12) Perform a reference drive. ✔
□
13) X-ray safety measurement with beam on sample, on all four sides of
the detector and behind the detector, in both measurement and parking
positions. The measured radiation should meet local requirement.
Recommended <2 uSv/h. ✔
□
2. Adjustment of Imaging
1) Set up the SCU/SBU using the service software. Save the setting
to flash. (It will be also saved automatically on the customer pc.) □
2) Calibrate the scale bar (set up the pixel size). □
3) Functional test of “Imaging System”. □
3. Adjustment of spectrometer
@Mn
Mn Kα
Zeropeak
Objects
Hypermap
Feature
Jobs
Report
Printing (PDF)
BRUKER Nano GMBH TELEFON: +49 30 6709 90 -0 AMTSGERICHT BLN. CHARLOTTENBURG HRB 99662 BANKVERBINDUNG:
AM STUDIO 2D TELEFAX: +49 30 6709 90 -30 GESCHÄFTSFÜHRER: DEUTSCHE BANK AG
12489 BERLIN E-MAIL: Support.BNA.EMEA@Bruker.com DIPL.-ING. THOMAS SCHÜLEIN BLZ 100 700 00
GERMANY WWW.BRUKER.COM KONTO 915 162 200
5. Customer Training □
The following section provides a rough guideline for customer training. Agenda should be
adjusted based on the actual needs.
Image mode
Setting up conditions of imaging
Image capturing
In/Out functions
Image processing
Spectra mode
Physical background
Calibration of the spectrometer
Setting up conditions of spectra acquiring
Spectra acquiring
Scaling of spectra displaying
Identification of the elements
Quantification and creation of a quantification method
In/Out functions
Spectra quantification
Automatic mode
Manual mode
User defined quantification
Standard based quantification, if available
Objects mode
Point analysis
Multiple object analysis
Linescan with QLine
Mapping with QMap
Report function
Creating reports
Creating templates
Print to PDF
Export to WORD, if available
Energy-Channel calibration
BRUKER Nano GMBH TELEFON: +49 30 6709 90 -0 AMTSGERICHT BLN. CHARLOTTENBURG HRB 99662 BANKVERBINDUNG:
AM STUDIO 2D TELEFAX: +49 30 6709 90 -30 GESCHÄFTSFÜHRER: DEUTSCHE BANK AG
12489 BERLIN E-MAIL: Support.BNA.EMEA@Bruker.com DIPL.-ING. THOMAS SCHÜLEIN BLZ 100 700 00
GERMANY WWW.BRUKER.COM KONTO 915 162 200