Professional Documents
Culture Documents
UserCom37 TA EN
UserCom37 TA EN
for a certain time and continuously mea- ity is greater than about 0.7. The mois- two thermostats. One is used to thermo-
suring its mass using a balance. ture content that corresponds to a water stat the balance and to cool the furnace
activity of 0.7 can vary between a few flange and furnace jacket, the other to
The measurement time should be long percent and several tens of percent de- thermostat the humidity interface.
enough for the mass of the sample to pending on the product. A sorption curve
reach a constant value. The RH is then allows the corresponding water content METTLER TOLEDO offers two humid-
increased in steps keeping the tempera- in a product to be determined for a par- ity generators that can be coupled to the
ture constant and allowing sufficient ticular water activity. TGA/DSC 1. The two humidity generators
time for the sample mass to stabilize differ mainly in the way the RH is con-
each time at the new RH level. The equi- Experimental details trolled. Both types of humidity genera-
librium moisture content of the sample is Sorption isotherms can be measured by tor were used for the measurements de-
determined for the different values of RH attaching a specially designed humidity scribed in this article.
and plotted as a function of the RH. generator to a TGA/DSC 1 (Figure 1).
Temperature dependence
This yields a so-called sorption isotherm. The humidity generator produces a well- of relative humidity
Under these conditions (equilibrium), the defined gas flow with a precise RH. The The relative humidity depends strongly
RH corresponds to the water activity (aw) gas flow from the humidity generator on temperature because the vapor pres-
of the water in the sample. Water activity passes into the furnace chamber of the sure of water is very temperature depen-
is related to the partial pressure of water TGA/DSC 1 through a humidity interface dent. This is illustrated in Figure 2, which
above the sample and is not the same as attached to the outlet of the TGA furnace. shows the corresponding RH levels at dif-
the water content. It is a measure of the ferent temperatures for initial RH levels
relatively free or unbound water present T he f low rat e is t y pica lly about of 30%, 40%, 60% and 80% at 40 °C.
in a raw material, food or product that 100 mL/min. About 5 mL/min dry gas
is available to support processes such as is also required as protective gas for the If, for example, the RH of air at 40 °C is
non-enzymatic browning (Maillard re- balance. To prevent condensation, the 40% (red curve), the RH of the same air
action), fat oxidation, the growth of mi- temperature of the interface must be at 30 °C is 68% (see the horizontal red
arrow). At about 23 °C, the RH of this air
Figure 1.
Schematic diagram Temperature and Humidity Sensors is 100%. This means that at lower tem-
Control Unit
of a system for peratures part of the water vapor would
measuring sorption
condense (dotted red line) – which ex-
isotherms by TGA.
plains why drops of water form in the
Dry Flow
refrigerator (after opening it) or on cold
Protective Gas
Mass
Mixing Unit
Saturator
The pronounced temperature depen-
TGA Furnace Humidity Generator dence of RH means that a small un-
certainty in temperature leads to a
Gas Outlet
large uncertainty in the RH. This is
demonstrated in Figure 3. It shows the
Table 1. temperature dependence of the relative
Temperature LiCl MgCl26 H2O Mg(NO3)26 H2O NaCl
Deliquescence uncertainty of the RH for a temperature
relative humidity for in °C in % in % in % in %
uncertainty of 1 K.
different salts at dif-
ferent temperatures.
20 11.3 33.1 54.4 75.5
30 11.3 32.4 51.4 75.1 Example: Let us assume that the RH
40 11.2 31.6 48.4 74.7 at 60 °C is 50%. An uncertainty of the
50 11.1 30.5 45.4 temperature of 1 K (i.e. the tempera-
ture is 60 ± 1 °C) results in a relative
60 11.0 29.3
uncertainty of RH of 4.6% (see arrows in
70 10.8 27.8 Figure 3).
2 METTLER TOLEDO UserCom 37
This gives a value of 50 ± 2.3% for the Figure 2.
Dependence of the
RH. Small uncertainties in the tem- relative humidity on
perature (in this example 1 K) therefore temperature.
cause large uncertainties in the RH. This
effect is more pronounced at low temper-
atures than at high temperatures.
Figure 3.
Table 1 lists some of the salts used as cal- Relative change of
ibration substances together with the cor- RH at a particular
responding values of the deliquescence temperature when
the temperature is
relative humidity at different tempera- changed by 1 K.
tures. There are two possible experimen-
tal approaches.
Method 1:
The sample is exposed to a relative hu-
midity in the TGA/DSC 1 that is higher
than the deliquescence relative humidity
of the substance used. Under these cir-
cumstances, the sample absorbs water
and the sample mass increases. The RH
in the furnace chamber is then reduced
in steps. When it reaches the deliques-
cence relative humidity, the sample be-
gins to lose water and the sample mass
decreases. Figure 4.
Determination of
The maximum in the TGA curve is the deliquescence
relative humidity
therefore reached when the RH in the of MgCl26 H2O
furnace chamber corresponds to the at 27.5 °C using
Method 1 (see text
deliquescence relative humidity of the
for details).
sample. This is illustrated in Figure
4, which shows results obtained using
MgCl26 H2O.
