ULTRA MAG™
ELECTROMAGNETIC FLOW METER
SUBMITTAL PACKAGE

From

McCROMETER
3255 WEST STETSON AVENUE
HEMET, CA 92545

Phone (951) 652-6811


Toll Free (800) 220-2279
Fax (951) 652-3078

www.mccrometer.com

Date:

Page i 30107-10 Rev. 2.7 | 02NOV2022


 Project Name:

Purchase Order No.:


Date:
Customer Name:

Submitted By:

Equipment Being
Purchased:

Included Items: Ultra Mag sensor (UM06 or UM08 as noted above)


UltraLiner™ technology with NSF fusion bonded epoxy coating
Stainless Steel Electrodes
2-Year Warranty
Selectable Items: Electrodes Material:
Flanges:
Converter (select one):
ProComm Meter Mount Converter Converter Power: AC DC
ProComm Remote Mount Converter
Converter Power: AC DC
with _____ feet of cable
ProComm GO converter Battery power
Optional Output: HART
Modbus RS-485
ETP
AMR
Built-In Verification
Data Logger

When completing this submittal package, it must also include the Ultra Mag Specification Data Sheet, which is also an
electronic document. A copy can be downloaded from the McCrometer web site with the following link.
Lit. No. 30124-74
The Specification Data Sheet is a macro-enabled Excel spreadsheet that contains multiple drop down fields for multiple
product configurations. After it is filled out, save the spreadsheet under a different name and submit the file with the
submittal package.
See the following specification sheet and drawings for sizing and option details.

Page 1 30107-10 Rev. 2.7 | 08NOV2022



Contents
Ultra Mag Flow Meter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Ultra Mag Flow Meter Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sensor Grounding and Electrical Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
ProComm Converter Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
ProComm Converter Cable Connection from Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
ProComm Converter Power Hook-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4-20mA Hook-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Opto-Isolated Pulse Output and Opto-Isolated Input Hook-Up . . . . . . . . . . . . . . . . . . . . . 12
ProComm GO Converter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
ProComm GO Converter Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
ProComm GO Converter Cable Connections from Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
ProComm GO Converter Battery Installation and Connection . . . . . . . . . . . . . . . . . . . . . . . 18

Page 2 30107-10 Rev. 2.7 | 08NOV2022


Specification Sheet
Ultra Mag Flow Meter
FLOW METER
Ultra Mag FlowSPECIFICATIONS
Meter Specifications
Pipe Sizes
2”, 3”, 4”, 6”, 8”, 10”, 12”, 14”, 16”, 18”, 20”, 24”, 30”, 36”, 42”, 48”

Flow Direction Measurement


Forward and reverse flow indication and forward, reverse, net totalization are standard with
all meters
Accuracy
• Standard: +/- 0.5% of measured value ±0.006 ft/s (±0.0018 m/s)
• Optional: +/- 0.2% of measured value ±0.006 ft/s (±0.0018 m/s)
• Battery powered: 1% of measured value ±0.006 ft/s (±0.0018 m/s)
IMPORTANT NOTICE ON FLOW METER ACCURACY: The flow meter, the cable and the
electronics are factory calibrated for accuracy as a single unit. Changing the cable length
with the Splice Kit changes the accuracy of the meter and invalidates the calibration
certificate.
Accuracy Tests
Multiple point wet flow calibration of every complete flow tube with its signal converter.
If desired, the tests can be witnessed by the customer. The McCrometer test facilities are
traceable to the National Institute of Standards & Technology. Uncertainty relative to flow is
±0.15%.
Pipe Run Requirements
2” & 3” wafer style 3D upstream / 1D downstream
4” and larger flanged 1D upstream / 0D downstream

Repeatability
±0.05% or ±.0008ft/s (±0.25mm/s), whichever is greater

Conductivity
5 µs/cm

Liner
UltraLiner NSF approved, fusion bonded epoxy

Electrodes
Type 316 stainless steel, others optional

Electrical Connections
• Compression gland seals
• Quick-Connect

Sensor Cable Lengths


Standard 25’/7.6 m McCrometer supplied submersible cable with each remote mount unit.
Optional Up to 500’/152.4 m, or 25’/7.6 m max for battery powered.
Quick Connect Available in standard cable lengths:
Feet: 25, 50, 75, 100, 125, 150, 175, 200, 500
Meters: 7.6, 15.25. 22.5, 30.5, 38.1, 45.75, 53.3, 61, 152.4
Custom cable lengths at additional cost.

