Professional Documents
Culture Documents
Xray Generation Troubleshooting
Xray Generation Troubleshooting
Xray Generation Troubleshooting
Contents
Sec 1 Overview ............................................................................................................................................. 1
Sec 2 Troubleshooting Flow Diagrams ......................................................................................................... 2
Sec 2.1 Main T/S Work Flow................................................................................................................. 2
Sec 2.2 Rotor T/S Work Flow................................................................................................................ 3
Sec 2.3 Filament T/S Work Flow........................................................................................................... 4
Sec 2.4 kV/mA TS Work Flow (page1 of 2) .......................................................................................... 5
Sec 2.4 kV/mA TS Work Flow (page 2 of 2) ......................................................................................... 6
Sec 3 Troubleshooting Methodology ............................................................................................................ 7
Sec 4 Diagnostic Functions........................................................................................................................... 8
Sec 4.1 HVDC ....................................................................................................................................... 8
Sec 4.2 Rotor ........................................................................................................................................ 9
Sec 4.3 Filament ................................................................................................................................. 12
Sec 4.4 kV/mA..................................................................................................................................... 16
Sec 5 HV Bleeder Usage ............................................................................................................................ 25
Sec 6 Power & Grounding........................................................................................................................... 25
Sec 7 Typical Diag Results ......................................................................................................................... 26
Rotor Diag – Typical results screens .................................................................................................. 26
Filament Test – Typical results screens.............................................................................................. 35
KV Loop Diag – Typical results screens ............................................................................................. 37
KV/MA Diag – Typical results screens ................................................................................................ 42
Audible Results: .................................................................................................................................. 47
Sec 8 Quick Review Highlights: .................................................................................................................. 50
Sec 1 Overview
The purpose of this document is to provide a convenient guide to defining Xray Generator problems and
identifying root cause. The goal is to correctly identify the failing part in one hour to avoid costly and time-
consuming miss-diagnosis of components, particularly HV Tanks and Xray Tube. This has a big impact
on Customer downtime and it is difficult to retrieve parts and reinstall originals if a wrong diagnosis is
made. Xray tube life is also a significant consideration for Customers when comparing CT scanners.
Please forward suggestions or corrections to michael.macneil@med.ge.com
N Y
Y Rotor
Rotor errors
Flowchart
Y Filament
Filament errors
Flowchart
Y kV / mA
kV / mA errors
Flowchart
Rotor Control
1 Diag Typical results
(Stationary Gantry)
Load Problem?
Accelerate rotor
Y Output Amps High Y
Listen at Gantry Rotor
Failure ? Accel Freq Low Relpace Tube
during rotor brake dragging ?
Brake Freq Low
& coast to stop
Internal DC bus low
N N
Optional:
HVDC Supply Low? Accelerate rotor
Output Amps Low Gantry 120V OFF Power OFF HVDC
Accel Freq Low Let rotor coast to & 120 VAC
HVDC errors stop - unpowered
HEMIT Xfmr
Secondary
Ohm Check
L to C ~1.2 ohms
TS HVDC HEMIT (rotor Xfmr)
S to C ~1.2 ohms
Pri: J10 Anode tank
L to S ~1.3 ohms
Sec: Anode HVcable
Drive Problem? Y
Replace HEMRC AC Replace HEMIT
Output Amps Low HEMIT bad?
Drive Anode tank
Accel Freq Low
N
Tube Stator
Rotor Control HVDC Warning Ohm Check
L to C ~2.3 ohms
2 Diag Lock Out Tag Out Tube stator
S to C ~2.3 ohms
system power input windings at Anode
(Rotating Gantry) L to S ~2.3 ohms
HV well
Replace Tube
Filament
Typical results
Diag
Ohm Check A
Filament Xfmr Primary
Ohm Check A
Y any test lead polarity OK
Filament Power OFF HVDC Remove J2 on
XS1: TP12-13 ~ 0.3 ohm
Open errors ? & 120 VAC Cathode Tank
XS2: TP12-11 ~ 0.3 ohm
Measure at CA
XL1: TP15-16 ~ 0.3 ohm
Tank Meas Bd.
N XL2: TP15-14 ~ 0.3 ohm
N N
Ohm Check B & C
Ohm Check B HV Cable / Xray Tube
Pull HV cable at note test lead polarity
Replace mA Bd
Cathode Tank +C to -S ~ 0.1 ohms
Measure at HV +C to -L ~ 0.2 ohms
cable pins -C to +S ~ 0.7 ohms
& bang on tube -C to +L ~ 0.7 ohms
Problem
resolved ?
Ohm Check C
High Y High Y
Measure in Xray Replace Tube
N Resistance ? tube well Resistance ?
& bang on tube
N N
Ohm Check D
Phone for Support
HV Cable Ohm Check D
Reinstall HV cable C to C ~ 0 ohms HC Cable Y
Measure HV cable Replace HV cable
at Cathode tank S to S ~ 0 ohms Bad ?
continuity (tighten
L to L ~ 0 ohms HV cable end pins)
N
Ohm Check E - Check condition of collars on HV cable pins
Ohm Check E & F
Remove HV cable Filament Xfmr Sec. - Check condition of springs in cable well pin sockets
at Tube Cathode - Reconnect cable at Tube Cathode & Repeat Ohm Check B
Measure at C to S ~ 0.3 ohms - IF Tube well connection can't be resolved Replace the Tube
HV cable pins C to L ~ 0.3 ohms
(tube end)
Ohm Check F
High Y High Y Replace Cathode
Measure in CA
Resistance ? tank well Resistance ? Tank
(filament xfmr sec)
N
N
Ohm Check D
- Check cable continuity from
Measure HV cable HC Cable Y
OBC backplane to CA tank Replace HV cable
continuity (tighten Bad ?
