Scanner Hardware Stopped Scan - Replace Collimator CAM (A - B) Motor and Coupling Assembly
Scanner Hardware Stopped Scan - Replace Collimator CAM (A - B) Motor and Coupling Assembly
Scanner Hardware Stopped Scan - Replace Collimator CAM (A - B) Motor and Coupling Assembly
Troubleshooting
https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=70832&sliceId=1 1/5
13/4/23, 11:44 Scanner Hardware Stopped Scan : Replace Collimator CAM (A/B) Motor and coupling assembly.
Cam A and B not driving and the Filter is also not driving
1. Check the incoming 48VDC on the CCB from test point 9 to 10. If the 48VDC is missing
troubleshoot backwards to the rotating 48VDC supply.
2. If the 48VDC is present and F1 and F2 are not blown on the CCB check the 24VDC from
test point 16 to 10. If the 24VDC is missing order the CCB, a cam motor and collimator filter
assembly.
3. Replace the CCB first. If after replacement the filter does not drive, replace the filter
assembly. If one of the cams do not drive replace the cam motor.
https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=70832&sliceId=1 2/5
13/4/23, 11:44 Scanner Hardware Stopped Scan : Replace Collimator CAM (A/B) Motor and coupling assembly.
Resolution
Details:
It is the service persons’ responsibility to use the appropriate system documentation to effect repairs
and prove the service action had the desired or intended effect, with no adverse effects.
Refer to Class C, Dir 5193575-800, Rev 5 >>Replacement > Gantry > Collimator > Cam Drive Motor
and Flex Coupling Replacement, which can be found on the Distributed Technical CDROM/DVD, or
online at:
Links to any needed documentation is found in the Product details as an attachment called
"ProductDocumentLinks."
NOTE: If the encoder counts are just a few ticks away from the commanded position. Check
and tighten motor/coupling set screws first.
Part Number
https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=70832&sliceId=1 3/5
13/4/23, 11:44 Scanner Hardware Stopped Scan : Replace Collimator CAM (A/B) Motor and coupling assembly.
5271284-3 (replaces 5271284-2, 5271284, 2404178) Collimator Motor with anti fretting
compound and Coupling Assembly
Comments:
-- FE ASSIST-- 5271284-3 is a Second Opinion part
Error Code
With this problem or issue, you may find one or more of the following errors:
Error: 260101023 The CCB has notified the system of an error condition which prevents scanning
Error: 260133154 CAM A Motion With Out Command Exceeded The Warning Tolerance
Error: 260133155 CAM (A or B) Motion With Out Command Exceeded The Error Tolerance
Error: 260133156 CAM A/B Move Timeout Timer expired - move did not complete in time
Error: 260133167 - (CAM A/B or Filter A/B) drive monitoring detected no motion when motion
Error: 260133169 drive monitoring detected the current motor position is out of the tolerance band for a successfull
move
Error: 260133176 Did not detect an encoder index pulse when homing CAM A/B
Error: 260133178 The current CAM positions are invalid for a position translation
Error: 260133224 The collimator has detected motor slip during z-axis tracking
Error: 260133223 CAM A Failed to Move To Correct Position For Beam Tracking.
Error: 260133164 CAM A/B Move Is In The Retry Tolerance. The Requested Move Will Be Attempted Again. Retry
No: X
Error: 260142520 CAM A drive monitoring detected the current motor position is out of the tolerance band for a
successfull move
Modality / Product
https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=70832&sliceId=1 4/5
13/4/23, 11:44 Scanner Hardware Stopped Scan : Replace Collimator CAM (A/B) Motor and coupling assembly.
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Optima 660
Revolution Discovery CT - AKA GSI or HD
Revolution EVO - Optima CT660 M40
Component
- ReqRef
- R0
- REM
https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=70832&sliceId=1 5/5