Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

SYSTEM

E
Anti-HBc
7C17
34-3095/R4

Anti-HBc
Customer Service
United States: 1-877-4ABBOTT
International: Call your Abbott Representative

This package insert must be read carefully prior to product use. Package insert instructions must be followed accordingly. Reliability of
assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

Note Changes Highlighted

Key to symbols used

List Number Lot Number

For In Vitro Diagnostic Use


Control Number

Store at 2-8°C
Reaction Vessels
©2003, Abbott Laboratories

Sample Cups
Expiration Date

Septum
Serial Number

Reagent Lot
Replacement Caps

Legal Manufacturer
Consult instructions for use

Authorized Representative

See REAGENTS section for a full explanation of symbols used in reagent component naming.

ABBOTT
Max-Planck-Ring 2
65205 Wiesbaden
Germany
+49-6122-580

Abbott Laboratories
Diagnostics Division
Abbott Park, IL 60064 USA Printed in USA

7C17-1C-00_Eng_ReIn.p65 1 12/5/2003, 10:08 AM

Color: PMS 329 CVC


SIZE: 8 1/2 x 11
FILE: Alms
EDITOR: Ann Marie
PUBLISHER: Craig
NAME • 1 or 4 Bottle(s) (10.0 mL per 100 test bottle/
ARCHITECT Anti-HBc 50.9 mL per 500 test bottle) Anti-HBc Assay Diluent containing
protein stabilizers in MOPSO buffer. Preservative: Antimicrobial
INTENDED USE Agents.
The ARCHITECT Anti-HBc assay is a Chemiluminescent Other Reagents
Microparticle Immunoassay (CMIA) for the qualitative detection of
antibody to Hepatitis B core antigen (anti-HBc) in human serum and ARCHITECT i Pre-Trigger Solution
plasma. The ARCHITECT Anti-HBc assay is intended to be used as • Pre-Trigger Solution containing
a screen for blood and plasma to prevent transmission of 1.32% (w/v) hydrogen peroxide.
hepatitis B virus (HBV) to recipients of blood and blood components
and as an aid in the diagnosis of HBV infection. ARCHITECT i Trigger Solution