- Check or replace K78 relay
on the OBC backplane HV cable end pins)
- IF the problem persists N
Replace the Tube - Check condition of collars on HV cable pins
- Check condition of springs in cable well pin sockets
- Reconnect cable at Cathode Tank & Repeat Ohm Check E
- IF Tank well connection can't be resolved Replace the Tank
~ 1.5 minutes
1
No load kV
Basic Test Typical results
(kV Loop Diag)
HVDC Warning
AN or CA Y Lock Out Tag Out
kV=0V ? Check fuse in
HV Inverter
Amplify sound with a
N stethoscope if possible
Repeat Failing
HV Spits Y Test while
or overcurrent Typical results
Listening to HV
errors?
Tanks
N
~ 10 minutes
2
Xray Test HV Tank Y
Low, Med, Hi Pwr Typical results Internal Replace HV Tank
(kV mA Diag) Arcing ?
mA values Y
abnormal ? B Y Remove input to
Fail ? suspect HV Tank
(P1, P2 conn)
N
~ 15-30 minutes N
3 While testing:
No load kV Inspect HV cable Rerun kV Loop
Position HV Tanks
140 kV 30sec and tube well for Diag on failing side
upside down and rotate
(kV Loop Diag) signs of arcing. (Anode or Cathode
gantry back & forth a bit
Clean tube well only mode)
HV Spits Y
or overcurrent HV Cable Y Replace HV
errors ? Fail ?
OK? Inverter
N ~ 10 minutes
Y N
4 While testing:
Xray Test Keep gantry in the same
Re-install HV HVDC Warning
140 kV 100mA cable & retest Lock Out Tag Out
position as the previous Replace HV tank
(kV mA Diag) IF still failing: system power input
No Load test
Replace Tube
HV Spits Y
or overcurrent C
errors ? Optional: Optional:
Suspect Return system to
Repeat testing For Intermittent Spits
intermittent tube Customer use or
N Scope Tanks (TP1) Try 2 Heat Soak &
problem replace Tube
Scope & HV Bleeder Seasoning cycles
Xray Test
(kV mA Diag)
140 kV 100mA
N
kV arcing or
overcurrent Suspect Anode Tank or Tube EHS Warning: The presence of Xray limits our
Anode Y More likely the Anode Tank, since errors ability to listen for arcing within the HV Tanks.
overcurrent ? only occur with Xray load at the same A remote audio recorder may be a solution.
stationary gantry position as no load test
EHS Policy requires avoiding unnecessary
N Xray exposure. If you must enter the room
Suspect AN or CA Tanks or Tube during Xray, ensure the aperture/filter is closed
Y More likely a Tank, since errors only and wear Personal Protective Equipment, to
HV Spits? include a lead apron and lead collar throat
occur with Xray load at the same
stationary gantry position as no load test protection.
Test sequence
1. IF Line Loss error and PDU fuses blown.
o Power OFF; Warning: Lock Out Tag Out system power input
o Ohm check HVDC PS output. +(Red) to –(Black); +(Red) to Ground; –(Black) to Ground;
o Ohm check the HVDC bridge rectifier
2. IF 183182 errors occur with very low HVDC Rail voltage, check for a failed softstart resistor on
the PDU control board.
3. Check gesyslog for any correlation of HVDC errors to Anode, Cathode, or Rotor overcurrent
errors. If load overcurrents return to the respective Rotor or kV/mA troubleshooting.
4. Check regulation of the mains input at the PDU control Bd.
(This validates the PDU Xfmr output as well )
5. IF mains regulation at PDU control is out of spec. check facility line regulation at the input to the
PDU
Theory
Advanced system service manuals provide a good description of HVDC power supply operation.
Sample Errors
183262 OBC Generator hemrc_manager.c 1.14 XRAY GENERATION : HVDC Backup
Contactor Intlk Pri/Most Severe Backup contactor opened without firmware command.
185553 OBC Generator The HEMRC drive has detected a Line Loss. This can occur
when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: HEM-IT,
HEMRC/AC Drive, or X-Ray tube stator drawing too much current therefore loading down the HEMRC
input power.
NOTE 1: Results-frequency output If the rotor is open or frozen the frequency output may still increase to
180-190 during acceleration before it drops back and the test fails. Remember the diagnostic can only
measure output frequency, not the actual frequency of the rotor itself.
Typical Rotor diag results for an open rotor, cable, etc.
Diag results during accel show a higher than normal output current but the output frequency remains
around 100 Hz. After about 5 results output the drive goes directly to the IDLE mode without a BRAKE
cycle and the output frequency is 0 Hz. This is a good indication the tube rotor was not spinning at all,
because it would not be 0 Hz output freq so quickly unless the rotor was extremely bad or never spinning
in the first place. Don’t confuse the high freq sound from the drive with the sound of actual rotor rotation.
Note 2: When evaluating the HEMIT using resistance checks, the workflow indicates to only check the
HEMIT secondary because this easy to do and a more likely failure point. The primary can checked if
desired at J10 of the HEMIT Anode transformer but it is difficult to do and may not be worth the effort in
typical cases. [Typical readings: J10 1 to 2 ~1.2 ohms; J10 2 to 3 ~1.2 ohms; J10 1 to 3 ~1.2 ohms]
IF test results indicate a load problem (HERMIT or Tube); At the gantry listen for noise or vibration that
might indicate tube rotor bearing problems. After the rotor completes the braking cycle try and determine
if the rotor comes to a stop too quickly. An optional test is to run up the rotor then turn off the gantry 120
VAC power so all fans are quiet and the rotor will coast freely with no brake cycle. If the rotor stops too
quickly there is likely rotor bearing problems. Don’t do this too often though, because the braking cycle is
designed to decelerate quickly thru the rotors resonant frequency zone to avoid the extra vibration that is
not good for long term tube life.
Theory
Advanced service manuals contain good theory on the HEMRC operation.