Page 7 30107-09 Rev. 5.9 | 26OCT2022


Page 3 30107-10 Rev. 2.7 | 08NOV2022
Specification Sheet
Ultra Mag Flow Meter
Ultra Mag
FLOW FlowSPECIFICATIONS
METER Meter Specifications
(CONT.)
IP Rating
Standard model • Quick Connect (NEMA 6P/IP68 with remote converter)
• Compression gland seals (NEMA 6P/IP68 with remote converter)
HL model • Quick Connect (IP67)
• Compression gland seals (IP67)
Sensor Submersibility Depth
With standard strain
9 m (30 ft.)
relief cable
With optional quick
1.8 m (6 ft.)
connect cable
Head Loss
None. No obstruction in line and no moving parts

Warranty
Meter 2 year warranty
Liner Lifetime guarantee

Pressure Range
AWWA Class D (150 psi Rating) (Standard)
ANSI Class 150# (285 psi Rating)
ANSI Class 300# (300 psi Rating)
AWWA Class F (300 psi Rating)

Velocity Range
.2 to 32 FPS

Temperature Range
Sensor Operating: -10 to 60°C (14 to 140°F)
Sensor Storage: -15 to 60°C (5 to 140° F)
Certifications and Approvals
Standard Model • ISO 9001:2015 certified quality management system
• Certified by MET to UL 61010-1 / CSA C22.2 No. 61010-1
• Certified to NSF / ANSI Standards*
HL Model • ISO 9001:2015 certified quality management system COMPONENT

• Certified by MET: Safety: UL61010-1 / CSA C22.2 No. IA


PM
O R
61010-1, Third Edition: Safety of Electrical Equipment For & T

Measurement, Control, and Laboratory Use NSF/ANSI/CAN 61®


• Certified by MET: Standards: ANSI / ISA12.12.01 / CSA C22.2 &
NSF/ANSI 372

No. 213, Nonincendive Electrical Equipment


• Class I and II, Division 2
• Class III, Divisions 1 and 2 Hazardous (Classified) Locations
• Certified to NSF / ANSI Standards*
System Options
• Hastelloy® electrodes
• Additional sensor cable up to 475’
• Annual verification / calibration
• Stainless steel ID tag

* Ultra Mag is certified by IAPMO R&T to NSF/ANSI 61 for material safety and NSF/ANSI 372 for low lead content.

Page 8 30107-09 Rev. 5.9 | 26OCT2022


Page 4 30107-10 Rev. 2.7 | 08NOV2022
be used to establish a connection to earth ground.

Plastic or internally coated pipe - Grounding rings can be installed to establish a connection to earth ground
See the Ultra Mag IOM Manual, Lit. # 30119-03, for more information on grounding configurations using grounding rods
and grounding rings.

Ultra Mag Flow


DIMENSIONS Meter
AND Dimensions
WEIGHTS
2" and 3" Models Body Style
Flow Ranges DIMENSIONS Est. Shipping
GPM (Lay Lengths) Weight (lbs.)**
Meter Pipe Size Meter
Standard A B C D E
Type (Nominal) Pipe ID
.2 to 32 FPS
Min - Max UM06* UM08* UM06* UM08* UM06* UM08*
Use model shown below for dimensions
Wafer 2" 1.625 2 - 310 4.5 4.5 4.0 6.5 7.25 n/a n/a 9.6 10.1
style 3" 2.625 5 - 700 4.5 4.5 4.0 7.0 7.75 n/a n/a 11.3 11.8
| * Note: UM06 relates to AWWA Class D; UM08 relates ASNI #150, #300, AWWA Class F
** For remote mount meters, add 4 lbs for ProComm converter.

A
Grounding rings are 0.0625" Specification Sheet
(1.5875 mm) thick. Ultra Mag Flow Meter
DIMENSIONS AND WEIGHTS (CONT.)
4" to 12" Models Body Style
Flow Ranges GPM DIMENSIONS Est.Shipping
Meter
Pipe Size Standard (Lay Lengths) Weight (lbs.)**
Pipe ID
(Nominal) .2 to 32 FPS A B C
UM06* UM08* Min - Max UM06* UM08* UM06* UM08* UM06* UM08*
4" 3.834 3.76 8 - 1,140 13.40 13.40 9.00 10.00 7.28 78 108
6" 5.782 5.732 19 - 2,660 14.60 14.60 11.00 12.50 8.25 82 138
8" 7.782 7.732 33 - 4,870 16.10 17.25 13.50 15.00 9.25 115 195
10" 9.782 9.732 52 - 7,670 18.50 18.50 16.00 17.50 10.5 144 247
Page 10 30107-09 Rev. 5.9 | 26OCT2022
12" 11.782 11.732 74 - 11,180 19.70 19.70 19.00 20.50 11.5 193 342
| * Note: UM06 relates to AWWA Class D; UM08 relates ASNI #150, #300, AWWA Class F
** For remote mount meters, add 4 lbs for ProComm converter.

