Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

BRIEF REPORT: CLINICAL SCIENCE

Flow Rate of Cerebrospinal Fluid Through a Spinal


Needle Can Accurately Predict Intracranial Pressure
in Cryptococcal Meningitis
Tom H. Boyles, MA, BMBCh, MRCP(UK), MD, DTM&H, Cert ID (SA),*
Elizabeth Gatley, MBChB, Dip HIV Man (SA),* Sean Wasserman, MBChB, MMed, FCP (SA), Cert ID (SA),*
and Graeme Meintjes, MBChB, FRCP(UK), FCP(SA), MPH, PhD*†‡

CSF flowing from a spinal needle. The optimal cutoff value for
Background: Patients with HIV-associated cryptococcal menin- defining high pressure using a standard 22-G spinal needle is
gitis (CM) commonly present with raised intracranial pressure (ICP). $40 drops/min. These findings have the potential to improve CM
Aggressive management of raised ICP reduces mortality but requires management in resource-limited settings.
manometers, which are unavailable in most resource-limited settings.
The law of Poiseuille states that the rate of flow of liquid through Key Words: cryptococcal meningitis, manometer, intracranial
a tube is directly proportional to the difference in pressure between pressure, HIV
each end, and it may be possible to indirectly determine ICP by (J Acquir Immune Defic Syndr 2017;74:e64–e66)
measuring flow of CSF through a spinal needle rather than using
a manometer.

Methods: A convenience sample of CM patients requiring lumbar


puncture (LP) (with 22-G spinal needle) for ICP measurement and INTRODUCTION
control were enrolled. ICP was first measured using a narrow bore Cryptococcal meningitis (CM) is a common oppor-
manometer. After removing the manometer, the number of drops of tunistic infection among adults with HIV worldwide and is
CSF flowing from the spinal needle in 15 seconds was counted. responsible for more than 130,000 deaths per year in sub-
Saharan Africa alone.1,2 The mortality rate in sub-Saharan
Results: Thirty-two patients had 89 LPs performed (range, 1–23). Africa has been estimated at 70% compared with 55% in
Fifty-four had high opening pressure with a CSF flow rate of 16–200
other low- and middle-income countries and 10%–20% in
drops/min, and 35 had normal pressure with a CSF flow rate of 8–
high-income countries.3,4 Elevated intracranial pressure
140 drops/min. Area under the fitted receiver operator character
(ICP), defined as cerebrospinal fluid (CSF) opening pres-
curve was 0.89. A flow rate cutoff to define high pressure of $40
sure (OP) of .250 mm H2O, is associated with an increased
drops/min correctly classified 75 of 89 LPs (accuracy 84%).
mortality,4–6 and recent studies have suggested that thera-
Conclusions: It is technically feasible to indirectly estimate CSF peutic lumbar punctures (LP) to control CSF pressure are
pressure to an accuracy that is clinically useful by counting drops of associated with improved survival.7,8 Therefore, routine
measurement of ICP and aggressive management of raised
pressure with repeated LP is recommended in treatment
Received for publication June 9, 2016; accepted September 6, 2016.
From the *Division of Infectious Diseases and HIV Medicine, Department of guidelines.9,10
Medicine, Faculty of Health Sciences, University of Cape Town, Cape The standard method of ICP measurement is with
Town, South Africa; †Clinical Infectious Diseases Research Initiative, Institute a manometer; however, these are unavailable in most low-
of Infectious Disease and Molecular Medicine, Faculty of Health Sciences, resource settings.11 A validated alternative is to use intrave-
University of Cape Town, Cape Town, South Africa; and ‡Department of
Medicine, Khayelitsha District Hospital, Cape Town, South Africa.
nous giving sets with a tape measure,7 but even this method
G.M. is supported by the Wellcome Trust (098316), the South African requires resources that may not always be available. The law
Research Chairs Initiative of the Department of Science and Technology of Poiseuille states that
and National Research Foundation (NRF) of South Africa (Grant No
64787), NRF incentive funding (UID: 85858), and the South African 8mlQ
Medical Research Council through its TB and HIV Collaborating Centres Dp ¼
Programme with funds received from the National Department of Health pr4
(RFA# SAMRC-RFA-CC: TB/HIV/AIDS-01-2014).
G.M. received remuneration for attending a fungal advisory board meeting of where Dp is the difference in pressure between the 2 ends of
Gilead Sciences and has received speaker remuneration from Sanofi a fluid filled tube of length (l) and internal radius (r), m is the
Aventis. The other authors have no funding or conflicts of interest to viscocity of the fluid, and Q is the volumetric flow rate.12 We
disclose. hypothesized that when using a spinal needle with constant
Correspondence to: Tom H. Boyles, MA, BMBCh, MRCP(UK), MD,
DTM&H, Cert ID (SA), Groote Schuur Hospital, Observatory, Cape length and internal radius, it may be possible to indirectly
Town, South Africa 7937 (e-mail: tomboyles@yahoo.com). determine ICP by measuring the number of drops of CSF
Copyright © 2016 Wolters Kluwer Health, Inc. All rights reserved. flowing from the end of the needle.

e64 | www.jaids.com J Acquir Immune Defic Syndr  Volume 74, Number 3, March 1, 2017

Copyright Ó 2017 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
J Acquir Immune Defic Syndr  Volume 74, Number 3, March 1, 2017 CSF Pressure by Drop Rate

METHODS
Study Population
A convenience sample of adult patients with CM
confirmed by CSF cryptococcal antigen testing and/or culture
was identified at 3 hospitals in Cape Town, South Africa:
Groote Schuur, New Somerset, and False Bay. Patients gave
written informed consent to participate in the study. The study
was approved by the University of Cape Town Human
Research Ethics Committee (Ref 643/2014).