Sample Errors
183316 OBC Generator The HEMRC failed to reach the Run speed in the allotted time.
183318 OBC Generator The HEMRC output current is out of 35.0 percent Warning tolerance
183319 OBC Generator The HEMRC output current is out of 55.0 percent MAXIMUM tolerance
185505 OBC Generator Could not send HEMRC message in the allotted time. HEMRC serial link
is down. HEMRC_ISO_+24V indicates OK possible causes: HEMRC CAN jumper, interconnects, CAN
devices
185548 OBC Generator The HEMRC operating frequency is below the minimum value.
185553 OBC Generator The HEMRC drive has detected a Line Loss. This can occur when the
input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: HEM-IT,
HEMRC/AC Drive, or X-Ray tube stator drawing too much current therefore loading down the HEMRC
input power.
219800 OBC Generator The HEMRC AC Drive reported fault code: F2 - Auxiliary Fault. The
interlock between the Chopper Control circuit on the HEMRC I/F Board and the Drive is open. Possible
Chopper Control fault or connections to the I/F board. Also check the fuse on the chopper resistor pan.
219800 OBC Generator The HEMRC AC Drive reported fault code: F4 - Undervoltage Fault The
Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 120 VAC in gantry
or power interruption. Also, may indicate excessive run or braking power required due to sluggish tube.
Problem/Solution
1) Rotor failure rate on Performix Ultra tubes increases when used heavily at 0.5 sec scanning
183553 183318 183319 Errors
The Performix Ultra tube used LightSpeed Plus, Ultra, and 16 systems have larger rotor bearings than the
Performix ADV tube. A trend observed in July 2002 showed a higher percentage of bearing failures on the
Ultra tubes when 0.5 second gantry rotation scan techniques are used frequently. The trend was only
seen on tubes with > 100 million mAs or 8 months life and high use of 0.5 sec gantry rotation speeds. Do
Not jump to this conclusion without adequate investigation and testing.
IF: Filament Open errors; Follow the flowchart to perform the sequence of ohm checks
IF: Filament power supply errors; Check the filament power supply located in the HEMRC assembly.
IF: Other filament errors i.e. filament inverter errors; Replace the mA Control Bd
Theory
The mA Control Board contains an inverter to produce filament heating current. Do not confuse tube mA
with filament current. Tube mA is a result of tube KV and filament heating current, so the bd. also has
feedback circuits for closed loop tube mA control. The filament/mA control function provides 10 - 440 mA
capability with 5% accuracy. Protection circuitry for; filament open circuit, mA imbalance, over-current,
and inverter faults is provided in addition to monitoring of mA by firmware during exposures.
Filament Current:
The filament inverter is a parallel resonant AC current supply with a 16KHz switching frequency to provide
a current source for the tube filaments. The inverter output current is commanded by FILDE produced on
the mA control bd. The inverter maintains output current equal to FILDE by using closed loop filament
current feedback sensed by T1 at the output of the mA circuit board. Inverter output is connected to a
relay located on the OBC backplane to select either the small or large tube filament. The filament select
relay output is connected to a filament isolation transformer in the Cathode tank by a cable harness from
the OBC backplane to the Cathode tank. The Cathode tank large and small focal spot filament
transformers provide HV isolation between the HV Filaments and the Low Voltage Filament Inverter.
Cathode tank filament transformer output is connected to the tube filaments by the Cathode HV cable.
Note: An older mA bd 46-288886G1 T3202BM used on HSA and CTi systems used Anode mA for
feedback control for use with Zeus tubes only.
Generator calibration
The generator calibration table contains open loop kV and filament current command values, sometimes
referred to as pre-scan values. Generator calibration tool adjusts these pre-scan values to produce
correct tube mA and KV for the first 4ms of exposure before closed loop mA and kV takes over. If pre-
scan values are too far out of calibration the result can be excessive kV or damaging filament current or
tube mA before closed loop feedback brings the values under control.
The two utilities for generator calibration Auto mA Cal and New Tube Seed Shift are often misunderstood.
These two utilities perform the same generator calibration with only one difference. Auto mA Cal uses
existing calibration values as a starting point for calibration. New Tube Seed Shift replaces existing
values with known safe starting values before calibration. Whenever a tube or HV component is replaced,
New Tube Seed Shift should be used to protect components from potentially damaging starting values.
Sample Errors
183119 OBC Generator FIL/MA fault detected: Filament current limit exceeded.
183121 OBC Generator FIL/MA fault detected: Open filament circuit.
183124 OBC Generator FIL/MA fault detected: Inoperative filament inverter.
183125 OBC Generator FIL/MA fault latch 1 would not reset.
214335 OBC diagExec Diagnostics XRAY GENERATION : MA Loop A filament inverter error
has been detected
NOTE 1: kV Loop Diag - kV accuracy On many systems the KV Loop Diag results can show incorrect
measured AN or CA kV values, usually Anode kV 5-30% low. The purpose here is only to test if there is 0
kV, HV spits, or inverter overcurrents with no load kV. The kV/mA Diag Xray Test performed later will
produce accurate readings for evaluation and decisions.
Running the test with HV cables pulled from the tube will report correct values as will running anode or
cathode only mode, but it is important to run both anode and cathode together at this point in the testing.
NOTE 2: ~13 kV is the normal result for Anode and Cathode in HV DC Bus:Test Mode. (HVDC=50V).
Test steps 2-4 must select HV DC Bus:Normal Mode.
IF: Anode or Cathode kV=0V; Check HV inverter fuse Warning: Lock Out Tag Out system power input
IF: No HV spits or inverter overcurrents occur move on to the Xray Test sequence 2.
IF: HV spits or inverter overcurrents do occur, repeat the test at the failing kV while listening to the HV
tank/s to determine the source of arcing. If an overcurrent occurs, the error message will indicate which
HV tank (anode or cathode) to listen to, otherwise listen to both HV tanks one at time.