Method 2:
The sample is exposed to a relative hu-
midity in the TGA/DSC 1 that is lower
than the deliquescence relative humidity
of the material used. Under these condi-
tions, the sample does not absorb water
and the sample mass remains constant.
The RH in the furnace chamber is then
METTLER TOLEDO UserCom 37 3
increased in steps. When it reaches the
deliquescence relative humidity, the
sample begins to absorb water and the
sample mass increases.
TA Tipp
4 METTLER TOLEDO UserCom 37
arrow). Figure 7 also shows that a hys- Formula 1
aw k 1 2 1 2 1
teresis effect occurs between the sorption = –1 + 1– a + = α · a2w + β · aw + γ
and desorption processes. Sorption hys-
w m0 c aw m0 c w m0· k · c
teresis is in fact a phenomenon that oc-
curs quite generally; the reason for this is Formula 2
k=√
2
not yet fully understood. β – 4αγ – β β 1
, c= + 2, m0 =
2γ γ⋅k γ⋅k⋅c
Different models are used to describe
sorption isotherms (for details see the
Figure 8.
books listed under Literature). In prac- GAB fit for the mea-
tice, the Guggenheim–Anderson–de sured sorption data
Boer (GAB) model is often used. In its of wheat starch.
METTLER TOLEDO UserCom 37 5
Mobile communication
possibilities with the STARe System
New in our Sales Program
E-mail notification
In the last issue of UserCom, we introduced the new e-mail and
SMS functionality that informs you about the status of your in-
strument. You know whether your measurement series has been
completed without having to look in the laboratory. You define
in advance the events you want to be informed about. The fol-
lowing information can be transmitted:
• Measurement completed
• Error on TA module during the experiment,
measurement stopped
• Warning on TA module during the experiment,
possible malfunction
• Waiting for sample insertion
• Waiting for sample removal
6 METTLER TOLEDO UserCom 37
The new METTLER TOLEDO
micro and ultramicro balances
The METTLER TOLEDO XP2U and XP6U ly structured customizable user interface
ultramicro balances set new standards in for easy, fast, and error-free operation.
terms of weighing accuracy and speed. Numerous quality management func-
The balances are designed to boost effi- tions support regulatory requirements.
ciency and reliability as well as support Even sample sizes of less than 1 mg
network compatibility. conform to the strict process tolerances
and satisfy the current international USP
Excellent repeatability standards.
Thanks to an improved weighing cell,
new electronics and optimized signal Built-in safety functions
processing, the ultramicro balances Together with extensive antistatic mea-
combine excellent reproducibility, higher sures and the optional LabX balance
capacity and a short stabilization time. software for reliable and safe data man-
The XP2U and XP6U offer unsurpassed agement, the XP2U and XP6U ultramicro
repeatability of up to 0.15 µg; reduction balances provide a complete weighing
of sample weight requirements allows solution that satisfies the highest user
considerable cost reductions. requirements.
METTLER TOLEDO UserCom 37 7
Complete deformulation of your sample
Procedures like failure analysis, process • Right-click functionality for simple The OMNIC Mercury software extracts
Applications
validation and competitive investigations searches everything – what and when – giving
often require that a sample be deformu- • Transfer to an OMNIC™ Specta™ for you a full profile of your sample:
lated to identify components and to un- multi-component mixture searching • Total gas evolution profile showing the
derstand processing differences. • One-Click analysis of the entire data overall IR signal
set through Mercury TGA software • Component search results plus re-
• Extensive gas phase spectral libraries sidual
for searching • Time profiles for each component re-
vealing underlying chemistry
Ideally, deformulation via TGA-IR entails • Visual comparison window permitting
getting three pieces of information: the analyst to determine completeness
• Weight loss for quantitation of the result
• Identification of off-gassing material • Total analysis time reduced from 30 to
for compositional analysis 60 minutes to less than 2 minutes
• Time profiles for component evolution • Consistent results from user to user
to probe processing conditions • Makes TGA-IR a practical tool
The first item is extracted from the TGA Mercury TGA: One Click and Done Defor-
data through weight loss curves. The se- mulation Analysis.
cond is extracted by searching the spect-
ra, while the third comes from profiling
a particular component in time.
The Thermo Scientific™ Nicolet™ iS™50
FT-IR provides excellent tools for this Critically, the TGA-IR gas evolutions Nicolet iS™50 Modules
application: do not occur one at a time. Individual The Nicolet iS™50 modules expand
the materials analysis capability
• Excellent quality spectra across the components can be evolved severely over-
further:
mid-IR lapped with other gases, and the compo-
• Simple linkage to Thermal Gravimetric nents do not evolve with the same time • Built-in, diamond mid/far-IR ATR
Analysis (TGA) systems profile. - Bulk material analysis before
the TGA experiment
• Optimized gas cell providing rapid re-
- Residue analysis after TGA-IR
sponse to off-gassing materials This can provide crucial information
for inorganics, ash
• 3-Dimensional graphic viewing of about processing, like heat treating of a
time-based data sets for manual review rubber. • iS™50 Raman module
- Mount the TGA-IR module in
Figure 1. an external sample com-
Mercury TGA reveals partment iS™50 AEM
what (lower right) - Gain morphology insights and
and when (lower
complementary search results
left) and offers a
visual comparison to FTIR
tool (upper right) – • iS™50 ABX Automatic Beamsplit-
everything you can
ter Exchanger
get from your data
in one click! - Automate multi-range: Raman,
mid-far IR-ATR with one touch
8 METTLER TOLEDO UserCom 37
The strange behavior of a bouncing modeling putty
Nicolas Fedelich
“Horst die Hüpfknete” is the German name for a make of bouncing modeling putty used as a toy for children.
