Professional Documents
Culture Documents
Untitled
Untitled
1.4 Calibration................................................................................................... 19
1.4.1 Step 1:VelCal............................................................................................................ 19
1.5 Scan..............................................................................................................32
1.6 CalibrationCheck........................................................................................ 34
1.6.1 Fusion............................................................................................................................34
1.6.2 Volumetric..................................................................................................................... 38
1
1.7.3 TOFD view measurement.......................................................................................... 44
1.8 Report...........................................................................................................46
2
Chapter 1 Software Operation
Double-click , and a window like Fig. 1-1 will pop up after the software is started.
3
1.2 Network Connection
Network connection is the precondition for automated ultrasonic testing. A good network
connection status guarantees the normal operation of the testing system. Users can
The wired network connection is to use a network cable to realize the interconnection
between the ultrasound module and the host PC. The operation steps of the wired
1) Use a network cable to interconnect the network port of the workstation (the inside of
the workstation is wired to the router) and the interface of the ultrasound module. See
Fig. 1-3.
2) Plug the monitor into the VGA port of the ultrasound module. In Ultrasonic module
subnet mask of the network settings. The subnet mask generally defaults to
255.255.255.0.
4
Fig. 1- 4 Ultrasonic module network settings
3) After SuporUp enters the AUT mode, see Fig. 1-5, click the menu
Network-Setting. In the pop-up (Fig. 1-6) window, input the IP address of the
4) After completing the above settings, the button in the upper left of the
SuporUp menu bar is displayed in green connection status, and the ultrasound
5
module screen displays "Network controlling...".
disconnected.
The Wi-Fi network connection is a connection method that only needs to be equipped with
a Wi-Fi network card and a router. The operation steps of the Wi-Fi network setting are
as follows:
1) Insert the Wi-Fi network card into the ultrasound module interface.
additional one.
2) Plug the monitor into the VGA port of the ultrasound module. On the ultrasound
module, go to Setting-Info-WIFI. See Fig. 1-7. In the pop-up Wifi window, check
Auto IP and select the Wifi name (The name of the router that comes with the
uncheck [Auto IP], and fix IP through IP settings, such as "wired connection".
3) After the connection is completed, uncheck "Auto IP", the status is shown in Fig. 1-8
as an example. The current display shows that the GL-AR750S-6ef-5G is connected,
6
Fig. 1-8 Wifi connection
4) Open the SuporUp software, and set the network IP address under
5) After completing the above settings, the SuporUp button shows that the
host PC is connected, and the ultrasound module screen displays " Network
controlling...".
After starting the SuporUp software, go to Workpiece menu under Emulator, and input
the workpiece parameters according to the actual size of the workpiece to be tested in the
7
red box in Fig. 1-10.
After entering Zones menu, the zones can be set up according to the workpiece drawing
of the calibration test block. See Fig. 1-11 for the interface, which is mainly divided into
Take Fig. 1-12 as an example for zonal setting, the setting steps are as follows:
8
Fig. 1-12 Drawing of calibration test block
a) According to the drawing, enter the zones data of the calibration test block into
the zonal setting column for automatic zones. See Fig. 1-13. Input in the
Note:
1) For TRMode, P/E is for emission and reception by itself, and P/C is for
2) The height of the cap area shall be set to simulate the zonal result normally.
b) After the automatic zonal setting is completed, check the zones and reflector
parameters in the zone information area (Fig. 1-14). If the parameters are not
9
the same as those in the drawing, click "Height" and "InspY" column to make
Note: After completing the zonal setting, the zones simulation diagram simulates
the zones and calibration reflector. The red and green lines and dots simulate the
zones and reflector of the fusion area. The blue line and dots simulate the zones
The user enters the probe menu, and completes the parameter setting of the probe and
"TFD-70-*-I".
Note: TOFD probe and wedge models are set according to the models used in
actual testing.
