Professional Documents
Culture Documents
Physics Expt Manuals
Physics Expt Manuals
Apparatus require:-
Telescope, Adjustable rectangular slit, with micrometer screw, Strips with
equidistant lines.
Procedure:-
Several black strips of equal width “a” and equal spacing “b” are drawn on a white
paper. This design / paper is placed a few meters away from the telescope, with its plane
perpendicular to the axis of the telescope. The design is focused in the telescope, now a
variable graduated slit is mounted on the telescope objective, with its length parallel to the
strips. The slit is fully closed and now it is opened gradually. A condition will come when
the white and black strips will appear to overlap each other (not clearly visible or in other
words the paper will appear to be equally illuminated. The slit width (A) is noted from the
main scale and the circular scale. Now the slit is opened and we will be able to see both
black and white strips clearly. Now again the slit is closed gradually till the design shows
uniform shade, not black and white strips i.e the black and white stripes overlap each
other. The slit width (B) is again noted. Let this critical width of the slit be “W=A+B/2”.
Let “d” be the total diameter of the telescope objective and D be the distance from the
objective of the telescope to drawing..
Ray diagram:-
Telescope
Eye piece ▐ Slit
-----------------
▌ ß--- D ---à
Rectangular
Screw DRAWING OF
STRIPS.
Formula: -
a+b
For aperture W smallest angle resolved =
D
a+bW
For aperture d, smallest angle resolved. q min =
D d
q is the angle when the two images are just resolved. The factor 1.22 is
introduced because the aperture d is circular, while W is for a rectangular slit.
Plot a graph between slit width W and Distance D by taking suitable scale.
A Straight line may appear.
Calculate q min by substituting a, b, & d .we get
a+b W
q min =1.22 ´ ´
D 2.45
W
q min = K
D
[i.e. : θminimum X D = KW ]
Observation Table:-
For critical width (W)
Least count of micrometer screw = …… cm
Precautions:-
(1) Adjustment of the slit and the lines on the strip should be parallel.
(2) The distance D should be used carefully from the objective of the
telescope.
(3) The micrometer screw of the slit should be used carefully for reading
the position of just telescope.
(4) The graph should be properly scaled
Viva-Voce:-
Description of Apparatus
If now the distance between the two bridges of the sonometer is so adjusted that
the natural frequency of vibration f of the wire equals to that of current, there will be a
resonance and wire will vibrate with large amplitude. At this stage
1 T
f=
2l m
Where l is the resonant length of the wire between the bridges, m is mass per unit
length of the wire. T(=Mg) is the tension in the wire, M is mass placed on the hanger and
g is acceleration due to gravity.
Procedure
(a) Place the bar magnets across the sonometer wire with opposite poles facing each
other.
(b) Adjust the vibrating length of wire between the bridges by shifting knife edge D
(if edge C is kept fixed) until you observe the vibrations with large amplitude in
the wire.
(c) Measure the length measurement between two knife edges C and D.
(d) Repeat the above procedure by increasing the weight by 100 gm and adjust the
length again between the bridges.
(e) Take four sets of observations each by gradually increasing the load
(f) Now observe the same by gradually decreasing the load.
Observations
Calculations
T= M g, g = 980 cm/s2
1 T
f=
2l m
Result
Mean frequency of A.C. mains = say x Hz (By experiment)
Frequency of A.C. Mains in India (Standard value) = 50 Hz
x − 50
% error = ×100
50
Precautions
(a) The wire must be uniform without kinks or unevenness.
(b) The knife edges should be shifted slowly so that correct resonance length is
found.
(c) The length of wire should be noted at vibrations of maximum amplitude in the
wire.
(d) Magnets with opposite poles facing each other should be placed.
Questions:
1 T
f= ,
2l m
Description of Apparatus
The electrical vibrator consists of a solenoid through which a thin rod of steel, is
passed. The rod is clamped at one end and at its free a hook is attached to which a string
under tension is connected. The solenoid is connected with a 25 watt lamp and is used
directly with a.c. mains. A permanent horse shoe magnet in also mounted on the base
board. The steel rod is passed through the pole pieces .When a.c. is fed through solenoid,
the rod is magnetized longitudinally with the polarity reversing with change in signs of
the current. Owing to the interaction of this with the field due to the permanent magnet,
the rod vibrates with the frequency of a.c. The length of the steel rod can be adjusted so as
to get resonance, indicated by large amplitude of vibration of its free end .A continuous
Procedure
(a) A rod AB is kept in a solenoid along the axis with clamping screws X, Y. The end B is
free to vibrate. N-S are the pole pieces of a permanent magnet. The thread is tied at B.