A Grounding rings are 0.0625" thick.


14+" Models Body Style
Flow Ranges GPM Page 5
DIMENSIONS Est. Shipping
30107-10 Rev. 2.7 | 08NOV2022
Pipe Size Meter Standard (Lay Lengths) Weight (lbs.)**
(Nominal) Pipe ID .2 to 32 FPS A B C D E
B

Ultra Mag Flow Meter Dimensions


A Grounding rings are 0.0625" thick.
14+" Models Body Style
Flow Ranges GPM DIMENSIONS Est. Shipping
Pipe Size Meter Standard (Lay Lengths) Weight (lbs.)**
(Nominal) Pipe ID .2 to 32 FPS A B C D E
Min - Max UM06* UM08* UM06* UM08* UM06* UM08*
14" 13.63 90 - 16,070 21.70 22.75 11.875 21.00 23.00 20.135 14.56 321 476
16" 15.50 118 - 20,900 23.60 25.25 14.25 23.50 25.50 21.635 15.32 390 645
18" 17.50 150 - 26,480 23.60 25.25 14.25 25.00 28.00 23.635 16.32 446 750
20" 19.50 185 - 32,720 25.60 28.25 16.06 27.50 30.50 25.6975 17.35 588 874
24" 23.50 270 - 47,180 30.70 35.75 21.75 32.00 36.00 29.51 19.25 769 1,568
30" 29.25 420 - 73,620 35.80 41.75 25.25 38.75 43.00 35.6975 22.35 1,261 2,317
36" 35.25 610 - 105,930 46.10 46.10 28.63 46.00 50.00 42.76 25.88 1,696 2,915
42" 41.25 830 - 144,370 48.05 *** 36.25 52.75 *** 48.135 28.57 *** ***
48" 47.25 1,080 - 188,430 50.00 *** 36.25 59.50 *** 54.135 31.57 *** ***
| * Note: UM06 relates to AWWA Class D; UM08 relates ASNI #150, #300, AWWA Class F
** For remote mount meters, add 4 lbs for ProComm converter.
*** Consult factory

C D

B
A **Grounding Rings are 0.125" thick.

Page 11 30107-09 Rev. 5.9 | 26OCT2022

Page 6 30107-10 Rev. 2.7 | 08NOV2022


Sensor Grounding and Electrical Interference
Electrical Grounding
The sensor body must have electrical contact with the media. This is achieved via a grounding ring or grounding button.
NOTE: The grounding ring is optional only on 4” through 12” models. For best performance, grounding rings are
recommended for all sizes.
Always ensure that the converter and the sensor are grounded (earthed) correctly. The grounding of the sensor and
converter ensures that the equipment and liquid have an equal potential. For most installations the quality of grounding
by the provided cabling assures the sensor is properly grounded and additional grounding of the sensor is not required.
However, in instances where this is not the case, i.e. the equipment and fluid do not have an equal potential, such as where
the installation location and/or media is subjected to electrical interference, additional grounding steps may be required.
Consult an electrician experienced with instrumentation installations to determine if electrical interference is present. For
further information on installation environments and sensor grounding, please contact McCrometer Technical Support.

Lines With Cathodic Protection


On meters installed on a line with cathodic protection it may be necessary to insulate the meter from the line. Consult your
cathodic protection vendor for instructions.
Specification Sheet
Ultra Mag Flow Meter
METER GROUNDING RECOMMENDATIONS
Grounding the meter body for safety according to national
(NEC) or local electrical codes is recommended on ALL meter
installations.

For best performance, grounding the fluid column is


recommended when the meter is installed in an electrically
noisy environment, such as with VFD pumps or nearby electrical
systems with insufficient grounding.

Conductive or uncoated pipe - The uncoated pipe flange can


be used to establish a connection to earth ground.