IF: sizzling or arcing is heard from within a HV tank replace that HV tank [Audible Results]
IF: spits or overcurrents occur but no sizzling or arcing is heard in either tank, or the source of
arcing is uncertain, retest with the HV cables removed from the Anode & Cathode HV tanks.
Remember to fill the Tank HV wells with oil to about 3 cm from the top to prevent arcing during
the test and don’t rotate the gantry.
IF: spits stop with only the HV tanks in the circuit, inspect the HV cable/s for arcing and
inspect the tube HV cable well/s for arcing and clean if necessary. Reinstall HV cable/s
and repeat the test at the failing kV to confirm the problem still occurs only with the tube
in circuit before replacing the Tube.
Sequence (2): Xray Test – Low Medium & High Power (10 min.)
Testing must include generator operating modes (low, med, & high power) and both filaments
Start with low power as instructed to minimize damage from major failure
Identify the problem type: mA/kV out of tolerance, mA imbalance, certain power range, or certain filament
Diag > X-Ray Generation > KV mA [ TYPICAL RESULTS ] [KV/MA TS WORK FLOW]
Note: Log To File:True [default selection] Must always be “True”
Common Selections: Xray Test Type:Manual Gantry Enable:ON Gantry Speed: 2 sec
1. kV:80 mA:40 Exp Time(sec):2 Focal Spot:Small
2. kV:120 mA:200 Exp Time(sec):2 Focal Spot:Small
3. kV:120 mA:200 Exp Time(sec):2 Focal Spot:Large
4. kV:120 mA:400 Exp Time(sec):2 Focal Spot:Large
IF: Anode or Cathode kV = 0 then Lock Out Tag Out power and check HV Inverter fuse.
Note 1: For Performix Adv/Ultra tubes, Anode mA will normally be 5-10% less than Cathode mA.
Note 2: Watch Exposure duration time on the results screen. Results for less than 1 second of
exposure duration may be un reliable, especially if the exposure was aborted early due to errors. Also
keep in mind the first 4 ms of exposure is open loop kV and mA control that depends entirely on stored
generator calibration pre-scan values.
Note 3: kV measured is only what the system thinks the kV is via the HV tank internal bleeders,
measurement bds, and feedback gain adj. on the kV bd. In closed loop mode the system feedback circuit
will attempt to keep this measured value equal to the kV commanded. It is uncommon, but possible to get
the kV gain pots out of adjustment as much as ± 15kV.
Note 4: Compare Test results to T/S. Compare VCNT, Inverter frequency, duty cycle, & measured kVs
to T/S KV. Compare mA measured, filament current, and inverter currents to T/S MA
IF: No HV spits or inverter overcurrents occur move on to the Xray Test sequence 4. (140 kV 100mA)
IF: HV spits or inverter overcurrents occur, repeat the test at the failing kV while listening to the HV tank/s
to determine the source of arcing. If an overcurrent occurs, the error message will indicate which HV tank
(anode or cathode) to listen to, otherwise listen to both HV tanks one at time.
IF: sizzling or arcing is heard from within a HV tank replace that HV tank [Audible Results]
IF: spits or overcurrents occur but no sizzling or arcing is heard in either tank, or the source of
arcing is uncertain, retest with the HV cables removed from the Anode & Cathode HV tanks.
NOTE 1: Always run KV loop initially with Both selected ( Anode and Cathode ).
It is just as important while listening to HV tanks to know if a tube spit occurs without any audible arcing
from the tanks to help identify a bad tube, as it is to hear an arc within a tank to identify a bad tank.
Both: Tests Tube , HV Tanks, and HV Cables
Anode only: Tests Tube-Anode to Casing only, Anode tank, Anode HV cable,
Cathode only: Tests Tube-Cathode to Casing only, Cathode tank, Cathode HV cable
IF: HV spits or overcurrents occur with Xray load at the same stationary gantry position with tanks upside
down, but can’t be reproduced in the same position with extensive testing at 140 kV no load, then the HV
tanks are the most likely failure.
IF: HV spits can only be produced with normal rotating Xray scans, then the problem may still be in a HV
tank or Tube. However at this point in the troubleshooting process a reasonable effort has been made to
isolate the root cause without success. Replace the tube as the most likely failing item.
IF: All inverter fiber LEDS light OK continue to isolate the problem between a failing Inverter or HV tank
IF: an Inverter LED does not light in sequence;
o Check the fiber optic cable connections at inverter and OBC backplane.
IF the fiber cable connections were not fully inserted
o Push the connectors in properly and rerun the fiber optic test.
IF the LEDs are now OK; Return to the Xray Test to verify operation
IF the fiber cable connections were already OK continue below to swap fiber cables
o Swap the suspect fiber cable with one of the good ones and rerun the fiber test
IF the fiber cable is bad replace it and rerun the fiber optic test
IF the fiber cable & connections are good replace the kV bd and rerun the fiber optic test
IF the problem continues replace the OBC backplane
Theory
HV Spits: This system doesn’t distinguish between HV tank, cable, or tube arcs so we should never
assume spits are “Tube spits” as is reported in some error messages. HV Spits are counted by the kV
control bd when measured kV drops faster than a specified rate indicating a “spit” and can be caused by
HV tanks, HV cables or the tube. Of course the spit counter on the kV bd could also be defective although
I believe this is rare. One difference we can observe is that spits in the tube can sometimes be improved
by running a heat soak and seasoning cycle. HV cable arcs will generally just get worse and more
frequent quickly as a carbon track forms from the arcs. With HV tanks, carbon tracks from arcs may get
diluted in the HV oil allowing intermittent arcs to go on for a long time, however a heat soak and
seasoning will not improve intermittent spits caused by tanks like it sometimes can with a tube.
Note: TS flowcharts in this document apply only to the 2 newer kV Bd versions. Older HSA or CTi
scanners should upgrade to one of the 2 newer versions.
