Professional Documents
Culture Documents
H1187A Student
H1187A Student
H1187A Student
System Operation
Course Number H1187A
Student Manual
P Series Micro GC and Data
System Operation
Course Number H1187A
Student Manual
ii
Table Of Contents
Module Descriptions 3
Micro GC 11
Carrier gas and gas control 11
The injector 11
The separation system 12
Backflush Modules 12
Detection system 12
Data system 13
Separation Processes 15
Qualitative analysis 16
Quantitative analysis 16
Micro GC Hardware 25
P-Series Micro GCs 26
External Inputs/Outputs 27
Agilent M-Series Micro GCs 28
Laboratory Exercises 34
Module 2 Review 37
iii
MODULE 3: EZCHROM INTRODUCTION 39
Introduction to EZChrom 40
Windows Compatibility 41
Software Installation 43
MTI GC Setup 49
MTI GC Verification 51
Initiating EZChrom 56
Instrument Menu 59
Math Menu 61
Lab Exercises 64
iv
Part 4: Viewing the Contents of a File. 82
Module 3 Review 90
GC Injectors 136
GC Hardware 139
v
Heated versus Unheated Injector 175
Columns 182
GC Hardware 183
Detectors 230
vi
Module 7 Review 246
Reporting 269
Sequences 273
Troubleshooting 292
vii
Communication Error Numbers 304
viii
MODULE 1: Book Introduction
This module describes the contents of this book. The sections are:
• Using this Book
• Module Descriptions
• Instructional Materials Required
1
MODULE 1: Book Introduction
Using this Book
2
MODULE 1: Book Introduction
Module Descriptions
Module Descriptions
3
MODULE 1: Book Introduction
Module Descriptions
4
MODULE 1: Book Introduction
Module Descriptions
Module 6: Columns
Micro GC Attributes
Column Types
Fundamentals of Separation Theory
Common Micro GC Applications
Reference for Module 6: GC User Manual; Chapter 5
Lab Exercises
Lab 1: Varying Temperature and Flow rate
Lab 2: Calculating Column Efficiency
Lab 3: View various applications from Micro GC Website.
Module 7: Detector
Fundamentals of the Thermal Conductivity Detector
Reference for Module 7: GC User Manual; Appendix A
Lab Exercises
Lab 1: Modifying the Detector Sensitivity
5
MODULE 1: Book Introduction
Module Descriptions
6
MODULE 2: Intro to Agilent P-Series
Micro GC
Figure 1
8
MODULE 2: Intro to Agilent P-Series Micro GC
Basic Gas Chromatography
Electrometer
Gas Trap Detector
Column
PC
Sample Vent
Carrier Gas
Figure 2
9
MODULE 2: Intro to Agilent P-Series Micro GC
Basic Gas Chromatography
Definitions
• Gases
Carrier Gas -- Pressurized gas used to transport the sample through
the system
• Sample Introduction
Introduces the sample to the carrier gas stream with minimal
disruption of the gas stream
• Column
Achieves separation of the components in the sample
• Detector
Recognizes and responds to sample components as they elute from
the column
• Data Acquisition
Converts the detector signal to a picture chromatogram and provides
manual or automated determination of the identity and amounts of the
sample components
Figure 3
The first major component of the GC is the carrier gas. A carrier gas supply is
always present and usually consists of helium, nitrogen, or argon. The function of
the carrier gas is to carry the sample through the system. The gas one chooses
depends on the specific application. The gases are commonly supplied by
compressed gas cylinders. Gas generators are also an option as a gas source for
bench top units.
The next major component part is the sample introduction process. It usually
consists of an injection port but can also include valves. The purpose of the inlet
is to introduce the sample into the carrier gas stream.
The next major component part is the column. The purpose of the column is to
provide the separation of the components in the sample mixture.
The Thermal Conductivity detector is a device that senses the presence of
components different from the carrier gas and converts the information to an
electrical signal.
The data acquisition device is a PC that uses software to convert the detector
signal to a peak chromatogram and report.
10
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC
Micro GC
Micro GC Diagram
Pressure
Carrier In Regulator Sample Loop
Switching Valve
Pressure
Transducer
Sample In Vacuum
Pump Sample Vent
Out
Micro Injector
Analytical Column
Column Heater
Reference Column
Figure 4
The injector
The second major component is the injector. The injector introduces a measured
amount of sample into the inlet of the analytical column. The Micro GC contains
a timed injector micro machined from a silicon wafer using manufacturing
techniques borrowed from the semiconductor industry. The injection sequence
begins when the carrier gas for the reference column flows through the injector.
11
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC
Two flow restriction channels moderate the flow of carrier gas to both the
reference and analytical columns. To take a sample, the sample microvalve in the
injector opens allowing the internal vacuum pump to pull sample gas through the
sample loop and out the vent. The sample microvalve then closes, the vacuum
pump turns off and a switch valve connects to the outlet of the pressure regulator,
thereby pressurizing the sample loop. Because of the flow restriction channels, the
pressure in the sample loop is slightly higher than the pressure at the inlet of the
analytical column. Therefore, when the inject microvalve opens, the sample from
the loop flows into the channel leading to the analytical column. The amount of
sample injected depends on the length of time the inject microvalve is open. After
a user defined time (typically about 40 msec), the inject valve is closed and the
switch valve connects to the vacuum pump. The injected sample flows into the
analytical column, where the sample is separated into its components.
Backflush Modules
A GC module with a precolumn backflush-to-vent configuration has two
columns—a short precolumn and a longer analytical column. The backflush valve
lies between the precolumn and the analytical column. The stationary phase and
length of the precolumn is chosen so that, at the operating temperature of the
columns, undesirable sample components that interfere with the analyses are
retained on the precolumn stationary phase. The sample components for
quantification have minimal affinity for the stationary phase of the precolumn,
and pass quickly through the precolumn. After the last sample component for
quantification exits the precolumn and enters the analytical column, the
backflush-to-vent valve located between the precolumn and the analytical column
opens. The carrier gas flow through the precolumn is reversed and backflushes the
undesirable sample components off the precolumn. The sample components for
quantification continue to flow through the analytical column for separation and
detection. At the end of the analytical run, the backflush GC module is fully
purged of all sample components and is ready for the next analysis.
Detection system
The fourth Micro GC component is a detector. A detector monitors the carrier gas
and senses a change in its composition when a component in the sample elutes
from the column. One of the most common, reliable and easy to use detectors is
12
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC
the universal micro machined Solid State Detector (SSD). The SSD is a micro
machined version of the Thermal Conductivity Detector (TCD). The design of
this detector is derived from a Wheatstone Bridge, which is an electronic device
that compares the electrical resistance of two branches in a circuit. In an SSD,
pure carrier gas elutes from the analytical column and passes over two matched
filaments. The bridge compares their resistance against those of the reference
side. When both sides have pure carrier gas flowing over the filaments, the bridge
is balanced and the output is zeroed. When a component elutes, the thermal
conductivity on the sample side of the bridge changes. Therefore, the resistance of
the analytical side changes and the output indicates a peak is eluting. A regulated
power supply controls the voltage being fed to the filaments at a constant level.
The bridge also has a balance control and an autozero circuit to zero the detector
output between runs. The user can select high, medium or low sensitivity levels
which changes the gain level of the electronics. This is unlike standard TCDs
where sensitivity increases by increasing filament temperature and corresponds
with a decrease in detector life.
Data system
The final or fifth component of a Micro GC is the data system. The main purpose
of a gas chromatographic system is to generate both qualitative and quantitative
data. Data systems provide both a visual recording of the detector output and an
area count of the detector response. The detector response is used to identify the
sample composition and measure the amount of each component by comparing
the area counts of the sample to the analysis of a known calibration standard.
Computer based data systems can communicate with the analyzers. This added
versatility allows the analyst to store and set operating conditions for an analysis
in the computer, making the data system and the analyzer an integrated
instrument.
13
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC
Micro GC Column
Analytical
Column
Reference
Column
Heater/
Shroud
Figure 5
Columns are 4 to 14 meters long and 0.15 to .5 mm in diameter. For the function
of the Thermal Conductivity detector, both a reference and analytical column is
used.
14
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
Separation Processes
Typical Chromatogram
Figure 6
15
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
Qualitative analysis
Qualitative analysis is designed to identify the components of a sample. Gas
chromatography is not an absolute qualitative tool and usually must rely on other
techniques for peak identification. However, a Micro GC containing two GC
modules can separate the components of a sample with two different stationary
phases, simultaneously. The result is peak identification and confirmation in a
single analysis. To analyze qualitatively, a calibration standard of known
composition must be analyzed first. Pure compounds elute at the same time under
exactly the same analytical conditions. If you match the calibration standard peaks
with those in your samples, you can be fairly sure that you have those compounds
in your sample.
Quantitative analysis
Quantitative analysis is designed to determine the amount or proportions of the
components of a sample based on the detector response. The detector response
(area or height) is proportional to the amount of a component in a sample.
However this response varies slightly with each component. Therefore, it is
necessary to experimentally determine the area (or height) concentration
relationship by running a calibration standard containing the components of
interest. The conditions in which the detector response for the calibration standard
must be the same as those in the unknown samples.
16
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
Flow
A
Figure 7
The column is responsible for the separation of the components in the mixture.
At the time of entering the column, the components are in a homogeneous
mixture. As the components pass through the column, certain components have
more affinity for the material inside the column and thus are retained longer than
other components. The components separate based upon this differential
interaction with the column material. Those components eluting first have the
least affinity for the column material, and those eluting last have the most
interactions with the column.
There are many different types of columns available. One chooses the column
type relative to the type of components requiring separation.
Gas chromatography is an analytical technique used for the separation of the
sample components between two phases; a mobile phase (the carrier gas) and a
stationary phase (the column packing or coating). The separation begins when a
gaseous sample fills the sample loop of the injector. The sample is then injected
from the loop onto the column. Carrier gas (mobile phase) transports the sample
down the length of the column. Because the different components in a sample
have different affinities or solubility, each component spends different amounts of
time in the column packing/coating. When in the packing/coating, a component
does not progress down the column. When diffused out of the packing/coating,
they progress down the column at the same rate as the carrier gas. As a result, the
components spend different amounts of time in the two phases, separating as the
carrier gas moves them through the column.
17
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
18
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
Figure 8
In gas chromatography there is a mobile phase, the carrier gas, and a stationary
phase, the column material. The stationary phase can be a solid material or a
liquid material (a very viscous liquid like a polymer).
19
MODULE 2: Intro to Agilent P-Series Micro GC
Separation Processes
Column Types
Open (Capillary)
Packed
Wall Coated
Open Tube
(WCOT)
Micro
PLOT WCOT
Packed
Length 0.25 - 10 cm 4-8m 4 - 14 m
I.D. 0.5 mm 0.15 - 0.32 mm 0.1 - 0.32 mm
Figure 9
Traditional GC columns were metal or glass tubing packed with a solid material.
Micro GC’s are configured with capillary or open-tubular columns. Capillary
columns are made of fused silica glass in very narrow diameters with their inside
walls coated with a stationary phase of solid or liquid.
20
MODULE 2: Intro to Agilent P-Series Micro GC
Gases and Plumbing
Carrier Gases
10
Figure 10
The gases used should not be reactive with the sample components; therefore,
inert gases are used. Gas purity is an important consideration to avoid
background interference problems. The vendor for your gas cylinders or
compressors can provide the purity levels for the gas provided.
21
MODULE 2: Intro to Agilent P-Series Micro GC
Gases and Plumbing
11
Figure 11
Molecular Sieve traps are recommended for use on all gases. Additional traps
may also be required. Indicator traps for moisture are available. However, caution
should be used since today’s detectors are usually more sensitive to moisture
content than the color change of the indicator.
Plastic tubing should not be used, only copper or stainless steel. Prior to use, the
tubing should be flushed with a solvent like methanol and carrier gas dried.
Filters or traps should be changed regularly; a conservative recommendation is
every 3 cylinders (supply, not internal MicroGC cylinder).
Ambient temperature changes and vibration can cause fittings to leak over time
and thus should be leak checked regularly.
22
MODULE 2: Intro to Agilent P-Series Micro GC
Gases and Plumbing
12
Figure 12
Minimum line pressures should be used in order to insure accurate pressure and
flow control.
23
MODULE 2: Intro to Agilent P-Series Micro GC
Gases and Plumbing
Sample Connections
13
Figure 13
24
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Micro GC Hardware
Micro GC System
14
Figure 14
The MicroGC system includes the GC, PC (desktop or laptop) and printer.
25
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Micro GC Models
15
Figure 15
There are several MicroGC models available. For Information on the various
models, see the Agilent website: http://www.agilent.com
26
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Sampling
Unheated model: Analyze gas samples containing multiple compounds with
boiling points up to 150°C.
Heated model: Analyze gas samples containing multiple compounds with boiling
points up to 220°C.
Injection System
Silicon micro-machined injector
Manual or automatic injection
Internal sampling pump
Injection volumes: 4 µL fixed, 0.5 µL to 15 µL user-selectable
Input 1/16" bulkhead union with built-in 5-µm filter
Detector
Silicon micro machined thermal conductivity detector with 240 nL internal
volume
Minimum detectable quantity: 1 ppm for most compounds
Linear dynamic range: 105±5%, 106±10%
Repeatability: ±2% RSD at constant temperature and pressure
Carrier Gas
Internal carrier gas supply: refillable 300 mL tank, 1800 psi certified for 40 hours
continuous operation.
External carrier gas supply: He, H2, N2, or Ar, 80 psi input to 1/8" Swagelok
fitting
Power
Unheated model: Internal 12 V rechargeable lead-acid battery or 12 VDC
6-8 hours/30 watts maximum.
