Professional Documents
Culture Documents
Characterization Tools of Nanomaterials (Part 2)
Characterization Tools of Nanomaterials (Part 2)
Nanomaterials
(Part 2)
The high resolution of EM images results from the use of electrons (which have very short wavelengths) as the source of
illuminating radiation.
The electron microscopes – the transmission EM (TEM), the scanning EM (SEM) and reflection electron microscope (REM).
The transmission electron microscope is used to view thin specimens (tissue sections, molecules, etc) through which electrons
can pass generating a projection image.
TEM is analogous to light microscope, TEM is used, to image the interior of target.
Conventional scanning electron microscopy depends on the emission of secondary electrons from the surface of a specimen.
SEM image is formed by scanning a focused electron beam onto the surface of the specimen in a raster* (low energy
secondary electrons, high energy back scatter electrons, X-rays and even photons) pattern.
Size of the raster at the specimen is much smaller than the viewing screen, the final picture is a magnified image of the
specimen. (*a rectangular pattern of parallel scanning lines followed by the electron beam on a computer monitor)
Advantage of Electron microscope
1. Magnification and higher resolution – as electrons rather than light waves are used, it can be used to analyze
structures which cannot otherwise be seen. The resolution of electron microscopy images is in the range of up to 0.2
nm, which is 1000x more detailed than light microscopy.
2. Diverse applications – Electron microscopy has a diverse range of applications in many different fields of research
including technology, industry, biomedical science and chemistry. Examples of applications include semiconductor
inspection, computer chip manufacture, quality control and assurance, analysis of atomic structures, and drug
development.
3. High-quality images – With proper training, an electron microscope operator can use the system to produce highly
detailed images of structures which are of a high quality, revealing complex and delicate structures that other
techniques may struggle to reproduce.
• A scanning electron microscope (SEM) is a type of electron microscope that produces images of a sample by
scanning the surface with a focused beam of electrons.
• The electrons interact with atoms in the sample, producing various signals that contain information about
the surface topography and composition of the sample.
• The electron beam is scanned in a raster scan pattern, and the position of the beam is combined with the
intensity of the detected signal to produce an image.
• In the most common SEM mode, secondary electrons emitted by atoms excited by the electron beam are
detected using a secondary electron detector (Everhart–Thornley detector).
• The number of secondary electrons that can be detected, and thus the signal intensity, depends, among
other things, on specimen topography. Some SEMs can achieve resolutions better than 1 nanometer.
• Specimens are observed in high vacuum in a conventional SEM, or in low vacuum or wet conditions in a
variable pressure or environmental SEM, and at a wide range of cryogenic or elevated temperatures with
specialized instruments.
• 3D image of material and the sample should not be hydrated.
Specimen must be coated with gold or
palladium
SEM arrangements
Topography
Crystallography or
magnetic properties
Electron source
High current density (more
electron per unit area per unit
time)
A Wehnelt cap has the shape of a topless, hollow cylinder. The
bottom side of the cylinder has an aperture (through hole)
located at its center, with a diameter that typically ranges from
200 to 1200 μm. The bottom face of the cylinder is often made
from platinum or tantalum foil.
Detection of secondary electrons:
• The most common imaging mode collects low-energy (<50 eV) secondary electrons that are
ejected from conduction or valence bands of the specimen atoms by inelastic scattering
interactions with beam electrons.
• Due to their low energy, these electrons originate from within a few nanometers below the
sample surface.
• The accelerated secondary electrons are now sufficiently energetic to cause the scintillator to
emit flashes of light (cathodoluminescence), which are conducted to a photomultiplier outside
the SEM column via a light pipe and a window in the wall of the specimen chamber.
• Backscattered electrons (BSE) consist of high-energy electrons originating in the electron beam, that are
reflected or back-scattered out of the specimen interaction volume by elastic scattering interactions with
specimen atoms.
• Heavy elements (high atomic number) backscatter electrons more strongly than light elements (low
atomic number), so appear brighter in the image. BSEs are used to detect contrast between areas with
different chemical compositions.
• Dedicated backscattered electron detectors are positioned above the sample in a "doughnut" type
arrangement, concentric with the electron beam, maximizing the solid angle of collection.
• BSE detectors are usually either scintillators or semiconductor types (CMOS). When all parts of the
detector are used to collect electrons symmetrically about the beam, atomic number contrast is
produced.
• Strong topographic contrast is produced by collecting back-scattered electrons from one side above the
specimen using an asymmetrical, directional BSE detector; the resulting contrast appears as illumination
of the topography from that side.
Transmission Electron Microscopy
• Transmission electron microscopy (TEM) is a microscopy technique in which a beam of electrons is
transmitted through a specimen to form an image.
• The specimen must be an ultrathin section less than 100 nm thick or a suspension on a grid.
• An image is formed from the interaction of the electrons with the sample as the beam is transmitted through
the specimen. The image is then magnified and focused onto an imaging device, such as a fluorescent
screen, a layer of photographic film, or a sensor such as a scintillator attached to a charge-coupled device.
