C306 Drug Box Controler


Specifications

User Manual

Thank you very much for purchasing our drug box controler.
Please read the “Operating Instructions” and “Warranty” before
operating this unit to assure proper operation. After reading these
documents, be sure to store them securely together with the “Warranty”
within reach for future reference.

Warning: Before operating the unit, be sure to read carefully and


fully understand important warnings in the operating instructions.

BIOBASE BIODUSTRY (SHANDONG) CO.,LTD.

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

C306 Drug Box Controler Specifications


1.Description

The controller controls the medication tank and has two temperature sensors (one control cabinet
temperature, one display, one of which can override its function when one of them fails), a door
switch alarm that controls cooling, evaporating the fan motor, And the cabinet lights; Buzzer
alarm, power alarm display temperature.

1.1 Technical Parameters

Work Temp.:-45.0~99.9oC
Temp.Accuracy:±0.5oC(@-10~20 oC),other ±1oC
Working Environment:-20~60 oC,humidity 20%RH~80%RH,No condensation
Storage environment: -30 ~ 70 oC, humidity 20% RH ~ 80% RH, non-condensing
Power supply : 220VAC ± 20%, 50 / 60Hz
consumption: not more than 5W
Film button
ROHS instruction: Fulfillment
Relay capacity
Refrigeration Relay: 30A / 240VAC (single phase 1.5HP / 240VAC motor and below)
Fan relay: 10A / 250VAC
Lamp relay: 5A / 250VAC
1.2 Drug box controller design test standards
Electromagnetic compatibility (EMC) test and measurement techniques Electrostatic discharge
immunity test
Electromagnetic compatibility Radiated electromagnetic fields Radiation immunity test GB / T
17626.3-2006
GB / T17626.4-2008 Electromagnetic compatibility test and measurement technology Electrical
fast transient burst immunity test
GB / T17626.5-2008 Electromagnetic compatibility test and measurement techniques surge
(impact) immunity test
GB / T17626.11-2008 Electromagnetic compatibility testing and measurement techniques Voltage
dips, short interruptions and voltage variations Immunity test
GB4343.2-1999 Requirements for Electromagnetic Compatible Household Appliances, Power
Tools and Similar Appliances - Part 2: Immunity - Product Class Standards
GB / T13639-2008 industrial process measurement and control system with analog input digital
indicator
GB / T2423.1-2001 Environmental testing for electric and electronic products Part 2: Test

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

methods Test A: Low temperature


GB / T2423.2-2001 Environmental testing for electric and electronic products Part 2: Test methods
Test B: High temperature
GB / T2423.4-1993 Basic environmental testing procedures for electric and electronic products
Test Db: Alternating hot and humid test method
GB / T2423.8-1995 Environmental testing for electric and electronic products Part 2: Test methods
Test Ed: free fall

2、 Control Description

The control panel consists of a red LED display and touch-type membrane keys, as shown in
Figure 1

2.1 Over-temperature alarm signs

If the cabinet temperature display sensor exceeds the set alarm temperature, over-temperature
alarm, over-temperature alarm indicator light

2.2 Door ajar alarm signs

If within the set time, the door switch is not closed, the door open alarm indicator light.

2.3 Defrost status sign

Defrost box when the state, the defrost status indicator light.

2.4 Cabinet temperature display

According to the sample temperature sensor display cabinet, showing the internal temperature of
the box for the 3 half-digit display, all-time was -88.8.

2.5 Mute button

Buzzer alarm box (including cabinet temperature over-temperature alarm, door alarm, sensor

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

failure alarm, etc.), press the mute button, the buzzer to stop tweets. (Press the mute button to turn
off the beeper of abnormal status alarm only. If the trouble is removed, the next time the abnormal
beeper needs to continue alarming.)

2.6 Set button

In normal mode, press the SET key for more than 3 seconds to display the set temperature value
for more than 5 seconds to access the parameter menu. (Need a password to enter the setup menu)
When the instrument is powered on, press the "3" key for more than three seconds at the same
time, so that the parameters will be restored to their default values. The digital tube will display
"CF".
In the parameter setting mode, press this key to alternately display the parameter value and the
parameter name, and save the setting if the pressed time is more than 3 seconds. Set the set
temperature, press the button to save the set value.

2.7 Up button

In the parameter setting mode, you can move to the next parameter, or increase the parameter
value.

Set the set temperature, increase the set temperature value.

2.8 Down button

Normal working hours, press this button for 3 seconds then blinking display control temperature

for 5 seconds.

