Professional Documents
Culture Documents
2060 MARGA - Application Manual - V2
2060 MARGA - Application Manual - V2
Application Manual
Metrohm Applikon B.V.
Schiedam
The Netherlands
Phone +31 (0)10-2983555
Fax +31 (0)10-4379648
041-info@metrohm.com
www.metrohm.com/en/products/process-analyzers/
2060 IC MARGA
Application Manual
/
November 2020
Rob Proost
Metrohm Applikon B.V.
Schiedam
Table of contents
1 General description 1
1.1 Safety Instructions ................................................................ 1
1.1.1 General notes on safety ........................................................... 1
1.1.2 Electrical safety ........................................................................ 1
1.1.3 Working with glass analyzer parts ............................................ 2
1.1.4 Tubing and capillary connections ............................................. 3
1.1.5 Potentially hazardous solvents and chemicals ........................... 3
1.1.6 Recycling and disposal ............................................................. 4
2 Analysis 5
2.1 Operating Principle ............................................................... 5
2.2 Eluents and Reagents ........................................................... 8
2.2.1 Chemicals and Apparatus ........................................................ 9
2.2.2 Cleaning procedure containers ............................................... 10
2.2.3 Eluents .................................................................................. 11
2.2.4 Reagents ............................................................................... 12
2.2.5 Internal Standard ................................................................... 13
3 Hardware 15
3.1 Analyzer Cabinets ............................................................... 15
4 Program 19
4.1 Program ............................................................................... 19
5 Typical results 20
5.1 Retention times (t R ) ......................................................... 20
5.2 Typical Chromatograms ..................................................... 21
6 Maintenance schedule 23
6.1 Maintenance schedule ....................................................... 23
6.2 Eluent replacement ............................................................. 24
6.3 Reagent and Standard replacement ................................. 26
7 Troubleshooting 28
7.1 Bacterial growth ................................................................. 28
7.2 Missing or no peaks in chromatograms ........................... 29
7.3 Anion IC - Anion Chromatograms ..................................... 31
7.4 Cation IC - Cation Chromatograms ................................... 32
■■■■■■■■ III
2060 IC
Table of contents ■■■■■■■■■■■■■■■■■■■■■■
8 Appendix 33
8.1 Extension 2060 MARGA standard application with Flu-
oride ..................................................................................... 33
8.2 Wet rotating denuder module (V2LH000343) ................. 39
8.3 Steam jet aerosol collector module (V2LH000022) ......... 43
8.4 Ion chromatography ........................................................... 48
IV ■■■■■■■■
2060 IC
■■■■■■■■■■■■■■■■■■■■■■ General description
1 General description
WARNING
WARNING
This analyzer has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
analyzer, the following instructions must be observed carefully.
WARNING
WARNING
■■■■■■■■ 1
Safety Instructions ■■■■■■■■■■■■■■■■■■■■■■
WARNING
Mains voltage
WARNING
CAUTION
WARNING
WARNING
2 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ General description
CAUTION
Leaks in tubing and tubing connections are a safety risk. Tighten all
connections well by hand. Avoid applying excessive force to tubing
connections. Damaged tubing ends lead to leakage. Appropriate
tools can be used to tighten or loosen connections.
Check the connections regularly for leakage. If the analyzer is used
mainly in unattended operation, then weekly inspections are manda-
tory.
WARNING
WARNING
WARNING
Do not use flammable liquids inside the wet part of the ana-
lyzer and avoid the generation of flammable gasses due to
chemical reactions.
■■■■■■■■ 3
Safety Instructions ■■■■■■■■■■■■■■■■■■■■■■
4 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Analysis
2 Analysis
The following scheme shows the air flow and liquid flow path of the
2060IC MARGA. The green lines represent the air flow, the blue lines the
liquid flow.
■■■■■■■■ 5
Operating Principle ■■■■■■■■■■■■■■■■■■■■■■
6 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Analysis
■■■■■■■■ 7
Eluents and Reagents ■■■■■■■■■■■■■■■■■■■■■■
The addition of internal standard to the WRD and SJAC samples makes
accurate anion and cation IC data validation possible.
The air flow is regulated by a critical orifice which is integrated in the air
outlet connector.
