Technical Manaual Mocom M10

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 72

97071268

Rev. 02
2021-05

EN
Dental units M10 2019 - TECHNICAL MANUAL

Contents
1. GENERAL WARNINGS ............................................................................................................................................ 3
1.1. FOREWORD ......................................................................................................................................................... 3
1.2. SAFETY INSTRUCTIONS .................................................................................................................................... 4
2. PACKING AND TRANSPORT .................................................................................................................................. 6
2.1. TRANSPORT AND STORAGE ............................................................................................................................ 6
2.2. DAMAGE DURING SHIPMENT ........................................................................................................................... 7
3. PREINSTALLATION ................................................................................................................................................. 8
3.1. DENTAL UNIT POSITION .................................................................................................................................... 8
3.2. ELECTRICAL HOOK-UP ...................................................................................................................................... 8
3.3. HYDRAULIC AND PNEUMATIC CONNECTIONS (AIR - WATER - DRAINAGE - SUCTION) ........................... 9
4. DEVICE INSTALLATION AND SETTING............................................................................................................... 11
5. ACCESSORIES AND MULTIMEDIA INSTALLATION ........................................................................................... 12
6. GUIDELINES - SAFETY TEST ............................................................................................................................... 13
6.1. CHECKS REQUIRED FOR THE DRAFTING OF CONFORMITY REPORT .....................................................13
6.2. CONFORMITY REPORT ....................................................................................................................................14
6.3. APPLICATION EXAMPLE WITH INSTRUMENT ESA 620 ON DENTAL UNIT AND/OR DENTAL CHAIR ......15
6.4. INDICATIONS FOR THE SOLUTION OF NON-CONFORMITIES DETECTED IN THE TEST .........................22
7. FUNCTIONING ....................................................................................................................................................... 24
8. KEYPADS AND DISPLAY SYMBOLS .................................................................................................................... 25
9. SERVICE MENU ..................................................................................................................................................... 26
10. HOW TO QUIT SHUTDOWN / FAULT CONDITIONS DURING DISINFECTING CYCLE .................................... 35
11. ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING ................................................................................ 37
12. SCHEDULED MAINTENANCE AND TECHNICAL SERVICE ............................................................................... 38
13. SYSTEM RESET .................................................................................................................................................... 39
14. ELECTRONIC BOARDS......................................................................................................................................... 40
14.1. DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD ...............................................................................40
14.2. BASIC HYDROGROUP BOARD ....................................................................................................................48
14.3. BASIC INSTRUMENT’S TABLE BOARD .......................................................................................................53
14.4. HYGIENE BOARD ..........................................................................................................................................58
14.5. ASSISTANT’S MODULE BOARD ..................................................................................................................61
14.6. MULTI-FUNCTION FOOT CONTROL BOARD .............................................................................................63
14.7. POWER FOOT CONTROL BOARD ...............................................................................................................66
14.8. VENUS PLUS L MCT LAMP CONTROL BOARD ..........................................................................................68

2 CONTENTS EN
Dental units M10 2019 - TECHNICAL MANUAL

1. GENERAL WARNINGS
1.1. FOREWORD
The device described in this manual is manufactured by CEFLA s.c. - via Selice Provinciale 23/A - 40026 Imola (BO)
Italia, which is the manufacturer, in compliance with the applicable European Directives detailed in the declaration of
conformity.

These technical instructions are addressed to the personnel in charge of repair and/or maintenance operations of
CEFLA s.c. equipment and contain all the necessary information.

CEFLA s.c. shall be responsible for the safety, reliability and efficiency of the equipment provided that:

 installation, any modifications, settings or repairs are made by authorised technical personnel using CEFLA s.c.
original spare parts
 In case of installation in medical locations: the electrical installation of the relevant location complies with IEC
60364-7-710:2002 Standards (Standards on electrical installations of medical locations) or with the equivalent
Standards in force in the country of installation
 the equipment is used as outlined in User Manual
 in case of installation of an X-ray unit: The room where the X-ray unit is installed complies with the official Directives
on protection from radiation in the country of use

EN GENERAL WARNINGS 3
Dental units M10 2019 - TECHNICAL MANUAL

1.2. SAFETY INSTRUCTIONS


Information that draws the user’s attention to the need of preventing dangerous situations and ensuring the correct
and safe use of the device is reported in the text according to the following layout:

SYMBOL DESCRIPTION

WARNING: General obligation

WARNING: It is mandatory to wear safety gloves

WARNING: It is mandatory to wear safety shoes

WARNING: Use eye protections

WARNING: Use protective clothing

WARNING: It is mandatory to ensure that there is an earth connection

WARNING: See the instruction manual

WARNING: Disconnect the device from the power mains

WARNING: See the technical manual

DANGER: It indicates a situation in which the failure to follow the instructions may lead to a fault of the
equipment or injuries to the user

DANGER: High voltage

DANGER: Crushing of hands

DANGER: Crushing of feet

DANGER: X-ray emission

DANGER: Moving parts

DANGER: Class 1 LASER radiation

4 GENERAL WARNINGS EN
Dental units M10 2019 - TECHNICAL MANUAL

SYMBOL DESCRIPTION

DANGER: Class 2 light source.

DANGER: Hot surface

DANGER: Biological risk

DANGER: Sharp objects

DANGER: Corrosive substances

DANGER: Flammable materials

DANGER: Tipping over

DANGER: Heavy object. Two persons are required

WARNING!

ALWAYS DISCONNECT power supply before performing any maintenance or cleaning operation.

The packed device must be handled using, where possible, suitable mechanical means (forklift, pallet
truck, etc.) and following the indications on the package.
In case of manual handling, the device must be lifted by several persons using the suitable available
means and, if possible, it must be moved with a truck or similar means.

Wear the proper protections.

EN GENERAL WARNINGS 5
Dental units M10 2019 - TECHNICAL MANUAL

2. PACKING AND TRANSPORT


2.1. TRANSPORT AND STORAGE
Instructions on how to handle, store and open the package are given on the outside of the cardboard box.
These instructions must be carefully followed.

1) Transport and store only in the direction indicated by the


arrows

2) Do not overlap more than two units

3) Avoid collisions

4) Protect against humidity

5) Do not use hooks to handle

6) The required storage conditions are indicated:


a) temperature from - -10° to +70° C

b) relative humidity from 10 to 90%

c) atmospheric pressure from 50 to 106 kPa

It is recommended to transport and store the device at a temperature not lower than that indicated on
the packing. A prolonged exposure to a low temperature can damage the device.

The packed device must be handled using, where possible,


suitable mechanical means (forklift, pallet truck, etc.) and
following the indications on the package.

In case of manual handling, it must be lifted by two or more


persons using the suitable available means.

6 PACKING AND TRANSPORT EN


Dental units M10 2019 - TECHNICAL MANUAL

Packaging Procedure PRECAUTION

Box 1 - remove the metal staples - do not handle the dental chair by
fastening the cardboard to the its backrest
wooden base and lift up - pay attention not to lift the dental
- loosen the two screws fixing the chair from its white plastic parts
dental chair to the wooden base
Box 2 - remove the metal staples - only remove the dental unit from
fastening the cardboard to the the wooden base when you
wooden base and lift up decide to attach it to the dental
chair

2.2. DAMAGE DURING SHIPMENT


When the device is received, check the packing container for any damage suffered.

If the packages are found to be damaged on delivery, accept them with reserve by signing the delivery note and
indicating that the “CONTENTS ARE ACCEPTED BUT THEY NEED TO BE CHECKED”.

If the contents are actually damaged, notify the shipping agent and request insurance compensation for damage
within five work days. The claim is to be made by the person who commissioned the shipping agent. I.e.:
 if the goods are delivered “carriage forward”, the receiver shall notify the shipping agent
 if the goods are shipped “carriage free” or “free delivered”, the consigner shall notify the shipping agent and file a
claim for damages. In this case, CEFLA s.c. shall be informed as soon as possible.

The damaged parts returned to CEFLA s.c. for replacement shall be placed inside the same damaged package.

Return to CEFLA s.c. shall be “carriage free” (at the expense of the consigner), while shipment back to the customer
shall be at the expense of CEFLA s.c. (carriage free).

EN PACKING AND TRANSPORT 7


Dental units M10 2019 - TECHNICAL MANUAL

3. PREINSTALLATION
3.1. DENTAL UNIT POSITION
Before installing the dental unit and dental chair, prepare the surgery considering the overall dimensions (see the
installation layout available on EXTRANET).
In addition, take all the required internal and external electrical and water connections close to the connections box.
To perform this operation correctly, it is essential to refer to the 1:1 scale installation layout (available on
EXTRANET).

