Download as pdf or txt
Download as pdf or txt
You are on page 1of 21

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/319041719

Dental implants: An overview

Article  in  Dental Update · July 2017


DOI: 10.12968/denu.2017.44.7.596

CITATIONS READS

15 21,242

5 authors, including:

Abdulhadi Warreth Matteo Cremonese


Ajman University Dublin Dental University Hospital
29 PUBLICATIONS   259 CITATIONS    1 PUBLICATION   15 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

The seal-ability of a single-cone root canal obturation method View project

Dental Implants. An overview View project

All content following this page was uploaded by Abdulhadi Warreth on 09 November 2018.

The user has requested enhancement of the downloaded file.


ImplantDentistry

Abdulhadi Warreth
Najia Ibieyou, Ronan Bernard O'Leary, Matteo Cremonese and Mohammed Abdulrahim

Dental Implants: An Overview


Abstract: Dental implants are widely used and are considered to be one of several treatment options that can be used to replace missing
teeth. A number of implant-supported treatment options have been used successfully to replace a single tooth and multiple teeth, as well
as a completely edentulous jaw. However, as the number of patients who have dental implants is increasing, dental personnel are more
likely to see patients with implant-supported restorations or prostheses. Nevertheless, dental implants may fail as a result of mechanical
complications, such as screw loosening or due to biological causes like peri-implant diseases. As a result, dental personnel should be able
to recognize these complications and the factors that have negative effects on the success of such implant-supported restorations or
prostheses. Therefore, a basic knowledge of dental implants is necessary for every dental student, hygienist and dentist.
CPD/Clinical Relevance: Maintenance of implant-supported restorations and prostheses requires long-term follow-ups. It is the
responsibility of the patient to maintain good oral hygiene and also of the dental personnel who look after the patient to ensure a durable
restoration and prosthesis.
Dent Update 2017; 44: 596-620

Dental implants (also known as oral or integration is influenced by several factors, implants is known as peri-implant tissue
endosseous implants) have been used to such as implant material, bone quality and is comprised of soft (mucosa) and
replace missing teeth for more than half and quantity, and the implant loading hard (bone) tissues. The peri-implant
a century. They are considered to be an condition.2,3 soft tissue has similar features to the soft
important contribution to dentistry as As the use of dental implants tissue that surrounds teeth.7-10 It consists
they have revolutionized the way by which has become much more common, dental of a junctional epithelium and connective
missing teeth are replaced with a high personnel are more likely to see patients tissue. The junctional epithelium is
success rate.1-3 This success depends on the who have implant–supported/retained attached to the implant and/or abutment
ability of the implant material to integrate restorations. Nevertheless, dental implants surface through a hemi-desmosomal
with the surrounding tissue. However, this are affected by diseases in a similar manner attachment. Connective tissue is present
to teeth and may also fail after several apical to the junctional epithelium and
months or years in service.4-6 Therefore, coronal to the crest of alveolar bone.10
Abdulhadi Warreth, BDentSc, it is not unreasonable to suggest that Connective tissue fibres are found to be
MDentSc(TCD), PhD(TCD), MFD RCSI, the implant and the peri-implant tissue positioned close to the implant surface
Department Restorative Dentistry, should be examined on a routine basis in but not attached to it, and predominantly
Ajman University, Al–Fujairah Campus, a similar manner to that which is carried arranged in a circular manner. Connective
United Arab Emirates, Najia Ibieyou, out for periodontal examination.7 So, when tissue fibres also arise from the crest of
BDentSc, MDentSc(TCD), PhD(TCD), a deviation from the norm is found, the alveolar bone and from the periosteum
Postgraduate student, Institute of treatment may be carried out in practice or and are oriented parallel to the implant/
Molecular Medicine, Trinity College, by a specialist, depending on the severity abutment surface and extend towards
Dublin, Ronan Bernard O’Leary, of the condition. Accordingly, the dentist the oral epithelium. Thus, the junctional
Fifth Year Dental Science, Matteo should be equipped with basic knowledge epithelium and connective tissue form
a protective seal between the oral
Cremonese, Third Year Dental Science, of dental implants. Hence, it is the aim of
environment and the peri-implant bone
Dublin Dental University Hospital, this article to provide this basic information
which plays a vital role in the success
Trinity College, Dublin and Mohammed which is needed by every dental student and
of the implant treatment outcome. The
Abdulrahim, BDentSc MDentSc(TCD), dentist alike.
junctional epithelium and the connective
PhD(TCD), Oral Medicine Department,
tissue are collectively known as the
Faculty of Dentistry, Benghazi University, Implant-soft tissue interface biologic width, which is comparable to
Benghazi, Libya.
The tissue that surrounds that found around teeth.11
596 DentalUpdate July/August 2017
ImplantDentistry

Implant-bone interface and categorized into four classes, as described


osseointegration in Figure 2 and Table 1. Some factors which
For dental implants to succeed, affect osseointegration are discussed below
intimate contact between the peri-implant and summarized in Table 2.
bone and the implant surface should
be achieved and maintained. Therefore, Implant placement methods
an integration between the implant Surgical implant placement may
surface and the bone is required for be carried out in one- or two-stage methods
the success of any implant system. This (Figure 3). The one-stage method is also
integration is known as osseointegration, known as the non-submerged method.
and is defined as a direct structural Using this technique, the bone is prepared to
and functional connection between receive the implant. The implant is fitted into
ordered living bone and the surface of the prepared bone (osteotomy). However,
a load-carrying implant.12 Under light the coronal part of the implant is kept
microscopy, successful osseointegration above the bone crest, protruding through
shows direct apposition of bone on the soft tissue, and is exposed to the oral
implant surface (Figure 1). However, when environment during the healing stage.22 The
the bone-implant interface is examined restoration can be attached immediately
using electron microscopy, the implant after the implant placement surgery or may
surface is found to be separated from also be delayed.
the surrounding bone by an amorphous The advantages of the one-stage
layer, a granular electron-dense layer, or a method include:16
layer of uncalcified collagen fibrils13,14 with  The avoidance of a second surgical
a thickness that ranges from 100 nm to procedure;
400 nm.13 Nevertheless, this layer appears  The lack of a micro-gap between the
not to have a negative impact on the implant and the abutment at the alveolar
success of the osseointegration. Inversely, bone crest level, resulting in a less crestal
when the connection between implant bone resorption; Figure 1. A histological image of bone-implant
surface and bone is mediated by a layer of interface. Bone formation around the implant
 The prosthetic procedure is simplified and
labelled with different chelating agents (fluoro-
connective tissue, osseointegration fails to less chair time per patient is required; and
chromes). The implant is the large black area.
occur.5,15,16  A non-loaded, immediate, or delay-
It is important to mention that,
as a result of the absence of periodontal
ligaments between the implant and its
surrounding bone, when the implants are
loaded, they move within the bone due to
bone elastic deformation.6 Furthermore,
osseointegrated implants cannot be
moved orthodontically.
Several factors are
reported to play a role in obtaining
osseointegration.17,18 As an example, poor
bone quality was found to be associated
with a high implant failure rate when
compared with bone of a high quality.19 A B D
C
Clinical studies have reported that dental
implants in the maxillary arch (especially Figure 2. The classification of bone according to its quality: Class I (A), Class II (B), Class III (C) and Class
for the posterior maxilla) have lower IV (C).
survival rates than those in the mandibular
arch.19 This is usually attributed to the
differences in bone quality between  Type I: almost the entire bone is composed of homogeneous compact bone;
the two arches.20 Bone quality, as  Type II: a thick layer of compact bone surrounds a core of dense trabecular bone;
classified by Lekholm and Zarb,21 is  Type III: a thin layer of cortical bone surrounds a core of dense trabecular bone; and
based on radiographic assessment as  Type IV: a thin layer of cortical bone surrounding a core of low density trabecular bone.
well as resistance during the implant
Table 1. Classification of bone according to its quality.21
drilling procedure. Accordingly, bone is
July/August 2017 DentalUpdate 597
ImplantDentistry

loaded protocol can be implemented. undue trauma which can negatively affect the implant platform below the bone crest
One of the drawbacks that the healing. However, bone with optimum may be implemented. The countersink
may be associated with this surgical quality and quantity is a prerequisite for allows the placement of the cover-screw
protocol is that the implant is exposed this method to be used. Nevertheless, level with the bone crest. The raised flap is
to the oral environment, which may lead the method can be clinically successful. then repositioned and sutured to conceal
to contamination of the surgical site. Examples of the implants that can be placed the cover-screw and the implant (Figure
Furthermore, the implant may be exposed to using the one-stage technique include 3). After a few months, the second stage
the Solid-Screw Implant® (Straumann surgery is carried out. In this stage, the
UK, Crawley, W Sussex), AdVent Implant® implant site is re-opened, the cover-screw is
(Zimmer, FLA, USA) and Single-stage Implant accessed and then replaced with a healing
 Bone quality and quantity System® (BioHorizons, AL, USA). abutment, which is also known as a sulcus
 Implant shape In contrast, the two-stage former or transmucosal abutment (Figure 4).
 Implant surface macro-structure Afterwards, the healing abutment is replaced
method is also known as the submerged
 Implant micro-structure (roughness) with a provisional or final restoration. This
technique (Figure 3). In this method, two
 Material biocompatibility surgical protocol is suitable for use when the
surgical procedures are carried out. The
 Surgical techniques quality of bone is not optimum and when
first surgery involves installing the implant
 Heat generation during the implant bone graft materials are used in conjunction
into the bone, and a cover-screw (also
placement surgery with the implant. Examples of an implant
known as a sealing-screw) is attached to
 Implant primary (initial) stability
the implant platform. A countersink bone system used for the two-stage procedure
 Implant loading
preparation that allows for placement of include the Fixture MK III® (Nobel Biocare,
Uxbridge, UK), MAX 2.5® Implant (Bicon Inc,
Table 2. Some factors affecting osseointegration.
Boston, MA, USA) and OSSEOTITE® 2 Certain
Implant (BIOMET 3i, Maidenhead, UK).
It is important to mention that
the cover-screw is used to prevent tissue
growth into the implant or over its platform.
It is attached to the implant using a screw-
driver with a light finger force. It is essential
to confirm that the cover-screw is fully
seated and no gap is left between the cover-
screw and the implant platform. The cover-
screw has a low profile which facilitates
the suturing procedure and allows the two
edges of the cut mucosa to be brought
close together without undue tension. If
there is too much tension, it may deteriorate
and preclude the healing.23 Conversely, the
healing abutment has a high profile and
protrudes through the peri-implant mucosa
Figure 3. A schematic presentation of an implant placed according to the one-stage (left) and two- to the oral cavity. Therefore, the healing
stage (right) implant placement methods. Note the transmucosal part (the neck) penetrating the peri- abutment is available in different lengths,
implant mucosa in the one-stage method. depending on the distance between the
implant platform and the surface of the
peri-implant mucosa. It is also available in
a variety of diameters, which is selected
according to the implant diameter. The
cover-screw and the healing abutment are
shown in Figure 4.

