INDUSTRY
Bidfnow
Is the Steam o.k.? The Air Detector Knows the Answer
‘A the international arena sterilisation of
medical devices is being continually
regulated by new standards , of existing
standards are being revised as needed
This is particularly true in the case of the
standards relating to steam stetisation. In-
temational standard ISO 17665 "Steri-
ization of Health Care Products", whichis
presently available as a draft version, will
‘soon replace European standard EN 554
regulating “Validation and routine moni=
toring for sterilisation with moist heat"
‘The second edition of European standard
285 "Steriisation ~ Steam steriisers —
Large sterlsers” will soon be published.
‘Standard prEN 13060, which is completely
now and applies to the same area but in
this case to small stetlisers up to 60 litres,
was recently put 2 vote, Publication of
this standard is soon expected. All three
draft standards make reference toa device
known as “air detector" which hitherto
hhas elicited little attention in Germany or
in other counties,
Butjust what isan air detector?
Before we answer this question, we must
first ofall consider the quality of steam 2s.
a sterlant. In this paper steam of @ good
‘quality is understood to moan a form of
steam that contains no, or only minute
‘quantities of, air or of other non-con-
30081LETTERS
Snow
the requisite concentration range, this is
surely an acceptable method if one as-
‘sumes thatthe BD laundry package serves
2 a reference for a worst-case scenario.
‘Measurements conducted by the Ad
Hoc Group of the TC 102 WG 2 + 3 Stan
ardisation Committee, which revised Eu
ropean standard 285, have demonstrat-
€ed that the BD testis not suitable for all
lumened instruments or MIS instruments.
Ieis under discussion to incorporate the
hollow device process challenge device
(PCD) Hollow A as specified in EN 867-5
4s an additional type test into standard
EN 286. Whereas the BD test needs ap-
prox. 100 - 200 ml NCGs to exceed the
2 temperature range, volumes of less
than 1 mi NCG are already critical for an
MIS instrument, since this volume cove
ers large surfaces within an MIS instru-
‘ment, which then cannot be propery ster-
lised. This means that the inert gas con-
centrations that are hazardous for MIS in-
struments are more than 100-fold small
fer that than those applicable to the BD
laundry package, representing solid and
porous instruments. Cen the air detector
measure NCG concentrations that are
smaller by even two orders of magnitude?
Non-condensable gases are composed
of air and primarily CO,, The CO, is gen
erated by decomposition of hydrogen car-
bonete salts when water is heated, pro-
vided that the salts have not been previ
‘ously removed by means of an appropri
ate water treatment procedure. Asis well
known, there is outgasing of CO, and sir
(on heating the feedwater. However, CO;
dissolves once again on coming into con
tact with water and gives fise to carbon
dioxide (aerated water) at low tempera-
tures. In the air detector steam condens-
ces and, in the presence of NOGs, a two-
phase mixture is formed, composed of
‘water and gas bubbles. The gas bubbles
are used to measure NCGs. Whereas the
air dissolves in the condensate only very
slowly, alatge portion of the gaseous CO;
immediately dissolves in water. The greater
the pH value, the faster wil this dissolve.
Hence a large proportion of the CO, dis-
‘appears, thus eluding detection. Can you
ive any information on the order of mag-
nitude of the detectable CO; proportion?
We would be grateful if you could
please respond to these queries. Knowing
the possibilities and limitations of an air de-
tector is surely in the interest of all users
‘of such a system. +
[~enaimat sevice —TVatume 2 3000
Author's Reply
‘oer Eittor| |
Elem 10
Bee comer |
Witte sees ihe 26 stator
Dr Kaiser makes reference to the
timepoint at which inet gases occur t0 the
measurement sensitivity of hollow de~
vices and to detection of CO, indeed,
these points merit discussion when fo-
‘using on steam sterilisation. However,
the use of the air detector tends to be
‘based on a pragmatic approach: the air
detector simulates a medical device which,
due tots geometry and cooling properties,
hampers steam penetration. To reinforce
this effect, the air detector is placed at
the most critical site within the steriiser
‘which harbours the highest partion of non-
condensable gases. tis checked, and re-
set, for proper functioning at the time of
validation. At tis juncture it ust be bone
in mind that routine validation of the ster-
llsation process at annual intervals has
become an established practice in hospi-
tals in the meantime. When carrying out
testing within the framework of valida-
tion, thermoelectric measurements are
performed in the medical devices ~ as in
the air detector. Here observance of the
delta 2K criterion plays a pivotal role be-
cause the temperature difference mea-
‘sured is a good meesure of steam pene-
348
tration, which is of paramount importance.
Since at the time of adjustment (calibra:
tion), the air detector correlates with this,
‘measure on using the 8D test packege
(checked) (see European standard EN
285) andit simulates a aifficut-to-steriise
‘medical device, its “Go or no 90” signal
is decisive in respect of steam penetration.
This is all the more true in view of the
fact that this test is not conducted once
annually or daily but for each and every
sterilisation load. For routine operation the
air detector is a robust test system that
conforms to EN 285.However, the Ad Hoc
Group of the TC 102, WG 2+ 3 Stan-
daidisation Committee, of EN 285, has
acknowledged that the BD test is not nec-
essarily representative of hollow device
systems. This fact will probably lead to a
tightening of EN 285 in that a special ho!-
low device test system willbe introduced
(Hollow A), However, this does net at all
impinge of the use of the air detector.
Standardisation institutes and industry
have a joint task: technology and stan-
‘dards must be tallored to each other. With-
‘out anticioting future events, itis con
celvable that the air detector will be
equippedin the form of a Hollow A device
as outlined in EN 8675, This would assure
‘a direct correlation between the air de-
tector and MIS hollow devices, endowed
with corresponding sensitivity. But we
‘must first of all avait further develop-
‘ments and amendments to EN 285 which
has just been revised. +