Download as pdf or txt
Download as pdf or txt
You are on page 1of 84

PantOs 16 xp & PantOs DG xp

Panoramic and Cephalometric


Dental X-Ray Equipment
Service & Installation Manual

Blue X Imaging Srl Via Idiomi 1/8-33 20090 Assago ITALY e-mail bluex@bluex.it
PantOs 16/DG XP - Service & Installation Manual

Blue X Imaging S.r.l.


Via Mario Idiomi 1/8-33
20090 Assago ITALY

tel.+39.0245712171
fax +39.0245703385
e-mail bluex@bluex.it
www.bluex.it

PantOs 16 xp Ceph, PantOs DG xp


Panoramic and Cephalometric Dental X-Ray Equipment
Service & Installation Manual – English Edition

Version 3.1 January 2010

Printed 1/30/2010 1:21:00 PM


Code 69 680 00210

Blue X Imaging is committed to Total Quality.


Thank you for notifying us any error found in this document.

2/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

Symbols on Control Panels


12 13 14 15
1 16
3
2
6 6
17

7 8 9 10 11

1 Unit ON 10 ACKNOLDGEMENT
and Alarm Reset

BACKWORD
2 Unit READY 11 Carriage Shift
(referred to patient)
FORWORD
3 X-ray Emission 11 Carriage Shift
(referred to patient)
PROGRAM
4 PROGRAM Selection 12 1 through 7 for Pan,
8 through 10 for Ceph
PATIENT Size:
Carriage
5 Small, Medium, 13
Displacement in mm
Large, Extra-Large
INCREASE Collimator type:
6 /DECREASE 14 PAN A, PAN B, AP1,
kV (left side) LLA1, LLA2, AP XL
mA, mAs (right side) (Not Pan Solo units)
Alarm Indication
START Arm Sequen- 15
7 Error Message
tial Movement
On Display
X-ray Indicator
Alignment LIGHTS
8 ON for 30 s 16 (Ionizing Radiation
Emission)

9 TEST Mode 17 X-RAY Radiography


without Radiation Push Button

Version 3.1 6968000210 3/84


PantOs 16/DG XP - Service & Installation Manual

Table of Contents
1.  INTRODUCTION ............................................................................................... 6 
1.1  Purpose............................................................................................... 6 
1.2  Product Documentation...................................................................... 6 
1.3  Notice to Installers ............................................................................. 7 
1.4  Warning ............................................................................................. 7 
1.5  Environmental Data ........................................................................... 7 
1.6  Safety Issues....................................................................................... 8 
2.  GENERAL ........................................................................................................... 9 
2.1  Film System ....................................................................................... 9 
2.2  Digital System .................................................................................. 10 
2.3  Control Panels .................................................................................. 11 
2.4  Tools for System Verification .......................................................... 13 
3.  BASIC INSTALLATION .................................................................................. 14 
3.1  Room Preparation ............................................................................ 14 
3.2  Unpacking ........................................................................................ 15 
3.3  Assembling the column .................................................................... 16 
3.4  Raising the System ........................................................................... 19 
3.5  Ceph Arm ......................................................................................... 20 
3.6  Carriage Release .............................................................................. 20 
3.7  Electrical installation........................................................................ 21 
3.8  Mode Setting .................................................................................... 23 
4.  DIGITAL SYSTEM........................................................................................... 24 
4.1  Architecture...................................................................................... 24 
4.2  Host Computer (PC workstation) ..................................................... 24 
4.3  System Software .............................................................................. 24 
4.4  Digital Panoramic Software Installation .......................................... 25 
4.5  Digital Panoramic Sensor Calibration .............................................. 27 
5.  ACCEPTANCE TEST ....................................................................................... 31 
5.1  System Test ...................................................................................... 31 
5.2  Radiation Check in Panoramic Mode .............................................. 32 
5.3  Radiation Check in Cephalometric Mode ........................................ 32 
6.  FUNCTIONAL DESCRIPTION ....................................................................... 33 
6.1  Block Diagram ................................................................................. 33 
6.2  Mains Connection Module ............................................................... 34 
6.3  Power Supply Unit ........................................................................... 35 
6.4  Power Switching Unit ...................................................................... 37 
6.5  Regulation Board ............................................................................. 39 
6.6  Control Unit ..................................................................................... 41 
6.7  Tube-Housing Assembly.................................................................. 43 
6.8  Digital Panoramic Sensors ............................................................... 44 
6.9  Panoramic USB Sensor .................................................................... 45 
6.10  Ceph Imager ..................................................................................... 46 

4/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

7.  SERVICE FUNCTIONS ................................................................................... 47 


7.1  Offset Values Setting (Pan Calibration) ........................................... 48 
7.2  Languages ........................................................................................ 48 
7.3  Ceph Formats ................................................................................... 48 
7.4  Exposure Counter (Tube Head Statistics) ........................................ 48 
7.5  Exposure Setting .............................................................................. 49 
7.6  Battery Replacement (Battery Change)............................................ 49 
7.7  Software Information ....................................................................... 49 
7.8  CAM Setting .................................................................................... 49 
7.9  Thermal Settings .............................................................................. 50 
7.10  Key Buttons Test.............................................................................. 50 
7.11  LED Test .......................................................................................... 50 
7.12  Automatic Cycling ........................................................................... 50 
7.13  Radiation Test .................................................................................. 51 
7.14  Test position sensors ........................................................................ 52 
7.15  Test Voltages ................................................................................... 52 
7.16  Ceph Alignment ............................................................................... 52 
8.  SERVICE PROCEDURES ................................................................................ 53 
8.1  Plastic and Metal Covers.................................................................. 53 
8.2  Replacement of Low Voltage Transformer ...................................... 55 
8.3  Replacement of Power Supply Unit ................................................. 56 
8.4  Replacement of Control Unit Board ................................................ 57 
8.5  Replacement of Regulation Board ................................................... 58 
8.6  Replacement of Power Switching Unit Board ................................. 59 
8.7  Replacement of Tube Housing Assembly ........................................ 60 
8.8  Replacement of Position Sensors ..................................................... 65 
9.  ALARM MESSAGES ....................................................................................... 68 
9.1  Warning Messages ........................................................................... 68 
9.2  Error Messages................................................................................. 68 
9.3  Mains Fuse Blows at X-ray On ........................................................ 72 
9.4  Other Alarm Messages .................................................................... 72 
10.  MAINTENANCE .............................................................................................. 73 
11.  SPARE PARTS .................................................................................................. 74 
11.1  Positioning & Test Items.................................................................. 74 
11.2  Power Supply & Control Modules ................................................... 75 
11.3  Mechanical tems .............................................................................. 77 
Appendix A Alarm Messages ..................................................................................... 78 
Appendix B Icons ........................................................................................................ 79 
Appendix C Technical Data ........................................................................................ 80 
Appendix D Labels ..................................................................................................... 81 
Appendix E Cooling Curves ....................................................................................... 82 
Appendix F Fuse Table .............................................................................................. 83 

Version 3.1 6968000210 5/84


PantOs 16/DG XP - Service & Installation Manual

1. INTRODUCTION
1.1 Purpose
The family of products for extra-oral dental radiology includes traditional
systems with film cassette or digital ones for direct diagnostic image ac-
quisition on computer.
A PantOs 16 xp / Ceph
These models are traditional units equipped with cassette for use of films
• either for chemical processing
• or phosphor plates for laser scanning and digital acquisition.
Two versions are available
• Panoramic Solo (Pan Solo) and
• Panoramic and Cephalometric (Pan Ceph) with arm for teleradiogra-
phy.
Each system can be equipped with temporo mandibular resting bars, on
request.
A panoramic cassette holder for 24x30 cm cassette is optionally available.
B PantOs DG xp / Ceph
These models are based on the use of digital imaging receptors for direct
acquisition of diagnostic image and subsequent computer processing.
Diagnostic functionality is the same of film systems both for panoramic
and for cephalometric modes.
Each one of these systems can be equipped with temporo mandibular rest-
ing bars, on request.
1.2 Product Documentation
The Operator’s and the Service and Installation Manuals supplied with the
system are integral part of the product and must be kept with the product.
The original language of the manuals is English.
Quick Reference Data is provided at the end of each manual
• Appendix A: Error Messages
• Appendix B: Icons
• Appendix C: Technical Data
• Appendix D: Labels
• Appendix E: Cooling Curves
• Appendix F: Fuse Table

6/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

1.3 Notice to Installers


Obligations of the installer are:
• To make sure that the line voltage specified by the Manufacturer of the
equipment is available and within the specified range.
• For safety reasons verify that a proper switch is available to disconnect
the equipment from mains when needed during installation.
• To install and test the equipment with due diligence according to the in-
stallation instructions from the Manufacturer.
1.4 Warning
X-ray equipment produces ionizing radiation that may be harm-
ful if not properly controlled. It is therefore recommended that
the equipment be operated by trained personnel only, in accor-
dance with existing law.
Even if compliant to specifications of electromagnetic compatibility, it is
recommended not to use the equipment in presence of external electro-
magnetic fields, such as those generated by cellular phones, which might
interfere with the electronic circuits of the system.
1.5 Environmental Data
Applicable ranges of temperature, humidity, and atmospheric pressure are
reported here below both for operation and transport conditions.

Ambient Transport & Storage


Temperature from 10 to 40 °C from –20 to +50 °C
Relative Humidity from 30 to 75% 10 to 100%
Pressure from700 to 1060 hPa from 500 to 1060 hPa

In case of PantOs DG xp Ceph the environmental data are limited as be-


low indicated:

Ambient Transport & Storage


Temperature from 15 to 27 °C from –18 to +49 °C
Relative Humidity from 30 to 75% 10 to 90%
Pressure from700 to 1060 hPa from 500 to 1060 hPa

Version 3.1 6968000210 7/84


PantOs 16/DG XP - Service & Installation Manual

1.6 Safety Issues


• Electrical.
ƒ Trained and qualified technicians only are authorized to remove cov-
ers and have access to power circuits.
ƒ Power supply lines must comply with safety legislation and have
ground terminals for protective earth connection.
ƒ Switch the equipment off and possibly disconnect it from mains be-
fore cleaning or disinfecting the unit.
ƒ FOR DIGITAL SYSTEMS. Requirements for the hardware, like
computers and other additional devices to be used for image acquisi-
tion, processing, and printing:
(a) ) it is recommended that such hardware be compliant with the
requirements of the safety standard IEC EN 60601-1
(b) if not,
(b1) such hardware must stay outside the patient area, i.e. at an
horizontal distance greater that 1.5 m from the patient or more
than 2.5 m from the floor, when placed above the patient,
(b2) it is recommended that such hardware be compliant with
the requirements of the standard IEC EN 60950

• Mechanical.
• Make sure that fingers or other parts of the patient or of the operator are
not pinched during the movement of the unit.
• Explosion.
ƒ The equipment cannot be used in presence of flammable gases or
vapors.
• Radiation.
ƒ Trained and qualified personnel only are authorized to operate the
equipment always complying with existing law for Radiation Protec-
tion.
ƒ Safety distance. The operator is invited to stay 3 meters from the
column (in the designated significant zone of occupancy) or exit the
room, always with patient in view, ready to terminate the exposure in
case of need.
ƒ Make sure that the equipment is not left unattended.
• Environmental.
ƒ The equipment contains components, which must be disposed-of fol-
lowing existing law.