• Trigger Solution containing 0.35N


SUMMARY AND EXPLANATION OF TEST
sodium hydroxide.
The ARCHITECT Anti-HBc assay utilizes microparticles coated with
recombinant Hepatitis B Virus Core Antigen (rHBcAg) for the ARCHITECT i Wash Buffer
detection of anti-HBc. Anti-HBc determinations can be used to
• Wash Buffer containing phosphate buffered
monitor the progress of hepatitis B viral infection. Anti-HBc is found
saline solution. Preservative: Antimicrobial Agent.
in serum shortly after the appearance of Hepatitis B Surface Antigen
(HBsAg) in acute hepatitis B infections. It will persist after the WARNINGS AND PRECAUTIONS
disappearance of HBsAg and before the appearance of detectable
antibody to HBsAg (anti-HBs).1-7 In the absence of information about • For In Vitro Diagnostic Use.
any other hepatitis B virus (HBV) markers, it must be considered • Package insert instructions must be followed. Reliability of
that an individual with detectable levels of anti-HBc may be actively assay results cannot be guaranteed if there are any
infected with HBV or that the infection may have resolved, leaving deviations from the instructions in this package insert.
the person immune.8 Anti-HBc may be the only serological marker
Safety Precautions
of hepatitis B viral infection and potentially infectious blood.9-15
The presence of anti-HBc does not differentiate between acute or
chronic hepatitis B infection. • CAUTION: All human sourced materials should be
considered potentially infectious. It is recommended that
BIOLOGICAL PRINCIPLES OF THE PROCEDURE human specimens be handled in accordance with the OSHA
The ARCHITECT Anti-HBc assay is a two-step immunoassay, using Standard on Bloodborne Pathogens.16 Biosafety Level 217
Chemiluminescent Microparticle Immunoassay (CMIA) technology, or other appropriate biosafety practices18,19 should be used
for the qualitative detection of anti-HBc in human serum and plasma. for materials that contain or are suspected of containing
infectious agents.
In the first step, sample, assay diluent, and HBc antigen coated
paramagnetic microparticles are combined. Anti-HBc present in the • These precautions include, but are not limited to the following:
sample binds to the HBc antigen coated microparticles. After 1. Wear gloves when handling specimens or reagents.
washing, anti-human acridinium-labeled conjugate is added in the 2. Do not pipette by mouth.
second step. Following another wash cycle, Pre-Trigger and Trigger 3. Do not eat, drink, smoke, apply cosmetics, or handle
Solutions are added to the reaction vessel (RV). The resulting contact lenses in areas where these materials are handled.
chemiluminescent reaction is measured as relative light units 4. Clean and disinfect all spills of specimens or reagents
(RLUs). A direct relationship exists between the amount of using a tuberculocidal disinfectant such as 0.5% sodium
anti-HBc in the sample and the RLUs detected by the hypochlorite, or other suitable disinfectant.20,21
ARCHITECT i* optical system. 5. Decontaminate and dispose of all specimens, reagents and
The presence or absence of anti-HBc in the specimen is determined other potentially contaminated materials in accordance with
by comparing the chemiluminescent signal in the reaction to the local, state and federal regulations.22,23
cutoff signal determined from a previous calibration. If the • ARCHITECT i Trigger Solution contains sodium hydroxide
chemiluminescent signal in the specimen is greater than or equal to (NaOH) and is classified per applicable European Community
the cutoff signal, the specimen is considered reactive for anti-HBc. (EC) Directives as: Irritant (Xi). The following are the appropriate
For additional information on system and assay technology, refer to Risk (R) and Safety (S) phrases.
the ARCHITECT System Operations Manual, Section 3. R41 Risk of serious damage to eyes.
S25 Avoid contact with eyes.
*i = immunoassay S26 In case of contact with eyes, rinse immediately
with plenty of water and seek medical advice.
REAGENTS S35 This material and its container must be disposed
of in a safe way.
Reagent Kit, 100/500 Tests S36/39 Wear suitable protective clothing and eye/face
NOTE: Some kit sizes are not available in all countries; please protection.
S46 If swallowed, seek medical advice immediately
contact your local distributor. and show this container or label.
ARCHITECT Anti-HBc Reagent Kit (7C17)
Information for European customers: For product not classified as
• 1 or 4 Bottle(s) (6.6 mL per 100 test bottle/ dangerous per European Directive 1999/45/EC - Safety data sheet
27.0 mL per 500 test bottle) Hepatitis B Core (E. coli, available for professional user on request.
Recombinant) Antigen Coated Microparticles in MOPSO buffer
• For a detailed discussion of safety precautions during system
with protein stabilizers. Minimum concentration: 0.08% solids.
operation, refer to the ARCHITECT System Operations Manual,
Preservative: Antimicrobial Agents.
Section 8.
• 1 or 4 Bottle(s) (5.9 mL per 100 test bottle/ Handling Precautions
26.3 mL per 500 test bottle) Anti-Human (Mouse, Monoclonal)
• Do not use reagent kits beyond the expiration date.
Acridinium-Labeled Conjugate in MES buffer with protein
stabilizers. Minimum concentration: 0.049 µg/mL. Preservative: • Do not pool reagents within a reagent kit or between reagent
Antimicrobial Agents. kits.

