Site Preparation Drawings

Table of contents
Sheet Type Subject

General
1 C-1 Cover sheet / Table of content
2 A-1 Site layout
3 AD-1 Schematic cross section(s)
Building works
4 X-1 Floor provisions (all disciplines)
5 X-2 Ceiling provisions (all disciplines)
6 X-3 Wall provisions (all disciplines)
7 X-4 Wall provisions (all disciplines)
Technical details
8 XD-1 Details 1 (all disciplines)
9 XD-2 Details 2 (all disciplines)

Brilliance CT Big Bore Radiology


All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

, Panama
Project
Typical configuration shown (not site specific) Revision history
H
G
F

Table of contents
E
D
C

Cover sheet
B

Sheet Subject
A
0 17-01-2018
Rev Date Drw Chk Proj. phase Description

Important notes

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

Quote No
Order No
Drawing
Philips contact Customer ID Technical ID

Philips Healthcare A3 mm

C-1
Sheet 1/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m Brilliance CT Big Bore
Mass Heat Noise
Resp No Description [kg] [W] [dB(A)]
Examination area
1m
A 1.1 Scanner Gantry (Brilliance CT big bore) 2043 5270 69
A 1.2 Patient table 404 - -
A 1.3 Laser Positioning Light wall mounted (LAP Dorado 3) 26 - -
A 1.4 Laser Positioning Light ceiling mounted (LAP Dorado 3) 26 - -
A 1.5 LAP Carinasim Touch screen ceiling mounted + PC 16 - -
Control area
A 2.1 Server Cabinet (CRC) 142 1040 55
A 2.2 Operating station (without desk) 11 290 -
B 2.3 Emergency power OFF switch - - -
(2700) (5600) A 2.4 LAP Carinasim Touch screen monitor + PC 5 - -
Technical area
Y A 3.1 Transformer (Teal Isotran Line Match) 274 650 50
Dimension origin
at finished wall, A 3.2 Console UPS 3kVA (optional) 70 249 55
1.3 3.1 near floor level.
General
B 4.1 Lead glass window - - -
B 4.2 Mains switch - - -
2.1 Examination area B 4.3 X-ray ON warning light - - -
Technical area 4.2 B 4.4 Earth reference terminal - - -
2.4
4.4

2000
1.4
2.2 4.1 3.2
Responsibilities

Brilliance CT Big Bore Radiology


A Delivered and installed by philips
X X B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Control area D (Pre) Delivered by customer and installed by Philips

(4600)
E Existing

1.2 1.1
2.3 Site requirements (release 4.2)
Mains power
· Supply configuration with Isotran: 3 phase delta, 3 wire (U1, V1, W1, PE1)into LM
4.3 3 phase wye, 4 wire (U2, V2, W2, N2, PE2, PE4 - electrode ground) output to CT gantry.
· Nominal Line Voltage with Isotran: 200, 208, 240, 380, 400, 415, 440, 480, 500 VAC delta
input to LM 480 VAC wye output to CT gantry.

, Panama
· Maximum/momentary power: 110 kVA.
· Mains resistance: 160 mOhm line to line max. 80kW scan at 480 VAC

Project
Remote service diagnostics
To establish this feature, a RJ45 type Ethernet connector must be
installed with access to the customer's network (100/1000 Mbps)

500 1980 Environmental requirements


4.3 1.5 1.3 Y Examination room:
Operating temperature range is 18°C-24°C
at 35%-70% relative humidity (non-condensing)
Temperature change should not exceed 4.1°C per hour.

Site layout
Sheet Subject
Control room:
Operating temperature range is 15°C-24°C
at 35%-70% relative humidity (non-condensing)
Temperature change should not exceed 5°C per hour.

Technical room:
Operating temperature range is 15°C-28°C
at 20%-80% relative humidity (non-condensing)
Temperature change should not exceed 5°C per hour.
Legend

0
Place to be defined

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

Walls Project notes


locally by customer.
· The feasibility of this project is based on site information (like drawings) provided to

Technical ID
New walls

17-01-2018
Philips. Philips cannot assume any liability for the accuracy of this information and the

Quote No
Order No
Drawing
consequences.
Existing doors/walls to be removed
· The equipment proposed in this room layout produces X-ray radiation. It is the
Beams or other building construction elements customer's responsibility to comply with the local regulation. Normally X-ray shielding has
to be applied to avoid hazards to patients, visitors and staff. Consult the Philips contact for A3 1 : 50 mm
Columns (materials and strength unknown) details and support.
A-1
Preferred service area Sheet 2/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Support structure. Designed and


built by customer / contractor
Support structure. Designed and Support structure. Designed and
built by customer / contractor built by customer / contractor

Suspended ceiling Suspended ceiling

2740 (min. 2590)

2740 (min. 2590)


?