The material exhibits two quite different types of behavior at room temperature: On the one hand, it can be
easily shaped and stretched and flows like a viscous liquid. In contrast, if it is rolled into a ball and allowed
to fall to the ground, it behaves elastically and bounces back up like a tennis ball.
METTLER TOLEDO UserCom 37 9
The second glass transition exhibits very
strong frequency dependence: the peak
temperature of G'' in the measured fre-
quency range shifts by more than 8 K.
Applications
10 METTLER TOLEDO UserCom 37
A DSC measurement indicated that the Table 1.
Effect DSC DMA at 1 Hz
Summary of the
putty consisted of silicone rubber (about
Onset G’ Peak G’’ DSC and DMA mea-
90%) and another component (about 10%). surements of the
Glass transition 1 in °C –121.2 –116.2 –114.6 putty.
This second material was directly de- Crystallization in °C –78.6 –84.6 –84.2
tected by DMA measurements and had Melting in °C –53.8 –50.9 –50.2
an unusually broad glass transition of Glass transition 2 in °C Not observed –15.6 –13.6
about 40 K. DMA measurements at differ-
ent frequencies clearly demonstrated why Figure 6.
the putty exhibits both elastic and flow Isothermal frequen-
cy sweep show-
properties at room temperature. ing the frequency
dependence of the
The measurements showed that, at low shear storage mod-
ulus (G') and loss
frequencies, the loss modulus was signif- modulus (G'') of the
icantly greater than the storage modulus. putty at 25 °C.
This explains the flow properties of the
putty. At higher frequencies, the storage
modulus was considerably greater than
the loss modulus and the putty behaves
elastically.
METTLER TOLEDO UserCom 37 11
Study of the curing behavior
of a trifunctional epoxy resin
Applications
This article describes how the curing behavior of a highly crosslinked epoxy resin system was investigated
using conventional DSC and TOPEM® measurements. The results showed that the curing behavior is much
easier to study by TOPEM® than by conventional DSC. Furthermore, the glass transition temperature of the
completely cured material can only be determined by TOPEM®.
Methods
The conventional DSC experiments were
performed using a METTLER TOLEDO
12 METTLER TOLEDO UserCom 37
DSC821e and theTOPEM ® measure- Table 1.
Curing time 0.5 1 1.5 2 3 5 6
Glass transition
ments with a DSC823e. TOPEM® is an in h temperatures after
advanced temperature-modulated DSC Tg in °C 23.3 94.5 139.2 158 175.8 194.2 198.1 curing for different
technique in which stochastic tempera- times at 150 °C.
Vitrification no no no yes yes yes yes
ture pulses are superimposed on the un-
derlying DSC temperature program. The The postcuring reaction was completed The heating runs are shown in Figure 2.
amplitude of these temperature pulses is by about 280 °C. The exothermic in- Table 1 presents an overview of the iso-
constant (typically 0.1 K). crease of the heat flow above 280 °C thermal curing times used and the mea-
indicates that the sample begins to de- sured glass transition temperatures. The
The duration of the pulses is varied ran- grade. In principle, it should now be pos- results in the table indicate that the vit-
domly within a certain range (typically sible to determine the glass transition of rification time at 150 °C is between 1.5
10 to 40 seconds) [3]. This technique has the completely cured sample from the and 2 hours.
been used with great success in applica- second heating run (blue). The curve
tions involving glass transitions [3], solid- however shows no indication of a glass This data is also presented graphically
solid transitions [3], the aging behavior of transition. in Figure 3. It shows the relationship be-
polymers [4], the thermal degradation of tween the glass transition temperature
polystyrene [5] and the vitrification and The reason for this is that the network and the curing time. The vitrification
devitrification of thermosets (see User- density of the completely cured sample time of about 102 minutes at the curing
Coms 29 and 30 [6, 7], and [8–12]). is so high that it is not possible to detect temperature (here 150 °C) can be read
the glass transition by DSC. The curve off from the graph (see the black ar-
Procedure shows that degradation begins at about rows). Several measurements are there-
The investigation of the curing behavior of 260 °C. fore needed to determine the vitrification
TGAP was carried out in several steps: time by conventional DSC.
• Isothermal curing for 3 hours at 150 °C The vitrification time of the material at
• Rapid cooling 150 °C was determined by conventional Results from the TOPEM®
• First heating run (DSC: 50 to 300 °C DSC by performing a series of isothermal measurements
at 10 K/min; TOPEM® : 100 to 290 °C curing experiments. Identical samples TOPEM ® is a temperature-modulated
at 2 K/min) were first cured isothermally for differ- DSC technique in which the heat flow
• Second heating run after cooling the ent times and then cooled. The glass and the specific heat capacity of a sample
sample rapidly (only by conventional transition temperature of each precured can be simultaneously measured. The
DSC, same conditions as for the first sample was then measured in a series of method is therefore ideal for monitoring
heating run) separate heating runs. the vitrification and curing process of a
material [3].
