Professional Documents
Culture Documents
LTP100 Wman 04 B
LTP100 Wman 04 B
LTP100W
User's and Installation Guide
1. GENERAL DESCRIPTION...............................................................................................................3
2. THEORY OF OPERATION..............................................................................................................6
2.1 ANALOG DATA ACQUISITION .................................................................................................................6
2.2 MODES OF OPERATION.........................................................................................................................8
2.3 CONTROL ALGORITHMS.........................................................................................................................9
3. LTP100W CONFIGURING.............................................................................................................11
3.1 APPLICATION SOFTWARE LOADING.......................................................................................................11
3.2 GENERAL CONSIDERATIONS.................................................................................................................11
3.2.1 Entering SET-UP...................................................................................................................13
3.2.1.1 Normal startup (Warm startup)..........................................................................................13
3.2.1.2 First startup after program downloading (Cold startup)...................................................13
3.2.2 Installation and Set-up quick instructions.............................................................................14
3.2.3 DOSING type detailed menu choices....................................................................................15
3.2.4 SETUP detailed menu choices...............................................................................................17
3.3 FACTORY .........................................................................................................................................19
3.4 TEST................................................................................................................................................19
4. CALIBRATION MODE MENU......................................................................................................20
4.1 CALIBRATION VIEW AND EDIT PROCEDURE (E).....................................................................................21
4.2 LOAD CELL CALIBRATION MENU (C)..................................................................................................22
4.3 CALIBRATION SETUP (S).....................................................................................................................24
5. OPERATION’S COMMANDS........................................................................................................25
5.1 BATCH AND ANALOG WEIGHT TRANSMITTER........................................................................................26
5.1.1 Batch MANUAL Mode Menu.................................................................................................27
5.1.2 Batch AUTOMATIC Mode Menu..........................................................................................28
5.1.3 Batch Parameters Data Entry...............................................................................................30
5.1.4 Batch Recipe Data Entry.......................................................................................................31
5.2 LOSS IN WEIGHT AND BELTS ...............................................................................................................32
5.2.1 Loss-in-weight and Belts MANUAL Mode Menu..................................................................34
5.2.2 Loss-in-weight and Belts AUTOMATIC Mode Menu............................................................37
5.2.3 Loss-in-weight and Belts RECIPE Mode Menu....................................................................40
5.2.4 Manually entering commands in MANUAL and AUTOMATIC modes................................41
5.2.5 Loss-in-weight and Belt Regulation Parameters Data Entry................................................42
5.2.6 Loss-in-weight and Belt Recipe Parameters Data Entry......................................................43
6. CONNECTION TO AN OUTSIDE COMPUTER.........................................................................46
6.1 BATCH DOSING MESSAGES................................................................................................................46
6.2 LOSS-IN-WEIGHT AND BELT DOSING MESSAGES...................................................................................47
7. SPECIFICATIONS FOR RS485 LINE CONNECTION............................................................50
7.1 SPECIFICATIONS OF RS485 CONNECTION INTERFACE ...............................................................................50
7.2 HARDWARE FEATURES OF RS485 INTERFACE OF ”LTP100W”............................................................50
2
1. General Description
The LTP100W is a computer based industrial electronic unit, the primary function of
which is to provide accurate and reliable dosage of material components, both solid
and liquid, in continuous as well as batch based mixing processes. Primary industry
segments served are Chemical Plants, Soaps and Detergents, Food and Beverage,
Pharmaceutical and others.
The figure below shows a picture of the unit. Physical dimensions and other
mechanical and environmental characteristics are detailed in the Specification
chapter.
In what follows, the reader is supposed to by familiar with machinery and techniques
as well as sensors and actuators commonly found in industrial dosing processes.
When used to run batch processes, the LTP100W starts the material’s loading (by
asserting a digital output), then continuously acquires the batch weight provided by a
suitable sensor, calculates the net weight, automatically stops the material’s loading
when the recipe value is reached and unloads the batch into the mixer. Various
techniques and algorithms are used in order to compensate for several inherent
sources of errors typically found in batch machinery.
Conversely, the basic concept outlining the LTP100W unit when controlling
continuous dosing processes is to constantly acquire the weight measurement
provided by a Load Cell (or similar instrument) with accuracy high enough to reliably
calculate the material’s flow rate. This is done by digitally obtaining the derivative of
the weight curve in “Loss in Weight” systems and by multiplying the weight per unit of
length by the tape speed in belt based systems. In the latter case the belt speed can
be supplied in various forms, most typically as encoders pulse streams or speed
transmitters analog signals (0-10V, 0-20 or 4-20 mA). In either case, various digital
processing algorithms are applied to filter out unavoidable sources of mechanical
and electrical noise. The calculated flow rate is then used as feedback signal in the
built-in regulator, which, in turn, drives the material’s unloading actuator.
3
Basically, the LTP100W unit consists of four main parts: Power Supply module,
Micro-Controller Electronic Board, Alphanumeric Display and Membrane Keyboard
(see the Specification chapter for details). The Electronic board is based on the
powerful and well-rehearsed Hitachi microprocessor H8/3003 carrying both flash
memory (for program code and static information) and battery upheld RAM memory
(for dynamic information).
The other sub-modules are industrial standard devices, which do not require
description here (see specifications for details).
The built in supplied software provides input/output testing, guided installation and
setup, friendly user interface, easy data entry procedures, self-conducting sensor
calibration procedures as well as high accuracy data acquisition and automatic
process control.
• Stand-alone - when the Operator enters all process inputs and commands
directly from the built-in keyboard, monitoring the process on the built-in
screen
• Remotely controlled - when the LTP100W is connected to an outside
computer, both for data entry, process commands and process monitoring
purposes
Both cases share the same theory of operation, software functions and control
performance. The operational procedures are equivalent in power (although in
Remotely controlled mode they can be developed to any degree of friendliness,
including graphical rendering, depending on the Host Computer power and software).
Since the operation in Remotely controlled mode is, to a certain extent, application
dependent and, therefore, needs specific case-by-case description, only a brief but
complete reference of the communication messages with a Host Computer will be
4
provided in this document in a separate chapter. This will allow the reader to
understand the related concepts and possibly develop specific changes. A complete
User Interface Application for a general dosing plant, featuring both batch and
continuous processes, based on a Windows NT Client-Server structure is available
as a model case.
For the above reasons, the remaining of this document will be mainly dedicated to
the description of the Stand-alone procedures.