WARNING!
- Installation must always be permanent with anchoring to the floor (Standard IEC EN 60601).
- the (continuous) floor should meet the load-bearing capacity set forth by "DIN 1055". The total weight
of the dental unit with a 190 Kg patient is approximately 395 kg

3.2. ELECTRICAL HOOK-UP


WARNING!
The system must have an effective earth ground that conforms to local regulations.
In particular, in Italy, this is to be done in compliance with the IEC 60364-7-710:2002 specifications that require the
installation upstream the system of a differential switch with the following characteristics:

If the voltage is = 230V (+/- 10V) - 50HZ


- rated current: 10A (D TURN)
- differential sensitivity: 0.03A

If the voltage is = 100/115V (+/- 10V) - 60HZ


- rated current: 16A (D TURN)
- differential sensitivity: 0.03A

Power supply:
- normally, the equipment is supplied for 230V - 50Hz. Different power supply voltage values (e.g. 115V / 60Hz) are
available on request
- absorbed power: 1500W
- power supply: 3x1.5mm type H05 RRF
WARNING!
The colour of the 3 wires must be as set forth by the standards.
- auxiliary control: 2x0.5mm type H05 RRF
- suction system control: 3x1mm type H05 RRF

8 PREINSTALLATION EN
Dental units M10 2019 - TECHNICAL MANUAL

3.3. HYDRAULIC AND PNEUMATIC CONNECTIONS (AIR - WATER - DRAINAGE -


SUCTION)
It is recommended to fit a cock, to stop both water and air delivery, so as to disconnect the dental unit from the other
appliances of the dental surgery. Should this not be possible, disconnect the unit using two cocks inside the
connections box.

Air:
- inlet pressure between 6-8 Bar
- actual flow rate: 82 litres/minute
- free of impurities
- dry
- free of oil
- if needed, install a heavy duty filter before the air supply: 40-50 micron
- connection at floor level: 1/2 Gas female fitting
- before installing, drain the duct

Water:
- inlet pressure between 3-5 Bar
(IF THE PRESSURE EXCEEDS 5 BAR, INSTALL A PRESSURE REDUCER AT THE INLET OF THE DENTAL
UNIT)
- actual flow rate of approx. 10 litres/minute
- free of impurities
- untreated potable water hardness no greater than the values indicated in the following table (where the water
shows higher values the installation of a water treatment unit is advised):

PRECAUTION:
If “ion exchange resin water softeners” are used the treated water hardness rating must not be less than
15°f in order to protect certain machine components.
There are treatments, such as ”reverse osmosis”, that reduce water hardness by removing the mineral
salts present in the water without any softening process; with this method treated water with hardness
ratings of less than 15°f does not cause damage to the machine.
Other “carbon” or “magnetic” devices do not alter water hardness.

in mg/L
in °T (°dH) (CaCO3 = in mg/L (CaO =
WATER HARDNESS in °f (French in °I (English
(German calcium calcium oxide
CLASSIFICATION degrees) degrees)
degrees) carbonate concentration)
concentration)
VERY SOFT <7 < 3.92 < 4.9 < 70 < 39.2

SOFT 7 < 14 3.92 < 7.84 4.9 < 9.8 70 < 140 39.2 < 78.4

QUITE HARD 14 < 22 7.84 < 12.32 9.8 < 15.4 140 < 220 78.4 < 123.2

MILDLY HARD 22 < 32 12.32 < 17.92 15.4 < 22.4 220 < 320 123.2 < 179.2

HARD 32 < 54 17.92 < 30.24 22.4 < 37.8 320 < 540 179.2 < 302.4

VERY HARD > 54 > 30.24 > 37.8 > 540 > 302.4

EN PREINSTALLATION 9
Dental units M10 2019 - TECHNICAL MANUAL

WATER HARDNESS UNIT OF MEASURE CONVERSION TABLE

UNIT OF MEASURE UNIT OF MEASURE CONVERSION COEFFICIENTS

°f
- x 0.56 x 0.7 x 10 x 5.6
(French degrees)