The other end is passed over pulley and fastened with pan. On switching alternating
current from a.c. mains passes through the solenoid. The rod gets magnetized with
polarity changing with a.c. frequency (50 Hz.).
(b) Tie the thread at B and pass it over pulley to fasten it with pan.
(c) Switch on the vibrator by connecting plug to a.c. mains carefully and put some weight
on pan to get loops.
(d) Adjust the length of the thread to get loops say 4 to 8 of maximum amplitude. Note
down the number of loops, weight placed on pan, weight of pan and length of the
thread between point B to pully.
(e) Repeat the same for different weights and have at least 6 set of observations.
(f) Calculate the frequency of a.c.. mains (f) by using the formula given above.
Observations
Mass per unit length of the thread (m) =
(Given at the bottom of the pan)
Acceleration due to gravity (g) = 980 cm /s2
Applied Mass of Total Tension No. of Length Length of
S.No. Mass pan Mass M applied loops of the one loop
(A) gm (B) gm = A+B T= Mg (n) thread ℓ = L /n
gm (L)
1
2
3
4
5
6
Calculations
1. ƒ1 =
2. ƒ2 =
3. ƒ3 =
4. ƒ4 =
5. ƒ5 =
6. ƒ6 =
Result
The mean frequency of A.C. Mains = ----------------------------- Hz.
= -----------------
Precautions
Questions:
(Do not write in your practical file while submission)
Thermometer
Oven
−E g
I0 = C e k BT …………..(2)
where Eg is the energy gap of the semiconductor and C is constant depending upon the
semiconductor doping and other parameter, and it is represented with usual meaning of
parameters as
−Eg
C = A e D p p + Dn n p N C N V e kT ……………….(3)
L p N D n Ln N A
ln I0 = lnC – E g
×
1
kB T
or log10 I0 = log10 C − Eg /2.3026 kBT …………………(4)
Equation (4) implies that the reverse saturation current I0 in a given semiconductor diode
is dependent only on the temperature T. It varies inversely with temperature. A plot of
log10 I0 vs 103 /T gives a straight line having a slope = 5.04 Eg . Hence band gap
Diode reverse 3
S.No Saturation
Temperature Temperature 1 × 10
Calculations:
Log10 I0
103/ T
Fig.3 Graph for band gap determination
Eg = (Slope/5.04 ) eV
= ………… eV
Result:
The width of the forbidden gap in germanium semiconductor is found to be ……eV
Precautions:
1. The thermometer should be properly placed .
2. The maximum temperature of the diode is not allowed to go beyond 700C
Objective: To plot the variation of magnetic field along the axis of a current
carrying circular coil & estimate the radius of circular coil by the
graph.
Apparatus required: Stewart & Gee (S&G) type tangent galvanometer, rheostat,
cummutator, key, connecting wires, spirit level
2pnir 2
F= ,
10( x 2 + r 2 ) 3 / 2
Where n is the number of turns in the coil, and r is the radius of coil.
If F is perpendicular to earth’s magnetic field H, the deflection θ is related to F as
F=H tanθ i.e.
2pnir 2
H tanq = .
10( x 2 + r 2 ) 3 / 2
Procedure:
(1) You will note that in the magnetic compass provided to you, the big thin needles
pointing on scales are pointers whereas the thick, small and black object is a freely
suspended magnet, (see Fig.1) which will always rest in north – south direction.
Thus the long needles will always point towards east – west directions.
(2) Now through natural observations locate geographical east and west directions and
then place the S&G galvanometer along with magnetic needle on the table such
that the long arms (pointers) of the magnetometer are in the east - west directions. This
enables us to place our apparatus in a magnetic meridian (i.e. along earth’s
magnetic field i.e. north – south direction).
(3) Place the magnetic compass at the centre of the coil and look from top of the
circular coil and rotate the apparatus such that the magnet and coil lies in the same
plane. Now our apparatus is in magnetic meridian.
(4) Now rotate the magnetic compass slowly such that the two long needles (pointers)
are at zero- zero (see Fig.1). If there is slight error in any of the needles reading level
the apparatus with the help of leveling screws and the spirit level provided. Once the
pointers are at zero – zero and the apparatus is in magnetic meridian we can
connect the circuit as shown in the circuit diagram. Care should be taken that the
magnetic coil shouldn’t be disturbed.