Results of the conventional Vitrification occurs when the glass
DSC measurements transition temperature of the material Figure 4 displays the results for the iso-
Figure 1 displays the DSC curves ob- reaches and exceeds the curing tempera- thermal curing of a sample of TGAP/DDS
tained from the isothermal curing of a ture (in this case 150 °C). at 150 °C. The upper diagram shows the
TGAP sample at 150 °C (above) and the Figure 2.
first and second heating runs that fol- Heat flow curves for
lowed (below). The isothermal curing samples of TGAP/
DDS cured for differ-
curve (red) shows that at this tempera- ent times at 150 °C
ture the curing reaction is completed and then heated at
after about 160 minutes but there are no 10 K/min.
METTLER TOLEDO UserCom 37 13
modulated heat flow curve (blue) from low heating rate, curves such as those tinues while the sample is slowly heated.
which the total heat flow curve (red) is displayed in Figure 5 are obtained. The This leads to the exothermic heat flow
obtained. This curve corresponds to the upper diagram shows the total heat flow (red curve in the upper diagram of
heat flow curve of a conventional DSC curve and the lower diagram the specific Figure 5).
measurement. heat capacity curve, cp0.
Applications
14 METTLER TOLEDO UserCom 37
Conclusions Figure 4.
Isothermal mea-
If the isothermal curing behavior is in- surement of TGAP/
vestigated using conventional DSC, the DDS for three hours
measured isothermal DSC curves show at 150 °C using
TOPEM®. Above:
no indication of possible vitrification. the modulated heat
flow curve (blue)
Vitrification during isothermal curing scaled with a factor
of 0.15 and total
can only be identified by measuring the heat flow curve
glass transition of the cured material in (red). Below: the
a second dynamic measurement – in this specific heat capac-
ity curve (green).
case, a postcuring peak occurs immedi-
ately after the glass transition.
METTLER TOLEDO UserCom 37 15
An investigation using TOPEM®, a temperature-modulated DSC and ments. Macromol. Chem. Phys. 211
new temperature modulated tech- dielectric analysis. Macromol. Chem. (2010) 57–65.
nique. J. Thermal Anal. Calorim. 91 Phys. 209 (2008) 2003–2011. [12] Fraga, I.; Montserrat, S.; Hutchinson,
(2008) 687–695. [11] Fraga, I.; Montserrat, S.; Hutchinson, J.M.: Vitrification and devitrification
[10] Fraga, I.; Montserrat, S.; Hutchinson, J.M.: Vitrification and devitrification during the non-isothermal cure of
J.M.: Vitrification during the isother- during the non-isothermal cure of a a thermoset: A TOPEM study. J.
mal cure of thermosets: Compari- thermoset. Theoretical model and Thermal Anal. Calorim. 99 (2010)
Applications
The level of thermal comfort nowadays needed in buildings has led to an increase in energy consumption in
the residential and service sectors. Thermal energy storage (TES) is an important alternative for reducing
high energy consumption. Phase change materials (PCMs) have been studied as suitable materials for stor-
ing thermal energy due to their high heat storage capacity.
Introduction released again below the temperature of of the materials. For this reason, thermal
The authors of this article have devel- the phase change. analysis of PCMs is necessary at an early
oped and tested a phase change mate- stage of the design of building because it
rial (PCM) incorporated in a polymer Paraffin materials can be used as PCMs is a key point for the thermal character-
matrix. Unexpected difficulties however to climatize rooms due to their heat stor- ization of the building envelope.
arose in the measurement of the thermal age capacity of about 130 kJ/kg and their
properties of this material using differ- melting temperature of 15 to 30 °C [1]. In DSC
ential scanning calorimetry (DSC). The most countries, this temperature range Differential thermal analysis (DSC) is
thermal effects of the polymer matrix is accepted as being optimal for thermal one of the most powerful techniques [3]
and the PCM overlapped. comfort in buildings. for measuring thermal properties such
as the specific heat capacity (cp), and the
A new method was therefore developed Consequently, PCMs based on paraffin enthalpy and temperature of phase tran-
to overcome this problem. This consist- materials are integrated as composite sitions. This has been discussed by Cas-
ed of using a reference crucible contain- materials in thermal insulation systems. tellón et al. [4]. The main limitations of
ing the polymer matrix as a blank in- These materials increase the thermal in- the technique when used for this particu-
stead of the usual empty crucible. This ertia of the building envelope and com- lar application are the sample require-
enabled a clear signal of the PCM to be bined with insulation material reduce ments, namely small size, purity, and
obtained. energy consumption and temperature homogeneity [5]. These are important
fluctuations [2]. limitations because many PCM samples
PCM are mixtures of different components.