The anion IC, cation IC, gas and aerosol results as well as all other hard-
ware parameters can be monitored using the 2060 and MagIC software.
NOTICE
NOTICE
8 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Analysis
NOTICE
Before refilling the containers with fresh eluents and reagents, con-
tainers should be cleaned and disinfected to prevent contamination.
See procedure Cleaning procedure containers (see chapter 2.2.2,
page 10)
NOTICE
For preparation of the described eluents and reagents use of Pro Ana-
lyse grade chemicals and UltraPure water (resistance of >18.2 MΩ) is
required.
Chemicals
■ Phosphoric acid (H3PO4) 85% (1.685 g/ml ~ 14.6 mol/l).
■ Nitric acid (HNO3), 2 mol/l (ready to use from Merck or Sigma).
■ Sodium carbonate (Na2CO3), 0.5 mol/l (ready to use from Merck or
Sigma).
OR
Sodium carbonate monohydrate (Na2CO3.H2O) in case you need/want
to prepare the stock solution for the anion eluent yourself.
■ Hydrogen peroxide (H2O2) 30% or 35%.
■ Potassium iodate (KIO3).
■ Lithium Bromide (LiBr)
■ Acetone (Propanon)
■ UltraPure water (resistance of >18.2 MΩ)
■ Hellmanex™ III, alkaline glass cleaning solution (dilute 100x before use,
1.5 ml in 150 ml Ultra Pure Water).
Apparatus
■ Volumetric flask of 250 ml with stopper.
■ Adjustable (recently calibrated) pipettes (range 0.00 -10.00 ml).
■ Analytical balance with an accuracy of 1 mg.
■ Balance with a maximum of at least 12 kg and accuray of 10 gram.
■ Plastic funnel to fit in volumetric flasks.
■ Plastic sample containers of 100 ml, 2 l, 5 l and 10 l.
■■■■■■■■ 9
Eluents and Reagents ■■■■■■■■■■■■■■■■■■■■■■
Cleaning containers
3 Empty container.
6 Empty container.
NOTICE
9 Empty container.
NOTICE
10 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Analysis
2.2.3 Eluents
Carefully read this complete section before you continue with the prepara-
tion of reagents.
2 Fill up to the 250 ml mark with ultrapure water and mix gently.
Anion IC eluent
NOTICE
Only use clean anion eluent containers for the preparation of anion
eluent. Cleaning procedure containers (see chapter 2.2.2, page 10)
OR
2 Add approx. 200 ml with ultrapure water, close with a cap and mix
gently till all salts are dissolved.
■■■■■■■■ 11
Eluents and Reagents ■■■■■■■■■■■■■■■■■■■■■■
3 Bring all liquid to the 10 l anion eluent container and rinse the flask 3
times with ultra pure water, add all rinses to the anion eluent con-
tainer.
4 Add 1ml potassium iodate (KIO3), supplement stock solution, into the
10 l anion eluent container.
Cation IC eluent
NOTICE
Only use clean cation eluent containers for the preparation of cation
eluent. Cleaning procedure containers (see chapter 2.2.2, page 10)
1 Pipette 20.0 ml nitric acid solution (HNO3 2 mol/l) into the 10 l cation
eluent container.
2.2.4 Reagents
Carefully read this complete section before you continue with the prepara-
tion of reagents.
NOTICE
12 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Analysis
NOTICE
Concentration: 1% H2O2
■■■■■■■■ 13
Eluents and Reagents ■■■■■■■■■■■■■■■■■■■■■■
Concentration: 4000.0 mg/L LiBr contains 320.0 mg/L Li+ and 3680.0
mg/L Br-
Storage: This solution is stable for 24 months if closed and stored in a
dark place at room temperature.
3 Close the volumetric flask with the stopper and mix gently.
NOTICE
Only use clean internal standard solution containers for the prepara-
tion of internal standard solution. Cleaning procedure containers (see
chapter 2.2.2, page 10)
Concentration: 4000.0 µg/L LiBr contains 320.0 µg/L Li+ and 3680.0 µ/L
Br-
14 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Hardware
3 Hardware
■■■■■■■■ 15
Analyzer Cabinets ■■■■■■■■■■■■■■■■■■■■■■
IMPORTANT: The tubing used to feed the steamer (black color), connect-
ing the tubing pump (2A) and the steamer (1B) is called the Steamer Feed
Tubing (part nr: V0M5217130).
