Checkout 500
Checkout 500
GE Healthcare
Customer: __________________________________________________________________
Configuration ()
Tools Required
1. Leakage tester 4. ESD Kit
Inspect the following for excess wear and/or any visual signs of damage.
Cleaning ()
1. External surfaces 3. Print head/Roller
Calibration ()
(There is no On-site calibration required for this product. Setup should be verified for desired performance.)
1________V Wall Receptacle Test (210 to 230 VAC between the two hot contacts)
2. Ω Ground Continuity (Impedance of protected connection) Test (≤ 0.1Ω or ≤ 0.2Ω w/power cord)
-1-
3. Leakage Open Closed
Normal Reversed Normal Reversed
Ground (Earth) wire leakage to ground test µA µA NA NA
{Maximum Allowable Leakage Current (µA) – 300}
ECG
Source Risk (Patient cable leakage to ground) Leakage µA µA µA µA
{Maximum Allowable Leakage Current (µA) – 50}
Checkout Procedure
Operational Checks
Diagnostic Tests
F. Display Diagnostics.
G. Speaker Tests
H. Keyboard Test
I. Writer Diagnostics
Results / Comments
All the above Checks / Tests has been done and the Device is found to be performed as per the Specifications.