°T (°dH)
x 1.79 - x 1.25 x 17.9 x 10
(German degrees)

°I
x 1.43 x 0.8 - x 14.3 x8
(English degrees)
mg/L (CaCO3 = calcium
carbonate x 0.1 x 0.056 x 0.069 - x 0.56
concentration)
mg/L (CaO = calcium
x 0.179 x 0.1 x 0.125 x 1.79 -
oxide concentration)

= mg/L = mg/L
= °f = °T (°dH) = °I
(CaCO3) (CaO)

- connection at floor level: 1/2 Gas female fitting


- before installing, drain the duct

Drainage:
- minimum flow rate of 10 litres per minute
- pipe internal diameter of 40mm, with outlet at floor level
- min. inclination: 1cm/m.

Suction:
- flow rate no less than 450 litres/minute
- pipe internal diameter of 40mm, with outlet at floor level

PRECAUTION
A good suction system should have 45° and not 90° bends.

10 PREINSTALLATION EN
Dental units M10 2019 - TECHNICAL MANUAL

4. DEVICE INSTALLATION AND SETTING


See the installation manual of the concerned device, available on EXTRANET.

EN DEVICE INSTALLATION AND SETTING 11


Dental units M10 2019 - TECHNICAL MANUAL

5. ACCESSORIES AND MULTIMEDIA INSTALLATION


See the installation manual of the concerned device, available on EXTRANET.

12 ACCESSORIES AND MULTIMEDIA INSTALLATION EN


Dental units M10 2019 - TECHNICAL MANUAL

6. GUIDELINES - SAFETY TEST


Checks to be carried out upon first installation, periodical checks (every year) and tests to be carried out after
repair/updates/settings of medical electrical equipment with reference to Standard IEC 62353

Safety standards for medical electrical equipment classified according to IEC 60601 in Class 1, Type B.
(ME equipment connected to mains power supply through a permanent connection, provided with protective earth
conductor).

NOTE If measurements cannot be carried out in compliance with Standard IEC 62353, they can be carried
out in compliance with Standard IEC 60601.

6.1. CHECKS REQUIRED FOR THE DRAFTING OF CONFORMITY REPORT


a) Visual inspection
The purpose of visual inspection is to ensure that the equipment in use still complies with the specifications
released by the manufacturer through the following checks:
• label data
• presence of documentation (e.g. updated user manual in printed or electronic format, instructions)
• conditions of warning plates
• condition of power cables and connectors
• condition of earth cables and correct tightening
• condition of machine protections, covers and casings
• condition of electrical board protections and correct value of the fuse on the power supply line
• conditions of light indications and acoustic signals (warning lights, LEDs, display, etc.)
• no liquid leaking out from ducts or on the bottom of the device

b) Electric measure test (to be carried out with special instruments)


• measurement of earth conductor resistance
• measurement of earth conductor leakage currents
• measurement of earth-leakage currents of the medical electrical equipment and patient
• measurement of insulation resistance between network and earth conductor
• measurement of insulation resistance between applied part and earth conductor

c) Functional checks
• check of movement controls from doctor’s and assistant’s console
• check of usage controls of the instruments applied
• check of light indications and/or acoustic signals
• check of emergencies

EN GUIDELINES - SAFETY TEST 13


Dental units M10 2019 - TECHNICAL MANUAL

6.2. CONFORMITY REPORT

14 GUIDELINES - SAFETY TEST EN


Dental units M10 2019 - TECHNICAL MANUAL

6.3. APPLICATION EXAMPLE WITH INSTRUMENT ESA 620 ON DENTAL UNIT


AND/OR DENTAL CHAIR
 Measurement of protective earth conductor resistance
 Measurement of leakage currents in earth conductor, casing and patient
 Measurement of insulation resistance between network and earth
 Measurement of insulation resistance between applied part and earth

Disconnect power supply cable from dental unit and connect it (Phase - Neutral - Earth) to the relevant instrument
socket (e.g.: ESA 620). Turn on the main switch of the dental unit.

NOTE It is advisable to follow the instructions below: the device under test must be on and must be pre-
heated in advance through an operating cycle of at least 30 minutes, until it reaches the operating
temperature.

From PC or tester screen, select the type of test with reference to Standard IEC 62353 (among the methods
specified for the test of earth-leakage currents, the direct method is equivalent to Standard IEC 60601).