Inverters
The Inverters are designed to invert the HVDC from the PDU to provide a high frequency output to the
anode and cathode HV transformers (Tanks). The inverters and tanks form a tuned circuit where
maximum VAC is output to the tanks at approximately 19.0 kHz. Individual inverter output capacitors are
adjusted at the factory to tune the inverter for 19.0 kHz. Actual base frequency and adjusted frequency
are recorded on a label in the inverters. Inverter IGBTs are controlled entirely by the frequency and duty
cycle of fiber optic signals received from the kV control Bd. For medium to high power exposures the
inverters are designed to operate at 100% duty cycle with kV controlled by frequency. However in practice
one inverter and tank combination will always produce a bit more kV for the same frequency than the
other side, so the higher output side will have to reduce duty cycle a bit to keep its kV correct. In our
system with a Performix tube this is typically the Anode side that is a bit higher output so it runs at a
slightly lower duty cycle (appx. 85%). However this is not always the case and with a slight difference in
inverter factory tuning and HV tank combination the Cathode output can sometimes be higher and run at
the lower duty cycle with no problems. For low power exposures frequency alone is not enough to keep
the kV under control and both inverters will have to run at a much lower duty cycle.
While inverter frequency and duty cycle control kV, inverter current is mostly a function of tube mA and
can be used to help decide the true mA in situations where a mA metering circuit failure is in question.
Failure Notes
Note: Older HSA or CTi scanners with Zeus tubes (MX165) used Anode mA for closed loop control on
46-288886G1 mA Bds. For Zeus tubes Anode mA and Cathode mA are equal so it doesn’t matter if
Anode or Cathode mA is used for the closed loop mA control. These scanners can use either type mA
control bd.
Warning: Some, older schematics, theory diagrams and/or troubleshooting directions have the anode vs.
cathode feedback control confused because of the mA bd differences.
Procedure to check the mA bd installed.
• Perform a Scan HW reset; (bd rev info in the log will get overwritten during normal scanning)
• rsh sbc; nbsClient obcr; y (to see log) ; Control-C (to exit nbsClient)
Example: > 3472<t7: =====| MA Board Revision: 2154834-01 C
> 3472<t7: =====| KV Board Revision: 2143147-01 B
mA Bd. 46-288886G1 T3202BM for ZEUS(MX165) Tube only. Uses Anode mA feedback for
closed loop mA control. It must not be used for Performix tubes (MX200). As these Bds went out
of stock they were replaced by the newer, backward compatible 2154834 Bd.
mA Bd. 2154834 for Performix or Zeus tubes uses Cathode mA feedback for closed loop mA
control. This change is required for compatibility with the Performix X-Ray Tube (MX200). Also
backward compatible with Zeus tubes.
Sample Errors
183109 An OC OBC Generator gen_abort.c 1.31 Function: XRAY GENERATION
: KV Loop Scan: 493/2/2 Type: Helical Exception Level: Secondary Time: 11:40:49:970
Log Series: 132 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4
Solution:
Found cathode HV tank bad, but when removing inverter from tank mount position, noticed a high amount
of arcing had been going on between inverter and tank chassis due to concentration of carbon tracking on
both assemblies. Carbon tracks extended from right upper bolt hole over to center of mounting position
and back to lower right bolt hole, forming a 'V' pattern between the two bolt positions.
Srv Eng Review enclosure from CTCge66828
Lots of arcing found between the inverter plate and HV tank. The arcing was produced due to space
created between the two surfaces. The inverter was tested and it worked correctly. The Tank had a bad
cathode as described by the FE. It is a manufacturing requirement to loc-tite (242) the bolts that hold the
inverter to the HV tank. This is to prevent the bolts from getting loose. Proper manufacturing procedure
must be followed in the field as well when replacing this component.
Marcela Gonzalez CT Xray Gen Eng.
The above problem was caused by a faulty Cathode tank. The Cathode tank mA measurement circuit
was measuring lower than actual mA, but since the Cathode mA controls closed loop feedback it will
adjust filament current until the cathode mA as measured equals the selected value. The Anode mA was
simply reporting the true tube mA that was driven too high due to the failing cathode measurement.
Note that both Inverter currents are similar indicating the true Anode and Cathode mA should be similar.
3) Cathode only or Anode only mA overloads on higher mA only. Not related to kV.
This was a known problem seen on the original old KV cont bd ( 46-321064G1 T3202BP )
Failure only occurs at higher mA at any kV. This problem has often been seen with the older KV control
bds, and is caused by KV bd diode failures in the rectification of the Anode or Cathode Inverter current
feedback to the KV bd. The KV bd reports a higher current than actual. If these old style bds fail they
should be replaced with one of the newer versions.
When measuring kV be sure to use the bleeder, scope, and cable setup. The 100 kV 50 ma technique is
selected to minimize high frequency kV ripple and scope aliasing effects, without being so low as to
produce high overshoot.
The HV bleeder is most accurate at 100 kV. It was also designed for older 3 phase 12 pulse generators
with a KV ripple of only 720 Hz. The high frequency generator operates from 19 to 32 KHz and when a
scope is connected directly to the bleeder it has a tendency to amplify the high frequency kV ripple
producing aliasing on the scope waveform. To minimize this effect the capacitance of the bleeder to
scope cable is needed to attenuate the higher frequency. The low 50 mA also minimizes ripple without
being so low as to cause too much kV overshoot.
Grounding
Tube and HV component life for HSA, CTi, LS is not as dependent on hospital grounding as older 9800,
HLA scanners. However grounding within the CT System itself remains critical for tube life, and is
necessary to protect other sensitive electronics from HV disturbances caused by the tube and/or
generator.
HV Ground Path Connections:
Tube – Rotating slipring ground ring –Stationary SR brush block – Stationary Gantry frame –
Table/Gantry raceway ground buss – PDU
Many systems may report inaccurate AN or CA kV values, often Anode kV 5-30% low.
For kV or mA accuracy decisions use kV/mA Diagnostic results.