Plastic or internally coated pipe - Grounding rings can be installed to establish a connection to earth ground
See the Ultra Mag IOM Manual, Lit. # 30119-03, for more information on grounding configurations using grounding rods
and grounding rings.

DIMENSIONS AND WEIGHTS


2" and 3" Models Body Style
Flow Ranges DIMENSIONS Est. Shipping
GPM (Lay Lengths) Weight (lbs.)**
Meter Pipe Size Meter
Standard A B C D E
Type (Nominal) Pipe ID
.2 to 32 FPS
Min - Max UM06* UM08* UM06* UM08* UM06* UM08*
Use model shown below for dimensions
Wafer 2" 1.625 2 - 310 4.5 4.5 4.0 6.5 7.25 n/a n/a 9.6 10.1
style 3" 2.625 5 - 700 4.5 4.5 4.0 7.0 7.75 n/a n/a 11.3 11.8
| * Note: UM06 relates to AWWA Class D; UM08 relates ASNI #150, #300, AWWA Class F
** For remote mount meters, add 4 lbs for ProComm converter.

Page 7 30107-10 Rev. 2.7 | 08NOV2022

C
Specification Sheet
ProComm Ultra Mag Flow Meter
DIMENSIONS
PROCOMM CONVERTER
10.0 DIMENSIONS
ProComm Converter METER MOUNT DIMENSIONS
Dimensions
Meter Mount Converter Dimensions Height 6.9” (20.1 cm)
Width7.2” (18.3 cm)
Depth 6.2” (15.7 cm)

6.2"

6.9"

3" 4.2" 3.1" 3.1"


7.2" 6.2"

140º
10.5"

5.6" 3.1"

Page 44 30124-60 Rev. 1.7 | 26OCT2022

Page 18 30107-09 Rev. 5.9 | 26OCT2022


Page 8 30107-10 Rev. 2.7 | 08NOV2022
ProComm Converter
ProComm Converter Cable
Wiring Connection from Sensor
Connection

All connections are made on the terminal board. To access the terminal board, loosen the four screws
on the front of the converter and lift the converter open.
All electrical cables enter the converter
through compression fittings located
on the side of the converter. Ensure that
all compression glands are properly
tightened and all unused fittings are
plugged so the case remains sealed.

Power
Coils and
ground STANDARD OUTPUTS
(AVAILABLE ON ALL UNITS)
DIGITAL INPUTS & OUTPUTS 4-20 OUTPUTS
Outputs
Electrodes
OUT2 C SH OUT2 (+) 4-20 SH
OUT1 C IN + OUT1 (+) 4-20 +24V
OUT 1-2 E IN - OUT 1-2 (-) 4-20 0V

Terminal Cable Wire Color OPTIONAL OUTPUTS


(SPECIFY AT TIME OF ORDER)
#1 (E1) A Blue MODBUS OUTPUT HART OUTPUT AMI OUTPUT
#2 (E2) A Pink
#3 (C) A Black
#4 (SH) A Green/Yellow
B SH OUT1(+) DATA SH
Chassis Lug B Black A OUT1(-) CLOCK

#11 (SH) B Green/Yellow


CHASSIS ELECTRODES COILS POWER OPTIONS
#12 (C2) B Yellow GROUND
CAUTION: THIS IS NOT A DUAL POWER UNIT. AC POWER
#13 (C1) B Red SHALL NOT BE CONNECTED TO DC UNIT OR VICE VERSA.

AC 100-240V DC 12-48V
N L - +
BLACK JUMPER
GREEN / YELLOW *

GREEN / YELLOW
YELLOW
BLUE
PINK

RED

* NOT APPLICABLE GREEN CONNECTOR RED CONNECTOR


TO SPI SENSOR AC POWER ONLY DC POWER ONLY

Copyright © 2020 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer. Any published pricing, technical
data, and instructions are subject to change without notice. Contact your McCrometer representative for current pricing, technical data, and instructions.

3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA


TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078
www.mccrometer.com
9
Page 6 30107-10
30120-71 Rev. 2.7
1.9 | 08NOV2022
13AUG2020
3.7 Converter Power Wiring Diagram

WARNING!
ProComm Converter
! Hazardous supply voltagePower
can shock,Hook-Up
burn, or cause death.