Heated model: Internal 12 V rechargeable lead-acid battery or 12 VDC
4 hours/60 watts maximum.
External Inputs/Outputs
Analog output: 0 to ±1 VDC and 0 to ±10 VDC
27
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
GC ready out
GC start
External wait ready
External wait ready
RS-232 for instrument status, control and data
Environmental
Operating temperature range: 0 °C to 50 °C
Humidity range: 0 to 95%, non-condensing
Physical
Dimensions
Unheated model: 15 cm (6") x 36 cm (14") x 36 cm (14")
Heated model: 15 cm (6") x 36 cm (14") x 41 cm (16")
Weight
Unheated model: 10.4 kg (23 lb)
Heated model: 13.2 kg (29 lb)
Versatile
The M Series micro GCs are ideal for analyzing vent gas, aqueous and soil
sample headspace, landfill gas, natural gas and furnace gas, as well as performing
leak detection, environmental screening, storage tank analysis, well logging, and
waste-water VOC monitoring in the lab, field or on-line.
28
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
16
Figure 16
Front panel
The front panel of the P200/P200H includes the following:
Power on/off
This button turns power on or off to the entire system including the detector
filaments. The green light on the button acts as an internal battery charge and
power cutoff circuitry indicator.
Caution! Do not power up the GC unless the carrier gas is on. The presence of air
inside the detector cell will cause a rapid rise in the filament temperature. This may result
in irreversible damage to the detector.
Serial port
The serial port of the front panel is the same as the serial port on the back panel of
the P200/ P200H. Two ports are available for convenience purposes. This serial
port is made through a female DB-9 connector, which conforms, to the RS-232
29
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
WARNING Contact with the gas sample inlet on the P200H after it has reached the set
temperature will cause burns.
Carrier on/off
This valve opens and closes the carrier flow from the internal tank to the system
at 80psi.
30
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Cylinder
Fill Port
Carrier
Out
17
Figure 17
The back panel of the P200/P200H is split into two parts to correspond to two
separate compartments inside the unit.
Power 12 VDC
A round black female connector allows you to plug the 12 VDC recharger to the
P200/P200H. If connecting to voltage greater than 115 volts, make sure the power
supply is for the correct line voltage. The 12 VDC power supply recharges the
sealed lead-acid battery inside the P200/P200H whenever the power switch is
turned off. For battery charging instruction, see Chapter 3 of the User’s Manual.
Whenever possible run your P200 with the 12 VDC power supply connected to
the power line. This prolongs the life of your internal battery and eliminates the
need to recharge it often. When the internal battery needs replacing, the
procedures, which should be followed, are covered in Chapter 7 of the User’s
Manual.
31
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Vents
The back panel of the P200 contains four color-coded vents for both column A
and column B (sample is red and reference is green) and a fifth sample vent
(white) coming from the sample pump. For protection and to avoid
contamination, the vents are closed using Luer-Lock plugs when the unit leaves
the factory.
WARNING The Luer-Lock plugs must be removed when the unit is operating. Save plugs
to use for long term storage.
Caution Consult either the BTU or the EZChrom User’s guide for instructions on how to
set column head pressure.
32
MODULE 2: Intro to Agilent P-Series Micro GC
Micro GC Hardware
Right panel
The right back panel corresponds to the compartment containing the internal
battery and the internal carrier gas cylinder.
Caution Make sure power is off before disconnecting the carrier gas jumper line.
Caution Your unit has been set for either hydrogen/helium or nitrogen/ argon as the
carrier gas. If argon or nitrogen is used, certain switch settings in the motherboard will
need to be changed to avoid damage to the detector filaments. See Chapter 6 of the
User’s Manual for instructions.
Auxiliary 12 VDC
The auxiliary 12 VDC allows you to run your P200/ P200H from any other 12
VDC power source such as a vehicle cigarette lighter plug.
Carrier outlet
The carrier outlet on the right back panel connects a stainless steel jumper tube to
the carrier inlet on the back left panel. When the front panel carrier gas knob is
turned on, carrier gas is released through this outlet at 80 psi.
Carrier inlet
The carrier inlet on the left back panel connects to a stainless steel jumper tube to
the carrier outlet on the back right panel. When the front panel carrier gas knob is
turned on, carrier gas flows into the inlet at 80 psi.
33
MODULE 2: Intro to Agilent P-Series Micro GC
Laboratory Exercises
Laboratory Exercises
Lab Exercises
18
Figure 18
34
MODULE 2: Intro to Agilent P-Series Micro GC
Lab 1: Hardware Familiarization and Set Up
35
MODULE 2: Intro to Agilent P-Series Micro GC
Lab 2: Filling the Internal Carrier Gas Cylinder
36
MODULE 2: Intro to Agilent P-Series Micro GC
Module 2 Review
Module 2 Review
1) List the primary components necessary to operate the Micro GC system.
a)
b)
c)
d)
e)
f)
g)
h)
i)
j)
_______________________ ________________________
37
MODULE 2: Intro to Agilent P-Series Micro GC
Module 2 Review
b)
|
c)
d)
6) Why is it important to turn on the carrier gas prior to powering on the GC?
b)
9) How long will the portable Micro GC operate using the internal carrier gas
cylinder?
38
Module 3: EzChrom Introduction
Introduction to EZChrom
Introduction to EZChrom
• Introduction to Windows
• Loading EZChrom Software
• Loading the GC Tools
• Verifying Installation, Setup, and Status
• Introduction to EZChrom Menu
Figure 19
The purpose of this module is to introduce the use of the EZChrom software.
40
Module 3: EzChrom Introduction
Windows Compatibility
Windows Compatibility
Figure 20
41
Module 3: EzChrom Introduction
Windows Compatibility
GC instrument requirements
Firmware version 17.4 or later.
42
Module 3: EzChrom Introduction
Software Installation
Software Installation
Figure 21
The next few pages will show the screens encountered when loading the software.
43
Module 3: EzChrom Introduction
Software Installation
Figure 22
44
Module 3: EzChrom Introduction
Software Installation
Figure 23
45
Module 3: EzChrom Introduction
Software Installation
Figure 24
Program groups are automatically created in Win 3.1, Win forWG, and WinNT.
But icons must be created in Win95.
46
Module 3: EzChrom Introduction
Software Installation
Figure 25
47
Module 3: EzChrom Introduction
Software Installation
Windows Explorer
Figure 26
This slide shows the path where the software modules are located.
48
Module 3: EzChrom Introduction
MTI GC Setup
MTI GC Setup
MTI GC Setup
Figure 27
A second floppy disk containing the GC Setup software must also be loaded the
same way as the Operation software.
With the GC system hooked up and powered on, start the GC Setup program.
This will establish the communication between the GC and PC.
49
Module 3: EzChrom Introduction
MTI GC Setup
MTI GC Setup
10
Figure 28
This slide shows the selection of the com port and the identification and
communication between the GC and PC.
50
Module 3: EzChrom Introduction
MTI GC Verification
MTI GC Verification
MTI GC Verification
11
Figure 29
51
Module 3: EzChrom Introduction
MTI GC Verification
52
Module 3: EzChrom Introduction
MTI GC Verification
MTI GC Verification
12
Figure 30
The first window you will see is the Introduction Window. In general, the GC
Verification windows will have three text items:
Explanation
A short explanation of the purpose or use of a particular window.
Request
If you are expected to do anything in particular, the Request item tells you
what the program wants you to do.
Question
If GC Verification needs any information from you, a Question item will pose
a question, which you answer by clicking a button in the window.
This window asks you to verify that anAgilentGC is connected to your computer.
Click [Yes] when you have connected your GC to the serial port of your
computer and are ready to proceed. If you click [Quit], GC Verification will
terminate. If you click [No], GC Verification Tool will tell you that it cannot
proceed without a GC.
53
Module 3: EzChrom Introduction
MTI GC Verification
MTI GC Verification
13
Figure 31
Typical sessions
In general, there are two types of sessions you will encounter when you are using
GC Verification. If your GC is OK, there is a simple sequence of windows, which
will be displayed as the condition of the GC is verified. If there is a problem, there
are a large number of possible window display sequences, depending on the type
of problem, which has occurred.
What happens if your GC is OK?
Connect your GC and click [Yes]. GC Verification checks each serial port on
your computer system for an M200. During this time a window is displayed. If
anAgilentGC is found, the window displays its serial number in the Status box
and, when you click [OK], GC Verification proceeds with checking the GC.
In the figure above, the “180505” represents the serial number of the GC. Your
Status box should display the serial number of your particular GC.
If your GC is OK, the final window is shown indicating successful verification.
See the User’s Manual for the EZChrom for further explanations if your GC has problems
(pages 139-158).
54
Module 3: EzChrom Introduction
Data File Path
14
Figure 32
In Explorer or File Manager, one can review the path for storage of data and
methods.
55
Module 3: EzChrom Introduction
Initiating EZChrom
Initiating EZChrom
Initiate EZChrom
15
Figure 33
56
Module 3: EzChrom Introduction
Initiating EZChrom
Initiate EZChrom
16
Figure 34
EZChrom 200/400 will then proceed to open and draw the windows, which
comprise the data system. Two display windows (four display windows, one each
for Channels A, B, C, D will appear with EZChrom 400) will appear in EZChrom
200 (i.e., one each for Channels A and B).
At this time, a default method and data file named after your instrument serial
number will be recalled and displayed. When you access the window represented
in the Figure above, the method and file names will be replaced by the serial
number of your particular instrument (i.e., the method EXAMPLE.MET will read
XXXXX.MET and the file name EXAMPLE.1 will read XXXXX.1, where
XXXXX represents the serial number for your instrument).
57
Module 3: EzChrom Introduction
Initiating EZChrom
Initiate EZChrom
In Win95 Must move menu bar and can size and position windows.
17
Figure 35
If using Windows95, one must manually resize and position the windows.
58
Module 3: EzChrom Introduction
Instrument Menu
Instrument Menu
Instrument Menu
18
Figure 36
The instrument status can be determined by selecting the Instrument menu item
and then selecting Status. Parameters cannot be changed in this screen.
59
Module 3: EzChrom Introduction
Method and Data Menus
19
Figure 37
The Method menu allows one to choose instrument and method parameters. We
will explore these options in detail in a later module.
60
Module 3: EzChrom Introduction
Math Menu
Math Menu
Math Menu
20
Figure 38
Purpose
Subtracts a user-selected chromatogram from the currently displayed
chromatogram. After analyzing, the reports and graphical display are updated.
Syntax
• Select Math or type <Alt> T.
• Select Catalog, or enter a path and file name to select a chromatogram. This
chromatogram will be subtracted from the currently displayed data set.
• Select [OK] to initiate the subtraction.
Comments
This feature could allow subtraction of large solvent peaks allowing easy
identification and integration of low level, overlapping peaks.
61
Module 3: EzChrom Introduction
Calib and Help
21
Figure 39
Purpose
Sets up EZChrom for a two channel calibration. It also provides the capability of
selecting a calibration level (from 1-8) for users who desire a multilevel
calibration.
Syntax
• Select Calib or type <Alt> C
• Enter a number between 1 and 8 to select a calibration level.
• Select [OK]. To stop a calibration prior to analysis, select [Cancel]
• Select Analyze. This will take the calculated peak areas and assign them to the
appropriate peak name and amount value.
62
Module 3: EzChrom Introduction
Calib and Help
Comments
Calibration does not begin until data is actually collected or stored data is
analyzed. Level 1 is all that is required for normal use. Higher levels are used to
construct linear or nonlinear (point-to-point) calibration plots.
Help
Displays a screen showing the current version of EZChrom.
We will explore the calibration options in a later module.
63
Module 3: EzChrom Introduction
Lab Exercises
Lab Exercises
Lab Exercises
22
Figure 40
64
Module 3: EzChrom Introduction
Lab Exercise 1: Starting Up the System
Note: Verify with your instructor as to where (the file path) the software should be
installed. Some PC’s may have dual operating systems; i.e. Windows for Workgroups
and Windows 95.
5) Repeat the procedure for loading the GC Tools Disk. Verify that the serial
number on the disk matches the serial number of the GC.
6) If working in Windows95, the procedure for creating desktop shortcuts can
be found in Lab Exercise 4.
65
Module 3: EzChrom Introduction
Lab Exercise 1: Starting Up the System
66
Module 3: EzChrom Introduction
Lab Exercise 2: Introduction to Windows NT
67
Module 3: EzChrom Introduction
Lab Exercise 2: Introduction to Windows NT
7) Select the Programs folder from Start. Point and click on the Windows Help
icon. The Figure below shows the Contents tab. If you are unfamiliar with
Windows NT, examine the contents of the Help topics.
8) When you have examined the various Help topics, close out the window by
clicking on the “X” icon in the top right corner of the Help window.
9) Select the Start icon, then the Programs folder. Select Windows NT
Explorer. The Figure below shows an example of the Explorer program.
68
Module 3: EzChrom Introduction
Lab Exercise 2: Introduction to Windows NT
10) Note the directory structure of the hard drive. In the upper right hand corner
of the screen, there are three icons. Point to and click the window shaped
icon in the center.
What happened?
11) Re-start Explorer. Verify that the tool bar appears as shown in the Figure
above. If it does not, select “View”, “Toolbar” and it should appear. Do not
“click”, but place the cursor on the bottom right corner of the “down arrow”
in box that appears in the small window under the menu items. Text should
appear which identifies the function of that arrow. See Figure 2.4 for an
example. Continue to place the cursor at the bottom right of each of the
icons and observe the definition of the function of each icon.
69
Module 3: EzChrom Introduction
Lab Exercise 2: Introduction to Windows NT
Change Directory
Changing directories is a way of navigating through your computer’s files.