The basic cabinet contains 12 wet part module positions, here called 1A
(left top) till 3D (right bottom).
1A: SJAC Module with vacuum sensor.
This module is used for the electronics of the SJAC sensor, the electronics
for the SJAC heater and the vacuum sensor of the vacuum/air pump.
1B: Blind plate.
Here the Steam Jet Aerosol Collector with heater, level sensor and filter are
positioned.
1C: WRD module.
This is the Wet Rotating Denuder module with rotation mechanism, level
sensor and sample filter
1D: Blind plate.
2A: Multi speed tubing pump module.
This module is the SJAC steamer feed tubing pump that feeds the absorb-
ance liquid for the Steam Jet Aerosol Collector. The tubing pump flow is
0.85 ml/min for default critical orifice of 1 m3/h. (Or 0.42 ml/min for
optional critical orifice of 0.5 m3/h).
2B: Double pipette module.
This module contains two glass Sample mixing vessel modules, one to mix
and store the aerosol liquid sample (at the left) and one to mix and store
the gas liquid sample (at the right) both collected in the previous hour.
2C: Burette Module with 20 ml dosing unit.
This module is the Wet Rotating Denuder (WRD) burette for collecting gas
sample (connection port-3), adding internal standard LiBr (connection
port-4), air for mixing in sample mixing vessel module (connection port-2)
and WRD pipette (connection port-1).
2D: Multi speed tubing pump module.
This module is the Wet Rotating Denuder (WRD) fill tubing pump that
feeds the absorbance liquid, controlled by the WRD level sensor, to main-
tain stable water level in the WRD. The flow (when continues running) is
6.9ml/min.
3A: Multi speed tubing pump module.
This module is the Steam Jet Aerosol Collector (SJAC) fill tubing pump that
feeds the absorbance liquid, controlled by the SJAC level sensor, to main-
16 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Hardware
tain stable water level in the SJAC. The flow (when continues running) is
6.9ml/min.
3B: Burette Module with 20ml dosing unit.
This module is the Steam Jet Aerosol Collector (SJAC) burette for collecting
aerosol sample (connection port-3), adding internal standard LiBr (connec-
tion port-4), air for mixing in sample mixing vessel module (connection
port-2) and SJAC pipette (connection port-1).
3C: Burette Module with 10ml dosing unit.
This module is the Ion Chromatograph (IC) burette for transfer of gas liq-
uid sample (connection port-1), or aerosol liquid sample (connection
port-2), or external standard (connection port-4) to the cation and anion
IC (connection port-3).
3D: Blind plate.
■■■■■■■■ 17
Analyzer Cabinets ■■■■■■■■■■■■■■■■■■■■■■
2: On the bottom, left half of the IC cabinet is the cation Ion Chromato-
graph (IC) located, with high-pressure pump, purge valve, injector with
200 µl sample loop and pulsation dampener.
3: On the bottom right half of the IC cabinet is the anion IC located, with
high pressure pump, purge valve and injector with 100 µl sample loop,
pulsation dampener, chemical suppressor with peristaltic pump for regen-
eration and eluent CO2 remover.
4: This is the chemical suppressor regeneration (peristaltic) pump. During
analysis this pump is active/on, the speed is set to 1 in MagIC (corresponds
with 0.1 ml/min).
5: This chemical suppressor is a cation exchanger: All cations are
exchanged for hydrogen ions (H+), suppressing the eluent conductivity
(carbonate eluent becomes water and CO2) and strengthen the anion ions
to be determined.
6: The CO2 suppressor consists of a special membrane, which removes
CO2 eluent from the liquid before detection, reducing the eluent conduc-
tivity and improving the linearity of the calibration lines of the components
to be analyzed.
7: At the right side of the anion IC the column oven is located. Inside the
column oven there is room for two analytical column with pre-columns,
one for anion and one for cation analysis. The column oven houses two
preheating coils to preheat both eluents to the application temperature
(default 45 °C).
