InstReq multiNC UVHS Pharma UV en

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Announcement of Readiness for Installation

Announcement of Readiness for


Installation
multi N/C UV HS, multi N/C pharma UV

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 1/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

Table of contents
1 Explanatory note............................................................................................................................................................. 3

2 Installation conditions .................................................................................................................................................... 4


2.1 Installation location.......................................................................................................................................... 4
2.2 Power supply..................................................................................................................................................... 4
2.3 Gas supply ......................................................................................................................................................... 5
2.4 Control computer.............................................................................................................................................. 6

3 Installation preparation and required reagents ............................................................................................................ 7

4 Device layout and space requirements .......................................................................................................................... 8

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 2/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.de Version: A (03/2022)
Announcement of Readiness for Installation

1 Explanatory note
The device may only be installed by our customer service or by persons authorized by Analytik Jena.
It is imperative that at least one person designated to operate the device is present during the briefing given by the ser-
vice technician.
Please provide us with confirmation that all the installation requirements have been met. We will be happy to send you
more detailed information if required.
Analytik Jena GmbH
Konrad-Zuse-Straße 1
D-07745 Jena
Phone: +49 3641/77 7407, Fax: +49 3641/77 7449
E-mail: service@analytik-jena.com
With your signature you confirm that all the prerequisites for the complete installation of the device will have been ful-
filled before the arrival of our service technician.
If this is not the case and as a result additional costs are incurred (for example, it is necessary to arrange a second visit),
we reserve the right to charge for this according to our price list.
Please complete in block capitals:

Company

Department

Address
(street, number,
zip code, city)

Phone

E-mail

Customer number

Order number

Please confirm the details on this page by entering the date and your signature:

Date

Signature

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 3/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

2 Installation conditions

2.1 Installation location

Guaranteed Not guaranteed


The installation location is on the first floor or accessible by a
suitable elevator.
If "no", please describe the conditions:

It is possible to reach the installation building with a trans-


porter.
If "no", please describe the conditions:

The device will be delivered directly to the final device location


by a transportation company.
The delivery by this company requires the presence of a com-
petent person.
The device must be positioned in such a way that allows easy
access from all sides. Alternatively, a heavy duty roller table
can be used.
The installation location is free of
¡ Direct sunlight or heat radiation from heaters
¡ High dust contamination
¡ Heavy mechanical impacts and vibration
¡ Magnetic fields
¡ Drafts
¡ Corrosive vapors, with the air in the room preferably be-
ing low in TOC and NOx
The installation location meets the following environmental
conditions:
¡ +10 to 35 °C (air-conditioning recommended)
¡ max. 90 % relative humidity at 30 °C
¡ 0.7 to 1.06 bar
¡ Storage temperature: +5 to 55 °C

2.2 Power supply

The device operates on single-phase alternating current. The installation of the electrical equipment in the laboratory
must comply with the IEC 38 standard. Uninterruptible power supply units must be used in the event of severe power
disturbances.
4 grounded power outlets are required for the basic device (with PC, monitor and liquid autosampler).

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 4/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

An additional power outlet must be provided for each of the following optional modules:
¡ HT 1300 solids module
¡ Manual TIC solids module

Guaranteed Not guaranteed


Voltage 100 to 240 V

Frequency 50/60 Hz

Typical average power con- 150 VA


sumption
Maximum power consump- 200 VA
tion
Fuse protection (mains side) 10 A H

If the electrical connection conditions differ, please describe


the conditions:

HT 1300 solids module

Guaranteed Not guaranteed


Voltage 230 V ± 10 %

Frequency 50/60 Hz

Typical average power consumption 700 VA

Maximum power consumption 1000 VA

Fuse protection (mains side) 10 A H

2.3 Gas supply

The gas supply to the basic device can be nitrogen or argon. Please select an option. Oxygen is required for the optional
solids modules.

Guaranteed Not guaranteed


Basic device carrier gas
Option 1: Nitrogen (≥5.0)

Option 2: Argon (≥4.6)

Inlet pressure 400 to 600 kPa

Flow rate 15 l/h

Connection For gas hose with ID 4 mm, OD 6 mm

Gases for optional solids modules


HT 1300 solids module Oxygen (≥2.5)

Man. TIC solids module Oxygen (≥2.5)

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 5/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

Guaranteed Not guaranteed


Connection: one T-piece respectively for gas hose with diameter ID 4 mm , OD 6 mm for
oxygen supply for HT 1300/man. TIC solids module

2.4 Control computer

The control computer can be supplied by Analytik Jena, but it must be ordered as a separate item. The control com-
puter must meet the following minimum requirements:

Guaranteed Not guaranteed


Processor Min. 3.2 GHz
Disk drive Min. 40 GB
RAM Min. 4 GB
Screen resolution Min. 1280 x 780
USB port Min. 1 USB 2.0 interface, for connecting
the basic device
CD/DVD drive For software installation

Operating system Windows 10 or higher, Windows 7/8.1 is


supported, 32 or 64 bit

Internet connection during installation is recommended. When connected to the Internet, updates to the operating sys-
tem as well as to additional software (Microsoft Office) are ensured. Remote access to the device is then also possible
for remote diagnosis. Customer consent is always required for remote diagnosis.
The analysis device can be operated without Internet connection!
Administrator rights are required on the control computer for the installation.
If the control computer is to be integrated into a network, this must take place before installing the software. Installa-
tion within a virtual environment constitutes an exceptional case. Consult with the Analytik Jena Service department or
the providing partner organization for this.