Implant stability
Implant stability (lack of mobility)
is divided into primary and secondary. The
primary, also known as initial stability, is
achieved during implant placement surgery.
It is believed that primary stability plays a
Figure 4. An image of a cover–screw (left) and healing abutments (middle and right)
vital role in reaching osseointegration, upon
598 DentalUpdate July/August 2017
ImplantDentistry

which secondary stability depends. are American Dental Association (ADA) used to overcome angle mismatching
Implant stability is produced approved. The most commonly used implant problems.
by close contact between the implant and systems include Nobel Biocare, Straumann, Implants are also available as
the host bone. The factors that may affect AstraTech, Bicon, BioHorizon, BIOMET 3i, hollow and solid. Hollow implants allow
primary stability may be categorized into Intralock, and Zimmer. All are constructed more contact with bone but are weaker
three factors; those related to surgical site on the same basic concepts but there are than solid implants, which makes them
(local) or related to implant or surgical differences in the patented technology and more susceptible to mechanical failure and
method used in placement of the implants. materials. fracture. An example of a hollow implant
Local factors, such as bone quality and In general, dental implants may is the Hollow Cylinder Implants® made by
volume, may affect the degree of bone-to- be classified as a one- or two-piece implant. Straumann and ITI (Basel, Switzerland).
implant contact and consequently affect Irrespective of the implant type
primary stability. As an example, larger bone- The one-piece implant and for descriptive purposes, the implant
to-implant contact fractions were observed In the first type, the implant and usually consists of an implant body and
in bone sites of higher density. The implant the abutment are formed as a single solid neck. The implant body is the part of the
factors include shape, length, diameter unit. In this case, there is no screw-joint implant that is buried in the osteotomy.
and surface texture. For instance, tapered between the implant and the abutment. The coronal part of the implant is denoted
implants lead to higher insertion torque The lack of a screw-joint is considered an as the neck, through which the abutment/
values than cylindrical implants, which advantage as there is no screw-loosening, attachment is connected to the implant.
was considered to be due to the greater dangerous fracturing or micro-motions The coronal part may be smooth (one- and
frictional surface of the tapered implants between the abutment and the implant. two-piece) and placed above the crest of the
and associated with high primary stability The one-piece implants may be used when bone, or roughened (two-piece), in which
(see below). A surgical technique, such as narrow implants are indicated, such as in the the platform is usually placed below or level
that which leads to bone condensation replacement of the maxillary lateral incisors with the crestal bone. When the coronal part
during implant placement surgery or a and lower incisors, or when bone volume is smooth and placed above the crest of the
mismatch between the osteotomy and is limited and the use of standard implants bone and penetrates peri-implant mucosa,
implant diameter (with the implant diameter is not suitable. These types of implants are it is known as the transmucosal part. The
being slightly greater than the osteotomy), installed only with the one-stage implant surface of the transmucosal part is usually
results in satisfactory primary stability.24 placement method. Examples of a one- highly polished and is available in different
Also, the use of implants with self-taping piece implant are the one-piece 3.0 Dental lengths. It may also have a straight or a bevel
blades results in a lower primary stability in Implant® (BioHorizons) and Y-TZP Ceramic profile and may be augmented with micro-
medium-density bone when compared with Implant® (Nobel Biocare). grooves in order to optimize healing around
those without such blades.25 However, this the implants.
issue is contradictory. Placing the smooth (machined)
The two-piece implant
Secondary stability represents part of the implant below the bone crest
The two-piece implant type may lead to its resorption.29 However,
integration of the implant as a result of new
consists of an implant to which an abutment fewer crestal bone changes were observed
bone formation through its remodelling.26-28
or a restoration/attachment is connected, when the smooth part was located above
Therefore, this stability depends on bone
usually with a screw. It is more commonly the crestal bone level, irrespective of the
activities and factors that influence such
used than the one-piece implant type. With implant type; one- or two-piece implants.29
activities throughout the patient’s life.27
this implant type, both the one- and the Accordingly, it has been recommended that
The general consensus is that peri-implant
two-stage implant surgery protocol can be the smooth-rough border should coincide
bone is in a continuous active remodelling
implemented. with the alveolar bone crest.29
state which maintains osseointegration and
Angled implants in which their
provides secondary stability.27-29
coronal part is angled in relation to the
It is important to mention that,
main implant body are also available. These
Features to consider when
when the implant is inserted into the host
angled implants are useful in the anterior
choosing an implant system
bone, spaces may exist in the bone-implant Five features can be used to
region when placing non-angled implants
interface. These spaces are initially filled describe the dental implant body: shape,
in their optimum position is not possible. An
with blood that comes from injured blood surface macro- and micro-structure, length
example of angled implants is the Co-axisä
vessels, forming a fibrin network which is and diameter. These features are important
implant (the Southern Implants, UK) in
the important step towards the formation of when an implant system is chosen.
which the neck is at an angle to the long
osseointegration.
axis of the implant body. It is useful to use
when the long axis of a prospective implant 1. Shape (geometry)
Dental implant types is not along the long axis of the potential Implant shape may generally
In the worldwide market, restoration. An angled abutment, such as be tapered or parallel (straight-walled).
there is a wide range of dental implant Regular Neck synOcta® angled abutment The tapered type in general has more
systems available, but only a few brands (Straumann), is also available and can be primary stability than the parallel type.30
July/August 2017 DentalUpdate 601
ImplantDentistry

The use of tapered implants results in lateral edentulous space and in softer bone, such
compression of bone and increased stiffness as type IV (Figure 2), where primary stability
of the interfacial bone, which is reported is not always easy to achieve.32 They may
to increase the implant primary stability.13 also be used immediately or early after
Tapered implants were found to require a tooth extraction.32,33 The use of a tapered
higher insertion torque and less insertion implant with a wide platform achieves
time than parallel implants. A higher a satisfactory emergence profile of the
insertion torque gives a better implant restoration.
Figure 5. A representation of the most commonly primary stability.31 Tapered implants are also
used implant threads: V-shaped thread (left); used to avoid damaging the converging
square thread (middle) and a reverse buttress 2. Surface macro-structure (threads)
roots of adjacent teeth that bind the The implant macro-structure
(right).
is represented as threaded or non-thread
(thread-less). The threaded type is the
most commonly used implant design.
The threads are usually incorporated into
the implant design to improve the initial
stability and dissipate interfacial stress in
a more favourable way. As the threaded
implants provide better mechanical and
biological outcomes, non-thread implants,
such as cylinder (press-fit) implants, are
less likely to be used and are replaced by
the threaded type. Thread features such as
thread depth, thread thickness, face angle,
pitch and helix angle are considered to
be factors that determine the functional
thread surface and affect the biomechanical
load distribution of the implant.
Figure 6. Bone resorption at alveolar crest occurs after tooth extraction which may preclude the use of There are three thread shapes
a long implant as the crestal bone has to be trimmed down to maintain at least one millimetre of bone which are most regularly used when a
buccally and lingually at the bone crest region. dental implant is described (Figure 5). These
are V-shaped, square-shaped or reverse
buttress.34,35 An animal study conducted
Implant length: a long implant should be considered whenever the condition permits.
by Steigenga and colleagues36 revealed
Implant diameter: ideally, the implant should be approximately the same diameter as the effects of thread type on peri-implant
the root of the tooth it is replacing. bone formation. The study showed that
a. Wide implant: implants with a square thread design had
i. Poor quality bone; significantly more bone-implant contact
ii. Limited ridge height with adequate mesio-distal and bucco-lingual width; and and greater reverse-torque measurements
iii. Immediate implant placement (after tooth extraction). than observed when the V-shaped and
b. Narrow implant: reverse buttress thread designs were tested.
i. Used to replace maxillary lateral incisors or mandibular incisors; A threaded implant may also
ii. Limited edentulous space; be classified as a self-taping or pre-taping
iii. Limited ridge width (to avoid ridge augmentation surgery); implant.37 A self-taping implant is an
iv. When it is not possible to achieve good emergence profile with a wide implant body; implant which is designed to make its
and own threads as it is being placed into the
v. Converging adjacent tooth roots. prepared osteotomy. On the other hand, in
pre-taped implants, threads are prepared
Tapered implant: on the surface of the osteotomy using a
i. In type IV bone, where primary stability is difficult to achieve; tap drill (taper). The produced threads will
ii. Narrow or concave bone; accommodate the threads of the implant.
iii. Converging adjacent roots; and The pre-taping method is sometimes
iv. Immediate and early implant placement. recommended, such as in the case of dense
bone (type I and II) (Figure 2). However, pre-
Table 3. Some implant features that should be considered when an implant is selected, and their
taping implants achieved lower primary
indications.
stability than the self-taping implants.38
602 DentalUpdate July/August 2017
ImplantDentistry

3. Surface texture (micro-structure) quality which will accommodate the implant 93.1% for 5 mm implants and 98.6% for 9.5
Implant surface texture describes (Figure 6). As implant primary stability is a mm implants.51 Furthermore, short implants
the roughness of the implant surface. function of contact between the implant may fail at an earlier stage than standard
Therefore, the implant surface is either surface and bone, the longer the implant, implants,20 as peak failure rates of short
smooth (machined) or can be of a variety the greater the surface contact and primary dental implants were 4−6 years, and 6−8
of roughness. A rough-surfaced implant stability. However, the increase in implant years for the standard implants.52
has a larger surface area than that of its stability does not occur linearly to the It is important to note that bone
counterpart smooth implant. It is found increase of the implant length. For instance, resorption following tooth extraction may
to be associated with positive healing of a 10 mm implant has about 30% more result in the thinning of the alveolar bone
peri-implant tissue and encourages the surface area than a 7 mm implant, while a crest, which may preclude placement of
formation of osseointegration.39 The increase 13 mm implant has 20% more surface area an implant with an adequate length and
in surface area distributes forces to which than a 10 mm implant.46 diameter, as shown schematically in Figure
the implant is exposed in a more favourable The bone of the edentulous 6. Therefore, bone mapping and a CT-scan or
manner. It also provides better primary ridge may not be sufficient for placing an Cone-Beam Computed Tomography may be
stability than that attained when the implant implant with the optimum length. Therefore, required.
surface is smooth.40 Histomorphometric and several techniques have been suggested to
removal torque studies with roughened compensate for the deficiency in the residual 5. Implant diameter
implant surfaces have revealed greater ridge, either before or simultaneously with The implant diameter is
bone apposition41 and higher removal implant placement. Among these methods measured from the crest of the widest
torque values than implants with smoother are guided bone regeneration, block grafts, thread to the same point on the opposite
surfaces.42 sinus lifting procedures, inferior alveolar side of the implant.53 According to the
In general, two methods for nerve repositioning methods, and bone diameter, implants may be classified as mini
the alteration of implant surface texture distraction.47 These surgical methods are when diameter is ≤2.7 mm; narrow when the
have been described in the literature: successful and can be used to increase bone diameter is >2.7 mm but ≤3.75 mm; regular
subtractive and additive methods. In the height.47 However, they are not without risks when it ranges from 3.75−5 mm; and wide
subtractive method, the implant surface and may lead to several complications and when the diameter is >5 mm.
is roughened by removal of its surface
undesirable treatment outcomes.5,47 This may The implant diameter plays
materials usually by blasting and/or acid
encourage the dentist and patient to avoid an important role in the success of oral
etching.42-44 In the additive method, a
such surgical methods and to use short implants and has a major impact on the
biocompatible material, such as titanium or
implants, therefore the implant is installed implant’s ability to withstand occlusal load.54
hydroxyapatite, is added to the surface42 (see
with less invasive surgical procedure and Selecting an implant of a suitable diameter
below). Some examples of rough surface
the cost is reduced. Nevertheless, when a is governed by the dimensions of the
implants include: grit blasting with titanium
short implant is used, factors that affect the edentulous space (bucco-lingual and mesio-
oxide produced by Astra Tech (Mannheim,
osseointegration, such as implant shapes, distal) (Figure 7), as well as the bone quality.
Germany); Sand-blasted Large-grit Acid-
surface texture, and thread designs, should Moreover, it is also affected by the type of
etched (SLA®) implants from Straumann
be carefully selected to achieve a satisfactory tooth being replaced.
(Basel, Switzerland); Acid-etched Implants®
long-term outcome.47,48 However, earlier An increase in the diameter of
from BIOMET 3i (Florida, USA); and Plasma-
studies have reported that shorter implants an implant is associated with an increase
sprayed® (molten titanium sprayed on the
are unpredictable and fail more frequently in its surface area. For instance, increasing
implant surface) produced by Straumann
than longer implants.46,49 In addition, the diameter in a 3 mm implant by 1 mm
and Dentsply Sirona Implants (Weybridge,
longer implants had statistically higher increases the surface area by 35% over the
UK).
survival rates when compared with shorter same length.55 Also, a 3.75 x 10 mm implant
It is important to note that, if
implants.50 For instance, it has been reported has 61% less surface area than a
the rough implant surface is exposed to the
that survival rates after two years were 6 mm diameter implant of the same
oral environment, it may encourage plaque
accumulation and interfere with its removal,
and subsequently may induce peri-implant
disease (see below).6,45