5
2. Theory of Operation
In this chapter we talk about the most significant functions making up the structure of
the LTP100W unit with the purpose of establishing the rationale and the meaning of
terms extensively adopted in subsequent chapters dealing with detailed Operational
Procedures.
The key function performed by the LTP100W, providing the basis for all other
available features, consists of precise data acquisition from weight sensors.
- Load Cell equipped with dedicated built in power supply and signal
conditioning electronics, usually providing either 0-10V or 4-20mA output
signal, to be connected to the precision, 24 bit resolution input port available
in the LTP100W for such signal types. This input connection covers also the
very rare cases when weight is measured by means of traditional mechanical
scales.
- Load Cell with no dedicated, built in electronics. In this case the LTP100W
provides a dedicated and fully de-coupled precision (0-5V) power supply to
feed the load cell resistors bridge and an advanced, state of the art integrated
component to sense the actual bridge voltage by means of current free wires
and directly acquire the bridge output millivolts with a 24 bit resolution.
In both cases, the acquisition rate is quartz controlled and can be selected at
installation time in the range 15.38 to 200 Hz. The default rate, which is suitable for
the vast majority of cases, is 25 Hz, corresponding to the acquisition of one sample
every 40 ms.
In addition, in both cases, the acquisition ICs provide built-in digital filters having the
purpose to get rid of possible high frequency noise and residues of any power AC
ripple.
The subsequent step in the data processing sequence consists of scaling the raw
data to standard units of measurement, most typically Kilograms, by means of a
conversion curve stored in the LTP100W. This is implemented as a look-up table
made up of at least 2 and no more than 1000 conversion points, using 8 bytes
floating point format and arithmetic. This ensures that truncation errors do not
appreciably decrease the overall accuracy in the scaling process. The Application
Engineer has the choice of either entering manually the conversion curve data (on
the basis of theoretical grounds) or to perform automatic calibration. In this case, he
will be required by the LTP100W itself to apply standard objects of precisely known
weight to the Load Cell and enter the corresponding numeric value at the keyboard.
6
The LTP100W will then accurately read the Load Cell generated input voltage, thus
obtaining corresponding points of the conversion curve.
In order to obtain clean and noise free values the system contains a six poles
numeric linear filter, which provides both the instantaneous weight and the material
flow rate. At installation, the Application Engineer is required to select the filter type
(either Butterworth, Bessel or other of his choice) as well as the corresponding cut-off
frequency, taking into account that the filter should get rid of all residual noise without
appreciably degrading the steady-state rate value. To do so, some proficiency in
Signal Processing techniques would be required but, at any rate, it has been found
both theoretically and experimentally that a Bessel type filter at 0.1 Hz performs
satisfactorily in the majority of practical cases and is provided by the LTP100W as
default.
The output rate from the filter is, obviously, the same as the acquisition rate (default
25 Hz) and the corresponding values are refreshed on the screen once a second.
At the same time, the LTP100 calculates in real time the linear regression of a
previously selected number of weight samples, using a minimum squared error
technique (the Application Engineer can select such number of samples or accept
the LTP100W default value of 50).
As the result, the values of the weight and of the flow rate are calculated concurrently
in two ways. The values coming out from the linear filter tend to be cleaner but
slower than those coming out from the mean square calculation are; therefore, they
are used internally for different purposes, including checking each other.
In Belt Equipped dosing systems, the belt speed needs to be measured, along with
the weight per unit of length. While the latter is measured following the technique
described in the previous paragraphs, the speed is measured in a different manner.
In the most significant cases, the belt speed is measured by an encoder. This
provides a pulse sequence the frequency of which is proportional to the belt speed.
The LTP100W unit obtains a new speed value every 0.2 seconds (by default). When
reaching the end of such a sampling interval the LTP100W waits for the next encoder
pulse and then immediately divides the number of received pulses by the time
spanned, measured in nanoseconds by the system clock. This way, since the time
accuracy is provided by a crystal quartz (thus typically better than ten parts over one
million), the overall speed accuracy depends primarily on the encoder quality and its
installation fixtures, without appreciable degradation caused by the acquisition
process.
7
2.2 Modes of Operation
Although the most basic functions are always active once the LTP100W is running,
some others are logically grouped together and can be started by the Operator if the
Unit has been previously placed in a proper operational mode.
When the LTP100W is turned on, it performs several hardware tests, checks the
integrity of some basic parameters and initializes its sub-units and components, then
places itself into the READY mode and displays the application’s main menu.
Although the main menu is slightly different for the three process types (batch, loss-
in-weight and belt) the menu concept is consistently the same. The Operator has to
choose among three possible actions:
In the first case, the LTP100W moves to the CALIBRATION mode, in which the
Operator can view in sequence all points of the weight calibration curve actually
stored in memory. If he determines that such curve is not suitable for the case, he
can generate a new curve either loading a previously defined default curve or by
performing a full system re-calibration using standard reference weights. The
LTP100W provides a full systematic guidance to do so. Upon leaving the
CALIBRATION mode the system returns to the READY mode.
In the second case, the Operator has to choose among three operational modes,
MANUAL, AUTOMATIC and RECIPE. They will be described shortly, while detailed
procedures and commands will be given in the subsequent chapter 4 - Operation’s
Commands.
In the third case, the LTP100W enters the SETUP/TEST mode whose procedures
and commands will be described in chapter 3 - Installation, later in this document.
8
In RECIPE mode the LTP100W performs the same functions of the AUTOMATIC
mode, the only difference consisting in the fact that, in RECIPE mode, the regulator’s
set point is taken from the recipe table rather than from the Operator’s command.
Obviously, he can override the recipe value at any time by placing back the system to
AUTOMATIC mode.
The first one is application dependent, so its functions are substantially spread
throughout a number of other functions provided for discrete batch processes. For
this reason we do not describe them here, but rather in a subsequent chapter, where
a detailed list of batch control parameters is given.
The second one is configured as a number of black boxes providing classical closed
loop control. Before using such boxes, the User must define the Process Variable (be
it weight, rate, speed, pressure, temperature, level or whatever else) and, obviously a
Set point of the same nature. The LTP100W provides three default regulators,
although one should take in mind that, given the black box nature of such
components, other regulators can be added or substituted easily any time this is
considered appropriate.