7C17-1C-00_Eng_ReIn.p65 2 12/5/2003, 10:08 AM


• Prior to loading the ARCHITECT Anti-HBc Reagent Kit on the SPECIMEN COLLECTION AND PREPARATION FOR
system for the first time, the microparticle bottle requires mixing ANALYSIS
to resuspend microparticles that may have settled during • Human serum (including serum collected in serum separator
shipment. For microparticle mixing instructions, refer to the tubes) or plasma collected in potassium EDTA, sodium citrate,
PROCEDURE, Assay Procedure section of this package insert. sodium heparin, lithium heparin, ACD, CPDA-1, CP2D, CPD,
• Septums MUST be used to prevent reagent evaporation and and potassium oxalate may be used in the ARCHITECT Anti-HBc
contamination, and to ensure reagent integrity. Reliability assay. Other anticoagulants have not been validated for use with
of assay results cannot be guaranteed if septums are not the ARCHITECT Anti-HBc assay. Follow the collection tube
used according to the instructions in this package insert. manufacturer’s instructions for processing serum or plasma.
• To avoid contamination, wear clean gloves when placing a • The ARCHITECT i System does not provide the capability to
septum on an uncapped reagent bottle. verify specimen type. It is the responsibility of the operator to
• When handling conjugate vials, change gloves that have verify the correct specimen types are used in the ARCHITECT
contacted human plasma/sera, since introduction of human Anti-HBc assay.
IgG will result in a neutralized conjugate. • Performance has not been established using cadaver specimens
• Prior to placing the septum on an uncapped reagent bottle, or body fluids other than human serum or plasma.
squeeze the septum in half to confirm that the slits are open. If • Use caution when handling patient specimens to prevent cross
the slits appear sealed, continue to gently squeeze the septum contamination. Use of disposable pipettes or pipette tips is
to open the slits. recommended.
• Once a septum has been placed on an open reagent bottle, do • This assay was designed and validated for use with human serum
not invert the bottle as this will result in reagent leakage and or plasma from individual patient and donor specimens. Pooled
may compromise assay results. specimens must not be used since the accuracy of their test
• Over time, residual liquids may dry on the septum surface. These results has not been validated.
are typically dried salts and have no effect on assay efficacy. • Do not use heat-inactivated specimens.
• For a detailed discussion of handling precautions during system • Do not use grossly hemolyzed specimens.
operation, refer to the ARCHITECT System Operations Manual,
• For optimal results, inspect all samples for bubbles. Remove
Section 7.
bubbles with an applicator stick prior to analysis. Use a new
Storage Instructions applicator stick for each sample to prevent cross contamination.
• For optimal results, serum and plasma specimens should
• The ARCHITECT Anti-HBc Reagent Kit, Calibrators, be free of fibrin, red blood cells, or other particulate matter.
and Controls must be stored at 2-8°C in an upright position and Such specimens may give inconsistent results and must be
may be used immediately after removal from 2-8°C storage. transferred to a centrifuge tube and centrifuged at least
10,000 RCF (Relative Centrifugal Force) for 10 minutes.
• When stored and handled as directed, reagents are stable until
the expiration date. • Ensure that complete clot formation in serum specimens has
taken place prior to centrifugation. Some specimens, especially
• The ARCHITECT Anti-HBc Reagent Kit may be stored on-board
those from patients receiving anticoagulant or thrombolytic
the ARCHITECT i System for a maximum of 30 days. After
therapy may exhibit increased clotting time. If the specimen is
30 days, the reagent kit must be discarded. For information on
centrifuged before a complete clot forms, the presence of fibrin
tracking on-board time, refer to the ARCHITECT System
may cause erroneous results.
Operations Manual, Section 5.
• Gravity separation is not sufficient for specimen preparation.
• Reagents may be stored on or off the ARCHITECT i System. If
Specimens must be separated from clots or red blood cells
reagents are removed from the system, store them at 2-8°C (with
using centrifugation as recommended by the tube
septums and replacement caps) in an upright position. For
manufacturer.
reagents stored off the system, it is recommended they be stored
in their original trays and boxes to ensure they remain upright. If • Specimens may be stored on or off the clot or red blood cells for
the microparticle bottle does not remain upright (with a up to 7 days at 2-8°C.
septum installed) while in refrigerated storage off the system, • If testing will be delayed more than 7 days, remove serum or
the reagent kit must be discarded. After reagents are removed plasma from the clot, serum separator, or red blood cells, and
from the system, you must initiate a scan to update the on-board store frozen (-20°C or colder).
stability timer. • Specimens must be mixed THOROUGHLY after thawing by
Indications of Reagent Deterioration LOW speed vortexing. Specimens containing particulate
When a control value is out of the specified range, it may indicate matter must be centrifuged prior to running the assay.
deterioration of the reagents or errors in technique. Associated test • Centrifuged specimens with a lipid layer on the top must be
results are invalid and will require retesting. Assay recalibration may transferred to a sample cup or secondary tube. Care must be
be necessary. For troubleshooting information, refer to the taken to transfer only the clarified specimen without the lipemic
ARCHITECT System Operations Manual, Section 10. material.
• No qualitative differences were observed between experimental
INSTRUMENT PROCEDURE
controls and the 24 nonreactive or 24 spiked reactive specimens
• The ARCHITECT Anti-HBc assay file must be installed on the subjected to 6 freeze-thaw cycles; however, multiple freeze-thaw
ARCHITECT i System from the ARCHITECT i Assay CD-ROM cycles should be avoided.
prior to performing the assay. For detailed information on assay
file installation and viewing and editing assay parameters, refer • Specimens from heparinized patients may be partially coagulated
to the ARCHITECT System Operations Manual, Section 2. and erroneous results could occur due to the presence of fibrin.
To prevent this phenomenon, draw the specimen prior to heparin
• For information on printing assay parameters, refer to the therapy.
ARCHITECT System Operations Manual, Section 5.
• For a detailed description of system procedures, refer to the
ARCHITECT System Operations Manual.