10
Fig. 1-15 "Probe" menu
Enter the SoundBeam menu. See Fig. 1-16 for the screen. The SoundBeam settings of all
channels are completed under this menu. It is mainly divided into three parts: SoundBeam
simulation, channel selection and parameter setting. The calibration of the zonal setting of
Channel selection
11
There are two ways for SoundBeam setting. One is to use "BeamAutoSimu"; the other is
Method 1 - Set up the pitch, and click "BeamAutoSimu". The parameters of all
channels can be automatically set, and the parameter setting of the entire workpiece can
FocusMode, StartEle, EndEle, Aperture, EchoLeg and RefAngle subject to the selected
channel, so as to meet the requirements of detecting reflectors in the current zone. The
Taking the setting of Fusion(US) cap zone as an example. The steps are as follows:
1) Click the channel selection, and set the channel to Fusion (US) Cap.
2) Set the parameters in the parameter setting column. Set DistanceE to -15mm,
the default TRMode "P/E", Focus Mode "Focus", Aperture "20", and EchoLeg
"2".
StartEle to make the SoundBeam fall on the Cap reflector. See Fig. 1-17.
12
4) Adjust Focus. See the figure above, enable the small yellow dot (focus) on the
Taking the setting of US Fill F6 zone setting as an example. The steps are as follows:
1) Click the channel selection bar and set the channel to “Fusion (US) Fill 6”.
Aperture (R) are “12”. Set EchoLeg (T) to “2”, EchoLeg (R) to“3”.
3) As shown in Fig. 1-18, set Refraction Angle (T) to about 50°, Refraction Angle
(R) to about 60°, adjust StartEle (T) and StartEle (R) to enable the SoundBeam
of the transmitting end and the SoundBeam of the receiving end intersect on the
4) Adjust Focus to enable the small yellow dot (focus) on the reflector of the
Take the zonal setting of US HotPass HP2 as an example. The steps are as follows:
1) Click the channel selection bar and set the channel to Fusion (US) HotPass 2.
13
EchoLeg (T) "2".
3) Set RefAngle to about 45°, adjust StartEle to make SoundBeam fall on the
4) Adjust Focus to enable the small yellow dot (focus) on the reflector of the
a) Click the channel selection bar and set the channel to Fusion (US) LCP.
b) The default TRMode is "P/C", FocusMode "Focus", and Aperture "12". Set
c) Set RefAngle (T) and EchoLeg (R) to about 60°. Adjust StartEle (T) and
StartEle (R) so that SoundBeam intersects with the LCP reflector. See Fig.
1-20.
14
Fig. 1-20 LCP SoundBeam setting
d) Adjust Focus so that the small yellow dot (focus) on the SoundBeam simulation
Take the US Root area setting as an example. The steps are as follows:
a) Click the channel selection bar and set the channel to Fusion (US) Root.
EchoLeg "1".
c) Set RefAngle to about 52.5°. Adjust StartEle to make SoundBeam intersect with
15
Fig.1-21 SoundBeam setting of Root
d) Adjust Focus so that the small yellow dot (focus) on the SoundBeam simulation
1.3.4.6 Volumetric
a) Click the channel selection bar and set the channel to "Volumetric (US) V4".
b) The default TRMode is "P/E", FocusMode "No Focus", Aperture "12", and
EchoLeg "2".
c) Set RefAngle to about 45°, and adjust StartEle to make SoundBeam fall on the
16
1.3.4.7 Root Volumetric
Take the US Root Volumetric MAP zone setting as an example. The steps are as follows:
a) Click the channel selection bar and set the channel to "Volumetric (US) Root
Volumetric".
b) The default TRMode is "P/E", FocusMode "No Focus", Aperture "12", and
EchoLeg "1".
d) Set RefAngle to about 52.5°, and adjust StartEle to make SoundBeam fall on
a) Click the channel selection bar and set the channel to "TOFD".
b) As shown in the figure below, the default LayerStart is 0mm, and LayerEnd is
22mm, which is the thickness of the workpiece. The default CalcRule is -6dB.
17
Fig. 1-24 TOFD channel setting
1.3.4.9 CouplingMonitor
EchoLeg 1.
After setting the US channel parameters, click the Sync button on the parameter setting
bar, the parameters are copied to the downstream synchronously, and all the SoundBeam
18
1.3.5 Step 5: Encoder setting
The user enters the ultrasound parameter interface to set the encoder and scan
Fig. 1-26.