4. Implant length
Implant length is determined by
the distance between the top surface of the
implant platform and the apex. In general,
the length of the standard implant ranges
from 7−18 mm.33 Selection of an implant Figure 7. The implant should be placed in the site that was previously occupied by the tooth being
replaced, and surrounded by an adequate amount of bone. Two implants may be used to replace a
of the required length is governed by the
molar tooth, which results in the dissipation of the occlusal forces in a satisfactory manner (right).
available vertical bone height, width and
July/August 2017 DentalUpdate 603
ImplantDentistry

length.33 Furthermore, an increase in the emergence profile, the implant platform is narrow implants are an alternative. For
diameter and a change in the threads may usually placed at about 2 mm apical to the example, narrow implants are suitable
lead to an increase in the implant surface cemento-enamel of the adjacent teeth. If an for replacing maxillary lateral incisors
area of more than 300%. This increase in implant is placed deeply below the crest of and mandibular incisors. They are also
the surface area may lessen stresses to bone, the crown height is increased, which suitable when bone quantity is insufficient,
the crestal bone areas and reduce both may lead to mechanical failure of implant or when the roots of adjacent teeth are
crestal bone loss and early loading implant components and compromise aesthetic converging. They may also be used with a
failure.55 treatment outcomes. When the implant is removable implant-supported overdenture.
It is important to mention that, placed more superficially, restoration may be However, the use of an implant with a small
when the implant is installed, it should be in deemed impossible and aesthetic treatment diameter is not without disadvantages,
close contact with the surrounding bone of outcome is also compromised.33 such as mechanical failure of the implant
not less than 1 mm thickness on its buccal When a molar tooth is replaced, component. Furthermore, obtaining a
and lingual surface, and preferably 1.5 mm the use of two implants may be an option, as good emergence profile of the restoration
or more between the implant surface and its dissipation of occlusal loads are favourable. may also be a problem. Hence, a detailed
adjacent tooth (Figure 7). For instance, when However, placement of implants close to examination of each patient’s condition
an implant of 4 mm is selected, the bucco- each other is associated with difficulty in should be taken before a specific implant is
lingual and mesio-distal dimensions of the obtaining an optimal emergence profile, selected, and alternative treatment options,
edentulous space should be a minimum of interferes with oral hygiene and leads to such as a fixed (conventional or resin-
6.0 and 7.0 mm, respectively. However, it has chronic inflammation and bone resorption. bonded) prosthesis, may be considered.
been suggested that, in the aesthetic zone, Short and wide implants may It is important to distinguish
maintaining a minimum of 3 mm of bone be used to compensate for the decrease in between the implant diameter and platform
between adjacent implants is beneficial, the vertical bone height of the edentulous diameter as they may not be equal. The
as bone height as well as the inter-dental space when surgery cannot be considered. implant platform represents the part of the
papilla are more likely to be maintained.11 They may also be used when the quality of implant that is connected to the prosthetic
Consequently, implants with a smaller the bone bed is not optimal.56 Wide implants (abutment) counterpart. Table 3 displays
diameter at the implant-abutment interface can be used to increase implant stability,57 examples of implant features that should be
may be used when multiple implants are to thus improving stress distribution within considered when an implant is selected.
be placed.11 the surrounding bone.47 Furthermore, the
The diameter of the roots use of a wide diameter implant may reduce
is usually estimated at 2 mm apical to the stress on the retained screws. Wide
Implant materials
the cemento-enamel junction. With this implants are also used for the replacement of The most commonly used
measurement, an implant with a diameter posterior teeth and immediately after tooth materials in dental implants are either bio-
that matches, or is slightly smaller than, the extraction (Table 1).58 inert, such as commercially pure titanium (Cp
tooth being replaced is selected. In order Several situations do not allow Titanium) and titanium alloy, or bio-active
to obtain a restoration with an optimal the use of wide diameter implants59 and ceramics such as hydroxyapatite, tri- and
tetra-calcium phosphate and bio-glass.18
For more than five decades,
titanium was the most commonly used
material in dental implants due to its bio-
compatibility, as well as its mechanical and
physical properties, such as resistance to
corrosion, high strength and low weight.60
Depending on its oxygen content, Cp
titanium may be categorized into four
grades; grade I contains the least oxygen
while grade IV contains the most (0.18%
versus 0.4%).18 Titanium alloy consists of 90%
titanium, 6% vanadium, and 4% aluminium
and is classified as grade V.34
Titanium is a non-noble metal
which has the ability to form a very adherent
self-repairing and protective surface oxide
layer, which prevents further titanium
corrosion. This layer forms immediately
when the titanium is exposed to oxygen.
Figure 8. A schematic representation of the screw-joint connections: the external connection and the
The formed oxide layer on Cp titanium is
butt joint (left) and the internal connection and the slip joint (right).
similar to that which is formed on titanium
604 DentalUpdate July/August 2017
ImplantDentistry

alloys.34 Titanium dioxide (TiO2) forms the material and the substrate is required to of the area is important. The restorative
main constituent of this oxide layer, however, withstand functional stresses and to avoid space should be carefully investigated as it
other oxides, such as Titanium oxide (TiO) fragmentation of the coating materials.70 It is extends between the two adjacent teeth and
and Titanium pentoxide (Ti2O5) may also found that hydroxyapatite mechanical failure accommodates the prospective restoration.
exist. Incorporation of other chemical occurs primarily at the interface between
elements, such as carbon, traces of nitrogen the metal substrate and hydroxyapatite coat Abutment-implant connections
or chlorine, into the oxide layer have been (adhesive failure), irrespective of the implant (interfaces)
reported.61 design. This may have a negative effect on
When an implant is put to
The release of metallic ions implant osseointegration.71 Nevertheless,
function, it is connected with the restorative/
from the titanium implant surface may the risk for hydroxyapatite-coat degradation
prosthetic components. The connection
occur and increase as the implant surface and loosening (delamination) are still a
type can be classified as internal or external.
area increases.62 It has been suggested remaining concern.
In the internal connection systems, the
that ionic release may interfere with the With improvement in technology,
normal peri-implant bone mineralization ceramic materials are extended for use apical part of the abutment is inserted into
and remodelling, which could lead to as implant substrates. This is because an access hole in the implant platform. In
the failure of the implant.63 Furthermore, ceramics such the yttrium-stabilized the external systems, a protrusion located
titanium release may induce hypersensitivity tetragonal zirconia polycrystalline has above the implant platform is inserted into
in susceptible patients, which may have an improved mechanical properties, superior a recess in the apical part of the abutment
undesirable impact on implant success.64 wear and corrosion resistance, with a high (Figure 8). The connection is also classified as
However, this issue is still debatable flexural strength. These characteristics a slip joint; when there is a space between
and more clinical and further laboratory may make them a potential alternative opposing mating surfaces, and a friction
investigations are required.24,64 Nevertheless, to conventional titanium implants for fit when such space does not exist. The
available literature indicates that Cp titanium supporting overdentures.72 Three types of connection may be further categorized as a
has a long-term successful performance. zirconia-containing ceramic systems are bevel (conical) joint or a butt joint (Figure 8).
In addition, the surface of the titanium most commonly used in dentistry; yttrium- The connection may have
implant, which was previously contaminated stabilized tetragonal zirconia poly-crystals, an anti-rotational component, such as
in the peri-implantitis case, was found to alumina-toughened zirconia and zirconia- hexagonal, octagonal, cone hex, cylinder
reintegrate with bone which was treated to toughened alumina. However, these non- hex, cam tube and pin/slot or be without
remove the contaminant.65-67 metallic materials are expected to replace Cp an anti-rotational device, such as a cone
Cp titanium and titanium alloys titanium and its alloys.73 Nevertheless, based (Morse taper). The function of the anti-
can make up the entire implant or can be on their systematic review of literature, rotational component is to stabilize and
used as a substrate to which a coating of Andreiotelli and colleagues74 concluded prevent abutment rotation.77,78 Likewise,
bio-active material, such as hydroxyapatite, that ceramic, in particular zirconia, implants the connection usually has a screw but is
is attached. are not yet suitable as an alternative to sometimes screw-less and relies entirely on
To speed up the healing titanium implants. Nevertheless, they the friction fit for its stability, such as Bicon®
process and osseointegration, implant potentially could be a successful material dental (Bicon Inc, Boston, MA, USA).
surfaces are coated with ceramics.68 The for use in implants, but this has not yet The first implant connection type
ceramics may be bio-active, such as calcium been supported by clinical investigations.75 used with a dental implant was described
phosphates, or inert, such as aluminium However, ceramics such as zirconia are used by P-I Brånemark.12 It was an external hex,
oxide and zirconium oxide. Examples of nowadays as abutments and crowns as they therefore consisting of six sides, each two
calcium phosphate coating materials are have good clinical outcomes.73 adjacent sides make a 60-degree angle
hydroxyapatite and fluorapatite.68 The It is not unreasonable to and had a height of 0.7 mm. The hex was
bio-active ceramics are reported to act as conclude that the prospective implant originally used to carry and insert the
osseoinductive materials which encourage should be selected carefully and a restorative implant into the prepared host bone
and accelerate bone apposition around the driven approach should be implemented to (osteotomy). The hex was not aimed for
implants. Furthermore, coatings that have avoid an unwanted result.76 Thus, thorough use as an anti-rotational device, as the
similar properties to that of the extra-cellular investigation should be carried out to implants were mainly used to restore
matrix provide a favourable environment for guarantee the best possible outcome. The completely edentulous dental arches with
osteoblasts, osteoclasts and their progenitor edentulous area should be viewed in three implant-supported overdentures with
cells, that are responsible for the healing dimensions: mesio-distal, bucco-lingual and multiple implants. Consequently, rotational
of bone.69 Therefore, an early and strong corono-apical. The mesio-distal dimension displacement of the overdenture was not
implant stability is achieved and the risk of of the edentulous space should also be an issue. However, as the use of dental
implant failure is reduced.69 thought of as two interrelated spaces implants progressed and extended for use
Ceramics are initially used in the (inter-radicular and restorative). The inter- in replacing single and multiple missing
additive methods in which ceramic coatings radicular space holds the implant and can teeth, the use of a guiding index and an
are added to the metal implant. However, be found between the roots of the two anti-rotational device is needed. To fulfil this
high bond strength between the coating adjacent. Hence, a precise radiograph image requirement the original external hexagonal
606 DentalUpdate July/August 2017
ImplantDentistry

