Rendered PDF File Page 1 of 3 DOC0877690, Rev:1

Electronic Signature Information

Name DOC0877690

Revision 1

Type Controlled Document

Title Mac 400 Pm Check list

Originator 305002478_thirumalai__sudhakar

Name Reason For Change File Size (Bytes)


Mac 400-Maintenance This document consist of PM check list of MAc 400 used in India
64176
Report.pdf region service

Route Signer Status Comments Completion Date


R-4379075 305007379_sanjay arudi__ranganath Approve Approved 16 Nov 2010 09:37:59 GMT

* Printed versions are For Reference Only *


Released
Rendered PDF File Page 2 of 3 DOC0877690, Rev:1

Maintenance Inspection Report


GE Healthcare

Equipment: MAC 400 ECG

Customer: __________________________________________________________________

Equipment Serial Number: _____________________ Software Revision: _______________

System ID : _____________________ Dispatch No. : ___________________

Configuration ()

Basic _____________________ Meas& Interpretation__________________

Tools Required
1. Leakage tester 4. ESD Kit

2. Standard hand tools 5. ECG Simulator

3. Multimeter (4 ½ digit display) 6. Patient Cable

Visual Inspection ()

Inspect the following for excess wear and/or any visual signs of damage.

1. General 2. Connectors 3. Cable Insulation

Cleaning ()
1. External surfaces 3. Print head/Roller

Calibration ()
(There is no On-site calibration required for this product. Setup should be verified for desired performance.)

1. System Setup 2. Service Set-up

Electrical Safety Checks ()

1________V Wall Receptacle Test (210 to 230 VAC between the two hot contacts)

2. Ω Ground Continuity (Impedance of protected connection) Test (≤ 0.1Ω or ≤ 0.2Ω w/power cord)
Released

-1-
Rendered PDF File Page 3 of 3 DOC0877690, Rev:1

3. Leakage Open Closed


Normal Reversed Normal Reversed
Ground (Earth) wire leakage to ground test µA µA NA NA
{Maximum Allowable Leakage Current (µA) – 300}

Chassis (Enclosure) leakage to ground test µA µA µA µA


{Maximum Allowable Leakage Current (µA) – 300}

ECG
Source Risk (Patient cable leakage to ground) Leakage µA µA µA µA
{Maximum Allowable Leakage Current (µA) – 50}

Checkout Procedure

1. Perform the applicable checkout procedure.

Operational Checks

A. Power-up Self-test OK.


B. Record Rhythm OK .
C. Record ECG Functional.

Diagnostic Tests

F. Display Diagnostics.
G. Speaker Tests
H. Keyboard Test
I. Writer Diagnostics

Results / Comments

All the above Checks / Tests has been done and the Device is found to be performed as per the Specifications.

Inspected By: Accepted BY:

Field Engineer Details Customer Details

Name _____________________________________________ ____________________________________________

Signature _____________________________________________ ____________________________________________

Date: _____________________ _____________________

References: 2036039-001 MAC 400 Resting ECG Service Manual


Released