Inthe alan eee
a) Measurement of the current in the alarm cir-
cuit
Purpose of the teats it must be guaranteed that,
even in the case of a malfunction, the current in
the circuit does not exceed the value of 1 mA DC.
Bovisone
~
Sonata
b) Trip test
Purpose of the test: to check that the insulation
‘monitoring device is functioning correctly, in
other words, that the alarm signal is activated
when the insulation resistance value drops below
50ka.
<
Instrument: Miliammeter.
Procedure: measurement of the current in the
alarm circuit can be carried out in conditions of
dead short earth fault by inserting the milliamme-
ter in series with the conductor that connects the
device to the equipotential node and connecting
‘one of the conductors of the isolated circuit di-
reetly to earth (ig. 4.2).
[=3]
Instrument: Rheostat,
After disconnecting the loads, each conductor of
the circuit powered by the secondary of the insu-
lating transformer is connected - one at a time -
with the equipotential node using the rheostat
(fig. 4.4) Fault simulation is implemented by re-
ducing the resistance of the rheostat to a value R
Seen cornet)