HV Manual; Both; Test Mode; 80kV 2 sec - 604520QXI LS_2x Performix Ultra
Wed Feb 4 13:31:39 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 25.8KV 80.0KV 26.1KV
2. Cathode KV: 12.8KV 40.0KV 12.9KV
3. Anode KV: 13.0KV 40.0KV 13.1KV
4. Cathode MA: 0.1mA 0.0mA 0.1mA
5. Anode MA: 0.1mA 0.0mA 0.1mA
6. Cathode inverter current: 0.375A -- 0.375A
7. Anode inverter current: 0.550A -- 0.550A
8. Approx. KV inverter frequency (VCNT): ( 0.09V) 19.7KHz
9. Cathode inverter duty cycle: 100%
10. Anode inverter duty cycle: 100%
11. HVDC Bus voltage: 66V -- 67V
12. Exposure duration: -- 2000mS 2007mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 88H
HV Manual; Both; Normal Mode; 80kV 2 sec 604520QXI LS_2x Performix Ultra
Wed Feb 4 13:33:10 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 77.3KV 80.0KV 78.4KV
2. Cathode KV: 37.5KV 40.0KV 37.0KV
3. Anode KV: 39.9KV 40.0KV 40.9KV
4. Cathode MA: 0.3mA 0.0mA 0.2mA
5. Anode MA: 0.3mA 0.0mA 0.2mA
6. Cathode inverter current: 0.125A -- 0.175A
7. Anode inverter current: 0.250A -- 0.250A
8. Approx. KV inverter frequency (VCNT): ( 4.44V) 29.2KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 680V -- 678V
12. Exposure duration: -- 2000mS 2063mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 89H
HV Manual; Both; Normal Mode; 120kV 2 sec 418862LS LS_4x Performix Ultra
Thu Nov 27 16:10:00 2003 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 115.3KV 120.0KV 114.1KV
2. Cathode KV: 60.2KV 60.0KV 59.8KV
3. Anode KV: 55.0KV 60.0KV 54.7KV
4. Cathode MA: 0.3mA 0.0mA 0.2mA
5. Anode MA: 0.3mA 0.0mA 0.1mA
6. Cathode inverter current: 0.175A -- 0.050A
7. Anode inverter current: 0.125A -- 0.000A
8. Approx. KV inverter frequency (VCNT): ( 6.36V) 30.5KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 691V -- 692V
12. Exposure duration: -- 2000mS 2055mS
13. Exposure number: -- 5 5
14. Status register (Address = FFCFF9H): -- 89H
HV Manual; Both; Normal Mode; 140kV 10 sec 604520QXI LS_2x Performix Ultra
Wed Feb 4 13:34:25 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 134.1KV 140.0KV 133.8KV
2. Cathode KV: 64.5KV 70.0KV 65.4KV
3. Anode KV: 69.6KV 70.0KV 70.0KV
4. Cathode MA: 1.3mA 0.0mA 0.7mA
5. Anode MA: 1.3mA 0.0mA 0.8mA
6. Cathode inverter current: 0.500A -- 1.525A
7. Anode inverter current: 0.675A -- 1.150A
8. Approx. KV inverter frequency (VCNT): ( 6.61V) 30.5KHz
9. Cathode inverter duty cycle: 1%
10. Anode inverter duty cycle: 1%
HV Manual; Both; Normal Mode; 140kV 10 sec 418862LS LS_4x Performix Ultra
Thu Nov 27 16:21:14 2003 High voltage status
140KV 15s Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 135.2KV 140.0KV 136.1KV
2. Cathode KV: 69.7KV 70.0KV 67.7KV
3. Anode KV: 65.4KV 70.0KV 64.1KV
4. Cathode MA: 0.6mA 0.0mA 1.0mA
5. Anode MA: 0.5mA 0.0mA 0.7mA
6. Cathode inverter current: 0.250A -- 0.125A
7. Anode inverter current: 0.175A -- 0.000A
8. Approx. KV inverter frequency (VCNT): ( 5.36V) 30.5KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 689V -- 691V
12. Exposure duration: -- 10000mS 10056mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 89H
HV Manual; Both; Test Mode; 80kV 2 sec 604682CTI CTi, Performix Adv, KV Bd.
2143147
Tue Feb 17 13:06:03 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 26.4KV 80.0KV 26.4KV
2. Cathode KV: 13.0KV 40.0KV 13.0KV
3. Anode KV: 13.3KV 40.0KV 13.2KV
4. Cathode MA: 0.2mA 0.0mA 0.2mA
5. Anode MA: 0.1mA 0.0mA 0.2mA
6. Cathode inverter current: 0.550A -- 0.550A
7. Anode inverter current: 0.550A -- 0.550A
8. Approx. KV inverter frequency (VCNT): ( 0.14V) 19.8KHz
9. Cathode inverter duty cycle: 100%
10. Anode inverter duty cycle: 100%
11. HVDC Bus voltage: 66V -- 66V
12. Exposure duration: -- 2000mS 2006mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 88H
HV Manual; Both; Normal Mode; 80kV 2 sec 604682CTI CTi, Performix Adv, KV Bd.
2143147
Tue Feb 17 13:06:27 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 68.4KV 80.0KV 69.0KV
2. Cathode KV: 40.7KV 40.0KV 41.4KV
HV Manual; Both; Normal Mode; 120kV 2 sec 604682CTI CTi, Performix Adv, KV Bd.
2143147
Tue Feb 17 13:07:50 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 110.1KV 120.0KV 112.7KV
2. Cathode KV: 60.5KV 60.0KV 62.3KV
3. Anode KV: Normal result > 49.6KV 60.0KV 50.6KV
4. Cathode MA: 0.3mA 0.0mA 0.5mA
5. Anode MA: 0.3mA 0.0mA 0.3mA
6. Cathode inverter current: 0.250A -- 0.550A
7. Anode inverter current: 0.250A -- 0.175A
8. Approx. KV inverter frequency (VCNT): ( 3.86V) 27.9KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 665V -- 665V
12. Exposure duration: -- 2000mS 2049mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 89H
HV Manual; Both; Normal Mode; 140kV 10 sec 604682CTI CTi, Performix Adv, KV Bd.
