Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 2

Laminar Air Flow Specification

 The cabinet work chamber has the maximum large size with the minimal external cabinet dimensions.

 Laminated safety glass.

 UV light should not work when blower is on.

 Pullout UV unit.

 Main body should be made of stainless steel [(304 grade).

 Table top and working zone should be made of stainless steel [(304 grade)

 Cleanliness level should be as per specifications provided by the latest guideline.

 Noise Level should be less than 65 db

 Average air flow should be as per specifications provided inthe latest guideline (down flow).

 HEPA filters should be with 99.99% efficiency for 0.3 micron with integral metal guards & filter frame

gaskets and manufactured in class 1000 super clean air-conditioned environment for longer life.

 Air Pressurization system should be statically and dynamically balanced, fitted with special vibration

reducing system to suit low noise and vibrations.

 Front door should be made of polycarbonate/ toughened glass.

 Side walls should be made of stainless steel [(304 grade)

 Should have Fluorescent light

 Should have support stand

 UV lamp should be in working zone (40 micro watts/ square cm at 254 nm or better

 Should have Switches & Electrical sockets outlets.


 Should have Pressure Monitors like Magnehelic gauge are required to indicate pressure drop across

HEPA filter.

You might also like