14) Scroll down and Point and Double-click on the Programs sub-directory.
Then select the Windows NT sub-directory.
15) Note the files and folders. Click on the “-“ box next to the open Windows
NT folder in the directory tree.
What happened?
MTI Directory
16) Scroll down to the MTI subdirectory and Double-click on it to open the
folder. Note the default subdirectories.
70
Module 3: EzChrom Introduction
Lab Exercise 2: Introduction to Windows NT
Backing up Files
It is important to regularly backup your important files. We will use a floppy disk
for backup, however there are several other alternatives.
17) Insert a blank floppy into the A: drive. In Explorer, scroll up to the 3 1/2
Floppy (A:) icon in the directory tree and click on it. If the floppy is not
formatted, you will be prompted to format the disk. Follow the prompts for
a Full format.
18) Creating a Folder – To create a new folder on the drive, Point and Click on
the File menu. Select New, then Folder. Type “Data Files” as the name of
the folder. Open the A: drive folder in the left window so that the “Data
files” folder is visible in the directory tree.
19) Copying files – Click on the View menu item in Explorer and verify that the
option “Details” is selected. Copy any file. Point and click on one of the
files. Press and hold the Ctrl key on the keyboard while pointing and
clicking on a second file. If all went well, both files would appear
highlighted in reverse video. Scroll up the left window until the A: drive
icon is visible. With both files highlighted, hold down the left mouse button
while clicking on one of the highlighted files and drag them to the “Data
Files” folder on the A: drive. Release the mouse button, and this should
copy the files to the A: Floppy.
20) Verify that these files were copied by selecting the A: floppy icon and
viewing its contents.
71
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
Note: Do not use any screen savers other than the standard Windows NT options. Other
screen saver programs can utilize system resources and could crash
yourAgilentChemStation.
72
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
Using Shortcuts
5) To launch an application represented by an icon on the desktop, simply
double-click on it. Select the Recycle Bin icon and double-click on it. If
there are contents in the list, select File, and Empty Recycle Bin. When a
file is deleted it is not removed from the disk drive to free up space until the
Recycle Bin is emptied.
6) Right click on the Start button on the bottom left corner of the screen.
Notice that a short menu appears and one can start Explorer by selecting
Explore.
7) If you are not sure where a file is located you may search for it in Explorer
or you may right click on the Start button and select Find. Type
"Config.sys" in the Named: field. Select C: in the Look in: field. Click
Find Now or press Enter. The file location is indicated. To close the Find
window, click on the X icon located in the upper right hand corner.
73
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
Tip
15) You can also add a program to the top of the Start menu by dragging the
program's icon to the Start button. Locate “Notepad” in WinNT.
16) Click and drag it down to the Start button at the bottom left of the screen.
Close out the open windows and click on the Start button to see Notepad
listed in the menu.
Creating Shortcuts
To put a shortcut on the desktop
17) Now go back to Explorer. Open Windows NT, then Accessories, then and
Pinball. Locate the file name “Pinball” with the file type “Application”.
Click on the file name and drag it to the desktop. (If your desktop is not
visible because Explorer is the full size of the screen, select the multiple
window icon in the top right corner of the screen.). An icon shortcut should
be created.
18) Double-click on the icon to start the game.
19) Now go back to Explorer. Select “Tools”, “Find”. Search for the Start
Menu in the WinNT directory. Locate the folder for MTI.
20) Select one of the programs in the right window. Right click on it and select
Create Shortcut. A second short cut is created. Click on it and drag it to the
desktop (If your desktop is not visible because Explorer is the full size of the
screen, select the multiple window icon in the top right corner of the screen.)
Keyboard Shortcuts
Menu items can be selected using keyboard operations rather than the mouse.
21) With Explorer open, Hold down the Alt key and press the “f” key on the
keyboard. Note that the File menu items appear. Menu items can be
selected and activated by typing the underlined character present in the menu
item name. If the menu is activated unintentionally with Alt, pressing Alt a
second time can turn it off again just.
22) Now select Alt “e” and note that the Edit Menu items appear. Using the
control key (Ctl) activates additional keyboard shortcuts. These are shown
74
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
next to the menu items: Ctl+X will activate the Cut feature, Ctl+C will
activate the Copy feature, etc.
75
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
76
Module 3: EzChrom Introduction
Lab Exercise 3: Working with Windows NT
Never use Scandisk.exe or Defrag.exe on a WinNT 4.0 NTFS partition. If this is done,
one will have to re-format the hard drive and re-load WinNT.
Virus Scanning
McAfee virus scanner is provided in ChemStation bundles. Many different
utilities are available for virus scanning. You should run a disk scanning utility if
you download programs from the Internet or exchange programs through e-mail
or flexible disks.
The programs must be updated regularly to protect your system. McAfee provides
an update monthly on its web site.
77
Module 3: EzChrom Introduction
Optional Lab Exercise 4: Intro to Windows 95
78
Module 3: EzChrom Introduction
Optional Lab Exercise 4: Intro to Windows 95
5) Select the Programs folder from Start. Point and click on the Windows Help
icon.
The Figure below shows the Contents tab. If you are unfamiliar with Windows
95, examine the contents of the Help topics.
6) When you have examined the various Help topics, close out the window by
clicking on the “X” icon in the top right corner of the Help window.
7) Select the Start icon, then the Programs folder. Select Windows Explorer.
The Figure below shows an example of the Explorer program.
79
Module 3: EzChrom Introduction
Optional Lab Exercise 4: Intro to Windows 95
8) In the upper right hand corner of the screen, there are three icons. Point to
and click the window shaped icon in the center. What happened? What's
the function of the other two icons?
9) Re-start Explorer. Verify that the tool bar appears as shown in the next
Figure. If it does not, select “View”, “Toolbar” and it should appear. Do not
“click”, but place the cursor on the “down arrow” in box that appears in the
small window under the menu items. Text should appear which identifies the
function of that arrow. Continue to place the cursor over each of the icons
and observe the definition of the function of each icon.
80
Module 3: EzChrom Introduction
Optional Lab Exercise 4: Intro to Windows 95
Change Directory
Changing directories is a way of navigating through your computer’s files.
11) Scroll down and Point and Double-click on the Windows sub-directory.
Note the files and folders. Click on the “-“ box next to the open Windows
folder in the directory tree.
What happened?
12) MTI - Scroll to the MTI subdirectory. Double-click on it to open the folder.
Note the default subdirectories under MTI.
Backing up Files
It is important to regularly backup your important files. We will use a floppy disk
for backup, however there are several other alternatives.
13) Insert a blank floppy into the A: drive. In Explorer, scroll up to the 3 1/2
Floppy (A:) icon in the directory tree and click on it. If the floppy is not
formatted, you will be prompted to format the disk. Follow the prompts for
a Full format.
14) Creating a Folder – To create a new folder on the drive, Point and Click on
the File menu. Select New, then Folder. Type “System Files” as the name
of the folder. Open the A: drive folder in the left window so that the “System
files” folder is visible in the directory tree.
15) Copying files – Click on the View menu item in Explorer and verify that the
option “Details” is selected. Point and double click on the C: drive icon.
Scroll down the directory in the right window and locate the file
AUTOEXEC.bat (MS DOS Batch File). Point and click on this file.
Scroll down and locate the file CONFIG.SYS. Press and hold the Ctrl key
on the keyboard while pointing and clicking on this file. If all went well,
both files would appear highlighted in reverse video. Scroll up the left
window until the A: drive icon is visible. With both files highlighted, hold
down the left mouse button while clicking on one of the highlighted files and
drag them to the “System Files” folder on the A: drive. Release the mouse
button, and this should copy the files to the A: Floppy.
16) Select the Windows folder in the C: directory tree, locate the Win.ini file in
the right window list. Copy it onto the A: floppy in the “System Files”
folder.
81
Module 3: EzChrom Introduction
Optional Lab Exercise 4: Intro to Windows 95
17) The three files that were copied are important to have backed. In the event of
an operational problem with your PC, these files may be useful in restoring
its operation. Verify that these files were copied by selecting the A: floppy
icon and viewing its contents.
Printing a File
20) In the File menu, select Print to get a hardcopy of the file.
82
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
Note: Do not use any screen savers other than the standard Windows 95 options. Other
screen saver programs can utilize system resources and could crash yourChemStation.
83
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
Using Shortcuts
To launch an application represented by an icon on the desktop, simply double-
click on it.
6) Select the Recycle Bin icon and double-click on it. If there are contents in
the list, select File, and Empty Recycle Bin. When a file is deleted it is not
removed from the disk drive to free up space until the Recycle Bin is
emptied.
7) Right click on the Start button on the bottom left corner of the screen.
Notice that a short menu appears and one can start Explorer by selecting
Explore.
8) If you are not sure where a file is located you may search for it in Explorer
or you may right click on the Start button and select Find. Type
"Config.sys" in the Named: field. Select C: in the Look in: field. Click
Find Now or press Enter. The file location is indicated. To close the Find
window, click on the X icon located in the upper right hand corner.
84
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
Creating Shortcuts
9) To create a shortcut to an application Launch Explorer. Go to the Windows
folder and find the Start Menu subdirectory. Open the Programs folder and
locate the Explorer program. Click on the Explorer item and drag it to the
Start Menu folder in the directory tree.
10) Now left click on the Start button and note where the Explorer short cut
appears. Double-click on the shortcut and note that a second Explorer is
launched. Close out that program.
11) Now go back to Explorer, Programs, and Start Menu. Locate the folder
for MTI EZChrom. Select EZChrom 200 in the right window. Right click
on EZChrom200 and select Create Shortcut. A second short cut is created.
Click on it and drag it to the desktop (If your desktop is not visible because
Explorer is the full size of the screen, select the multiple window icon in the
top right corner of the screen.)
Keyboard Shortcuts
12) Menu items can be selected using keyboard operations rather than the
mouse. With Explorer open, Hold down the Alt key and press the “f” key
on the keyboard. Note that the File menu items appear. Menu items can be
selected and activated by typing the underlined character present in the menu
item name. If the menu is activated unintentionally with Alt, pressing Alt a
second time can turn it off again just.
13) Now select Alt “e” and note that the Edit Menu items appear. Using the
control key (Ctl) activates additional keyboard shortcuts. These are shown
next to the menu items: Ctl+X will activate the Cut feature; Ctl+C will
activate the Copy feature, etc.
85
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
17) An alternate way to activate a window is Alt-Tab, which toggles among the
windows that are activated. Try Alt-Tab and notice the difference between
this and Alt-Esc.
18) Close all the opened applications.
86
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
DEFRAG
Over time, as programs read from and write to a hard disk, information stored on
the disk can become fragmented – that is, files are stored in noncontiguous
sectors. Fragmentation does not affect the validity of the information – the files
are still complete when they are opened. But it takes much longer for the
computer to read and write fragmented files than it does for unfragmented files.
Never run a defragmentation while an application is open. Never interrupt a
defragmentation in progress.
21) Click the Start button, click Run, then type defrag. If you have difficulty
finding the defrag program, select right click on Start, then Find, and type in
defrag. When the file is found simply double click on the defrag.exe file to
start the program.
22) In the Select Drive dialog box, specify C: and then click OK. The Disk
Defragmenter displays a dialog box telling you whether defragmentation is
recommended for this disk or not. If this disk has low fragmentation, the
Disk Defragmenter will not recommend defragmentation.
23) Showing details while the Disk Defragmenter is running causes it to take
longer than it does when showing only summary information or running it
minimized. For quickest performance, minimize the Disk Fragmenter
window while the utility is running.
To see defragmentation information for a particular drive:
24) In My Computer, right click the drive’s icon (C:), and then click Properties
25) Click the Tools tab. The Tools properties dialog box shows the number of
days since the last complete defragmentation process ran on the drive. You
can also run Disk Defragmenter from this dialog box.
87
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
ScanDisk
SCANDISK can be used to correct errors in the file system by searching for lost
file fragments and allow re-allocation of fragments to the file system. ScanDisk is
a full-featured disk analysis and repair program. ScanDisk runs automatically
when you start Windows 95 Setup.
ScanDisk checks and fixes problems in the following areas on hard disk drives,
floppy disk drives, RAM drives, and memory cards.
• File Allocation Table (FAT)
• Long filenames
• File system structure (lost clusters, cross-linked files)
• Directory tree structure
• Physical surface of the drive (bad sectors)
To run ScanDisk click Start button, then Run, and type scandisk.
88
Module 3: EzChrom Introduction
Lab Exercise 5: Working with Windows 95
Many different utilities are available for virus scanning. You should run a disk
scanning utility if you download programs from the Internet or exchange
programs through e-mail or flexible disks.
89
Module 3: EzChrom Introduction
Module 3 Review
Module 3 Review
1) When is it safe to turn on the GC?
a)
b)
c)
5) How does one determine the compatibility of the operating system and the
revision of software?
90
Module 3: EzChrom Introduction
Module 3 Review
7) Why is it important to know the path for data and method storage on the
hard drive?
91
Module 3: EzChrom Introduction
Module 3 Review
92
Module 4: Data Acquisition and Analysis
Figure 41
94
Module 4: Data Acquisition and Analysis
Data Acquisition and Analysis
Setup Acquisition
Parameters
Integration
Optimization
Calibration Setup
Reporting Options
Figure 42
95
Module 4: Data Acquisition and Analysis
Data Acquisition and Analysis
• Open up a method
• Edit method to fit project needs
• Download the method to the GC
• Setup the run file pathways
• Run blanks to check baseline
• Analyze the Sample
Figure 43
96
Module 4: Data Acquisition and Analysis
Data Acquisition and Analysis
Figure 44
97
Module 4: Data Acquisition and Analysis
Data Acquisition and Analysis
Figure 45
There are several alternatives for collecting and injecting the gas sample into the
GC.
