Annex 4 - Specification&quantities - Final
Annex 4 - Specification&quantities - Final
4 Patient Monitor 13
Multi-parameter monitor
Built in battery (Battery is accessible).
HR/PR, ECG/RESP-NIBP-SPO2-2TEMP
System and malfunction alarms with visual and audible alarm indicators
Audible and visual alarm for all setted parameters with message indicator.
The monitor should be supplied with:
should be supplied with all accessories & all connection for adult chilled and another one for pediatrics
Wall Mounted Base Tray With All Necessary Parts To Fix On Wall.
7 ICU Beds 5 Functions with matters PLEASE PROVIDE QUOTATION AND SPECIFICATION FOR BOTH electrical operation and manual operation 9
10 Syringe Pump 23
Permits the use of standard disposable Syringes (sizes 10 – 50 ml) and extension sets, with auto syringe size sensing.
Infusion volume 0.1 to 999 ml in 0.1ml increments in lower ranges
Digital display of volume infused. Infusion rate, alarm prompts etc.
Purge/ bolus facility
Occlusion Pressure limit (from 0 to 750 mmHg)
Mains/Battery operated, Battery can sustain 30 - 60 mins.
Mobile Stand W/ holder
11 Infusion Pump 38
The unit should be microprocessor based programmable peristaltic volumetric infusion pump
Display: 6 digit, display for infusion rate setting, total infusion time, total volume infused, infusion status on alarm
Infusion range: 0.1 ml/hr to 299.9ml/hr (in setps of 0.1 ml)
Volume limit: 0 to 9999 ml (in steps of 1 ml)
It should have Bolus facility
It should have audio visual alarm for : Zero rate selection, Drops not detected, infusion rate high, infusion rate low, Infusion end, Low battery
Mains/Battery operated, Battery can sustain 30 - 60 mins.
Mobile Stand W/ holder
The model capable of working with standard and non-dedicated infusion lines will be favored
12 X-Ray Implant 1
Power Supply 400 V -3ph 50/60Hz
50 kW Generator Power minimum
Floating Top Bucky Table 4 way with Electromagnetic Locks
Table With Bucky Device, X-Ray Grid & Patient Band.
Fluor Rail Tube
Rotating Anode X-Ray Tube Dual Focuses 150KV Min.Tube Voltage And High Speed Rotation.
X-Ray Generator High Frequency Power Output Max 20 kw , max125KV & mA not less than 500 mA.
Anatomical Program
Vertical Bucky Chest Stand With X-Ray Grid .
Digital Panel
2 Flat Panel Detector
Scintillator CsI (Directly Deposit)
Active area 17” x 17”
Image transfer WIFI (wireless communication)
Battery Operating Time > 2 hours
Complete workstation with monitor high resolution and DICOM connection
4. Dry imager /Printer
The system must have a dry Printer
It must be DICOM compatible allowing multiple modalities to be connected
The system must be able to print at least 60 films/ hr of the largest size
The system must deliver its first film within 80 seconds from the request sent
The imager must have spatial resolution of 300 - 500 ppi minimum
The system must have at least three online film sizes : 10” X 12”, 11” X 14” or 14” X 17” films.
15 Portable Ultrasound 1
Portable Ultrasound with trolley and UPS
System weight: 6 kg max and Display 10" min
Broadband, Multifrequency Imaging: 2D / Tissue Harmonic Imaging / M-Mode Velocity Color Doppler / Color Power Doppler PW, PW Tissue Doppler and CW
Doppler angle, correct after freeze
DICOM Image Management (TCP/IP): Print and Store
Transducer
Broadband and Multifrequency: Linear Array, Curved Array, Phased Array
16 Steam autoclave 2
Autoclave Vertical 360 L min, Single door .
Micro-processor operated unit .
Stainless Steel chamber and door
Sterilization Temperatures : 121°C / 134°C
(Vacuum – Steam pressure –dry ) Phases
Over pressure safety valve and low water cut-out with visual and audible alarm indication shall be as standard
Power Supply: 400V -3ph 50/60Hz
Supplied with:
Set of 1 sterilisation containers adapted to the equipment proposed
Set of two door gasket
With suitable softener for unit function
if machine need specific input water Pressure must be attached with suitable pump
OTHER REQUIREMENTS
1.All electrical, mechanical, plumbing, civil and installation works, including materials should be done by the supplier.
All ball joints, clamps, rods and fixations should be made from stainless steel
. Flat, square head rest, (table width), infinite slope adjustment, integral foam padding, 1 piece each
. Leg plates, split type, including integral foam padding, 1 piece each
. Patient restraint / body strap, 1 piece
. Arm supports, 2 pieces
. IV rod, 1 set, adjustable height Stainless Steel.
. Shoulder rest / support with pad, 2 pieces
Orthopaedic trolley, including n. 2 double joint bars, n.2 telescopic height adjustable columns that support the traction group; n. 2 columns to support and adjust the
different orthopaedic rests
Pelvis support for plaster adults /children
21 Ortopedic Drill 3
Portable and battery system
for universal use with attachments for acetabulum reaming, intramedullary drilling, drilling and K-wire application
Jacobs chuck attachment
K-WIRE attachment
Battery Operating Time > 2 hours
battery and charger if is possible compatible with Dermatome
22 Dermatome 2
Portable and battery system
Quick and simple blade changes
Change cutting without tools
Change cutting thickness without tools
Battery Operating Time > 2 hours
battery and charger if is possible compatible with Ortopedic Drill
32 Centrifuge 3
CENTRIFUGE 4x100ML, 6000RPM,
Four place swing out rotor
Rectangular buckets
Rectangular insert 12 x 5-7 ml SET OF 4
Reduction plate 12 x 1.5 ml SET OF 4
Rectangular insert 7 x 5-7ml Vacuntainer SET OF 4
NOTE
All equipment needing consumables must allow the possibility to use generic and/or locally made consumables and/or disposables.
Machine with close Or open system, the supplier must supply free quantity of reagents with machine and must mention how many
free kits and tests of reagents would be supplied. The cost of solutions & consumables per Test should also be offered.
Installation & Commissioning must be done by manufacturer engineer
Users training, by Specialist from the Supplier.
Except where explicitly indicated, each appliance must have a power 220-240V 50/60 Hz