1.4 Calibration
Calibration directly affects the final judgment of the test results. Calibrating on the
calibration test block according to the correct calibration steps is an indispensable step
The SoundPath calibration is carried out by using the P/E parameters of HotPass
zone, and the velocity calibration is realized by taking the arc on the phased array
SoundPath calibration generally uses a test block with two known radius of curvature
19
Fig. 1-27 VelCal
Steps:
"Gain" to make the echo amplitude around 80%. Move the probe back and forth
3) Click the “Record” menu to set the value of “Record” to 1, which means to record
the echo position of CalibRefPoint1. Repeat step 2) to make the AWaveEcho of
CalibRefPoint2 pass through the center of the gate and get the highest wave of
CalibRefPoint2.
Click “Record” menu, which means to record the echo position of CalibRefPoint2 and
complete the calibration. If the calibration value of MaterialVel is not accepted, click
20
“ClearCalib” and repeat steps 2)~3).
After entering the sensitivity calibration, the interface is as shown in Fig. 1-29:
1) After selecting the channel for calibration (taking "Fusion (US) F4 zone" as an
example), the remote control controls the PA probe to move to the vicinity of the
calibration reflector, and observe the gate echo in the A-wave until the probe stops at
2) Adjust the parameters of StartEle (T) and StartEle (R) until it is determined that the
echo amplitude under the current StartEle is the highest. Click AdjustGain to
automatically change the gain of echo amplitude to 80%. See Fig. 1-30:
21
Fig. 1-30 Adjust StartEle
Note: The small red dot in the gate of View A is the peak record point for
recording the highest amplitude. Click “ResetPeak” and the software will
3) Click DelayCal to calibrate the calibration echo. After DelayCal, the values of S,
D, R, P are updated. See Fig. 1-31, where the delay deviation P is set to 0.
Note: The delay deviation P represents the relative gate position, which is the
distance between the echo signal and the weld fusion line. When P value is 0, it
22
4) Adjust the remote control and control the probe to pass the calibration reflector
and its adjacent upper and lower reflectors (in the same zone) to test whether it
meets the requirements. (In the example, the adjacent reflector is lower than the
Note:
a) The upper reflector in the top zone F6 and the lower reflector in the bottom
lower than the calibration reflector, but only need to be lower than the
"Aperture" and then repeat steps 2) -step 4) till the calibration pass.
displayed strip chart interface can more intuitively read the echo
23
Axis A Axis B
H2
H1
Axis C
Name Description
H1 Represents the height of the color block. It represents the position where
the reflected signal appears in the time gate. The color of the color block
amplitude is less than 20%, Green for 20%≤amplitude≤40%; Red for 40%
1) Taking the "Fusion (US) HotPass2" channel as an example, the remote control
controls the PA probe to stop at the highest echo position of the calibration
reflector.
2) Adjust StartEle until it is determined that the echo amplitude under the current
StartEle is the highest. Click AdjustGain to adjust the echo amplitude to 80%.
4) Adjust the remote control to control the probe to pass the calibration reflector and
its adjacent upper and lower reflectors (in the same zone), and test whether the
required. If it meets the requirements, see Fig. 1-34, the sensitivity calibration of
24
this channel is completed.
Note:
a) The upper reflector in the uppermost Zone HP2 and the lower reflector
Aperture and then repeat steps 2) -step 4) till the calibration passes.
1) Select "Fusion (US) LCP", the remote control controls the PA probe to stop at
2) Adjust StartEle until it is determined that the echo amplitude under the current
StartEle is the highest. Click AdjustGain to adjust the echo amplitude to 80%.
4) Adjust the remote control to control the probe to pass the calibration reflector and
its adjacent upper and lower reflectors, and test whether the adjacent reflectors
meet the requirements of the echo amplitude shall be lower than the calibration
reflector. If it meets the requirements. See Fig. 1-35, and complete the sensitivity
25
calibration of this channel.
Note:
Aperture and then repeat steps 2) -step 4) till the calibration passes.
The sensitivity calibration procedure of the Root zone is the same as that of LCP. See
2.4.1.3 for details, and the calibration verification is shown in Fig. 1-36.
Note: Since there is no lower reflector in the Root zone, when performing sensitivity
adjacent upper reflector shall be lower than the echo amplitude of the calibration
reflector.
The procedure of Cap sensitivity calibration is the same as that of LCP. See 2.4.1.3 for
26
details, and the calibration verification is shown in Fig. 1-37.