(5) With the help of the circuit diagram (see Fig.2) connect the circuit and adjust the
current in the circuit (by varying the sliding port of rheostat) such that the pointers
deflection reads around 70-75 degrees. Place this as reading at x = 0.
(6) Now move the compass slightly by 2 cm or 4cm towards left and note down the
reading of the compass in forward direction and reverse direction ( by changing the
position of the keys in the commutator) in the observation table. In this way note
down the reading after every two / four centimeters and reach the extreme left of
the apparatus. Here it must be noted that all the values of deflection (θ) will be almost
similar and if it is not so it is an indication that the apparatus is not perfectly placed
in magnetic meridian or not properly leveled.
(7) Now continue this process till you reaches the extreme end (LHS) or the deflection
drops to 2°- 4°.
(8) Now again bring the compass in the center and note the reading at x = 0. Start
moving towards RHS and repeat the above mentioned process.
(9) Now plot a graph between the distance (on x axis) and tanθ on the y axis by taking
suitable scale (see Fig.3)
(10) The graph will be like as shown in the figure 3.
(11) Now mark the points of inflexion (points where the graph changes its concavity /
slope) say P and Q.
(12) Measure the distance between P and Q. This is roughly twice the radius of the coil.
Half of this distance gives us the estimated radius of the coil.
Observation Table
S.No. Distance of the Deflection in the magnetic needle (on moving tan θ
needle from the towards RHS end)
center of the Current one way Current in reverse Mean θ = (θ1
apparatus direction +θ2 +θ3 +θ4 )
x = --- cm θ1 θ2 θ3 θ4 /4
1 0 cm
2 2 cm
3 4 cm
4 6 cm
5 8 cm
6 10 cm
7 12 cm
8 14 cm
9 16 cm
10 18 cm
11 20 cm
12 22 cm
13 24 cm
14 28 cm
15 30 cm
S.No. Distance of the Deflection in the magnetic needle (on moving towards tan θ
needle from the LHS end)
center of the Current one way Current in reverse Mean θ = (θ1
apparatus direction +θ2 +θ3 +θ4 )/4
θ1 θ2 θ3 θ4
1 0 cm
2 2 cm
3 4 cm
4 6 cm
5 8 cm
6 10 cm
7 12 cm
8 14 cm
9 16 cm
10 18 cm
11 20 cm
12 22 cm
13 24 cm
14 28 cm
15 30 cm
Results:
(1) A study of the variation of magnetic field along the axis of the circular coil was done.
The graph showing this variation of magnetic field is plotted and it was concluded that the
magnetic field is maximum at the center of the coil and the field decreases as we move
away from the center of the coil on either side.
(2) The radius of the coil was estimated from the graph drawn and it comes out to be
________ cm.
Precautions
1.The coil should be carefully adjusted in the magnetic meridian.
2.All the magnetic materials and current carrying conductors should be at a considerable
distance from the apparatus.
3.Parallax should be removed while reading the position of the pointer.
4.The curve should be drawn smoothly and points of inflexion should be determined very
carefully.
Questions
(Please do not write in your practical file for submission)
1. Does earth have any magnetic field of its own?
Ans. Yes the earth itself is a magnet. Its north geographical pole is close to a magnetic
south pole which is why the north pole of a compass needle points north. The earth’s
magnetic axis is not quite parallel to its geographic axis ( the axis of rotation), so a
compass reading deviates somewhat from geographic north. This deviation which varies
with location is called magnetic declination or magnetic variation.
APPARATUS REQUIRED:-
Lummer Brodhum photometer, optical bench, two sources of light,
semi transparent glass plate.
Illuminating power:-
The illuminating power of a source is the quantity of light falling per second on a unit area
placed at a unit distance from the source in a direction normal to the rays. It is measured in
candle power.
Intensity of Illumination.
The intensity of illumination at a point is defined as the light per second on a unit aria of the
surface placed at a point under consideration.
Intensity of illumination = Illuminating power / (distance) 2
r1 r2
S1 S2
B
G1 G2
G4
G3
Figure 1
S2
S1
Figure2
Glass plate
Source Source
S1 S2
← r1' → ← r2 ' →
Figure 3
When the distances of the sources are adjust such that the two parts are equally bright, i.e. the
light falling from both the sources is equal, we have
P1 P2 P1 r12
= or = , (1)
r12 r22 P2 r22
where r1 and r2 are the distances of two sources S1 and S2 respectively at the point of where
the two sources produces equal illumination.