7C17-1C-00_Eng_ReIn.p65 3 12/5/2003, 10:08 AM


• When shipped, specimens must be packaged and labeled in • The minimum sample cup volume is calculated by the system
compliance with applicable state, federal, and international and is printed on the Orderlist report. No more than 10 replicates
regulations covering the transport of specimens and infectious may be sampled from the same sample cup. To minimize the
substances. Specimens may be shipped ambient, at 2-8°C (wet effects of evaporation verify adequate sample cup volume is
ice), or -20°C or colder (dry ice). Do not exceed the storage time present prior to running the test.
limitations listed above. Prior to shipment, it is recommended • Priority: 75 µL for the first Anti-HBc test plus 25 µL for each
that specimens be removed from the clot, serum separator, or additional Anti-HBc test from the same sample cup.
red blood cells. • 울 3 hours on-board: 150 µL for the first Anti-HBc test plus
• No qualitative performance differences were observed between 25 µL for each additional Anti-HBc test from the same sample
experimental controls and the 23 nonreactive or 23 spiked cup.
reactive specimens tested with elevated levels of bilirubin • > 3 hours on-board: additional sample volume is required.
(< 20 mg/dL), hemoglobin (< 500 mg/dL), triglycerides Refer to the ARCHITECT System Operations Manual,
(< 3,000 mg/dL), or protein (< 12 g/dL). Section 5 for information on sample evaporation and volumes.
• No qualitative performance differences were observed between • If using primary or aliquot tubes, use the sample gauge to
ensure sufficient patient specimen is present.
experimental controls and 24 nonreactive or 24 spiked reactive
specimens tested with red blood cells at < 0.4% v/v. • ARCHITECT Anti-HBc Calibrator and Controls should be
mixed by gentle inversion prior to use. To obtain the
PROCEDURE recommended volume requirements for the ARCHITECT
Anti-HBc Calibrator and Controls, hold the bottles vertically,
Materials Provided: and dispense 5 drops of the Calibrator or 4 drops of each
• 7C17 ARCHITECT Anti-HBc Reagent Kit Control into each respective sample cup.
• Load samples.
Materials Required but not Provided:
• For information on loading samples, refer to the ARCHITECT
• ARCHITECT i System System Operations Manual, Section 5.
• 7C17-01 ARCHITECT Anti-HBc Calibrator 1 • Press RUN. The ARCHITECT i System performs the following
• 7C17-10 ARCHITECT Anti-HBc Controls functions:
• ARCHITECT i • Moves the sample to the aspiration point.
• Loads a reaction vessel (RV) into the process path.
• ARCHITECT i
• Aspirates and transfers sample into the RV.
• ARCHITECT i
• Advances the RV one position and transfers assay diluent
• ARCHITECT i and microparticles into the RV.
• ARCHITECT i • Mixes, incubates, and washes the reaction mixture.
• ARCHITECT i • Adds conjugate to the RV.
• ARCHITECT i • Mixes, incubates, and washes the reaction mixture.
• Adds Pre-Trigger and Trigger Solutions.
• Pipettes or pipette tips (optional) to deliver the volumes specified
on the patient or control order screen. • Measures chemiluminescent emission to detect the presence
of anti-HBc in the sample.
• For information on materials required for maintenance
• Aspirates contents of RV to liquid waste and unloads RV to
procedures, refer to the ARCHITECT System Operations solid waste.
Manual, Section 9.
• Calculates the result.
Assay Procedure • For optimal performance, it is important to follow the routine
• Before loading the ARCHITECT Anti-HBc Reagent Kit on the maintenance procedures defined in the ARCHITECT System
system for the first time, the microparticle bottle requires mixing Operations Manual, Section 9. If your laboratory requires more
to resuspend microparticles that may have settled during frequent maintenance, follow those procedures.
shipment:
Specimen Dilution Procedures
• Invert the microparticle bottle 30 times.
Specimens cannot be diluted for the ARCHITECT Anti-HBc assay.
• Visually inspect the bottle to ensure microparticles are
resuspended. If microparticles are still adhered to the bottle, Calibration
continue to invert the bottle until the microparticles have been • To perform an ARCHITECT Anti-HBc calibration, test 3 replicates
completely resuspended. of Anti-HBc Calibrator 1. A single sample of all levels of Anti-HBc
• Once the microparticles have been resuspended, remove controls must be tested to evaluate the assay calibration. Ensure
and discard the cap. Wearing clean gloves, remove a septum that assay control values are within the ranges specified in the
from the bag. Squeeze the septum in half to confirm that the Control package insert. Calibrators should be priority loaded.
slits are open. Carefully snap the septum onto the top of the
bottle. • Once an ARCHITECT Anti-HBc calibration is accepted and
• If the microparticles do not resuspend, DO NOT USE; stored, all subsequent samples may be tested without further
contact your local Abbott representative. calibration unless:
• Order calibration, if necessary. • A reagent kit with a new lot number is used
• For information on ordering calibrations, refer to the • Controls are out of range
ARCHITECT System Operations Manual, Section 6. • For detailed information on how to perform an assay calibration,
• Order tests. refer to the ARCHITECT System Operations Manual, Section 6.
• For information on ordering patient specimens, calibrators, QUALITY CONTROL PROCEDURES
controls, and general operating procedures, refer to the
ARCHITECT System Operations Manual, Section 5. NOTE: It is recommended that the ARCHITECT Anti-HBc Positive
Control and Negative Control be run in order to verify the calibration.
• Load the ARCHITECT Anti-HBc Reagent Kit on the
ARCHITECT i System. Verify that all necessary reagents are The recommended control requirement for the ARCHITECT
present. Ensure that septums are present on all reagent bottles. Anti-HBc assay is a single sample of each control tested once every
24 hours each day of use. If the quality control procedures in your