8/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

2. GENERAL
2.1 Film System

CONTROL PANEL
CASSETTE HOLDER
PANORAMIC

CEPH ARM

CEPHALOSTAT

CASSETTE HOLDER
CEPHALOMETRIC

CHIN REST HOLDER

COLLIMATOR WHEEL

TUBE HOUSING ASSEMBLY

COLUMN SAFE ZONE OF


OCCUPANCY
FOR OPERATOR

3m
SELF STANDING BASE 10 foot

Version 3.1 6968000210 9/84


PantOs 16/DG XP - Service & Installation Manual

2.2 Digital System

CONTROL PANEL

PANORAMIC SENSOR

CEPH
ARM

CEPHALOSTAT
DIGITAL
CEPH CHIN REST
IMAGER HOLDER
COLLIMATOR WHEEL

TUBE HOUSING ASSEMBLY

COLUMN

SAFE ZONE OF
OCCUPANCY
FOR OPERATOR

3m
10 foot
SELF STANDING
BASE UNIT

10/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

2.3 Control Panels


A Radiation Control
12 13 14 15
1 16
3
2
6 6
17

7 8 9 10 11

10 ACKNOLDGEMENT
1 Unit ON
and Alarm Reset
BACKWORD
2 Unit READY 11 Carriage Shift
(referred to patient)
FORWORD
3 X-ray Emission 11 Carriage Shift
(referred to patient)
PROGRAM
4 PROGRAM Selection 12 1 through 7 for Pan,
8 through 10 for Ceph
PATIENT Size:
13 Carriage
5 Small, Medium,
Displacement in mm
Large, Extra-Large
INCREASE 14 Collimator type:
6 /DECREASE PAN A, PAN B, AP1,
kV (left side) LLA1, LLA2, AP XL
mA, mAs (right side) (Not Pan Solo units)
Alarm Indication
START Arm Sequen- 15
7 tial Movement
Error Message
On Display
X-ray Indicator
Alignment LIGHTS
8 ON for 30 s 16 (Ionizing Radiation
Emission)
TEST Mode X-RAY Radiography
9 without Radiation 17 Push Button

Version 3.1 6968000210 11/84


PantOs 16/DG XP - Service & Installation Manual

B iPan HS USB Interface

Digital Detector
18 indicator (green light)
When blinking:
digital sensor is transmit-
18 ting data to iPan USB in-
19 terface
USB Interface
19 (orange light)
When blinking: the inter-
face is transmitting data
to host computer

Green Light Orange Light

CONNECTORS (BACK VIEW)

Digital Detector USB Interface USB CABLE

CAT-5 CABLE

USB cable connected, digital sensor


OFF ON
NOT detected
ON OFF Error condition
USB cable connected,
ON ON
digital sensor detected

The digital sensor is transmitting image


Blinking
data to the USB interface

The interface is transmitting image data


Blinking
to host computer

USB cable NOT connected or USB driver


OFF OFF
not installed

12/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

2.4 Tools for System Verification


These tools are provided on request and have to
be ordered.
• Needles phantom to check the panoramic pro-
jection which is used mounted on the chin
rest.
• Alignment plate.
• Phosphor plate X-
ray detector for
digital or for analog
systems to
check the rad-
iation beam
on the image
receptor.

• Here below the marking of the left (L) ndicators on the lower edge of
the intensifying phosphor screen on the X-ray tube side.

EXTERNAL SIDE

PANORAMIC CASSETTE 15X30


WITH INTENSIFYING SCRRENS

TUBE SIDE

Version 3.1 6968000210 13/84


PantOs 16/DG XP - Service & Installation Manual

3. BASIC INSTALLATION
3.1 Room Preparation
Room Width for Pan Ceph
240 cm / 95 inch

Column
to Wall
85 cm
33 inch
Room
Depth
150 cm
59 inch
Room-
minimum Height
110 cm 225 cm
44 inch 89 inch

140 cm / 55 inch
Room Width for Pan Solo

A visit to the installation site has to be planned to prepare the room.


• Minimum room size:
ƒ Pan Solo: cm 140x110x225 (inch 55x44x89)
ƒ Pan Ceph: cm 240x110x225 (inch 95x44x89)
ƒ Wall support at 210 cm (82” 11/16) from the floor.
ƒ The use of self standing base requires that the same be secured to the
floor with adequate bolts.
• Electrical issues:
ƒ The connection to the mains line has to be intercepted by a safety
switch to disconnect the equipment from the mains when needed.
ƒ The connection to the mains line has to be made on the wall on the
back of the column at about 100 cm (40”) from the floor.
ƒ In case of need the exposure hand-switch can be placed remotely us-
ing a 4 wire cable (kit for remote mounting is available).
• Radiation Protection.
ƒ The system is equipped with lead barriers to cut residual radiation.
Additional shielding to the walls is depending on local regulations.

14/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

• Computer and Data Communication (for Digital Systems only)


ƒ The connection to the computer for data acquisition is made with a
cable 25 foot long (about 7.6 m) supplied with the system.
• Computer (for Digital Systems only)
ƒ The computer too be compatible in relation to hardware and software
(operating system) requirements and safety.
3.2 Unpacking
A Packaging Size and Weight
System Column Ceph Ceph Standing
Overhead Pan Plus Arm Imager Base
Width 120 cm / 47” 240 cm / 95” 150 /60” 120 cm / 47” 120 cm / 47”
Depth 80 / 32” 30 cm / 12” 60 cm / 24” 80 / 32” 100 cm
Heigth 60 cm / 24” 36 cm / 14” 60 cm / 24” 60 cm / 24” 20 cm / 8”

B Unpack the components and check the following:


• Each item is in good condition and did not suffer damages during trans-
portation.
• All the items for the specific configuration are available.
C Computer hardware (for Digital versions only)
• The computer hardware is not part of the supply with the radiological
equipment but it is an essential module to complete system installation
of digital versions.
D Mounting steps are listed below:
• Prepare the column.
• Attach the optional self standing base if in the configuration.
• Slide the overhead assembly on the column.
• Raise the column with overhead assembly.
• Secure the column to the wall or alternatively secure the self standing
base to the floor (MANDATORY).
• Complete the installation with accessories and perform connection to
the mains line.

A detailed description of the mounting steps follows.

Version 3.1 6968000210 15/84


PantOs 16/DG XP - Service & Installation Manual

3.3 Assembling the column


• Take the body of the column out of the
package and lay it down on the floor on
two wooden blocks with its back side up-
wards (front down).

Free the coun-


terweight from
the blocking
woods.
When applica-
ble mount the
additional
weight pro-
vided with the
system over-
head carriages
(refer to the Weight Table below).

WEIGHT PantOs 16 xp PantOs 16 xp PantOs DG xp PantOs DG xp


TABLE Ceph Ceph
System Type Pan Solo Analog Pan Ceph Analog Pan Solo Digital Pan Ceph Digital
Unit of
kg lb kg lb kg lb kg lb
Measure
Basic coun-
70.0 154.3 70.0 154.3 70.0 154.3 70.0 154.3
terweight
Additional
0 0 16.5 36.4 0.0 0.0 28.2 62.2
Weight
Total Balanc-
70.0 154.3 86.5 190.7 70.0 154.3 98.2 216.5
ing Weight

• Verify proper connection of cables from


the balancing arm with holding cylinders
(counterweight side) through the pulleys
(counterweight roller guide)
ƒ Presence of seeger rings blocking the
cylinders for the terminals of the sus-
taining cables
ƒ Correct insertion of sustaining cables
on the pulleys of the counterweight
roller guide, entering anterior and exit-
ing posterior.

16/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

• Pass the sustaining cables through the


rectangular holes at top of the body of the
column on the back.

• Connect the sustaining cables to the sus-


taining plate.
ƒ Remove the blocking strip form the
holding plate (3 Allen screws).
ƒ Insert the terminals of the sustaining
cables. BLOCKING
STRIP
ƒ Place back the blocking strip on the
holding plate).

• Insert the counterweight into the body of


the column with the holes for safety bolts
upward and block it with 2 safety bolts

HOLDING
PLATE

ANGLED
SAFETY PLATE
BOLTS

• Pull the holding plate and block it to the body of the column with the
angled plate.
• Protect the paint of the column with a plastic sheet between cables and
body of the column.

Version 3.1 6968000210 17/84


PantOs 16/DG XP - Service & Installation Manual

• Rotate the column laying on the back ready to mount the carriages of
the system.
• Verify the presence of 2 safety bolts blocking the counter-
weight and preventing its movement during system assembly.
DO NOT REMOVE SAFETY BOLTS UNTIL SYSTEM
ASSEMBLY IS COMPLETED.
• Disconnect the 2 spacing blocks from the holding plate and have the 4
bolts M8x40 ready to connect to the frame of the body of the system.
The 4 nuts are not used.
• Slide in the body of the system carefully, making sure the magnets of
the electrical breaking system on the left side are not interfering.
• Fix the holding plate with the sustaining cables on the back of the ver-
tical carriage using the 2 spacing blocks and the 4 bolts M8x40 pro-
vided.
• The angle plate must be in place to stop the vertical carriage from slid-
ing during the raising of the system.

LEVEL ADJUSTMENT
LEVEL ADJUSTMENT

FLOOR SECURING POINTS

• In case the self standing base has to be mounted, the column foot (base
block) with 6 bolts, each with rubber bumper, has to be removed from
the bottom part of the column.
• The self standing base is blocked to the column
with 6 bolts M 10x25.
• 4 screws are available for level adjustment.

• The plate has to be secured to the floor


with 4 bolts in the predefined holes
covered with plastic caps.

18/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

3.4 Raising the System


• The wall holder has to be mounted at
210 cm (82” 11/16) from the floor.
• The room place necessary from mid plane
of the column is of about 85 cm (34”) on
the right side for the operator to enter.
• Fix the wall holder to the wall by proper screws.
ƒ Steel or nylon anchors for solid concrete or solid bricks
ƒ Anchors for wood on wooden walls
ƒ Chemical anchors for perforated bricks
ƒ The sizes of the core of the bolts are listed below.
In case the wall is not in condition to withstand the indicated load,
the adoption of reinforcing plates and connection with bolts through
the wall can be considered.
• Raise the system and bring the column close to the wall holder.
Next fix the column to the wall holder with the two M 8x20 screws
without tightening.
• Assure proper leveling the side rails of the column preventing left or
right tilting as well as forward or backward tilting.
When leveling is fine, tightening of the bolts can be done.
• The same type of bolts can be used to secure the self standing base to
the floor, where the normal load is lower than for wall connection.
Recommended Bolts
Diameter Class Core Section mm2
M 8X1.25 ISO 8.8 36.6
M 8X1 ISO 8.8 39.2
5/16” – 18 UNC SAE- Grade 5 33.8
5/16 –24 UNF SAE- Grade 5 37.41

INSUFFICIENT WALL OR HARDWARE STRENGHT MAY


CAUSE THE WALL HOLDER TO PULL OUT FROM THE
WALL AND THE FULL SYSTEM TO FALL ON TO THE
PATIENT OR THE OPERATOR CAUSING INJURIES.

SIMILAR CONSIDERATION APPLIES TO THE MOUNTING


WITH SELF STANDING BASE TO BE SECURED TO THE
FLOOR, WHERE THE FLOOR AND THE USED HARD-
WARE MUST BE SUFFICIENTLY STRONG FOR THE
UNIT NOT TO DISCONNECT AND FALL DOWN.