98
Module 4: Data Acquisition and Analysis
Data Acquisition and Analysis
Figure 46
When starting a run, one should hear the sample pump, see the run time displayed
on the PC and see the chromatogram being drawn.
99
Module 4: Data Acquisition and Analysis
Creating a Method
Creating a Method
The following slides will review the major steps taken when
creating a method.
Figure 47
In the following slides, we will identify the specific steps in creating a method
using the EZChrom software.
100
Module 4: Data Acquisition and Analysis
Creating a Method
Figure 48
When starting the software for the first time, the Example.Met is loaded. One
should modify the parameters of this method and save the method with a new
name.
Prior to making any modifications to the method, select the Method menu, Save
Method As, and give the method a new name.
101
Module 4: Data Acquisition and Analysis
Creating a Method
Figure 49
102
Module 4: Data Acquisition and Analysis
Creating a Method
Note: There are some operational considerations when using Backflush modules.
10
Figure 50
103
Module 4: Data Acquisition and Analysis
Creating a Method
cool the columns to about 40°C. Then, turn the column head pressure knob
counter clockwise until the carrier gas is set to deliver zero psig. It will take about
10 minutes for the pressure in the columns to decline to below 5 psig.
Caution To protect the detectors, always turn the detectors off before decreasing
column head pressure to zero psig.
The manifold delivers a column head pressure of about zero when, looking down
the edge of the BFM GC back panel, the black knob is 3/8 inch from the manifold
stem. The manifold delivers a column head pressure of around 30 psig when the
black knob becomes difficult to adjust.
104
Module 4: Data Acquisition and Analysis
Creating a Method
11
Figure 51
It is necessary to Send the Current Method from the PC to the GC in order for
the parameters to be recognized by the GC.
105
Module 4: Data Acquisition and Analysis
Creating a Method
Check the Instrument Status to verify that the new setpoints are being implemented.
12
Figure 52
Select Instrument Status in order to confirm the implementation of the new set
points. One can watch any temperature changes as the zones heat up or cool
down.
106
Module 4: Data Acquisition and Analysis
Creating a Method
For definitions of the parameters in the table see pages 35 -37 of the User’s Manual.
13
Figure 53
Purpose
Contains compound names and elution times, which allows EZChrom 200/400 to
properly identify and quantify integrated peaks. It also provides the option of
grouping multiple peaks, which elute in a user-specified window. User-defined
response factors can be defined and referenced here.
Syntax
Select Peak Table or type E.
Comments
In most cases, the only information that needs to be entered in this table is peak
names, concentration units and Level 1 Cal Amounts. Retention times can be
entered graphically. The Peak Table contains the following information:
Peak Name
In this portion of the table, the names of the compounds which are to be analyzed
are entered. The peak names are entered for each component by typing them in.
When an integrated peak falls in a defined retention time window, it is assigned
the name linked to that window.
Retention Time
This is the amount of time it takes for the apex of a named peak to elute from the
column. The retention time can be entered graphically or with the keyboard. (In
107
Module 4: Data Acquisition and Analysis
Creating a Method
the Calibration Setup table there is an option to update the retention time at the
end of a run. If this is selected and the retention time changes, then the retention
time and the retention time window will be updated).
Retention Time Window
If a peak apex lies within a defined region of time, even though the retention time
of the peak may not match the retention time listed, it will be identified as the
compound defined by the preset retention time window.
Concentration Unit (CU)
This is a text entry, which identifies the concentration units of the various
compounds. Any three character text entries are allowed. The unit is not a
conversion factor and has no mathematical significance. Each component may
have a different concentration unit. The concentration units are only used for
external standard reports.
Level 1 Cal Amount
This is the amount of a compound that the level one external calibration standard
contains. When a Level 1 calibration is run, the entered amounts are assigned to
the areas of any integrated peaks, which elute in the appropriate retention time
windows. Obviously, accurate calibration standards must be used and retention
time windows must be properly set for EZChrom 200/400 to accurately quantitate
real samples. EZChrom 200/400 can use up to eight levels of calibration but for
routine analysis one level is usually sufficient. Each level corresponds to a gas
standard concentration and a point on a calibration plot.
Relative Response Factor (RRF)
When an external standard is unavailable, an estimated response factor may be
specified. Enter a number, which will be multiplied by a calibrated response
factor.
Peak
This entry is used to reference the RRF discussed above to a particular peak in the
Peak Table. Choose a compound, which has a similar thermal conductivity to the
uncalibrated peak.
108
Module 4: Data Acquisition and Analysis
Creating a Method
MolSieve or OV1
column
PoraPlot U
Column
14
Figure 54
The peak table should be edited with information relevant to the analysis.
109
Module 4: Data Acquisition and Analysis
Creating a Method
15
Figure 55
110
Module 4: Data Acquisition and Analysis
Collecting the Data
16
Figure 56
111
Module 4: Data Acquisition and Analysis
Collecting the Data
Typical Chromatograms
17
Figure 57
When the run is completed, the data is first stored on the hard drive in the default
directory c:\HP\ezchrom\200\chrom under the run ID that you provided. The data
set is then automatically analyzed using the default timed events provided with
the NEW method. At this point take the opportunity to carefully examine the
analyzed chromatograms using the zoom and unzoom techniques. During your
examination, keep in mind the following questions:
Have all of the peaks been detected?
Are the peaks correctly integrated?
In the next few slides we will review data analysis techniques used in order to be
able to answer “Yes” to the above questions.
112
Module 4: Data Acquisition and Analysis
Collecting the Data
18
Figure 58
You may choose to save the data with another name or in a different file path.
113
Module 4: Data Acquisition and Analysis
Data Analysis
Data Analysis
Setup Acquisition
Parameters
Integration
Optimization
Calibration Setup
Reporting Options
19
Figure 59
We will now investigate the process for optimizing the graphics and integration.
114
Module 4: Data Acquisition and Analysis
Graphic Manipulation
Graphic Manipulation
Zoom Box
Press
Pressand
andHold theleft
Holdthe leftmouse
mousebutton
button
andat
and atthe
thesame
sametime
timedrag
dragthe
thecursor
cursor
totothe
theright
rightand
anddown
downtotoform
formaa
rectangular
rectangularbox
boxaround
aroundthe
theportion
portionyou
you
wish
wishtotozoom.
zoom.
20
Figure 60
Press and hold the left mouse button and at the same time drag the
cursor to the right and down, just below the baseline. This should draw a
rectangular box around a portion of the chromatogram.
Release the mouse button. EZChrom 200/400 will now draw the area inside the
zoom box so that it fills the lower screen.
115
Module 4: Data Acquisition and Analysis
Graphic Manipulation
Vertical Lines
Use
Usethese
thesecursors
cursorsasasmobile
mobile
reference
referencepoints
pointstotoobserve
observepeak
peak
amplitude
amplitudeandandretention
retentiontimes.
times.
Click
Clickthe
theright
rightmouse
mousebutton
buttonand
and
scan
scanthe
thecursor
cursoracross
acrossthe
the
chromatogram
chromatogramand andsee
seethe
thetime
time
and
andamplitude
amplitudedisplayed.
displayed.
Click
Clickthe
theleft
leftmouse
mousebutton
buttontoto
freeze
freezethe
thefirst
firstvertical
verticalcursor.
cursor.
21
Figure 61
Use these cursors as mobile reference points to observe peak amplitude and
retention times. They are also extremely useful in simplifying the selection and
definition of integration parameters and in the peak identification routine.
• Place the single-headed arrow cursor in either the channel A or B lower
window.
• Click the right mouse button. A single vertical line should appear. When
the mouse is moved to the left or right, the vertical cursor should move
accordingly. When using a laptop computer, the mouse must be moved slowly
for the vertical line to remain visible.
• Scan the cursor across the chromatogram and observe the string of text in the
lower display window. You should see the time and amplitude vary. Dragging
the mouse to the right should result in a larger time value. (This is because
data is collected from time 0, left to right). The value of the amplitude is the
actual instrument output in microvolts at the displayed time.
• Click the left mouse button. This will freeze the first vertical cursor.
• Drag the mouse to the right. The frozen cursor should remain, while a second,
mobile cursor will appear. The values for time and amplitude now displayed
are in reference to the first vertical cursor. That is to
say, the first vertical cursor now has been assigned a time and amplitude value
of 0.
116
Module 4: Data Acquisition and Analysis
Graphic Manipulation
Drag
Dragthethemouse
mousetotothetheright.
right.AAsecond
second
mobile
mobilecursor
cursorwill
willappear.
appear.Values
Valuesfor
fortime
time
and
andamplitude
amplitudenownowsetsetininreference
referencetotothe
the
first
firstvertical
verticalcursor
cursor(first
(firstvertical
verticalcursor
cursor
values
valueshave
havebeen
beenset
settotozero).
zero).
Click
Clickon onthe
theright
rightcursor
cursortotofreeze
freezethe
the
second
secondvertical
verticalcursor.
cursor.
To
Toremove
removethethevertical
verticalcursor,
cursor,select
select
Display,
Display,Cursor.
Cursor.
22
Figure 62
• Click the right mouse button. This freezes the second vertical cursor. The
region between the two vertical cursors is called the selected region. Being
able to select specific regions in a chromatogram is essential in graphically
setting up peak identification tables and integration parameters.
• Click the right mouse button again. This unfreezes the second vertical cursor.
• Click the left mouse button. This will cause the two vertical cursors to merge
and be mobile.
• Click the right mouse button. This will freeze the single vertical cursor
remaining, while also reactivating the single headed cursor. The single
headed cursor may now be used in any of the ways previously described.
• To remove the vertical cursor from the screen, click on Display in the
appropriate channel A or B window, then click on CURSOR.
Practice selecting various portions of the chromatogram. At first you may find it
difficult to obtain the desired cursor in an appropriate frozen or unfrozen state.
Don’t get frustrated; after a short time the manipulation of the mouse clicks will
become habit.
117
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
23
Figure 63
After carefully examining the analyzed chromatograms using the zoom and
unzoom techniques, answer the following questions.
118
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
If the answer to either of the questions was no, then the Timed Events tables
must be optimized in order for an accurate analysis to be made. It may be that
the default timed event parameters were sufficient to find and integrate the peaks
in such a simple application. In general, though, this will not be the case.
119
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
Timed Events
INTG Off – Integration Off Min Area – Minimum Area
Turns Integration Off Defines the smallest integrated area that
for a specified period of time. should be identified as a peak. Allows small
Prevents integration over selected peaks or baseline defects to be ignored
regions during peak search.
24
Figure 64
120
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
Refer to pages 41 - 45 of the User’s Manual for procedures for using these events.
25
Figure 65
121
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
26
Figure 66
122
Module 4: Data Acquisition and Analysis
Optimizing Timed Events
27
Figure 67
• Select the Analyze command from the lower menu bar. EZChrom will
analyze the chromatogram according to the parameters just set. Zoom in
closely on each peak to assure that the entire peak is within its predetermined
hash marks. If the peak does not have an upward hash mark at its beginning
and a downward hash mark at its end, it has not been integrated.
• Open the External Standard reports for channels A and B by clicking on
Reports, then on External Standard (See Figure 33). They should display the
peak names in the order of elution time and have large amount values for all
of the peaks listed in the reports. This is because the method is not yet
calibrated.
• If any of the peaks in the report have an amount of zero, and has an
integration baseline drawn underneath it, the retention time has been
incorrectly set.
• The Peak Table entries are verified when all the names you have added to the
Peak Table are present in the External Standard report.
123
Module 4: Data Acquisition and Analysis
Lab Exercises
Lab Exercises
Lab Exercises
28
Figure 68
124
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
125
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
126
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
evidenced by the tops of peaks rising above the upper limits of the channel
window display. (See page 117 through page 120 respectively for
explanations of how to unzoom and how to adjust attenuation.) After
adjusting the instrument parameters, run the GC again. If the analysis is still
unsatisfactory, repeat this step. Be careful not to be fooled by the display.
Always set the display Attenuation to 2048, and Fully Unzoom the
lower display when checking for detector saturation.
6) Once the instrument settings have been optimized, the integration parameters
tables should be adjusted. During the course of completing Step 5, you may
have noticed that some of the peaks in the chromatogram were not
integrated, or were integrated incorrectly. To correct this deficiency, PKWD
events must be added to the Timed Events tables.
Observe the last analyzed dataset on the screen. Note the first peak, which is not
properly integrated (the calculated baseline does not extend from the true
beginning of the peak to its true end). Using a Vertical Cursor, determine
the time at which the missed peak begins to elute. Now, add a PKWD event, with
a value of 0.2, and a time slightly less (approximately 0.5 seconds) than the
beginning of the missed peak, to the appropriate Timed Events table.
After the PKWD event has been added, Reanalyze the data. The peak, which
was previously not integrated, should now be properly integrated. If it is not,
double the PKWD value to 0.4 and reanalyze. Eventually, the peak will be
successfully integrated. Repeat the above steps for both channels and all
subsequent peaks, which are still incorrectly integrated, remembering to double
the value for each successive PKWD.
At this time, you may also wish to extend the Intg Off event to include
solvent or composite peaks, which are not of interest.
7) Set the Retention Times and Retention Time Windows
graphically (as described in Chapter 4 of the User’s Manual) for all peaks of
interest in both channels. If you now reanalyze the data, the External
Standard report should display the names and area counts for all integrated
peaks in the chromatogram.
8) Select the Calib command from the lower menu bar and run the GC. Data
will be acquired, and a calibration will be done. Inspection of the External
Standard reports should show all compounds, which were in the calibration
gas, and at the correct concentrations. The Method is now complete. Save
the Method to the hard drive.