7C17-1C-00_Eng_ReIn.p65 4 12/5/2003, 10:08 AM


laboratory require more frequent use of controls to verify test results, TABLE I
follow your laboratory-specific procedures. Ensure that assay control ARCHITECT Anti-HBc Precision
values are within the ranges specified in the Control package insert. Panel Total No. Grand Mean Intra-assay Inter-assaya Totalb
Members Replicates (S/CO) SD %CV SD %CV SD %CV
Verification of Assay Claims
For protocols to verify package insert claims, refer to the ARCHITECT 1 120 0.25 0.020 7.7 0.026 10.4 0.056 22.1
2 120 0.92 0.041 4.4 0.046 5.0 0.072 7.8
System Operations Manual, Appendix B. The ARCHITECT Anti-HBc
3 120 1.70 0.079 4.7 0.082 4.9 0.111 6.5
assay belongs to method group 5. 4 120 6.39 0.663 10.4 0.691 10.8 0.728 11.4
5 120 8.98 0.390 4.3 0.440 4.9 0.507 5.6
RESULTS 6 120 11.50 0.720 6.3 0.733 6.4 0.799 6.9
The ARCHITECT i System calculates the Anti-HBc Calibrator 1 mean Positive Control 120 1.68 0.074 4.4 0.081 4.8 0.110 6.6

chemiluminescent signal from 3 Anti-HBc Calibrator 1 replicates and a Inter-assay variability contains intra-assay variability.
stores the result. b Total variability contains intra-assay, inter-assay, inter-lot, and inter-site variability.