2143147
Tue Feb 17 13:08:32 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 130.8KV 140.0KV 132.9KV
2. Cathode KV: 70.2KV 70.0KV 71.7KV
3. Anode KV: Normal result > 60.4KV 70.0KV 61.5KV
4. Cathode MA: 0.7mA 0.0mA 0.5mA
5. Anode MA: 0.6mA 0.0mA 0.3mA
6. Cathode inverter current: 0.375A -- 0.725A
7. Anode inverter current: 0.300A -- 0.925A
8. Approx. KV inverter frequency (VCNT): ( 6.58V) 30.5KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 666V -- 666V
12. Exposure duration: -- 10000mS 10049mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 89H
HV Manual; Both; Normal Mode; 140kV 2 sec 604727CT CTi Performix Adv
Wed May 29 09:26:26 2002 High voltage status
140KV Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 120.6KV 140.0KV 120.5KV
2. Cathode KV: 70.9KV 70.0KV 72.0KV
3. Anode KV: Normal result > 49.6KV 70.0KV 48.3KV
4. Cathode MA: 0.6mA 0.0mA 0.5mA
5. Anode MA: 0.5mA 0.0mA 0.3mA
6. Cathode inverter current: 0.500A -- 0.975A
7. Anode inverter current: 0.500A -- 1.350A
8. Approx. KV inverter frequency (VCNT): ( 6.42V) 30.5KHz
9. Cathode inverter duty cycle: 0%
10. Anode inverter duty cycle: 0%
11. HVDC Bus voltage: 541V 550V 540V
12. Exposure duration: -- 2000mS 2032mS
13. Exposure number: -- 1 1
14. Status register (Address = FFCFF9H): -- 89H
Tube Type Low KVA Med KVA Med KVA High KVA
Small Fil Small Fil Large Fil Large Fil
Performix Ultra LS_2x 80kV 40mA 120kV 200mA 120kV 200mA 120kV 400mA
Performix Ultra LS_4x 120kV 200mA 120kV 200mA 120kV 400mA
Performix Adv CTi 80kV 40mA 120kV 200mA 120kV 200mA 120kV 400mA
Performix: Anode mA is normally 5-10% lower than Cathode; Cathode mA = true tube mA;
Low power; Small filament; 80kV 40mA 2s 604520QXI LS_2x Performix Ultra
Wed Feb 4 13:37:45 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 79.7KV 80.0KV 79.7KV
2. Cathode KV: 39.8KV 40.0KV 39.7KV
3. Anode KV: 39.9KV 40.0KV 39.8KV
4. Cathode MA: 40.4mA 40.0mA 40.4mA
5. Anode MA: 37.3mA 40.0mA 37.3mA
6. Cathode inverter current: 6.775A -- 6.850A
7. Anode inverter current: 6.525A -- 6.525A
8. Approx. KV inverter frequency (VCNT): ( 3.60V) 27.4KHz
9. Cathode inverter duty cycle: 41%
10. Anode inverter duty cycle: 36%
11. Filament current: 5.573A 5.482A 5.568A
12. HVDC Bus voltage: 663V -- 664V
13. c: -- 2000mS 2007mS
14. Exposure number: -- 1 1
Med power; Small filament; 120kV 200mA 2s 604520QXI LS_2x Performix Ultra
Wed Feb 4 13:40:48 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.5KV 120.0KV 119.6KV
2. Cathode KV: 59.6KV 60.0KV 59.5KV
3. Anode KV: 60.0KV 60.0KV 60.0KV
4. Cathode MA: 200.7mA 200.0mA 200.7mA
5. Anode MA: 184.3mA 200.0mA 184.3mA
6. Cathode inverter current: 31.575A -- 31.625A
7. Anode inverter current: 29.300A -- 29.300A
8. Approx. KV inverter frequency (VCNT): ( 2.40V) 24.7KHz
9. Cathode inverter duty cycle: 86%
10. Anode inverter duty cycle: 73%
11. Filament current: 6.364A 6.146A 6.361A
12. HVDC Bus voltage: 633V -- 630V
13. Exposure duration: -- 2000mS 2007mS
14. Exposure number: -- 1 1
Med power; Small filament; 120kV 200mA 2s 418862LS LS_4x Performix Ultra
Thu Nov 27 16:27:29 2003 High voltage status
Average Selected Last
No. Device Value Value Sample
Med power; Large filament; 120kV 200mA 2s 418862LS LS_4x Performix Ultra
Thu Nov 27 16:30:32 2003 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.5KV 120.0KV 119.5KV
2. Cathode KV: 59.6KV 60.0KV 59.6KV
3. Anode KV: 59.8KV 60.0KV 60.1KV
4. Cathode MA: 201.4mA 200.0mA 201.4mA
5. Anode MA: 185.8mA 200.0mA 185.9mA
6. Cathode inverter current: 31.925A -- 31.925A
7. Anode inverter current: 29.300A -- 29.375A
8. Approx. KV inverter frequency (VCNT): ( 2.48V) 24.9KHz
9. Cathode inverter duty cycle: 85%
10. Anode inverter duty cycle: 73%
11. Filament current: 5.778A 5.670A 5.773A
12. HVDC Bus voltage: 633V -- 634V
13. Exposure duration: -- 2000mS 2007mS
14. Exposure number: -- 1 1
High power; Large filament; 120kV 400mA 2s 418862LS LS_4x Performix Ultra
Thu Nov 27 16:34:36 2003 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.6KV 120.0KV 119.9KV
2. Cathode KV: 59.4KV 60.0KV 59.4KV
3. Anode KV: 60.1KV 60.0KV 60.4KV
4. Cathode MA: 400.8mA 400.0mA 401.2mA
5. Anode MA: 369.6mA 400.0mA 369.3mA
6. Cathode inverter current: 63.675A -- 63.550A
7. Anode inverter current: 58.375A -- 58.425A
8. Approx. KV inverter frequency (VCNT): ( 1.32V) 22.4KHz
9. Cathode inverter duty cycle: 100%
10. Anode inverter duty cycle: 97%
11. Filament current: 6.198A 5.987A 6.195A
12. HVDC Bus voltage: 599V -- 598V
13. Exposure duration: -- 2000mS 2006mS
14. Exposure number: -- 1 1
Low power; Small filament; 80kV 40mA 2s 604682CTI CTi Performix Adv
