

RET 60mm, CET 60mm, Electrode cover for RET 60mm

12. About A/S


● On the letter of guarantee

The letter of guarantee is attached to this user manual at the end page. Please check the store and

purchase date, read it carefully, and store it. The guaranteed period for the main body and plug of the

machine is one year from the purchase date.

● Before requesting repairing work

Before requesting repairing work, please read this user manual and check it again. When there is still a

problem, please contact the store where you purchased the product or the company's customer center

mentioned below.

● When contacting us for breakdown

Please contact the company inquiry center mentioned below or a store where you purchased the product,

and request for repairing work.

● About repair after the period of guarantee is expired

Please consult with the store where you purchased the product or our company. When the product will

continue to work through the repair, you can get a repair at a cost according to your opinion. At this time,

you should pay for the freight charge.

After we check deposit of cost, we will deliver the repaired product to you or the store by delivery service.

● Precaution when you receive repair

34
When you request for repair to the store or the company and the cost for repair is under 5 thousand won,

the repair work will be arranged automatically as usual. When the cost for repair is over 5 thousand won,

the repair work will be arranged after getting approval of the store or the customer.

● As for other inquires, please contact

K1MED CO.,LTD
Customer support center : +82-2-871-0657

35
Regulation for free repair

1. If the machine is broken down even when a customer uses it according to the
user's manual and precautions, the customer can receive free repair according to
the regulation for free repair. When you request for free repair, please submit the
product and the letter of guarantee to the store where you purchased the product.
2. Even within the period of guarantee, you should receive the repair at a cost in
the following situations.
① Breakdown or damages by errors in use not according to the user manual
② Breakdown or damages by inappropriate repair or modification
③ Breakdown or damages caused by dropping the product after purchase
④ Breakdown or damages caused by using the product for other purpose than
for family use (e.g. for commercial use)
⑤ Incase that you cannot submit the letter of guarantee
⑥ In case that you did not fill in the date of purchase, the name of a store or
the name of a customer in the letter of guarantee, or the contents are modified.
⑦ Breakdown and damages occurred when the produce is installed in a car or
a ship
⑧ Breakdown and damages due to fire, earthquake, flood, a stroke of lightning
and other natural disasters, abnormal voltage, and the use of power supply not
designated
⑨ Breakdown and damages caused by conducting wrong repair or
modification except by the company
3. This letter of guarantee is valid only in Korea.
4. As this letter of guarantee will not be issued again, please carefully store it.
When you request for free repair, please separate this letter of guarantee from the
user manual and submit it with the product.
● This letter of guarantee promises free repair on the basis of terms and conditions
described in this manual. Accordingly, this letter of guarantee is not to limit the
customer's legal rights. When it is not clear whether free repair is available or not,
please contact the store or the customer support center on the back side of this
letter.
● Please refer to relevant contents on the A/S as for repair after the period of
guarantee is expired or the period of holding components for repair.

36
GUARANTEE

This letter of guarantee means that you can get free repair as of the contents in the
back side of this letter when the product is broken down during the period
described below since the purchase date.
Please refer to the previous page for details.

Name of a product / Diathermy treatment system (SNE-7100)

Manufacturing No. /

Term of guarantee /

Purchase date /

● Name of a customer

Address :

Tel : ( ) Mobile phone :

● Name and address of the store

Tel : ( )
※ Please check the store and the contents of the guarantee, and fill in the blank.