Phase change materials (PCMs) are The thermal behavior of composite ma-
ideal for storing thermal energy due to terials normally used in building enve- Paraffinic PCMs
their large heat storage capacity at the lopes is difficult to characterize due to The authors have developed a new mate-
phase change: The thermal energy is the large dimensions and heterogeneity rial for the building envelope [6]. The
16 METTLER TOLEDO UserCom 37
composite material is a shape-stabilized Table 1.
Components Characteristics Percentage weight Material compo-
paraffinic PCM containing 10 to 12 nents and composi-
Polymer matrix (PM) Ethylene/octene 20
weight percent PCM in a polymer ma- copolymer + EVA tion of the compos-
trix. To determine the concentration ite material studied.
Filler (EAFD) Inorganic residue from 70
of the PCM in the polymer matrix, the
electric arc furnace dust
material was analyzed by DSC using
the conventional measurement method Paraffin (PCM) RT-21: Tm = 20.14 °C; 10
∆Hm = 96.22 kJ/kg
with an empty crucible as the reference
sample.
Table 2.
Experiment Reference sample Experiments and
The main goal of this study was to design Run 1 Empty crucible runs performed in
and evaluate a new experimental DSC this study and the
Run 2 Empty crucible different reference
method to separate the PCM signal from samples.
that of the polymer matrix. More infor- Run 3 Polymer matrix
mation on this work has been published Run 4 Polymer matrix with EAFD
by Barreneche et al. [7].
the sample material. This improved the Furthermore, the values for the enthalpy
Experimental details PCM signal; the peaks were sharper and of fusion from Runs 1 and 2 listed in
better defined at the phase transition and Table 3 do not correspond to the esti-
Materials the matrix overlap in the DSC curve was mated values. A value between 9.6 and
The samples investigated were composite reduced. 11.5 kJ/kg is expected for the composite
materials consisting of a polymer matrix, material investigated if an enthalpy of
an inorganic filler and a PCM in different Measurements and resultats fusion of 96.22 kJ/kg and a melting tem-
formulations. Details of the samples are Table 2 summarizes the experiments perature of 20.14 °C is assumed for the
presented in Table 1. performed and the corresponding blank PCM investigated.
samples. Two measurements were first
The composite material was manu- performed with empty crucibles as refer- As can be seen in the DSC curves in
factured on a large scale. An internal ence samples as is usual in conventional Figure 1, the inaccurate values obtained
Banbury mixer was used to obtain ho- DSC analysis (Runs 1 and 2, repeat mea- for the enthalpy of fusion could have to
mogeneous material. The material was surements). do with the unclear peak shape leading
laminated at room temperature using a to incorrect integration values. Since the
two-roll mill. The heat flow curves are shown in results obtained were unsatisfactory, two
Figure 1. The curves were not clearly further experiments were performed in
Due to its complex composition, it was defined and indicated that the melting which non-conventional reference sam-
difficult to analyze and understand the process took place between 11 and 30 °C. ples were used. Figure 2 shows the DSC
thermal properties of the material using This is not in agreement with the expect- curves of the two samples measured us-
DSC. The expected enthalpy of fusion due ed behavior (between 18 and 24 °C). ing different reference samples, namely
to the PCM content was about 10 kJ/kg Figure 1.
but the PCM peak was overlapped by ther- DSC curves of two
mal effects of the matrix and filler. heating runs of the
composite material.
Empty crucibles
Methodology were used as ref-
Samples weighing about 15 mg were erence samples
(Run 1 and 2).
measured in 100 μL aluminum cru-
cibles in nitrogen using a flow rate of
80 mL/min. The DSC measurements
were performed in the temperature range
10 to 35 °C at heating and cooling rates
of 0.5 K/min [3].
METTLER TOLEDO UserCom 37 17
the polymer matrix in Run 3 and the The peaks of the phase change transition curve response signal. To eliminate the
polymer matrix with EAFD in Run 4. The are sharper and better defined and the overlap due to the matrix signal, two
DSC signal was clearly improved by the PCM measurement signal is improved. different blank reference samples were
use of these unconventional reference An optimization of the DSC curves was used; one was the pure polymer matrix,
samples. observed when the new reference samples the second the polymer matrix and the
Applications
Acknowledgements
This work was partially funded by
the Spanish government (ENE2011-
28269-C03-02), the European Union
(COST Action TU0802) and the Program
“PPV University de Barcelona-Santand-
er”. The authors would like to thank the
Catalan Government for the quality ac-
creditation given to their research group
GREA (2009 SGR 534) and research
group DIOPMA (2009 SGR 645).
Table 3.
Evaluation of the
Enthalpy of fusion Melting Type of References
enthalpy of fu- in kJ/kg temperature in °C reference used [1] Cabeza L.F, Castell A, Barreneche C,
sion and melting de Gracia A, Fernández A. I. Materials
temperature of the
Run 1 6.16 21.44 Conventional used as PCM in thermal energy sto-
composite using Run 2 5.85 21.49 Conventional rage in buildings: A review. Renewable
different types of & Sustainable Energy Reviews 2011;
reference samples. Run 3 10.03 20.92 Polymer matrix 15: 1675–1695.