127
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
Backflush
Pressure Reducing
Valve
Sample Flow Restriction
Flow Detector
During
Injection
Analytical
Carrier Gas Pre-Column Column
supply
ForeFlush
Valve Inject Valve
128
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
4) Set the run time to 160 seconds. Select a shorter run time if you are certain
that all of the sample components for quantification will be detected within
the shorter run time. A long run time facilitates the detection of carry-over
peaks and the passage of undesirable sample components from the
precolumn to the analytical column.
5) Analyze a sample. Determine if all the sample components for quantification
appear in the chromatogram. If all the sample components for quantification
are not detected, go to step 7.
6) Change the Method and BFM operating parameters to optimize the analysis.
Adjust the column temperature, column head pressure, or "Backflush At"
time to optimize the analysis. Potential improvements to the analysis
include: detecting all the sample components for quantification; reducing the
run time needed to get peaks for all the sample components for
quantification; improving peak separation; and reducing the "Backflush At"
time. In general, shorter "Backflush At" times and lower column head
pressures yield better chromatography. Increasing the column temperature
generally improves peak shape for the late-eluting peaks. On the downside,
increasing the column temperature can undesirably reduce the resolution of
the early-eluting peaks in the chromatogram.
a) If peaks for all the sample components for quantification appear in
the chromatogram, monitor the peak area of the last peak for
quantification. The "last peak for quantification" means the peak
with the longest "retention time" among the peaks used for
calculating sample results. If the size or shape of the last peak for
quantification causes inaccurate peak integration, then monitor the
peak area of another peak near the last peak. Reduce the
"Backflush At" time and rerun the sample. Compare the previous
run peak area for the monitored peak to the peak area in the rerun
of the sample. If the peak areas agree within 2%, then the shorter
"Backflush At" time is preferred. On the other hand, if the previous
run peak area of the monitored peak quantification is more than
2% greater than the peak area from the rerun then increase the
"Backflush At" time and rerun the sample. Make small adjustments
(tenths of a second) when you think you are close to the optimum
time, and larger adjustments (1 to 5 seconds) when you are far
from optimum time. Find the minimum "Backflush At" time that
delivers all the sample components for quantification to the
analytical column. Determine the value of this minimum
"Backflush At" time, and then add some time increment (e.g., 5%
to 10%) as a buffer against system variability. For example, if 3
seconds is the minimum time, select 3.3 seconds for the
"Backflush At" time. Remember that the "Backflush At" time has a
resolution of 0.1 second. Verify the repeatability of the GC with
the selected "Backflush At" time. Hewlett-Packard suggests
analyzing the same sample eight successive times. The peak area
129
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
130
Module 4: Data Acquisition and Analysis
Lab 1: Air Analysis using the Molsieve/Poraplot U Backflush GC
131
Module 4: Data Acquisition and Analysis
Lab 2: Butane Analysis using the MolSieve/Poraplot U GC
132
Module 4: Data Acquisition and Analysis
Lab 3: Natural Gas Analysis using the OV1/Poraplot U GC
133
Module 4: Data Acquisition and Analysis
Module 4 Review
Module 4 Review
1. When do changes to the method become active on the GC?
3. Why is it important to know how to use the of the vertical line cursors?
134
Module 5: Injectors
GC Injectors
GC Injectors
Figure 69
The purpose of this module is to review the key considerations of the injectors.
136
Module 5: Injectors
GC Injectors
Injection Systems
Purpose:
To allow the insertion of a sample into the gas chromatograph in a
repeatable, reproducible manner. The sample should be representative of
the bulk, and unless specifically desired should be inserted without
chemical change.
Injection Types:
• Unheated - Variable (Timed)
• Unheated – Fixed
• Unheated Backflush
• Heated Variable (Timed)
• Heated Fixed
Figure 70
137
Module 5: Injectors
GC Injectors
Injector Hardware
Figure 71
Two common types of instruments are the dual and single channel designs.
138
Module 5: Injectors
GC Hardware
GC Hardware
Figure 72
139
Module 5: Injectors
GC Hardware
Major Components
GC Channel or Module
Figure 73
This schematic shows the dual channel GC with the injector pneumatics module
140
Module 5: Injectors
GC Hardware
GC Module
6“
Detector
Vents SSD
Heater
Control 4“
Carrier
Input &
Valve
Activation
Injector
Column Heater Column Spiral Analytical
and Reference Columns
Figure 74
This diagram shows the relative size and arrangement of the GC components.
Inside each module is a complete gas chromatograph consisting of a micro
injector, columns (analytical and reference), column oven, universal micro solid
state detector and heaters. The module can be easily exchanged. However,
individual components can only be exchanged at the factory.
Observe that each module has an outside label. This clearly identifies the column
type, length, heater scale and the offset. Scale and offset will be discussed later,
since they need to be considered or are important when exchanging modules.
141
Module 5: Injectors
GC Hardware
Gas Samples
Figure 75
The P200 carrier gas moves the sample through the column. The carrier gas
should be dry and of high purity. The thermal conductivity of your carrier gas
should differ from components you want to analyze. Because of its high thermal
conductivity, the best choice for most applications is helium. Choose a gas with a
purity of at least 99.995%. When analyzing low ppm of oxygen and nitrogen, the
helium should be at least 99.999% pure. If you are analyzing helium or hydrogen
in a sample, nitrogen or argon are good carrier gas selections.
Caution Your P200/ P200H is configured at the factory to use helium/ hydrogen or
nitrogen/ argon as the carrier gas. Use of another carrier gas generates filament
temperatures that are too high and that result in damage to the detector if run at the
standard voltage levels. The GC controller board will adjust the filament voltage when
using argon or nitrogen so that filament temperatures are comparable to those when
using helium. These adjustments are discussed in Chapter 6–Theory of Operation.
Caution Hewlett-Packard does not recommend using hydrogen in the internal tank of
the P200.
The use of a carrier gas other than helium or hydrogen pose problems that result
in a loss in sensitivity. This is caused by the following:
142
Module 5: Injectors
GC Hardware
143
Module 5: Injectors
Injection Process
Injection Process
Pressure
Carrier In Regulator Sample Loop
Switching Valve
Pressure
Transducer
Sample In Vacuum
Pump Sample Vent
Out
Column Heater
Reference Column
Figure 76
144
Module 5: Injectors
Injection Process
145
Module 5: Injectors
Injection Process
Pressure
Carrier In Regulator Sample Loop
Switching Valve
Pressure
Transducer
Sample In Vacuum
Pump Sample Vent
Out
Column Heater
Reference Column
Figure 77
146
Module 5: Injectors
Injection Process
Analytical
Carrier Gas Pre-Column Column
supply
ForeFlush
Valve Inject Valve
10
Figure 78
This schematic shows the flow of gases for the backflush process.
A GC module with a pre-column backflush-to-vent configuration has two
columns—a short pre-column and a longer analytical column. The backflush
valve lies between the pre-column and the analytical column. The stationary
phase and length of the pre-column is chosen so that, at the operating temperature
of the columns, undesirable sample components that interfere with the analyses
are retained on the pre-column stationary phase. The sample components for
quantification have minimal affinity for the stationary phase of the pre-column,
and pass quickly through the pre-column. After the last sample component for
quantification exits the pre-column and enters the analytical column, the
backflush-to-vent valve located between the pre-column and the analytical
column opens. The carrier gas flow through the pre-column is reversed and back
flushes the undesirable sample components off the pre-column. The sample
components for quantification continue to flow through the analytical column for
separation and detection. At the end of the analytical run, the backflush GC
module is fully purged of all sample components and is ready for the next
analysis.
147
Module 5: Injectors
Injection Process
WARNING Carrier gas pressure surges greater than 1.4 psig/sec may damage the
analytical columns.
It is critical in BFMs to set the column head pressure knobs to deliver zero
psig before startup
If full pressure is applied to the BFM, the pre-column phase dislodges from the
column wall and flakes away causing a flow restriction. Flow restrictions are
known to change the retention times, peak elution, Backflush at Times and even
destroy the detector. Therefore, it is necessary to always back-out the column
head pressure knobs for BFMs before turning on the carrier gas.
To ensure the integrity of the analytical columns in BFMs, do not subject the
columns to pressure surges. Before connecting a BFM to carrier gas, set the
column head pressure knob to zero psig by turning the knob counter clockwise.
After the carrier gas is connected, slowly turn the knob clockwise to increase the
148
Module 5: Injectors
Injection Process
Shutdown Procedure
As standard practice with BFMs, the column head pressure should be set to zero
psig when finishing up for the day. To protect the GC, turn the detectors off and
cool the columns to about 40°C. Then, turn the column head pressure knob
counter clockwise until the carrier gas is set to deliver zero psig. It will take about
10 minutes for the pressure in the columns to decline to below 5 psig.
Caution To protect the detectors, always turn the detectors off before decreasing
column head pressure to zero psig.
The manifold delivers a column head pressure of about zero when, looking down
the edge of the BFM GC back panel, the black knob is 3/8 inch from the manifold
stem. The manifold delivers a column head pressure of around 30 psig when the
black knob becomes difficult to adjust.
149
Module 5: Injectors
Injection Process
Sam ple C C O O C
Com pression O O O C C
11
Figure 79
This chart shows the valve positions during the injection process.
150
Module 5: Injectors
Injection Process
Injsolenoid position
Sam ple in
M anifold (includes
pressure regulator) Carriergas out
12
Figure 80
This diagram shows how the injection manifold is connected to the injector.
151
Module 5: Injectors
Injection Process
Injector Function
13
Figure 81
152
Module 5: Injectors
Injector Types
Injector Types
Injector Types
14
Figure 82
There are several choices of injectors that can be used. The application
determines which injector is best.
153
Module 5: Injectors
Injector Types
• Fixed (1uL)
– better repeatability
– less flexibility with unknown samples
– backflush available: 1 or 0.4uL (specifically for Alumina PLOT)
• Variable/Timed
– offers greater flexibility
– 30uL volume
15
Figure 83
154
Module 5: Injectors
Injector Types
Physical Dimensions
Length: 1.166 +/- 0.002 inches
Width: 0.920 +/- 0.002 inches
Thickness: 0.118 +/- 0.002 inches
Silicon Micro-machined
Injector
16
Figure 84
This picture shows the actual hardware of the injector and the micro channels
where the gases flow.
155
Module 5: Injectors
Injector Types
Micro-Injector
17
Figure 85
156
Module 5: Injectors
Injector Types
Injector Restrictions
• Pressure Restrictions
– Maximum Column Head Pressure: 55 psi
– Maximum helium inlet pressure: 90 psi
– Maximum sample inlet pressure: 55 psi
• Temperature Restrictions
– Maximum injector temperature: 110°C
• Particulate Restrictions
– Maximum particulate diameter: Dp50 = 25 mm
18
Figure 86
There are operational restrictions in order to have the best performance and to
avoid problems.
157
Module 5: Injectors
Injector Types
70
Concentration (ppm) 65
60
Curve Fit Equation:
55 y = 10.494Ln(x) + 24.117
R = 0.99
50
45
10 20 30 40 50 60 70
19
Figure 87
This chart shows the relationship between injection time and concentration
detected. Doubling the injection time doubles the response only for the shorter
injection times. Therefore, there are minimal gains when using longer injection
times.
158
Module 5: Injectors
Theory of Injector Operation
Theory of Operation
20
Figure 88
159
Module 5: Injectors
Theory of Injector Operation
VALVE
OPEN
Helium Under
Positive Pressure
VALVE
CLOSED
21
Figure 89
160
Module 5: Injectors
Theory of Injector Operation
T Channel Valves
T Channel valves control gas flow at intersecting channels
VALVE VALVE
OPEN CLOSED
Helium Under
Positive
Pressure
22
Figure 90
The next few diagrams show the function of the “T” Channel Valve.
161
Module 5: Injectors
Theory of Injector Operation
T Channel Valves
When one end of the T channel is closed, the flow becomes unidirectional
VALVE Closed
OPEN End
Helium Under
Positive Pressure
23
Figure 91
162
Module 5: Injectors
Theory of Injector Operation
Inject Position
Valve
Switching
Pre-column Analytical
Column
Carrier Gas
24
Figure 92
163
Module 5: Injectors
Theory of Injector Operation
Valve Open
Valve
Switching
To Sample Sample In
Pump
To Analytical Column
25
Figure 93
164
Module 5: Injectors
Theory of Injector Operation
Valve Closed
Valve
Switching
To Sample Sample In
Pump
To Analytical Column
26
Figure 94
165
Module 5: Injectors
Theory of Injector Operation
Valve
Switching
Pre-column Analytical
Column
Carrier Gas
27
Figure 95
166
Module 5: Injectors
Theory of Injector Operation
Sample Channel
Sample Inlet
Sample Valve
Sample Channel
Sample Inlet Channel
Column Inlet Channel
28
Figure 96
167
Module 5: Injectors
Theory of Injector Operation
Open
End of
Sample
Chamber Vacuum
Pump
29
Figure 97
168
Module 5: Injectors
Theory of Injector Operation
Sampling
Sample
Pump On
Sample
Valve Open
Sample
From
Bulkhead
Fitting
30
Figure 98
169
Module 5: Injectors
Theory of Injector Operation
Sample Dwell
31
Figure 99
170
Module 5: Injectors
Theory of Injector Operation
6000000
5000000
Area Response
4000000
5000000-6000000
4000000-5000000
3000000
3000000-4000000
2000000-3000000
2000000
500 1000000-2000000
400 0-1000000
1000000
300
Injection Time
0 200
0 100 200 300 100
400 500
Dwell Time
32
Figure 100
The dwell time must be optimized with respect to the injection time.