A connections were modified and are now with external ones.79 However, the implant
available in different heights including neck should be strong enough to resist
0.9, 1.0 and 1.2 mm and with various sizes. such loads. Nevertheless, when the internal
Furthermore, several types of internal connection is used with a narrow implant,
connections were also introduced and are the connection is exposed to vertical or
widely used nowadays. oblique loads. Although the screw itself may
In general, when the connection be protected from loading, the implant neck
is an internal type, the occlusal load is may not be able to resist such a load and will
Figure 9. Measurement of rotational freedom. A
usually dissipated through the implant mechanically fail80.81 as most of the occlusal
passive fit of the abutment (blue) into a recess
(hexagonal) in the implant platform (a dotted
body and the screw is more likely to forces are transferred to the implant walls.81
circle). The space between the two components is be protected from the imposed load.
represented by the red area. The rotational freedom Loose screws were reported to occur less Screw-joint
degree during abutment rotation is indicated by the frequently with internal connections than
When the implants and the
letter ‘A’.
restoration/prosthesis are connected
a b together by a screw, the connection is
known as a screw-joint.16,77,82 For example,
when the single restoration (crown) is screw-
retained, one screw-joint is usually found to
connect the restoration to the implant. When
the restoration is cement-retained, there
is also one screw-joint, but it is between
the abutment and the implant (see below).
The screw-joint is also found with the fixed
implant-supported prosthesis in a similar
way as that described for the cement- and
screw–retained single implant-supported
restoration. In the fixed implant-supported
overdentures (FISOs), there is a screw-
joint between the frame-work and the
implants, whereas in the removable implant-
Figure 10. An intra-oral radiograph showing a single implant-supported crown replacing the right supported overdentures (RISOs), there is a
second molar (a). The cuspal inclinations are lowered and flattened, but the occlusal table is widened screw-joint between the attachment system
which creates a cantilevering effect and exposes the restoration, the screw and the implant to high
and the implant.6,16 The attachment systems
tipping forces that may lead to their mechanical failure. A diagram of an implant-supported restoration;
are discussed later in the article. In some
the implant is oriented so occlusal loading is directed along its long axis (b).
situations when a screw-retained restoration
is used, there may be two screw-joints: one
1. Implant-abutment interface design/type. between the implant and the abutment,
2. Rotational freedom (misfit). and one between the abutment and the
3. Manufacturing allowances (tolerance). restoration/prosthesis.
4. The settling (embedment). When the screw is tightened,
5. Repeated opening and closing of the screw. there are two opposing forces that act on
6. The applied torque value: over and under torqueing the screw. the implant platform and the abutment or
7. Loading of restoration. restoration/attachment that form the joint.
8. Prefabricated metal- and costume-made cylinders. One of these forces tries to hold the joint
9. The casting process: together and is known as the clamping
a. Casting alloy;
force. The other force is called the separating
b. Investment; and
force as it tries to disengage the screw-joint
c. The finishing/polishing method.
components away from each other. Hence,
10. Screw design and materials:
the two forces are acting against each other.
a. Shank or shank-less screws (a shank-less screw is usually less resilient than that with
As a tightening torque is applied to the
a shank);
screw, a tension (pre-loaded) is generated
b. Shape and diameter of screw’s head;
in the screw. Consequently, the screw shank
c. Materials from which a screw is made of such as gold, titanium and gold-coated
and threads are tense and an elastic recovery
screws.
is generated, thus creating the clamping
force between the mating surfaces.16,77,82
Table 4. Factors that affect screw-joint stability.
To obtain an effective clamping
July/August 2017 DentalUpdate 607
ImplantDentistry

force, the tension created in the screw gold-plated surface. Both screw types were when excessive torque is applied, slippage
material should be less than that of the found to be associated with lower friction between the screw threads (male) and
material’s elastic limit (Young’s modulus) coefficients and greater pre-load values than the implant internal threads (female)
so no permanent plastic deformation or the conventional gold alloy and titanium occurs, which consequently leads to screw
screw fracture occurs. Maximum screw-joint alloy screws.86 Likewise, higher pre-loads loosening.94 Inversely, too little torque or a
stability can be achieved with a maximum were associated with gold-coated screws lower torque value which cannot produce
pre-load when the proportional limit of the when compared with that obtained from the required screw pre-loading needed to
screw is approached. Thus, to obtain this, screws made of uncoated gold or titanium hold the mating surfaces together exhibits
the applied torque should be 75% of the alloy for all insertion torques, as well as when greater micro-motion at the screw-joint,95
torque required to cause screw permanent the screws were re-tightened.87 which consequently causes screw loosening
deformation. In order to hold the implant Manufacturing tolerance is and may lead to its fatigue and fracture.
components together, a maximum clamping another factor that affects the screw- Therefore, it is vital to use the manufacturer’s
force and a minimal separating force are joint stability. It is defined as unplanned recommended tightening torque, which
required. Therefore, the clamping force deviations from the theoretical dimension should be within the elastic range of the
overcomes the separating force. of the shaft and its mating recess as some screw’s materials, as mentioned earlier.96,97
deviations from a perfect fit are expected, It is also essential to ensure consistent
Factors affecting screw-joint but not planned. Hence, this indicates an tightening torque values are applied.
insignificant value of misfit between the Therefore, torque gauges (control) should
stability
matting surfaces. This misfit allows for what be used and manual torqueing should be
Lack of screw-joint stability is is known as rotational freedom (play) to avoided.91 It is also important to calibrate
reflected in loosening of the screw. It is occur. The rotational freedom is calculated the torqueing devices to obtain consistent
considered as one of the most common by the formed angle between the clockwise torqueing.98
problems associated with the use of implant- and anti-clockwise rotation of the anti- Torqueing the screw should be
supported restorations.83 One of many rotational components of the screw-joints carried out carefully and a counter-torque
factors that play a role in the stability of (Figure 9). The rotational freedom may vary device should be used to avoid disturbing
the screw-joint is the friction coefficient of from 1.6 to 5.3 degrees.88 The most stable the osseointegration. Hence, the use of a
the materials used in the fabrication of the and predictable screw-joint may be expected counter-torque device is recommended as
implant components, such as the abutment, when the rotational freedom is lower it reduces transmission of the tightening
implant and screw. The friction coefficient than two degrees.89 Hence, the produced torque to the implant-bone interface. On
has an effect on the generated pre-loading. rotational freedom affects the stability of the average, about 90% of the recommended
Tightening torque and consequently the screw-joint. pre-load tightening torque is transmitted
developed pre-load is inversely affected by Furthermore, the presence of to the implant-bone interface when the
the friction between the mating surfaces.84 a micro-roughness on the implant and counter-torque device is not used. This value
In general, during screw torqueing, friction abutment mating surface, which is worn is reduced to only 10% when the counter-
occurs between the implant surface and the away as a result of screw torqueing, leads torque device is used.94
opposing abutment surface, between the to what is called settling (embedment Overloading of the restoration
head screw and the abutment surface and relaxation). Consequently, part of the may lead to screw loosening and failure.
between the screw threads (male) and the clamping force is lost and the screw Therefore, the occlusion should be adjusted
implant threads (female). As such, when a becomes loose. The mean loss of pre-load and occlusal forces should be directed along
screw is tightened, only 10% of the torque may be up to 40% of the original pre-load the long axis of the implant, whenever
is converted into screw pre-load, while the value 15 hours after screw torqueing.90 possible (Figure 10). This can be achieved
other 90% of the tightening torque is lost as To reduce the settling effect, it has been by construction of a restoration in which
friction.84,85 In order to maximize pre-loading, suggested that the implant screws should its occlusal morphology is constructed
the friction between mating surfaces should be retightened ten minutes after the initial according to the mechanical principals
be reduced. This can be achieved by coating torque application as a routine clinical that favour this concept. For instance, the
the mating surfaces with other materials, procedure.91.92 All screw types were reported cuspal inclination should be flattened and
such as carbon film or the screw with to display some decline in pre-load with the incisal guidance made shallow to avoid
tungsten carbide. This process is known as repeated tightening. This decline occurs bending moments caused by the lateral
dry lubrication and the coating material is irrespective of the insertion torque and component of the occlusal forces.99 The
denoted as a dry lubricant. Both carbon and abutment type.87 As screws lose pre-load occlusal table of the prospective restoration
tungsten carbide coatings were reported to following placement, their re-tightening may be reduced by 30−40% of the tooth
reduce the friction coefficient and improve is required from time to time during the being replaced (Figure 10) and cantilevering
pre-loading.84 Torq-Tite® abutment screws restoration’s life. the restoration should be avoided. Use
(Nobel Biocare, Uxbridge, UK) are made The screw pre-load should of an occlusal splint is recommended for
of titanium alloys and are coated with a be high enough to maintain the joint patients with parafunctional habits such as
carbon layer and Gold-Tite® abutment screws integrity and reduce the possibility of the bruxism. The implant should be placed in
(BIOMET 3i) are titanium screws with a screw loosening and fracturing.93 However, the site that was previously occupied by the
608 DentalUpdate July/August 2017
ImplantDentistry

tooth being replaced, and surrounded by limits its resorption.102 It also shifts the stress (prostheses) may be used to replace a
an adequate amount of bone (Figure 7). between the implant and abutment away single or multiple missing teeth, as well
It should also be oriented along the long from the cervical bone-implant interface, as completely edentulous mandible and
axis of the tooth being replaced and within which may also help in maintaining the maxilla. Therefore, when a patient whose
the occlusal table. However, when a molar crestal bone level. missing teeth were replaced with an implant-
tooth is replaced, the use of two implants A recent meta-analysis,102 supported restoration attends the dental
may be considered in order to dissipate the including 13 human randomized clinical clinic, one of the following restoration/
occlusal loads satisfactorily, as mentioned trials (RCTs), has shown a significantly less prosthesis is usually present:
earlier (Figure 7). mean crestal change at platform-switching  An implant-supported single restoration
Some of the other factors that implants, compared with when the implant (crown) (Figure 11);
may affect the screw-joint stability are platform dimensions matches the abutment  A fixed implant-supported prosthesis;
displayed in Table 4. (0.49 mm versus 1.01 mm). However, the  A removable implant-supported partial
use of platform-switch did not preserve denture (Figure 12); and
Platform switching concept the crestal bone better than when the  A fixed or removable implant-supported
This concept was based on switching concept was not used, when prosthesis (overdenture) (Figure 13).
clinical observations where the implant thin mucosal tissues on crestal bone were
platform diameter was wider than the present.103 Furthermore, the stress within the
abutment.100 It is assumed that, when this screw-joint was found to increase when the a
principle is used, the crestal bone loss platform-switching concept is implemented.
after implant placement is less than when This may lead to failure of the screw-joint
the implant platform and the abutment connection.104,105 Therefore, this concept
pose a similar diameter.100 This concept should be used with substantial care.
is theoretically explained on the bases
of moving the micro-gap between the Types of restorations/
platform and the abutment inward from prostheses for missing teeth
the outer edge and consequently away Implant-supported restorations
from the bone.101 It also results in an
increase in horizontal soft tissue dimension,
which may protect the bone crest and a b

b C

Figure 11. A clinical image of a missing upper


right centre incisor (1.1) replaced with a single Figure 13. Clinical views of an upper edentulous
cement-retained, implant-supported crown. The Figure 12. (a, b) Clinical images of multiple maxilla restored with a RISO. (a) Four dental
abutment (a) and the restoration (b) is made missing maxillary teeth restored with a partial implants placed in the anterior region. (b)
of porcelain fused to metal. The papilla failed denture which gains its support/retention from The implants are connected with a CAD/CAM
to fill the inter-dental space on the mesial and the teeth, alveolar ridge as well as from an designed and fabricated bar. Four locator
distal aspect of the restoration. This may have a implant placed in the right canine region. The attachments (matrices) are attached to the bar.
negative effect on the aesthetic outcome if the fitting surface of the denture showing the patrix (c) The fitting surface of the RISO showing the
patient has a high lip-line. of a locator attachment. patrices of the attachment.

610 DentalUpdate July/August 2017


ImplantDentistry

Treatment options for principle, this type of restoration resembles by several implants (usually four or more).
replacement of missing teeth with dental that described for a single-implant When such prostheses are indicated, it is a
implants are shown in Figure 14. supported crown: cement- or screw-retained favourable option for many patients. The
1. An implant-supported single restoration restorations. volume of the prosthesis, and consequently
(crown) 3. A removable implant-supported prosthesis the tissue coverage by the prosthesis, are
When a single tooth is replaced, In certain clinical situations, reduced. However, this type of prosthesis is
the restoration is usually either cemented multiple missing teeth cannot be restored more expensive than removable ones. It also
to the abutment or screwed to the implant with a fixed implant-supported restoration. requires more implants to support and retain
(Figure 11). This is known as a cement- Instead, they are restored with a removable the prosthesis.
FISOs are of two basic types:
retained restoration and a screw-retained prosthesis which is fundamentally similar to
hybrid and porcelain fused to metal.
restoration, respectively. As mentioned that which is used in replacing a completely
The hybrid prosthesis is made of a metal
earlier, in the cement-retained restoration, edentulous jaw with a removable implant-
substructure, acrylic and denture teeth. The
the abutment is attached to the implant supported overdenture (RISO) (Figure 12). In
porcelain fused to metal prosthesis is made
body through a screw-joint and the this case, in addition to the available teeth,
of a metal substructure and porcelain in a
restoration is cemented to the abutment one or more implants with attachment
similar way to that used in the fabrication of
in a similar fashion to that which is used systems are usually used. The attachment
the conventional porcelain-fused-to-metal
in the conventional crown. Therefore, the systems are discussed later in the article.
restoration. It is more expensive than the
abutment is used to connect the crown
hybrid and is difficult to make, but it is the
to the implant. In the screw-retained 4. Implant-supported overdenture for better option when the vertical restorative
implant restorations, the restoration and completely edentulous jaws space is limited.
the abutment are a single unit which When the jaw is completely Conversely, the RISOs are
is attached to the implant directly by a edentulous, there are two treatment options removable prostheses that can be removed
screw.16,106,107 for its restoration; namely a fixed or a and replaced by the patients. They are used
removable implant-supported overdenture in combination with attachment systems
2. A fixed implant-supported prosthesis (fixed (FISO or RISO). A FISO is when the prosthesis (see below).
bridge) is permanently fixed to the implants through The number of implants
This is when multiple teeth are screw-joints between the prosthesis and the used with the RISOs may be reduced.
missing and replaced with a prosthesis implants.108 This is so it cannot be removed For instance, in the case of edentulous
that cannot be removed by the patient. In by the patient. The prosthesis is supported mandible, the number may be reduced to

Missing tooth Missing multiple Missing multiple teeth Completely edentulous jaw
adjacent teeth but not adjacent

Single Fixed Removable implant-


implant- implant- supported denture
supported supported
crown bridge

Fixed implant-supported Removable implant-supported


overdenture (FISO) overdenture (RISO)

Hybrid

Screw-retained Cement-retained
Porcelain fused to metal

Figure 14. Treatment options for replacement of missing teeth with dental implants.

July/August 2017 DentalUpdate 611


ImplantDentistry

two implants, which are usually placed in a similar fashion as that covered when the the restoration form one unit. In addition,
the anterior region of the mandible. The conventional complete denture is used. there are five types of abutments which are
two-implant supported overdenture option available for use in single and fixed implant-
is recommended as the first-choice standard The abutments supported restorations.16,35 A summary of
of care for an edentulous mandible.108-110 these abutment types can be found in Table
The restorations that consist of
When two-implant supported overdentures 5.
crowns or fixed prostheses (bridges), and
are used, the attachments permit movement that are supported by implants, may be
of the overdenture during function and divided into two types, depending on how Screw-retained restorations
allow the mucosa of the residual ridge to they are connected to the implants; cement- In this case, the retention of
be involved in dissipating the imposed retained and screw-retained. As mentioned the restoration relies on the retaining
force. Therefore, it is important to note earlier, in the cement-retained restoration screw. Nevertheless, the restoration can be
that, in order to obtain good support the abutment is required to connect the removed and/or replaced when required,
from the residual ridges, the RISOs should restoration to the implant, while in the without damage or need for a new
extend to cover the supporting tissues in screw-retained restoration the abutment and restoration. The adaptation between the

Custom-made abutments
 They are made of a plastic/wax pattern with/without a metal-machined interface ring;
 The pattern is made (wax) or adjusted (plastic) to the required form, shape and angle;
 The pattern is then used to create a metal abutment in a similar procedure to the conventional lost-wax technique;
 An abutment plastic/wax pattern is attached to the implant analogue, which is submerged in a working cast;
 The restoration is then made to fit the abutment also in the conventional method;
 UCLA plastic patterns are an example of these types of abutments;
 They require an impression of the implant platform.

Pre-machined (prefabricated/ready-made) modifiable metal abutments


 They are prefabricated abutments;
 They are adjustable and modifiable intra- and extra-orally;
 They cannot be used when the implant is placed in an improper position or with improper angulation;
 An impression of the abutment, not the implant, is taken using a manufactured impression coping;
 The conventional crown and bridge procedures are used when provisional or final restorations are made.

Pre-machined (pre-fabricated/ready-made), non-modifiable metal abutments


 They are pre-fabricated abutments that cannot be modified or altered;
 The abutment that is suitable for the specific clinical condition is selected;
 The abutment is attached to the implant body;
 An impression of the abutment, not the implant, is taken using a manufactured impression coping;
 The conventional crown and bridge procedures are used when provisional or final restorations are made.

All-ceramic abutments
 They are made entirely of ceramic;
 They are available in ready-made or customizable forms;
 They are indicated for use in cases when aesthetics are essential, and when thin biotype gingiva exists so that metal show through is
avoided.

CAD/CAM milled abutments


 They are made from a block of titanium or ceramic;
 An implant platform level impression may be required depending on the manufacturers;
 A working cast is fabricated then scanned optically to generate exact 3D images of the region;
 The information is sent to the milling machine to produce the abutment;
 It eliminates certain negative factors that may be associated with the conventional method of abutment fabrication, such as an
improper fit and incorporation of porosity;
 This type of abutment is more expensive than the other abutment types.

Table 5. Different abutment types.

612 DentalUpdate July/August 2017


ImplantDentistry

restoration and the underlying implant is identical to those used in conventional implants together. The other attachment
usually better than that in the case of its restorations. However, it may not be possible types may be used individually and also
cement-retained counterpart. It can be used to remove the cement-retained restorations in combination with the bar system. The
when the vertical restorative space is limited if permanent cementing media is used. attachments are attached to the implant by
as the retention depends on the screw, but Therefore, restorations have to be cut in screws, resulting in a screw-joint. Features
is contra-indicated when mouth opening order to remove them. The removal of excess of attachment systems used for RISOs are
is limited, as the use of the different tools cement may be not possible, which may displayed in Table 6.
required for screwing and torqueing the result in soft tissue problems and to peri-
screws may not be possible. However, the implant disease (see below).16,106,107 Therefore, Peri-implant tissue response
use of a screw-retained restoration may be its use should be avoided when the implant- to bacterial insult and peri-
considered when the implant platform is abutment connection is deeply embedded implant diseases
situated deep sub-mucosally, as complete sub-mucosally, which may preclude its
Despite their high success
removal of cement is not always possible removal. Furthermore, removing the
rate, implant failures are also reported to
when a cement-retained restoration is used. cement is not a predictable procedure and
occur. Several factors that have already
The screw type is not indicated when the may cause the abutment/restoration to be
been mentioned earlier which influence
screw hole is pointed at the labial surface scratched,111 leading to plaque accumulation.
such success should be considered when
as this compromises the aesthetics. Hence, Marginal adaptation between the abutment
treatment is planned.7 The implant may fail
the implant should be placed in its optimal and the restoration may also be inferior
before it is put to function as a result of its
position and angulation to avoid negative to that obtained when the screw-retained
failure to integrate with the peri-implant
effects on aesthetics, otherwise an angled restoration is used. It is also not suitable
tissue during the healing stage. This type of
abutment may provide an acceptable when the vertical restorative space is limited,
failure is categorized as an early failure. The
alternative. In the posterior region, the as retention may be compromised.
implant may also lose its integration and
occlusal morphology of the restoration may
fail at a later stage, months or even years
be difficult to obtain as the hole through The attachment systems after implant placement. This is known as
which the screw is tightened occupies
An attachment is defined as a late failure.113 The criteria for dental implant
a major part of the occlusal table of the
mechanical device used for the fixation, success are displayed in Table 7.
restoration. Furthermore, the access hole
retention and stabilization of a dental One of the complications that
may weaken the porcelain and lead to its
prosthesis.112 It is used with implant- is reported to affect the peri-implant tissue
fracturing. It is important to mention that,
supported removable partial dentures is caused by the inflammatory response of
if screw loosening of one restoration occurs
and overdentures. The attachment usually this tissue to bacteria that forms a biofilm on
in a fixed-implant supported restoration, a
consists of two parts. One part is attached to the implant surface.114,115 It occurs when the
cantilevering effect can arise and put the
the implant, while the other part is attached balance between the host’s defence and the
other abutment, implants, screw and the
to the prosthesis. Five types of attachment bacterial load shifts in favour of the bacteria.
peri-implant bone at risk as they are exposed
systems are available and compatible with This tissue response may be limited to the
to tremendous forces. Also, the screw
the main implant systems. The attachment peri-implant soft tissues (mucosa) or may
loosening is not an unreal problem with the
systems that are commonly used with RISOs also extend to and affect the peri-implant
screw-retained restoration. However, the
include: bar/clip, balls, locators, magnet bone and lead to its resorption.
ability to retrieve the restoration/prosthesis
and telescopic crown.108,112 The use of a Both tissue responses to
easily to allow its cleaning (and of the peri-
bar system allows splinting of two or more bacterial insult are collectively known as
implant tissues) is a significant advantage of
screw-retained restorations.
1. The different designs of the attachment systems are used to gain retention, support
Cement-retained restorations and stability of the overdenture.
The cement-retained restoration 2. They consist of a matrix (female) and a patrix (male):
is indicated when mouth opening is  The matrix accommodates the patrix; and
restricted, and when the implant angulation  The patrix frictionally fits and engages the matrix.
is not optimal without a major negative 3. The joint that is made between the patrix and the matrix may be rigid (when
effect on the aesthetic outcome of the no movements exist between the patrix and matrix) or resilient (when there are
restoration.106-108 The occlusal morphology movements).
can be easily constructed in the normal way, 4. The involved dental implants are either splinted or non-splinted.
as in conventional restorations. The materials 5. A bar is usually used to connect the implants (splinted).
and techniques used for the fabrication 6. Bars may be custom-made, pre-fabricated (ready-made) or CAD/CAM milled.
of the cement-retained restoration are 7. An individual attachment system is usually used in a non-splinted manner or combined
similar to those used in the fabrication of with a bar system.
conventional restorations. The trial stage and
Table 6. Features of attachment systems used for RISOs.
the final cementation procedure are almost
July/August 2017 DentalUpdate 613
ImplantDentistry

1. That an individual, unattached implant is immobile when tested clinically.


2. That a radiograph does not demonstrate any evidence of peri-implant radiolucency.
3. That vertical bone loss is less than 0.2 mm annually following the implant’s first year of service.
4. That individual implant performance is characterized by an absence of signs and symptoms, such as pain, infections, neuropathies,
paresthesia or violation of the mandibular canal.
Table 7. Criteria for dental implant success114

1. Bone remodelling after implant placement


2. Reformation of a ‘biologic width’ loss annually. In another study, an average of
3. Presence of rough/smooth interface 0.9 mm crestal bone was lost during the first
4. Presence of a micro-gap at implant-abutment/restoration interface year and no more than 0.07 mm annually in
5. Surgical trauma the following years.120,121 The exact cause of
6. Occlusal overloading this bone loss is still debatable. Nevertheless,
7. A ‘stress shielding’ phenomenon the current literature presents several factors
8. Incomplete removal of luting cement which may contribute to this loss, such as
9. Peri-implant disease surgical trauma, reformation of a ‘biologic
width’ and presence of a rough/smooth
Table 8. Factors that may contribute to or cause crestal bone loss.6
interface. However, the factors that are
most commonly cited to cause such bone
resorption are displayed in Table 8.
peri-implant diseases, and are classified as periodontal ligament, and to the surface
peri-implant mucositis or peri-implantitis.116 topographies of the involved implants which Role of the patient and the
In peri-implant mucositis, the inflammatory facilitate the spread of infection apically dental professionals
response is not essentially different from as well as laterally.6,117 The extent and the Each dental implant and
that which occurs in gingiva when it is composition of cells in the peri-implantitis, restoration/prosthesis should be evaluated
exposed to pathogenic bacteria and leads as well as its progression rate, may differ clinically and radiographically in a similar
to gingivitis.117 Therefore, in principle, peri- from that which is commonly seen in manner to the treatment of periodontal
implant mucositis resembles gingivitis. chronic periodontitis.6,7.117 For instance, the disease. Oral hygiene should be observed
The onset and progression of protective connective tissue capsule, which and regular check-ups should be
mucositis may be affected by a decrease was found to separate the periodontal lesion scheduled. Therefore, after a physiologic
in the vascularity and an increase in from the alveolar bone around teeth in the tissue remodelling period and at the time
collagen to fibroblast ratio in the peri- case of chronic periodontitis, does not exist of prosthesis installation, clinical and
implant connective tissue, and by the way around implants.7 Therefore, the self-limiting radiographic examinations of the peri-
they are arranged around the implant process is not present around implants, implant tissue should be carried out and
surface.117 Clinically, peri-implant mucositis which may provide an explanation for the used as a baseline to monitor any change
is characterized with bleeding on gentle fast development and progression of the in the tissue and to intervene if required.
probing. It is a treatable disease and the peri-implant disease. When any deviation from the norm is
damage is reversible. However, it may It should be mentioned that found, intervention is then considered and
progress into peri-implantitis if untreated.6,7 dental implants may fail as a result of these carried out. In general, oral hygiene should
There are no major differences in the diseases if they are not treated as they be monitored and different oral hygiene
bacteria that were found to be associated lead to bone resorption, and eventually aids should be demonstrated and the
with mucositis and peri-implantitis. This may to mobility and failure of the affected patient encouraged to use them as often as
indicate that mucositis is the origin of peri- implant.8-10 required.
implantitis.117 It is important to remember In general, care for dental
On the other hand, peri- that resorption of peri-implant crestal implants has two phases: patient self-care
implantitis occurs when both the peri- bone occurs within the first year of implant and professional clinical maintenance
implant mucosa and bone are affected. placement and continues to occur to a lesser aspects.122 It is the responsibility of the
It resembles chronic periodontitis in degree afterwards. It occurs irrespective of patient to maintain good oral hygiene.
natural teeth. However, some differences the implant placement method (sub-merged Patient self-care consists of a daily oral
do exist. For instance, the crestal bone or non-submerged). Based on a 15-year hygiene procedure in which toothbrush
loss occurs in a circumferential fashion retrospective study, Adell and colleagues119 (manual/powered and single tufted ones),
around the affected implant, unlike bone reported that crestal bone loss during auxiliary aids such as inter-proximal
resorption seen in chronic periodontitis. The the healing period and the first year after brushes, dental floss/tape and mouthrinses
circumferential shape of the peri-implantitis connecting the prosthesis, was about 1.5 may be used. A combination of these
lesions may be attributed to the lack of mm. Thereafter, there was only 0.1 mm bone aids, whenever it is necessary, should be
614 DentalUpdate July/August 2017
ImplantDentistry

considered and demonstrated. For instance, is made. Accordingly, a treatment plan When the screw is over-torqued
powered toothbrushes, which have different and continuous follow-up strategy are to a degree which places the screw material
interchangeable bristle heads (flattened, constructed. A summary of this protocol in tensile stress that exceeds its elastic limit,
rubber cup-like, short- and long-pointed in is presented in Table 9. However, the the screw may be plastically elongated.
shape) that suit different clinical situations management of peri-implant diseases is not This leads to screw loosening or even to its
may be used. When they are used properly, within this article’s scope. fracture. In the former situation, the screw
the result is a healthy environment around may be replaced, but in the latter situation
the implant. However, it is important Complications associated with the removal of the screw may not be
to mention that limiting the number of implant-supported restorations possible and the treatment is complicated,
auxiliary aids, their simplicity and the time and prostheses which is beyond the scope of this article. To
required for their use are important for Several biological and minimize the occurrence of screw-loosening
patients’ compliance as they play a vital role mechanical complications are reported or fracture, the recommended torque should
in this aspect.123 with the use of dental implants to support/ be implemented using a torque driver that
As already mentioned, dental retain restorations and prostheses. For ensures that the right amount of torque is
implants are affected by and may fail as a instance, screws used to connect different achieved.33
result of the peri-implant disease which combinations of the implant-supported Mechanical superstructure
can be detected only by regular clinical restorations/prostheses may become loose failure may also occur when the material’s
and radiographic examinations. Therefore, and need to be retightened or replaced. mechanical properties and/or thickness is
when an implant is affected by the peri- Screw loosening may be due to it not being not optimum or when the occlusal design is
implant disease, the patient should be made adequately torqued or over-torqued or due not correctly designed. The superstructure
aware of the situation and a treatment plan to micro-movements that occur as a result of failure may also occur as a result of lack
should be implemented and regular follow- the manufacturing tolerance.33,35 An under- of passivity when several implants are
ups arranged. However, there is a lack of torqued screw fails to deliver the tension connected together. The lack of passivity
consensus on how peri-implant disease that is required to produce the optimum may overload the implants and place the
is treated. Nevertheless, the Cumulative clamping force between the screw-joint superstructure under tremendous pressure,
Interceptive Supportive Therapy (CIST) components. Re-tightening is there for that may lead to its failure. To check for
protocol that was presented by Lang required. Screw re-tightening can be easily passivity a test called a ‘Sheffield test’ or
and colleagues123 may be followed when achieved when the restoration is a screw- a ‘one-screw test’ is usually carried out.
peri-implant disease is found. The CIST is retained type. However, when the restoration However, the passivity problem may be
a systemic comprehensive protocol. This is cement-retained, cutting the restoration avoided by the use of computer-aided
protocol is based on clinical parameters to gain access to the screw may be the only design/computer-aided manufacture (CAD/
such as peri-implant pocket depth (PIPD), solution, especially when permanent cement CAM) technology.
bleeding on probing (BoP) and peri-implant is used. When a provisional cement is used, Acrylic or porcelain veneer may
bone loss on which clinical diagnosis the use of crown removal may be tried.35 also fail when the bulk of these materials

Clinical parameters Clinical Diagnosis Treatment Protocols

*PIPD (shallow), Healthy peri-implant tissues No treatment is needed, just regular check-ups and
No plaque enhancement of oral hygiene
No **BoP
*PIPD (shallow) Mucositis A. Mechanical debridement and polishing using a
Plaque is present rubber cup and non-abrasive paste and regular check-
**BoP is present ups and enhancement of oral hygiene

*PIPD ≤5 mm Mucositis B. Treatment includes treatment A with antiseptic


cleaning

*PIPD >5 mm associated with bone loss of up to Peri-implantitis C. Same as treatment B in addition to the use of local
2 mm or systemic antibiotic
*PIPD >5 mm associated bone loss >2 mm Severe peri-implantitis D. Same as treatment C combined with surgery
(access flap, resective method or regenerative
technique)
*Peri-Implant Pocket Depth; **Bleeding on Probing
Table 9.The clinical parameters, diagnosis and a summary of the CIST protocol for treatment of peri-implant diseases.123

616 DentalUpdate July/August 2017


ImplantDentistry

are inadequate. For instance, when a is necessary. This intervention consists of promote the patient to look for treatment.
limited vertical restorative space does not the use of scalers with plastic tips to avoid Consequently, professional evaluation and
allow the use of the optimum thickness scratching the implant components.33 assessment are required to discover such
of the material. Depending on the degree conditions. This necessitates recall visits and
of mechanical damage of the restoration/ Conclusion check-ups which allow the dental personnel
prosthesis, fracture of porcelain may be to intervene in the proper time and to rescue
Dental implants are widely
repaired intra-orally using the Co-Jet® system the implant and its restoration/prosthesis.
used and considered as one of the options
(3M ESPE, St Paul, Mn, USA) and composite Therefore, the dental personnel should be
by which missing teeth are replaced. They
resin material. It is considered as a reliable prepared and able to diagnose and to deal
are used successfully to replace single
method for such repairing. Fracture of acrylic with such complications and to refer the
and multiple missing teeth as well as a
may also be repaired using composite resin patients when required.
completely edentulous jaw. The use of
materials. However, when the metal frame-
dental implants are increasing and dental
work is fractured, the only solution is its Acknowledgement
removal. professionals are more likely to see patients
who have implant-supported restorations/ The authors would like to thank
RISO attachment failure and Mr Emmet Ryan (Dublin Dental University
complications are mostly of a mechanical prostheses. Therefore, basic knowledge
of dental implants is necessary for dental Hospital) for providing the images in Figure
nature and include:35 11 and Dr Brendan Grufferty (Dublin Dental
 Fracture of the acrylic base, teeth and personnel. Several factors are known to
University Hospital) for providing the images
retentive clip; affect success of any implant system. These
in Figure 13.
 Reduction of retention as a result of wear factors may be related to features locally,
of the retentive elements or loosening of such as bone quality and quantity. Other
matrices and screws; factors are related to the surgical method References
by which an implant is placed or which are 1. Gokcen-Rohlig B, Yaltirik M, Ozer S, Tuncer
 Fracture or wear of the clip and matrix; ED, Evlioglu G. Survival and success of ITI
 Fracture of solder joints; and related to the implant system used, such implants and prostheses: retrospective
 Dislodgement of the attachments. as length and diameter of the implant. study of cases with 5-year follow-up. Eur J
Furthermore, dental implants are affected by Dent 2009; 3: 42−49.
Wear of the attachment 2. Baig MR, Rajan M. Effects of smoking
component is a problem that may peri-implant diseases which, if not treated, on the outcome of implant treatment: a
reduce the overdentures’ retention and, can cause the implant to fail. It requires literature review. Indian J Dent Res 2007;
continuous monitoring, regular check-ups 18: 190−195.
consequently, a replacement of the worn 3. Zupnik J, Kim S-W, Ravens D, Karimbux
attachment becomes a necessity. Less and may require professional interventions, N, Guze K. Factors associated with dental
prosthetic maintenance was required with the time of intervention being vital. implant survival: a 4-year retrospective
The success of any implant- analysis. J Periodontol 2011; 82:
the splinted (bar/clip) designs than with 1390−1395.
the unsplinted ones.124 Nevertheless, the supported restoration/prosthesis is 4. Abt E. Growing body of evidence on
use of bars may complicate the hygiene dependent on the interaction between survival rates of implant-supported fixed
the patient and the dental personnel. prostheses. Evid Based Dent 2008; 9:
matter125 and it may be associated with 51−52.
a misfit of the framework, which has the Therefore, maintaining good oral hygiene 5. Han HJ, Kim S, Han DH. Multifactorial
potential to generate unwanted stress on the and committing to regular check-ups are the evaluation of implant failure: a 19 year
responsibility of the patient. On the other retrospective study. Int J Oral Maxillofac
attachment, the implant, the retained screw Implants 2014; 29: 303−310.
and also the peri-implant bone. hand, it is the responsibility of the dental 6. Warreth A, Boggs S, Ibieyou N, El-Helali
Relining of the denture is also personnel to examine the implants and R, Hwang S. Peri-implant diseases: an
the restorations/prostheses clinically and overview. Dent Update 2015; 42: 166−184.
required regularly and may need to be 7. Renvert S, Giovannoli J-L. Peri-implantitis.
carried out every few years to compensate radiographically. It is also the responsibility Paris, France: Quintessence International,
for the changes in the alveolar ridge of the dental practitioner to demonstrate 2012.
8. Buser D, Weber HP, Donath K, Fiorellini
that may occur. Failures of the implant- and educate the patient on how to look after JP, Paquette DW, Williams RC. Soft tissue
supported fixed dental prosthesis also the implant and to tailor check-up recall reactions to non-submerged unloaded
occur. The failures include screw loosening visits according to the patient’s needs. titanium implants in beagle dogs.
J Periodontol 1992; 63: 225−235.
and fracture of the superstructure. Speech Mechanical failures associated 9. Weber HP, Buser D, Donath K, Fiorellini JP,
may be affected when tissue loss is severe. with implant-supported restorations/ Doppalapudi V, Paquette DW, Williams RC.
The compensation of lost tissue with prostheses, such as screw loosening or Comparison of healed tissues adjacent
to submerged and non-submerged
acrylic or porcelain is usually required. This fracture and chipping of porcelain veneer unloaded titanium dental implants. A
compensation may lead to an increase and fracture of the superstructure, are not histometric study in beagle dogs. Clin Oral
in plaque accumulation and tissue uncommon. Loss of retention of the implant- Implants Res 1996; 7: 11−19.
10. Cochran DL. The scientific basis for and
inflammation as the oral hygiene procedure supported overdenture are common clinical clinical experiences with Straumann
is compromised. Meticulous effort from the findings which may make the patient seek implants including the ITI Dental Implant
patient is required. Calculus deposition once treatment. On the other hand, plaque System: a consensus report. Clin Oral
Implants Res 2000; 11(Suppl. 1): 33−58.
formed cannot be removed by a daily oral accumulation and mucosal hyperplasia 11. Tarnow DP, Cho SC, Wallace SS. The effect
hygiene. Therefore, professional intervention in the per-implant site do not necessarily of inter-implant distance on the height

July/August 2017 DentalUpdate 617


ImplantDentistry

of inter-implant bone crest. J Periodontol guidelines. J Dent Oral Hyg 2010; 11: trabecular bone. J Biomed Mater Res 1995;
2000; 71: 546−549. 155−162. 29: 1567−1575.
12. Brånemark P-I. Osseointegration and 29. Schwarz F, Alcoforado G, Nelson K, 44. Le Guehennec L, Goyenvalle E, Lopez-
its experimental studies. J Prosthet Dent Schaer A, Taylor T, Beuer F, Strietzel FP. Heredia MA, Weiss P, Amouriq Y, Layrolle
1983; 50: 399−410. Impact of implant-abutment connection, P. Histomorphometric analysis of the
13. Sennerby L, Ericson LE, Thomsen P, positioning of the machined collar/ osseointegration of four different implant
Lekholm U, Astrand P. Structure of the microgap, and platform switching on surfaces in the femoral epiphyses of
bone-titanium interface in retrieved crestal bone level changes. Camlog rabbits. Clin Oral Implants Res 2008; 19:
clinical oral implants. Clin Oral Implants Foundation Consensus Report. Clin Oral 1103−1110.
Res 1991; 2: 103−111. Implants Res 2014; 25: 1301−1303. 45. Renvert S, Roos-Jansåker AM, Claffey N.
14. Clokie CML, Warshawsky H. 30. Romanos GE, Basha-Hijazi A, Gupta B, Nonsurgical treatment of peri-implant
Morphological and radioautographic Ren YF, Malmstrom H. Role of clinician’s mucositis and peri-implantitis: a literature
studies of bone formation in relation to experience and implant design on review. J Clin Periodontol 2008; 35:
titanium implants using the rat tibia as a implant stability. An ex vivo study in 305−315.
model. Int J Oral and Maxillofac Implants artificial soft bones. Clin Implant Dent 46. Misch CE. Short dental implants: a
1995; 10: 155−165. Relat Res 2014; 16: 166−171. literature review and rationale for use.
15. The glossary of prosthodontic terms. 31. Menicucci G, Pachie E, Lorenzetti M, Dent Today 2005; 24: 64−68.
J Prosthet Dent 2005; 94: 38. Migliaretti G, Carossa S. Comparison of 47. Monje A, Fu JH, Chan HL, Suarez F,
16. Warreth A, Fesharaki H, McConville R, primary stability of straight-walled and Galindo-Moreno P, Catena A, Wang HL.
McReynolds D. An introduction to single tapered implants using an insertion Do implant length and width matter
implant abutments. Dent Update 2013; torque device. Int J Prosthodont 2012; 25: for short dental implants (<10 mm)? A
40: 7−17. 465−471. meta-analysis of prospective studies.
17. Thomas KA. Hydroxyapatite coatings. 32. Alves CC, Neves M. Tapered implants: J Periodontol 2013; 84: 1783−1791.
Orthopaedics 1994; 17: 267−278. from indications to advantages. 48. Misch CE, Steignga J, Barboza E, Misch-
18. Sykaras N, Iacopino AM, Marker VA, J Periodont Rest Dent 2009; 29: 161−167. Dietsh F, Cianciola LJ, Kazor C. Short
Triplett RG, Woody RD. Implant materials, 33. Jacobs SH, O’Connell BC. Dental Implant dental implants in posterior partial
designs, and surface topographies: their Restoration: Principles and Procedures edentulism: a multicenter retrospective
effect on osseointegration. A literature 1st edn. New Malden, UK: Quintessence 6-year case series study. J Periodontol
review. Int J Oral Maxillofac Implants 2000; Publishing, 2001. 2006; 77: 1340−1347.
15: 675−690. 34. Misch CE. Contemporary Implant Dentistry 49. Bahat O. Treatment planning and
19. Higuchi KW, Folmer T, Kultje C. Implant 2nd edn. St Louis: Elsevier, 2008. placement of implants in the posterior
survival rates in partially edentulous 35. Warreth A, McAleese E, McDonnell P, maxillae: report of 732 consecutive
patients: a 3-year prospective multicenter Slami R, Guray SM. Dental implants and Nobelpharma implants. Int J Oral
study. J Oral Maxillofac Surg 1995; 53: single implant-supported restorations. J Ir Maxillofac Implants 1993; 8: 151−161.
264−268. Dent Assoc 2013; 59: 32−43. 50. Winkler S, Morris HF, Ochi S. Implant
20. Geckili O, Bilhan H, Geckili E, Cilingir A, 36. Steigenga J, Al-Shammari K, Misch C, survival to 36 months as related to length
Mumcu E, Bural C. Evaluation of possible Nociti FH Jr, Wang H-L. Effects of implant and diameter. Ann Periodontol 2000; 5:
prognostic factors for the success, thread geometry on percentage of 22−31.
survival, and failure of dental implants. osseointegration and resistance to reverse 51. Jokstad A. The evidence for endorsing
Implant Dent 2014; 23: 44−50. torque in the tibia of rabbits. J Periodontol the use of short dental implants remains
21. Lekholm U, Zarb GA. Patient selection 2004; 75: 1233−1241. inconclusive. Evid Based Dent 2011; 12:
and preparation. In: Tissue-Integrated 37. Rabel A, Köhler SG, Schmidt-Westhausen 99−101.
Prostheses. Osseointegration in Clinical AM. Clinical study on the primary stability 52. Monje A, Chan HL, Fu JH, Suarez F,
Dentistry 1st edn. Brånemark P, Zarb of two dental implant systems with Galindo-Moreno P, Wang HL. Are short
G, Albrektsson T. New Malden, UK: resonance frequency analysis. Clin Oral dental implants (<10 mm) effective? A
Quintessence Publishing Co, 1985: Investig 2007; 11: 257−265. meta-analysis on prospective clinical
195−205. 38. Yoon HG, Heo SJ, Koak JY, Kim SK, Lee trials. J Periodontol 2013; 84: 895−904.
22. Jones AA, Cochran DL. Consequences of SY. Effect of bone quality and implant 53. Lee JH, Frias V, Lee KW, Wright RF. Effect
implant design. Dent Clin North Am 2006; surgical technique on implant stability of implant size and shape on implant
50: 339−360. quotient (ISQ) value. J Adv Prosthodont success rate: a literature review. J Prosthet
23. Kim TH, Lee DW, Kim CK, Park KH, Moon 2011; 3: 10−15. Dent 2005; 94: 377−381.
IS. Influence of early cover screw exposure 39. Cochran DL. A comparison of endosseous 54. Allum SR, Tomlinson RA, Joshi R. The
on crestal bone loss around implants: dental implant surfaces. J Periodontol impact of loads on standard diameter,
intra-individual comparison of bone level 1999; 70: 1523−1539. small diameter and mini implants: a
at exposed and non-exposed implants. 40. Oue H, Doi K, Oki Y, Makihara Y, Kubo T, comparative laboratory study. Clin Oral
J Periodontol 2009; 80: 933−939. Perrotti V, Piattelli A, Akagawa Y, Tsuga K. Implants Res 2008; 19: 553−559.
24. Javed F, Romanos GE. The role of primary Influence of implant surface topography 55. Misch CE, Qu M, Bidez MW. Mechanical
stability for successful immediate loading on primary stability in a standardized properties of trabecular bone in the
of dental implants. A literature review. osteoporosis rabbit model study. J Funct human mandible: implications for dental
J Dent 2010; 38: 612−620. Biomater 2015; 6: 143−152. implant treatment planning and surgical
25. Kim YS, Lim YJ. Primary stability and 41. Novaes AB Jr, Souza SL, de Oliveria PT, placement. J Oral Maxillofac Surg 1999;
self-tapping blades: biomechanical Souza AM. Histomorphometric analysis of 57: 700−706.
assessment of dental implants in the bone-implant contact obtained with 56. Renouard F, Nisand D. Impact of implant
medium-density bone. Clin Oral Implants 4 different implant surface treatments length and diameter on survival rates.
Res 2011; 22: 1179−1184. placed side by side in the dog mandible. Clin Oral Implants Res 2006; 17(Suppl 2):
26. Morris HF, Winkler S, Ochi S, Kanaan A. Int J Oral Maxillofac Implants 2002; 17: 35−51.
A new implant designed to maximize 377−383. 57. Ivanoff CJ, Sennerby L, Johansson C,
contact with trabecular bone: survival 42. Klokkevold PR, Johnson P, Dadgostari S, Rangert B, Lekholm U. Influence of
to 18 months. J Oral Implantol 2001; 27: Caputo A, Davies JE, Nishimura RD. Early implant diameters on the integration of
164−173. endosseous integration enhanced by screw implants. An experimental study in
27. Davies JE. Mechanisms of endosseous dual acid etching of titanium: a torque rabbits. Int J Oral Maxillofac Surg 1997; 26:
integration. Int J Prosthodont 1998; 11: removal study in the rabbit femur. Clin 141−148.
391−401. Oral Implants Res 2001; 12: 350−357. 58. Langer B, Langer L, Herrmann I, Jorneus
28. Warreth A, Ibieyou N, MacCarthy D. 43. Wong M, Eulenberger J, Schenk R, L. The wide fixture: a solution for special
Bisphosphonates, oral implants and Hunziker E. Effects of surface topology bone situations and a rescue for the
osteonecrosis of the jaw: a review and on the osseointegration of implant in compromised implant. Part 1. Int J Oral

618 DentalUpdate July/August 2017


ImplantDentistry

Maxillofac Implants 1993; 8: 400−408. abutments compared with natural tooth: in gold prosthesis-retaining screws as
59. Davarpanah M, Martinez H, Tecuciana microcirculation monitoring as a novel a function of time. Int J Oral Maxillofac
J-F, Celletti R, Lazzara R. Small- bioindicator. Implant Dent 2015; 24: Implants 2004; 19: 124−132.
diameter implants: indications and 37−41. 91. Winkler S, Ring K, Ring JD, Boberick KG.
contraindications. J Esthet Dent 2000; 12: 74. Andreiotelli M, Wenz HJ, Kohal RJ. Are Implant screw mechanics and the settling
186−194. ceramic implants a viable alternative to effect: overview. J Oral Implantol 2003; 29:
60. Suba C, Velich N, Turi C, Szabó G. Surface titanium implants? A systematic literature 242−245.
analysis methods of biomaterials used in review. Clin Oral Implants Res 2009; 92. Kim KS, Han JS, Lim YJ. Settling of
oral surgery: literature review. J Craniofac 20(Suppl. 4): 32−47. abutments into implants and changes in
Surg 2005; 16: 31−36. 75. Özkurt Z, Kazazoğlu E. Zirconia dental removal torque in five different implant-
61. Mouhyi J, Sennerby L, Wennerberg implants: a literature review. J Oral abutment connections. Part 1: Cyclic
A, Louette P, Dourov N, van Reck Implantol 2011; 37: 367−376. loading. Int J Oral Maxillofac Implants 2014;
J. Re-establishment of the atomic 76. Fuentealba R, Jofré J. Esthetic failure in 29: 1079−1084.
composition and the oxide structure of implant dentistry. Dent Clin North Am 93. Tan KB, Nicholls JI. Implant-abutment
contaminated titanium surfaces by means 2015; 59: 227−246. screw-joint pre-load of 7 hex-top
of carbon dioxide laser and hydrogen 77. McGlumphy EA, Mendel DA, Holloway JA. abutment systems. Int J Oral Maxillofac
peroxide: an in vitro study. Clin Implant Implant screw mechanics. Dent Clin North Implants 2001; 16: 367−377.
Dent Relat Res 2000; 2: 190−202. Am 1998; 42: 71−89. 94. Lang LA, May KB, Wang RF. The effect of
62. Cooper LF. A role for surface topography 78. Maeda Y, Miura J, Taki I, Sogo M. the use of a counter-torque device on the
in creating and maintaining bone at Biomechanical analysis on platform abutment-implant complex. J Prosthet
titanium endosseous implants. J Prosthet switching: is there any biomechanical Dent 1999; 81: 411−417.
Dent 2000; 84: 522−534. rationale? Clin Oral Implants Res 2007; 18: 95. Gratton DG, Aquilino SA, Stanford CM.
63. Blumenthal NC, Cosma V. Inhibition 581−584. Micromotion and dynamic fatigue
of apatite formation by titanium and 79. Gracis S, Michalakis K, Vigolo P, Vult von properties of the dental implant-abutment
vanadium ions. J Biomed Mater Res 1989; Steyern P, Zwahlen M, Sailer I. Internal interface. J Prosthet Dent 2001; 85: 47−52.
23(A1 Suppl): 13−22. vs. external connections for abutments/ 96. Haack JE, Sakaguchi RL, Sun T, Coffey JP.
64. Siddiqi A, Payne AG, De Silva RK, Duncan reconstructions: a systematic review. Elongation and pre-load stress in dental
WJ. Titanium allergy: could it affect dental Clin Oral Implants Res 2012; 23(Suppl. 6): implant abutment screws. Int J Oral
implant integration? Clin Oral Implants Res 202−216. Maxillofac Implants 1995; 10: 529−536.
2011; 22: 673−680. 80. Akça K, Cehreli MC, Iplikçioğlu 97. Quek HC, Tan KB, Nicholls JI. Load fatigue
65. Javed F, Al-Hezaimi K, Almas K, Romanos H. Evaluation of the mechanical performance of four implant-abutment
GE. Is titanium sensitivity associated characteristics of the implant-abutment interface designs: effect of torque level
with allergic reactions in patients with complex of a reduced-diameter morse- and implant system. Int J Oral Maxillofac
dental implants? A systematic review. Clin taper implant. A nonlinear finite element Implants 2008; 23: 253−262.
Implant Dent Relat Res 2013; 15: 47−52. stress analysis. Clin Oral Implants Res 2003; 98. Goheen KL, Vermilyea SG, Vossoughi J,
66. Alhag M, Renvert S, Polyzois I, Claffey N. 14: 444−454. Agar JR. Torque generated by handheld
Re-osseointegration on rough implant 81. Segundo RM, Oshima HM, da Silva IN, screwdrivers and mechanical torqueing
surfaces previously coated with bacterial Burnett LH Jr, Mota EG, Silva LL. Stress devices for osseointegrated implants. Int J
biofilm: an experimental study in the dog. distribution of an internal connection Oral Maxillofac Implants 1994; 9: 149−155.
Clin Oral Implants Res 2008; 19: 182−187. implant prostheses set: a 3D finite 99. Khraisat A, Abu-Hammad O, Al-Kayed
67. Renvert S, Polyzois I, Maguire R. element analysis. Stomatologija 2009; 11: AM, Dar-Odeh N. Stability of the implant/
Re-osseointegration on previously 55−59. abutment joint in a single-tooth external-
contaminated surfaces: a systematic 82. Keating K. Connecting abutments hexagon implant system: clinical and
review. Clin Oral Implants Res 2009; to dental implants: ‘an engineer’s mechanical review. Clin Implant Dent Relat
20(Suppl 4): 216−227. perspective’. Irish Dentist 2001; July: Res 2004; 6: 222−229.
68. Gineste L, Gineste M, Ranz X, Ellefterion 43−46. 100. Monje A, Pommer B. The concept of
A, Guilhem A, Rouquet N, Frayssinet 83. Jung SW, Son MK, Chung CH, Kim HJ. platform switching to preserve peri-
P. Degradation of hydroxylapatite, Abrasion of abutment screw coated with implant bone level: assessment of
fluorapatite, and fluorhydroxyapatite TiN. J Adv Prosthodont 2009; 1: 102−106. methodologic quality of systematic
coatings of dental implants in dogs. J 84. Jo JY, Yang DS, Huh JB, Heo JC, Yun reviews. Int J Oral Maxillofac Implants 2015;
Biomed Mater Res 1999; 48: 224−234. MJ, Jeong CM. Influence of abutment 30: 1084−1092.
69. Förster Y, Rentsch C, Schneiders W, materials on the implant-abutment joint 101. de Almeida FD, Carvalho AC, Fontes M,
Bernhardt R, Simon JC, Worch H, Rammelt stability in internal conical connection Pedrosa A, Costa R, Noleto JW, Mourão CF.
S. Surface modification of implants in type implant systems. J Adv Prosthodont Radiographic evaluation of marginal bone
long bone. Biomatter 2012; 2: 149−157. 2014; 6: 491−497. level around internal-hex implants with
70. Wie H, Herø H, Solheim T, Kleven E, Rørvik 85. Satterthwaite J, Rickman L. Retrieval of a switched platform: a clinical case report
AM, Haanaes HR. Bonding capacity fractured abutment screw thread from an series. Int J Oral Maxillofac Implants 2011;
in bone of HIP-processed HA-coated implant: a case report. Br Dent J 2008; 204: 26: 587−592.
titanium: mechanical and histological 177−180. 102. Strietzel FP, Neumann K, Hertel M. Impact
investigations. J Biomed Mater Res 1995; 86. Drago CJ. A clinical study of the efficacy of platform switching on marginal peri-
29: 1443−1449. of gold-tite square abutment screws in implant bone-level changes. A systematic
71. Cook SD, Salkeld SL, Gaisser DM, Wagner cement-retained implant restorations. Int J review and meta-analysis. Clin Oral
WR. The effect of surface macrotexture Oral Maxillofac Implants 2003; 18: 273−278. Implants Res 2015; 26: 342−358.
on the mechanical and histologic 87. Byrne D, Jacobs S, O’Connell B, Houston 103. Linkevicius T, Apse P, Grybauskas S, Puisys
characteristics of hydroxylapatite-coated F, Claffey N. Pre-loads generated with A. Influence of thin mucosal tissues on
dental implants. J Oral Implantol 1993; repeated tightening in three types of crestal bone stability around implants with
19(4): 288−294. screws used in dental implant assemblies. platform switching: a 1-year pilot study. J
72. Osman RB, Elkhadem AH, Ma S, Swain J Prosthodont 2006; 15: 164−171. Oral Maxillofac Surg 2010; 68: 2272−2277.
MV. Titanium versus zirconia implants 88. Binon PP. Evaluation of three slip fit 104. Maeda Y, Satoh T, Sogo M. In vitro
supporting maxillary overdentures: hexagonal implants. Implant Dent 1996; 5: differences of stress concentrations
three-dimensional finite element analysis. 235−248. for internal and external hex implant-
Int J Oral Maxillofac Implants 2013; 28: 89. Binon PP, McHugh MJ. The effect of abutment connections: a short
e198−208. eliminating implant/abutment rotational communication. J Oral Rehabil 2006; 33:
73. Kajiwara N, Masaki C, Mukaibo T, Kondo misfit on screw-stability. Int J Prosthodont 75−78.
Y, Nakamoto T, Hosokawa R. Soft tissue 1996; 9: 511−519. 105. Cimen H, Yengin E. Analyzing the effects
biological response to zirconia and metal 90. Cantwell A, Hobkirk JA. Pre-load loss of the platform-switching procedure

July/August 2017 DentalUpdate 619


ImplantDentistry

on stresses in the bone and implant- 112. Burns DR. Mandibular implant reactions at osseointegrated titanium
abutment complex by 3-dimensional fem overdenture treatment: consensus and fixtures. Swed Dent J 1985; 28(Suppl):
analysis. J Oral Implantol 2012; 38: 21−26. controversy. J Prosthodont 2000; 9: 37−46. 175−181.
106. Hebel KS, Gajjar RC. Cement-retained 113. Sakka S, Baroudi K, Nassani MZ. Factors
associated with early and late failure of 121. Adell R, Lekholm U, Rockler B,
versus screw-retained implant restorations:
achieving optimal occlusion and aesthetics dental implants. J Investig Clin Dent 2012; Brånemark PI, Lindhe J, Eriksson B,
in implant dentistry. J Prosthet Dent 3: 258−261. Sbordone L. Marginal tissue reactions at
1997; 77: 28−35. 114. Albrektsson T, Zarb G, Worthington P, osseointegrated titanium fixtures (I). A
107. Newsome P, Reaney D, Owen S. Screw- Eriksson AR. The long-term efficacy of 3-year longitudinal prospective study. Int
versus cement-retained crowns. Irish currently used dental implants: a review J Oral Maxillofac Surg 1986; 15: 39−52.
Dentist 2011; March: 22−25. and proposed criteria of success. Int J Oral
108. Warreth A, Ramadan M, Bajilan MR, Maxillofac Implants 1986; 1: 11−25. 122. Kracher CM, Smith WS. Oral health
Ibieyou N, El-Swiah J, Elemam RF. 115. Socransky SS, Haffajee AD. Dental maintenance dental implants. Dent Assist
Fundamentals of occlusion and biofilms: difficult therapeutic targets. 2010; 79: 27−35.
restorative dentistry. Part I: basic Periodontol 2000 2002; 28: 12−55. 123. Lang NP, Berglundh T, Heitz-Mayfield
principles. J Ir Dent Assoc 2015; 61: 116. Marsh PD. Dental plaque: biological LJ, Pjetursson BE, Salvi GE, Sanz M.
201−208. significance of a biofilm and community Consensus statements and recommended
109. Feine JS, Carlsson GE, Awad MA, Chehade life-style. J Clin Periodont 2005; 32(Suppl
A, Duncan WJ et al. The McGill consensus 6): 7−15. clinical procedures regarding implant
statement on overdentures. Mandibular 117. Mombelli A, Lang NP. The diagnosis and survival and complications. Int J Oral
two-implant overdentures as first choice treatment of peri-implantitis. Periodontol Maxillofac Implants 2004; 19(Suppl):
standard of care for edentulous patients. 2000 1998; 17: 63−76. 150−154.
Montreal, Quebec, May 24−25, 2002. 118. Lang NP, Berglundh T. Periimplant 124. Stoumpis C, Kohal RJ. To splint or not
Int J Oral Maxillofac Implants 2002; 17: diseases: where are we now? − Consensus to splint oral implants in the implant-
601−602. of the Seventh European Workshop on
110. British Society for the Study of Prosthetic Periodontology. J Clin Periodontol 2011; supported overdenture therapy? A
Dentistry. The York consensus statement 38(Suppl 11): 178−181. systematic literature review. J Oral Rehabil
on implant-supported overdentures. 119. Adell R, Lekholm U, Rockler B, Brånemark 2011; 38: 857−869.
Eur J Prosthodont Restor Dent 2009; 17: PI. A 15-year study of osseointegrated 125. Warreth A, Byrne C, Alkadhimi AF,
164−165. implants in the treatment of the Woods E, Sultan A. Mandibular implant-
111. Agar JR, Cameron SM, Hughbanks edentulous jaw. Int J Oral Surg 1981; 10: supported overdentures: attachment
JC, Parker MH. Cement removal from 387−416.
restorations luted to titanium abutments 120. Adell R, Lekholm U, Brånemark PI, Lindhe systems, and number and locations of
with simulated subgingival margins. J, Rockler B, Eriksson B, Lindvall AM, implants − Part II. J Ir Dent Assoc 2015; 61:
J Prosthet Dent 1997; 78: 43−47. Yoneyama T, Sbordone L. Marginal tissue 144−148.

620 DentalUpdate July/August 2017


View publication stats

You might also like