Version 3.1 6968000210 19/84


PantOs 16/DG XP - Service & Installation Manual

3.5 Ceph Arm


• Cephalometric version requires the
Ceph arm on the left side of the col-
umn.
• Additional weight to be added to the
basic counterweight to balance.
The additional weight is delivered loose
and has to be assembled with the exist- CEPH
ing counterweight. ARM
• Once the column with the panoramic
overhead assembly has been raised and
has been secured to the wall or to the
floor, the Ceph arm can be bolted to the
holding plate on the back of the vertical
carriage using the 4 screws M8x16 and
washers supplied.
• Cephalostat, cassette holder, accesso-
ries and covers have to be mounted too for correct balancing.
3.6 Carriage Release
• The safety bolts on the back of the column (two M12x40) blocking the
counterweight can now be removed.
Do not dispose of them, but store in a safe place for future need.
WARNING. Due to minor weight differences, the carriage can slide up
until the unit is switched on and the electrical brakes block the carriage.
• Additional balancing weights are provided to balance the system. They
can be secured with
ƒ 2 bolts M 8x25 on to the holding plate for pan Solo systems
ƒ 2 bolts M 6x25 on to the Ceph arm for Pan Ceph systems

One transport bracket is


mounted on the top of the TRANSPORT
unit to prevent the move- BRACKET
ment of carriage during
transportation and has to be
removed before switching
the system ON.
TRANSPORT
BRACKET

20/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

3.7 Electrical installation


A Mains voltage configuration

PLUG ON LEFT SIDE FOR 115 V PLUG ON RIGHT SIDE FOR 230 V

The system is delivered with voltage selector


jumper and fuses NOT mounted. The system has
to be configured to work at 115 or 230 V depend-
ing on local mains line voltage. Access to the
F 501 Power Supply Unit placed on top of vertical car-
riage and to the right side of the body behind the
mirror is required. Refer to section 8.1, Plastic
HANDSWITCH
and Metal Covers, at page 53 for cover removal
SOCKET • Remove Vertical Upper Top cover, metal
shield on Power Supply Unit and mirror cover.
• 115 V configuration:
ƒ jumper on left side: MAINS CABLE SUP-
PORTING GUIDE
ƒ 3.15 AT on FU303
ƒ 16 AT on F501
• 230 V configuration:
ƒ jumper on left side:
ƒ 1.6 AT on FU303
ƒ 8 AT on F501
• Common to 115 V and
230 V fuses rated
6.3 AF on FU304 and
FU305

Version 3.1 6968000210 21/84


PantOs 16/DG XP - Service & Installation Manual

B Connection to Mains
• Proper cabling to be in place in the room for connection to the mains.
• Connection to be made behind the column at 40 inches (about 100 cm)
from the floor, with cable free to move up and down with the carriage.
• Mount the supporting guide for the mains cable.
• Perform polarity test to make sure that the “live” wire is actually the
“hot” one connected to the mains fuse:
ƒ Turn “OFF”. the mains room switch and system mains switch
ƒ Connect an AC voltmeter between the “fuse contact” and “ground”.
ƒ Turn “ON” the mains room switch and system mains switch.
If full line voltage is measured the wiring is correct.
If not, turn “OFF” the mains room switch and system mains switch,
reverse “live in” and “neutral in” wires, then repeat measurement.
Eventually the full line voltage is read between the “live” point on
the fuse contact and “ground”.
Line wires are either colored or numbered for easier identification.
The wire connected to the fuse contact has to be the “live” one.
ƒ Make sure no voltage is read between the “neutral” wire and
“ground”. If not check line voltage distribution.
ƒ Turn “OFF” the mains room switch and system mains switch
• Connect the power supply cable to the mains line block, permanently
• Apply fuse label for selected mains voltage close to the mains fuse
holder or check the applicable box on the existing label
• Be ready to turn “ON” the mains room switch and system mains switch
to supply the system when needed.
C Exposure Hand-Switch
• Connect the exposure hand-switch plug into
the relevant socket on the right side on the
vertical carriage.
• In case of use of remote hand-switch discon-
nect the coiled cord of the hand-switch and
connect the extension cable (4 wires cable).
YELLOW LIGHT

EXPOSURE HAND-
SWITCH CABLE EXPOSURE
PUSHBUTTON

22/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

3.8 Mode Setting


The dip switches on top of Control Unit board are set during manufactur-
ing to match the system configuration as indicated above.

To complete the configuration of Ceph versions the applicable image for-


mats are defined by means of a specific Service Function.
The jumper for the signal of the temperature sensor has to be in place.

DIP 1 DIP 2
0 OFF ANALOG 0 OFF PAN CEPH
1 ON DIGITAL 1 ON PAN SOLO

DIP 3 – 4 – 5 – 6 - 7- 8
0 OFF

JUMPER FOR SIGNAL


OF THERMAL SENSOR

Version 3.1 6968000210 23/84


PantOs 16/DG XP - Service & Installation Manual

4. DIGITAL SYSTEM
4.1 Architecture
CAT 5 USB
CABLE CABLE
PantOs iPan HS PC
INTERFACE WORKSTATION
HOST MACHINE

MAINS
POWER

The PantOs radiographic system for panoramic and cephalometric radiol-


ogy is connected to the PC Workstation through the iPan HS (High Speed
USB) interface. A CAT-5 cable connects the Host Machine to the PantOs
interface and a USB cable connects the iPan HS interface to the PC
workstation. The host computer should be dedicated to imaging and not
for general computing use.
4.2 Host Computer (PC workstation)
The recommended system requirements for PC workstations are:
• Compatible operating system (Windows 2000, XP, or Vista).
• 512 MB RAM or higher.
• CPU Pentium IV: 1 GHz or better.
• 40 GB free disk space (Storage requirements will vary depending on
the patient volume for each practice).
• The PC workstation, CE-approved computer system conforming to the
Low Voltage (73/23/EC) and EMC directive (89/336/ERC) connects to
the iPan HS interface via a compatible USB cable and serves as the host
for the OrisWin Pan software and utilities.
• For safety, connect the cable from the iPan HS Interface to Listed / Ap-
proved / Certified Information Technology Equipment (ITE) computer
equipment only.
• Monitor display settings for dedicated workstation CRTs should be at
least 0.25 dot pitch @ 1024x768; for flat screen LCD screens 450:1
contrast ratio @ 1280x1024 resolution is recommended.
4.3 System Software
The PantOS imaging system requires the following software and compati-
ble operating systems:
• Operating System: Windows 2000, XP, or Vista.
• Imaging Software: OrisWin Pan or application with TWAIN interface.
• Device Driver: iPan HS Interface Driver.

24/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

4.4 Digital Panoramic Software Installation


Do follow the installation sequence as reported here:
(1) Install the OrisWin Pan application software,
(2) Install the iPan HS Interface Driver,
(3) Connect the iPan HS USB interface to the computer,
(4) Use the iPan HS Upgrade Utility to verify proper installation,
(5) Start using the PantOs DG xp system.
A OrisWin Pan Software installation
• Turn the computer ON and load the OrisWin Pan installation CD
• Select the proper folder and start the setup process. The Welcome
screen is displayed. Click Next to start installation
• Provide Customer information and references to destination location
• Select type of installation (typical is recommended)
• Click Finish to terminate the installation procedure.
B iPan HS Interface Driver installation
• Exit running programs and verify that the iPan HS USB Interface is
NOT connected to the computer
• Insert the iPan HS Interface Driver CD. Start “iPan HS Interface Driver
Setup.exe” if not started automatically
• Click Next, click Install, click Finish
C iPan USB Interface connection to computer
• Connect the iPan HS Interface to the computer using a USB cable
• Click Next when the Add New Hardware wizard starts.
• Click Finish when the Add New Hardware wizard ends. The iPan HS
Interface driver is copied to the system
D Starting iPan HS Upgrade Utility
• Start > Programs > CDR DICOM for Windows > iPan HS Upgrade
Utility
• Verify that all version numbers reported are in green. In case not verify
proper connection of plugs and cables. Call Technical Service in case of
problems.
E Controls and LED Indicators
Control lights are located on the exterior of the iPan HS Interface Driver
box and provide power and status information.

Version 3.1 6968000210 25/84


PantOs 16/DG XP - Service & Installation Manual

iPan USB Interface LED Indications


Digital Detector
18 indicator
GREEN
LIGHT (green light)
When blinking:
digital sensor is trans-
mitting data to iPan
USB interface
USB Interface
19 (orange light)
ORANGE When blinking: the in-
LIGHT
terface is transmitting
data to host computer

CONNECTORS (BACK VIEW)

Green Light Orange Light USB CABLE

CAT-5 CABLE

Digital USB
Detector Interface

USB cable connected, digital sensor


OFF ON
NOT detected
ON OFF Error condition
USB cable connected,
ON ON
digital sensor detected
The digital sensor is transmitting image da-
Blinking
ta to the USB interface

The interface is transmitting image data


Blinking
to host computer

26/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

4.5 Digital Panoramic Sensor Calibration


Start OrisWin Pan to perform sensor calibration.
• To perform the calibration place the X-ray filter (0.5 mm copper filter)
over the collimator to intercept the X-ray beam. This enables to take a
flat-field image of medium grey level. After performing the calibration
procedure, the calibration file is created. Remember to remove the X-
ray filter from the collimator before taking a non-flat-field image.
• Go to Calibrate and click on Ca-
librate mode. This will allow un-
corrected images to be taken for
creating a calibration file. Cali-
brate mode status can be checked
by clicking on Properties. In the
Acquisition box disable
(leave unchecked) the boxes
Calibrate Image and Subtract
Dark Image. The Mapping
box to be Linear. Set
ƒ Image Width to 3070
ƒ Image Height to 1603,
ƒ Initial Timeout to 15
ƒ Early Stop Timeout to 7.

• The Calibration box to be


set with:
ƒ Gap Width (lines): 3
ƒ Calibration
Threshold: 99
ƒ Dark Threshold: 99
ƒ Extra Bad Rows: to be
unchecked
ƒ Clip saturated Pixels: set
checked

Version 3.1 6968000210 27/84


PantOs 16/DG XP - Service & Installation Manual

• Take a dark image without X-ray emission (white picture):


ƒ Select Standard Panoramic Projection on the system control panel.
ƒ Press the TEST button to prevent X-ray emission.
ƒ Under OrisWin Pan click on the New Image icon in the toolbar or
select File in the menu and click on New Image.
Make sure this is the first
image (Image 1 on the tool-
bar).
ƒ Press X-ray push button and
complete the panoramic ro-
tation.
A white image is displayed
with a black horizontal line
at the center.
• Take a flat field image with X-ray emission:
ƒ Select Standard Panoramic Projection on the system control panel
ƒ Press the TEST button on the system to exit the test condition.
ƒ Make sure the 0.5 mm copper filter is intercepting the X-ray beam
(collimator).
ƒ Set 61 kV and 4 mA on the system control panel.
ƒ Under OrisWin Pan click on the New Image icon in the toolbar or
select File in the menu and click on New Image. Make sure this is
the second image (Image 2 on the toolbar).
ƒ Press X-ray push button and take a panoramic exposure.
Make sure you are in a safe position during X-ray emission
to prevent unnecessary exposure to radiation.
ƒ The image taken contains a vertical dark band in the center of the
image, created by the re-
duction in the speed of the
X-ray source as it travels
behind the spine. Pixel
values outside the vertical
dark band should fall be-
tween 30% and 60%. If not
erase the picture and take a
new one having adjusted
the technique factors (kV and mA) for darker (increase) or lighter
(decrease).

28/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

• Go to Calibrate and Click Ca-


librate.