8: The cation conductivity detector (used with direct conductivity detec-
tion) is located behind the white thermic isolation cover. The cation detec-
tor is located at the top.
9: The anion conductivity detector (used with sequential suppressed con-
ductivity detection) is located behind the white thermic isolation cover.
The anion detector is located at the bottom.
18 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Program
4 Program
4.1 Program
NOTICE
The original programs (2060 and MagIC Net) are stored on the USB-drive
that came with the analyser.
■■■■■■■■ 19
Retention times (tR) ■■■■■■■■■■■■■■■■■■■■■■
5 Typical results
This chapter shows and explains the results you might expect from your
2060IC MARGA analyzer.
NOTICE
The results of your analyzer and the expected results may differ and
can be caused by:
■ Use of different separation columns.
■ Quality of eluents, reagents and standards.
Status of the complete analyzer as a result of regular maintenance
or lack of regular maintenance.
Column lifetime
The column lifetime depends on the eluent quality and the general main-
tenance of the entire 2060IC MARGA including the sampling device. The
level of air polution influences (decreases) the column life time.
The typical lifetime of a column is two to four months. It is recommended
to replace the column every three months.
20 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Typical results
■■■■■■■■ 21
Typical Chromatograms ■■■■■■■■■■■■■■■■■■■■■■
Aerosol phase
22 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance schedule
6 Maintenance schedule
This chapter describes all the maintenance steps for the 2060IC MARGA
analyzer. The maintenance of a Metrohm Process Analytics analyzer
depends on many variables:
■ Kind of sample and used application
■ Number of cycles per hour (or day)
■ Importance to the customer (e.g. following a trend or the need for
absolute/exact values)
CAUTION
Basic Cabinet
Table 7 Wet Part Cabinet
Description 2 weeks 1 month 6 months 12 months
■■■■■■■■ 23
Eluent replacement ■■■■■■■■■■■■■■■■■■■■■■
CAUTION
NOTICE
24 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance schedule
6 Hold the cap with the tubing and filter in the air.
7 Insert tubing into the fresh eluent container and tighten cap.
6 Hold the cap with the tubing and filter in the air.
7 Insert tubing into the fresh eluent container and tighten cap.
■■■■■■■■ 25
Reagent and Standard replacement ■■■■■■■■■■■■■■■■■■■■■■
CAUTION
Wear single use powder free gloves when replacing reagents and
standards.
6 Hold the cap with the tubing and filter in the air.
26 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance schedule
6 Hold the cap with the tubing and filter in the air.
6 Hold the cap with the tubing and filter in the air.
7 Insert tubing into the fresh internal standard container and tighten
cap.
■■■■■■■■ 27
Bacterial growth ■■■■■■■■■■■■■■■■■■■■■■
7 Troubleshooting
This chapter describes troubleshooting for the standard 2060IC MARGA
application.
NOTICE
NOTICE
Wear single use powder free gloves when handling liquids or per-
forming maintenance or repairing the system.
Peroxide is used to prevent bacterial grow, but when the bacterial grow is
there a higher percentage (0.1%) peroxide is only working at the very
beginning of the bacterial presence. 0.1% per-acetic acid is needed to kill
the bacteria.
28 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Troubleshooting
CAUTION
Do not keep the system in analysis when you use higher concentra-
tions of peroxide or per-acetic acid. It will destroy your columns!!!
Clean the complete system and do not forget the denuder caps: quite
often you will find somekind of bacteria based slurry. When present, take
out the denuder seals and clean the caps and seals manually with a tissue,
before you start the cleaning procedure rinse without analyses (no injec-
tions). All components including inlet, tubings, WRD, SJAC, dosino’s and
injection loops (fill position) must be cleaned and rinsed. After cleaning
the system against bacteria, first rinse the system thoroughly with normal
absorbance liquid before you restart the measurement to save your col-
umns.
■■■■■■■■ 29
Missing or no peaks in chromatograms ■■■■■■■■■■■■■■■■■■■■■■
SJAC liquid level too low. Clean SJAC level sensor with a tissue.
Missing Lithium/ Internal standard container Prepare fresh internal standard and replace
Bromide peak in is (almost) empty. container.
chromatogram
Possible leakage (air or liq- Check all connections for possible leakages (air
uid). or liquid).