The power supply line must be equipped with external surge protection for current overload (fuse or circuit breaker with
limiting capacity not greater than 10A). It must be easily accessible for the operator and clearly identified.
Power connection is made using the power terminal block on the upper right side of the terminal board.
NOTE: The terminal block unplugs from the circuit board for easy connection. Connect earth ground to the protective
grounding terminal before making other connections. The power supply of a standard converter is 100-240VAC, 45-66Hz
at maximum 20W. DC converter is available as an option.

Neutral Wire Power Supply


(typically a (typically a
white wire) white wire)
Ground Wire Ground Wire
(green or yellow/ Line Wire (typically a Power Supply +
green wire) (typically a black wire) green wire) (typically a red wire)

Figure 15. AC Power Figure 16. Optional DC Power


Supply Terminal Block Supply Terminal Block

Page 11 30124-60 Rev. 1.7 | 26OCT2022

Page 10 30107-10 Rev. 2.7 | 08NOV2022


ProComm
4-20mA Hook-Up
CONNECTING WIRES TO TERMINALS
3.3 4-20mA Hook-Up
Isolated 4-20mA current loops are used to output flow data to external devices. Maximum load impedance is 1,000Ω, and
the maximum voltage without load is 27VDC. The converter has the capability to detect a loss of load on this output. To
disable this function set the value “mA Val. Fault” under the ALARMS menu to zero (see section 5.4). A graphical example of
the usage of the current loop with external device is shown below:

+ Terminal 24 FORWARD FLOW


OUT1 4-20mA
-

4-20mA Devices

+ Terminal 25 REVERSE FLOW


OUT2 4-20mA
-

Terminal 23 COMMON
OUT 1 / OUT2 4-20mA

Figure 12. 4-20mA Hook-Up

If the external device requires a voltage input, a precision resistor placed across the input terminals of the external device
will change the current to voltage. Calculate the required resistor using Ohm’s law (V = I x R). For example, a 250Ω resistor
will provide an input voltage of one to five volts with the transmitter range being set from 4mA to 20mA. An additional 4
to 20mA loop output is available.

I
IMPORTANT
The converter powers the 4-20mA loops. Do not use external power for the 4-20mA loop as it may cause
permanent damage to the converter.

Page 11 30107-10 Rev. 2.7 | 08NOV2022

Page 9 30124-60 Rev. 1.7 | 26OCT2022


ProCommPulse Output and Opto-Isolated
Opto-Isolated Input Hook-Up
CONNECTING WIRES TO TERMINALS
Opto-Isolated Pulse Output
3.4 Opto-Isolated Pulse Hook-Up
Output Hook-Up
The outputs are open collector transistor outputs used to communicate with or activate external devices.
• Opto-isolated output with collector and emitter terminals floating and freely connectable
• Maximum switching voltage: 40 VDC
• Maximum switching current: 100mA
• Maximum saturation voltage between collector and emitter 1.2V@100mA
• Maximum switching frequency (load on the collector or emitter, RL=470Ω, VOUT=24VDC): 1250Hz
• Maximum reverse current bearable on the input during an accidental polarity reversion (VEC): 100mA
• Isolation from other secondary circuits: 500 V

I
IMPORTANT
OPTO
Digital outputs are not isolated ISOLATED
from OUTPUTS
each other. All digital outputs MUST use the same power source.

COMMON
EMITTER
17

OUT 1
18

OUT 2
19

Figure 13. Opto-Isolated Pulse Output Diagram

Opto-Isolated Input Input


3.5 Opto-Isolated
• Opto-isolated input
• 500 V isolation
• 2-40 VDC on voltage
• Input programming per input menu, will perform functions set to ON.