ƒ Check the parameters in the Ca-


libration File dialog box and,
in case, adjust the input parame-
ters for Crop Image and Ex-
clude Dark Band.
The coordinates of a point are
indicated at the bottom when
clicking on the point.
ƒ Click Create file. This creates
the calibration file, then the cali-
bration mode is terminated.

• Recall “Calibrate” / ”Proper-


ties” and make sure the set
up has been changed for im-
age acquisition with “Cali-
brate Image” and “Subtract
dark image” selected (boxes
checked) and Mapping set to
“Nonlinear (traditional)”.
Should it not, perform ma-
nual selection and make it
accepted.

• Take a picture to ve-


rify positive conclu-
sion of sensor cali-
bration. Only images
that have been cali-
brated will have the
R (right) and L (left)
markers superim-
posed on them.

Version 3.1 6968000210 29/84


PantOs 16/DG XP - Service & Installation Manual

Acquisition Window
iPan HS USB Interface Device Driver name
Scan direction Left to Right
3100 / 3060 Image width in pixels
Image height in pixels, including space be-
1603
tween sensors (3)
Default 15
Initial Timeout Number of seconds to wait before starting da-
ta acquisition after beginning of rotation
Default 7. Maximum number of seconds to
Early stop timeout
read a column before timing out
Calibrate Image Use calibration data to adjust image taken
Equalize Image Optimize image contrast
Apply Spot Remover Remove black and white spots
Subtract Dark Image Use offset data to adjust image taken
Acquire 12 bit image Increase grey level resolution
Turns on and off automatic histogram calcula-
Map Auto Threshold
tion for mapping
Mapping Type of mapping of grey levels

Calibration Window
File path Directory of the destination folder
3 Gap Width between sensors as number of
lines
Calibration Threshold Pixels at this level and above are not calibrated
Dark Percentage of grey level above which dark
Threshold subtraction is not applied
Extra Expands to 5 lines the interpolation area in the
Bad Rows gap between sensors
Clip Saturated Pixels Saturated pixels are not subject to calibration

30/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

5. ACCEPTANCE TEST
5.1 System Test
A Switching System On
• The unit is connected to the mains and
can be turned ON with the switch below MAINS
SWITCH
the chin support body.
• The electrical brake engages.
• Initialization is performed and count-
down is indicated on the display.
• “Do Reset” is requested. Press
START Arm or EXPOSURE push button to reset the system.
B Positioning the Carriage
• By pressing the brake pushbutton, the brake disengages and it is possi-
ble to adjust the vertical carriage at proper position.
• By releasing the brake push-button the brake engages and the move-
ment of the carriage is blocked.
C Aiming Lights
The lights for patient alignment are made by three laser beams.
WARNING. DO NOT STARE INTO BEAM OR VIEW DI-
RECTLY WITH OPTICAL INSTRUMENTS
• Test of the lights for
positioning of the
patient. MIDSAGITTAL
LIGHT
ƒ Press the LAMP
push-button on
the keyboard.
ƒ Three laser beams
are generated:
midsagittal ver- FRANKFURT AND
LATERAL LIGHTS
tical line, Frank-
furt horizontal
line and side (foc-
al trough) vertical
line.
ƒ The lights go off automatically after 30 s.

Version 3.1 6968000210 31/84


PantOs 16/DG XP - Service & Installation Manual

5.2 Radiation Check in Panoramic Mode


A Enable Service Function X-RAY TEST Continue
• For Pan Ceph systems make sure the tube-head is rotated in favour of
the panoramic cassette or imaging detector.
• Select collimator for Standard Adult Panoramic.
• Activate the service function for test of radiation, 70 kV and 6.3 mA are
adequate.
B Beam Check in Panoramic Mode
• Place the phosphor plate on the secondary collimating window on the
cassette holder (film system) or on the image receptor (digital system).
• Keep the room dark to see the marked lines. Watch the light of the
beam from below the X-ray generator, at a convenient distance and
with proper radiation shielding.
• Expose for few seconds. The irradiated field has to be symmetric to the
vertical median line of the detector, and has to fall in between the upper
and lower horizontal marks.
• For corrections of Panoramic field deviations refer to section 8.7, Re-
placement of Tube Housing Assembly, at page 60 and following.
5.3 Radiation Check in Cephalometric Mode
A Enable Service Function Tarature Ceph
• Select Ceph Mode by rotating the tube-head towards the Cephalostat
• Values of 61 kV and 5 mAs are usually adequate for checking the beam
using regular films.
B Beam Check in Ceph Mode
• Select the AP collimator (18x24S or 8x10S on display)
• Place a loaded cassette on the cassette holder in the AP position
(18x24S or 8x10S).
• Rotate the Cephalostat with ear supports directed along the beam and
opened to the maximum extent.
• Expose and check (1) the alignment of the exposed area on the film
(unexposed edge should be seen all around the picture) and (2) that the
ear supports for side radiographs (LL projection) are aligned.
• For corrections of Cephalometric field deviations refer to section 8.7.

32/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6. FUNCTIONAL DESCRIPTION
6.1 Block Diagram
• The mains line is connected to the terminal block on the right side of
the Vertical Carriage. The mains power is fed to the Power Supply Unit
located on top of the unit.
• Supply and signal conductors bring power from the Power Supply Unit
to the Power Switching Unit located in the rotating carriage. The power
supply board is providing necessary low voltages to feed the circuits.
Two auxiliary circuits are specifically dedicated to the laser lights and
the electric brakes.
• The power switching unit is connected to the tube-head and to the Reg-
ulation Unit with a flat cable.
• The Regulation Unit is located above the control panel, which is a part
of the system control unit.
• The system control unit drives the control panel, the power regulator
and the motors making use of the information from the position sensors.

DUAL POWER POWER TUBE-HEAD


SUPPLY SWITCHING X-RAY GEN-
115/230 V ERATOR
MAINS
POWER X-RAY
EMISSION

BRAKE LASER
SYSTEM LIGHTS REGULATOR KV & mA
FEEDBACK

CONTROL
PANEL
ROTATION
X-RAY
CARRIAGE PantOs 16 XP PUSHBUTTON
TRANSLATION
PantOs DG XP
SYNCRO
SENSOR POSITION SYSTEM SIGNALS DIGITAL
CONTROL IMAGING
UNIT SYSTEM

IMAGING
CASSETTE TRANSLATION DATA

SENSOR IMAGING
POSITION
COMPUTER

Version 3.1 6968000210 33/84


PantOs 16/DG XP - Service & Installation Manual

6.2 Mains Connection Module

The connecting blocks, mains switch,


fuse holder, varistors and cables are
placed on the right side of the vertical
carriage.
The supply line then moves up on the
top part of the vertical carriage where
reaches the dual voltage supply board
through the EMI filter and the low
voltage transformer.

On the same place the connector


XC503 for the X-ray pushbutton is
also located.

The wires of the switch are number 3


and 4 (those in the middle of the con-
nector).

34/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6.3 Power Supply Unit

LOW VOLTAGE TRAFO MAINS


INPUT

FU303

PLUG ON RIGHT SIDE FOR 230 V

MAINS
LEFT SIDE FILTER
FOR 115 V

CONNECTOR FOR
FU304 FU305 ADDITIONAL X-RAY
EXPOSURE LIGHTS

OUTPUT

At installation the unit can be set to work at 115 or 230 V after proper in-
sertion of configuration plug and fuses (see table below). Mains power is
fed via the interference filter to the low voltage transformer and via the re-
lay to the line voltage rectifier. The stabilized voltages of +15V, +24V and
non-stabilized of +38V are generated. Specific low voltage supply mod-
ules are available for the electrical brakes and for the laser lights.
During the preparation phase (filament heating) a signal from the control
unit causes the relay to close. During exposure the mains voltage rectified
is connected to the Power Switching Unit. The yellow light on the expo-
sure hand-switch (push button) and the buzzer are switched-on too.
System Fuses
Fuse Code Use 230 V Version 115 V Version
F 501 Mains Power Input 8 AT 16 AT
FU 303 Low Voltage Circuit 1.6 AT 3.15 AT
FU 304 Output Power to Switching 6.3 AF 6.3 AF
FU 305 Output Power to Switching 6.3 AF 6.3 AF

Version 3.1 6968000210 35/84


PantOs 16/DG XP - Service & Installation Manual

The equipment is protected by fuse F 501 (input power from mains), lo-
cated on the top part of the mains connection module, on the right side of
the vertical carriage. The circuit of the low voltage transformer is pro-
tected by fuse FU 303. The Output Power circuit (to the Power Switching
Unit) is connected by fuses FU 304 and FU 305

DUAL VOLTAGE SUPPLY BOARD 115-230 V

36/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6.4 Power Switching Unit

300 VDC
INPUT

CONTROL OF
FILAMENT
SUPPLY
CONTROL OF
HIGH VOLTAGE
GENERATION
OUTPUT
TO FILAMENT
TRANSFORMER OUTPUT TO HV
TRANSFORMER

The power switching module is divided in two parts: on the left side the
circuits for the control of the filament supply, on the right side those for
high voltage generation.
• During preparation and exposure the rectified mains voltage (300 VDC)
is fed into the unit.
• Pulses with a frequency of 80 kHz pass to the filament regulator driver.
The output voltage of the regulator depends on the duty cycle of the
driving pulses.
• Feedback from the regulator output is fed to the regulation unit.
• Pulses with the frequency of 40 kHz pass to the filament inverter driver.
Output of the inverter goes to the filament transformer in the tube-head.
• The inverter current is sensed by a dedicated resistor and the signal is
fed to the regulation unit.
• During preparation, the pulses with the frequency of 20 kHz pass to the
High Voltage (HV) inverter driver.
During exposure the pulses pass also to the HV regulator driver.
The output voltage of the regulator depends on the duty cycle of the
driving pulses.
• Feedback from the regulator output is fed via the opto-coupler to the
regulation unit.

Version 3.1 6968000210 37/84


PantOs 16/DG XP - Service & Installation Manual

• Output of the inverter is connected to the HV transformer in the tube-


head. The inverter current is sensed by a current transformer and rele-
vant information passed to the regulation unit.

38/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6.5 Regulation Board

PROBES 8-11

SA101

PROBES 1-7

The purpose of the regulation module is to control the level of the tech-
nique factors kV and mA by checking the feedback signals coming from
the tube-head, and comparing them with the reference signals set by the
control unit. The necessary information is passed to the voltage regulators
on the power switching unit.
• The level of the anode voltage (kV) is controlled by acting on the input
voltage to the high voltage transformer (tube-head).
• The level of the anode current (mA) is controlled by acting on the fila-
ment supply.
• Calibration is required only for the part controlling the filament (and
therefore the resulting anode current in mA) and has to be performed
every time either the tube-head, or the power switching, or the regula-
tion unit itself is replaced (due to tolerances of the components).
• Calibration of the kV level is done automatically by the control unit.
• The Switch SA101 ON during calibration, on OFF for normal use.
Two set of probes are available for testing purposes as indicated below.

PROBES 1-7 PROBES 8-11


1 Filament regulator feedback 8 Irradiation (L)
2 HV regulator feedback 9 HV failure (L)
3 Filament regulator driver 10 Filament check (0.5 – 2 V)
4 Filament inverter driver 11 Reference kV (6.1 – 8.5 V)
5 Filament inverter driver 12 Reference mA (4.0 – 10 V)
6 HV inverter driver
7 HV inverter driver

Version 3.1 6968000210 39/84


PantOs 16/DG XP - Service & Installation Manual

SA 101 “ON” ONLY


FOR CALIBRATION

40/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6.6 Control Unit

The control unit represents the heart of the system.