In parameter setting mode, move to the previous parameter or decrease the parameter value.

Set the set temperature, reduce the set temperature.

2.9 Forced defrost key

Normal operation, press the forced defrost key for more than 5 seconds, you can enter or exit the
defrosting state.

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

2.10 Light switch key

In normal operation, press the light switch key, you can open or close the cabinet

light.

3、Input and output description:

3.1 Relay output

Refrigeration Relay: 30A / 240VAC (single phase 1.5HP / 240VAC motor and below).
Fan relay: 10A / 250VAC.
Lamp relay: 5A / 250VAC

3.2 Power supply

220VAC ± 20%, 50 / 60Hz

3.4 Sensor input

Cabinet temperature control sensor: measuring the temperature inside the cabinet for compressor
start-stop control.
Cabinet temperature display sensor: measuring cabinet temperature for the thermostat display, if
the cabinet temperature display sensor exceeds the set alarm temperature, over temperature alarm,
high temperature alarm indicator flashes, buzzer alarm. Control and display sensor functions are
interchangeable.
Door switch detection port: connect the door switch, if within the set time, the door switch is not
closed, the door alarm indicator flashes, the buzzer sounds the output.

3.5 Temperature display after power off

After the power supply is cut off, the controller control function is invalid, and the backup power
supply is used. If the backup power is charged for 10 hours, the power-down symbol "Pd" will be
displayed alternately with the current temperature and alarm, then stop for 30 seconds, display
again, and the digital tube will stop displaying after 8 hours of cyclic operation.

4 .Control parameter
The control parameters are divided into basic parameters and advanced parameters.Firstly,the
basic parameters is “St” and it is used into setting the working temperature by users.The advanced
parameters have the password protected system and you need to enter the correct password to
access.
4.1Access to basic (St) parameters

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

The setting method:

Under normal circumstances, press the “Setting Key” for 3 seconds, then the digital tube will
show the current setting point.
Press the “Increasing Key ” to increase the number or press the “Decreasing key ” to decrease
the number until it reach the expected number.

Press the “Setting Key ” again to confirm the new number, save the new number and return
to the normal working mode.
4.2 Access to advanced type parameter

Press the “setting key ” to more than 5S, and the digital tube will display parameter code "PS"
(password).

Press the “setting key ” to display "00" into the password setting mode.
Press the “Increasing Key or Decreasing key to make the number on "-15”.

d. Press the “setting key ” to confirm the password ,the digital tube will display "/ 0"
e. Press the “Increasing Key or Decreasing key to change the number.

f.①Press the “setting key ” to confirm the parameter type and this parameter type’s first
parameter will display,
②Press the “Increasing Key or Decreasing key to change the number.

③Press the “setting key ” to display the corresponding parameter.


④Press the “Increasing Key ” to increase the number or press the “Decreasing key ” to
decrease the number.

⑤Press the “setting key ” to save the modified number temporarily and return to the display
parameter.
If you want to modify the parameter,please do the procedure①--④ again.

⑥If you press the“ Setting Key ” more than 3 seconds,it will save the modified parameters
and return to the display parameter category.
g.If you want to change the parameters, please do theprocedures e--f again.

h.If you press the “Setting Key ” more than 3 seconds or you do not press the key in 60
seconds ,it will exit the parameter setting program.
Warning:
If you do not press the key in 60 seconds,the changes of all the parameters (temporarily stored in
RAM) will not be valid and it will revert to previous settings.
When you set the parameter r1 , r2 , St,AL,AH , press the key in the short time to adjust the
parameter number by 0.1, and hold the parameter for a long time, then adjust the parameter
number continuously every 0.5 second.

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

5 .The types and descriptions of parameters


5.1 Temperature sensor calibration
Temperature sensor measurements can be calibrated and stabilized by using the following table .

Parameter description Min Max Unit Default


The stability of
/0 1 15 - 4
sensor measurement
The offset of Cabinet
/1 temperature display -5.0 5.0 oC 0.0
sensor
The offset of Cabinet
/2 temperature control -5.0 5.0 oC 0.0
sensor

/0 The stability of sensor measurement


This parameter defines the coefficient of temperature measurement stability.The smaller number
speeds up the reaction time of the sensor and it is sensitive to interference.The larger number
slows down the reaction time of the sensor, but it is not sensitive to interference, which is the
stable value.