Now if we place a screen or as in our case a semitransparent glass plate with
absorption coefficient k (then the transmission coefficient of the glass plate will be 1-k where
we have neglected the reflection coefficient r (as r+a+t=1)) is put, near the photometer
against source S1. Let with the glass plate we obtain equal illumination at distances (say)
r1 ' and r2 ' , then the Eq.(1) will be modified to
(1 - k ) P1 r1'2
= '2 . (2)
P2 r2
Using Eq.(1) in (Eq.2) we get
(i) The photometer upright is mounted nearly in the middle of an optical bench and
adjusted in such a way that photometer head becomes normal to the line joining the
two sources as shown in Fig.(3).
(ii) Remove the shutter from one end of photometer and allow light to fall on one face of
the head and observe carefully through the eye piece whether the central or outer
region of the screen is more illuminated. Close the shutter and open the shutter from
the other end and again make the same observation. This will give us an idea that due
to which source the two regions of the screen are illuminated.
(iii) Now remove the shutters from both sides of photometer and allow the light to fall on
each face of the head.
(iv) The positions of one lamp and photometer head are noted. For one set of observation,
position of the photometer head should not be disturbed.
(v) Now move the other lamp from one end of the bench towards photometer till the field
of view in photometer is equally illuminated. Note down the position of the lamp.
(vi) Now interpose the glass plate between photometer and source S1. Move S1 till the
field of view in photometer is equally illuminated. Note the distance of source from
photometer.
(vii) Repeat the above procedure by changing the distance of source S2 from photometer.
(viii) Repeat the above procedure and take at least five sets of observations.
VIVA-VOCE:-
Q. (1) What do you mean by photometer?
Ans: The branch of the optics which deals with the comparison and measurement of the
quantity of radiant energy emitted or received or absorbed .
Q. (2) Define illuminating power.
Ans: The illuminating power of a source is the quantity of light falling per second on the
unit area placed a unit distance form the source in a direction normal to the rays .It is
measured in candle power.
Q. (4) What is the sensitivity of the eye in comparing luminance of two surfaces
Ans: The eye can distinguish between two surfaces placed side by side and seen
simultaneously which differ in brightness by one percent?
S.No. Position of Without glass plate With glass plate r1=|x-a| cm r2=|b-x| cm r2 ' =|x-c| cm r2 ' =|d-x| cm r '12 r22 Mean k
photometer Position Position Position Position
k = 1-
r ' 22 r12
‘x’ cm of source of source of source of source
S1 S2 S1 S2
a cm b cm c cm d cm
1
2
3
4
5
6
Calculations:
Lens
He-Ne
LASER
Theory :
He-Ne laser: The word LASER stands for light amplification by stimulated
emission of radiation. A laser provides an output which is coherent and of high
irradiance. It has a good directionality and monochromatic in nature. In He-Ne
laser population inversions is produced through inelastic collisions between excited
He atoms and the Ne atoms. This is possible because the levels E4 E6 of neon
atoms have almost the same energy as the levels F2 and F3 of helium atoms as
shown in energy level diagram below. The most prominent laser line is 6328Ǻ
emitted by transition from E6 to E3.
Helium
19 - Through atomic collisions
- Neon
17 - F3 E6 3..39 mm
- E5
15 - F2 E4 1.15 mm 6328 Ǻ
Energy - E3
eV 13 - Excitation by Spontaneous
E
- collision with 2 emission (6000 Ǻ)
11 - electrons
- De-excitation by
0 collision.
E
F1 1
pass by the edge of narrow aperture or slit. this effect is approximated by q = l / d where θ
is the diffraction angle, l the wavelength of radiant energy, and d the aperture diameter.
When an single slit is placed in the path of a laser beam a well-defined diffraction pattern
is obtained. The width of the slit b is obtained from the formula
ml
b=
sin q n
q = ± sin -1 (2 l / b ) .Since sinθ cannot exceed unity, the maximum value at m is integer
which is less than b/l.
Procedure
(i)The diffraction pattern is obtained at the focal plain of the lens. The single slit is put in
(ii) The scale is adjusted to be obtain the symmetrical diffraction pattern about the central
maximum
(iii)The distances between the central maximum and the first, second, third minima are
measured accurately on either side of it with the help of the transparent side.
(iv) The distance D between the slit and the screen is noted.