• It controls input from the operator, to set up the system when selecting
proper technique factors and type of projection, and output and control
of X-ray emission and mechanical movements during exposure.
• A number of error messages is handled to react to improper commands
or to signal non correct or unsafe working conditions.
• More in detail:
ƒ Driving motors for carriage rotation and carriage translation during
the execution of panoramic projections.
ƒ Driving motors for cassette on analog systems.
ƒ Digital Sensor synchronization for digital systems.
ƒ Analog inputs for supply voltage control (38 V).
ƒ Analog output for Pulse Width Modulation (PWM) control of high
voltage (kV) and filament current (resulting in mA control).
ƒ Digital output for X-ray on and heating alarm off.
• Reset DRK. By keeping pressed the PLUS and the MINUS keys
on the left side for 5 seconds a reset is performed of the control unit
board.

Version 3.1 6968000210 41/84


PantOs 16/DG XP - Service & Installation Manual

42/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

6.7 Tube-Housing Assembly

FEEDBACK
INTERFACE
1 2 3 4 5 6 7

• Power for high voltage generation is


supplied through points 4 and 5.
High voltage is generated by means
of a transformer plus a cascade mul-
tiplier.
• Filament supply is given through
points 2 and 3. The X-ray tube fila-
ment is supplied by a dedicated trans-
former.
• The feedback signals of both the
anode voltage (point 6 to kV point on
the feedback interface board) and the
anode current (point 1 to mA point
on the feedback interface board) are
supplied to the regulation unit.
• A sensor is available to generate a
temperature feedback (7) handled by
the control unit, to inhibit X-ray
emission when the temperature is out
of limit.

Version 3.1 6968000210 43/84


PantOs 16/DG XP - Service & Installation Manual

6.8 Digital Panoramic Sensors


A Digital System Architecture

Control Unit XP Exposure Pan


Enable Syncro

XC1 XC14 XC7


FLAT CABLE
XC101 Regulation XC4 XC1
Unit JED25 Pan Syncro
XC2 J6
Exp Enable
XC102 XC3 J3
CMOT2/3
Ceph Select
J7
XC202
Power Supply to Digital Module
XC204 J5
Power
Switching Power
XC203 XC302 Supply
Unit JED26
JED39

XC310 XC304
Tube Digital
Main Imaging
Housing Power Supply PanCeph
Assembly Trigger Cable Power Supply
to Mux Cable

J6 J7
SEL/Arm
Cable
J2 J8
Ceph Power Cable Ceph
J5 Mux J4 J1
Board USB Sensor
Ceph Data Cable Module Cable
J1
J3 J2 J4
Analog 10x28
CAT5 Sensor Ribbon
Cable Cable Cable
Ceph
Imager iPan HS Panoramic
Interface USB
Sensor
USB
Cable

PC

Digital sensors for panoramic and cephalometric imaging.

44/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual
Panoramic USB SensorThe panoramic imaging subsystem
subsystem is composed of CCD sensor, Acquisition board, local power
supply board, an data transmission board to send image information to the
computer

38 VDC
POWER
SUPPLY

SYNCRONIZATION COMPUTER READY


READOUT SIGNALS ENABLE SIGNAL

J5

J1

CAT 5 CABLE
TO iPAN BOX

J1 - CAT 5 CABLE
TO iPAN BOX

J5 - 38 VDC
POWER SUPPLY

J8 - PanCeph Rotation J7 - Ceph Mode


SEL/Arm Select

Version 3.1 6968000210 45/84


PantOs 16/DG XP - Service & Installation Manual

6.10 Ceph Imager


In case of Ceph configuration the system is extended with the digital Ceph
Imager and the multiplexer board to selectively send the digital images
from the panoramic or from the cephalometric sensor to the computer via
iPan HS Interface.

MULTIPLEXER BOARD iPan HS USB


J3
PAN
SENSOR

J4
CEPH
SENSOR

J1
J5 J6 J2 J7

Power Supply Pan Ceph Trigger


to Ceph

Main Power Supply to Mux SEL/Arm

CEPHALOSTAT MULTIPLEXER
BOARD ON THE
LEFT SIDE

CEPH IMAGER

46/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

7. SERVICE FUNCTIONS
By keeping pressed the RETURN key for 5 s the system enters Service
Mode. Service function can be selected with plus and minus keys
on the right side of the display.
Functions of the other buttons are:

(TEST) to exit service menu.


(RETURN) to go back to the beginning of the service menu.
(PLUS on the right side of the display) to select the service function.
(MINUS on the right side of the display) to select the service func-
tion.
(PROGRAM) to enter selected program. Keep pressed to store the
current value (not valid in “Exposure Setting” where the LASER
key is used instead ).
(START Arm Movement) to start some service functions.

Service Functions appear as listed below when scrolling downward with


the MINUS key.
Functions for Ceph mode appear only for systems configured as such (re-
fer to section 3.8, Mode Setting, at page 23).

Service Functions
1 PAN Calibration
2 Languages
3 Ceph Formats (Ceph only)
4 Tube Head Statistics
5 Exposure Settings
6 Battery Change
7 Software Info
8 Cam Setting
9 Thermal Setting
10 Keyboard Test
11 Leds
12 Automatic Test
13 X-ray Test
14 Encoders Test
15 Voltages Test
16 Calibration CEPH (Ceph only)

Version 3.1 6968000210 47/84


PantOs 16/DG XP - Service & Installation Manual

7.1 Offset Values Setting (Pan Calibration)


The program is intended to perform mechanical alignment of the
machine and compensate for mechanical tolerances.
WARNING. The determination of the Offset values must be
performed in sequence: Axe Cassette (not for digital systems),
Axe Rotation, and Axe X.
• AXE CASSETTE to set the offset value for cassette movement
• AXE ROTATION to set the offset value for rotation movement
• AXE X to set the offset value for X-axis movement (forward direction
is moving towards column and backward direction is far from column)
For further information on parameters’ changes refer to Technical Service.
7.2 Languages
The function allows for the selection of the proper language for the mes-
sages on the display among English, Italian, French, Spanish, German, and
3 available slots. The list is scrolled with the PLUS or the MINUS
keys, then selection is done with the PROGRAM key.
7.3 Ceph Formats
The function is intended to set the formats of Ceph collimators.
This function can be accessed when the dip number 2 of Control Board is
in Pan/Ceph position and allows for selection of configurations.
• Ceph Mode on Analog System (Dip Switch 1 to OFF):
ƒ Metric 18x24 cm (vertical)
ƒ Inches 8x10” (vertical)
ƒ Inches 10x8” (horizontal)
• Ceph Mode on Digital System (Dip Switch 1 to ON):
ƒ Digital type A: 24x24 cm (10x10”)
The value displayed is set during manufacturing and in case the need for
change arises please refer to Technical Service.
7.4 Exposure Counter (Tube Head Statistics)
Press PROGRAM to enter. The number of exposures panoramic (P),
cephalometric (C), and Total (T) is displayed.

48/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

7.5 Exposure Setting


Values of technique factors for each patient type (4 sizes) can be set by
changing kV and mA in panoramic mode (programs 1, 3, 4, 5, 6, 7 sepa-
rate from program 2 – 8 positions) or kV and mAs values in Ceph mode
(programs 8, 9, and 10 separate – 12 positions).
Keep pressed the LASER key to store the selected value.
The TEST key to reset all values to default conditions.
RETURN key and TEST key to exit.
7.6 Battery Replacement (Battery Change)
This function supports the replacement of the battery for the non volatile
memory NVRAM on the control unit board with LCD display.
In case the need please refer to Technical Service.
7.7 Software Information
SOFTWARE INFO to retrieve the code of the version of software loaded
into the memory of the control unit.
Press the PROGRAM key to display the first level code of the loaded
version, and press a second time to get the second level code.
7.8 CAM Setting
mm 88-90 pixel 900-940

This function is used to align the panoramic projection with the bite block
and the lateral light by checking and setting the distance of the two pins of
the test phantom.
The current value of CAM Setting appears on display when entering this
function. Set up is fine when the pin distance is of 89 +/- 1 mm (88-90) for
film systems, 920 +/-20 pixels (900-940) for digital systems.
• mA MINUS key on the right, to increase the pin distance.
• mA PLUS key on the right, to decrease the pin distance.
Keep pressed the PATIENT key to store the value.

Version 3.1 6968000210 49/84


PantOs 16/DG XP - Service & Installation Manual

7.9 Thermal Settings


Function to enable or disable the use of the thermal sensor and set the limit
temperature.
• First part: Enable / Disable
• Second part: Editor thermal values.
Refer to Technical Service in case of need. RETURN key to exit.
7.10 Key Buttons Test
TEST KEY function to display the code of the pushbutton key. The cor-
responding codes are reported in the table below.

2sx 2sx 2dx 2dx 4 13 5 6 7 8 9 11

Keep RETURN key pressed to exit.


7.11 LED Test
Test LED to let all lights on the panel to blink and the display as well.
Use ARM MOVEMENT key to start

7.12 Automatic Cycling


Select FULL ROTATION to have the unit performing a number of pano-
ramic cycles according to the selected program, using the technique fac-
tors set before entering the function.
A 100 PAN NO RX
Function to perform 100 panoramic movements without X-ray emission.
• Use ARM MOVEMENT key to start.
• Use EXPOSURE key on the hand-switch to terminate.

50/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

B 10 PAN + RX
Function to perform rotation with radiation using the technique factors in-
dicated on the display, as set prior to entering service mode.
• Use EXPOSURE key on the hand-switch to start.
• Use ALARM RESET key to terminate.
7.13 Radiation Test
Select X-RAY TEST to perform irradiation without motion:
• PULSE mode (pre-defined time), by setting of kV and mAs.
• CONTINUOUS mode (time set by the operator), by setting kV and
mA.
A Setting kV and mAs
kV and mAs at different mA values can be set within the defined ranges.

• To set kV use kV PLUS and MINUS keys on the left


kV 61 64 67 70 73 76 79 82 85

• To set mA use mA PLUS and MINUS keys on the right


mA 4 6.3 8 10

• Once set the mA value, use LEFT and RIGHT keys to set mAs
value
mAs 2 2.5 3.2 4 5 6.3 8 10 12.5 16 20

• mAs can be set at different mA value


mAs
2 2.5 3.2 4 5 6.3 8 10 12.5 16 20
mA
4 0.5 0.63 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0
5 0.4 0.5 0.63 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0
6.3 0.32 0.4 0.5 0.63 0.8 1.0 1.25 1.6 2.0 2.5 3.2
8 0.25 0.32 0.4 0.5 0.63 0.8 1.0 1.25 1.6 2.0 2.5
10 0.2 0.25 0.32 0.4 0.5 0.63 0.8 1.0 1.25 1.6 2.0

Irradiation can thus be obtained at different exposure times.

Version 3.1 6968000210 51/84


PantOs 16/DG XP - Service & Installation Manual

B Setting kV and mA
• To set kV use kV PLUS and MINUS keys on the left
kV 61 64 67 70 73 76 79 82 85

• To set mA use mA PLUS and MINUS keys on the right


mA 4 6.3 8 10

Irradiation can thus be obtained for the desired time.

