KV Meter RMI 245

Use toggle and select to enter set up information.

Modality Waveform
1 Rad 1 Single phase half wave
2 Mammo 2 Single phase full wave
3 Fluoro 3 Three phase 6 pulse
4 Three phase 12 pulse
5 Constant potential

Press mode: display rdy (Listo para recibir exposicion)

Take up to a maximun of ten exposures (use toggle button to cycle to stored values)

To return to setup mode press Mode (hill erase all stored readings)

Displays:

--- exposure too low


ooo exposure too high
: meter is out of calibration
lobat battery low
.1.1._ meter in set up mode
rdy ready for first exposure