No steam gener- Absorbance liquid contain- Replace with a full fresh prepared container.
ated. ers (almost) empty.
30 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Troubleshooting
Baseline conductiv- IN/OUT suppressor not con- Correct connections - label IN is for column,
ity changes in every nected correctly. label OUT for CO2 remover (MCS)
chromatogram
Retention time (tR) Leakage on high pressure Replace both piston seals in anion pump
shifts in every chro- pump
matogram
No leakage Replace inlet and outlet valves in anion pump
■■■■■■■■ 31
Cation IC - Cation Chromatograms ■■■■■■■■■■■■■■■■■■■■■■
Retention time (tR) Leakage on high pressure Replace both piston seals in cation pump
shifts in every chro- pump
matogram
No leakage Replace inlet and outlet valves in cation pump
32 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
8 Appendix
NOTICE
Carefully read this complete section before you continue with the
preparation of reagents.
Column
■ 6.1006.620 - Metrosep A Supp 7 - 150/4.0.
■ 6.1011.030 - Metrosep RP 2 Guard/3.5.
NOTICE
For preparation of the described eluents and reagents use of Pro Ana-
lyse grade chemicals and UltraPure water (resistance of >18.2 MΩ) is
required.
Apparatus
■ 6.2832.000 - MSM rotor A.
■ 6.2842.020 - Adapter sleeve for Suppressor Vario.
■ Volumetric flask of 250 ml with stopper.
■ Adjustable (recently calibrated) pipettes (range 0.00 -10.00 ml).
■ Analytical balance with an accuracy of 1 mg.
■ Balance with a maximum of at least 12 kg and accuray of 10 gram.
■ Plastic funnel to fit in volumetric flasks.
■ Plastic sample containers of 250 ml.
■■■■■■■■ 33
Extension 2060 MARGA standard application with Fluoride ■■■■■■■■■■■■■■■■■■■■■■
Chemicals
■ Phosphoric acid (H3PO4) 85% (1.685 g/ml ~ 14.6 mol/l).
■ Nitric acid (HNO3), 2 mol/l (ready to use from Merck or Sigma).
■ Sodium carbonate (Na2CO3), 0.5 mol/l (ready to use from Merck or
Sigma).
OR
Sodium carbonate monohydrate (Na2CO3.H2O) in case you need/want
to prepare the stock solution for the anion eluent yourself.
■ Hydrogen peroxide (H2O2) 30% or 35%.
■ Oxalic acid dihydrate (C2H2O4.2H2O)
■ UltraPure water (resistance of >18.2 MΩ)
2 Fill up to the 250 ml mark with ultrapure water and mix gently.
NOTICE
Only use clean anion eluent containers for the preparation of anion
eluent. Cleaning procedure containers (see chapter 2.2.2, page 10)
2 Add approx. 200 ml with ultrapure water, close with a screw cap and
mix gently till all salts are dissolved.
34 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
4 Add approx. 200 mL with ultrapure water, close with a screw cap
and mix gently till all salts are dissolved.
5 Bring all liquid into the 10 l anion eluent container and rinse both
250 ml plastic containers 3 times with ultra pure water, add all rinses
to the anion eluent container.
6 Add 1 ml Oxalic acid (C2H2O4) stock solution into the 10 l anion elu-
ent container.
■■■■■■■■ 35
Extension 2060 MARGA standard application with Fluoride ■■■■■■■■■■■■■■■■■■■■■■
F aerosol concentration
36 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
■■■■■■■■ 37
Extension 2060 MARGA standard application with Fluoride ■■■■■■■■■■■■■■■■■■■■■■
38 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
5 Level sensor
The WRD consists of two concentric glass tubes, forming an annulus pas-
sage. These tubes rotate at 8 revolutions per min, forming a continuous
liquid film on the inside of the outer cylinder and the outside of the inner
cylinder. Air is drawn in and, due to high diffusion coefficients, close to
100% of acid gases and ammonia are stripped from the air mass. The
resulting solution of gases is continuously sampled. Due to the geometry
and air velocity a laminar flow of the air within the annulus is realized.