Note: Since Cap does not have an upper reflector, when performing sensitivity
lower reflector shall be lower than the echo amplitude of the calibration reflector.
1) Taking "Volumetric (US) V3" as an example, the remote control controls the PA
2) Adjust StartEle until it is determined that the echo amplitude under the current
StartEle is the highest. Click AdjustGain to adjust the echo amplitude to 80%.
4) Adjust the remote control to control the probe to pass through the calibration
reflector and its adjacent upper and lower reflectors, and test whether the
adjacent reflector meets the requirements of the echo amplitude of the adjacent
reflector shall be lower than the calibration reflector. If it meets the requirements,
27
Fig. 1-38 Volumetric (V3) sensitivity calibration and verification
Note:
Aperture and then repeat steps 2) -step 4) till the calibration passes.
The sensitivity calibration procedure of Root Volumetric is the same as Volumetric. See
2.4.2.1 for details, and the calibration verification is shown in Fig. 1-39.
Note: Since Root Volumetric does not have a lower reflector, when performing
the adjacent upper reflector shall be lower than the echo amplitude of the
calibration reflector.
28
1.4.4 Step 4: CouplingMonitor calibration
1) Ensure that the probe and the wedge are well coupled on the workpiece.
4) Click AdjustGain, and the software will automatically adjust the BW to 80%
If the selected channel type is TOFD, the user needs to perform CalibZero and CalibPCS.
1.4.5.1 CalibZero
When the user enters the CalibZero interface, the interface is as shown in Fig. 1-40.
1) After the probe is well coupled with the wedge, dock the probes properly. See Fig.
1-41:
29
Fig. 1-41 Docking probes
2) Move the relative position of the two TOFD probes and observe the butt echo in
the gate of view A. See Fig. 1-42. When the echo amplitude is the highest, keep
3) Click CalibZero, and Probe Delay will update automatically. Measure "d", and enter
the value of d/2 into "X_Value". See Fig. 1-43. Complete CalibZero.
30
1.4.5.2 CalibPCS
When the user finishes CalibZero, it enters the next step CalibPCS. The interface is as Fig.
1-44.
1) Install the TOFD probe on the crawler, adjust the TOFD arm, and measure the
3) Adjust Range and DispDelay to make the LW (lateral wave) and BW (bottom
wave) completely and clearly displayed in the A view window. Place the PCS red
line of A view at the first echo peak of the calibration echo. Take Fig. 1-45 as an
example. If RefPt is set to 0, place the PCS red line on the first peak of the LW.
31
4) Click CalibPCS, the "PCS" value is updated. If the error between the updated
"PCS" value and the old "PCS" value is within the allowable range, see Fig. 1-46,
then CalibPCS is completed. If the error is greater than the allowable range,
5) After completing the calibration, adjust Gain until the maximum amplitude of the
LW (lateral) reaches 40% -80% as the testing sensitivity. And move the
1.5 Scan
After the calibration is finished, all channels need to be scanned as a whole on the scan
interface, and the parameters can be applied to field testing after verifying the accuracy of
As shown in Fig. 1-47, enter the scan interface. Confirm that the encoder cable of the
crawler is well connected to the encoder port of the ultrasound module. Click to
start the scan.
32
Fig.1-47 Scanning screen
During the scanning process, the position and map of each channel are displayed as
Click to stop and the software will end the scan. Then click the File-Save menu, as
shown in Fig. 1-49, to save the record file.
33
Fig. 1-49 Saving file
1.6 CalibrationCheck
The CalibrationCheck function is used to check and calibrate the overall sensitivity of all
channels.
CalibrationCheck before applying it to the field testing. 2) After the field testing for a
period of time, perform the sensitivity check again in accordance with the standard.
1.6.1 Fusion
1) After opening the record file, see Fig. 1-50,In the AUT view, right-click to select
"Zoom", and left-click to drag the mouse to realize the zonal zoom. Move
to be measured.
Note: The MeasureLine should avoid recording other echoes in the channel to
34
be measured that are higher than the amplitude of the calibration echo to
2) Long press Ctrl, click the title bar of the channel to be tested, and select multiple
channels. See Fig. 1-51. The background color of the channel is highlighted by
Note: The specific menus and description can be found in Table 2-2.
