REFERENCE MANUAL
ACU-RITE
®
The SENC 50 Precision Glass Scale Linear Encoder provides the For future ordering information or warranty service, record the linear
accuracy and reliability of an ACU-RITE measuring system in a
®
encoder catalog and serial numbers.
very small cross-section with digital output (analog output available).
Catalog # Serial # Manufacture Code #
Features and options include:
• Resolutions of 0.5, 1.0 or 5µm. Axis # 1: ______________________________________
• Accuracy grades of ± 3 or ± 5 µm / 1000mm. Axis # 2: ______________________________________
• Position Trac or 50mm interval Reference Mark. Axis # 3: ______________________________________
• Vinyl or Armor cables of 5, 13, & 19 ft. lengths. Axis # 4: ______________________________________
• Mounting fasteners, center supports, and backup spar.
Date of Purchase ______________________________________
• Installation tools, brackets, and accessories.
Distributor: ______________________________________
The SENC 50 is especially useful for applications with limited space
such as X/Y stages and compact metrology equipment. Contact Address: ______________________________________
your Authorized Distributor for assistance with selection of product Telephone: ______________________________________
options and accessories.
Shipping carton
contents
A
A) SENC 50 linear encoder
C B) Reference Manual
C) Encoder and Cable Mounting
Hardware
2
ACU-RITE
®
// .005 A
• Mount encoders close to ways to insure system accuracy. Gap
-A- = Machine travel B
• Mount with lip seals down and away from the work area. .059 ±.005”
• Brackets should be short and rigid. Center support surface [1.50]
• Surfaces must be in good condition, clean and free of Reading head required
Scale assembly
dirt and paint. assembly
• Do not remove shipping brackets until instructed. // .005 A
• Shipping bracket spacers can be used to insure .059” gap. // .005 A
• Mount from either side to obtain desired cable exit direction. Equal Equal
• Allow clearance for shipping bracket removal. • Limit equipment travel to less than measuring length.
3
ACU-RITE
®
Reading
head for M4 Strain relief Ø .30 Ø .16 Align both
assembly Hex nut and cable [7.6] [4.1] surfaces
.140
.463 Armor Vinyl
[3.55] 1.85 .170
[11.76] [47.00] [4.32]
2.0 [50] Approx. .02
2.775 Ø .160 min bend radius [0.5]
.175 [4.57]
[70.50] [4.00] min. of armor
Typ.
2.350 .280 .840
.213
[59.69] [7.11] [21.34]
[5.41] .150 1.352 [34.34]
Typ. [3.81]
Spar
.170
.138 .652 [4.32]
.295 .159
[7.50] M3 x .25 [6.31] [3.50] [16.56]
[4.04] Typ.
deep min.
4
ACU-RITE
®
Spar
Encoder retaining fastener
clip
Spar leveling
screws
5
ACU-RITE
®
// .005 // .030
section.
• General steps for small X & Y stage.
• Adjust drill depths and fastener lengths as required.
CL
• Contact your Authorized Distributor if assistance is required.
Mark center of axis
Keep head and scale Underside of end caps to be Align top of scale case to
Scale case case flush with parting line of within .015” of axis travel
CL surfaces flush upper slide.
Scale case
• Center the reading head on the scale case by sliding the head and • Locate the linear encoder along the axis parting line.
shipping brackets together along the scale case. • Mark location of one end mounting hole in the scale case.
8
ACU-RITE
®
Drill / tap
CL
for 6-32 (M3).
Repeatability Test
Readout
• Zero the display and indicator.
• Move axis to the end of it’s travel, then return to dial zero.
• Check function of the linear encoder. • Readout should read zero ± 1 count.
10
ACU-RITE
®
Connect a ground wire from the terminal on the back of the readout
to the machine or earth ground. Attach a ground wire from the
machine to a solid earth ground.
With the encoder attached to the machine and the cable connected
to the readout, check shielding by measuring resistance between
connector housing and scale unit. Desired value: 1 Ω max.
11
ACU-RITE
®
If you experience difficulties with your installation, there are methods with other machine structures through the length of the linear
to analyze the difficulty. Operating difficulties can be caused by encoder travel.
either the linear encoders or the readout. • Check for loose fasteners. If you find loose fasteners, first confirm
Checking the readout that the linear encoder is installed to the tolerances specified and
then retighten the fasteners as required.
Follow the steps below to determine if your difficulties are associated • Confirm that the linear encoder is installed to the required
with the readout: tolerances by checking the alignment tolerances specified on Page
• Insure that the linear encoder connectors are correctly seated. 3 “Key Points”. If the installation does not meet the tolerances,
• Swap linear encoder cables at the readout to see if the problem reinstall the linear encoder according to the procedures in
is still shown in the same display. “Installation”.
• If the problem remains in the same display, the readout is in • Perform a Repeatability Test as described on Page 10, “Checking
error. Your Installation”. If the linear encoder is installed to the required
• If the problem follows the connection change, the linear encoder tolerances, the bracketry and linear encoder have been checked
may be in error. for interferences and loose fasteners, and the linear encoder fails
• Follow this procedure for any intermittent problem experienced. the repeatability test, the linear encoder is likely at fault.
If the readout is at fault, refer to “What to do” to arrange for the Do not attempt to repair the linear encoder. The units are only
parts necessary to repair your system. If the linear encoders appear serviceable by assembly replacement. Attempted repair can
to be the cause, perform the next step. permanently damage the unit and void the warranty.
Checking the Linear Encoders What to do
Linear encoder difficulties can be caused by improper installation, If an ACU-RITE linear encoder or readout is found to be at fault,
®
loose or misaligned bracketry, or a damaged / failed linear prior to removing the linear encoders or readout contact your
encoder. Authorized Distributor or OEM/OEI for repair instructions.
Follow the steps below to determine the cause of your system
difficulties:
• Confirm that your brackets, and installation does not interfere
12
ACU-RITE
®
13
ACU-RITE
®
Digital Differential
Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9
N/C Green Yellow Pink Red White Black Gray Brown
N/C Channel Channel Channel Channel Ground Vcc, + 5.1 Channel Channel
A+ A- B+ B- ± 0.1 VDC R+ R-
1 5
6 9
Analog Differential
Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9
White w/Grn Stripe & Brown Green Yellow Orange Red N/C Black Blue Violet
Ground Channel Channel Channel Channel N/C Vcc, + 5.0 Channel Channel
A+ A- B+ B- ± 0.1 VDC R+ R-
14
ACU-RITE
®
Output Signals
IOH=(High level output current) = 20mA
VOH=(High level output voltage) >2.5Vdc
0° 90° 360°
0° 360° IA, B:7-16 µApp
1 1 Channel B or
Channel A+ Channel R+
0 0 1.0 Vpp
1 Channel A
Channel A- 1 Channel R-
0 0
90°
1 1 Count IR:2-8 µApp
Channel B+ or
0 (Phased) Channel R
1 1.2 Vpp
Channel B-
0
IOL=(Low level output current) = -20mA
VOL=(Low level output voltage) < 0.6Vdc
Reference Mark signals Square-wave signal Differential current or 1 volt peak to peak output
Signal level TTL-level 2-8µApp or 1.2 Vpp
w/100 K Ohm load
Power Supply 5.1 ± 0.1 VDC @ 220 mA max. 5.0 ± 0.1VDC @ 75 mA max.
15
ACU-RITE
®
16
ACU-RITE
®
ACU-RITE
®