Aerosols pass through to the WRD.
It is desired that the WRD module is positioned on the upper row of the
wet part. This improves both the accessibility of the modules as well as the
performance of the instrument; less dead volume due to shorter tubing.
Connections
■ Wet connections
– The sample outlet can be connected to a sampling module (eg.
pump, burette module etc.) with an M6 flared tubing.
– The WRD feed can be connected to an absorbance tubing pump
which is controlled by the level sensor.
■■■■■■■■ 39
Wet rotating denuder module (V2LH000343) ■■■■■■■■■■■■■■■■■■■■■■
■ Air connections
– The air inlet can be connected to an air inlet device using a 1/2""
PE tubing. Make sure that the inlet tube ferrules are installed cor-
rect. The inlet tube must be pushed completely into the denuder
cap. This is to avoid dead volume and disturbance of the laminar
flow.
■ Electrical connections
– The WRD motor is connected to a digital output on the wet part
I/O controller.
– The level sensor is connected to a digital input on the wet part
I/O controller.
Basic maintenance
CAUTION
Wear single use powder free gloves during the following procedures
2 Remove Denuder
Carefully remove the glass denuder from the wet part by removing
the right denuder cap.
Inspect the Denuder Bearing Caps for debris or particles. If required
demount the denuder caps from the wet part and rinse with ultra-
pure water from a rinse bottle.
40 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
tissues to clean the ends of the denuder of any Teflon powder left
behind from wear of the seals.
1 Insert one end of the denuder into the first cleaning cap assembly.
2 Remove the PVC screw cap of the second cleaning cap assembly and
mount it on the other end of the denuder.
3 Stand the denuder vertically on a flat and stable surface and fill with
150 ml of sufficient cleaning reagent.
Hellmanex™ III, alkaline glass cleaning solution (dilute 100x before
use, 1.5 ml in 150 ml Ultra Pure Water).
4 Close the top cleaning cap assembly with its PVC screw cap.
9 Fill with 150 ml ultrapure water and replace the end cap.
■■■■■■■■ 41
Wet rotating denuder module (V2LH000343) ■■■■■■■■■■■■■■■■■■■■■■
CAUTION
The WRD sample filter is located at the left side of the denuder.
2 Rinse the new filter twice with 5 ml ultrapure water using a syringe.
NOTICE
When rinsing the filter with water, be sure that no air bubbles
introduced into the filter.
3 Unscrew the female luer lock located on the right side of the current
WRD sample filter.
4 Disconnect the male luer located on the left side of the current WRD
sample filter.
5 Screw the female luer lock located on the right side onto the new
activated WRD sample filter.
6 Connect the male luer located on the left side onto the new activa-
ted WRD sample filter.
Specifications
Contact material Borosilicate glass, PET-P, PTFE (seal)
42 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
■■■■■■■■ 43
Steam jet aerosol collector module (V2LH000022) ■■■■■■■■■■■■■■■■■■■■■■
13 Level Sensor
44 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
Connections
■ Wet connections
– The Sample Filter can be connected to a sampling module (WRD
and SJAC) with a luer -M6 adapter and M6 flared tubing.
– The SJAC level addition connection can be connected to an
absorbance tubing pump, which is controlled by the Level sen-
sor, with an M6 flared tubing.
– The SJAC steamer is connected to an absorber liquid pump with
a Norprene® tubing (V0M5217130; Norprene® tubing ID
1.6mm OD 4.8mm)
– The Air out is part of the Air flow regulation. The Air flow is
regulated by a Critical Orifice, which is integrated in the air out-
let connector.
■ Electrical connections
– The SJAC steamer is connected to a PID control output on the
Wet part I/O controller.
– The Level sensor is connected to a digital input on the Wet part
I/O controller. This sensor can regulate a digital output.
■■■■■■■■ 45
Steam jet aerosol collector module (V2LH000022) ■■■■■■■■■■■■■■■■■■■■■■
5 6
Basic maintenance
CAUTION
Wear single use powder free gloves during the following procedures
2
CAUTION
NOTICE
46 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
CAUTION
The SJAC sample filter is located at the left front side of the SJAC.