WARNING. Wait for X-ray tube cool down after each exposure.
Cool down diagrams are reported on Appendix E.
5 min waiting time is required after each panoramic
of 20 s exposure at 85 kV, 10 mA (about 13000 J).
7.14 Test position sensors
Select this function to perform test of position and motion sensors.
Control of the motion is done with LEFT and RIGHT keys.
• Axe Rotation.
• Axe Cassette.
• Axe X.
7.15 Test Voltages
TEST VOLTAGE to get indication of actual levels of low voltage supply
at 38, 24, 15, and 5V with indication on the display.
7.16 Ceph Alignment
TARATURE CEPH to set manually the alignment position of the rotating
arm in cephalometric mode (angle of the rotation carriage).
The values displayed are set in manufacturing.
Please refer to Technical Service for possible changes.

52/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

8. SERVICE PROCEDURES
8.1 Plastic and Metal Covers

A
F
B

E
K
D

G C

L
H
N J
I
M

A Vertical Upper Top cover Two Allen screws on top and two on the back
B Vertical Upper Bottom cover Two Allen screws on the back and one in front
C Mirror Cover Four Allen screws on the back
D Display cover Two Phillips screws on top
E Drive Lower Cover Two Phillips screws at bottom, tube head side
Remove D, F, J, I, G, H
F Driver Upper Metal Cover Four Allen and four Phillips screws
Remove D and G first
G Drive External Cover Two Phillips
H Sensor/Cassette Cover Four Phillips (in case of system with cassette
holder 2 Phillips in addition)
I Inner Tube Head Cover Two Phillips screws
J Outer Tube Head Cover Two Phillips screws
K Cephalostat Upper Cover Clamped. Pull upward to remove
L Cephalostat Lower Metal Cover Five Allen screws
M Ceph Imager Inner Cover 12 Phillips screws (remove outer cover first)
N Ceph Imager Outer Cover 6 Phillips screws

Version 3.1 6968000210 53/84


PantOs 16/DG XP - Service & Installation Manual

O Metal Shield on Power Supply Unit Two Phillips screws on top


P Metal Shield on Control Unit Four Phillips screws in front
Q Metal Shield on Regulation Board Two Phillips screws at sides
R Metal Shield on Power Switching Board Four Phillips screws on top

54/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

8.2 Replacement of Low Voltage Transformer


• Remove Vertical Upper Top Cover and Metal Shield on Power Supply
(see § 8.1 at page 53).

PRIMARY SIDE SECONDARY SIDE


Ub S2
BLACK
BROWN

WHITE
Nb S1
Ua TF
RED
ORANGE

BLUE
Na TF

LOW VOLTAGE TRANSFORMER

• The low voltage transformer has


two primary windings which are
combined to connect to 115 or
230 V mains.
• The wires on primary side are
connected to XT301 on the
power supply following the in-
dicated colors.
• TF on block XT301 refer to the
wires of the thermal fuse.
• The wires at low voltage on
secondary side are connected to
block XT302.

Version 3.1 6968000210 55/84


PantOs 16/DG XP - Service & Installation Manual

8.3 Replacement of Power Supply Unit


XC301
XT301 XT302

XC308
VT303
XT305

TEST
15V
38V
24V

EXP
ZH
SA301

XC303 XC304 XC302

• Remove Vertical Upper Top Cover and Metal Shield on Power Supply
(see § 8.1 at page 53)
• XC305: input of mains voltage from output of mains filter
• XC303: 300 VDC output voltage to power switching unit
• XC304: exposure lights (24 VDC, used for digital ceph)
• XC302: multipole cable
• TEST 15, 38, 24: test point for low voltages 15, 24, and 38 VDC
• ZH: Filament supply signal
• EXP: Exposure signal
• XC301: Hand Switch connection with ground wire to XC302
• XC308: laser supply at – 3.1 V
• VT303: jumper to supply electric brakes at 24 V
• SA301: ON for normal use, OFF to prevent exposures on demo units
• Perform filament calibration to compensate supply variances

56/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

8.4 Replacement of Control Unit Board

DISPLAY

CONNECTOR
CONTROL PANEL XC11 XC9

SERIAL XC12
MICROSWITCHES XC14
PORT
FOR MODE SETTING XC10
JUMPER XC20
SIGNAL
THERMAL
MOTOR SENSOR XC18
X AXIS XC19
MOTOR MOTOR INPUT
ROTATION CASSETTE 38VDC

TO TO TO TO
XC5 XC6 XC7 OUTPUT XC101
38VDC

• Remove Display Cover, Driver Upper Metal Cover and Metal Shield on
Regulation Board, unlock the control unit module, swing it out open
and remove Metal Shield on Control Unit (see § 8.1 at page 53).
• Disconnect all the cables, mount the new control unit board, and then
connect back all the cables as indicated.
• Configure the unit with mode setting (see § 3.8 at page 23).
• New firmware to be uploaded with the loading tool via the serial port
and null modem cable. Code of loaded software can be read with ser-
vice function (see § 7.6 at page 49)
• Load Rotation and Translation Offset values for panoramic (see § 0 at
page 47
• Load CAM setting value (see § 7.8 at page 49)
• Load Rotation and Translation Offset values for Ceph mode (see § 7.16
at page 52)
• Perform connections of cables as indicated.

Version 3.1 6968000210 57/84


PantOs 16/DG XP - Service & Installation Manual

8.5 Replacement of Regulation Board

PROBES 8-11
XC 101

SA101

XC 102 PROBES 1-7

• Remove Display Cover (D), Driver Upper Metal Cover (F) and Metal
Shield on Regulation Board (Q), unlock the control unit module, swing
it out open (see § 8.1 at page 53)
• Disconnect all the cables, mount the new regulation board, and then
connect back all the cables
• Perform filament calibration (see § 8.7B at page 61).

58/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

8.6 Replacement of Power Switching Unit Board


XC202
XC203 XC201

XC204

XT202 XT201

• Remove Display Cover (D), Driver Upper Metal Cover (F), Metal
Shield on Regulation Board (Q), unlock the control unit module, swing
it out open and remove Metal Shield on Power Switching Board (R),
Inner Tube Head Cover (I) and Outer Tube Head Cover (J) (see § 8.1 at
page 53).
• XC201: 300VDC power supply to HV converter
• XC202: Flat cable from/to regulation board
• XC203: Multipole cable
• XC204: 38VDC supply to digital sensor
• XT201: Supply to HV transformer in the Tube Housing Assembly
• XT202: Filament supply (ZH1, ZH2) and feedback signals (mA, kV)
• Disconnect all the cables and replace the board
• Filament calibration is required (see § 8.7B at page 61).

Version 3.1 6968000210 59/84


PantOs 16/DG XP - Service & Installation Manual

8.7 Replacement of Tube Housing Assembly


The procedure includes:
• Hardware replacement
• Filament calibration
• Beam alignment for Pan Solo systems
• Beam alignment for Pan Ceph systems
A Hardware replacement
• Remove Display Cover, Driver Upper Metal Cover, Metal Shield on
Regulation Board, unlock the control unit module, swing it out open
and remove Metal Shield on Power Switching Board (see § 8.1 at page
53).
• Disconnect all wires of the Tube Housing Assembly (THA) from the
Power Switching Unit (PSWU) and ground wire from the frame.
ƒ Filament Supply (two wires: ZH1, ZH2)
ƒ Ground connection (GND)
ƒ mA feedback (mA)
ƒ kV feedback (kV)
ƒ TER Thermal Sensor
ƒ High Voltage Supply (two wires plus ground : UN1, UN2, GND)
• Dismount the tube-head and mount the new one.
• Connect the wires to the PSWU.

FEEDBACK POWER SWITCHING UNIT


SIGNALS
FILAMENT kV & mA
SUPPLY

THERMAL
SENSOR HIGH VOLTAGE
TER SUPPLY

60/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

B Filament Calibration
This calibration is required every time a Tube-Head or a Regulation Unit
or a Power Switching Unit is replaced.
TEST
POINT
mA

GROUND
SWITCHES
SA101

TRIMMER TEST
mA POINT
RP103 FDB

TRIMMER
FDB
RP104

WARNING: wait for cooling time between one exposure and


the other. Make sure the voltmeter to be used is calibrated
• Enter Service program “X-RAY TEST”,
“CONTINUE” mode.
• Set the anodic voltage to 61 kV and anodic current to 10 mA.
• Switch to ON both switches SA101.
• Monitor the voltage to Ground and mA point.
• Tune trimmer RP103 (mA) for 6V during exposure.
• Switch to OFF both switches SA101.
At this point, only in case this procedure is done after replacement of
either the Regulation Unit or of the Power Switching Unit (or of both of
them) calibration of the feedback signal (FDB) has to be performed:
ƒ Monitor the voltage to Ground and FDB test point.
ƒ Set the anodic voltage to 85 kV and anodic current to 10 mA.
ƒ Tune trimmer RP104 (FDB) for 10V during exposure.
• Exit Service Program “X-RAY TEST CONTINUE” mode (return key).

Version 3.1 6968000210 61/84


PantOs 16/DG XP - Service & Installation Manual

C Panoramic Field Check


• Shoot X-ray, by using Service function “X-RAY TEST – CONTINUE
mode” at 70 kV. The room should be dark enough to be able to see the
marked lines. Watch the light of the beam from below the X-ray gene-
rator, at a convenient distance and with proper radiation shielding. Al-
ways wait for cooling time between one exposure and the other.
• Check the beam for rota- UP UP UP

tion and align it to be ver-


tical by loosening the
three screws around the
body of the collimator.
• The irradiated field in ver-
tical height must stay be-
tween the upper and lower
lines. Align it using the
screws on the collimator.
• The irradiated field must
be symmetrical in relation
to the central line of the
radiation detector. If not, a
horizontal shift has to be done by unlocking the tube head and acting on
the shift tuning screw for side movements.
D Panoramic Beam Rotation
• On a Pan Solo system the correction of rotation is done by loosening
the 4 screws holding the collimator and by rotating it as needed.
• On Pan Ceph system the correction of rotation is done by loosening the
3 screws holding the disk with collimators and by rotating it as needed.
E Panoramic Beam Vertical Shift
• Both on a Pan Solo or on Pan Ceph system the vertical correction of the
beam is obtained by raising or lowering the supporting plate of the col-
limator.
ƒ Pan Solo Vertical Alignment: Loosen the 4 screws of the collimator
move it up or down then tighten the screws to block it again.
ƒ Pan Ceph Vertical Alignment: select the Adult Pan slit, release the
blocking screws of the holding support then move up or down the
holding cylinder. At the end tighten all the screws loosen.

62/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

F Panoramic Beam Horizontal Shift


• Both on a Pan Solo or on Pan Ceph system the horizontal correction of
the beam is obtained by shifting on the left or on the right the tube
housing assembly by loosening the 4 screws holding it.
A shift tuning screw is available to fine tune the position.

SUPPORTING
PLATE

SHIFT
TUNING
SCREW

COLLIMATOR
HOLDING
SCREWS

G Cephalometric Field Check


• Select the AP collimator (18x24S or 8x10S on display)
• Place a loaded cassette on the cassette holder in the AP position.
• Rotate the Cephalostat with ear supports directed along the beam and
opened to the maximum extent.
• Expose and check (1) the alignment of the exposed area on the film
(unexposed edge should be seen all around
the picture) and (2) that the ear supports for
side radiographs (LL projection) are aligned.
H Correction of Ceph
Field Deviations
• Deviations in vertical direction are corrected
by raising or lowering the Cephalostat by
acting on the leverage screw on the mount-
ing plate SCREW FOR
CEPHALOSTAT
• Horizontal deviations are corrected fine tun- HEIGHT SETTING
ing of the angle of the rotation carriage by
means of the service function.

