Plot No.: 3802, Laxman Vihar-ll, Gurgaon.122006, Haryana, lndia


Phone: +9'1 991 1 052344, 9818768501
Email: mkt@mcl-pl.in

Cs-l"ilttstionW ol4 ''z


Unique Lab Report No ULR - CC21 0521 000051 514F
M/s Masu Brakes Pvt. Ltd.
Customer
Khasra No. 3063-66 -69-70-74-75-76, 306411, 24423-44,
Name &
42KM Stone, Delhi-Rohtak Road, N.H-10, Village- Mandothi,
Address.
Bahadu rgarh, Distt. - J hajjar, Haryana-124507
Calibration Certificate No lssue Date of Certificate Calibration Done At NABL Certificate No.
2'100005'1514 03 November 2021 Site cc-2105
Date of Calibration Next Due Date (as per customer request)
20 October 2021 19 October 2022 -"'
ROCKWELL HARDNESS TESTER
Details of lD No Range O to 1OO HRB/HRC Visual
DUC Serial No 10t12t083 Least Count 1 HRB/HRC *-- lnspection
L Make/Model Saroj Location Result: Ok
Environmental Conditions:- Temperature ('C):- 23.5 Relative Humidity (%):. 47
Calibration Procedure References:- CP/86 .lt 1qB6 (Part-2):2012, ISO 6508-2:2005
ard Used For ration a d

ui pment Name Hardeness Block Set Hardeness Block Set


ldentification No. HBS/03 HBS/04
SerialNumber 09207668, 1 0208857,09207608 1 020867 2,08207 022,1 02 0 I 5 30
Calibration Certificate No. FRI/1 0-20lH01 368 to H01 370 FRI/1 0-20lH0137 1 to H01 373
Calibration Date 17 October 2020 17 October 2020
bration Valid U 16 October 2023 16 October

Values of DUC Expanded


Repeatabilit
Scale Values of STD considering resolution Error Uncertainty
10% of least count v (r)
HRBW 31.36 31.7 0.34 0,1 1.5

HRBW 70.48 70.7 0.22 0.2 1.5

HRBW 90.42 90.8 0.38 0.1 1.5

HRC 27.04 27.2 0.16 0.2 1.0

HRC 41.00 41.4 0.40 0.2 1.0

HRC 62.22 62.7 0.48 0.'l 1.0


Remarks: None
Abbreviatioris: STD-Standrad Equipment, DUC-Device Under Calibration
Notes: 1. The reported expanded uncertainty is stated as the standard uncertainty in measurement multiplied by the coverage factor
(k=2),whichforanormaldistributioncorrespondsacoverageprobabilityofapproximately95%.
2. calibration.
This certificate is referes only to the particular item submitted for :

3. Results reported are valid at the time of and under stated conditions of measurement
4. This certificate shall not be reproduced except in full without permission of Head of Laboratory.
5. This certificate is for industrial and scientific purpose and can not be used in legal matters

Format Rev.-03, Rev. Date-O1/1 0/201 8

ft'#-.

Page 1 of 1