The Classical PID needs not to be described here since it is well known in technical
literature and widely applied in any kind of classical control devices. We need only to
mention here that the implemented PID equations assure smooth transitions, control
of integral windup and calculation accuracy, the latter being assured by the use of 8
bytes floating point arithmetic and advanced numeric techniques for derivation and
integration.
The PID+S regulator performs like the classical PID, but in case of set-point change,
the regulator’s output is also immediately changed by a quantity proportional to the
set-point change itself, thus providing a much quicker response to set-point changes
than the classical PID. In such occurrences, the regulator output is first kept
constant for a certain time in order to allow the plant to respond, then the closed loop
control is resumed. When this occurs, the integral value is automatically adjusted to
provide smooth behavior. Given these characteristics, the PID+S regulator is
particularly suitable for such Loss-In-Weight processes in which the weight filter
cutoff frequency has to be chosen exceptionally low.
The PEICS regulator maintains its output constant for the most part of the calculation
cycle, then allows it to grow for a certain time at a rate proportional to the difference
between the set-point and the average of the process variable during the previous
9
cycle. This technique is known to be particularly suitable for processes in which a
substantial delay exists between the actuator action and the feedback measuring
point, as in the case of very long belts.
10
3. LTP100W CONFIGURING
Once the LTP100W unit has been mechanically installed and wired to the relevant
sensors and control signals (this is an application dependant task not outlined in this
section, see the LTP100W hardware manual for details on "connector layouts" and
wiring specifications), the Field Engineer needs to configure the LTP100W unit
according to the application's needs. This is done by running a few simple
procedures built-in the unit's software.
Note. The LTP100W units are normally supplied already loaded with the latest
software version. Should one accidentally become corrupted, or otherwise being
desired (in due time) to load a newer version, the Field Engineer needs to download
it from a host Personal Computer. Please contact Logosystem to obtain the
necessary software files, should they have not been included in the original supply.
Under normal conditions the presently loaded version number will appear on the
screen as the second "screen layout" after turning on the unit (the first one shows a
"factory" code that you can normally disregard). Look for a string resembling the
following:
*** Version P-030614/1 ***
(Actual digits may vary).
If you do not get this, you definitely need to re-load the software.
To load (or re-load) the software, perform the following easy steps.
- Install the PC "LTP100W loader application" from the supplied Compact Disk
on any Windows based PC (if not already installed).
- Turn off the LTP100W unit.
- Connect any PC's serial port (most likely COM1) to the LTP100W's RS232
port, using the supplied cable.
- Run the "LTP100W loader application" on the PC, optionally enter the
selected COM port and the desired transmission baud rate (COM1 at
38.4Kbaud by default), then open the application file (named DOSING.BIN by
default) and press button "Download".
- Turn on the LTP100W unit. After a few seconds the transmission will begin.
Watch the downloading progress and verify than no errors are reported.
Upon completion of the "software loading", the program will start automatically by
itself. Please check that the Version Number matches the one just loaded.
11
The LTP100W program starts automatically by itself as soon as power is switched-
on. For reasons that will be made clear later, the LTP100W can start in two ways:
- Cold Start.
The program jumps directly to the Set-up section, without attempting to run
with possibly corrupted (or blank) essential parameters. A minimum set of
parameters will be supplied by default, while others will have to be entered by
the Field Engineer by means of the Set-up procedure. All counters and
"totalizers" will be reset to zero.
- Warm Start.
The program runs normally, using the application's parameters already in
memory.
After turn-on, a "never set-up" (or corrupted) LTP100W unit will detect a "blank data"
condition of its memory and will automatically perform a Cold Start, inviting you to
perform Set-up.
You can force the LTP100W unit to perform a Cold Start by pressing the DEL key
within two seconds after the version number has been displayed.
If a Cold Start is being performed, the LTP100W will inform you and wait for your
confirmation to proceed (remember that all counters will be reset, hence do not
perform unnecessary Cold Starts).
If a Cold Start is not necessary, execute a Warm Start and let the LTP100W perform
its duties. Even in this case the Field Engineer might want to run Set-up again. This
can be done as described later.
Any LTP100W unit can be installed and set-up to drive any of the above mentioned
dosing systems, without the need of adjusting hardware settings or re-loading the
application software.
Along the SET-UP operations, the program will automatically load certain default
parameters on the basis of the selected dosing type. However, any default parameter
can be overridden by the Installation Engineer to accommodate specific plant
requirements.
The following picture shows the layout of the LTP100W’s alphanumeric display and
keyboard.
12
3.2.1 Entering SET-UP
Once turned on, the LTP100W performs a few hardware checks, and then displays
the following (or similar) initial screen layout.
13
Position the cursor
then <ENTER>
First select the DOSING type. Depending on the dosing type you choose you will
have to enter different sets of parameters.
If you choose a DOSING type different from the previously selected one (or if you
choose a DOSING type for the first time in a "blank" LTP100W), the Set-up
procedure proposes a "default" set of parameters, which past experience indicates
are normally suitable for the majority of cases. You can optionally accept such set of
parameters (by pressing <ESC>) or override them by entering appropriate numbers
in the fields. Once this is done the screen will indicate which DOSING type was
selected and whether the entered set of installation parameters is valid (that is,
corresponds to certain built-in criteria). If it is not, repeat the DOSING type selection
and carefully revise each field.
The Belt nominal speed and the corresponding Encoder Frequency are not defaulted
as "valid". You always have to override such fields in order to properly install Belts.
Occasionally you can satisfactorily run your dosing application even if the installation
set of parameters is shown to be invalid. The LTP100W gives you only a warning.
Conversely, if you are running under remote PC or PLC control, you have to choose
SETUP in the menu (next field to the right of the field "DOSING") and enter the
following:
14
This completes the installation / Setup process. Press <ESC> a certain number of
times, until the LTP100W restarts.
Although the DOSING type menu branch is "self explanatory" and straightforward,
we provide here a semi-graphical reference guide to properly navigate in the
DOSING type selection menu. Values appearing between brackets in the following
picture are the "default" ones. Values appearing, without brackets, in the screen
layouts are the ones actually present in memory.
To make a selection among menu choices, use the arrow keys to point to one of the
fields, then press <ENTER>.
