First, press the Func key to enter the function menu, enter "65864720" and enter the correction
interface.
1. Short circuit correction
First insert the check plate O/S calibration port to the instrument A PORT, then press Enter to
enter the short circuit correction interface, adjust the check code to the status OK, you can also
press the auto correction key to automatically correct.
2. Conduction measurement and correction
First insert the check plate on the check port to the instrument A PORT, and then press Enter to
enter the conduction measurement and correction interface, adjust the check code, so that the
actual value is close to the standard value, you can also press the auto correction key to
automatically correct.
3. DC high voltage correction
After placing the multimeter gear on the highest dc high voltage, insert the probe into the high
voltage output port below the test machine screen, and then press Enter to enter the DC high
voltage correction interface, adjust the check code, make the actual value close to the standard
value, or press the automatic correction key to automatically correct.
4. Insulation measurement and correction
First insert the checker insulation calibration port to the instrument A PORT, adjust the upper
and lower keys, the instrument will actually measure the insulation impedance value of each
point close to the standard value, you can also press the automatic correction key to
automatically correct.
5. AC voltage correction
After placing the multimeter gear on the highest level of AC high voltage, insert the probe into
the high voltage output port below the test machine screen, and then press Enter to enter the
AC high voltage correction interface, adjust the check code, so that the actual value is close to
the standard value, and you can also press the automatic correction key to automatically
correct.