Guaranteed Not guaranteed


Internet access is available during installation of the control software.

NET Framework 3.5 can be activated for installation.

Administrator rights are ensured during installation. Alternatively, the customer's IT de-
partment can provide support.
Network connection LAN

WLAN

Optionally, a Microsoft Office package can be included in the order. The customer must have a Microsoft ID for licens-
ing.

Yes No
The installation of an Office package is included in the order.

A Microsoft ID for the Microsoft Office license is provided.

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 6/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

3 Installation preparation and required reagents


The customer must provide the following reagents and standard solutions for installation:

Reagents and standards multi N/C UV HS multi N/C pharma UV


Sodium peroxodisulfate solution (80 g/l)

Sulfuric acid H2SO4 (2 mol/l)

10 % phosphoric acid

TOC water in accordance with Ph.Eur. and USP (TOC <100 /


ppb)
Potassium hydrogen phthalate (TOC) /

Saccharose, p-benzoquinone (TOC) /

Sodium carbonate, sodium hydrogen carbonate (TIC) /

Calcium carbonate (HT 1300, man. TIC solids module) /

All chemicals must meet the purity grade p.a. (pro analysi – analytically pure).

Offer from Analytik Jena


Analytik Jena offers the following reagents and standard solutions. Please order these solutions separately.

Reagents and standard solutions offered


multi N/C pharma UV 899-CLUVSA: TOC calibration kit 0.6 mg/L C - UV, sucrose solution stabilized with H2SO4;
899-JPPT2SA: JP SST kit, UV, JP performance test kit stabilized with H2SO4;
899-LIN1SA: Linearity test kit, UV, 0.25 mg/L C sucrose for UV digestion TOC systems;
899-TOCBNS: TOC stock solution para-benzoquinone USP, 100 mg/L C;
899-TOCSNS: TOC stock solution sucrose USP, 100 mg/L C;
899-USPS1SA: USP TOC SST kit, UV, SST test kit stabilized with H2SO4;
899-W500SA: TOC preparation water, UV stabilized with H2SO4 for UV digestion TOC systems
multi N/C UV HS 899-TC1MNS: TOC stock solution 1000 mg/L C Potassium hydrogenphthalate (KHP);
899-TIC1MNS: TIC stock solution 1000 mg/L C Sodium carbonate and sodium hydrogen carbon-
ate 50:50 Mixture;
899-TOCC2A: NPOC calibration kit, UV - 0.5 mg/L C, KHP, drinking water calibration;
899-W500SA: TOC preparation water, UV stabilized with H2SO4 for UV digestion TOC systems

Guaranteed Not guaranteed


All the reagents required for installation are provided by the customer.

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 7/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)
Announcement of Readiness for Installation

4 Device layout and space requirements


The basic device has been designed as a compact bench-top device. The required space depends on all components
that make up the measuring station.
Further components of the measuring station:
¡ The PC, monitor and printer may be placed on a separate side table.
¡ An acid-resistant waste container can be placed on or under the bench.
Optional modules:
¡ The AS 10e and AS 21hp autosamplers are attached to the right side wall of the basic device. Alternatively, the au-
tosamplers can be placed next to the device on the right.
¡ The AS vario, AS vario ER and EPA Sampler autosamplers must be positioned to the right of the device.
¡ The HT 1300 and the manual TIC solids module are placed to the left of the basic device.

Component Dimensions (width x height x depth) Weight


Basic device 513 x 464 x 550 mm 30 kg
AS 10e autosampler 260 x 390 x 320 mm 4.5 kg
AS 21hp autosampler 260 x 390 x 320 mm 4.5 kg
AS vario, AS vario ER autosamplers 350 x 470 x 400 mm 15 kg
EPA Sampler 500 x 560 x 400 mm 15 kg
HT 1300 solids module 510 x 470 x 550 mm 22 kg
Manual TIC solids module 300 x 470 x 550 mm 10 kg

Figure 1 Space required for multi N/C UV HS with modules

Analytik Jena GmbH Phone +49 3641 77 70 Announcement of Readiness for Installation multi N/C UV HS, multi N/C 8/8
Konrad-Zuse-Strasse 1 info@analytik-jena.com pharma UV
07745 Jena ⋅ Germany www.analytik-jena.com Version: A (03/2022)

You might also like