Input example: Push button


switch
10K +
15 24V

-
16

Figure 14. Opto-Isolated Input Diagram

Page 12 30107-10 Rev. 2.7 | 08NOV2022


ProComm SPECIFICATIONS
ProComm GO Converter Specifications
9.0 SPECIFICATIONS
Physical Specifications
Electronic Housing Diecast aluminum, powder coated enclosure w/ tamper resistant seal, 6½” x 6½” x 43/8” tall
Converter
See "Dimensions" section for meter mount and remote mount converter dimensions.
Dimensions
Battery: Standard: three 3.6V lithium-thionyl chloride (Li-SOCI2) D size
batteries with two AA backup batteries
Power
AC Power: 100-240VAC/45-66Hz (4W)
DC Power: Linear power supply 10-35VDC (4 W)
Electrical • Optional shielded cable for 10-32VDC/4-20 mA output
Connections • Optional shielded cable for pulse out
Performance and Operational Specifications
Battery Life Five-year expected battery life, five-year battery warranty
Location Indoor or outdoor use
Operating: 2000 meters
Altitude
Storage: 12,000 meters
Operating
-4° to 140° F (-20° to 60° C)
Temperature
Storage Temperature -4° to 140° F (-20° to 60° C)
Relative Humidity 0% to 100%
IP Rating IP67 Die cast aluminum converter
Digital output: Digital pulse (open collector) output for volumetric
- Two isolated digital pulse (open collector) outputs for volumetric
- AMI output
Outputs
Analog output: 4-20mA: Galvanically Isolated, 16 Bit resolution. All power
configurations (including battery).
Note: 9-30 VDC loop power required (not supplied via converter)
Display and Measurement
• 2-Line LCD display (no backlight) • Flow rate and velocity (to 5 digits of precision)
• Non-volatile memory • Two alarms: low battery and empty pipe
Display
• Anti-reverse totalizer (standard) (optional)
• Total (to 9 digits of precision) • Opening lid activates display
Digits 5 Rate, 9 Total
GPM Gallons per minute IGM Imperial gal per minute CFM Cubic feet per minute
MGD Mega gal per day MI9 Miners inch (9G) B5M Barrels per minute (55G)
CFS Cubic feet per second MI1 Miners inch (11.22G) B5H Barrels per hour (55G)
MLD Megaliters per day APD Acre feet per day B5D Barrels per day (55G)
Units
LPS Liters per second KLH Kiloliters per hour B4M Barrels per minute (42G)
CMH Cubic meters per hour LPH Liters per hour B4H Barrels per hour (42G)
LPM Liters per minute CMM Cubic meters per minute B4D Barrels per day (42G)
GPH Gallons per hour CFM Cubic feet per minute

Page 13 30107-10 Rev. 2.7 | 08NOV2022


Page 20 30125-72 Rev. 1.2 | 26OCT2022
ProCommGO Converter Specifications SPECIFICATIONS
ProComm
GAL Gallons B42 Barrel (42G) MH1 Miners Inch Hour (11.22G)
CUF Cubic Feet B46 Barrel (46G) MD1 Miners Inch Day (11.22G)
AFT Acre Feet B55 Barrel (55G) MH9 Miners Inch Hour (9G)
CUM Cubic Meters IMG Imperial Gallon MD9 Miners Inch Day (9G)
Totalizer Units
LIT Liters AIN Acre Inch KGL Kilo Gallons
MML Megaliter TON Ton (Short) MGL Mega Gallons
MTT Metric Ton (KL) MM1 Miners Inch Minute (11.22G) IN3 Cubic Inch
B31 Barrel (31G) MM9 Miners Inch Minute (9G)
Data Logger Standard with all models, minimum of five years of data stored

Other Specifications
• Data Logger - included as standard with five years of data storage at default (12hr)
Options and
interval. (Cable sold separately)
Accessories
• AC, DC, and battery powered with battery backup powered available
• IEC 61010-1, Pollution Degree II
Safety
• Overvoltage protection Category III
Certifications
• ISO 9001:2015 certified quality management system
Standard Model • Certified by MET to UL 61010-1 / CSA C22.2 No. 61010-1
• ISO 9001:2015 certified quality management system COMPONENT

• Certified by MET: Safety: UL61010-1 / CSA C22.2 No. IA


PM
O R
61010-1, Third Edition: Safety of Electrical Equipment For & T

Measurement, Control, and Laboratory Use ®


HL Model NSF/ANSI/CAN 61

• Certified by MET: Standards: ANSI / ISA12.12.01 / CSA C22.2 &


NSF/ANSI 372

No. 213, Nonincendive Electrical Equipment


• Class I and II, Division 2
• Class III, Divisions 1 and 2 Hazardous (Classified) Locations

Page 14 30107-10 Rev. 2.7 | 08NOV2022

Page 21 30125-72 Rev. 1.2 | 26OCT2022


ProComm DIMENSIONS
ProComm GO Converter Dimensions
10.0 DIMENSIONS
Meter mount converter

6.44"
4.34"

6.94"
6.94"

Remote mount converter

4.8”

6.4”

8.8”

8.8”

Page 15 30107-10 Rev. 2.7 | 08NOV2022


Page 22 30125-72 Rev. 1.2 | 26OCT2022
ProComm
ProCommGO Converter Cable Connections
INSTALLINGfrom
THESensor
CONVERTER AND CABLES
Example Remote Mount Configuration
2.2 Example Remote Mount Configuration
After installing the sensor, mount the converter and connect the cable to the meter’s junction box and the converter’s
external connection. Figure 4 below shows an example of a remote mount installation with an optional Smart Output
connection.
You will need to prepare the location where you will install the remote converter. The location cannot be further from the
flow meter than the length of the 25’ cable. This must be planned in advance because the cable cannot be lengthened.
Doing so will alter the calibration accuracy between the meter and the converter and void the warranty.