Version 3.1 6968000210 63/84


PantOs 16/DG XP - Service & Installation Manual

I Ear Plugs Horizontal Shift

The shift of the inner (smaller) HORIZONTAL


ring to the right is obtained by TUNING SCREW
loosening the horizontal tuning
screw on the top of the cephalos-
tat .
J Ear Plugs Vertical Shift

VERTICAL
TUNING SCREW

The raise of the inner (closer)


ring is obtained by loosening the
vertical tuning screw.

K Ceph Soft Tissue Filters


Two positions for Latero Lateral (LL) exposures with soft tissue filter.
The actual shift of the copper filter can be adjusted to the desired extent
for Adult (A1) or Child (A2) by loosening the blocking screws on the
proper window of the collimator disk .

The following table indicates typical distance of copper filters at center.


Collimator Type Format Measure (mm)
cm 18x24 A1 7.8
cm 18x24 A2 6.8
inch 10x8 A1 13
inch 10x8 A2 12

64/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

8.8 Replacement of Position Sensors


The procedure includes:
• Description of sensors
• Position Sensor for Rotation
• Position sensor for Translation
• Position Sensor for Pan Ceph collimator
A Electro-Optical Sensors
The sensors to monitor the movement of the
rotation carriage are made of three electro-
optical devices on a small board.
Each one sensor is made of a light source
(light emitting diode) and of a device sensi-
tive to light (photo-transistor).
The presence of light detected by the photo-
transistor, or the absence of the same due to
a non transparent object intersecting the
A
beam of light, define two conditions.
In the side picture A, B, and C are the posi- B
C
tions where the three sensors are located.
Using a plate with proper profile we can
generate 3 sets of signals depending on
presence or absence of light for the 3 sen-
sors.
In order to generate the necessary move-
ment signals for the rotating carriage a met-
al disk is used with shaped external profile
and a full set of small holes to interact with
the sensitive area of the sensors.
Sensor A works aligned on the holes all
around the plate to generate a pulsed signal
during carriage rotation.
Sensor B is positioned to detect the middle
position of the carriage, when it reaches half of the rotation.
Sensor C is used to detect both beginning (starting point) and end of car-
riage rotation (final point).

Version 3.1 6968000210 65/84


PantOs 16/DG XP - Service & Installation Manual

ROTATION
SENSOR
XC10

XC9 TRANSLATION
SENSOR

B Replacement of Position Sensor Rotation


• Switch the system OFF
• Remove Vertical Upper Top Cover (see § 8.1 at page 53).
• Disconnect connector XC9 (rotation).
• Remove the two holding screws and replace the sensor module.
• Align the electro optical device (A) on to the holes of the encoder disk.
• Place back connector XC9 and switch the unit ON.
• Enter Service Menu TEST SENSOR (see § 7.14 at page 52) and select
AXE ROTATION.
• Verify the OFFSET value to be correct and in case load new value.
• Record date of change and new offset value.
C Replacement of Position Sensors for Translation
• Switch the system OFF
• Remove Vertical Upper Top Cover (see § 8.1 at page 53).
• Disconnect connector XC10 (translation).
• Remove the two holding screws and replace the sensor module.
• Align the electro optical device (A) on to the holes of the linear guide.
• Place back connector XC10 and switch the unit ON.
• Enter Service Menu TEST SENSOR (see § 7.14 at page 52) and select
AXE X.
• Verify the OFFSET value to be correct and in case load new value.

66/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

D Replacement of Position Sensors


for Pan Ceph Collimator
• Switch the system OFF.
• Remove Display Cover (D),
Driver Upper Metal Cover (F),
Metal Shield on Regulation
Board (Q).
• Disengage tube head rotation by
removing the blocking pin.
• Rotate the tube head to remove BLOCKING PIN
the cover screws on left and right
sides.
• Remove Inner Tube Head Cover
(I). COVER SCREWS RIGHT SIDE
• Remove the Knob of Frankfurt
Light and remove Outer Tube
Head Cover (J).
• Remove two screws holding the
position sensor and replace it
• Put back the blocking pin and
switch the system ON for test
and sensor alignment. COVER SCREWS LEFT SIDE
• Once set proceed in reverse order
to put back covers.

Version 3.1 6968000210 67/84


PantOs 16/DG XP - Service & Installation Manual

9. ALARM MESSAGES
9.1 Warning Messages
Warning messages inform about the occurrence of certain events which
might require the user to operate.
A Low Battery
The internal battery is generating low voltage and needs to be replaced in
about one month from first warning message.
B X-ray generator high temperature
The system is cooling down and the requested exposure would exceed the
loading capacity. Wait until ready.
C CAM init
Wait for system to complete initialization until ready.
D Adjust Arm Position
The rotation arm was moved out of position with unit switched OFF and
cannot be initialized.
• Turn the system OFF.
• Place manually the arm in “patient entrance” position by rotating and
shifting the carriage slowly.
• Turn system ON and perform initial reset.
E Wait PC ready
The communication line to the computer is missing or the application pro-
gram is not in acquisition mode. Perform necessary correction.
This warning message is not active in test mode (service functions).
9.2 Error Messages
In the following indications are given on possible corrective actions in re-
lation to the error messages generated or other abnormal situations.
A Error 1 kV reference signal out of range
Measure the voltage value at test point 11 (reference kV, 10 kV/V from
6.1 to 8.5 V) on the Regulation board. Replace the Control Unit if the sig-
nal is not corresponding to the display setting.
B Error 2 mA reference signal out of range
Measure the voltage value at test point 12 (reference mA, 1 V/mA from 4
to 10 V) on the Regulation board. Replace the Control Unit if the signal is

68/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

not corresponding to the display setting.


C Error 3 Carriage rotation encoder failure
Position sensor to be replaced or aligned.
D Error 4 Carriage displacement encoder failure
Position sensor to be replaced or aligned.
E Error 5 Carriage rotation sensor failure
Position sensor to be replaced or aligned.
F Error 6 Carriage displacement sensor failure
Position sensor to be replaced or aligned.
G Error 7 Cassette displacement sensor failure
Check the micro-switch on the cassette carriage and align or replace it.
H Error 9 Tube-head temperature exceeding limit
If the tube-head is at high temperature wait for it to cool down.
I Error 11 Exposure aborted during irradiation
This is usually a consequence of the “dead man functionality” which ter-
minates irradiation upon release of the exposure hand-switch. In case the
error appeared without specific request by the operator a malfunction
could be present in the cable of the hand-switch, in its connector or in the
Control Unit itself. Fix or replace accordingly.
J Error 12 Exposure aborted before irradiation
This is usually a consequence of the “dead man functionality” which ter-
minates irradiation upon release of the exposure hand-switch. This error is
generated by interruption during movement before irradiation. In case the
error appeared without specific request by the operator a malfunction
could be present in the cable of the hand-switch, in its connector or in the
Control Unit itself. Fix or replace accordingly.
K Error 13 38 V DC supply out of range
Check the value with a volt meter on the test point of the Power Supply
Board and replace it if the value is not in the range 34 to 42 V.
Conversely, if the electrical value is in the range, with the service function
“Test Voltages” verify on the display the abnormal readout of the control
unit and replace it in case. Check also the integrity of the connecting
cables.

Version 3.1 6968000210 69/84


PantOs 16/DG XP - Service & Installation Manual

L Error 14 24 V DC supply out of range


Check the value with a volt meter on the test point of the Power Supply
Board and replace it if the value is out of the range 23to 25 V. Conversely,
if the electrical value is in the range, with the service function “Test Vol-
tages” verify on the display the abnormal readout of the control unit and
replace it in case. Check also the integrity of the connecting cables.
M Error 15 15 V DC supply out of range
Check the value with a volt meter on the test point of the Power Supply
Board and replace it if the value is out of the range 14.5 to 15.5 V. Con-
versely, if the electrical value is in the range, with the service function
“Test Voltages” verify on the display the abnormal readout of the control
unit and replace it in case. Check also the integrity of the connecting
cables.
N Error 16 5 V DC supply out of range
Check the value with a volt meter on the test point of the Regulation Unit
Board and replace it if the value is out of the range 4.75 to 5.25 V. Con-
versely, if the electrical value is in the range, with the service function
“Test Voltages” verify on the display the abnormal readout of the control
unit and replace it in case. Check also the integrity of the connecting
cables.
O Error 17 High voltage failure
This message is generated when the inverter current or tube voltage are in-
creasing abnormally.
Failure on the Power Switching Unit or in the tube-head itself is possible.
Replace the Power Switching Unit board first and the tube-head in case.
P Error 19 Exposure time exceeded
This can be caused by a malfunction in the control unit. Replace it.
Q Error 20 Exposure aborted after irradiation
This is usually a consequence of the “dead man functionality” which ter-
minates irradiation upon release of the exposure hand-switch. This error is
generated by interruption during movement after irradiation. Re-take is not
needed, unless in the case of quitting the exposure in the middle of the
TMJ procedure.
In case the error appeared without specific request by the operator a mal-
function could be present in the cable of the hand-switch, in its connector
or in the Control Unit itself. Fix or replace accordingly.

70/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

R Error 21 No tube current


No or low anode current. This error can be caused
• by interruption of High Voltage power supply (right side of the Power
Switching Unit board)
• or by interruption of the filament power supply (left side of the Power
Switching Unit board)
• or by interruption of the driving circuits (in the regulation board)
• or by fault in the tube-head (input transformer or multiplying circuits).
Check the supplying circuit in the bushing and the flat cable between the
regulation and power switching boards.
On the Power Supply unit check the switch SA 301 (must be ON), fuses
FU 304 and FU 305. If they are OK, switch exposure and check the driv-
ing signals for HV regulator and inverter (points 1, 2, 6 and 7 on the Regu-
lation Unit) and the HV inverter supplying voltage (between points D3
and S5 in the Power Switching Unit) – should be in range from 50 to 100
volts (depending on kV and mA setting). Then measure the primary circuit
of HV transformer.
If all is correct, the circuit between the secondary of HV transformer and
tube is broken.
Depending on the location of the fault, consider to replace the Power
Switching Unit board, the Regulation Unit board, the connecting cables or
the tube-head.
This error can be also activated by sparking in the tube, when a very short
over-current signal switches off the HV regulator only, but is not trapped
by the Control Unit in order to activate error 17.
In this case try to perform the tube seasoning (de-gassing by making a
number of exposures at low kV) or replace the tube-head if necessary.
S Error 23 Filament current out of range
It appears when filament current value is exceeding the permitted range.
Power switching unit to be replaced first, regulation unit second, and con-
trol unit in case.
T Error 25 Thermal sensor failure
The thermal sensor is broken or disconnected. Verify continuity and pres-
ence of thermal sensor jumper on the control unit board. Fix accordingly.
In case of failure of the thermal sensor inside the tube-head, the tube-head
should be replaced. Alternatively the functionality for minimization of
waiting time for cool down can be disabled through “Thermal Settings” in
Service Mode and control loading manually using the cooling curves.

