Professional Documents
Culture Documents
61 678 PDF
61 678 PDF
By MasayukiTanaka,TakakiyoNakazawaand ShuhjiAoki
UpperAtmosphere
ResearchLaboratory,
TohokuUniversity,Sendai980,Japan
(Manuscript
receivedMarch3 1983,in revisedformJune5, 1983)
Abstract
The technique for high quality measurements of the atmospheric CO2 concentration has
been developed by improving many facets of the commercial nondispersive infrared CO2
analyzers. The precision of better than *0.05ppm was attained for both continuous and
flask-sample analyses. The standard gases of CO2-in-air mixture have been prepared by the
gravimetric method using an extremely precise balance; the absolute accuracy of the CO2
concentrations were estimated to be better than *0.3ppm. The problems of great impor-
tance in the flask sampling technique, i.e., possible deterioration risk of air samples stored
in the flasks and effect of water vapor condensed in the sample flask on measured values
of the CO2 concentration, have been examined.
Tanaka et al. (1983a), and those of continuous the reference cell of the VIA analyzer from
measurements at a ground station will be given enclosed N2 gas to a constantly flowing reference
in Tanaka et al. (1983b). In this paper, tech- gas to set optical bias for accurate measurements
nical aspects of our monitoring program are in an arbitrary concentration range of 50ppm
described in detail. span; (5) rearrangement of the electric system
of the EIA analyzer to convert the original range
2. Non-dispersive infrared (NDIR) CO2
of 0-500ppm to an arbitrary range of about
analyzers
150ppm by using a differential amplifier; and
The non-dispersive infrared CO2 analyzers of (6) insertion of interference filters (*max=4.3*m,
the types EIA and VIA, manufactured by 1/2=0.3*m), which pass light only* in a spec-
Hitachi-Horiba Co., Japan, have been adopted tral region of the 4.3*m CO2 band, into both
for continuous measurements and analyses of sample and reference beams to reduce spectral
flask samples, respectively. The optical system contaminations due to atmospheric water vapor
of these analyzers is composed of double light absorption, especially by the 2.7 and 6.3*m
sources, an optical balancer, parallel reference bands.
and sample cells, an optical chopper and a single Since the spaces for the optical chopper and
detector with parallel absorption cells (Fig. 1). the optical balancer are open to the analyzer
The following facets of these analyzers have housing, the fluctuation of CO2 content of the
been improved to satisfy the precision required air enclosed in the housing becomes an addi-
for the atmospheric CO2 monitoring: (1) Sta- tional source of the instrumental drift. To avoid
bilization of the electric voltage supplied to the such a defect, the analyzers were purged of CO2
light sources to minimize the fluctuation of the by continuously flowing the CO2-free N2 gas
intensity of the light sources; (2) stabilization of with a pressure slightly higher than the environ-
the voltage and frequency of the electric sources ment.
supplied to the chopper motor to minimize the
irregularity in the rotation rate of the optical 3. Standard gases
chopper; (3) improvement of the temperature To maintain the consistency of measurements
control of the analyzer with a precise propor- for a long time, the standard gases are classified
tional regulation; (4) alternation of the gas in into three categories i.e., the primary, secondary
and working standards. All standard gases are
CO2-in-air mixtures and stored in 47 l high pres-
sure gas cylinders of chrome-molybdenum-steel.
Therefore, the correction for the carrier gas
effect is not necessary. The working standard
gases are used for the calibration of the analyzer
in usual measurements, and their concentrations
are compared with the secondary standard gases
before and after their uses. The secondary
standard gases are also compared carefully with
the primary standard gases for about once a
year. The primary standard gases were calibrated
by gases prepared by the gravimetric method
with an extremely precise balance of accuracy
better than 3*=*4.6mg (* being the rms error)
for a wide range from 1mg to 100kg. For this
purpose, the original gas of about 1 % concen-
tration was first prepared by mixing the air
which was free from CO2 and H2O (<0.1ppm
for CO2 and dewpoint temperature of -70*
for H2O) with accurately known amount of CO2
and, then, the 9 calibration gases ranging from
310 to 450ppm were obtained by dilution. An
Fig. 1 Non-dispersive infrared gas analyzer. example of the preparation of the calibration
680 Journal of the Meteorological Society of Japan Vol. 61, No. 4
gases is given below. The 10l chrome-moly- Table 1 Comparison of the original and cor-
bdenum-steel cylinders were used for this pre- rected CO2 concentrations of the pri-
mary standard gases (units: ppm).