/1The offset of Cabinet temperature display sensor


This parameter is used to calibrate the measurement temperature of the cabinet display system.The
numerical value of the symbol indicates that the temperature of the cabinet shows the temperature
of the sensor increase (positive) or decrease(negative) offset.
/2 The offset of Cabinet temperature control sensor
This parameter is used to calibrate the measurement temperature of the cabinet control system.The
numerical value of the symbol indicates that the temperature of the cabinet control the temperature
of the sensor increase (positive) or decrease(negative) offset.

5.2 Temperature control parameter


According to the actual measurement temperature and parameter setting of the cabinet temperature
control sensor to control to start and stop the compressor.
Parameter Description Min Max Unit Default
Set
St r1 r2 oC 5.0
temperature
Set min
r1 -45.0 r2 oC 2.0
temperature
Set max
r2 r1 90.0 oC 14.0
temperature
Temperature
rd gap between 0.1 10.0 oC 1.0
the lower limit
ru Temperature 0.1 10.0 oC 0.5

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

gap between
the higher limit
The description of parameters
Set temperature
The users set the working parameters
R1, set minimum temperature
Allowed Minimum setting working temperature number
R2 ,set maximum temperature
Allowed set maximum working temperature number
Rd, ru temperature gap
When the temperature of the cabinet temperature controls the temperature of the sensor <= st-rd,
the compressor stops working;When the temperature of the cabinet temperature controls the
temperature of >=St+ ru, the compressor starts to work.
Note: When the temperature control sensor is out of control, the cabinet
temperature shows that the sensor replaces the temperature control sensor to
control the compressor.When the temperature control sensor of the cabinet and
the temperature of the cabinet show that the sensor is out of order, the controller
can control the compressor's opening (reference parameter C2) by setting the
proportional period or memory ratio.

5.3 The protection of compressor


This part is to protect the normal operation of the box and the normal operation of the compressor.
Parameter Description Min Max Unit Default
The compressor’s delay
c0 after 0 100 Min 1
charged with electricity.
c1 Minimum downtime 0 100 Min 0
compressor operation mode
when the cabinet
c2 temperature control sensor 0 1 - 0
and the cabinet temperature
display the sensor faults.
The compressor’s working
c4 time when Control sensor 0 99 Min 30
is breaking down.
The compressor’s
c5 downtime when Control 1 99 Min 15
sensor is breaking down.
c0: The compressor’s delay after charged with electricity.
When the equipment is charged with electricity , the compressor and fan do not start
immediately and they need to delay at least c0 (unit: Min).

c1:Minimum downtime
The interval from compressor to restart requires at least c1 (unit: Min) to start.

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

C2:Compressor operation mode when the cabinet temperature control sensor


and the cabinet temperature display the sensor faults.
When the controller operate normally , the compressor’s starting time and stopping time in 24
hours is recorded.When the temperature control sensor and the cabinet temperature display sensor
fail, if c2 = 0, the compressor will operate according to the storage compressor in the controller;If
c2 = 1, the compressor starting time is c4, the stopping time is c5.

c4:The compressor’s working time when Control sensor is breaking down.


When the temperature control sensor and the cabinet temperature display sensor fail, the
compressor’s starting time work together with c2 and c5.

C5:The compressor’s downtime when Control sensor is breaking down.


When the temperature control sensor and the cabinet temperature display sensor fail, the
compressor’s stopping time work together with c2 and c4.

5.4 Defrosting
This part is the parameters related to defrosting.
parame
Description Min Max Unit Default
ter
d0 Defrosting Mode 0 1 - 0
d1 Defrosting Cycle 1 199 Hour 8
dp Defrosting’s Maximum Time 0 199 Min 10
d0:Defrosting Mode
d0 = 0 shielding defrost function
d0 = 1 defrosting cycle,and it is determined by d1.

d1
Defrosting Cycle,interval between two defrostings

dp
defrosting maximum time is end time.

5.5 Alarm
In order to avoid misjudgments and ensure the controller to be running normally .Alarm
parameters are set in the controller in that way the user is prompted when the fault occurs.

