SSI Technologies - Application Note DT-AN4

Digital Fluid-Trac™ (DFT series)


Troubleshooting Guide

For correct operation, the Digital Fluid-


Trac™ DFT must be mounted properly with
no obstructions between the sensor and
media being monitored.
1) If using the ¾” adapter, verify the
yellow acoustic absorber disc (see
photo below) is properly installed
inside the adapter.

Press and release


Refresh/Select to
update the level
reading.
A level sensor that is mounted or sitting
other than level will create a reduction in
sensor performance. As shown in the figure
*Measurement Technology below, the amount of returned sound energy
The Digital Fluid-Trac™ Model DFT drum is dependent on mounting and/or sitting
level gauge uses ultrasonic technology to angle.
generate a high-frequency sound wave and
measures the time for the echo to reflect off
the target fluid’s surface and return. The
distance from the Digital Fluid-TracTM sensor
to the liquid is calculated based on the
speed of sound.
The Digital Fluid-Trac™ Model DFT is Incorrect Correct
factory programmed to work on ANSI MH2
standard 55 gallon, 30 gallon, or 15 gallon
drums.
*The Digital Fluid-Trac™ is NOT intended for use with
gasoline or other volatile liquids that produce a vapor
pressure under normal operating temperature (0°C - 50°C).
SSI TECHNOLOGIES, INC. © Copyright August 12th 2016
Controls Technologies Division SSI Technologies, Inc.
3200 Palmer Dr. All Rights Reserved
Janesville, WI 53547-5011 Revision 3
Phone: (608)757-2000
www.ssitechnologies.com
Page 1
SSI Technologies - Application Note DT-AN4
Digital Fluid-Trac™ (DFT series)
Troubleshooting Guide

2) Verify the Digital Fluid-Trac™ DFT 5) If the display doesn’t show 55 gallons,
mounting is parallel to the liquid. The drum you will have to place the unit into the
must be stored in an upright position. 55-gallon drum mode. Refer to
Ensure that barrel lid is not extremely “Measurement Mode Selection”
warped, creating an unlevel condition. procedure below.

Measurement Mode Selection

Distance Reading a) Enter the Select Mode: Press and


hold the Refresh/Select button down
for 7 seconds. The display will show
“Liter/Gallon” and flash (units mode).

b) Release the Refresh/Select button


and then press and hold for 7
seconds to select liters or gallons.
The chosen setting will be displayed
3) Verify there are no obstructions between and the barrel volume will begin to
the sensor head and the media being flash (volume mode).
measured. If something is between the
sensor and the media being tested, the c) Press and release the button to scroll
sound wave will be reflected off the through the other settings until
object and not the media resulting in desired mode is flashing. For gallons
incorrect level readings. setting – 55, 30, or 15 gallons can be
selected. For Liter setting – 208, 114,
4) Verify the unit is set to measure 55 gallon
or 57 liters can be selected.
drums.
a) Place the Digital Fluid-Trac™ flat d) Press and hold the Refresh/Select
on a table top. button for 5 seconds to select the
b) Press the Refresh/Select button. desired volume setting.

c) The display should read 55 e) After desired volume mode stops


gallons (±4 gallons, depending on flashing, release the button. The unit
which adaptor(s) being used) will then go back into level monitoring
mode.

SSI TECHNOLOGIES, INC. © Copyright August 12th 2016


Controls Technologies Division SSI Technologies, Inc.
3200 Palmer Dr. All Rights Reserved
Janesville, WI 53547-5011 Revision 3
Phone: (608)757-2000
www.ssitechnologies.com
Page 2
SSI Technologies - Application Note DT-AN4
Digital Fluid-Trac™ (DFT series)
Troubleshooting Guide

f) The level reading can be refreshed by c) The display should read


momentarily pressing the Refresh / approximately 35 gallons or
Select button. The back light will light 14.2mA depending on output
up and will be on for 15 seconds. type.
When the back light shuts off, the
level reading will be displayed. Simulate an Empty Reading
(0 Gallon / 4mA)
6) If the previous four conditions are met, a) Place the Digital Fluid-Trac™
perform the following bench test. Note 32 inches above the table top.
that current output (mA) is specific to the Make sure the sensor is level
DFT-200 series barrel gauge. with the table top.
b) Press the Refresh/Select
Simulate a Full (55 gallons / 20mA) button.
Reading c) The display should read
a) Place the Digital Fluid-Trac™ approximately 0 gallons or
flat on a table top. 4mA, depending on output
type
b) Press the Refresh/Select
button.
c) The display should read 55 7) For DFT 200 series only
gallons (±4 gallons) or 20mA (DFT-200, DFT-210, DFT-220)
(±1.2mA), depending on
which output and adaptor(s) a) If cable is connected in back
of unit, but backlight is off, it
being used.
indicates loss of power or
Simulate a 35 Gallon (14.2mA) Reading ground from power supply.
a) Place the Digital Fluid-Trac™ Display may have erroneous
12 inches above the table top. reading during this condition.
Make sure the bottom of the b) When using batteries
sensor is perpendicular to the “REFRESH/SELECT” button
table top. must be pushed to update unit
b) Press the Refresh/Select to current barrel level reading.
button.

SSI TECHNOLOGIES, INC. © Copyright August 12th 2016


Controls Technologies Division SSI Technologies, Inc.
3200 Palmer Dr. All Rights Reserved
Janesville, WI 53547-5011 Revision 3
Phone: (608)757-2000
www.ssitechnologies.com
Page 3