Calculation Specificity
The ARCHITECT Anti-HBc assay calculates a result based on A total of 2709 serum and plasma specimens from volunteer whole
S/CO. blood donors, a low prevalence population for HBV infection, were
• Cutoff Calculation: Calibrator Mean RLU value x 1.1 = Cutoff evaluated by three clinical sites (Table II). The presence of anti-HBc
RLU in ARCHITECT anti-HBc reactive specimens was supported by the
presence of other HBV markers including HBsAg, IgM anti-HBc,
• S/CO = Sample RLU/Cutoff RLU
anti-HBc, anti-HBe, and HBV DNA. The initial and repeat reactive
Interpretation of Results rates were 0.85% (23/2709) and 0.81% (22/2709), respectively.
• Specimens with S/CO values < 1.00 are considered nonreactive There were a total of 22 repeatedly reactive specimens of which 13
by the ARCHITECT Anti-HBc assay and need not be tested were positive, 1 was indeterminate, and 8 were negative following
further. supplemental testing.
• Specimens with S/CO values > 1.00 are considered reactive by Twenty-seven of 334 specimens obtained from hospital patients were
the ARCHITECT Anti-HBc assay. repeatedly reactive of which 26 were positive and 1 was negative
following supplemental testing. In 50 plasma specimens from
• All initially reactive specimens should be retested in duplicate. If
matched serum and plasma pairs, 1 specimen was repeatedly
both retest values are nonreactive, the specimen must be
reactive and was positive following supplemental testing. In
considered nonreactive for anti-HBc. If either of the retest values
190 specimens from individuals with medical conditions unrelated
is reactive, the specimen must be considered repeatedly reactive
to HBV infection and specimens containing potentially interfering
for anti-HBc by the criteria of ARCHITECT Anti-HBc.
substances, 15 specimens were repeatedly reactive of which 7 were
NOTE: For details on configuring the ARCHITECT i System to use positive and 8 were negative following supplemental testing.
Grayzone Interpretations, refer to the ARCHITECT System
TABLE II
Operations Manual, Section 2, subsection Assay Settings, Configure
Reactivity of the ARCHITECT Anti-HBc Assay in Specimens from
assay parameters dialog window - Interpretation. Volunteer Whole Blood Donors, Hospital Patients, Plasma Specimens
Flags from Matched Serum/Plasma Pairs, Individuals with Medical
Conditions Unrelated to HBV Infection, and in Specimens Containing
• Some results may contain information in the Flags field. For a Potentially Interfering Substances
description of the flags that may appear in this field, refer to the
Number of Positive by
ARCHITECT System Operations Manual, Section 5. Supplemental Testinga
Number (% of Repeatedly
LIMITATIONS OF THE PROCEDURE Category Tested IR (% of Total) RR (% of Total) Reactive)

• If the anti-HBc results are inconsistent with clinical evidence, Volunteer Whole Blood Donors
Serum 1347 16 (1.19%) 15 (1.11%) 10 (66.67%)
additional testing is suggested to confirm the result. Plasma 1362 7 (0.51%) 7 (0.51%) 3 (42.86%)

• For diagnostic purposes, results should be used in conjunction Total Donors 2709 23 (0.85%) 22 (0.81%) 13 (59.09%)

with patient history and other hepatitis markers for diagnosis of Hospital patients 334 28 (8.38%) 27 (8.08%) 26 (96.30%)

acute or chronic infection. Plasma Specimens from


Matched Serum/Plasma Pairs 50 1 (2.00%) 1 (2.00%) 1 (100.00%)
• Samples containing particulate matter or red blood cells must be
Medical Conditions Unrelated
centrifuged prior to running the assay. to HBV Infection and Potentially
Interfering Substancesb 190 15 (7.89%) 15 (7.89%) 7 (46.67%)
• Performance has not been established using cadaver specimens
or body fluids other than human serum or plasma. IR = Initially Reactive; RR = Repeatedly Reactive
• Do not use heat-inactivated specimens. a A specimen was defined as anti-HBc positive if the anti-HBc (comparator method)
was reactive and one or more of the following HBV markers were detected: HBsAg,
• Specimens from heparinized patients may be partially coagulated IgM anti-HBc, anti-HBe, HBV DNA PCR, or anti-HBc (by additional ELISA methods).
and erroneous results could occur due to the presence of fibrin. b Category included the following: anti-CMV positive (10), anti-HSV (10), anti-HAV (10),
IgM anti-HAV (10), HBV vaccine recipients (10), rubella antibody positive (10),
To prevent this phenomenon, draw the specimen prior to heparin toxoplasma antibody positive (10), E. coli infections (10), yeast infections (10), syphilis
therapy. positive (10), anti-nuclear antibody positive (10), rheumatoid factor (10), multiple
myeloma (10), pregnant females (50), and alcoholic liver disease (10).
SPECIFIC PERFORMANCE CHARACTERISTICS Sensitivity
Precision A total of 349 specimens from 182 individuals known to be positive
The precision of the ARCHITECT Anti-HBc assay was determined for total anti-HBc, 20 individuals known to be positive for IgM
during clinical studies using two reagent lots. A panel composed of anti-HBc, 18 individuals with acute HBV infection, 48 individuals with
six unique members was tested in replicates of four with each reagent chronic HBV infection, 14 individuals from recovered HBV infection,
lot once daily for five days at three sites. Each daily run also included and 67 individuals at increased risk for HBV infection were tested.
the ARCHITECT Positive Controls tested in duplicate at the beginning Of the 349 specimens, 314 (89.97%) were repeatedly reactive and
and end of the run. The intra-assay and inter-assay standard of these 314 specimens, 313 (99.68%) were positive by supplemental
deviations (SD) and percent coefficients of variation (%CV) were testing (Table III).
determined with a variance component analysis24 for a random
effects model25 (Table I).