To enter numeric parameters, use the numeric keys, then:
- <ENTER> to fix the parameter's value into memory
- <ESC> to scroll to the next parameter
- <DEL> to clear the parameter's entry fie
15
DOSING
NORMAL LEVEL
LIQUID SOLID CONTROL LIQUID SOLID
ONE TWO
MOTOR MOTORS
Inlet Valve Open Timeout [sec]:[10] Motor 1 Startup Timeout [sec]:[4] Nominal Belt speed [m/min]:[invalid] Charge Valve Open Timeout [sec]:[20]
Inlet Valve Close Timeout [sec]:[10] Motor 1 Stop Timeout [sec]:[4] Encoder freq. at that speed [Hz]:[invalid] Charge Valve Close Timeout [sec]:[20]
Outlet Valve Open Timeout [sec]:[15] Motor 2 Startup Timeout [sec]:[4] Motor Startup Timeout [sec]:[4] Motor Startup Timeout [sec]:[4]
Outlet Valve Close Timeout [sec]:[15] Motor 2 Stop Timeout [sec]:[4] Motor Stop Timeout [sec]:[4] Motor Stop Timeout [sec]:[4]
Outlet Valve Open Timeout [sec]:[15]
Outlet Valve Close Timeout [sec]:[15]
3.2.4 SETUP detailed menu choices
After selecting the SETUP menu entry, you get the message: "Starting setup..." and,
after two seconds, a "screen layout" resembling the following:
The first line shows the five menu entries available for selection. The one being
actually pointed is marked by the cursor position (in the picture above BR1 is
supposed). You scroll among the five entries using the "arrow" keys. You make a
selection by pressing the <Enter> key.
The second line provides a description of the action performed by the pointed entry.
If <ENTER> is pressed:
Password : .....
If password is correct:
Address = .....
(available choices)
18
SETUP menu item 5: PLC
If you select "Scaled to PLC analog", enter the PLC scaling factor.
3.3 Factory
The selection of the third item "FACTORY" in the menu is password protected and
reserved for factory use. Do not attempt to enter such procedure.
3.4 Test
19
The selection of the fourth item "TEST" in the menu is mostly dedicated to Factory
personnel and to User's Field Engineers who have received advanced training on the
LTP100W hardware. A brief description of the available "hardware testing"
procedures is provided as a summary reference. If you are not properly trained, or
unsure about such procedures, please do not attempt to use them. Safety
precautions need to be enforced when testing. Severe damage to the plant
equipment could result from improper use of such testing procedures.
Upon selecting the TEST menu-item the message "Starting LTP-100 test..." will
appear on screen followed, after two seconds, by the following screen layout:
RX232
This is a testing program to try out the programming port serial line.
RX485
This is a testing program to try out the Host Computer/PLC communication port.
Keyb
This is a testing program to try out the keyboard.
Out
This is a testing program to try out and individually switch on and off (periodically) the
output relays.
Inp
This is a testing program to try out the digital inputs.
(NOTE: the inputs are displayed using negative logic notation: 0 = input asserted, 1 =
input not asserted)
20
The Calibration procedures are common to the three types of dosing processes.
Hence they are described here only once, although they can be reached from the
menus of each of them.
Once the LTP100W has reached CALIBRATION mode (from the initial menu) it
shows Figure C1.
Figure C1
In Figure C1 the first line shows that the menu pertains to the CALIBRATION
procedures. The following keypad commands are allowed:
ESCape, E, C, S
ESCape return to the initial menu, stop the charging device and close the
discharging valve
E Go to the calibration view and edit procedure
C Go to the Load Cell calibration menu
S Go to the Calibration Setup menu
This procedure allows the Operator to view and, optionally, modify the curve used to
convert raw acquisition data (in the range 1 –16777216) into Kg (in the proper
range).
Figure C2
You can either solidly translate the whole curve up or down by a certain offset (by
pressing <Shift>O) or translate each individual point coordinate up, down, left or right
(by pressing <Shift>P).
Caution should be taken when changing the calibration points to any extent. The
accuracy of the whole weighting process may be impaired. If you did not receive
proper training, or you feel unsure about the "theory of operation" (described above),
avoid proceeding further in this procedure.
21
Both procedures are straightforward and do not require detailed explanation. Just
follow the screen instructions.
If you enter a new curve offset, the LTP100W will acknowledge your change with the
following (or slightly similar) messages:
Offset requested
Tables updated
View and Modify in 4 seconds
In four seconds the screen will show in turn the numeric values of each point of the
curve. You can limit yourself to inspect them or make further changes.
Figure C3
The menu in Figure C3 allows the Operator to either calibrate the Load Cell and the
acquisition system using pre-defined default values by entering the keypad “D”, or to
perform an experimental calibration session using standard weights by entering the
keypad “S”, or to return to the previous menu by pressing the keypad ESC.
Use the first option (Default) to quickly enter a default curve. This allows you to check
out the system and verify if the "load cell" is properly connected. This case is
straightforward and does need any further explanation. Just enter the number of
points (min 2, max 20) making up the curve.
Use the second option to precisely calibrate the weight acquisition system using
standard weights. The following layout appears on screen:
22
Load Cell Calibration (standard weights)
- Point 02 -
Enter net Weight[Kg]:.....
Negative value to abort, ESC to exit...
Figure C4
The Operator has to add the standard calibration weight (usually provided by the
mechanical equipment manufacturer) to the Load Cells, then enter the corresponding
value (in Kg), then press ENTER.
The LTP100W will begin to define a new calibration curve point (in the instance of
picture C4 will be point 02), following three steps and keeping the Operator informed
by appropriate messages:
Once a point has been calibrated, the Operator can either end the procedure and
permanently store the new curve by pressing ESC (as long as at least two points
have been defined) or abort the calibration procedure entirely by entering any
negative number. In this case (and also in case only one point has been defined) any
point previously defined in the calibration session will be disregarded and the pre-
existing calibration curve will remain unchanged. The following two pictures show the
two possible outcomes.
Figure C5 Figure C6
23
4.3 Calibration Setup (S)
This menu allows the Operator to view and, optionally modify some calibration
constants (described below). Such constants are given a default value at each Cold
Startup and any time the Process type is changed by means of the Setup menu
option (for instance a LTP100W unit, previously set-up for Belts is re-configured for
Loss-in-weight or Batch). The default values have been chosen to fit the vast majority
of cases and should be changed only in rare occasions (by trained personnel).
