Jurnal CBCT 1

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Azin Parsa Bone quality evaluation at dental

Norliza Ibrahim
Bassam Hassan
implant site using multislice CT,
Paul van der Stelt micro-CT, and cone beam CT
Daniel Wismeijer

Authors’ affiliations: Key words: bone density, bone volume fraction, dental implants, multislice computed tomog-
Azin Parsa, Norliza Ibrahim, Bassam Hassan, Paul raphy, micro-computed tomography, cone beam computed tomography
van der Stelt, Section of Oral Radiology,
Department of General and Specialized Dentistry,
Academic Center for Dentistry Amsterdam Abstract
(ACTA), Amsterdam, The Netherlands
Norliza Ibrahim, Department of General Dental
Objectives: The first purpose of this study was to analyze the correlation between bone volume
Practice and Oral & Maxillofacial Imaging, Faculty fraction (BV/TV) and calibrated radiographic bone density Hounsfield units (HU) in human jaws,
of Dentistry, University of Malaya, Kuala Lumpur, derived from micro-CT and multislice computed tomography (MSCT), respectively. The second aim
Malaysia
Daniel Wismeijer, Section of Implantology and was to assess the accuracy of cone beam computed tomography (CBCT) in evaluating trabecular
Prosthetic Dentistry, Department of Oral Function bone density and microstructure using MSCT and micro-CT, respectively, as reference gold
and Restorative Dentistry, Academic Center for standards.
Dentistry Amsterdam (ACTA), Amsterdam, The
Netherlands Material and methods: Twenty partially edentulous human mandibular cadavers were scanned by
three types of CT modalities: MSCT (Philips, Best, the Netherlands), CBCT (3D Accuitomo 170, J
Corresponding author: Morita, Kyoto, Japan), and micro-CT (SkyScan 1173, Kontich, Belgium). Image analysis was
Azin Parsa, DDS, MSc
Section of Oral Radiology, Department of General performed using Amira (v4.1, Visage Imaging Inc., Carlsbad, CA, USA), 3Diagnosis (v5.3.1, 3diemme,
and Specialized Dentistry, Cantu, Italy), Geomagic (studioâ 2012, Morrisville, NC, USA), and CTAn (v1.11, SkyScan). MSCT,
Academic Center for Dentistry Amsterdam (ACTA)
CBCT, and micro-CT scans of each mandible were matched to select the exact region of interest
Gustav Mahlerlaan 3004, 1081 LA Amsterdam,
The Netherlands (ROI). MSCT HU, micro-CT BV/TV, and CBCT gray value and bone volume fraction of each ROI were
Tel.: +31 20 598 0834 derived. Statistical analysis was performed to assess the correlations between corresponding
Fax: +31 20 598 0333
measurement parameters.
e-mail: a.parsa@acta.nl
Results: Strong correlations were observed between CBCT and MSCT density (r = 0.89) and
between CBCT and micro-CT BV/TV measurements (r = 0.82). Excellent correlation was observed
between MSCT HU and micro-CT BV/TV (r = 0.91). However, significant differences were found
between all comparisons pairs (P < 0.001) except for mean measurement between CBCT BV/TV and
micro-CT BV/TV (P = 0.147).
Conclusions: An excellent correlation exists between bone volume fraction and bone density as
assessed on micro-CT and MSCT, respectively. This suggests that bone density measurements could
be used to estimate bone microstructural parameters. A strong correlation also was found between
CBCT gray values and BV/TV and their gold standards, suggesting the potential of this modality in
bone quality assessment at implant site.

Introduction one of the most important factors influencing


primary implant stability (Ozan et al. 2007;
Primary implant stability is the key factor for Tolstunov 2007). Among the bone character-
the long-term success of an implant treat- istics, bone mineral density (BMD) and tra-
ment by improving osseointegration (Fuh becular microstructure are the strongest
et al. 2010). Primary instability of an implant predictors for bone strength (Muller 2003).
induces movements during healing. This mi- However, these two parameters need to be
cromotion leads to fibroplasia as a biological simultaneously assessed to provide better
response at bone tissue surrounding the estimation of bone strength (Teo et al. 2007;
Date: implant. The replacement of bone by fibrous Diederichs et al. 2009). Several radiographic
Accepted 9 November 2013
tissue and loss of osseointegration cause modalities have been used for bone quality
To cite this article: implant failure (Lioubavina-Hack et al. 2006). assessment. For bone microstructure, micro-
Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D.
Bone quality evaluation at dental implant site using Bone quality, which refers to the combina- computed tomography (micro-CT) was rec-
multislice CT, micro-CT and cone beam CT.
tion of all bone characteristics that influence ommended as gold standard for assessing
Clin. Oral Impl. Res. 00, 2013, 1–7
doi: 10.1111/clr.12315 bone resistance to fracture (Fyhrie 2005), is bone morphology and micro-architecture

© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 1
Parsa et al  CT bone quality assessment at implant site

(Burghardt et al. 2011; Ibrahim et al. 2013a). between BV fraction derived from CBCT and Image processing
However, it is limited to ex vivo small bone CT numbers from MSCT has been reported All CT data sets were converted to Digital
samples and cannot be employed for patients. (Naitoh et al. 2010a; Gonzalez-Garcıa & Imaging and Communication in Medicine
Multiple X-ray projections with different Monje 2012). However, the relation between (DICOM3) format. As the scan orientation
angles in micro-CT allow a precise three- BV fraction and radiographic bone density in differs between micro-CT and the other two
dimensional (3D) reconstruction of the bone human jaws remains controversial (Stoppie systems, the z-axis of CBCT and MSCT
samples and assessment of bone trabeculae et al. 2006; Aksoy et al. 2009). Therefore, images was flipped to match that of micro-
(Martin-Badosa et al. 2003). Micro-CT is used the first purpose of this study was to ana- CT for further procedures. Micro-CT data
to measure several histomorphometric vari- lyze the correlation between bone volume sets were large in size therefore was not pos-
ables including bone volume (BV), total vol- fraction (BV/TV) and calibrated radiographic sible to be flipped by our workstation. Image
ume (TV), bone volume fraction (BV/TV), bone density (HU) in human jaws, derived analysis was performed using Amira (v4.1,
trabecular thickness (Tb.Th), trabecular num- from micro-CT and MSCT, respectively. The Visage Imaging Inc., Carlsbad, CA, USA),
ber (Tb.N), and trabecular separation (Tb.Sp) second aim was to assess the accuracy of 3Diagnosis (v5.3.1, 3diemme, Italy), Geoma-
(Odgaard 1997). CBCT in evaluating trabecular bone density gic (studioâ 2012, Morrisville, NC, USA), and
For bone density, multislice computed and microstructure using MSCT and CTAn (v1.11, SkyScan). MSCT images were
tomography (MSCT) is an established clinical micro-CT, respectively, as reference gold imported to 3Diagnosis software. Two cylin-
modality in which calibrated Hounsfield standards. drical shape virtual probes (with diameter
units (HU) can accurately be converted to and height of 0.7 and 8 mm, respectively)
BMD measurements (Shahlaie et al. 2003; were inserted at the edentulous region within
Material and methods
Shapurian et al. 2006). However, higher radia- the cancellous bone, with 3 mm buccolin-
tion exposure risk to patients in comparison gual distance between them (Fig. 1a, b).
Sample preparation and radiographic
with other modalities remains a main con- evaluation These probes were used as indicators to facil-
cern for applying MSCT for assessing bone Twenty partially edentulous human mandib- itate the selection of exact region of interest
quality (Ekestubbe et al. 1992, 1993; Freder- ular cadavers not identified by age, sex, or (ROI) from MSCT, CBCT, and micro-CT. For
iksen et al. 1995; Dula et al. 1996). Cone ethnic group were obtained from the func- MSCT, the probes were visible in a single
beam computed tomography (CBCT), due to tional anatomy department. The cadavers cross-sectional slice as the voxel size of
increased accessibility to dental practitioners, were sectioned at the mid-ramus level and MSCT scans was 0.67 mm, which is thick
more compact equipment and reduced cost fixed in formaldehyde (formaldehyde 74.79%, enough to allow the probes to be visible in
and radiation dose, has widely replaced medi- glycerol 16.7%, alcohol 8.3%, and phenol one slice. Subsequently, a rectangular area
cal CT for oral and maxillofacial imaging. 0.21%) and stored. A declaration was was drawn between the two probes from the
Several studies reported high geometric accu- obtained from the Functional Anatomy slice of interest to define the ROI for density
racy of CBCT for linear measurement (Nai- department to use this human remains mate- measurements. The mean HU values from
toh et al. 2004; Lou et al. 2007; Lagravere rial for research purposes. The restorative each ROI were calculated. All ROIs were
et al. 2008), while its reliability in bone qual- materials which can produce artifact such as totally within the cancellous bone, excluding
ity evaluation remains controversial. Only amalgam filling and metal crowns were cortical bone, inferior dental canal, and any
few studies suggested that CBCT could be removed from dentitions. The mandibles large bone defect.
applied to assess trabecular bone microstruc- were scanned by three types of CT modali- For CBCT, a volume-based 3D registration
ture (Liu et al. 2007; Corpas et al. 2011). ties: MSCT (Philips, 120 kVp, 222 mA, algorithm using Geomagic software was
Additionally, CBCT does not represent cali- 1.128 s, 0.67 mm isotropic voxel size, Best, applied to transform the inserted probes from
brated voxel gray values expressed in HU the Netherlands), CBCT (3D Accuitomo 170, the MSCT data sets to the CBCT scans. A
(Hua et al. 2009). Yet, many attempts have 90 kVp, 5 mA, 30.8 s, 4 9 4 cm FOV, 0.08 standard triangulation language (STL) surface
been conducted to assess the feasibility of mm isotropic voxel size, J Morita, Kyoto, file of the MSCT and CBCT scans were
converting CBCT gray values to actual den- Japan), and micro-CT (SkyScan 1173, 130 matched, and the probes were transferred
sity measurements. High correlation between kVp, 61 mA, 35 min, 35 lm isotropic voxel from MSCT scans to the exact region on
HU derived from MSCT and CBCT voxel size, Kontich, Belgium). In MSCT scans, the CBCT’s (Fig. 1c). As a result, new CBCT data
gray values has been demonstrated, hinting occlusal plane of each mandible was set sets which include the probes were obtained.
at the potential of CBCT in bone density perpendicular to the floor with zero gantry Using 3Diagnosis, eight consecutive slices
assessment (Aranyarachkul et al. 2005; tilt, whereas in CBCT scans, it was set passing through the probes were selected
Lagrav ere et al. 2006; Naitoh et al. 2009, parallel to the floor according to manufac- from each CBCT data set to calculate the
2010b; Nomura et al. 2010; Parsa et al. 2012; turer’s recommended protocol. The edentu- mean gray values (radiological density). This
Reeves et al. 2012; Cassetta et al. 2013). lous region of each mandible was located at is because slice thickness in CBCT is
However, the excessive scattering and tech- the center of FOV in CBCT scans. Owing to 0.08 mm which approximately amounts to
nology-specific artifacts produced in CBCT the large gantry of applied micro-CT 89 thinner than the equivalent slice thick-
have been denoted as the perpetrator for the (140 mm in diameter, 200 mm in height), ness in MSCT. A rectangular region was also
unreliable BMD measurements (Yoo & Yin mandibles were not sectioned to smaller drawn between the two probes similar to
2006; Hua et al. 2009; Araki & Okano 2011; samples. To prevent the possible micromove- MSCT and gray values from corresponding
Nackaerts et al. 2011; Schulze et al. 2011). ments during the scanning due to the large anatomical locations were derived.
High correlation between bone volume size of the samples, a cylindrical shape Styro- Cone beam computed tomography radiolog-
fraction (BV/TV) provided by micro-CT and foam was used to fix and mount the sample ical density of each mandible’s ROI was con-
voxel gray value from CBCT, and also into the holder. sidered as the mean of eight calculated gray

2 | Clin. Oral Impl. Res. 0, 2013 / 1–7 © 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd
Parsa et al  CT bone quality assessment at implant site

(a) (b) (c)

Fig. 1. (a) Three-dimensional reconstruction of a mandible multislice computed tomography (MSCT) scan with inserted probes. (b) Close-up view of probes and (C) three-dimen-
sional reconstruction of a mandible cone beam computed tomography (CBCT) scan with transferred probes.

values. Subsequently, the selected ROIs were correlation coefficient was used to assess the CBCT and micro-CT BV/TV measurements
saved as a bitmap (BMP) image files to allow linear relation between corresponding (r = 0.82). Excellent correlation was observed
the trabecular microstructure evaluation. measurement parameters. Finally, a Bland– between MSCT HU and micro-CT BV/TV
Using Amira, each micro-CT scan was Altman plot was used to assess the accuracy (r = 0.91). Bland–Altman analysis showed the
cropped to have a smaller sample including of CBCT in measuring trabecular BMD and bias in measuring BV/TV between CBCT and
the ROI. Due to large micro-CT data sets, bone microstructural density by plotting the micro-CT is smaller (4.44/lm) than measur-
the superimposition of CBCT and micro-CT difference between the measurements of ing the density between CBCT and MSCT
scans was performed as follows: maximum CBCT against MSCT density and micro-CT (154.65HU) (Fig. 3a, b). The 95% measure-
alignment of both data sets was obtained by BV/TV against the means of the compared ment errors are between 21.31 to 30.19 for
manually matching and superimposition of measurements. BV/TV and 29.74–279.56 for density measure-
isosurfaces generated in Amira software. Sub- ment. The differences of CBCT and micro-
sequently, sixteen micro-CT slices (corre- CT BV/TV measurements were minimal
spondence to the eight CBCT slices) were Results (4.44/lm), suggesting strong agreement.
selected and saved as a 16-BMP image file
(65536 gray values). Then, these bmp files Excellent intra-observer reliability (ICC ≥
were exported to CTAn software for trabecu- 0.97) was revealed for repeated measurements Discussion
lar microstructure evaluations (Fig. 2a). A in the three systems. Therefore, the mean of
rectangular ROI for trabecular was selected two measurements was calculated for further It has been proven that the success of an
on each data set slice by slice (Fig. 2b). All analysis. The mean HU of the selected ROI inserted implant strongly depends on the
images were thresholded using an automated ranged from 60 to 507.6 (mean 222.85 & quality, beside the quantity, of the sur-
histogram analysis and binarized (Fig. 2c) to standard deviation [SD] 140.5), while CBCT rounded bone (Jaffin & Berman 1991; Jemt
allow the measurement process. On micro- gray values ranged from 161.6 to 665.6 (mean et al. 1992). In jawbones, density measure-
CT data sets, the ROI was again verified by 377.49 & SD 127.4). The negative HU derived ments derived from MSCT HU are highly
carefully comparing slices with CBCT’s (as from MSCT for case 4, 16, and 20 (Table 1) reliable (Schwarz et al. 1987; Shapurian et al.
reference). This was performed to reduce bias may indicate fat in trabecular spaces (Parsa 2006). However, bone density alone does not
which may have been introduced during the et al. 2012). Calculated BV/TV of the same fully represent bone quality and should be
manual superimposition of the two data sets. ROI ranged from 2.24 to 75.83 (mean 32.35 & considered together with bone micro-archi-
All measurements were performed twice SD 18.81) for micro-CT and from 3.73 to tecture to estimate bone strength and frac-
with 1-month interval by a trained maxillofa- 68.72 (mean 36.79 & SD 23.17) for CBCT ture resistance (Diederichs et al. 2009).
cial radiologist. (Table 1). Paired t-test showed significant dif- Histomorphometrically, bone volume frac-
ferences (P < 0.001) between all comparison tion, which is the trabecular BV per tissue
Data analysis pairs except for mean measurement between volume (TV) expressed in percentage, is the
Statistical analysis was performed using SPSS CBCT BV/TV and micro-CT BV/TV (P = most important parameter (Parfitt et al.
(v17.0, SPSS Inc., Chicago, IL, USA). To 0.147). In all selected ROIs, CBCT showed a 1987). Micro-CT is accepted as a gold stan-
determine the intra-observer reliability of the higher density than MSCT HU and a higher dard modality for trabecular microstructure
radiological and microstructural density mea- BV/TV than that of micro-CT. The normal assessment, but it cannot be employed in the
surement, intraclass correlation coefficient distribution of measurements was confirmed clinic (Burghardt et al. 2011). In this study,
(ICC) was used. The Shapiro–Wilk test was by visually inspecting the histogram and the our aim was to investigate the possible corre-
used to verify the normality of the data. result of the Shapiro–Wilk test (P > 0.05). lation between bone quality measurements
Paired t-test was used to assess the mean dif- Therefore, the use of the t-test and Bland– of clinically applicable scanners in compari-
ference between MSCT and CBCT density Altman test is justified. Strong correlations son with micro-CT.
measurements and between CBCT BV/TV were observed between CBCT and MSCT A study on porcine vertebral cancellous
and micro-CT BV/TV, while Pearson’s density measurements (r = 0.89) and between bone revealed a high correlation between HU

© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 3 | Clin. Oral Impl. Res. 0, 2013 / 1–7
Parsa et al  CT bone quality assessment at implant site

Table 1. Mean results of MSCT, micro-CT, and


(a) CBCT density (gray value) and bone volume
fraction (BV/TV) measurements
CBCT Micro-CT CBCT
Mandible MSCT gray BV/TV BV/TV
No. HU value (%) (%)
1 390.9 513.3 37.11 44.43
2 205.1 444.6 41.40 46.17
3 507.6 665.6 75.83 63.58
4 16.6 204.7 3.91 3.73
5 136 324.4 17.44 9.98
6 202.4 270.8 23.99 24.87
7 245.2 368.5 31.68 24.68
8 327.1 417.5 47.62 50.41
9 341 442.9 49.34 37.94
10 316 474.3 48.68 62.94
11 270 379.3 50.86 48.58
12 210.2 278 44.64 68.72
13 320 559.1 44.12 63.28
(b) 14 204.4 344.9 22.65 26.31
15 285.7 357 34.57 66.29
16 27.6 161.6 2.24 4.33
17 220 483.7 24.87 55.30
18 240.4 376.7 26.97 23.21
19 139.2 285.3 15.97 6.68
20 60 197.7 3.16 4.45

CBCT, cone beam computed tomography; HU,


Hounsfield units; MSCT, multislice computed
tomography.

precise HU measurement would be achiev-


able (Stoppie et al. 2006). Our results showed
a strong correlation between BV/TV and HU
in human mandibular trabecular bone, regard-
less to gender and thickness of surrounding
cortical bone (r = 0.91). This confirms the
(c)
possibility of prediction of bone volume frac-
tion from MSCT bone density measurement.
The usefulness of this prediction can be
emphasized by the limitation of micro-CT in
clinical settings.
Cone beam computed tomography has sev-
eral advantages over MSCT in terms of more
compact equipment, small footprint for the
clinic, and relatively reduced scan costs.
Additionally, lower radiation dose levels to
the main organs of the head and neck region
have been cited as one of the most important
advantages of CBCT over MSCT (Kau et al.
2005; White 2008; Carrafiello et al. 2010).
Fig. 2. (a) Images of micro-CT and cone beam computed tomography (CBCT) were compared slice by slice from the Due to these advantages, the use of this
same anatomical region. (b) A rectangular region of interest (ROI) was selected for each data set. (c) Images were bi- modality in dental implant planning is grow-
narized and (d) processed to allow the microstructural measurements. ing so fast and it is more accessible to the
dental practitioners than before. Therefore,
derived from CT images and BV/TV from using human zygomatic and jawbones, a high the validity of CBCT in bone quality assess-
micro-CT and suggested the use of HU correlation was found only in female subjects ment has been studied broadly. The majority
from medical CT for the prediction of micro- (Nkenke et al. 2003; Aksoy et al. 2009). Thus, of these studies have focused on the bone
architecture (Teo et al. 2006). Our results sup- they suggested that only female trabecular density measurement and found CBCT a reli-
port these findings that correlation between BV/TV can be predicted from BMD. In con- able modality for bone density measurement
MSCT HU and Micro-CT BV/TV is high trast, another study found a high correlation (Aranyarachkul et al. 2005; Lagravere et al.
(r = 0.91). However, the mean of calculated between BV/TV and HU in trabecular bone 2006; Naitoh et al. 2009, 2010b; Nomura
BV/TV in mentioned study deviated from our surrounded by a thin layer of cortical bone et al. 2010; Parsa et al. 2012; Reeves et al.
findings in human mandibles. This could be regardless to gender (Stoppie et al. 2006). This 2012; Cassetta et al. 2013). The high
due to different samples, ROI selections and study suggested that with the development of correlation between measured CBCT gray
different scanner systems. In similar studies MSCT scanners and imaging software, more values and CT numbers in our study

4 | Clin. Oral Impl. Res. 0, 2013 / 1–7 © 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd
Parsa et al  CT bone quality assessment at implant site

(a) deviated from those of gold standards. This


deviation arises from increased scattering,
noise level, and artifacts specific to the scan-
ner technology which operates at lower peak
kilovoltage and tube loading setting than
MSCT and micro-CT, resulting in a reduced
signal-to-noise ratio (Schulze et al. 2011). A
higher noise level in comparison with MSCT
can cause more inconsistencies in voxel gray
values (Aranyarachkul et al. 2005; Araki &
Okano 2011). Additionally, as the acquired
volume in CBCT is larger than collimated
fan beam in MSCT, the influence of these
artifacts is excessively exacerbated (Nacka-
erts et al. 2011; Schulze et al. 2011).
Unlike the majority of other studies on
bone volume fraction, our bone samples were
(b) not harvested for micro-CT scans. As such,
in our sample, the possible deviation between
the planned and excised ROI, which might
arise during the trepanation procedure, was
eliminated (Stoppie et al. 2006). Additionally,
in the present study, a fully automated and
observer independent 3D matching algorithm
was employed for MSCT and CBCT scans
registration to ensure that all measurements
are exactly from the same site up to voxel
accuracy. However, due to the manual align-
ment of CBCT and micro-CT data sets, there
is a possibility for observer error and selec-
tion of not identical regions. As micro-CT
data sets are large and therefore computation-
ally expensive, technical limitations prohib-
ited applying the 3D registration algorithm
for automated alignment. Technical advance-
ments in the future might resolve this issue.
Fig. 3. Bland–Altman plots of (a) bone volume fraction (BV/TV) measurements between cone beam computed
tomography (CBCT) and l-CT, and (b) density measurements between CBCT and multislice computed tomography Finally, the difference in voxel size of CBCT
(MSCT). (0.080 mm), micro-CT (0.035 mm), and
MSCT (0.67 mm) can also contribute to the
(r = 0.89) may confirms the possible potential et al. 2013b). Our results also confirm the observed discrepancy in calculating BV/TV
of CBCT in radiographic density measure- reliability of CBCT in trabecular microstruc- and bone density. Voxel size in CBCT influ-
ment. However, the limit of agreement in ture assessment, based on a high correlation ences image quality among other factors
Bland and Altman plot (Fig. 3b) is huge between BV/TV measured by CBCT and including the unit itself, tube voltage, and
(29.74–279.56) with a high bias value micro-CT (r = 0.82). The positive bias value FOV selection (Kamburoglu et al. 2011). Gen-
(mean = 154.64). This indicates an unfavor- (4.44/lm) in the Bland and Altman plot erally, the smaller the voxels, the higher the
able strength of agreement. Thus, although (Fig. 3a) indicates that BV/TV was measured spatial resolution and therefore the sharper
the measurements is reliable (ICC > 0.97) smaller by CBCT. The small range between the images appear to be. However, small vox-
and validated between two compared systems the confidence interval for the measurement els result in decreased contrast-to-noise ratio
(r = 0.82), the density measurement using differences between the two systems was levels and they require higher exposure dose
CBCT is less accurate when compared to its small ( 21.31 to 30.19) indicates a strong to the patient (Davies et al. 2012). The higher
gold standard system (MSCT). It should be agreement between CBCT and micro-CT in the spatial resolution, the more technical
considered that CBCT density measurement measuring BV/TV. In present study, smallest demands are imposed on the imaging system
can be effected by scanning parameters and available FOV (40 9 40 mm) and high-resolu- as a whole and on the imaging detector in
the location of the ROI within the scanner tion scan mode were applied in CBCT specific to attempt to suppress noise and
(Nackaerts et al. 2011; Parsa et al. 2013). scans to achieve the highest possible spa- increase signal levels. CBCT suffers from
Using micro-CT as gold standard, the reli- tial resolution (0.08 mm isotropic voxel size). increased noise levels especially at smaller
ability of CBCT in trabecular microstructure Therefore, using different CBCT scanning voxel sizes due to low tube voltage, cone
assessment has been validated in human parameters, the results may differ. beam divergence phenomena, and inferior
mandibles, but BV/TV was not among the It should be emphasized that the CBCT detector efficiency when compared to MSCT
assessed microstructural parameters (Ibrahim bone quality measurements in our study and micro-CT (Hassan et al. 2010). However,

© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 5 | Clin. Oral Impl. Res. 0, 2013 / 1–7
Parsa et al  CT bone quality assessment at implant site

the potential influence of varying voxel size each model. Developing such a standard for absorb the beam causing extinction artifacts
on visibility of hard tissue structures such as manufacturing CBCT systems may help in rather than beam-hardening artifacts (Hara-
bone remains largely unknown. A recent sys- generalizing research findings in the future. mati et al. 1994). Resulted artifacts degrade
tematic review of the literature concluded The study was also limited as surrounding the quality of images and affect the gray
that there is a systematic lack of evidence anatomical structures including the tongue scales of normal anatomical structures close
regarding the impact of varying voxel size in and vertebra were absent. As a result, in to foreign bodies. The severity of mentioned
CBCT on diagnostic performance and that CBCT scans, partial object artifacts resulting effect is also dependent on the energy of
possibly different voxel sizes might lead to from structures placed outside the scan field applied X-ray beam, density, and geometry of
comparable diagnostic outcomes (Spin-Neto were not simulated. It has been previously artifact inducing materials (Schulze et al.
et al. 2013). Only one study could be identi- noted that artifacts resulting from partial 2010). Therefore, further CBCT studies on
fied which demonstrated a possible effect of sampling of objects outside the scan field assessing the effect of dental restorative
varying voxel size on cancellous bone mea- could result in a deviation in voxel gray val- materials on bone density and microstructure
surements in micro-CT (Yeni et al. 2005). ues with CBCT (Katsumata et al. 2009; Araki measurements are suggested.
However, it remains unknown whether the & Okano 2011). Gray values obtained from In conclusion, this present study demon-
same applies to CBCT. In this study, a con- the cadaver may also deviate from the clini- strates the reliability and validity of CBCT in
scious effort was made to optimize image cal situation. bone quality assessment. However, unlike
quality through selecting the scan protocols In present study, restorative materials the bone volume fraction measurement, the
and voxels sizes as recommended by the which could induce artifacts were removed accuracy for density measurement is unfavor-
manufacturer for the chosen FOV’s. Our from our samples. However, in normal clini- able. In assessing density using CBCT, the
results are limited to one CBCT system cal settings, presence of metallic materials in microstructural assessment (BV/TV) is there-
(Accuitomo 170), and results may vary on oral cavity is quite common. In CBCT scans, fore recommended. However, based on the
other systems. The design specifications of restorative materials with high atomic num- inconsistencies in CBCT designs, further
different systems still vary (De Vos et al. bers harden the X-ray beam causing streak studies are suggested on validation of differ-
2009). The lack of a technical standard for artifacts in reconstructed images (Schulze ent systems.
the development of CBCT systems has led to et al. 2011), while many high-density filling
a wide disparity in the physical parameters of materials such as amalgam completely

References
Aksoy, U., Eratalay, K. & T€ oz€
um, T.F. (2009) The Corpas, L.S., Jacobs, R., Quirynen, M., Huang, Y., computed tomography for dental implant surgery:
possible association among bone density values, Naert, I. & Duyck, J. (2011) Peri-implant bone tis- comparison with conventional tomography.
resonance frequency measurements, tactile sense, sue assessment by comparing the outcome of Dentomaxillofacial Radiology 22: 13–17.
and histomorphometric evaluations of dental intra-oral radiograph and cone beam computed Frederiksen, N.L., Benson, B.W. & Sokolowski,
implant osteotomy sites: a preliminary study. tomography analyses to the histological standard. T.W. (1995) Effective dose and risk assessment
Implant Dentistry 18: 316–325. Clinical Oral Implants Research 22: 492–499. from computed tomography of the maxillofacial
Araki, K. & Okano, T. (2011) The effect of sur- Davies, J., Johnson, B. & Drage, N. (2012) Effective complex. Dentomaxillofacial Radiology 24:
rounding conditions on pixel value of cone beam doses from cone beam CT investigation of the 55–58.
computed tomography. Clinical Oral Implants jaws. Dentomaxillofacial Radiology 41: 30–36. Fuh, L.J., Huang, H.L., Chen, C.S., Fu, K.L., Shen,
Research 17: 862–865. doi:10.1111/j.1600-0501. De Vos, W., Casselman, J. & Swennen, G.R. (2009) Y.W., Tu, M.G., Shen, W.C. & Hsu, J.T. (2010)
2011.02373.x Cone-beam computerized tomography (CBCT) Variations in bone density at dental implant sites
Aranyarachkul, P., Caruso, J., Gantes, B., Schulz, imaging of the oral and maxillofacial region: a in different regions of the jawbone. Journal of
E., Riggs, M., Dus, I., Yamada, J.M. & Crigger, M. systematic review of the literature. International Oral Rehabilitation 37: 346–351.
(2005) Bone density assessments of dental Journal of Oral and Maxillofacial Surgery 38: Fyhrie, D.P. (2005) Summary–Measuring “bone
implant sites: 2. Quantitative cone-beam comput- 609–625. quality”. Journal of Musculoskeletal & Neuronal
erized tomography. The International Journal of Diederichs, G., Link, T.M., Kentenich, M., Schwie- Interactions 5: 318–320.
Oral & Maxillofacial Implants 20: 416–424. ger, K., Huber, M.B., Burghardt, A.J., Majumdar, Gonzalez-Garcıa, R. & Monje, F. (2012) The reli-
Burghardt, A.J., Link, T.M. & Majumdar, S. (2011) S., Rogalla, P. & Issever, A.S. (2009) Assessment ability of cone-beam computed tomography to
High-resolution computed tomography for clini- of trabecular bone structure of the calcaneus assess bone density at dental implant recipient
cal imaging of bone microarchitecture. Clinical using multi-detector CT: correlation with sites: a histomorphometric analysis by micro-CT.
Orthopaedics and Related Research 469: 2179– microCT and biomechanical testing. Bone 44: Clinical Oral Implants Research 24: 871–879.
2193. 976–983. doi: 10.1111/j.1600-0501.2011.02390.x.
Carrafiello, G., Dizonno, M., Colli, V., Strocchi, S., Dula, K., Mini, R., van der Stelt, P.F., Lambrecht, Haramati, N., Staron, R.B., Mazel-Sperling, K., Free-
Pozzi-Taubert, S., Leonardi, A., Giorgianni, A., J.T., Schneeberger, P. & Buser, D. (1996) Hypo- man, K., Nickoloff, E.L., Barax, C. & Feldman, F.
Barresi, M., Macchi, A., Bracchi, E., Conte, L. & thetical mortality risk associated with spiral com- (1994) CT scans through metal scanning
Fugazzola, C. (2010) Comparative study of jaws puted tomography of the maxilla and mandible. technique versus hardware composition. Comput-
with multislice computed tomography and cone- European Journal of Oral Sciences 104: 503–510. erized Medical Imaging and Graphics 18: 429–
beam computed tomography. La Radiologia med- Ekestubbe, A., Thilander, A. & Gr€ ondahl, H.G. 434.
ica 115: 600–611. (1992) Absorbed doses and energy imparted from Hassan, B., Couto Souza, P., Jacobs, R., de
Cassetta, M., Stefanelli, L.V., Pacifici, A., Pacifici, tomography for dental implant installation. Spiral Azambuja Berti, S. & van der Stelt, P. (2010)
L. & Barbato, E. (2013) How accurate is CBCT tomography using the Scanora technique com- Influence of scanning and reconstruction parame-
in measuring bone density? A comparative pared with hypocycloidal tomography. Dento- ters on quality of three-dimensional surfaces
CBCT-CT in vitro study. Clinical Implant Den- maxillofacial Radiology 21: 65–69. models of the dental arches from cone beam com-
tistry and Related Research. doi: 10.1111/cid. Ekestubbe, A., Thilander, A., Gr€ ondahl, K. & puted tomography. Clinical Oral Investigations
12027. Gr€ondahl, H.G. (1993) Absorbed doses from 14: 303–310.

6 | Clin. Oral Impl. Res. 0, 2013 / 1–7 © 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd
Parsa et al  CT bone quality assessment at implant site

Hua, Y., Nackaerts, O., Duyck, J., Maes, F. & Surgery Oral Medicine Oral Pathology Oral assessment. The International Journal of Oral &
Jacobs, R. (2009) Bone quality assessment based Radiology Endodontics 104: 402–411. Maxillofacial Implants 27: 1438–1442.
on cone beam computed tomography imaging. Martin-Badosa, E., Amblard, D., Nuzzo, S., Elm- Parsa, A., Ibrahim, N., Hassan, B., Motroni, A., van
Clinical Oral Implants Research 20: 767–771. outaouakkil, A., Vico, L. & Peyrin, F. (2003) der Stelt, P. & Wismeijer, D. (2013) Influence of
Ibrahim, N., Parsa, A., Hassan, B., van der Stelt, P., Excised bone structures in mice: imaging at cone beam CT scanning parameters on gray value
Aartman, L.H. & Wismeijer, D. (2013b) Accuracy three-dimensional synchrotron radiation micro measurements at implant site. Dentomaxillofa-
of trabecular bone microstructural measurement CT. Radiology 229: 921–928. cial Radiology 42: 79884780.
at planned dental implant sites using cone-beam Muller, R. (2003) Bone microarchitecture assess- Reeves, T., Mah, P. & McDavid, W.D. (2012) Deriv-
CT datasets. Clinical Oral Implants Research. ment: current and future trends. Osteoporosis ing Hounsfield units using grey levels in cone
doi: 10.1111/clr.12163. International 14: S89–S95. beam CT: a clinical application. Dentomaxillofa-
Ibrahim, N., Parsa, A., Hassan, B., van der Stelt, P. Nackaerts, O., Maes, F., Yan, H., Couto Souza, P., cial Radiology 41: 500–508.
& Wismeijer, D. (2013a) Diagnostic imaging of Pauwels, R. & Jacobs, R. (2011) Analysis of inten- Schulze, R.K., Berndt, D. & d’Hoedt, B. (2010) On
trabecular bone microstructure for oral implants: sity variability in multislice and cone beam com- cone-beam computed tomography artifacts
a literature review. Dentomaxillofacial Radiology puted tomography. Clinical Oral Implants induced by titanium implants. Clinical Oral
42: 20120075. Research 22: 873–879. Implants Research 21: 100–107.
Jaffin, R.A. & Berman, C.L. (1991) The excessive Naitoh, M., Aimiya, H., Hirukawa, A. & Ariji, E. Schulze, R., Heli, U., Gross, D., Bruellmann, D.D.,
loss of Branemark fixtures in type IV bone: a (2010a) a) Morphometric analysis of mandibular Dranicshnikow, E., Schwanecke, U. & Schoemer,
5-year analysis. Journal of Periodontology 62: 2–4. trabecular bone using cone beam computed E. (2011) Artifacts in CBCT: a review. Dentomax-
Jemt, T., Book, K., Lind en, B. & Urde, G. (1992) tomography: an in vitro study. The International illofacial Radiology 40: 265–273.
Failures and complications in 92 consecutively Journal of Oral & Maxillofacial Implants 25: Schwarz, M.S., Rothman, S.L., Rhodes, M.L. &
inserted overdentures supported by Br anemark 1093–1098. Chafetz, N. (1987) Computed tomography: part II.
implants in severely resorbed edentulous maxil- Naitoh, M., Hirukawa, A., Katsumata, A. & Ariji, E. Preoperative assessment of the maxilla for endos-
lae: a study from prosthetic treatment to first (2009) Evaluation of voxel values in mandibular seous implant surgery. The International Journal
annual check-up. The International Journal of cancellous bone: relationship between cone-beam of Oral & Maxillofacial Implants 2: 143–148.
Oral & Maxillofacial Implants 7: 162–167. computed tomography and multislice helical Shahlaie, M., Gantes, B., Schulz, E., Riggs, M. &
Kamburoglu, K., Murat, S., Kolsuz, E., Kurt, H., computed tomography. Clinical Oral Implants Crigger, M. (2003) Bone density assessments of
Yuksel, S. & Paksoy, C. (2011) Comparative Research 20: 503–506. dental implant sites: 1. Quantitative computed
assessment of subjective image quality of cross- Naitoh, M., Hirukawa, A., Katsumata, A. & Ariji, tomography. The International Journal of Oral &
sectional cone-beam computed tomography scans. E. (2010b) Prospective study to estimate mandibu- Maxillofacial Implants 18: 224–231.
Journal of Oral Science 53: 501–508. lar cancellous bone density using large-volume Shapurian, T., Damoulis, P.D., Reiser, G.M., Grif-
Katsumata, A., Hirukawa, A., Okumura, S., Naitoh, cone-beam computed tomography. Clinical Oral fin, T.J. & Rand, W.M. (2006) Quantitative evalu-
M., Fujishita, M., Ariji, E. & Langlais, R.P. (2009) Implants Research 21: 1309–1313. ation of bone density using the Hounsfield index.
Relationship between density variability and Naitoh, M., Katsumata, A., Mitsuya, S., Kamemoto, The International Journal of Oral & Maxillofa-
imaging volume size in cone-beam computerized H. & Ariji, E. (2004) Measurement of mandibles cial Implants 21: 290–297.
tomographic scanning of the maxillofacial region: with microfocus x-ray computerized tomography Spin-Neto, R., Gotfredsen, E. & Wenzel, A. (2013)
an in vitro study. Oral Surgery Oral Medicine and compact computerized tomography for dental Impact of voxel size variation on cbct-based diag-
Oral Pathology Oral Radiology Endodontics 107: use. The International Journal of Oral & Maxillo- nostic outcome in dentistry: a systemic review.
420–425. facial Implants 19: 239–246. Journal of Digital Imaging 26: 813–820.
Kau, C.H., Richmond, S., Palomo, J.M. & Hans, Nkenke, E., Hahn, M., Lell, M., Wiltfang, J., Schu- Stoppie, N., Pattijn, V., Van Cleynenbreugel, T.,
M.G. (2005) Three-dimensional cone beam com- ltze-Mosgau, S., Stech, B., Radespiel-Troger, M. & Wevers, M., Vander Sloten, J. & Ignace, N. (2006)
puterized tomography in orthodontics. Journal of Neukam, F.W. (2003) Anatomic site evaluation of Structural and radiological parameters for the
Orthodontics 32: 282–293. the zygomatic bone for dental implant placement. characterization of jawbone. Clinical Oral
Lagravere, M.O., Carey, J., Toogood, R.W. & Major, Clinical Oral Implants Research 14: 72–79. Implants Research 17: 124–133.
P.W. (2008) Three-dimensional accuracy of mea- Nomura, Y., Watanabe, H., Honda, E. & Kurabay- Teo, J.C., Si-Hoe, K.M., Keh, J.E. & Teoh, S.H.
surements made with software on cone-beam ashi, T. (2010) Reliability of voxel values from (2006) Relationship between CT intensity, micro-
computed tomography images. American Journal cone-beam computed tomography for dental use architecture and mechanical properties of porcine
of Orthodontics and Dentofacial Orthopedics in evaluating bone mineral density. Clinical Oral vertebral cancellous bone. Clinical Biomechanics
134: 112–116. Implants Research 21: 558–562. 21: 235–244.
Lagravere, M.O., Fang, Y., Carey, J., Toogood, Odgaard, A. (1997) Three-dimensional methods for Teo, J.C.M., Si-Hoe, K.M., Keh, J.E.L. & Teoh, S.H.
R.W., Packota, G.V. & Major, P.W. (2006) quantification of cancellous bone architecture. (2007) Correlation of cancellous bone microarchi-
Density conversion factor determined using a Bone 20: 315–328. tectural parameters from microCT to CT num-
cone-beam computed tomography unit NewTom Ozan, O., Turkyilmaz, I. & Yilmaz, B. (2007) A ber and bone mechanical properties. Materials
QR-DVT 9000. Dentomaxillofacial Radiology preliminary report of patients treated with early Science and Engineering 27: 333–339.
35: 407–409. loaded implants using computerized tomography- Tolstunov, L. (2007) Implant zones of the jaws:
Lioubavina-Hack, N., Lang, N.P. & Karring, T. guided surgical stents: flapless versus implant location and related success rate. The
(2006) Significance of primary stability for osseo- conventional flapped surgery. Journal of Oral Journal of Oral Implantology 33: 211–220.
integration of dental implants. Clinical Oral Rehabilitation 34: 835–840. White, S.C. (2008) Cone-beam imaging in dentistry.
Implants Research 17: 244–250. Parfitt, A.M., Drezner, M.K., Glorieux, F.H., Kanis, Health Physics 95: 628–637.
Liu, S., Zhang, Z.Y., Li, J.P., Liu, D.G. & Ma, X. J.A., Malluche, H., Meunier, P.J., Ott, S.M. & Yeni, Y.N., Christopherson, G.T., Dong, X.N., Kim,
(2007) A study of trabecular bone structure in the Recker, R.R. (1987) Bone histomorphometry: D.G. & Fyhrie, D.P. (2005) Effect of microcom-
mandibular condyle of healthy young people by standardization of nomenclature, symbols, and puted tomography voxel size on the finite ele-
cone beam computed tomography. Chinese Jour- units. Report of the ASBMR Histomorphometry ment model accuracy for human cancellous bone.
nal of Stomatology 42: 357–360. nomenclature committee. Journal of Bone and Journal of Biomechanical Engineering 127: 1–8.
Lou, L., Lagravere, M.O., Compton, S., Major, P.W. Mineral Research 2: 595–610. Yoo, S. & Yin, F.F. (2006) Dosimetric feasibility of
& Flores-Mir, C. (2007) Accuracy of measure- Parsa, A., Ibrahim, N., Hassan, B., Motroni, A., van cone-beam CT-based treatment planning com-
ments and reliability of landmark identification der Stelt, P. & Wismeijer, D. (2012) Reliability pared to CT-based treatment planning. Interna-
with computed tomography (CT) techniques in of voxel gray values in cone beam computed tional Journal of Radiation Oncology, Biology,
the maxillofacial area: a systematic review. Oral tomography for pre-operative implant planning Physics 66: 1553–1561.

© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 7 | Clin. Oral Impl. Res. 0, 2013 / 1–7

You might also like