171
Module 5: Injectors
Theory of Injector Operation
Sample Compression
33
Figure 101
172
Module 5: Injectors
Theory of Injector Operation
6000000
5000000
4000000
Area Response
5 0 0 0 0 0 0 -6 0 0 0 0 0 0
3000000 4 0 0 0 0 0 0 -5 0 0 0 0 0 0
3 0 0 0 0 0 0 -4 0 0 0 0 0 0
2000000 2 0 0 0 0 0 0 -3 0 0 0 0 0 0
500 1 0 0 0 0 0 0 -2 0 0 0 0 0 0
400 0 -1 0 0 0 0 0 0
1000000
300
I n j e c ti o n T i m e
0 200
0 100 200 100
300 400
C o m p r e ssi o n T i m e 500
34
Figure 102
173
Module 5: Injectors
Theory of Injector Operation
Sample Injection
35
Figure 103
174
Module 5: Injectors
Heated versus Unheated Injector
36
Figure 104
This chart shows the affect of using a heated versus unheated injector. Sensitivity
is not enhanced for all components when a heated injector is used.
One should determine the need for a heated injector based upon the type of
components analyzed.
175
Module 5: Injectors
Column Flow Considerations
130
120
110
100 Inlet Pressure (PSI) @ 1 mL/min Helium
90
80 Inlet Pressure (PSI) @ 2 mL/min Hydrogen
70
60
50
Inlet Pressure (PSI) @ 1 mL/min Hydrogen
40
30
20
10
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Column Length [m]
37
Figure 105
Longer column require higher head pressures. Using hydrogen as the carrier
allows for the use of lower head pressures.
176
Module 5: Injectors
Lab Exercises
Lab Exercises
Lab Exercises
38
Figure 106
177
Module 5: Injectors
Lab: Setting Inlet Options
178
Module 5: Injectors
Module 5 Review
Module 5 Review
1) How does one determine the type of injector to use?
179
Module 5: Injectors
Module 5 Review
180
Module 6: Columns
Columns
Columns
• Micro GC Attributes
• Column Types
• Fundamentals of Separation Theory
• Calculating Column Efficiency and Resolution
• Understanding Column Selectivity
• Common Micro GC Applications
Figure 107
The purpose of this module is to review the key considerations of the columns.
182
Module 6: Columns
GC Hardware
GC Hardware
Major Components
GC Channel or Module
Solid state injector
•Silicon micro-machined
injector either unheated
(ambient) or heated
o
(< 90 C.) Pneumatics
Column
•Narrow bore capillary column
• 0.15 - 0.32mm I.D. Electronics
•isothermally heated(ambient to 180oC.
1 2
Thermal conductivity detector
Silicon micro-machined TCD
Figure 108
183
Module 6: Columns
GC Hardware
Speed
• much shorter analysis time
– 21 minutes for typical Natural Gas Analysis: maximum number of
samples per day is 68
– 2.67 minutes for Micro GC Natural Gas Analysis: maximum
number of samples per day is 539
• increase sample throughput
• use less power and carrier gas
– Micro GC with TCD requires only 3 mL per minute of carrier gas
(Typical GC with TCD requires 30 mL per minute of carrier gas)
Size
• portability for performing measurements at place needed
Sensitivity
• perform TCD measurements with up to 10X lower detection
Figure 109
184
GC.
Standard Deviation
Figure 110
M
et
h
0.0000
0.1000
0.2000
0.3000
0.4000
Et an 0.5000
0.6000
0.7000
0.8000
h e
Et an
h e
Pr yle
n
Pr opa e
op ne
Is yl
ob en
Repeatability
n- uta e
tr B n
an u e
s- tan
2
1- -B e
u
Is But te
ob e ne
ci u ne
s- ty
2 l
Is -B ene
op ut
n en ene
Comparable or better than traditional GC
1, -Pe tan
3
3- -B nta e
M u n
e t e
tr thy adi
an l- e
2- s 1- ne
M -2 B
et -P ut
hy en en
l e
2- 1-P -1- ten
M e Bu e
et nt t
ci hyl ene ene
s- -1
2- B
n- Pen ute
He te ne
xa ne
ne
[Mole %]
[Mole %]
Micro Std. Dev
Conventional GC,
4
Standard Deviation
185
Module 6: Columns
GC Hardware
Module 6: Columns
GC Hardware
GC Module
6“
Detector
Vents SSD
Heater
Control 4“
Carrier
Input &
Valve
Activation
Injector
Column Heater Column Spiral
Analytical and Reference Columns
Figure 111
The function of the TCD requires the use of two columns, an analytical and
reference column.
186
Module 6: Columns
GC Hardware
Analytical Column
Analytical
Column
Reference
Column
Heater
Figure 112
The very small columns allow for the speed and sensitivity of the Micro GC
technique.
187
Module 6: Columns
Column Separation
Column Separation
MOBILE STATIONARY
SAMPLE
PHASE PHASE
Figure 113
188
Module 6: Columns
Column Types
Column Types
Column Types
Open (Capillary)
Packed
Wall Coated
Open Tube
(WCOT)
Micro
PLOT WCOT
Packed
Length 0.25 - 10 cm 4-8m 4 - 14 m
I.D. 0.5 mm 0.15 - 0.32 mm 0.1 - 0.32 mm
Figure 114
There are two primary types of columns used in GC; traditional packed columns
and open tubular capillary columns. Packed columns are literally packed with a
material for adsorption or absorption and capillary columns have inside walls
coated with an adsorbing of absorbing material. Most applications today utilize
capillary type columns; however, there are some specific applications, especially
fixed gas analyses, where packed columns are still used.
The primary differences in packed and capillary columns are the materials of
construction, column length, and column inner diameter. Packed columns are
made of copper, stainless steel, or borosilicate glass. Capillary columns are made
of fused silica. Packed columns are much greater in diameter, typically 2- mm,
whereas capillary columns are 0.5 to .75 mm. Packed columns are usually about
0.5 to 10 meters long whereas capillary columns can be as long as 150 meters.
There are significant differences in column flow rate conditions for the two types
of columns.
Micro GC utilizes primarily the PLOT and WCOT columns.
189
Module 6: Columns
Column Types
Micro-Packed Columns
Adsorbent Packing
Carrier Gas
Porous with large surface area
Figure 115
190
Module 6: Columns
Column Types
Capillary Columns
Wall Coated Open Tubular (WCOT)
Liquid Phase
10
Figure 116
When the stationary phase is a viscous liquid, usually a polymer, the type of
separation is called Gas-Liquid chromatography. This type of separation is used
for about 90% of all the gas chromatographic applications.
In packed columns, the liquid stationary phase is coated onto a solid support and
the resulting material is packed into the column.
In capillary columns, the liquid stationary phase is coated onto the inside walls of
the column.
191
Module 6: Columns
Column Types
11
Figure 117
192
Module 6: Columns
Column Types
Considerations
12
Figure 118
193
Module 6: Columns
Column Types
Column Connection
PLOT columns union contains glass wool to filter out dislodged material.
13
Figure 119
It is very important to avoid any particulates in the system. Glass wool filters are
used to trap particulates.
194
Module 6: Columns
Fundamentals of Separation Theory
Flow A
14
Figure 120
195
Module 6: Columns
Fundamentals of Separation Theory
GOOD GOOD
POOR POOR
15
Figure 121
When characterizing separation the terms efficiency, resolution and selectivity are
discussed. Good efficiency or sharp peaks often leads to good resolution,
however in the diagram above, the first is an example of good efficiency but poor
resolution and the second is an example of good resolution but poor efficiency.
Factors like column diameter, length, column flow rate, and oven temperature can
have an effect on efficiency and resolution.
Selectivity refers to the type of stationary phase in the column. One must assure
that the components of interest will have the ability to interact with the type of
stationary phase of the column.
196
Module 6: Columns
Fundamentals of Separation Theory
R = n k
α-1
4 1 + k α
Factors Affecting
16
Figure 122
197
Module 6: Columns
Fundamentals of Separation Theory
Partitioning
Kd = Distribution Constant
17
Figure 123
198
Module 6: Columns
Fundamentals of Separation Theory
Partitioning Theory
18
Figure 124
199
Module 6: Columns
Fundamentals of Separation Theory
Partition Ratio
mass stationary
= partition ratio
mass vapor
retention time
Partition ratio is the relative amounts of component in the vapor phase versus
the stationary phase. This ratio can be deduced by knowing retention times.
tr - tm
partition ratio = = k
tm
tm = retention time on an unretained peak
tr = retention time of the solute
19
Figure 125
200
Module 6: Columns
Fundamentals of Separation Theory
Partition ratio (i.e., retention) is influenced by temperature and phase ratio (ß).
If ß then k or If ß then k
Phase ratio = relative amount of stationary phase to the radius of the column.
r
ß= 2df
r = radius of column
df = film thickness of column
20
Figure 126
201
Module 6: Columns
Fundamentals of Separation Theory
Choosing Columns
• The greater the film thickness for a given diameter column, the
more retained a compound is.
• As the diameter of a column increases for a given film
thickness, the less the retention.
• By knowing the phase ratio of a column, one can change from
one diameter column to another and always know what film
thickness is needed.
21
Figure 127
The bottom line is that when choosing and comparing columns, it helps to know
the phase ratio.
202
Module 6: Columns
Fundamentals of Separation Theory
Calculating Efficiency
Solute t R' 2
N(eff) = 5.545
Wh
Start t R ' = Retention Time
W h = width at half-height
N = effective theoretical plates
tR Time
22
Figure 128
The absolute retention time, the one printed on the report represents the total time
the component spends on the column from time of injection to time of detector.
Separation occurs because components spend different degrees of time in the
stationary phase. Therefore it would be useful to know how much time a
component interacts with the stationary phase as opposed to being in the mobile
phase.
The assumption is that at any given moment in time the component will be found
either in the stationary phase or traveling along with the carrier gas in the mobile
phase. Another assumption is that all components will spend the same amount of
time in the mobile phase and this time is directly related to the length of the
column. The amount of time spent in the mobile phase is equal to the time it
takes for a non-retained species to elute from the column. The assumption is that
there is no time spent in the stationary phase since there is no interaction with the
stationary phase. In many analyses the solvent has very little interaction with the
stationary phase, eluting first and very early in the analysis.
One can calculate the time spent in the stationary phase by subtracting the non-
retained species retention time from the reported retention time of a component.
This is known as the corrected retention time.
Separation Theory assumes that column is divided into a number of zones called
theoretical plates. A simplified definition would be to think of a theoretical plate,
203
Module 6: Columns
Fundamentals of Separation Theory
as each time there is an interaction between the component and the stationary
phase.
The number of theoretical plates can be calculated using the formula in the figure
on the previous page. The corrected retention time and the width of the peak is
substituted into the formula. One has a choice of which peak width to use, one at
the base of the peak, or one at the half-height. The constant 5.545 is related to the
use of the peak width at half-height.
The effective number of plates can then be related to the column length by
calculated the plates per meter or height equivalent to a theoretical plate (HETP).
The more efficient the column, the more plates per meter or the smaller the
HETP.
204
Module 6: Columns
Fundamentals of Separation Theory
Efficiency is a function of
B +C µ
HETP = A +
µ
the carrier gas linear
velocity or flow rate.
Plot of HETP vs. linear velocity is know as the Van Deemter plot.
The linear velocity value at the minimum of the curve is the optimum value for achieving
the best efficiency.
23
Figure 129
The Van Deemter plot relates the linear velocity or flow rate to the column
efficiency. From the plot, one can determine what column flow rate or linear
velocity to us for an analysis to yield the sharpest peaks. In methods
development, one may want to spend some time generating a plot as a tool for
determining the column flow to use.
In the Van Deemter equation: A = eddy diffusion term
B = longitudinal or ordinary diffusion term
C = resistance to mass-transfer term
Since capillary columns have no packing material to cause obstruction to flow,
there is no eddy diffusion term. Therefore, the HETP is fundamentally lower than
for packed columns. The longitudinal diffusion term is related to the type of
carrier gas used. The smaller molecular weight gases, like hydrogen as compared
to nitrogen offer more diffusivity and thus reduce the value of “B”. The “C” term
accounts for the resistance to mass transfer in the liquid phase. An obvious way
of reducing this term is to reduce the liquid film thickness.
205
Module 6: Columns
Fundamentals of Separation Theory
24
Figure 130
Based upon the VanDeemter plot, the suggested column head pressure ranges
should produce the best efficiency.
206
Module 6: Columns
Fundamentals of Separation Theory
25
Figure 131
When trying to optimize efficiency, one can try the above changes in order to
make improvements.
207
Module 6: Columns
Fundamentals of Separation Theory
Resolution
Resolution is a measure of the ability of a
column to separate two peaks.
RT = 4.41
Resolution is measured in terms of two
RT = 4.59
RT = 5.10
6.0E5 adjacent peaks which we want to separate.
Generally, the most difficult pair is chosen; if
these can be pulled apart successfully then
5.0E5
all of the others will be resolved as well.
R= (W + W )
3.0E5 1 2
2.0E5
R = 1.5 Means Baseline Separation
1.0E5
Doubling column length does not double resolution!
4 4.5 5 5.5 6
Time ( Min.)
The more efficient the column, the greater the chances of better resolution.
26
Figure 132
Resolution can also be calculated. The above diagram presents the formula using
retention times and peak widths of two adjacent peaks in question.
208
Module 6: Columns
Fundamentals of Separation Theory
OV-101 0.4µ
1.0
25 m x 0.25 mm
He 15 m x 0.25 mm
.8
Glass WCOT
.6 SE - 52
H2
.4 Isothermal, 150º C
.2
10 20 30 40 50 60 70 80 90
R = 1.17 R = 1.37
Average Linear Velocity (cm/sec)
Efficiency curves for a 25 m x 0.25 mm id WCOT column Effect of carrier gas on the resolution of n-
with 0.4 um of OV-101. heptadecane and pristane.