Brilliance CT Big Bore Radiology


Schematic cross section A-A Schematic cross section B-B
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

, Panama
Project
Schematic cross sections
B

Sheet Subject
A A

0 P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

Technical ID
17-01-2018

Quote No
Order No
Drawing
Situation (scale 1:100) A3 1 : 50 mm

AD-1
Sheet 3/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

(5600)

2700
Y
400 Dimension origin
at finished wall,
2 4 6 Wall duct see X-3
near floor level.

450
Wall duct see X-3

1420
6

2150
2000
6 Wall duct see X-3

340

730
5 3

Brilliance CT Big Bore Radiology


X X
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

(4600)
1

, Panama
Project
500 1980
Y

Floor provisions
(all disciplines)
Sheet Subject
Item list:
Resp No Description Qty Detail Text ref
Floor covering Open cable duct with cover
B 1 Base plate provision for CT - XD-1 § 2.2
Shroud layer Through filled with cement or epoxy B 2 Wall opening 1 - -
B 3 200 x 60 Floor duct flush mounted - - § 3.3
Legend

0
B 4 130 x 60 Floor duct flush mounted - - § 3.3
Building floor

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

B 5 200 x 90 Cable outlet - - § 3.3


Walls
B 6 Cable outlet - - -

Technical ID
New walls

17-01-2018

Quote No
Order No
Drawing
Existing doors/walls to be removed Typical floor duct installation (not to scale) Responsibilities
A Delivered and installed by philips Works symbology
Beams or other building construction elements
B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
Columns (materials and strength unknown) C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works X-1
Preferred service area E Existing Equipment works Sheet 4/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

(5600)

2700
Y
500 1980 Dimension origin
at finished wall,
near floor level.
Wall duct see X-3

300
Wall duct see X-3
3
2

2000
2150
4 1

Brilliance CT Big Bore Radiology


467
X X

467
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

(4600)
1500
4
Building ceiling

, Panama
Project
Ceiling hanger
Wall duct see X-3
False ceiling Open cable duct

1980
500
Y Typical ceiling duct installation (not to scale)

Ceiling provisions
(all disciplines)
Sheet Subject
Item list:
Resp No Description Qty Detail Text ref

B 1 Ceiling provisions for mounting plate 3 XD-2 § 2.3


B 2 Wall opening 1 - -
Legend

0
B 3 130 x 60 Cable duct above suspended ceiling - - § 3.3

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

B 4 Power outlet (Philips dedicated) 2 - § 3.2


Walls
B 5 Network connection 1Gb, RJ45 1 - § 3.4

Technical ID
New walls

17-01-2018

Quote No
Order No
Drawing
Existing doors/walls to be removed
Responsibilities
A Delivered and installed by philips Works symbology
Beams or other building construction elements
B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
Columns (materials and strength unknown) C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works X-2
Preferred service area E Existing Equipment works Sheet 5/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Ceiling duct see X-2

Ceiling duct see X-2


Suspended ceiling Suspended ceiling
Suspended ceiling

9 7 2 1 6
8 6 8

2740 (min. 2590)

2740 (min. 2590)

2740 (min. 2590)


?

?
2840

2840

2840
900
4 4 5

900
5

>720
>300

300

300
5 4 7 3

Brilliance CT Big Bore Radiology


Floor duct see X-1
Elevation A Elevation B Elevation C
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Wall

, Panama
Open cable duct with cover

Project
Typical wall duct installation (not to scale)
B
A
Item list:

Wall provisions
Resp No Description Qty Detail Text ref

(all disciplines)
B 1 Ø 19mm Conduit - - § 3.3

Sheet Subject
C B 2 Lead glass window 1 - § 1.6
B 3 Emergency power OFF switch 1 - § 3.2
B 4 Power outlet (Philips dedicated) 3 - § 3.2
B 5 Network connection 1Gb, RJ45 3 - § 3.4
B 6 70 x 70 Wall duct surface mounted - - § 3.3
Situation (scale 1:100)

0
B 7 120 x 70 Wall duct surface mounted - - § 3.3

P-PAN000112
B Wall provisions for mounting plate 4 XD-2 § 2.4
© Koninklijke Philips Electronics N.V. 2018

