3608/3606 calibration procedure

First, press the Func key to enter the function menu, enter "65864720" and enter the correction
interface.
1. Short circuit correction
First insert the check plate O/S calibration port to the instrument A PORT, then press Enter to
enter the short circuit correction interface, adjust the check code to the status OK, you can also
press the auto correction key to automatically correct.
2. Conduction measurement and correction
First insert the check plate on the check port to the instrument A PORT, and then press Enter to
enter the conduction measurement and correction interface, adjust the check code, so that the
actual value is close to the standard value, you can also press the auto correction key to
automatically correct.
3. DC high voltage correction
After placing the multimeter gear on the highest dc high voltage, insert the probe into the high
voltage output port below the test machine screen, and then press Enter to enter the DC high
voltage correction interface, adjust the check code, make the actual value close to the standard
value, or press the automatic correction key to automatically correct.
4. Insulation measurement and correction
First insert the checker insulation calibration port to the instrument A PORT, adjust the upper
and lower keys, the instrument will actually measure the insulation impedance value of each
point close to the standard value, you can also press the automatic correction key to
automatically correct.
5. AC voltage correction
After placing the multimeter gear on the highest level of AC high voltage, insert the probe into
the high voltage output port below the test machine screen, and then press Enter to enter the
AC high voltage correction interface, adjust the check code, so that the actual value is close to
the standard value, and you can also press the automatic correction key to automatically
correct.

Second, conduction impedance detection


1. Press the No. 3 key to enter the conduction impedance setting, the upper limit value is set to
1.1Ω , and the lower limit value is 0.9Ω to exit the screen
2. Press the Number 6 key to enter the test project settings, and only put the second item:
Conduction √
3, insert the check board on the verification port, press the Learn learning circuit, the machine
screen will appear 32 (P) point 1 (Net) circuit, the instrument each port must be in turn to learn
the test check
4. Press the Test test key, the instrument will test the actual conduction impedance of the 32P
loop:
The test results are good
Impedance values of around 1Ω , exceeding 1.1Ω or below 0.9Ω are bad

Third, insulation impedance detection


1. Press the No. 5 key to enter the high voltage setting, the DC voltage is adjusted to 500V, the
time is 0.1S, and the insulation impedance is 100MΩ
2. Press a menu to enter the test item settings, and only put item 5: DC insulation √
3. Insert the insulation port of the checker, press the Learn learning loop, and each port of the
instrument must be studied and tested in turn
4. Press the Test test key, the instrument will test the insulation impedance of each circuit, in
order:
The test results were poor
A02<1MΩ ,A05=8-12MΩ , A08=16-24MΩ ,A11=40-55MΩ ,A14=80-110MΩ

Fourth, AC leakage current detection


1. Press the No. 5 key to enter the high voltage setting, the AC voltage is adjusted to 500V, the
time is 0.1S, and the leakage current is 5mA
2. Press a menu to enter the test item settings, and only put item 4: AC pressure √
3. Insert the insulation port of the checker, press the Learn learning loop, and each port of the
instrument must be studied and tested in turn
4. Press the Test test key, the instrument will test the leakage current of each circuit, in order:
The test results were poor
A02>10mA,A05=4-5.5mA, A08=0.8-1.1mA,A11=0.4-0.6mA,

Fifth, resistance detection


1. Press the No. 4 key to enter the component editing and compile the resistance components
1 R A01-A02 1.000Ω 20%
2 R A05-A06 10.00Ω 10%
3 R A09-A10 100.0Ω 5%
4 R A15-A16 1.000KΩ 5%
5 R A21-A22 10.00KΩ 5%
6 R A27-A28 100.0KΩ 5%
7 R A31-A32 1.000MΩ 10%
2. Press the -6 key to enter the test item setting, and only put the 3rd item: passive component

3. Insert the checker resistance port and press the test key Test.
The test results are good

Sixth, capacitance detection


1. Press the No. 4 key to enter the component editing and compile the capacitor components
(C1)
1 C A01-A02 20.00pF 30%
2 C A03-A04 47.00pF 30%
3 C A05-A06 220.0pF 30%
4 C A07-A08 470.0pF 20%
5 C A09-A10 1.000nF 20%
6 C A11-A12 2.220nF 10%
7 C A13-A14 4.700nF 10%
8 C A15-A16 10.00nF 10%
9 C A17-A18 47.00nF 10%
10 C A19-A20 220.0nF 10%
11 C A21-A22 1.000uF 10%
12 C A23-A24 4.700uF 10%
13 C A25-A26 33.00uF 10%
14 C A27-A28 100.0uF 20%
15 C A29-A30 220.0uF 20%
16 C A31-A32 470.0uF 20%
2. Press the No. 6 key to enter the test project settings, and only put item 3: passive
components √
3. Insert the card capacitor (C1) port and press the test key Test.
The test results are good
4. Press the No. 4 key to enter the component editing and compile the polarity of the capacitor
(C2)
1 C A03-A04 920 uF 20%
2 P A03-A04 7.00 V 5%
5. Press the No. 6 key to enter the test project settings, and only put item 3: passive
components √
6. Insert the card capacitor (C2) port and press the test key Test.
The test results are good

Seven, diode detection


1. Press the No. 4 key to enter the component editing and compile the diode components
1 DV A29-A30 5.1V 10%
2 DV A31-A32 0.7V 10%
2. Press the No. 6 key to enter the test project settings, and only put item 3: passive
components √
3. Insert the calibration board capacitor (C2) port and press the test key Test.
The test results are good