27
Figure 133
The type of carrier gas used effects efficiency as already discussed in reference to
the Van Deemter plot. Using hydrogen allows for the use of faster flow rates or
linear velocity, which will result in shorter run times. However, if nitrogen is
used at the fast linear velocity of 58 cm/sec., resolution of the peaks is lost.
209
Module 6: Columns
Fundamentals of Separation Theory
• Isothermal Operation
– Oven is maintained at a constant temperature throughout the
analysis.
– Stop time is set with the initial time.
– Excessive broadening of later eluting peaks.
• Column Conditioning
– Mol Sieve 5A and Al oxide/KCL Plot columns become deactivated.
– Set Oven Temp to 180 degrees for min. 4 hours to overnight.
– Cleans out moisture and residues from column.
28
Figure 134
Only Isothermal analyses are available in Micro GC. Traditional GC offers the
alternative of temperature programmed analyses.
Plot columns may require periodic reconditioning.
210
Module 6: Columns
Typical Applications
Typical Applications
Typical Applications
Enterprise Application Enterprise Type Application
Type
29
Figure 135
This chart lists some typical applications for the Micro GC.
211
Module 6: Columns
Typical Applications
Application Notes
Natural Gas Literature Refinery Gas Literature
30
Figure 136
212
Module 6: Columns
Typical Applications
31
Figure 137
213
Module 6: Columns
Typical Applications
Measurement of C1-C3
Houston, Texas
2
1 Nitrogen 0.717%
Natural Gas Sample
2 Methane 92.685% Taken on-line
1 1034 BTU
3 Carbon Dioxide 1.807%
4
4 Ethane 3.598%
3
5 Propane 0.808%
5
32
Figure 138
214
Module 6: Columns
Typical Applications
1 i-Butane 0.094%
2
2 n-Butane 0.137%
1
3 i-Pentane 0.045%
4 n-Pentane 0.037%
3 4
5 Hexane 0.035%
3
6 Heptane 0.019%
4
7 Octane 0.008% 10 15 20
8 Nonane 0.010% Retention Time (Second)
5
Natural Gas Sample 6
7 8
Taken on-line
1034 BTU 10 20 30 40 50 60 70 80 90 100110120130140150
Retention Time (Second)
33
Figure 139
215
Module 6: Columns
Typical Applications
1 2&3
1. Composite
2. i-Pentane 1510 ppm
3. n-Pentane 1500 ppm
4. Hexane 500 ppm
5. Water
6. Heptane 200 ppm
7. 1-Propanethiol 150 ppm
8. 2-Propanethiol 120 ppm
4
9. 2-Me-2-propanethiol
100 ppm
5 6 78 9
0 20 40 60 80 100 120 140 160
Retention Time (Second)
34
Figure 140
216
Module 6: Columns
Typical Applications
Simple Compositions:
Sources:
• Distillation
• Catalytic reforming
• Fractionating / separating units
35
Figure 141
217
Module 6: Columns
Typical Applications
36
Figure 142
218
Module 6: Columns
Typical Applications
37
Figure 143
219
Module 6: Columns
Typical Applications
Complex Compositions
Sources
• Cracking catalysts
• Thermal cracking
• Fractionating units for cracking units
38
Figure 144
220
Module 6: Columns
Typical Applications
39
Figure 145
221
Module 6: Columns
Typical Applications
40
Figure 146
222
Module 6: Columns
Typical Applications
1. methane/N2 8. neopentane
2. ethylene/ethane 9. iso-pentane
3. propane/propylene 10. 1-pentene Helium carrier
4. iso-butane 11. n-pentane
5. iso-butylene/1-butene 12. cis-2-pentene 0.47 mL/min.
6. 1-butene 13. cyclopentene
7. cis-2-butene 14. n-hexane
41
Figure 147
223
Module 6: Columns
Typical Applications
1. methane/N2 8. neopentane
2. ethylene/ethane 9. iso-pentane
3. propane/propylene 10. 1-pentene
4.
5.
iso-butane
iso-butylene/1-butene
11.
12.
n-pentane
cis-2-pentene
Hydrogen carrier
6.
7.
1-butene
cis-2-butene
13.
14.
cyclopentene
n-hexane
0.77 mL/min.
42
Figure 148
224
Module 6: Columns
Lab Exercises
Lab Exercises
Lab Exercises
43
Figure 149
225
Module 6: Columns
Lab Exercise: Changing Column Parameters
226
Module 6: Columns
Module 6 Review
Module 6 Review
1) Why would one routinely monitor the number of theoretical plates or plates
per meter for a column?
5) Does the type of liquid phase effect the elution order of unsubstituted
aliphatic hydrocarbons (i.e., C-3 through C-8)?
6) What happens to the peak shape of later eluting peaks as the oven
temperature is reduced?
227
Module 6: Columns
Module 6 Review
228
Module 7: Detectors
Detectors
Detectors
Figure 150
The purpose of this module is to review the key considerations of the detectors.
230
Module 7: Detectors
Detectors
The Micro GC
GC Channel or Module
Solid state injector
•Silicon micro-machined
injector either unheated
(ambient) or heated Pneumatics
(< 90oC.)
Column
•Narrow bore capillary column
• 0.15 - 0.32mm I.D.
•isothermally heated(ambient to 180oC. Electronics
1 2
Thermal conductivity detector
Silicon micro-machined TCD
Figure 151
231
Module 7: Detectors
Detectors
GC Detector - A Definition
Figure 152
TCD
Filament temperature increases as analytes present in the carrier gas pass over it,
causing the resistance to increase.
232
Module 7: Detectors
Detectors
GC Module
6“
Detector
Vents SSD
Heater
Control 4“
Carrier
Input &
Valve
Activation
Injector
Column Heater Column Spiral
Analytical and Reference Columns
Figure 153
This diagram shows the position of the TCD relative to the column and injector.
233
Module 7: Detectors
Setting Detector Parameters
Figure 154
234
Module 7: Detectors
Detector Response Characteristics
Figure 155
235
Module 7: Detectors
Detector Response Characteristics
Figure 156
236
Module 7: Detectors
Detector Response Characteristics
Dynamic Range
Dynamic range is a measure of response vs. Quantity
Response is the signal produced by the sample.
Dynamic Range
Response increases reproducibly with
Response increased quantity.
Quantity
Quantity
Figure 157
237
Module 7: Detectors
Thermal Conductivity Detection
FLOW
Figure 158
238
Module 7: Detectors
Thermal Conductivity Detection
Wheatstone Bridge
10
Figure 159
The universal micro Solid State Detector (SSD) is a four filament resistive
Wheatstone Bridge where two branches of a circuit are joined to compare
resistance. When pure carrier gas passes through both branches of the bridge, the
cooling effect on the filaments is the same and the bridge, which was initially
balanced, will stay balanced. When a component elutes from the column and
touches one set of filaments, the cooling effect on the one set of filaments will
change. With the change in temperature the resistance of the filaments will
change and the bridge will become unbalanced. The bridge will become balanced
again when pure carrier gas passes again through the filaments.
The sample effluent from the analytical column flows over two
filaments and the flow from the reference column flows over the other two
filaments. In addition, the jumper in the controller board controls the current to
the filaments. This current change is for applications using argon or nitrogen as a
carrier gas.
239
Module 7: Detectors
Thermal Conductivity Detection
240
Module 7: Detectors
Thermal Conductivity Detection
11
Figure 160
241
Module 7: Detectors
Thermal Conductivity Detection
12
Figure 161
242
Module 7: Detectors
Thermal Conductivity Detection
Thermal Conductivity
The ability of a substance to conduct heat from a warmer to a cooler surface.
Hydrogen 41.6 2
Helium 34.8 4
Methane 7.2 16
Nitrogen 5.8 28
Argon 3.8 40
Pentane 3.1 72
Hexane 3.0 86
13
Figure 162
This table lists the relative thermal conductivity values for several gases. When
helium is used as a carrier gas, all components other than hydrogen have lower
thermal conductivity values. Therefore, the filament will heat up when the gas
stream is “contaminated” with sample effluent of any gas with a lower thermal
conductivity compared to helium. If hydrogen were present, the filament would
become cooler; therefore, a negative peak would result.
243
Module 7: Detectors
Lab Exercises
Lab Exercises
Lab Exercises
14
Figure 163
244
Module 7: Detectors
Lab Exercise: Modify the Detector Sensitivity
245
Module 7: Detectors
Module 7 Review
Module 7 Review
1) Why is it important to understand the relationship of thermal conductivity
values for various gases?
4) Why is it important to use very pure carrier gas that is free of particulates?
246
Module 8 Calibration, Sequences, and
Reporting
247
Module 8 Calibration, Sequences, and Reporting
Calibration, Sequences, Reporting
• Quantification Theory
• Calibration Theory
• Calibration Menus
• Multi Level Calibration
• Reporting Options
• Automated Sequences
• BTU Reporting
Figure 164
248
Module 8 Calibration, Sequences, and Reporting
Calibration, Sequences, Reporting
Setup Acquisition
Parameters
Integration
Optimization
Calibration Setup
Reporting Options
Figure 165
249
Module 8 Calibration, Sequences, and Reporting
Qualitative Analysis
Qualitative Analysis
Qualitative Analysis
Module 4 discussed:
• Create a Method (setting up instrument parameters)
• Fill out the Peak Table with expected peaks
• Run the Analysis
• Identify the Peaks
• Set up the Timed Events
• If necessary, adjust any GC parameters
• Optimize the Timed Events Table
• Set the peak retention times and retention time windows
graphically
Reference: User’s Manual HP EZChrom Chromatography Data System, pages 96-102.
Figure 166
250
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
Quantitative Analysis
What is Quantification?
Figure 167
After the peaks have been integrated and identified, the next step in the analysis is
quantification. Quantification uses peak area to determine the concentration of a
compound in a sample.
A quantitative analysis involves many steps, which are briefly summarized as
follows:
• Know the compound you are analyzing.
• Establish a method for analyzing samples containing this compound.
• Analyze a sample or samples containing a known concentration or
concentrations of the compound to obtain the response due to that
concentration. You may alternatively analyze a number of these samples with
different concentrations of the compounds of interest if your detector has a
non-linear response. This process is referred to as multi-level calibration.
• Analyze the sample containing an unknown concentration of the compound to
obtain the response due to the unknown concentration.
251
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
252
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
Quantification Calculations
Figure 168
253
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
44248397 99.99
1598210
Area % of Ethane = X 100 = 3.61
44248397
AREA (pk)
Σ AREA (pk)
Figure 169
254
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
44248397 99.99
Response factor takes into account that equal amounts do not yield equal
detector response.
Figure 170
255
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
Calibration Comparison
Area Percent
• No calibration mixture required
• For Accuracy: All peaks must elute and all peaks must be
detected.
External Standard
• Requires calibration mixture
• Not all peaks need elute or be detected
• Choose the components in the mixture to quantitate.
Normalization
RF (pk) X AREA (pk)
NORM % = X 100%
[RF (pk) X AREA (pk)]
Figure 171
In the Normalization method, response factors are applied to the peak areas (or
heights) to compensate for changes that occur in detector sensitivity for the
different sample components. The Norm% report is calculated the same way as
an ESTD report except that there is an additional step to calculate the relative
rather than absolute amounts of compounds.
The Norm% has the same disadvantages as the Area% report. Any changes that
affect the total peak area will affect the concentration calculation of each
individual peak. The Norm% report should only be used if all components of
interest are eluted/migrated and integrated. Excluding selected peaks from a
normalization report will change the reported results in the sample.
256
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
C A
C A
C A B
A B
B
B
1 0.5 1.5 1.5 1.0 2.0 2.0 1.5 3.0 3.0 2.0 4.0
AREA/HEIGHT
Level 4
Level 3
AREA/HEIGHT
Level 2
Level 1
Area
Response of
Unknown
CONCENTRATION CONCENTRATION
Calculated
Amount MULTI-LEVEL FOR
COMPONENT C
Figure 172
The above example shows the generation of single and multi-level calibration
curves. In the first diagram the results of four analyses are shown. Four
calibration standards of known amounts of three different components were
prepared and analyzed.
In the single point calibration, the area or height response for Component A in the
Level 1 calibration mixture is plotted versus the known concentration of
Component A. When a mixture containing an unknown amount of A is analyzed,
its area count is noted on the plot and its concentration is determined by the
extrapolation from the plot.
For the Multi-level calibration above, each of the area counts for the four different
concentrations of Component C is plotted versus their known amounts. When
complete one would have three calibration plots, one for each component.
Multilevel calibration is used when it is not sufficiently accurate to assume that a
component shows a linear response or to confirm linearity of the calibration
range. Each calibration level corresponds to a calibration sample with a particular
concentration of components. Calibration samples should be prepared so that the
concentration of each component varies across the range of concentrations
expected in the unknown samples. In this way it is possible to allow for a change
in detector response with concentration and calculate response factors
accordingly.
257
Module 8 Calibration, Sequences, and Reporting
Quantitative Analysis
258
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
Calibration Procedure
10
Figure 173
259
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
11
Figure 174
One can create the calibrated method as the samples are analyzed, or sometimes it
is more convenient to collect the data and then do the data processing on the
previously collected data files.
260
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
For definitions of the parameters in the table see pages 35 -37 of the User’s Manual.
Figure 175
It is assumed that the method has been created and a peak table exists. The
integration parameters should have been optimized. Verify that all the pertinent
peaks in the analysis are listed in the table.