8
B 9 Wall opening 1 - -

Technical ID
17-01-2018

Quote No
Order No
Drawing
Responsibilities
A Delivered and installed by philips Works symbology
B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works X-3
E Existing Equipment works Sheet 6/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

1 2
Suspended ceiling Suspended ceiling

2740 (min. 2590)

2740 (min. 2590)


(3200)

(3200)
3

>300
4

Floor duct see X-1


Elevation D Elevation E

Brilliance CT Big Bore Radiology


All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

, Panama
Project
Wall provisions 2
(all disciplines)
D

Sheet Subject
Item list:
Resp No Description Qty Detail Text ref

B 1 X-ray ON warning light 1 - § 3.2


B 2 Door switch 1 - -
B 3 Mains switch 1 - § 3.2

0
B 4 Power outlet (Philips dedicated) 1 - § 3.2

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

B 5 200 x 70 Wall duct surface mounted - - § 3.3


B 6 Earth reference terminal 1 - § 3.2

Technical ID
17-01-2018

Quote No
Order No
Drawing
E Responsibilities
Works symbology
A Delivered and installed by philips
Situation (scale 1:100) B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
Construction works
Electrical works
A3 1 : 50 mm

D (Pre) Delivered by customer and installed by Philips Mechanical works X-4


E Existing Equipment works Sheet 7/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Floor plate Floor plate
Total floor area
Y for waviness and
levelness boundary
>50.8mm embedment in concrete

>45mm embedment in concrete


Screed layer Screed layer

Hilti HIT-HY 200 with HIT-V

2743
Red Head RL-12 drop in anchor

610
110 / 8.8 X
bolt diameter: 12,7mm
bolt diameter: 12mm
drill bit size: 15.9mm
drill bit size: 14mm
Concrete tension
Concrete quality ≥C20 Floor waviness and levelness of exam room shall not
>13.8N/mm² exceed 6.3 mm when measured within 300 mm from
any point within the shaded area.

Gantry and patient table fixing detail Gantry and patient table fixing detail 610
(not to scale) (not to scale) 975
Delivered by Philips Not delivered by Philips 3658

3 1 2
Y Detail - Floor levelness and waviness (not to scale)

Base frame outline

Brilliance CT Big Bore Radiology


Cable entry area
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Base frame outline

1 2 4

183
140

, Panama
X

Project
Patient table baseplate Gantry anchor brackets
1105
632

1137 5 bolts total required. 8 bolts total required


836
951

1764
978
1103

Max. tension force = 14,7kN Max. tension force = 14,7 kN


1136

2407
Max. shear force = 20,4kN Max. shear force = 20,4kN

(all disciplines)
2604 Max. torque = 30Nm Max. torque = 30Nm

Sheet Subject
Details 1
Item list:
Resp No Description Qty Detail Text ref

294 181 A 1 Expansion anchor M12 13 - § 2.2

0
382 197 B 2 Drill hole for M12 anchor 13 - § 2.2

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

422 328 A 3 Base plate for gantry 1 - § 2.2

438 A 4 Base plate for patient table 1 - § 2.2

Technical ID
17-01-2018

Quote No
Order No
551 Always apply the Philips drill template

Drawing
Responsibilities
A Delivered and installed by philips Works symbology
B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
C (Pre) Delivered by Philips, installed by customer Electrical works

Detail - Scanner gantry/patient table base frame (scale 1:20) D (Pre) Delivered by customer and installed by Philips Mechanical works XD-1
E Existing Equipment works Sheet 8/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
220
60

1 2
220
60

12
75 75
1 2
12

Building structure
75 75

Support structure Mounting plate

75 75

Bottom of suspended ceiling

934

Brilliance CT Big Bore Radiology


934
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

220

Detail - Dorado Side view mounting plate


(scale 1:5)

, Panama
120
170

Project
120
170

Detail - Wall mounting plate for LAP laser Dorado

(all disciplines)
(scale 1:5)

Sheet Subject
Details 2
Detail - Ceiling mounting plate for LAP laser Dorado
(scale 1:5) Item list:

0
Resp No Description Qty Detail Text ref

P-PAN000112
© Koninklijke Philips Electronics N.V. 2018

B 1 M10 anchors Hilti 4 - § 2.4


B 2 Drilled holes M10 4 - § 2.4

Technical ID
17-01-2018

Quote No
Order No
Drawing
Responsibilities
A Delivered and installed by philips Works symbology
B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works XD-2
E Existing Equipment works Sheet 9/9
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED