Nursing Master Controller Blinking Alarm Abnormal Solution
Nursing Master Controller Blinking Alarm Abnormal Solution
solution
1 Pressing buttons of whole bed up and down simultaneously lasts 10sec, see if alarm is
off
abnormal
3 Checking all the connecting wire to see whether there is loosen,stripping,damage. Details as follows:
Column motors
wire interface
(if have)
interface
interface
interfae
normal
4 If loosened,then plug well, repress bottons of whole bed up and down
abnormal
6 7 8 9
Pull out controller and Pull out electric box If blinking alarm is off after reposition,but
Pull out left/right side change same and change same blinking alarm reoccure after reoperate nurse
rail signal wire controller,and model electric controller,then we need to reposition.
separately,Repressing
abnormal abnormal abnorma
repressing buttons of box,and repressing Reoperate other function,like head section
buttons of whole bed whole bed up and buttons of whole bed l rising to see whether there is
up and down down simultaneously up and down alarm.(operating knee section, whole bed and
simultaneously to to release alarm. simultaneously to so on), If alarm,then pull out the head
release checking whether release alarm. controller motor,and to operate other
alarm.Checking alarm is off(wait for checking whether functions to see whether alarm occure.( Pull
whether alarm off(wait 30mins) alarm is off(wait for out knee controller motor and whole bed
30mins.) 30mins) controller motor and so on, and wait for
30mins)
head sect,
whole bed
Knee sect.
knee sect.
whole bed
CO61 electric box
head sec.
nurse controller
LA31H
T style wire and side rail LCD nurse controller CB6S electric box Column motor /BL1 LA40M
signal wire map after
pulling out