During the Calibration process the cell load must be kept constant for a time long
enough to allow the acquisition stream to reach steady conditions, so that its average
value provides an accurate measure of the real cell load. Occasionally it may happen
that, due to excessive mechanical or electrical noise, the load cell acquired values
show some comparatively strong oscillation around the mean value, never reaching
a satisfactory steady state, or a sudden and normally unwanted disturbance removes
the acquisition from steady conditions. Although such events would normally be a
sign of defective components or installation problems and, therefore should be
evaluated and possibly fixed, the System Engineer could determine that they are
acceptable, given the circumstances. In such cases it may be necessary to modify
one or more of the constants the LTP100W uses to evaluate the reaching of the
steady conditions. Such constants are:
As in other similar editing procedures, use ESC to scroll down, passing from the
definition of one constant to the next. The present value of the constant being
considered is shown in line 2. Enter a new value in line 3 then press the "ENTER"
key. The entered value will replace the old one in line 2. Repeat as many times as
desired.
24
The following four figures show an instance of the data displayed on screen to
complete the procedure (default values shown).
Figure C7
Figure C8
Figure C9
Figure C10
The entered values (or the default ones, if not changed) will remain in effect even
after the unit is turned off and on, until they are changed again.
5. Operation’s Commands
In this chapter, we describe the menu commands that are available once the
LTP100W is set-up and running. Most of the commands are self-explanatory; hence
they do not need extensive description. In all menus the following rules apply:
ESC will always terminate the procedure and return one step up in the menu
hierarchy
ENTER will be required to accept numeric data entry
DEL will blank the data entry field and allow entering new data
CAPITAL any time a menu command contains one (and only one) capital letter,
pressing that letter will activate that command
25
Menus for the three types of processes are similar in concept but differ in specific
commands. Batch and Loss-in-weight processes have different sets of commands.
Conversely, Loss-in-weight and Belt processes share the same commands structure.
When the LTP100W unit is set-up as a Batch controller, it will also provide the weight
measured by the load cell as a 4-20 mA analog output, as in conventional weight
transmitters. The end-of-scale has to be selected by the Operator in the Parameter
procedure, while the analog output quantization accuracy is fixed to 1 part over
65535 (16 bit).
Initial Menu
Figure B-I1
The first line indicates that the menu pertains to the Batch Dosing process. The
Operator can now press anyone of the key characters shown “Capitalized” in lines 2
and 3. In this particular menu the ESC key has no effect, being the first menu level.
26
5.1.1 Batch MANUAL Mode Menu
Once the LTP100W has reached MANUAL mode (from the initial menu or from the
AUTOMATIC menu) it shows either Figure B-M1 or B-M2, depending on the
circumstances.
Shows menu
Figure B-M1
Shows data
Figure B-M2
In both pictures, the first line shows that the menu pertains to the Batch Dosing
process in MANUAL mode. Although they appear different, both allow the same
following commands (some not explicitly shown in the menu itself, being valid in all
operational menus, therefore assumed to be known by trained Operators):
ESCape return to the initial menu, stop the charging device and close the
discharging valve
+ Toggle line 2 and 3 to display menu items (as in Figure B-M1) or
process data (as in picture B-M2)
A go to the AUTOMATIC menu
P View or edit batch parameters (described later)
R View or edit batch recipe parameters (described later)
F1 Close the loading valve (for liquids) or stop the loading motor (for
solids)
F2 Open the loading valve (for liquids) or start the loading motor (for
solids)
F3 Inactive (for liquids) or start the loading motor at high speed (when
available for solids)
F4 Reset the loading valve (for liquids) or reset the loading motor (for
solids)
F5 Close the unloading valve (both liquids and solids)
27
F6 Open the unloading valve (both liquids and solids)
F7 Inactive
F8 Reset the unloading valve (both liquids and solids)
I Stop the second loading motor (when available for solids)
J Start the second loading motor (when available for solids)
K Stop the second loading motor at high speed (when available for
solids)
L Reset the second loading motor (when available for solids)
When Figure B-M2 is being displayed (as mentioned, use the “+” keypad to switch
from Figure B-M1 to B-M2 and back) the following information is refreshed twice a
second (from left to right):
Once the LTP100W has reached AUTOMATIC mode (from the initial menu or from
the MANUAL menu) it shows either Figure B-A1 or B-A2, depending on the
circumstances.
Shows menu
Figure B-A1
Shows data
Figure B-A2
In both pictures the first line shows that the menu pertains to the Batch Dosing
process in AUTOMATIC mode and, although they appear different, both allow the
same following commands: (some not mentioned in the menu itself, being valid in all
operational menus, therefore assumed to be known by trained Operators).
28
ESCape, + , M. P, R, S, U, I, J, K, L, F1, F2, F3, F4, F5, F6, F7, F8
ESCape return to the initial menu, stop the charging device and close the
discharging valve
+ Toggle line 2 and 3 to display menu items (as in Figure B-A1) or
process data (as in Figure B-A2)
M go to the MANUAL menu
P View or edit batch parameters (described later)
R View or edit batch recipe parameters (described later)
S Start Sequence. With this command, the Operator instructs the
LTP100W that he wants to start batch production. The Ltp100W
performs several checks on the plant status. If the plant is not ready
an error message is issued, else the cycle progressive number and
the step timer are cleared, the total production counter (totalizer) is
zeroed, the unit reaches a “ready” state and the following message is
displayed on the screen: AUTO MODE/NO ALARMS
U Start Pulse.
This command is provided for installation and testing purposes only.
Use with discretion. Do not use in production cycles.
With this command, the Operator directs the unit to perform a full
dosing cycle, composed of sequential steps, as required by the
particular process. Once the cycle is completed, if a fatal error
occurred an error message is issued and a FAULT digital output is
asserted, else a cycle condition code is issued and a DONE digital
output is asserted.
F1 Close the loading valve (for liquids) or stop the loading motor (for
solids)
F2 Open the loading valve (for liquids) or start the loading motor (for
solids)
F3 Inactive (for liquids) or start the loading motor at high speed (when
available for solids)
F4 Reset the loading valve (for liquids) or reset the loading motor (for
solids)
F5 Close the unloading valve (both liquids and solids)
F6 Open the unloading valve (both liquids and solids)
F7 Inactive
F8 Reset the unloading valve (both liquids and solids)
I Stop the second loading motor (when available for solids)
J Start the second loading motor (when available for solids)
K Stop the second loading motor at high speed (when available for
solids)
L Reset the second loading motor (when available for solids)
When Figure B-A2 is being displayed (as mentioned, use the “+” keypad to switch
from picture B-A1 to B-A2 and back) the following information is refreshed twice a
second (from left to right):
29
• Progressive cycle number since start of sequence (BATCH), cycle step in
progress (for instance CHARGING), time elapsed since step beginning (in
seconds), state of the loading and unloading devices (in line 2)
• Present load cell weight, last evaluated tare, last evaluated “fall down”, last
evaluated totalizer (in line 3)
• Last occurred alarm (in line 4)
You enter this procedure by pressing keypad “P” from either the MANUAL or the
AUTOMATIC menus.
"New" is safer
Figure B-P1
Type “N” if, presently, the LTP100W does not contain any valid batch parameters or
if you want to disregard old values entirely.
Type “O” if presently the LTP100W does contain valid batch parameters and you like
to see the old values before changing them.
Note. If you select “O” while the LTP100W does not contain valid batch parameters,
the procedure is likely to crash and the system to stop abruptly. Should this occur,
turn the LTP100W off and on again and select “N” this time.
The main Batch Parameters data entry Procedure looks like Figure B-P2.
Figure B-P2
30
Several batch-dosing parameters are defined in the LTP100W. The present value of
the parameter on display will appear in line 2, while a series of dots will appear in line
3. As shown, use keypad ESC to scroll parameters forward leaving them unchanged,
or enter a value in the dotted line followed by ENTER. As long as the entered value
is acceptable, it will replace the old one in line 2, while the cursor will return at the
beginning of the dotted line, ready to accept another value for the same parameter
should you decide to do so. Else, press ESC to definitely accept the entered value
and move to the next parameter. In case of typing error during data entry, use
keypad DEL to clear the input field. If the keyboard “MENU” is pressed, the
procedure is terminated and the program execution returns to the application menu.
The table of Batch Parameters is made up as follows:
You enter this procedure by pressing keypad “R” from either the MANUAL or the
AUTOMATIC menus.
"New" is safer...
Figure B-R1
Type “N” if, presently, the LTP100W does not contain any valid batch recipe
parameters or if you want to disregard old values entirely.
31
Type “O” if, presently, the LTP100W does contain valid batch recipe parameters and
you like to see the old values before changing them.
Note. If you select “O” while the LTP100W does not contain valid batch recipe
parameters, the procedure is likely to crash and the system to stop abruptly. Should
this occur, turn the LTP100W off and on again and select “N” this time.
The main Batch Recipe Parameters data entry Procedure looks like Figure B-R2.
Figure B-R2
A number of batch dosing recipe parameters are defined in the LTP100W. The
present value of the recipe parameter on display will appear in line 2, while a series
of dots will appear on line 3. As shown, use keypad ESC to scroll parameters forward
leaving them unchanged, or enter a value in the dotted line followed by ENTER. As
long as the entered value is acceptable, it will replace the old one in line 2, while the
cursor will return at the beginning of the dotted line, ready to accept another value for
the same recipe parameter should you decide to do so. Else, press ESC to definitely
accept the entered value and move to the next recipe parameter. In case of typing
error during data entry, use keypad DEL to clear the input field. If the keyboard
“MENU” is pressed, the procedure is terminated and the program execution returns
to the application menu. The table of Batch Recipe Parameters is made up as
follows:
32
wide extent, especially as far as the operational procedures are concerned. Hence,
the procedures described in this chapter are applicable in general to both systems.
The few exceptions will be dutifully noted.
The Belt systems will always deal with solid powders, while the Loss-in-weight ones
will sometimes handle liquid components. This will be taken into account by the
control software, but again the basic concepts will remain unaltered. The first menu
to appear on entry (Figure BLIW-I1) will show the type of process for which the
LTP100W is actually setup.
Initial Menu
Figure BLIW-I1
The first line indicates that the menu pertains to the Loss-in-weight process. The
Operator can now press anyone of the key characters shown “Capitalized” in lines 2
and 3. In this particular menu the ESC key has no effect, being the first menu level.
Here is the list of possible commands:
33
5.2.1 Loss-in-weight and Belts MANUAL Mode Menu
Once the LTP100W has reached MANUAL mode (from the initial menu or from the
AUTOMATIC menu) it shows either Figure LIWB-M1 or LIWB-M2, depending on the
circumstances.
Shows menu
Figure LIWB-M1
Loss-in-Weight Shows data
Figure LIWB-M2
Shows menu
Figure LIWB-M1
Belt Shows data
Figure LIWB-M2
34
In all four pictures, the first line, while indicating that the menu pertains to the
MANUAL mode, shows the actual manually set rate of discharge in percentage.
Although they appear different, all allow the same set of commands (some not
mentioned in the menu itself, being valid in all operational menus, therefore assumed
to be known by trained Operators):
ESCape return to the initial menu, stop the discharging motor and close the
recharging valve (only in case of LIWs)
+ Toggle line 2 and 3 to display menu items (as in picture LIVB-M1) or
process data (as in picture LIWB-M2)
A go to the AUTOMATIC menu
R go to the AUTOMATIC mode, but take the flow rate set-point from the stored
recipe
O accept a new value for the discharging motor speed
P View or edit regulation parameters (described later)
C View or edit recipe parameters (described later)
F1 Close the recharging valve (for LIWs) or stop the belt motor (for Belts)
F2 Open the recharging valve (for LIWs) or start the belt motor (for Belts)
F3 Inactive (both for LIWs and Belts)
F4 Reset the recharging valve (for LIWs) or reset the belt motor (for
Belts)
F5 Stop the LIW discharging motor. Inactive for Belts
F6 Start the LIW discharging motor. Inactive for Belts
F7 Inactive in all cases
F8 Reset the LIW discharging motor. Inactive for Belts
When picture LIWB-M2 is being displayed (as mentioned, use the “+” keypad to
switch from picture LIWB-M1 to LIWB-M2 and back) the following information is
refreshed once every two seconds (from left to right):
- line 2 Actual load cell weight in Kg, actual material flow rate in Kg/hour
- line 3 Totalizer value, whether the totalizer is enabled (ena) or disabled
(dis), whether the LIW is in the recharging state (loa) or in the normal
discharging state (unl), the recharging valve state (CLO=closed,
MOV=moving, OPE=open, LOC=selector in local, ALA=alarm), the
discharging motor state (STP=stop, STA=starting, RUN=running,
LOC=selector in local, ALA=alarm)
- line 4 Last occurred alarm. Possibly displayed alarms are:
35
5.2.1.2 Belt systems only
When picture LIWB-M2 is being displayed (as mentioned, use the “+” keypad to
switch from picture LIWB-M1 to LIWB-M2 and back) the following information is
refreshed once every two seconds (from left to right):
- line 2 Actual load cell weight in Kg, actual belt speed in meters/minute,
actual material flow rate in Kg/hour
- line 3 Totalizer value, whether the totalizer is enabled (ena) or disabled
(dis), the belt motor state (STP=stop, STA=starting, RUN=running,
LOC=selector in local, ALA=alarm)
- line 4 Last occurred alarm. Possibly displayed alarms are:
36
5.2.2 Loss-in-weight and Belts AUTOMATIC Mode Menu
Once the LTP100W has reached AUTOMATIC mode (from the initial menu or from
the MANUAL menu) it shows either picture LIWB-A1 or LIWB-A2, depending on the
circumstances.