261
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
Calibration Set Up
13
Figure 176
262
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
Uncalibrated Peaks RF
This option allows the assignment of a blanket response factor to all unidentified
peaks. This response factor is multiplied by the area of the unknown peaks to give
an amount value. The default is Zero.
Multiplication Factor
This is a multiplication factor that is applied to all quantitated peaks, both in
calibration standards, as well as, samples. This feature is useful when a
concentrator or dilutor is used in the sampling system. The default is One.
263
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
14
Figure 177
To enter the concentrations from your calibration standard gases into your Peak
Calibration Table:
• Select Method
• Select Peak Calibration.
• Select Channel A or B
• Enter the amount information for each level using Page Down as necessary.
The calibration amounts must be entered in order from least to most
concentrated. A level one calibration will require only one entry per
compound.
• To switch compounds, click on [Prev] or [Next].
• The area information can also be entered now, manually, or later while
performing a calibration on stored or newly acquired data.
• To plot the new calibration points, click on the Plot button. The slope of this
plot defines the response factor for that compound.
264
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
265
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
Multi-Level Calibration
15
Figure 178
Stored calibration
First, clear the statistics for a clean start.
• Select Data.
• Select Clear Statistics.
• Select [OK].
Now select the stored data file to use as a calibration standard.
• Select Data.
• Select Open.
• Enter the data file name manually, or double click on the data file name.
Now set up to calibrate.
• Select Calib.
• Enter the Calibration Level (The calibration level refers to which calibration
gas dataset is being analyzed: first, second, third, etc.).
266
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
• Select [OK].
To finish the calibration:
• Select Analyze.
• Repeat the preceding steps for each calibration gas standard dataset.
Run calibration
• Attach the first cal gas standard.
• Select Calib.
• Enter the Calibration Level.
• Select [OK].
• Start the GC.
Calibration will begin immediately. Prior to sampling, the EZChrom menu title
bar will display the following message:
267
Module 8 Calibration, Sequences, and Reporting
Calibration Procedure
Figure 179
Once the Calibration is set up, the Peak Table will list the response factors.
268
Module 8 Calibration, Sequences, and Reporting
Reporting
Reporting
Setup Acquisition
Parameters
Integration
Optimization
Calibration Setup
Reporting Options
17
Figure 180
269
Module 8 Calibration, Sequences, and Reporting
Reporting
Reporting
18
Figure 181
Caution Chromatograms generate many data points that occupy a lot of memory on
your storage media. If you are printing multiple chromatograms simultaneously, you may
overload the printer memory and lock up your system.
1. Select Method.
2. Select Print Options.
3. Check any of the reports you wish to print, at the completion of the analysis,
for both channels.
4. Select [OK]. Remember to check the Print box in the Start window
prior to beginning a run. The available options are:
Chromatogram (full scale and/or zoomed)
Hard copies of the chromatograms can be generated for channels A or B. If the
Zoomed option is selected, the zoomed chromatogram shown in the lower display
will be printed.
270
Module 8 Calibration, Sequences, and Reporting
Reporting
Any or all of the three available reports (Area %, ESTD, or Norm) for channels A
and B can be printed between autoruns.
271
Module 8 Calibration, Sequences, and Reporting
Reporting
Display Options
19
Figure 182
272
Module 8 Calibration, Sequences, and Reporting
Sequences
Sequences
Automated Sequences
Number of
Consecutive Runs
20
Figure 183
Running the GC and obtaining accurate results is a simple task once a good
Method has been developed.
• Select Start.
• A run window will appear which prompts for information regarding the
automated run sequence.
• In the Run ID space, enter a name for the data, which is to be collected and
stored. This may be a maximum of eight characters and DOS illegal
characters are forbidden.
• Specify the Number of Runs. If the number of runs is greater than one,
the run number is appended to the Run ID as a data file extension (for
example: run.1, run.2, etc.).
• Specify the Time Between Injections in seconds. Note what the
delay interval is between injections; make sure this interval is long enough to
accommodate the run time, time between runs, and any necessary printing
time.
273
Module 8 Calibration, Sequences, and Reporting
Sequences
• Check if the data is to be Saved and/or Printed after each run. Make
sure that printing selections have been made in Print Options before
selecting Print here.
• Check if the data is to be output in a DIF or PRN format.
The DIF format saves external standard report peak name and amount information
as ASCII characters delimited by tabs (EXCEL format). The filename used is the
Run ID followed by the extension .DIF. The PRN format saves the peak name and
amount report information as ASCII characters delimited with quotes and
numbers delimited by commas (Lotus format). The filename used is the Run ID
followed by the extension .PRN.
Check Extended if you would like to include the area and retention time in the
information saved to DIF or PRN files.
Recall allows one to reprocess data.
274
Module 8 Calibration, Sequences, and Reporting
BTU Reporting
BTU Reporting
BTU Reporting
BTU Software is required. A third window is available.
21
Figure 184
275
Module 8 Calibration, Sequences, and Reporting
BTU Reporting
276
Module 8 Calibration, Sequences, and Reporting
Lab Exercises
Lab Exercises
Lab Exercises
22
Figure 185
277
Module 8 Calibration, Sequences, and Reporting
Lab 1: Preparing Calibration Standards
Concentration = Concentration
278
Module 8 Calibration, Sequences, and Reporting
Lab 1: Preparing Calibration Standards
------------------ ------------------
Area Area
Level 1. . into .= .
Level 2. . into .= .
Level 3. . into .= .
Level 4. . into .= .
Level 5. . into .= .
279
Module 8 Calibration, Sequences, and Reporting
Lab 1: Preparing Calibration Standards
Type in the Run ID and select Save for saving the data.
280
Module 8 Calibration, Sequences, and Reporting
Lab 1: Preparing Calibration Standards
281
Module 8 Calibration, Sequences, and Reporting
Lab 2: Creating a Multi-Level Calibration
282
Module 8 Calibration, Sequences, and Reporting
Lab 2: Creating a Multi-Level Calibration
1) At the EZChrom Main Menu Bar, click on Data and then select Clear
Statistics. Then click on [OK]. All previous results that have
accumulated in the statistics registers are cleared and you are ready to begin
your analysis.
2) Click on Start at the Main Menu Bar. Enter 5 or more for the Number of
runs in the Run window. Save the data. Analyze your calibration sample at
least five times.
Make sure the number of runs found in Calibration Setup under the
Method menu equal the number of runs found in the Run window from the Start
Menu.
3) Click on the {Start} button in the Run window or press <Enter> to
begin the series of 5 runs.
4) After the last analysis of the sequence, click on Reports and then choose
External Standard. Observe the Relative Standard Deviation values
reported (%RSD) for all the peaks. The % SD should be equal to or less than
2.0% for satisfactory quantification. You may have to enlarge the window or
scroll down to see all the peaks.
If some of the peaks in your calibration are not reported (0.00 amounts), go to
Chapter 5 Troubleshooting of the User manual. If all peaks are reported and the %
RSD of all peaks is equal to or less than 2.0%, you can continue with the editing
of the peak table.
283
Module 8 Calibration, Sequences, and Reporting
Lab 2: Creating a Multi-Level Calibration
while dragging the mouse so that the whole entry is blackened. Now press the
<Del> key to remove the entry or simply type the new entry.
8) Using the values determined for your new calibration standard, reenter the
concentration for all other components in the Peak Table. All text and
numbers in the cells, other than the values in the instrument status window,
may be edited in this way.
9) Enter a "Y" under the BP column for those components that elute in a
component window and that you want to group. Normally these are isomers
of a particular component. Grouping is set up for hexanes, heptanes, and
octanes to generate a composite C 6 + calibration amount. Your calibration
sample should only have one peak for each of those components. Your
unknown samples will be more complex.
10) Click on [OK] when finished.
Caution Do not attempt to change the retention times of your calibration table
manually.
Part 3: Recalibrating
Once repeatability has been checked and found acceptable, your peak table has
been set up, and all peaks have been detected, run the calibration sample to obtain
new response factors through the following steps:
11) At the EZChrom main menu bar, click on Calib. You will access the
Calibration window. Enter 1 for the Calibration Level and click on [OK].
12) Connect your calibration standard to the flow controller. Make sure that the
transfer lines have been purged and your calibration sample is flowing into
the GC inlet.
13) Click on Start at the EZChrom Main Menu Bar. You will open the Run
ID window. Enter the number of runs (three runs are recommended), and
click on Save to store your runs.
14) Click on the [Start] button at the Run window or press <Enter>.
A ratio of areas to concentrations for your calibration standard (Response Factors)
are calculated for each component peak. These factors are used for the calculation
of your reports when unknown samples are analyzed.
15) Click on Method and Save As to store the new calibration in your
method file. The current method name is blackened and you must type a new
name to differentiate it from the existing one. Use the extension ".met" to
distinguish this file as a method file.
284
Module 8 Calibration, Sequences, and Reporting
Lab 2: Creating a Multi-Level Calibration
Caution If you click on Save instead of Save As, or if you store your new calibration
under the original filename, it will replace the original calibration with your new calibration.
16) Repeat the procedure for the remaining calibration 4 levels of standards.
Compound: _____________________________________
285
Module 8 Calibration, Sequences, and Reporting
Lab 3: Exploring Reporting Options
286
Module 8 Calibration, Sequences, and Reporting
Lab 3: Exploring Reporting Options
287
Module 8 Calibration, Sequences, and Reporting
Lab 4: BTU Reporting
288
Module 8 Calibration, Sequences, and Reporting
Module 8 Review
Module 8 Review
1) Describe the steps needed to prepare a single level calibration table.
b. Multiplication Factor
c. Uncalibrated Peaks RF
289
Module 8 Calibration, Sequences, and Reporting
Module 8 Review
290
Module 9: Troubleshooting
Troubleshooting
Troubleshooting
Figure 186
292
Module 9: Troubleshooting
Common Problem Areas
• Sample line
• Chromatograph
• Computer
• Communications
Figure 187
We will review some common problem areas associated with the Micro GC.
293
Module 9: Troubleshooting
Common Problem Areas
• Inlet pressure
• Temperature
• Clogged line
• High moisture
• Flow restrictions
• Pressure biasing
• Component failure
Figure 188
294
Module 9: Troubleshooting
Common Problem Areas
Hardware
– Power supply
• No power
• fluctuations
– RF interference
– Module
– Detector
– Carrier gas
– Heated inlet
– Pump
– Solenoid valve
Figure 189
295
Module 9: Troubleshooting
Common Problem Areas
• Sampling time
• Method Parameters
– Column temperature
– Injection time
– Run time
– Sensitivity
• File management
– Data file path
– Format
Figure 190
296
Module 9: Troubleshooting
Common Problem Areas
• No peaks
• Saturation
• Co-elution
• Carryover
• Baseline problems
• Retention time shift
• Out of calibration
• Negative peaks
• Integration parameters
Figure 191
297
Module 9: Troubleshooting
Common Problem Areas
• Cables
• Power
• Software conflicts
• GC setup
• Printing
• Lockups
Figure 192
298
Module 9: Troubleshooting
Common Problem Areas
• Phone line
• Modem
• Serial connection
• Communications software
Figure 193
299
Module 9: Troubleshooting
Common Problem Areas
Troubleshooting Strategies
Figure 194
300
Module 9: Troubleshooting
Routine Troubleshooting
Routine Troubleshooting
Routine Troubleshooting
10
Figure 195
You should refer to the troubleshooting tables available in the User Manuals.
301
Module 9: Troubleshooting
Routine Troubleshooting
1. Verify Flows
Check for reference and analytical column flow at vents on rear panel of GC
2. Verify Communication between GC & PC
Check for cable connections, Send/Receive Method to GC
3. Verify Pressure readings
Verify pressure readings are displayed on CRT
4. Verify Detector Autozero is within limits:
(±) 300 mv
5. Verify Detector noise is within acceptable limits:
2000µv within 2 second window.
6. Check All column heaters ramp to maximum in a timely manner:
CT max < 4 min
7. Check board configuration
Board can be configured in either the Ar mode or the He mode.
He mode with Ar carrier or Ar mode with He carrier is very bad,
will burn detector.
8. Zero inject baseline can show signs of micro-valve failure
11
Figure 196
302
Module 9: Troubleshooting
Monitoring Instrument Status
12
Figure 197
One should routinely check the instrument status to verify the instrument set
points.
303
Module 9: Troubleshooting
Communication Error Numbers
2 Character was not read from the hardware before the next
character arrived. The character was lost.
13
Figure 198
304
Module 9: Troubleshooting
GC Verification Tool
GC Verification Tool
GC Verification Tool
14
Figure 199
305
Module 9: Troubleshooting
GC Verification Tool
MTI GC Verification
15
Figure 200
306
Module 9: Troubleshooting
GC Verification Tool
MTI GC Verification
16
Figure 201
307
Module 9: Troubleshooting
MTI GC Setup
MTI GC Setup
MTI GC Setup
17
Figure 202
308
Module 9: Troubleshooting
MTI GC Setup
MTI GC Setup
18
Figure 203
309
Module 9: Troubleshooting
Maintenance Procedures
Maintenance Procedures
Maintenance Procedures
19
Figure 204
310
Module 9: Troubleshooting
Module Change Tool
20
Figure 205
311
Module 9: Troubleshooting
Lab Exercises
Lab Exercises
21
Figure 206
312
Module 9: Troubleshooting
Lab 1: Troubleshooting Problems
313
Module 9: Troubleshooting
Lab 2: Final Exam
Part 1: “I’ve run the Instrument, But Can I Get the Right
Answers?”
1) 1. Load your method.
2) 2. Run your unknown.
3) 3. Fill out your answer sheet.
4) 4. Take a deep breath, your done!
314