Tue Feb 17 13:10:11 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 79.9KV 80.0KV 79.7KV
2. Cathode KV: 40.0KV 40.0KV 40.0KV
3. Anode KV: 39.9KV 40.0KV 40.0KV
4. Cathode MA: 40.3mA 40.0mA 40.4mA
5. Anode MA: 36.8mA 40.0mA 37.0mA
6. Cathode inverter current: 6.950A -- 6.950A
7. Anode inverter current: 6.400A -- 6.400A
8. Approx. KV inverter frequency (VCNT): ( 3.59V) 27.4KHz
9. Cathode inverter duty cycle: 40%
10. Anode inverter duty cycle: 37%
Med power; Small filament; 120kV 200mA 2s 604682CTI CTi Performix Adv
Tue Feb 17 13:13:16 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.8KV 120.0KV 119.6KV
2. Cathode KV: 59.9KV 60.0KV 59.6KV
3. Anode KV: 59.9KV 60.0KV 59.7KV
4. Cathode MA: 200.3mA 200.0mA 200.5mA
5. Anode MA: 182.8mA 200.0mA 182.9mA
6. Cathode inverter current: 31.800A -- 31.800A
7. Anode inverter current: 29.175A -- 29.175A
8. Approx. KV inverter frequency (VCNT): ( 2.38V) 24.7KHz
9. Cathode inverter duty cycle: 85%
10. Anode inverter duty cycle: 76%
11. Filament current: 6.403A 6.192A 6.398A
12. HVDC Bus voltage: 632V -- 632V
13. Exposure duration: -- 2000mS 2007mS
14. Exposure number: -- 1 1
Med power; Large filament; 120kV 200mA 2s 604682CTI CTi Performix Adv
Tue Feb 17 13:16:17 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.9KV 120.0KV 119.8KV
2. Cathode KV: 59.9KV 60.0KV 59.8KV
3. Anode KV: 59.9KV 60.0KV 59.8KV
4. Cathode MA: 201.1mA 200.0mA 201.4mA
5. Anode MA: 183.5mA 200.0mA 183.2mA
6. Cathode inverter current: 31.925A -- 31.925A
7. Anode inverter current: 29.300A -- 29.300A
8. Approx. KV inverter frequency (VCNT): ( 2.35V) 24.6KHz
9. Cathode inverter duty cycle: 85%
10. Anode inverter duty cycle: 77%
11. Filament current: 5.910A 5.764A 5.907A
12. HVDC Bus voltage: 631V -- 632V
13. Exposure duration: -- 2000mS 2007mS
14. Exposure number: -- 1 1
High power; Large filament; 120kV 400mA 2s 902496CTIP CTi Performix Adv
Thu Apr 15 10:55:13 2004 High voltage status
Average Selected Last
No. Device Value Value Sample
---------------------------------------------------------------------
1. Total KV: 119.3KV 120.0KV 119.8KV
2. Cathode KV: 59.2KV 60.0KV 59.7KV
3. Anode KV: 59.9KV 60.0KV 60.3KV
4. Cathode MA: 400.6mA 400.0mA 401.1mA
Audible Results:
[ KV MA T/S WORK FLOW ]
Often listening closely to HV tanks can help diagnose generator problems. While major HV transformer
breakdowns can usually be heard when standing beside the gantry some intermittent or less serious
problems may require a stethoscope to amplify the sound. There is usually one available in most CT
rooms. If a stethoscope is not available a screwdriver blade held to a tank with the handle held to your ear
will help amplify sound within HV tanks.
Before you perform “No Load HV” testing, first use the stethoscope to listen to the HV tanks and become
accustomed to the normal background noises. Rotate the gantry a bit while listening to hear some of the
normal results below.
Normal Results
Under no load kV testing the tanks should be silent except for the background noises listed below
o Noise from tube fans if they are on.
o May hear occasional popping noises especially if rotating the gantry a bit.
o May hear some sloshing of the oil inside the tank when rotating the gantry a bit.
o Noise from the stethoscope tube rubbing on the gantry components
Failure results
HV Tanks:
Arcs are typically audible
o Typical HV tank arcing will sound like bacon frying, hissing with some arcing pops. (some
high freq content but predominantly <100 hz content) If there is only a very faint high
frequency (>1kHz) hissing sound it may just be a reflection from the tube.
o Multiple spits can sound like a rapid fire arcing sound.
Minor transformer breakdown or partial winding short
o Subtle humming during HV ON
Major transformer defects such as winding shorts can often be easily heard without any aid.
o Loud humming sound or even loud sound of oil bubbling.
HV cables
Warning: Never directly contact a HV cable to listen for arcs. Arcing in HV cables will not be
difficult to find because a carbon track will build with each arc, very quickly leading to either
constant arcing or a hard short. You also will be able to detect a burning smell. Years ago arcing
would occasional occur in a HV cable either at the end of the flexible cable under the horn or in
the termination. Since cable manufacturing was improved a long time ago HV cable arcs have
become very rare.
Site Examples:
There is a project underway to record audio samples as heard using a stethoscope to include in future
revisions.