[2] Cabeza L.F, Castell A, Medrano M,
Run 4 9.89 21.04 Polymer matrix/EAFD Martorell I, Pérez G, Fernández A.I. Ex-
18 METTLER TOLEDO UserCom 37
perimental study on the performance national Journal of Energy Research ración estabilizante en un material de
of insulation materials in Mediterrane- 2008; 32(13): 1258–1265. construcción]. P201131272. 2011.
an construction. Energy & Buildings [5] Lázaro A, Günther E, Mehling H, Hieb- [7] Barreneche C., Solé A., Miró M.,
2010; 42(5): 630–636. ler M, Marín M.M, Zalba B. Verification Martorell I, Fernández A.I., Cabeza
[3] Mehling H, Cabeza L. F. Heat and cold of a T-history installation to measure L.F. New methodology developed for
storage with PCM. An up to date intro- enthalpy versus temperature curves the differential scanning calorimetry
duction into basics and applications. of phase change materials. Measure- analysis of polymeric matrixes in-
Springer-Verlag Berlin Heidelberg; ment Science & Technology 2006; 17: corporating phase change materials.
Berlin (Germany), 2008. 2168–2174. Measurement Science and Technolo-
ISBN: 978-3-540-68556-2. [6] Fernández A.I, Cabeza L.F, Barreneche gy. 2012; 23: 085606, 5 pp.
[4] Castellón C, Günther E, Mehling H, C, Chimenos J.M,Espiell F, Segarra [8] Höhne G. W. H, Hemminger W.F,
Hiebler S, Cabeza L.F. Determination M, Solé C. Waste removal of EAFD by Flammersheim H.J. Differential scan-
of the enthalpy of PCM as a function means of its stabilization integrating it ning calorimetry, Springer–Verlag
of temperature using a heat-flux DSC into a building material [Inertización Berlin Heidelberg. Berlin (Germany),
– A study of different measurement de polvo de acería mediante su integ- 2003. ISBN: 3-540-00467-x.
procedures and their accuracy. Inter-
Thermoanalytical characterization
of oxo-biodegradable polymers
Dr. Melanie Nijman
One of the major problems of the plastics we use today is their great stability – they impact the environment
for years without undergoing decomposition. Recently, a number of research projects have been started to
develop plastics that degrade within a short time in the environment.
Introduction lysts and promote oxidization of the oxo-biodegradable packaging film was
Up until now, two different approaches polymer chains. Since oxidation can only determined from its melting curve. Pure
have been followed for making biologi- take place when oxygen is present, this PE-LD in the form of pellets was mea-
cally degradable polymers. type of decomposition only works when sured for comparison.
the plastic is exposed to an open-air en-
The first approach is to produce plastics vironment. If the plastic is disposed of in Small pieces were cut out from the film
from biomaterials such as maize (corn) a landfill under anaerobic conditions, and placed in the DSC or TGA crucibles. A
or wheat, which rapidly decompose to decomposition is no faster that with or- PE-LD pellet was cut into thin slices. The
smaller, biologically degradable com- dinary plastics. sample mass used for the DSC measure-
pounds after use. ments was approximately 3 mg. Appre-
The metal salts added are only present ciably more (approx. 30 mg) was used for
The other approach is to modify com- in catalytic amounts and do not have a the TGA measurements because only very
monly used polymers such as polyeth- negative effect on the environment. They small residues were expected.
ylene so that they degrade more quickly are salts of iron, magnesium, manga-
and can then be digested by microorgan- nese, zinc or nickel, which are in any DSC experiments were performed on both
isms. This article deals with this second case present in nature, for example in materials to measure the OOT (Oxidation
category of polymers, also referred to as metalloproteins. Onset Temperature). The onset of oxida-
oxo-biodegradable polymers. tion is determined by the stability of the
Experimental details polymers. If the stability of a polymer is
The degradation of such polymers is The polymers were investigated by per- artificially decreased, as in an oxo-bio-
accelerated by the addition of small forming DSC and TGA/DSC measure- degradable polymer, one expects a lower
amounts of metal salts that act as cata- ments. The polymeric composition of an oxidation temperature.
METTLER TOLEDO UserCom 37 19
Results and discussion OOT (Oxidation Onset Temperature) yielded an oxidation onset temperature
measurements were then performed to of 201.3 °C for the oxo-biodegradable
DSC measurements compare the stability of the two materi- film and 214.7 °C for the PE-LD pellet.
Figure 1 shows the DSC melting curves of als. The curves obtained are displayed The temperature difference is significant
the two materials. Both samples have very in Figure 2. Care was taken to make sure and very reproducible. The onset of oxi-
Applications
similar melting temperatures. The oxo- that the samples had about the same dation of the oxo-biodegradable PE film
biodegradable film is however a blend of mass and that they rested flat on the is in fact about 13 K lower than that of
PE-LD and PE-LLD. It exhibits two clear bottom of the crucible. the pure PE-LD pellet and indicates that
melting peaks at 112 °C and 121 °C. The it is much less stable.
mean degree of crystallinity is compa- The measurements were performed in
rable with the values usual for PE-(L)LD. oxygen (50 mL/min) and with no lid on TGA measurements
the crucible. The same heating rate of TGA measurements were also performed
The curves shown in Figure 1 are second 10 K/min was used as before. The mea- in addition to the DSC measurements.
heating runs of the individual samples surements were automatically stopped The resulting curves for the two mate-
in order to eliminate the effects of pro- when an exothermic heat flow of 5 W/g rials are displayed in Figure 3. A large
duction and storage history on the DSC was reached in order to protect the DSC difference can be seen between the de-
curves. The heating rate was 10 K/min furnace from decomposition products. composition of the two materials. The
with a gas flow of 50 mL/min nitrogen. Evaluation of the OOT measurements first part of the curves was measured in
nitrogen up to 600 °C.
Figure 1.
DSC measurements
(second heating The pure PE-LD sample is completely py-
runs) of an oxo-bio- rolyzed in nitrogen, with a characteristic
degradable (OBD)
film and a pure PE-
onset at 457 °C (extrapolated onset). At
LD pellet. 600 °C, the purge gas was switched from
nitrogen to air: oxidation of polymer
residues began immediately. All the TGA
measurements were performed at a heat-
ing rate of 20 K/min.
20 METTLER TOLEDO UserCom 37
a lower temperature than for the film. Table 1.
Onset Pyrolysis Oxidation Residue
Overview of the TGA
However, close examination of the simul- in °C in % in % in % weight losses of the
taneously measured DSC curves shows Oxo-biodegradable film, 471.4 91.73 0.60 7.67 oxo-biodegradable
that oxidation of the oxo-biodegradable sample 1 film and the PE-LD
pellet.
film begins earlier. Oxo-biodegradable film, 472.2 91.73 0.61 7.67
sample 2
The DSC measurements are much more PE-LD pellet, sample 1 457.6 99.99 – 0.01
sensitive to the onset of oxidation than the PE-LD pellet, sample 2 457.7 100.0 – –
TGA measurements. The rate of decompo-
sition in air is greater for PE-LD compared Figure 3.
with for the oxo-biodegradable film; in ni- TGA/DSC curves
of an oxo-biode-
trogen, the rates are about the same.
gradable film and a
PE-LD pellet mea-
The measurements in air also show that sured in nitrogen
to 600 °C and then
the oxo-biodegradable film leaves a resi- in air to 900 °C, at
due whereas the PE-LD pellets burn com- 50 mL/min.
pletely. In addition, the oxo-biodegradable
film still exhibits the endothermic effect
at about 600 °C in air. This effect is also
visible in the measurements in Figure 3.
References
[1] http://en.wikipedia.org/wiki/Oxo_Bio-
degradable
[2] Assessing the Environmental Impacts
of Oxo-degradable Plastics Across
Their Life Cycle, Loughborough
University (document EV0422) /
Department for Environment, Food
and Rural Affairs (DEFRA), UK.
METTLER TOLEDO UserCom 37 21
Exhibitions, Conferences,
Seminars and Training Courses
Dates
Webinars: For information on our current program of webinars, please visit www.mt.com/ta-webinars
Handbooks: Our specialist application handbooks provide valuable www.mt.com/ta-handbooks
background information
Applications: The applications database includes hundreds of interesting examples www.mt.com/ta-applications
Training courses: Improve your knowledge of thermal analysis
– have a look at our extensive training program www.mt.com/ta-training
or
– register for an online e-Training course www.mt.com/ta-etraining
TA News: You can find the latest product news and an overview of our www.mt.com/ta-news
thermal analysis services at
22 METTLER TOLEDO UserCom 37
Local TA Customer Courses:
TA-Kundenkurse Nähere Informationen unter www.mt.com/labtalk oder durch:
und Seminare Frau Janine Schindler, Mettler-Toledo GmbH, Giessen, Tel: ++49 641 507 405, e-mail: labtalk@mt.com
in Deutschland DSC-Workshop, SETA1302 17. und 18.09.2013 Giessen
und der Schweiz TGA-Workshop, SETA1304 08. und 09.10.2013 Giessen
Online-Seminare zu div. Theme div. Termine www.mt.com/ta-webinars
Cours et Renseignements et inscriptions par Christine Fauvarque, Mettler-Toledo S.A., Tél: ++33 1 3097 1439, e-mail: christine.fauvarque@mt.com
séminaires Congrès National du GFP 18–21 Novembre 2013 Lille (France)
d’Analyse Principe de la TMA/DMA 25 novembre 2013 Viroflay (France)
Thermique en
DSC: notions de base 26 novembre 2012 Viroflay (France)
France
DSC: perfectionnement 27 novembre 2013 Viroflay (France)
Principe de la TGA 28 novembre 2013 Viroflay (France)
Logiciel STARe: perfectionnement 29 novembre 2013 Viroflay (France)
Cursos y Para detalles de los cursos y seminarios, por favor contacte con:
Semianros de TA Francesc Català Mettler-Toledo S.A.E., Tel: ++34 93 223 76 00, e-mail: francesc.catala@mt.com
en España Aplicaciones del Análisis Térmico 18 de septiembre 2013 Vigo
Formación Usuarios TA 9 de octubre 2013 Barcelona
Aplicaciones del Análisis Térmico 23 de octubre 2013 Sevilla
Seminários e Para maiores informações sobre os seminários, favor entrar em contato com: Mr. Pierre Delmorio, Tel: +55 11 4166 7936, e-mail: pierre.delmorio@mt.com
workshops de TA Ms. Kíria Serranegra de Arruda Tel +55 11 4166 7435, e-mail: kiria.arruda@mt.com
no Brasil Aplicações Farmacêuticas 31.07.2013 Mettler Toledo - Avenida Tamboré, 418 – Alphaville - Barueri - SP
Bases de Análise Térmica 09.10.2013 Mettler Toledo - Avenida Tamboré, 418 – Alphaville - Barueri - SP
TA Courses and For details of training courses and seminars, please contact:
Seminars in Mahesh Tripathi, Mettler-Toledo India Private Limited, Mumbai 400 072, Tel: +91-22-28031 111 / 28031 370, e-mail: mahesh.tripathi@mt.com
India DSC and TGA, FP July 2013 Mumbai, India
TA Courses and For details of training courses and seminars, please contact:
Seminars Chatriya Suamsung, Mettler-Toledo (Thailand) Limited, Tel: +66-2-7230338, Fax: +66-2-7196479, e-mail: Chatriya.Suamsung@mt.com
Thailand DSC Existing Customer Training August 2013 Bangkok
Thermal Analysis for Pharmaceutical Application August 2013 Bangkok
and Validation in Thermal Analysis, Seminar
TA Courses and For details of training courses, workshop and seminars, please contact: SangKoo Lee or Kyungho Chung at Mettler-Toledo Korea, Tel: ++82 2 3498 3500,
Seminars in e-mail: sangkoo.lee@mt.com, Kyungho.chung@mt.com, Website: www.mt.com
Korea TA Academy in Q3 2013 July 2013 Seoul
TMA School in Q3 2013 July 2013 Seoul
TA Academy for the Province in Q3 2013 September 2013 DaeJeon
TA Academy in Q4 2013 October 2013 Seoul
DMA School in Q4 2013 October 2013 Seoul
Seminar
Public Seminar for Chemical Industry in Q3 2013 September 2013 Ulsan
TA Courses and For details of training courses and seminars, please contact: Lu LiMing at Mettler-Toledo Instruments (Shanghai) Co. Ltd.,
Seminars in Tel: ++86 21 6485 0435, Fax: ++86 21 6485 3351, e-mail: liming.lu@mt.com
China TA Infoday in Q3 2013 September 2013 Shanghai
TA Infoday in Q4 2013 October, November and December 2013 Guangzhou, Xian, Dalian, Nanjing, Qingdao, etc.
Advanced TA Technique Conference November 2013 Nanjing
TA User Meeting 2013 July 2013 Zhuhai
TA Training Courses in Q3 2013 End of June 2013 Shanghai, Chengdu
TA Training Courses in Q4 2013 End of December 2013 Beijing, Guangzhou
METTLER TOLEDO UserCom 37 23
Local TA Customer Courses:
Australia & New Zealand: Nadine Winkelmann, Tel: ++ 61 438 093 25, e-mail: nadine.winkelmann@mt.com
Ceská Republika: Helena Beránková, Tel: +420 226 808 152, e-mail: helena.berankova@mt.com
Finland: Oy G. W. Berg & Co Ab, Tel: +358 201 255 255, e-mail: gwb@gwb.fi
Latin America: Francesc Català, Tel: ++34-932 237 615 (Spain), e-mail: francesc.catala@mt.com
Malaysia: Tel: (+603)-78445888, e-mail: mtmymarketing@mt.com
Österreich: Geraldine Braun, Tel: +43-1-604 19 80 - 33, e-mail: geraldine.braun@mt.com
Polska: Piotr Witukiewicz lub Grazyna Czaplicka, Tel: +48 22 545 06 80,
e-mail: piotr.witukiewicz@mt.com, grazyna.czaplicka@mt.com, Strona: www.mt.com
Россия: Алексей Важинский, тел: +7 495 651 98 86, e-mail: alexey.vazhinskiy@mt.com
Singapore: Gina Chan, Contact no: +65 68900012/13, e-mail: Gina.Chan@mt.com
Sweden: Per Torenstam, Tel: ++46 8 702 5021, e-mail: per.torenstam@mt.com
Slovenia: Keith Racman, Tel: +386 1 547 4900, e-mail: keith.racman@mt.com
Taiwan: Karen Chen, Tel: +886-2-26578898 p102; e-mail: karen.chen@mt.com
UK: Lester Troughton, Tel: ++44 (0)7831 490 132, e-mail: lester.troughton@mt.com
USA & Canada: Michael Zemo, Tel: +1 614-438-4686, e-mail: michael.zemo@mt.com
For further information regarding meetings, products or applications, please contact your local
METTLER TOLEDO representative and visit our Website www.mt.com
Editorial Team
Dr. A. Hammer Dr. S. Giani Dr. D. P. May Dr. R. Riesen Dr. J. Schawe Dr. S. Domann
Chemist Chemist Chemist Chem. Engineer Physicist Journalist & Chemist
N. Fedelich Dr. M. Nijman Dr. E. Hempel Dr. M. Wagner Dr. M. Schubnell U. Jörimann
Chem. Engineer Chemist Physicist Chemist Physicist Electr. Engineer