paration. The mass of the CO2 gas with a purity
of 99.995% was measured by the precise bal-
ance to be 30.073g after correcting for the
fluctuation of the zero-point of the balance due
to the fluctuation of the air conditioning and
for the buoyancy for which the correction was
minimized by roughly balancing the volumes of
both sides. The carrier gas (i.e., dry air without
CO2) was then added to the CO2 gas, and the
mass of the total mixture was measured to be
1417.833g; the mass of the carrier gas being
1387.760g, by the same procedure as the CO2
gas. Assuming normal abundances of air mole- tion gases were corrected by a least square fit
cules and their isotopes (Miller, 1961), the technique assuming a quadratic relation between
molecular weights of the carrier gas and CO2 the response of each analyzer and the CO2 con-
were estimated to be 28.960 and 44.011, respec- centration. An assumption of quadratic relation-
tively, and accordingly the CO2 concentration of ship was found to be adequate for an arbitrary
the original gas prepared above was determined range of about 50ppm. In Table 1 are shown
to be 1.406% by volume. The mass of 33.977g the results of comparison of the original calibra-
of the original gas thus obtained was diluted by tion values and those corrected by the above
adding the carrier gas of 1285.965g to obtain mentioned procedure for both VIA and EIA
the calibration gas of the CO2 concentration of analyzers. Most of the corrections were less
359.3ppm by volume. than *0.2ppm. Although the concentration drift
Although the accuracy of the balance we used of the standard gases is a critical problem for
is very good as mentioned before, an overall the interpretation of the secular trend of CO2
accuracy of the gravimetry is degraded to be increase in the atmosphere, our primary stand-
better than *25mg due to adsorption of water ard gases have been confirmed to remain stable
vapor and especially adhesion of dust particles over 2 years by the comparison with the calibra-
on the outer wall of the cylinder. Thus, the tion gases prepared by the same procedure in
accuracy of the calibration gas is given by February 1981. However, further work will be
necessary for the more precise determination,
which is required for the examination of the
potential long-term drift in the concentration.
The cross comparison of our primary standards
for the preparation of the original gas, and
with the Scripps standards is also desired for
comparability of the data obtained.
The working and secondary standard gases
were used after aging over more than a month
to minimize the concentration drift. The usage
for the preparation of the diluted gas,and
of all gases is terminated at a pressure of about
30kg/ cm2 because the drift in concentration is
for an overall accuracy. The estimation of the likely to occur with further decrease of the pres-
accuracies for the other calibration gases is very sure in the cylinders.
similar to this example; the absolute accuracy
4. Flask sampling procedure
of the calibration gases is of the order of 0.1%
or, equivalently, about 0.3ppm. Uncertainties Air sample containers are cylindrical Pyrex
due to the residual CO2 in the carrier gas and glass flasks of 550ml capacity with greaseless
the molecular abundances are expected to be stopcocks fitted at the both ends. Cleaning of
smaller than the above limit. the flasks was first performed at the time of
In order to minimize the uncertainties in cali- their fabrication by washing in an aqueous de-
brated values, CO2 concentrations of the calibra- tergent solution activated by ultrasonic waves
August 1983 M. Tanaka, T. Nakazawa and S. Aoki 681
and by heating to about 460* for annealing. for the operation at the ground station for con-
Subsequent cleaning is usually done by the same tinuous measurements. The analyzer has the 12
procedure as above at about half year intervals inlets with solenoid valves; 8 channels are usual-
during the long-term use of flasks. However, the ly alloted to the sample air and remaining 4
flasks used for the air sampling on ships are channels to the standard gases. Tubing system
cleaned whenever they are used. Before sam- was made of a number of different materials
pling, the flasks are evacuated to a pressure be- including stainless steel, brass, glass and flexible
low 1 * 10-3mmHg and heated at approximately plastic for which permeability and/or exchange
100* for more than 5 hours. among different gases are of no problems. Es-
The air samples are pressurized up to about pecially, flexible plastic was used only for the
3 atm with an electric or manual diaphragm tubing of the air intake in order to avoid uptake
pump. The intake for the air sample is mounted and release of CO2 depending on the pressure.