Reference Method of Measuring ICP


Clinicians from the team primarily caring for the patient
performed LPs when clinically indicated. A 22-gauge (which
are typically color-coded black) spinal needle of 90-mm
length was used for all LPs. With the patient lying in a lateral
position, ICP was measured using a standard narrow gauge FIGURE 1. ROC curve for measurement of CSF pressure by
manometer by attaching it to the spinal needle and allowing flow rate of CSF through a spinal needle compared with
CSF to rise to a constant height. Raised ICP was defined as an a reference standard of manometry. Area under the ROC curve
OP .250 mm H2O. was 0.89.

classified (accuracy 84%). Sensitivity was 91%, specificity


Investigational Method for Estimating ICP 74%, positive predictive value 84%, negative predictive value
Immediately following reference measurement of ICP, 84%, positive likelihood ratio 3.53, and negative likelihood
the manometer was disconnected from the spinal needle and ratio 0.12 (Table 1).
the number of drops of CSF falling from the hub of the needle
in the first 15 seconds was counted. This was converted to
a flow rate in drops/min. DISCUSSION
This study shows for the first time that it is possible to
measure ICP indirectly by measuring the flow rate of CSF
Statistical Analysis through a spinal needle at LP in patients with CM. The area
Receiver operator character (ROC) analysis was used under the ROC curve of 0.89 when assessed in comparison to
to quantify the discriminative ability of flow rate to determine standard manometry suggests good predictive power. Using
an appropriate cutoff for the presence of high pressure as a standard 22-G spinal needle, a flow rate of $40 drops of
measured by manometry. The area under the curve and 95% CSF/min predicts raised versus normal ICP with an accuracy
confidence intervals were calculated. Diagnostic accuracy of 84% and a sensitivity of 91%, meaning that there were few
was determined by sensitivity, specificity, negative predictive cases with raised OP missed using the flow rate method.
value and positive predictive value, and test accuracy. The It is not surprising that an imperfect relationship
optimal cutoff was determined by performance characteristics between ICP and CSF flow rate was found as several
and clinical utility. All statistical analyses were performed assumptions, which may not be entirely valid, must be made
using StataCorp. 2013. Stata Statistical Software: Release 13. when applying the law of Poiseuille to the flow of CSF
College Station, TX: StataCorp LP. through a spinal needle. Importantly, viscosity of CSF is
assumed to be constant, but this is unlikely given the
variability in protein and glucose content of CSF in patients
RESULTS with CM13; unfortunately, many therapeutic LP samples were
From April 2015 to April 2016, 32 patients (31 HIV not sent for analysis because of resource constraints, so we
infected, 21 men) were recruited, and 89 LPs were are not able to correlate these variables with flow rate.
performed (range, 1–23 LPs per patient). OP as measured Cryptococcus neoformans has a polysaccharide capsule, and
by standard manometry was raised on 54 occasions with
CSF drop rate median of 92 drops/min and range 16–200
drops/min. OP was normal on 35 occasions with CSF drop TABLE 1. Flow Rate of CSF With a Cutoff Value of $40 Drops/
rate median of 28 drops/min and range 8–140 drops/min. Min Compared with a Reference Standard of Manometry
The area under the ROC curve for drop rate was 0.89 Raised ICP .25 Normal ICP #25
(Fig. 1). An optimal cutoff value of $40 drops/min was on Manometry on Manometry Total
chosen because this represents the greatest accuracy and
$40 drops/min 49 9 57
prioritized sensitivity over specificity, as false-negative results
,40 drops/min 5 26 32
are more likely to cause harm to a patient than false-positive
Total 54 35 89
results. Using this cutoff, 75 of 89 LPs were correctly

Copyright © 2016 Wolters Kluwer Health, Inc. All rights reserved. www.jaids.com | e65

Copyright Ó 2017 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Boyles et al J Acquir Immune Defic Syndr  Volume 74, Number 3, March 1, 2017

viscosity might also vary with fungal burden. It should also of patients with CM in low-resource settings, although sites
be noted that 24 of 89 LPs were from a single patient, which wanting to use this method may need to do their own
could have weighted the flow rate results if the CSF viscosity validation according to needle sizes available at their site.
in this individual varied from the norm. The size of CSF This means that ICP can be determined quickly without the
drops is also assumed to be constant but is likely to vary for need for manometers that are costly and not currently
similar reasons. Rate of flow is proportional to the forth available in most resource-limited clinical settings and has
power of the internal radius of the needle and therefore very the potential to improve the care of patients with
sensitive to any inconsistencies in the manufacture of spinal this condition.
needles. There were 5 false-negative results when OP was
high but flow rate was ,40 drops/min; one possible reason
might be that the bevel of the needle moved while removing REFERENCES
the manometer such that the entire needle opening was no 1. Boulware DA. 9th International Conference on Cryptococcus and
Cryptococcosis, 15–19 May 2014, Amsterdam, The Netherlands: Royal
longer within the subarachnoid space. Tropical Institute (KIT). Mycoses. 2014;57(suppl 1):1–108.
Guidelines suggest repeating LPs in patients with 2. Holmes CB, Losina E, Walensky RP, et al. Review of human immuno-
raised ICP when symptoms persist or recur9; however, there deficiency virus type 1-related opportunistic infections in sub-Saharan
is evidence that the use of therapeutic LPs is associated with Africa. Clin Infect Dis. 2003;36:652–662.
3. Dromer F, Mathoulin-Pelissier S, Launay O, et al; French Cryptococcosis
an improvement in survival regardless of initial ICP.8 It may Study Group. Determinants of disease presentation and outcome during
therefore be reasonable for future guidelines to recommend cryptococcosis: the CryptoA/D study. PLos Med. 2007;4:e21.
at least 2 LPs for all patients. If the novel method of ICP 4. van der Horst CM, Saag MS, Cloud GA, et al. Treatment of cryptococcal
measurement were used and any 1 of 2 LPs with flow rate meningitis associated with the acquired immunodeficiency syndrome.
National institute of allergy and infectious diseases mycoses study group
$40 drops of CSF/min was used to meet the definition of and AIDS clinical trials group. N Engl J Med. 1997;337:15–21.
raised ICP, this would reduce the chance of false- 5. Graybill JR, Sobel J, Saag M, et al. Diagnosis and management of
negative results. increased intracranial pressure in patients with AIDS and cryptococcal
An important limitation of the study is that the meningitis. The NIAID Mycoses Study Group and AIDS Cooperative
reference standard of measuring ICP was not performed by Treatment Groups. Clin Infect Dis. 2000;30:47–54.
6. Fessler RD, Sobel J, Guyot L, et al. Management of elevated intracranial
a single individual. LPs were performed by clinicians caring pressure in patients with cryptococcal meningitis. J Acquir Immune Defic
for patients, and there was inevitably a variation in the Syndr Hum Retrovirol. 1998;17:137–142.
experience of the clinicians. A second limitation is that 7. Meda J, Kalluvya S, Downs JA, et al. Cryptococcal meningitis manage-
measuring CSF pressure by manometry released CSF and ment in Tanzania with strict schedule of serial lumber punctures using
lowered the ICP when estimated subsequently using flow intravenous tubing sets: an operational research study. J Acquir Immune
Defic Syndr. 2014;66:e31–e36.
rate. This effect was minimized using narrow-gauge man- 8. Rolfes MA, Hullsiek KH, Rhein J, et al. The effect of therapeutic lumbar
ometers, but it may mean that the true flow rate related to punctures on acute mortality from cryptococcal meningitis. Clin Infect
high OP is slightly . 40 drops/min. Any error in this Dis. 2014;59:1607–1614.
direction favors over rather than underestimating OP. The 9. Govender N, Meintjes G, Bicanic T, et al. Guideline for the prevention,
diagnosis, and management or cryptococcal meningitis among HIV-
strength of the study was that the investigational method was infected persons: 2013 update. SAJHIVMED. 2013;14:76–86.
performed by a variety of clinicians with variable experience, 10. Perfect JR, Dismukes WE, Dromer F, et al. Clinical practice
which suggests that this method is generalizable to other guidelines for the management of cryptococcal disease: 2010 update
clinical settings. Future research will be necessary to by the infectious diseases society of America. Clin Infect Dis. 2010;
determine optimal cutoff points for flow rate of CSF using 50:291–322.
11. Jackson A, Hosseinipour MC. Management of cryptococcal meningitis in
spinal needles with different internal radius and length. sub-saharan Africa. Curr HIV/AIDS Rep. 2010;7:134–142.
Future work might also determine the influence of protein 12. Bingham E, editor. Experimental Investigations upon the Flow of Liquids
level and fungal burden in the CSF on flow rate. in Tubes of Very Small Diameter, by Dr Jean Leonard Marie Poiseuille
In conclusion, while measurement of ICP at LP in high- (1846) Translated by W. H. Herschel. Vol 1. 1st ed. Lancaster, PA:
Lancaster Press Inc; 1940.
resource or research settings should be performed by standard 13. Jarvis JN, Meintjes G, Williams A, et al. Adult meningitis in a setting of
manometry, the measurement by flow rate of CSF through high HIV and TB prevalence: findings from 4961 suspected cases. BMC
a spinal needle is accurate enough to be used for clinical care Infect Dis. 2010;10:67. 2334-10-67.

e66 | www.jaids.com Copyright © 2016 Wolters Kluwer Health, Inc. All rights reserved.

Copyright Ó 2017 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.

You might also like