2 Rinse the new filter twice with 5 ml ultrapure water using a syringe.
NOTICE
When rinsing the filter with water, be sure that no air bubbles
introduced into the filter.
4 Disconnect the male luer connector located on the outlet side of the
current SJAC sample filter.
6 Connect the male luer connector onto the new activated SJAC sam-
ple filter.
Specifications
Contact material Borosilicate glass, PET-P, SS, PTFE, Silicone
■■■■■■■■ 47
Ion chromatography ■■■■■■■■■■■■■■■■■■■■■■
48 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
Cation IC
IC Cabinet
■■■■■■■■ 49
Ion chromatography ■■■■■■■■■■■■■■■■■■■■■■
Absorber tubes
Fill the anion absorbing tube on the eluent container with cotton wool
and soda lime. The soda lime will absorb the CO2 from the air which pre-
vents lowering of the pH of the eluent (Na2CO3). If the pH of the anion
eluent lowers because of CO2 the separation of the components on the
anion column will change (become worse/slower).
Fill the cation absorbing tube on the eluent container only with cotton
wool to prevent dust falling in.
Part number: 6.2837.100 - CO2 adsorption cartridge CW
Eluent degasser
Unstable baseline or shifting retention times indicates eluent flow disturb-
ance. This can be caused by a malfunctioning eluent degasser.
Contact your Metrohm organisation for a repair.
50 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Appendix
IC Inline filter
The inline filter protects the columns, both separation and guard, for con-
taminations/ particles by the eluent. Change the filters every 6 months, or
when those are dirty and the eluent system pressure goes up. When the
suppressor liquid flow stops dripping in the waste bucket or slows down,
the inline filter has to be changed.
Part number: 6.2821.130 - Spare filter for inline filter, 10 pieces
Injection Valve
The injection loops can become dirty from sample in the inside of the
tube, interfering the chromatogram. Replace the anion and cation sample
loop every year for a new one.
Part number: V2LH001061 - Sample Loop PEEK 100 µl (used in anion IC)
Part number: V2LH001071 - Sample Loop PEEK 200 µl (used in cation IC)
Pre-Column
The 2060IC MARGA uses RP2 pre-column as guard columns. Change
those guard columns (for anion and cation) every month, or when the elu-
ent pressure moves slowly up. The RP2 housing can be reused; there is a
replacing set of 10 RP2 discs with filters.
Part number: 6.1011.030 - Metrosep RP 2 Guard/3.5
Part number: 6.1011.130 - Spare filter for RP 2 Guard/3.5, 10 pieces
Column
The quality of the separation column of anion and cation is normally
slowly decreasing till the end of it’s lifetime. This can be seen on peak
shape (fronting or tailing), broadening peaks, loss of capacity (lower reten-
tion times) or increasing back pressure. As long as the peaks are baseline
separated and the pressure is inside the limits, the column can still be
used.
An average lifetime is 2-4 months of continues analyses. Dirt and bacteria
will shorten it‘s lifetime.
Part number: 6.1006.640 - Metrosep A Supp 7 - 150/2.0, anion column
Part number: 6.0105.1220 - Metrosep C 6 - 150/2.0, cation column
MSM suppressor
The 2060 IC MARGA uses 2 mm ID separation columns. For proper opera-
tion a MSM LC-A (low capacity for anion separations) suppressor is used.
It exchanges all cations in the eluent and sample in H+ resulting in lower
baseline and enhancing sensitivity of the components.
■■■■■■■■ 51
Ion chromatography ■■■■■■■■■■■■■■■■■■■■■■
Using phosphoric acid, instead of sulphuric acid, the heavy metals present
in the airosols are also removed from the suppressor.
Part number: 6..2844.000 - MSM-LC Rotor A
MCS suppressor
The MCS (Metrohm CO2 Suppressor) is a special designed degasser to
remove the CO2 from the eluent and sample, effecting in strongly reduced
system-peak in anion chromatogram and improving linearity of the cali-
bration line.
The MCS uses a CO2 adsorption cartridge to adsorb the CO2. When the
CO2 adsorption cartridge becomes wet (moisture in the inside), it has to
be replaced.
Contact your Metrohm organisation for a repair.
52 ■■■■■■■■