Parameter Description Min Max Unit Default


(Cabinet
temperature
AL -20.0 50.0 oC 2.0
display sensor)Low
Temperature Alarm
(Cabinet
AH -20.0 50.0 oC 10.0
temperature

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

display
sensor)High
Temperature Alarm
(Cabinet
temperature
Ad display 0 99 Min 1
sensor)Alarm
Delay
charging alarm
AS 0 199 Min 30
delay
Door opening
AO 0 250 S 60
alarm delay
AL :(Cabinet temperature display sensor)Low Temperature Alarm
when the temperature in the cabinet (the temperature of the cabinet temperature display sensor)
is lower than AL, if the alarm delay is met parameter Ad, the buzzer chirps.
AH:(Cabinet temperature display sensor)High Temperature Alarm
When the temperature in the cabinet is higher than AH, if the alarm delay is met parameter Ad, the
buzzer sounds.
Ad:(Cabinet temperature display sensor)Alarm Delay
When the temperature reaches the alarm temperature, the alarm will not sound immediately until it
exceed the Ad time.If the alarm condition is relieved during the delay, the alarm delay is canceled
and the counter is cleared.
If Ad=0 is set, alarm immediately.
AS:Charging alarm delay
After charging,,the temperature of the cabinet (the cabinet temperature display sensor) alarm will
also be shielded, , only exceed the As time ,we can judge alarm.
AO:Door opening alarm delay
When the door opens, the door opening alarm sign on the control panel will flicker and the buzzer
sounds.

5.6 Fan management of evaporator


In order to make the evaporator fan work normally, this part of the parameter is set for the fan
management of the evaporator.The fan needs to be connected to the door switch. When the door
opens, the evaporator will stop.
Parame
Description Min Max Unit Default
ter
F0 Fan control type 0 1 - 0
Under normal circumstances, the evaporator fan can be managed in the following ways:
A, the fan runs all the time
B,the fan control with compressor simultaneously.
The description of parameters.
F0 = 0 fan runs all the time
F0 = 1 fan and compressor synchronous control
Normally, the fan runs according to the pattern set above.

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

5.7 Other parameters


Parameter description Min Max Unit Default
Software version
H3 - - - -
number
Machine Model
H4 0 1 - 0
selection
H3:Software version number
This parameter can be used to check the software version number, which cannot be set. It is the
data that comes with the system.
H4:Machine Model selection
Different machines have different internal default parameters . After modifying H4 , all
parameters are changed to the default after you press the “setting key ”.

6 .Alarm code

Alarm
Code Error Description

E0 Cabinet temperature display sensor failure (display)


E1 Cabinet temperature display sensor failure (control)
Cabinet temperature display sensor low temperature
Lo
alarm
Cabinet temperature display sensor high temperature
Hi
alarm
When the controller shows the above alarm code, check the corresponding equipment or parts
according to the code, or check whether the controller's parameter setting is appropriate.

7.Parameter List

Parameter Description Min Max Unit Defaults


/0 Sensors measure stability 1 15 - 4
The cabinet temperature shows the
/1 -5.0 5.0 C
o
0.0
sensor offset
Cabinet temperature control sensor
/2 -5.0 5.0 C
o
0.0
offset
St set temperature r1 r2 oC 5.0
r1 Set the minimum temperature -45.0 r2 C
o
2.0
r2 Set the maximum temperature r1 90.0 C
o
14.0
rd Lower temperature limit 0.1 10.0 C
o
1.0
c0 Compressor delay after power 0 100 Min 1
c1 Minimal downtime 0 100 Min 0
c2 Compressor operating mode when 0 1 - 0

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

the cabinet temperature control


sensor and the cabinet temperature
display sensor are both faulty
Compressor operating time when
c4 0 99 Min 30
sensor failure is detected
Control compressor failure
c5 1 99 Min 15
compressor downtime
d0 Defrost mode 0 1 - 0
d1 Defrost cycle 1 199 Hour 8
dp Defrost the maximum time 0 199 Min 10
Cabinet temperature sensor low
AL -45.0 90.0 C
o
2.0
temperature alarm temperature
Cabinet temperature sensor
AH temperature alarm temperature -45.0 90.0 C
o
10.0
display
Cabinet temperature sensor alarm
Ad 0 99 Min 1
delay
AS Alarm delay after power on 0 199 Min 30
AO Door open alarm delay 0 250 S 60
F0 Fan control category 0 1 - 0
H3 Software version number - - - -
H4 Model selection 0 1 - 0

8. Connection methods

www.biobase.cc
BIOBASE BIODUSTRY (SHANDONG) CO., LTD.

Biobase Biodustry (Shandong) Co., Ltd.

Registered address: No.2717 Gongyeyi Road, Mingshui Economic Developing Zone,


Zhangqiu, Jinan, Shandong,China

Manufacture address:No.2717 Gongyeyi Road, Mingshui Economic Developing Zone,


Zhangqiu, Jinan, Shandong,China

Production license NO.: Shandong Food and Drug Equipment Production Permit No.
20120119

TEL:+86-0531-81219888 FAX:+86-0531-81219880

EMAIL: EXPORT@BIOBASE.CN

www.biobase.cc