Continuous-spectrophotometric
AMP BUFFER (IFCC)
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument: BS-480
Reagent preparation
Reagent 1: Use the Reagent A.
Reagent 2: Use the Reagent B.
Basics
Test information Reagent Volume Sample volume
No * R1 200 Standard 5
Test information Alp R2 50 Increase
Full name Alp R3 0 Decreased
St.No 2 R4 0
Judgment Criteria
Absorbance 0 - 30000 Lin.Range 1.0-1200 Prozone
Incre.Test Lin.Limit Q1 Q2 Q3 Q4
Decre.Test Subs.Limit PC Abs
Reference
Gender Sample Type Low limit High Limit
Male Serum 0 115
Female Serum 0 105
Calibration
Calibration Calibrators
Rule Two point linear Calibrators Lot.On Std Conc Expire Position
Replicate 3 Dest. Water 0
K S1 (*)
Judjement Criteria
Sensitivity
Factor Diff
SD
(*) entered
Blank Abs by user
Error Limit
Corr. Coeff.
Vers.04/21