Version 3.1 6968000210 71/84


PantOs 16/DG XP - Service & Installation Manual

U Error 26 Battery fault


Battery voltage too low. Replace battery.
V Error 32 Control unit fault
Replace control unit board.
9.3 Mains Fuse Blows at X-ray On
Check resistors R321, R322, R323, R324, R325 (connected to the relay in
the Power Supply Unit board).
If defective, charging of capacitors C305, C306, C307, C308 could not be
done properly during the preheating time. The capacitors will be charged
directly as soon as the radiation is started with the consequence of burning
the fuse due to the very high charging current.
Causes of resistor breaking could be:
• Pulse overloading of the resistor during the capacitors charging.
• Too high leakage current of capacitors C305, C306, C307, and C308.
• Short circuit in the input of the power switching unit (JED26 - XC201).
Replace the Power Supply Unit board or in case the involved module
causing the short circuit to occur.
9.4 Other Alarm Messages
• Internal Software Error, with error code xxh ?? ISE ??
and address yyyyyyyh. Switch the system OFF C: xxh A:yyyyyyyh
and ON again to try to overcome the problem.
• Over-Current. Over Current
This occurs with a short circuit of a stepper mo- Axe: A B C
tor:
A = Axe Rotation, B = Axe X, C = Axe cassette
Switch the system OFF and remove the short circuit on the motor, then
switch the system ON.
• TRIPOS, microprocessor’s operating system error.
• Application abort.
Take a note of the message generated and refer to the Technical Service.

72/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

10. MAINTENANCE
Maintenance to the equipment must be performed at least once a year by
qualified personnel.

Always disconnect the system from mains supply using the


general mains switch for the room where the equipment is
located before starting maintenance, and connect back to
mains only for activities requiring it.

The following actions to be done:


• Check the condition of the cables for vertical movement of the carriage.
If some wires of the strand are broken, it is necessary to replace them.
• Perform lubrication of accessible parts subject to movement.
• Avoid using liquid or spray detergents which may enter into the equip-
ment and cause corrosion.
• Complete check of system performance (kV, mA, mAs, s).
• Verify proper alignment of aiming lights (laser beams) and clean the
output window of each source.

The laser aiming system is a Class I laser product. Avoid dur-


ing operation unnecessary exposure of the eyes of the patient
or of the operator to the laser radiation. Also pay attention
that the laser beams are not intercepted by any optical device.
• Check for proper working conditions of all mechanical and electrical
safety features.

WARNING. Any defective item affecting a safe use must be


repaired or replaced.

Only original certified components and spare parts must be used for re-
pairs or replacements.
Correction of damages to the identification labels to be handled with the
Manufacturer.
Any defect or non-compliance must be reported promptly to the Manufac-
turer or to its Local Agent.

Version 3.1 6968000210 73/84


PantOs 16/DG XP - Service & Installation Manual

11. SPARE PARTS

11.1 Positioning & Test


Items E Line Laser Module

A Bite Block

Code
76 680 10010 Code 76 190 24195

F Test Phantom

Code
76 680 10030
B Chin Rest

Code
76 680 10020 G Radiation Detector
Analog System

C Temporal
Resting Bars Code 76 680 10040
H Radiation Detector Digital
System
Code
76 680 30050 Code
76 680 10050

I Alignment Plate
D Ear Plug Ceph
Code
79 680 10100

Code
76 680 30060

74/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

11.2 Power Supply &


Control Modules E Power cable Inverter
A Dual Voltage Supply
Unit 115-230V

Code 76 680 20200


F Signal Cable Inverter

Code 76 680 20010


Version 13.11.2007 Code 76 680 20300

B Dual Voltage Supply G Low Voltage Transformer


Board 115-230V

Code 76 190 24376


For Dual Voltage Supply
Code 76 190 24760
Version 13.11.2007 H Mains Switch
C Mains Connection Module

Code
76 680 20090 Code 76 190 26190
I EMI Filter

D Mains Cable

Code
76 680 20400 Code 76 190 24730
20A 120/250V 50-60Hz

Version 3.1 6968000210 75/84


PantOs 16/DG XP - Service & Installation Manual

J Set of Fuses 115 V O Microswitch


Code Code
76 190 24475 76 680 20080
Set of 10
for each type

K Set of Fuses 230 V P Regulation Unit


Code
76 190 24470
Set of 10
for each type
Code 76 190 24113
L Control Unit Board Version 26.05.2004
Q Power Switching Unit

Code Codice76 680 20170


M Control Panel Code 76 190 24600
Version 15.11.2007
R Tube Housing
Assembly

Code 76 680 20030


N Position Sensor

Code
76 680 20020
Code 76 680 20160

76/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

S Hand Switch D Counterweight Cable


Code Holder
76 190 25590 Code
76 190 24710
T Coiled Cable
E Side Bearing
Code Code
76 190 25580 76 680 30010

U Kit Remote Hand-Switch F Eccentric Bearing

Code Code 76 680 30020


93 300 02000
G Ball Bearing Spare

Code
76 680 24860
11.3 Mechanical tems
A Counterweight Cables H Stepping Motor
Code
76 190 24385 Code
Set of 2 76 680 20060

B Counterweight Pulleys

Code
76 190 24640 I Magnetic Brake

C Adjustment Weights
Code
76 190 24740 Code 76 190 24650
2x0.5 kg
& 1x1 kg J Brake Pushbutton

Code
76 680 20070

Version 3.1 6968000210 77/84


PantOs 16/DG XP - Service & Installation Manual

Appendix A
Alarm Messages

Warning Messages
A Low Battery
B X-ray generator hot
C CAM init
D Adjust Arm Position
E Wait forPC ready

Num Error Messages


1 kV reference signal out of range
2 mA reference signal out of range
3 Carriage rotation encoder failure
4 Carriage displacement encoder failure
5 Carriage rotation sensor failure
6 Carriage displacement sensor failure
7 Cassette displacement sensor failure
9 Tube-head temperature exceeding limit
11 Exposure aborted during irradiation
12 Exposure aborted before irradiation
13 38 V DC supply out of range
14 24 V DC supply out of range
15 15 V DC supply out of range
16 5 V DC supply out of range
17 High voltage failure
19 Exposure time exceeded
20 Exposure aborted after irradiation
21 No tube current
23 Filament current out of range
25 Thermal Sensor Failure
26 Battery fault (low voltage)
32 Control unit fault

78/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

Appendix B
Icons
Compliance to Compliance to
IEC Type B Equipment European Commu- North American
nity Requirements Safety Requirements

The unit is ON OFF (disconnected See the Annexed


from mains supply) Documentation

X-ray indicator
ON (connected
The unit is READY (ionizing radiation
to mains supply)
emission)

Radiation Emission Radiography Digital Detector


Light Push Button indicator (green light)

Program selected
Program selection
1 through 7 for Pan, USB Interface
panoramic and cepha-
8 through 10 for (orange light)
lometric
Ceph
Patient size selection:
Standard panoram-
small, medium, large Fuse
ic
extralarge

Increase parameter
kV level
mA or mAs level
Child Panoramic ~ Alternate Current

Decrease parameter
kV level Half Panoramic Left Protective Earth
mA or mAs level

Carriage shift right Half Panoramic


Inherent Filtration
for patient positioning Right

Dispose properly
Carriage shift left
Anterior Dentition Do not abandon in the
for patient positioning
environment

Entity of carriage
TMJ Open and Fragile, Handle With
displacement
Closed Mouth Care
in mm

Type of collimator
selected (Pan, Child, Maxillary Sinus Fear of Humidity
AP, LLA, LLB, APXL)

Arm Sequential Ceph


Up, Do Not Overturn
Movement Latero Lateral

Patient alignment
Ceph
laser beams Stacking Limit Number
Antero Posterior
(15 s ON)

Test mode without rad-


Wrist
iation emission

Size Adult/Child:
Alarm Reset small, medium,
large extra-large

Version 3.1 6968000210 79/84


PantOs 16/DG XP - Service & Installation Manual

Appendix C
Technical Data
EXTRAORAL RADIOGRAPHIC SYSTEM
FDA Classification Class II (21CFR872-1800, 21CFR892.1650)
IEC Classification: Class I, type B
Power Supply: 230 V ± 10%, 50/60 Hz, 8 A, slow fuse
115 V ± 10%, 50/60 Hz, 16 A, slow fuse
Mains Resistance: < 2 ohm at 230 V, < 1 ohm at 115 V
X-ray generator: Multipulse at 40-80 kHz
Anode Voltage: 61 - 85 kV ± 10%, constant potential
Anode Current: 4 - 10 mA ± 10%, direct current (DC)
Focus size: 0.5 IEC 336
Inherent Filtration > 2.5 mm Al @ 85 kV
Leakage Radiation < 50 mR/h (0.43 mGy/h) @ 100 cm, 85 kV, 10 mA
Column Heigth 220 cm/87”
Displacement 86 cm/34”, from 90 to 176 cm (35 to 69”)
Vertical Movement Manual adjustment with electrical brake
Patient Positioning Motorized horizontal displacement
Centering Light 3 lasers: Lateral, Median sagittal, Frankfurt
Centering Reference Bite block and chin rest for edentulous, tunable
temporal resting bars (option)
Source Image Distance Pan 51.3 cm/20.2”, Ceph 165 cm/65”
Vertical Magnification Pan 27%, Ceph 10%
Panoramic Projections: Adult standard with constant vertical magnification
on dental arch(15 s), Child at reduced dose (13 s),
Half Dentition Left (8 s), Half Dentition Right (8 s),
Anterior Teeth (7 s), TMJ habitual occlusion and
opened at maximum extent (4 x 2.5 s), Frontal View
of Maxillary Sinuses (13 s)
Ceph projections Antero Posterior, Latero Lateral Adult,
Latero Lateral Child, Carpus
Cooling pause: Automatically controlled, 7 min maximum
Weight: Pan Solo 220 kg/485 lb, Pan Ceph 240 kg/529 lb
DIGITAL SENSORS
Image Size Pan 15x30 cm, 3000 x 1500 pixels
Image Size Ceph 25x25 cm, 2000x2000 pixels
A/D Conversion 12 bits
LASER LIGHTS
FDA Classification Class I
Wavelength 650 nm
Output Power < 0.11 mW at 100 mm

80/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

Appendix D
Labels

Version 3.1 6968000210 81/84


PantOs 16/DG XP - Service & Installation Manual

Appendix E
Cooling Curves
COOLING CURVE OF X-RAY TUBE
kJ

20

18

16

14

12

10

0
0 1 2 3 4 5 6

Time (min)

COOLING CURVE OF TUBE HOUSING ASSEMBLY

kJ
120

100

80

60

40

20

0
0 15 30 45 60 75 90 105 120 135 150 165 180

Time (min)

82/84 6968000210 Version 3.1


PantOs 16/DG XP - Service & Installation Manual

Appendix F
Fuse Table
FU 303
CONFIGURATION 230 V 115 V
CONNECTOR PLUG
T8A T 16 A
F 501 5x20 6.3x32
T 1.6 A T 3.15 A
FU 303 5x20 5x20
FU 304 FU 305 F 6.3 A F 6.3 A
FU 304 5x20 5x20
F 6.3 A F 6.3 A
FU 305 5x20 5x20
POWER SUPPLY UNIT IS TOP PLASTIC COVER FIXED WITH
AT TOP OF VERTICAL 2 ALLEN SCREWS ON TOP
CARRIAGE 2 ALLEN SCREWS ON THE BACK
TO ACCESS REMOVE
TOP PLASTIC COVER
AND TOP METAL COVER
TOP METAL COVER FIXED WITH
2 PHILLIPS SCREWS ON TOP
F 501

FUSE F 501 IS ON
THE RIGHT SIDE
BEHIND MIRROR
PLASTIC COVER

MIRROR PLASTIC
COVER IS FIXED
WITH 4 ALLEN
SCREWS
ON THE BACK

Version 3.1 6968000210 83/84


PantOs 16 xp Ceph
PantOs DG xp Ceph
Panoramic and Cephalometric
Dental X-Ray Equipment
Service & Installation Manual
English Edition
Version 3.1

*6968000210*

You might also like