• Transmission electron microscopes are capable of imaging at a significantly higher resolution than light
microscopes, owing to the smaller de Broglie wavelength of electrons. This enables the instrument to
capture fine detail—even as small as a single column of atoms, which is thousands of times smaller than a
resolvable object seen in a light microscope.
• Transmission electron microscopy is a major analytical method in the physical, chemical, and biological
sciences. TEMs find application in cancer research, virology, and materials science as well as pollution,
nanotechnology, and semiconductor research, but also in other fields such as paleontology and palynology.
Tungsten filament
(heated , 20KVolt)
Electromagnetic lens
A single beam of
electron to focus on
sample
Ultra thin
Electromagnetic lens
sample
• The information is gathered by "feeling" or "touching" the surface with a mechanical probe.
Piezoelectric elements that facilitate tiny but accurate and precise movements on (electronic) command
enable precise scanning. Despite the name, the Atomic Force Microscope does not use Nuclear force.
• For imaging, the reaction of the probe to the forces that the sample imposes on it can be used to form an
image of the three-dimensional shape (topography) of a sample surface at a high resolution.
• This is achieved by raster scanning the position of the sample with respect to the tip and recording the height
of the probe that corresponds to constant probe-sample interaction. The surface topography is commonly
displayed as a pseudo color plot.
Piezoelectric materials can produce electric energy upon application of mechanical stress. A commonly
known piezoelectric material is quartz. The mechanism involves development of electric charge due to
movement of electron upon application of stress.
• Several forces are measured in AFM, some of which include van der Waals forces, mechanical forces, chemical
bonding, electrostatic forces, magnetic forces, capillary forces, etc
• In contact mode, a tip that is attached to a cantilever is scanned over the sample surface, while the force between
tip and sample is measured.
• Contact mode involves the tip making continuous contact with the surface of the sample as it moves across the
sample.
• The contours of the sample are measured either by the deflection of the cantilever or by the feedback signal
maintaining a constant position of the cantilever.
• A more flexible cantilever is generally used in contact mode due to it being prone to noise and therefore to keep
the interaction force low.
• A piezoelectric element oscillates the cantilever. The detector then records the deflection of the cantilever relative
to its equilibrium position and converts it into an electrical signal.
• An AFM generates images by
scanning a small cantilever
over the surface of a sample.
• The cantilever has a sharp tip that scans over the sample surface, by forming an attractive force between the surface
and the tip when it draws closer to the sample surface. When it draws very close making contact with the surface of
the sample, a repulsive force gradually takes control making the cantilever avert from the surface.
• During the deflection of the cantilever away from the sample surface, there is a change in direction of reflection of
the beam, and a laser beam detects the aversion, by reflecting off a beam from the flat surface of the cantilever.
• The Atomic Force Microscope (AFM) takes the image of the surface topography of the sample by force by
scanning the cantilever over specimen.
• Depending on how raised or how low the surface of the sample is, it determines the deflection of the beam, which
is monitored by the Positive-sensitive photo-diode (PSDP).
• The microscope has a feedback loop that controls the length of the cantilever tip just above the sample surface,
therefore, it will maintain the laser position thus generating an accurate imaging map of the surface of the image.
In no contact mode, the cantilever is oscillated close to its resonant frequency, while the amplitude of the
oscillation is measured.
As the cantilever/tip assembly approaches the surface, the interaction changes the resonant frequency, which in
turns changes the oscillation amplitude.
The change in the amplitude is the interaction that will be probed and generate the image.
The position of the cantilever is measured with the help of a laser, which reflects on the top of the cantilever just
above the tip to a set of photodiodes.
AFM can generate high magnifications (up to atomic resolution) of almost any type of sample, as the interaction
force is generally small (less than the bonding force between atoms).
1. Modified tips which are used to detect the sample surface and undergo deflections
2. Software adjustments used to image the samples.
3. Feedback loop control – they control the force interactions and the tip positions using a laser deflector. the
laser reflects from the back of the cantilever and the tip and while the tip interacts with the surface of the
sample, the laser’s position on the photodetector is used in the feedback loop for tracking the surface of the
sample and measurement.
4. Deflection – The Atomic Force Microscope is constructed with a laser beam deflection system. The laser is
reflected from the back of the AFM lever to the sensitive detector. They are made from silicon
compounds with a tip radius of about 10nm.
5. Force measurement – the AFM works and depends highly on the force interactions, they contribute to the
image produced. The forces are measured by calculation of the deflection lever when the stiffness of the
cantilever is known. This calculation is defined by Hooke’s law, defined as follows:
F= -kz, where F is the force, k is the stiffness of the lever, and z is the distance the lever is bent.
Image formation
• This section applies specifically to imaging in Contact mode. For other imaging modes, the process is similar,
except that "deflection" should be replaced by the appropriate feedback variable.
• When using the AFM to image a sample, the tip is brought into contact with the sample, and the sample is
raster scanned along an x–y grid.
• Most commonly, an electronic feedback loop is employed to keep the probe-sample force constant during
scanning. This feedback loop has the cantilever deflection as input, and its output controls the distance along
the z axis between the probe support and the sample support.