Menu Description
MainAmp Set the allowable error range of the highest echo amplitude of the
UCover/DCover Set the allowable error range of the highest amplitude of the
35
POff Set the allowable measurement error of the delay deviation P value.
BaseAmp Set the BaseAmp of MainAmp for Fusion, Volumetric and TOFD.
4) Click the measurement menu. See Fig. 1-53, In this case, select
MainAmp+UpDownCov (Multi1).
Note: For the description of the measurement menus, see Table 2-3.
Sub-menus Description
MainAmp (Single) Search for the channel selected in the red frame, and measure the
UpCov (Single) For the channel selected in the red frame, set the highest echo
of the channel.
DownCov (Single) For the channel selected in the red frame, set the highest echo
MainAmp (Multi) Search for multiple selected channels, measure the highest echo
36
Sub-menus Description
MainAmp Search for Search for multiple selected channels, measure the
+UpDownCov highest echo amplitude of MainAmp and its upper and lower
MainAmp Search for multiple selected channels and measure the highest
+UpDownCov echo amplitude of MainAmp and its upper and lower adjacent
CalibrationCheck window. As shown in Fig. 1-56, the pass values are displayed
37
Fig. 1-56 CalibrationCheck window
Note: After the measurement is finished, if the UpDownCov does not meet the
1.6.2 Volumetric
Note: The MeasureLine should avoid recording other echoes in the channel to be
measured that are higher than the amplitude of the calibration echo to ensure the
2) Long press Ctrl and click the title bar of the channel to be tested. See Fig. 1-58,
38
select multiple channels.
2) Click the TOFD title bar. See Fig. 1-60, the TOFD channel is selected by the red
frame.
39
3) Click Toolbar-CalibrationCheck measurement menu, select MainAmp (Single)
CalibrationCheck window.
"DelayOffset" does not meet the requirements, click the Calib button under the
CalibrationCheck window. In the pop-up window of Fig. 1-61, click OK for One-key
calibration, scan the calibration test block again, and perform the CalibrationCheck
steps of 2.6.1~2.6.3 again until the calibration results of all channels meet the
requirements.
After the parameters are applied to the field testing, the stored record files can be opened
After opening the map under data analysis. See Fig. 1-62, slide the scroll bar on the left of
the ruler to browse the map as a whole, and then control the length of the scroll bar to
40
Fig. 1-62 Field testing map
To perform AUT measurement on the flaws of the map, the operation is as follows:
double-clicking on the channel name to zoom in the channel image to the entire
viewing area.
41
2) As shown in Fig. 1-65, check in front of the channel name or hold down the
“Ctrl” button, drag the mouse to select/deselect the channel, and select all the
3) Enter the Toolbar-Measure menu. See Fig. 1-66, select "CorrectMethod" as the
42
selected, the menu is active. Click the menu, and the
TOFD view. After clicking this menu, the measurement data will
4) After the setting is finished, click , and the AUT view MeasureResult will be
displayed in the MeasureResult-AUT Image menu under the menu bar window.
1) Right-click the view window and change the view layout to AUT+A.
2) Select the channel with the highest flaw amplitude in the AUT view, and move the
MeasureLine to the highest echo position. As shown in Fig. 1-68, the red box
selects the HP2D channel, and MeasureLine2 is placed at the highest amplitude
43
in the A-view gate.
Fig. 1-69.
If the flaw on the TOFD view is clearly displayed, the user shall measure the TOFD view.
1) Right-click the mouse in the view window and change the view layout to
"AUT+B".
2) After selecting the TOFD channel of the AUT view, B view shows the enlarged
44
3) Move the red LW line to the first peak (white) of the LW. See Fig. 1-71, move the
parabola of MeasureLine1 to fit the first trough (black) of the upper endpoint of
the flaw, and the parabola of MeasureLine2 fit the first peak (white) of the lower
1-72.
Note: If necessary, the user can also perform BImage measurement on other
window simultaneously.
After completing the measurement above, check the summary measurement data in
45
1.8 Report
After the measurement is finished, click General Report under the MeasureResult
window, and a pop-up window is shown as Fig. 1-74. After the user enters the basic
equipment information, click Report, and the software will automatically generate the
report.
46