Observations
Calculations:
Results:-The width of the single slit using the He-Ne laser source = ………..cm
Precautions:
ii)The scale on the screen should be kept normal to the incident beam.
Objective: - To plot the charging and discharging curves of two different
combinations of resistances and capacitors.
Circuit Diagram:-
R C X
Y
Z
+ -
Battery
Connect Y and Z – for charging
Connect Y and X - for Discharging
Observation Table:-
R 1 = ---- Ω, C 1 =---- μf.
1 5 1 5
2 10 2 10
3 15 3 15
4 20 4 20
5 25 5 25
6 30 6 30
7 35 7 35
8 40 8 40
9 45 9 45
10 50 10 50
Graph:-
Charging Curve
Charging
or
Discharging
Voltage (V)
(in voltage) Discharging Curve
| | | | | | | | | | | | |
5 10 15 20 25 30 35 40 45 50 55 60 65
Time(in sec.)
Scale:-
X axis = 10 division = -----Sec.
Y axis = 10 division = ….V.
Result: - It is clear from the graph that the current flow has an exponential nature in
both the cases of charging and discharging of a capacitor through a resistance.
Precautions:-1. The connections wire should be strength and not coiled in order to
eliminate to effect due to self induction.
2. Before starting the fresh observations, the condenser should be discharged by sorting
both end of the condenser.
3. Wires connected to the power should be well insulated and so placed that they are not
likely to be touched accidentally.
Questions:-
Q.1 what is capacitor?
Ans. A condenser is a device for storing charge. Basically, it consists of two conducting
plates separated by an insulating material called dielectric.
QαV
or Q/V = constant = C
This constant C is called the capacity of the conductor and is defined as the ratio of the
charge on the conductor to its potential.
Q.6 On what factors does the rate of charging and discharging of a capacitor depends?
Ans. The rate of charging and discharge and discharging of a capacitor depends on its
capacity C and effective resistance R in the circuit.
Equipment: Fiber Optic trainer, Optical fibers, Numerical Aperture Measurement pattern
Theory: Numerical aperture (NA) refers to the maximum angle at which the light
incident on the fiber end is totally internally reflected and is transmitted properly along the
fiber. Thus Numerical aperture of any optical system is a measure of how much light can
be collected by the optical system. It is the product of the refractive index of the incident
medium and the sine of the maximum ray angle.
NA = n0.sinөmax ; n0 for air is 1,
hence NA = sinөmax
For a step-index fibre, the numerical aperture in term of refractive indices is given by
NA= − ncladd
2 2
n core
Procedure:
The schematic diagram of the numerical aperture measurement system is shown in Fig 1 .
Fig.1
1. Connect power supply to the board.
2. Connect one end of fiber cable to the output socket of emitter circuit and the other end
to the numerical aperture measurement pattern. Take the other end of the fiber and
project the light output on to the screen (NA Measurement pattern) to obtain a bright
circular spot as shown in fig 1.
3. Determine the diameter W of the bright spot and the distance L from the fiber end to
the screen. For this hold the white scale-screen with 4 concentric circles (5, 10, 15&
20 diameter), vertically at a suitable distance to make the red spot from the fiber
coincide with diameter circle. Vary the distance between in screen and fiber optic
cable and make it coincide with one of the concentric circles. Note its distance.
4. Calculate the acceptance angle using the formula θ = tan − 1 ( W )
0
2L
Observation Table
1
2
3
4
Result: Numerical aperture of the available optical fibers is determined, and is -----.
Objective: To study the V-I characteristics of light emitting diode (LED).
Apparatus: Light emitting diode, 0-5V variable Supply, 0-10v Voltmeter, 0-50mA DC Ammeter.
Theory: In a PN junction charge carrier recombination takes place when the electrons cross from the
N layer to the P-layer. The electrons are in the conduction band on the P-side while holes are in the
valence band on the N-side. The conduction band has a higher energy level compared to the valence
band and so when the electrons recombine with a hole the difference in energy is given out in the
form of heat or light. In case of silicon or germanium, the energy dissipation is in the form of heat,
whereas in case of gallium-arsenide and gallium phosphide, it is in the form of light. This light is in
the visible region. Germanium and silicon, which have Eg about 1 eV cannot be used in the
manufacture of LED. Hence Gallium arsenide, Gallium phosphide which emits light in the visible
region is used to manufacture LED.
Procedure:
Fig .1 Fig.2
Observation Table :
Result: The V-I characteristics of given LED are studied by graph (as shown Fig.2).
Calculated
Viva Voce