• As long as the tip remains in contact with the sample, and the sample is scanned in the x–y plane, height
variations in the sample will change the deflection of the cantilever.
• The feedback then adjusts the height of the probe support so that the deflection is restored to a user-defined
value (the setpoint). A properly adjusted feedback loop adjusts the support-sample separation continuously
during the scanning motion, such that the deflection remains approximately constant. In this situation, the
feedback output equals the sample surface topography within a small error.
Applications
• The AFM has been applied to problems in a wide range of disciplines of the natural sciences, including solid-
state physics, semiconductor science and technology, molecular engineering, polymer chemistry and physics,
surface chemistry, molecular biology, cell biology, and medicine.
• Applications in the field of solid state physics include (a) the identification of atoms at a surface, (b) the
evaluation of interactions between a specific atom and its neighboring atoms, and (c) the study of changes in
physical properties arising from changes in an atomic arrangement through atomic manipulation.
• In molecular biology, AFM can be used to study the structure and mechanical properties of protein complexes
and assemblies. For example, AFM has been used to image microtubules and measure their stiffness.
• In cellular biology, AFM can be used to attempt to distinguish cancer cells and normal cells based on the
hardness of cells, and to evaluate interactions between a specific cell and its neighboring cells in a competitive
culture system. AFM can also be used to indent cells, and to study how they regulate the stiffness or shape of
the cell membrane or wall.
• In some variations, electric potentials can also be scanned using conducting cantilevers. In more advanced
versions, currents can be passed through the tip to probe the electrical conductivity or transport of the
underlying surface, but this is a challenging task with few research groups reporting consistent data
Advantages of Atomic Force Microscope Disadvantages of Atomic Force
Microscope
XPS analyzes the elements constituting the sample surface, its composition, and chemical bonding state by irradiating X-
rays on the sample surface, and measuring the kinetic energy of the photoelectrons emitted from the sample surface.
The energy of an X-ray with particular wavelength is known (for Al Kα X-rays, Ephoton = 1486.7 eV), and emitted electrons'
kinetic energies are measured. Hence, the electron binding energy of each of the emitted electrons can be determined by using
the photoelectric effect equation-
where E binding is the binding energy (BE) of the electron measured relative to the chemical potential, E photon is the energy of
the X-ray photons being used, E kinetic is the kinetic energy of the electron as measured by the instrument and ‘𝜙𝜙’ is a work
function-like term for the specific surface of the material, as an adjustable instrumental correction factor. It is a constant that
rarely needs to be adjusted in practice
X-ray photoelectron spectroscopy(XPS)
• XPS is routinely used to analyze inorganic compounds, metal alloys, semiconductors, polymers, elements, catalysts,
glasses, ceramics, paints, papers, inks, woods, plant parts, make-up, teeth, bones, medical implants, bio-materials,
coatings, viscous oils, glues, ion-modified materials, and many others.
• A typical XPS spectrum is a plot of the number of electrons detected at a specific binding energy. Each
element produces a set of characteristic XPS peaks. These peaks correspond to the electron configuration of the
electrons within the atoms, e.g., 1s, 2s, 2p, 3s, etc. The number of detected electrons in each peak is directly related to
the number of elements within the XPS sampling volume.
• To generate atomic percentage values, each raw XPS signal is corrected by dividing the intensity by a relative
sensitivity factor (RSF), and normalized over all of the elements detected. Since hydrogen is not detected, these
atomic percentages exclude hydrogen
Detection limits: Detection limits may vary greatly with the cross-section of the core state of interest and the
background signal level. In general, photoelectron cross sections increase with atomic number. The background
increases with the atomic number of the matrix constituents as well as the binding energy, because of secondary
emitted electrons.
The Raman Spectrophotometer
• Raman spectroscopy is a non-destructive laboratory technique, which provides information about chemical
structure, phase, polymorph, and crystallinity through interactions with molecules. To check Purity of Sample
Raman spectroscopy is related to scattering of radiation. The change in frequency during the time of the
scattering process, is due to the energy changed by interaction with molecular vibrations.
Scattered ray Δ𝝂𝝂 Raman =108/ λexcitation -108/ λRaman (wave length)
Results obtained from Raman spectrometers are graphically depicted as Raman spectra. Rotational Raman spectra,
vibrational Raman spectra, and Electronic Raman spectra can be obtained.
The intensity of the scattered light on the Y-axis is plotted against the energy or frequency of light on the
X-axis. The frequency is measured in wavenumber, which is the number of waves per cm, cm-1. X-axis
frequencies relative to that of laser are plotted as it is the shift in the energy of the light, which is the desired one.
With the Raman shifts and relative intensities of Raman bands of the material, the material can be identified.
Individual band changes as a band may narrow, broader or shift, or vary in intensity. These changes disclose
information about the stresses in the sample, changes in crystallinity, and the amount of material.
Variation of spectra with the position in the sample reveals the change in the homogeneity or uniformity of the
material.
Further, the spectrum can be analyzed at several arbitrary points and systemically measured at an array of points,
which validates the production of images of compression, stress and crystallinity.
Applications of Raman Spectroscopy:
END