Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 2

SAMPLING PROCEDURE BY ISOKINETIC

1. Isokinetic Sampling is the sampling condition in which the mean velocity of


the air entering the sample probe inlet is the same as the mean velocity of
the unidirectional airflow at that location.
2. The sample probe inlet size should be calculated from consideration of the
sample flow rate of the measuring instrument and the filter exit airflow
velocity so that the probe inlet air velocity approximates to the filter exit
airflow velocity (i.e. Iso-kinetic Sampling)

1. The sampling probe should be of square or rectangular configuration.


2. Dp as shown in above figure is calculated as below

Dp = qva  / (U x Wp)

Dp is the probe dimension parallel to the scan direction (cm)

qva is the actual sample flow rate of the measuring instrument  (photometer) in
cm3/s

U is the average filter exit airflow velocity (cm/s)

Wp is the probe dimension perpendicular to the scan direction (cm)

3. Prior to the generation of the aerosol, place the photometer probe


downstream of the HEPA filter until the instrument gets stabilized. The
photometer has a sample flow rate of 1 cfm. The scanning of the
downstream side will be read as a percentage penetration.
4. Hold the probe at a distance of 3 to 5 cm from the filter face.
5. Scan the filter and the filter perimeter by passing the probe in slightly
overlapping strokes over the filter so that the entire area of the filter
and sealing gasket is sampled.
6. Scan Rate: The probe traverse scan rate should be
approximately15/Wp cm/s.
7. Scanning should be performed over the entire downstream face of
each filter, the perimeter of each filter, the seal between the filter frame
and the grid structure, including its joints.
8. Verify the upstream concentration of challenge aerosol during the test
and ensure that the current concentration is comparable to that of
starting level. 

ACCEPTANCE:
Designated leaks are deemed to have occurred where a reading greater than
10-4 (0.01 %) of the upstream challenge aerosol concentration. Alternative
acceptance criteria may be agreed upon between the customer and the
supplier.
REPAIRS FOR FILTER SCAN TEST:
1. Factory repair should not block or restrict more than 1 % of the filter
face area and no single repair should exceed 13 cm 2 (2 in2)
2. Field repair should not block or restrict more than an additional 3 % of
the filter face area and no single repair should have a lesser dimension
exceeding 3.8 cm (1.5 in.)

REPORTING:
By agreement between the customer and supplier, the following information
and data should be recorded:

a) Test method

b) Type designations of each measurement instrument and apparatus used


and its calibration status;

c) Any special condition or departures or both from this test method and any
special procedures agreed on between the customer and the supplier;

d) Measured upstream aerosol concentrations with their sample point


locations and the corresponding time of measurement;

e) Sample flow rate

f) Calculated average upstream aerosol concentration and its distribution;

g) Calculated acceptance criteria applied for the downstream measurements;

h) Result of the downstream measurement for each clearly identified filter,


area section, or measuring location;

i) Final result of the test for each defined location;

j) If there is no leakage then the test passed, otherwise if there is a leakage


then report leak location, repair action, and the result of re-testing the location

You might also like