See section 3.3 See section 4.0 External


for connection for connection connection on
flow meter

See section 3.2 AC or DC


for connection power source

I IMPORTANT: The example shown above does not include grounding installation. The
meter will need to be grounded according to each meter's requirements.

Figure 4. Example remote mount configuration

2.3 Mounting the Converter


Note: This applies to the remote mount converter only.
If possible, mount the converter in an electronics shed or environmental
enclosure. The sun shield should be oriented in a direction to reduce
sun damage and ensure readability. This electronic unit is rated IP67 for
temporary flooding.
There is a stainless steel mounting plate attached to the back of the
ProComm GO converter, which is used to mount the converter to a solid,
flat surface using four bolts (Figure 5). Mounting plate feet are located at
the top, bottom, left, and right sides.

Figure 5. Mounting
converter to solid surface

Page 16 30107-10 Rev. 2.7 | 08NOV2022


ProComm GO Converter Cable Connections from Sensor
INTERNAL WIRE CONNECTION
ProComm
Wiring Diagrams
3.2 Wiring Diagrams
TERMINAL BLOCK ASSIGNMENTS

Coils Harness
Terminal Port Wire Color
YELLOW
8 4 Green/Yellow
11 4 Red
RED 14 4 Yellow

GREEN / YELLOW

Electrodes Harness

PINK Terminal Port Wire Color


BLACK 1 5 Green/Yellow
BLUE
GREEN / YELLOW
2 5 Blue
3 5 Black
4 5 Pink

Pulse Output Harness


Terminal Port Wire Color
PINK
BROWN 23 1 Pink
YELLOW 24 1 Brown
GRAY
25 1 Yellow
26 1 Gray

WHITE 4-20 mA Output Harness


GREEN Terminal Port Wire Color
21 1 White
22 1 Green

DC Power Harness
RED
Terminal Port Wire Color
BLACK
Negative 3 Black
Positive 3 Red

BLACK
AC Power Harness
Terminal Port Wire Color
WHITE
Chassis lug 3 Green
Neutral 3 White
ATTACHES TO
GREEN Load 3 Black
CHASSIS LUG
To complete AC power connection,
connect green grounding lug to
chassis as shown on previous page.

Page 17 30107-10 Rev. 2.7 | 08NOV2022


Page 7 30125-72 Rev. 1.2 | 26OCT2022
Phillips screwdriver
ding the 10. Pass the new zip tie through 5/32”
the Allen11. driver
Wrap the new zip tie around the
nge on the slots in the internal hinge.
Connectors new dry pack and cut off the
excess.
5. Unplug the batteries. 6. Cut the two zip ties securing the batteries.

ProComm GO Converter Battery Installation and Connection


cover
proof seal and remove the wire 2. Remove screws holding
crew. the lid to the housing
ot required for battery installation using a 5/32” Allen driver.
Lift the lid up and place
Security screw the lid upside hanging by
the green ground wire.

Cut wire
p ties 13. Place the small battery pack in 14. Place the large battery pack on the
the internal the center pocket, making sure the internal tray with wire leads Zip ties
the wire leads go out through passing through the slot.
the opening.
7. Remove the large battery pack. 8. Remove the small battery pack located underneath.
Pull the wires carefully from underneath the battery
tray.
batteries with fresh batteries, If you are installing batteries in a new meter
that does not have batteries already installed,
go to step 9.

Large Small
Page 1
battery 30125-57 Rev. 1.1 | 19AUG2022 battery
pack pack

packs to the internal tray with the 16. Plug the batteries into the circuit board.
ut off the excess of each of the zip • The large battery pack goes to connector B1.
• The small battery pack goes to connector B2.

Page 2 30125-57 Rev. 1.1 | 19AUG2022

B2 B1

Page 3 30125-57 Rev. 1.1 | 19AUG2022

Page 18 30107-10 Rev. 2.7 | 08NOV2022