YY 0645-2018

1 Blood Pump Flow rate


Accury: less than ±10% of Nominal Range

Test method of Blood


Pump Flow rate Pressure:-200mmHg±50mmHg

2 Fluid Removal

Overall precision of
less than ±200mL in 8h
weight loss:

3 Syringe Pump
Accury: ±0.2 mL/h or 5% of the displayed
value,take the larger one
4 Venous pressure monitor

Accuracy: less than ±10mmHg

couldn't be set below 10 mmHg in


lower alarm limit
treatment

5 Arterial Pressure Monitor

Accuracy: less than ±10mmHg

6 TMP
Accuracy: less than ±20mmHg

7 Working Noise

≤ 62dB
Performances of Acusmart Verdict Medica comment

We can check if, reducing maximum


< +- 10% up to -150mmHg of suction pressure pump speed to 300ml/min, we
F
<20% up to -250mmHg of suction pressure remains in that limit. Otherise a
software modification is needed+

The total treatment fluid indicated


here is the sum of replacement
volume + effluent volume.
Considering a CVVH at high flow
0.3% with respect to total treatment fluid F
rates, where you infuse 3000ml/h
and remove 3200ml/h, you get an
overall volume of 5l in 8h, that gives
a maximum error of 150g

<0.2 mL/h of displayed value of 5 mL/h Uncertain We will repeat internal test to
confirm requested accuracy

The accuracy can be bettere


expressed as the worst case between
±4mmHg or ±5%. Considerng the
normal operative range of [-200,
+200]mmHg, that definitions fulfulls
< ±15 mmHg at the displayed value of -300 or your needs. When pressure goes
300 mmHg; F below -200mmHg for a long time
< ±4 mmHg at the displayed value of 0mmHg there can be small deviations due to
small air leakages, but that does not
happen in CRRT or hemofiltration,
where pressure values below
200mmHg raise alarms and hence
pumps are stopped

The minimum venous alarm


threashold can be modified if
needed, although 10mmHg is
unknown Uncertain generally considered high: there are
patients where systemic pressure is
very low and even 5mmHg is too
high.

< ±15 mmHg at the displayed value of -300 or


300 mmHg; F See discussion for Venous Pressure
< ±4 mmHg at the displayed value of 0mmHg
This is a calculated pressure (Pven +
Ppre)/2 - Puf). Therefore theoretical
unknown Uncertain
error should be considered the worst
case between ±12mmHg or 15%

For sure below 62, but we have only


measurements done by ourselves,
inside our building, therefore
unknown Uncertain
overextimating the noise. If needed
we can ask for a measurement from
a certified lab
considering a CVVHD with high Considering a SCUF,
flow rates, let's say 6000ml/h of with a weight loss rate
dialysate and 6500 of effluen. In of 200ml/h:
this case 8h is not realistic: we Total fluid in 8h=
consider 4h as maximum time, 1600ml.
with a maximum error of 150g Error = 5g