Accuray 525r Product Data

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

D K Medical System

A c c u R a y- 5 2 5 R
PRODUCT DATA
500mA,125KV General Radiographic
X-Ray System

AccuRay-525R is a high frequency inverter typed radiographic x-ray system which


diagnoses in detail by selecting specifications and components, in accordance with
user's professions and purposes.

1. Generator (ACC-525R)
High Frequency X-Ray Generator
Specifications:
- Electric Power : single phase 220V, 40kVA
- Rating : 125kV 500mA
- Short time rating : 125kV at 320mA
80kV at 500mA
- kV range : 40-125kV (±5%)
- mA range : 10-500mA (±5%)
- mAs range : 0.1-630mAs
- Timer range : 0.001 -10 sec
- Display system : Digital LED display
- Anatomical programming : 216 programs
- Protections : Overload/filament stabilization protection
Circuits should be available
Self-diagnostic massage indication
- Dimensions and weight : Console 400 x 200 x 42 mm
Cabinet - 613 x 510 x 375 mm

2. X-ray tube assembly (E7239X)


Specifications:
- Maximum kVp rating : 125 kVp
- kW rating : 1.0mm-22.5kW
2.0mm-47kW
- Focal spot size : 1.0 x 2.0mm
- Anode heat storage capacity : 140KHU
- Anode cooling storage capacity : 60KHU / min
- Housing heat storage capacity : 1.250KHU
- Target angle : 16°
- Weight : 16 kg

3. Bucky table (TF-32)


Specifications:
- Electric Power : 110V, 60Hz
- Dimensions : 790W X 2.000L X 700Hmm
- Tabletop to film distance : 60mm
- Longitudinal travel : ±300mm
- Transverse travel : ±125mm
- Type : 4 way floating Top
- Size of table : 2000 x 790mm
- Exposure film size : 8 X 10" to 14 X 17
- Table height from floor : 700mm
- Bucky device travel : 402 mm
D K Medical System
A c c u R a y- 5 2 5 R
PRODUCT DATA
- Bucky unit
* Grid ratio : R = 10 : 1
* Grid density : N = 103 lines/inch
- Weight : 130 kg

4. X-ray tube support


1) SF-80
Specifications:
- Electric Power : 110V, 60Hz ( Except SF-80B)
- Type : floor to ceiling
- Balance : weight balance
- Vertical travel : 700-1.870mm
- Longitudinal travel : 1.835mm(SF-80), 1.835mm(SF-80B)
- Transverse travel : 948-1.318mm(R/F)
737-992mm(R)
737-992mm(SF-80B)
- Rotating : ±180°
- Rail
* ceiling : 2,300mm
* floor : 2,800mm
- Installation dimensions : 3.600W x 1.600D x 2.250 ~2.750Hmm
- Weight : 142 kg

5. Wall Bucky Stand


VB-20
Specifications:
- Electric Power : 110V, 60Hz
- Bucky unit to film distance : 45mm
- Type : Up down travel
- Exposure film size : 8″X 10″~14″X 17″
- Vertical travel : 370 - 1.660mm
- Vertical lock : Electro-magnetic lock
- Grid motions : Reciprocating motion by
motor driven
- Grid ratio and density : R=10:1, N=103 lines/inch
- Weight : 90 kg
- Stand height : 1.900H mm

6. Collimator
1) BL-50
Specifications:
- Electric Power : 24VAC, 150W
- Type : Manual , Rectangular
- kV range : 150kVp
- Max field X-ray beam: SID at 65cm less than 35cm × 35cm
- Min field X-ray beam : SID at 100cm less than 5cm × 5cm
- Rotating : ±180°
- Dimensions : 183W × 183D × 179H mm
- Cone rail dimensions : 148W × 164L mm
- Weight : 5 kg