Shows menu
Figure LIWB-A1
Loss-in-Weight Shows data
Figure LIWB-A2
Shows menu
Figure LIWB-A1
Belt Shows data
Figure LIWB-A2
In all four pictures, the first line, while indicating that the menu pertains to the
AUTOMATIC mode, shows the actual rate of discharge set-point in Kg. Although they
appear different, all allow the same set of commands (some not mentioned in the
37
menu itself, being valid in all operational menus, therefore assumed to be known by
trained Operators):
ESCape return to the initial menu, stop the discharging motor and close the
recharging valve (only in case of LIWs)
+ Toggle line 2 and 3 to display menu items (as in picture LIVB-M1) or
process data (as in picture LIWB-M2)
M go to the MANUAL menu
R Remain in AUTOMATIC mode, but take the flow rate set-point from the stored
recipe
F accept a new value for the discharging rate set-point
P View or edit regulation parameters (described later)
C View or edit recipe parameters (described later)
F1 Close the recharging valve (for LIWs) or stop the belt motor (for Belts)
F2 Open the recharging valve (for LIWs) or start the belt motor (for Belts)
F3 Inactive (both for LIWs and Belts)
F4 Reset the recharging valve (for LIWs) or reset the belt motor (for
Belts)
Reset the regulator’s “out of tolerance” error (if present)
F5 Stop the LIW discharging motor. Inactive for Belts
F6 Start the LIW discharging motor. Inactive for Belts
F7 Inactive in all cases
F8 Reset the LIW discharging motor. Inactive for Belts
Reset the regulator’s “out of tolerance” error (if present)
When picture LIWB-M2 is being displayed (as mentioned, use the “+” keypad to
switch from picture LIWB-M1 to LIWB-M2 and back) the following information is
refreshed once every two seconds (from left to right):
- line 2 Actual load cell weight in Kg, actual material flow rate in Kg/hour
- line 3 Totalizer value, whether the totalizer is enabled (ena) or disabled
(dis), whether the LIW is in the recharging state (loa) or in the normal
discharging state (unl), the recharging valve state (CLO=closed,
MOV=moving, OPE=open, LOC=selector in local, ALA=alarm), the
discharging motor state (STP=stop, STA=starting, RUN=running,
LOC=selector in local, ALA=alarm), the value of the motor control
signal in percentage (OUT=xx.x)
- line 4 Last occurred alarm. Possibly displayed alarms are:
38
DOSER IN TEST MODE
LOAD CELL FAULT
AUTO MODE/NO ALARMS
When picture LIWB-M2 is being displayed (as mentioned, use the “+” keypad to
switch from picture LIWB-M1 to LIWB-M2 and back) the following information is
refreshed once every two seconds (from left to right):
- line 2 Actual load cell weight in Kg, actual belt speed in meters/minute,
actual material flow rate in Kg/hour
- line 3 Totalizer value, whether the totalizer is enabled (ena) or disabled
(dis), the belt motor state (STP=stop, STA=starting, RUN=running,
LOC=selector in local, ALA=alarm), the value of the motor control
signal in percentage (OUT=xx.x)
39
5.2.3 Loss-in-weight and Belts RECIPE Mode Menu
Once the LTP100W has reached RECIPE mode (from the initial menu or from the
MANUAL menu, or from the AUTOMATIC menu) it shows either Figure LIWB-R1 or
LIWB-R2, depending on the circumstances.
Shows menu
Figure LIWB-R1
Loss-in-Weight Shows data
Figure LIWB-R2
Shows menu
Figure LIWB-R1
Belt Shows data
Figure LIWB-R2
40
This set of pictures show that the recipe mode is active (in line 1).
The RECIPE mode is actually undistinguishable from the AUTOMATIC one, except
that in the former the regulator set-point is taken from the recipe, while in the latter
from the Operator’s manual entry.
Pressing the keypad “A”, or manually entering a new rate set-point (using keypad
“F”), will force the LTP100W to go back to the AUTOMATIC mode.
All other RECIPE mode commands and procedures are the same as the
AUTOMATIC mode ones. Please refer to the previous chapter 4.2.2 for details.
As shown above, the Operator can control the output to the motor controller (usually
an inverter) while in MANUAL mode by means of keypad “O” and the rate regulator’s
set-point while in AUTOMATIC mode by means of keypad “F”. Both procedures
operate in a similar way, so they can be described together. In the former case, the
entered number must be between 0 and 100, while in the latter between 0 and the
maximum rate (defined as a parameter). Enter an appropriate number in the
indicated field, the press Enter. Once a new value has been accepted it replaces
immediately the old one, while the screen remains ready to accept another value,
continuing this way until the Operator presses the keypad ESC in order to leave the
data entry procedure. Pressing the keypad MENU instead of ESC will terminate the
data entry procedure immediately.
Note. While either procedure “O” or procedure “F” is active, the LTP100W
regulation functions are working normally, but the screen refreshing is
disabled. Because of this, the Operator should leave either procedure
immediately once changing values is no more immediately needed.
Figure LIWB-A3
Suppose the present regulator’s set-point is 31.5 Kg/hour. The Operator can enter a
new value on the dotted line, then press Enter, then ESC. The entered value will be
displayed in place of 31.5 . Screen refreshing is inactive while this menu is being
displayed.
41
5.2.5 Loss-in-weight and Belt Regulation Parameters Data Entry
You enter this procedure by pressing keypad “P” from either the MANUAL or the
AUTOMATIC menu.
The Loss-in-weight regulator parameters differ slightly from the Belt ones, however
the procedure concepts remain the same in both cases.
Figure LIWB-P1
Type “N” if, presently, the LTP100W does not contain any valid regulator parameters
or if you want to disregard old values entirely.