7C17-1C-00_Eng_ReIn.p65 5 12/5/2003, 10:08 AM


TABLE III 9. Seeff LB, Beebe GW, Hoofnagle JH, et al. A Serologic Follow-
Reactivity of the ARCHITECT Anti-HBc Assay in Selected Populations up of the 1942 Epidemic of Post-vaccination Hepatitis in the
with HBV Infection and at Increased Risk for HBV Infection United States Army. N Eng J Med 1987;316:965–70.
ARCHITECT Anti-HBc 10. Katchaki JN, Siem TH, Brouwer R, et al. Detection and
Assay Number of Number of Positive by Significance of Anti-HBc in the Blood Bank: Preliminary Results
Number Repeatedly Reactive Supplemental Testing of a Controlled Prospective Study. J Virol Methods 1980;2:
Category Tested (% of Total) (% of Repeatedly Reactive) 119–25.
Preselected total anti-HBc Positive 182 182a (100.00%) 182 (100.00%) 11. Lander JJ, Gitnick GL, Gelb LH, et al. Anticore Antibody
Preselected IgM anti-HBc Positive 20 20a (100.00%) 20 (100.00%) Screening of Transfused Blood. Vox Sang 1978;34:77–80.
Acute HBV Infection 18 18 (100.00%) 18 (100.00%)
12. Hoofnagle JH, Seeff LB, Buskell-Bales Z, et al. Hepatitis
Chronic HBV Infection 48 48 (100.00%) 48 (100.00%)
Cooperative Study Group. Type B Hepatitis After Transfusion
Recovered HBV Infection 14 14 (100.00%) 14 (100.00%)
with Blood Containing Antibody to Hepatitis B Core Antigen.
Increased Risk for HBV Infectionb 67 32 (47.76%) 31 (96.88%)
N Eng J Med 1978;298:1379–83.
TOTAL 349 314 (89.97%) 313 (99.68%)
13. Koziol DE, Holland PV, Alling DW, et al. Antibody to Hepatitis B
a Specimens were tested once. Core Antigen as a Paradoxical Marker for Non-A, Non-B Hepatitis
b Category included the following: intravenous drug users (34), hemophilia patients (33). Agents in Donated Blood. Ann Internal Med 1986;104:488–95.
Overall Specificity and Sensitivity 14. AuBuchon JP, Sandler SG, Fang CT, et al. American Red Cross
Overall specificity and sensitivity were estimated from the results of Experience with Routine Testing for Hepatitis B Core Antibody.
Transfusion 1989;29:230–2.
3632 serum and plasma specimens tested with the ARCHITECT
Anti-HBc assay at six clinical sites. 15. Stevens CE, Aach RD, Hollinger FB, et al. Hepatitis B Virus
Antibody in Blood Donors and the Occurrence of Non-A, Non-B
The overall specificity was estimated to be 99.42% (3248/3267) with Hepatitis in Transfusion Recipients: An Analysis of the
a 95% confidence interval of 99.09% to 99.65%. The overall Transfusion-Transmitted Viruses Study. Ann Intern Med
sensitivity was estimated to be 98.63% (360/365) with a 95% 1984;101:733–8.
confidence interval of 96.83 to 99.55%. 16. US Department of Labor, Occupational Safety and Health
Analytical Sensitivity Administration, 29 CFR Part 1910.1030, Occupational Exposure
to Bloodborne Pathogens; Final Rule. Federal Register
The sensitivity of the ARCHITECT Anti-HBc assay was evaluated 1991;56(235):64175–82.
with a six-member panel that was standardized against reference
17. US Department of Health and Human Services. Biosafety in
serum from the Paul-Ehrlich-Institute (PEI). The panel was tested
Microbiological and Biomedical Laboratories, Fourth Edition.
with two reagent master lots at three clinical sites. The ARCHITECT Washington, DC: US Government Printing Office, May 1999.
Anti-HBc assay sensitivity ranged from 0.537 to 0.542 PEI
18. World Health Organization. Laboratory Biosafety Manual.
Units/mL.
Geneva: World Health Organization, 1993.
Common Technical Specifications Sensitivity and Specificity 19. National Committee for Clinical Laboratory Standards. Protection
The diagnostic sensitivity and specificity of the ARCHITECT Anti-HBc of Laboratory Workers from Occupationally Acquired Infections:
assay were calculated according to the requirements of the Common Approved Guideline - Second Edition. NCCLS Document
Technical Specifications 2002/364/EC. Additional specimens were M29-A2. Wayne, PA: NCCLS, 2001.
tested in the categories of volunteer whole blood donor (n=2362) 20. CDC, Guidelines for the prevention of transmission of Human
and preselected total anti-HBc positive (n=119) to comply with the Immunodeficiency Virus and Hepatitis B Virus to health-care and
sample sizes required by the Specifications. The results from these public-safety workers. MMWR 1989,38, (S-6); 16S.
additional specimens were combined with those of the clinical 21. Sehulster LM, Hollinger FB, Dreesman GR, et al. Immunological
studies. and Biophysical Alteration of Hepatitis B Virus Antigens by
Sodium Hypochlorite Disinfection. Appl Envir Microbiol
The diagnostic sensitivity was determined to be 100.00% (401/401) 1981;42(5):762–7.
with a 95% confidence interval of 99.08% to 100.00%. There were
22. National Committee for Clinical Laboratory Standards. Clinical
five ARCHITECT false reactive specimens identified after testing of Laboratory Waste Management: Approved Guideline. NCCLS
the additional volunteer whole blood donor specimens. The Document GP5-A. Villanova, PA: NCCLS, 1993;13(22):1-18,
diagnostic specificity was determined to be 99.58% (5605/5629) with 29-42.
a 95% confidence interval of 99.37% to 99.73%. 23. US Environmental Protection Agency. EPA Guide for Infectious
Waste Management. Publication No. EPA/530-SW-86-014.
BIBLIOGRAPHY Washington, DC: US Environmental Protection Agency, 1986:
1. Hoofnagle JH, Gerety RJ, and Barker LF. Antibody to Hepatitis 1-1–5-5, R1–R3, A1–A24.
B-Virus Core in Man. Lancet 1973;ii:869–73.
24. Box GEP, Hunter WG, Hunter JS. Statistics for Experimenters:
2. Szmuness W, Hoofnagle JH, Stevens CE, et al. Antibody against An Introduction to Design, Data Analysis, and Model Building.
the Hepatitis Type B Core Antigen. Am J Epidemiol New York: John Wiley & Sons, Inc, 1978:510-39, 571-83.
1976;104:256–62.
25. SAS Institute Inc. The MIXED Procedure. In: SAS® Technical
3. Hoofnagle JH, Seeff LB, Buskell-Bales Z, et al. Serologic Report P-229, SAS/STAT ® Software: Changes and
Responses in HB. In: Viral Hepatitis (Vyas GN, Cohen SN, and Enhancements, Release 6.07. Cary, NC: SAS Institute Inc,
Schmid R., eds.) The Franklin Institute Press, 1978;219–42. 1992:289-366.
4. Krugman S, Overby LR, Mushahwar IK, Ling CM, et al. Viral
Hepatitis, Type B, Studies on Natural History and Prevention
Reexamined. N Eng J Med 1979;300:101–6. The following U.S. Patents are relevant to the ARCHITECT system
or its components. There are other such patents and patent
5. Zito DR, Gurdak RG, Tucker FL, et al. Hepatitis B Virus Serology:
Loss of Antibody to Surface Antigen. Am J Clin Path 1987;88: applications in the United States and worldwide.
229–31. 5,468,646 5,543,524 5,545,739 5,565,570 5,669,819 5,783,699
6. Gitlin N. Hepatitis B: Diagnosis, Prevention, and Treatment. Clin
Chem 1997;43(8B):1500–6. ARCHITECT is a registered trademark of Abbott Laboratories, Abbott
7. Koff RS. Viral Hepatitis. In: Schiff L, Schiff ER, editors. Diseases Park, IL USA.
of the Liver. 7th ed. Philadelphia: JB Lippincott, 1993:492–577.
8. Dodd RY, Popovsky MA and Members of the Scientific Section Abbott Laboratories
Coordinating Committee. Antibodies to Hepatitis B Core Antigen Diagnostics Division
and the Infectivity of the Blood Supply. Transfusion 1991;31: Abbott Park, IL 60064 USA November, 2003
443–9.

7C17-1C-00_Eng_ReIn.p65 6 12/5/2003, 10:08 AM

You might also like