so as to avoid unnecessary contaminations such The inner wall of the tubing system was washed
as expiration of the observers. In the air sam- by chemicals and dried out sufficiently before
pling on aircraft and ship, a certain care is also its arrangement. To achieve rapid gas exchange
taken to avoid the contamination from the en- in the system and to save the usage of standard
gine exhaust and cabin air. The sampling time gases, the tubing was made as short as possible.
is taken to be long enough to ensure a complete Additional attention was also paid to confirm
air exchange in the sample flasks. Our sample no leakage in the tubing and connection system.
flasks, not evacuated, require the air flow which Water vapor contained in the sample air is
corresponds to about four times their volume removed in two steps, by specially designed glass
for the complete air exchange. The flask filled traps cooled by electric freezers. The rear trap
with the air sample is protected from breakage is always operated at temperatures lower than
by a shock absorbing material and stored in a - 30* , while the temperature of the fore trap
dark inside of a thick paper box. is arbitrarily set according to the ambient air
temperature. The concentration of CO2 is there-
5. Measurement system and procedures
fore defined as part per million (ppm) by volume
5.1 Continuous measurement dry air. Aerosol particles are removed from the
Figure 2 shows the infrared analyzer system sample air by thread-glass and membrane filters
6. Accuracy of measurements
Griffith, D. W. T., 1982: Calculation of carrier gas book of Geophysics (revised edition), The Mac-
effects in non-dispersive infrared analyzers. I. millan Company, New York.
Theory. Tellus, 34, 376,-384. Pearman, G. I. and J. R. Garratt, 1975: Errors in
atmospheric CO2 concentration measurements
Guenther and R. B. Bacastow, 1982: Calculation arising from the use of reference gas mixtures
of carrier gas effects in non-dispersive infrared different in composition to the sample air. Tel-
analyzers. II. Comparisons with experiment. lus, 27, 62-66.
Tellus, 34, 385-397. Pearman, G. I., 1977: Further studies of the com-
Keeling, C. D., 1958: The concentration and iso- parability of baseline atmospheric carbon dioxide
topic abundances of atmospheric carbon dioxide measurements. Tellus, 29, 171-181.
in rural areas. Geochim. et Cosmochim. Acta, Tanaka, M., T. Nakazawa and S. Aoki, 1983a: Con-
centration of atmospheric carbon dioxide over
, ,R. B. Bacastow, A. E. Bainbridge, J r., Japan. J. Geophys. Res., 88, 1339-1344.
C. A. Ekdahl, P. R. Guenther, L. S. Waterman , and , 1983b:
and J. F. S. Chin, 1976a: Atmospheric carbon Atmospheric carbon dioxide variations in hilly
dioxide variations at Mauna Loa Observatory, district in the suburbs of Sendai, Japan, submit-
Hawaii. Tellus. 28. 538-551. ted to Tellus.
, J. A. Adams, Jr., C. A. Ekdahl and WMO, 1981: .Report of the WMO/UNEP/ICSU
P. R. Guenther, 1976b: Atmospheric carbon Meeting on Instrumennts, Standardization and
dioxide variations at the South Pole. Tellus, 28, Measurement Techniques for Atmospheric CO2.
552-564. Geneva, 8-11 September, 1981.
Miller, L. E., 1961: Atmospheric composition, Hand-
大 気 中 の 二 酸 化 炭 素 濃度 の 高 精 度 測 定
田 中 正 之 ・中 澤 高 清 ・青 木 周 司
東北大学理学部超高層物理学研究施設