Type “O” if presently the LTP100W does contain valid regulator parameters and you
like to see the old values before changing them.
Note. If you select “O” while the LTP100W does not contain valid regulator
parameters, the procedure is likely to crash and the system to stop abruptly. Should
this occur, turn the LTP100W off and on again and select “N” this time.
The main Regulator Parameters data entry Procedures look like Figure LIWB-P2.
Figure LIWB-P2
42
A number of regulator parameters are defined in the LTP100W. They can be
accessed sequentially. Once a particular parameter is considered, its present value
will appear in line 2, while a series of dots will appear on line 3. As indicated, use
keypad ESC to scroll parameters forward leaving them unchanged, or enter a value
in the dotted line followed by ENTER. As long as the entered value is acceptable, it
will replace the old one in line 2, while the cursor will return at the beginning of the
dotted line, ready to accept another value for the same parameter should you decide
to do so. Else, press ESC to definitely accept the entered value and move to the next
parameter. In case of typing error during data entry, use keypad DEL to clear the
input field. If the keyboard “MENU” is pressed, the procedure is terminated and the
program execution returns to the application menu. The table of Regulator
Parameters is made up as follows:
You enter this procedure by pressing keypad “C” from either the MANUAL or the
AUTOMATIC menu.
Figure LIWB-C1
43
Type “N” if, presently, the LTP100W does not contain any valid recipe or if you want
to disregard old values entirely.
Type “O” if presently the LTP100W does contain valid recipe parameters and you like
to see the old values before changing them.
Note. If you select “O” while the LTP100W does not contain valid recipe parameters,
the procedure is likely to crash and the system to stop abruptly. Should this occur,
turn the LTP100W off and on again and select “N” this time.
The main Recipe Parameters data entry Procedures look as shown in picture LIWB-
C2.
Figure LIWB-C2
A number of recipe parameters are defined in the LTP100W. They can be accessed
sequentially. Once a particular parameter is considered, its present value will appear
in line 2, while a series of dots will appear on line 3. As indicated, use keypad ESC to
scroll parameters forward leaving them unchanged, or enter a value in the dotted line
followed by ENTER. As long as the entered value is acceptable, it will replace the old
one in line 2, while the cursor will return at the beginning of the dotted line, ready to
accept another value for the same parameter should you decide to do so. Else, press
ESC to definitely accept the entered value and move to the next parameter. In case
of typing error during data entry, use keypad DEL to clear the input field. If the
keyboard “MENU” is pressed, the procedure is terminated and the program execution
returns to the application menu. The table of Recipe Parameters is made up as
follows:
44
Parameter name on screen Meaning Limits
45
6. Connection to an Outside Computer
The LTP100W is constantly ready to receive messages from the Host Computer.
After a message has been received without errors (either frame or syntax errors), the
LTP100W sends back an appropriate answer, in some cases just to acknowledge, in
others to send some kind of information.
The set of messages defined in the LTP100W allows full remote control of the unit,
including state changing, motors starting and stopping, parameters and recipe
downloading and data polling.
While a detailed structure description of all messages is beyond the scope of this
Guide, we list them here together with a brief description of their purpose, for better
comprehension of the Host Computer operational procedures.
Actual weight
Actual tare
Step timeout
Actual dosing step (tare stabilization, charging, waiting fall down,
discharging, waiting next cycle)
Dosing state (tare over limits warning, weight out of range
warning, charging timeout warning, motor/valve alarm, weight over
limits, dosing unit in test mode, tare over limits alarm, weight out of
range alarm, charging timeout alarm)
Charging device state (stop, starting, running at low speed,
running at high speed, in alarm, valve closed, valve open, valve
moving, valve in alarm)
Discharging device state (valve closed, valve open, valve moving,
valve in alarm)
Charging device alarm (timeout, fault)
Discharging device alarm (timeout, fault)
Fall down
Batch weight
46
10H,01H Batch Parameters download.
The Host computer downloads the following parameters:
47
04H,02H Regulation Parameters Polling.
The LTP100W returns the following information:
recipe set-point
recipe upper limit of deviation error
recipe lower limit of deviation error
48
The Host computer downloads the following parameters:
Table 1
PID States
PID Alarms
manual
automatic high deviation
cascade low deviation
recipe recharge timeout
direct/reverse
Table 2 Table 3
Motors states
Valves states Motors alarms
stop Valves alarms
closed timeout
starting
moving fault
running low speed open timeout
open belt misalignment
running high speed close timeout
alarm local mode
alarm
LTP100W in test
49
7. Specifications for RS485 line connection
In this chapter the specifications for connection and usage of LTP100W terminal are
described .
Data are transferred from one device to the other in both directions on one
single twisted two-pairs cable (TX - RX – reference signal ).
The communication line will therefore have a starting point that normally is
identified with the “MASTER” device and an arrival point within the max.
distance of cable coincident with a “SLAVE” device.
The communication drivers and the reference mass signal of the RS485
interface of LTP100W device are insulated in galvanic way from the internal
input of the terminal .
The communication drivers consider a max. total charge on the two lines
of 60 Ohms given by the terminators.
The reference signal has a 100 Ohms in-series resistance to the common
input cable of the driver, for protection against electrostatic discharges.
50
For the RS485 connection LTP100W terminal is equipped with two
connectors DB9, male for the arrival connection , female for the start
connection or for the end-of-line terminator .
PC
OR
PLC
R(A-) R(A-)
R(B+) R(B+)
SHIELD
GND GND
GND GND
The shield of cables starting from the LTP100W can be connected to the
metallic housing of DB9 connector , being sure that the female connector on
the terminal is earth-connected .
51
Example of RS 485 serial cable between PLC SIEMENS
MODULE CP441and LTP100W terminal
52
Example of RS 485 serial cable among PC with converter
and LTP100W terminal.
B- ------------------------------------------------- 9 (RA-)
intelligent
PC Converter A+ ------------------------------------------------- 8 (RB+)
RS 232/485
GND ------------------------------------------------- 5 GND
SHLD
1K2
+V
A+
B-
120
GND
1K2
53
An example of intelligent converter is the “ 485-25E “ of “Integrity
Instruments” with on-board micro-controller and a.m. terminator
embedded in the device .
54
The terminator consists of a plug with DB9 male connector
on whose pins three resistive networks are installed as
showed here below .
1K2
7 (+V)
8 R(B-)
9 R(A+)
150
4 (GND)
DB9 M
1K2
55