Follow the instructions displayed with reference to the visual check and dental unit data entry (e.g.: device serial
number).

If the test also includes the applied part (instruments), insert the grippers of the “APPLIED PARTS” on the relevant
measurement points: one gripper in the metal end part of the syringe [A]; one gripper on the tip of scaler, nose of
micromotor, quick coupling of the turbine [B] (in case of turbines with integrated LED, install the turbine on the
tubing); one gripper in 0V of power supply box [C]; one gripper in the monitor chassis [D].

NOTE If the monitor is not present, apply the gripper to V-OUT connector of the multimedia board, in
the point corresponding to 0 Volt, as shown in picture [D1].
In case there are tools also on the assistant’s side [E], manage them as described above for points [A]
and [B].

INITIAL SETTING

EN GUIDELINES - SAFETY TEST 15


Dental units M10 2019 - TECHNICAL MANUAL

NOTE Connections with the machine points must be kept for the tests described below.
Follow the instructions requested by the control instrument.

E.G.: ESA620 instrument connection

A: Syringe
B: Instruments
C: Secondary circuit
D: Multimedia circuit
E: Applied part
F: Equipment under test
WARNING: During measurement execution, hazardous voltage could remain in the device.

Press “Start” to start test sequences.

Depending on your instrument, you have to select all the tests indicated below:
A) measurement of earth conductor resistance
B) measurement of earth-leakage current
C) measurement of insulation resistance

A) Measurement of earth conductor resistance (to be carried out after setting the test current of 25 Amperes in the
tester)

16 GUIDELINES - SAFETY TEST EN


Dental units M10 2019 - TECHNICAL MANUAL

NOTE (Only with test carried out with reference to Standard IEC 62353):
• the measurement instrument must be able to supply a current of at least 200mA and the open circuit
supply voltage must not exceed 24V
• before carrying out the measurements, it is necessary to calibrate the power supply cable between
measurement instrument and equipment under test
• the measurement is carried out via probe between the metal parts of the equipment (see indicated
points) and the protective conductor connection in the equipment

Measurement method

Connect one gripper to the main earth point (connections box).

Connect the second gripper (if any, otherwise use the main one, by repositioning it), in sequence, in the various
points indicated:
• connections box metal part [A] • auxiliary socket [B]

EN GUIDELINES - SAFETY TEST 17


Dental units M10 2019 - TECHNICAL MANUAL

• dental chair base mechanical part [C] • inverter board support sheet [D]

• metal part on dental chair electric motors (seat and backrest, if any) [E] or on dental chair metal body [E]
DENTAL CHAIR WITH MOTORS
THAT
OPERATE AT MAINS VOLTAGE

E1: Motor

DENTAL CHAIR WITH MOTORS THAT


DO NOT OPERATE AT MAINS
VOLTAGE

18 GUIDELINES - SAFETY TEST EN


Dental units M10 2019 - TECHNICAL MANUAL

• hydrogroup metal part [F]

• instrument’s table metal part [G].

• integrated X-ray unit [H]

EN GUIDELINES - SAFETY TEST 19


Dental units M10 2019 - TECHNICAL MANUAL

B) Measurement of earth-leakage current

Select the specific test on the tester.

By performing the test with reference to standard IEC 62353, it is possible to select an alternative method, that
carries out the test without powering the device under test. Generally this test is carried out automatically by the
instrument that carries out the test.
NOTE Pay attention to properly set the test equipment for “earth conductor” earth-leakage currents

NOTE The limit of leakage current in earth conductor for permanently installed M.E. equipment is ≤ 5
mA (see notes in the manual of the instrument used to carry out this setting)

Measurement method - Alternative method

If the test is carried out with reference to standard IEC 62353, by selecting the direct method, that carries out the
test by powering the device under test, the check is comparable to the measurements carried out according to
standard IEC 60601.
NOTE The guidelines below refer to the test carried out with direct method

Measurement method - Direct method


NOTE The direct method is comparable to the
measurements carried out in accordance with
Standard IEC 60601-1, it allows detecting any
leakage of current when the equipment and its parts
(heaters, instruments, etc.) are in full operation.

• “Earth conductor”: start the test; it must be carried out both in normal use and in single failure conditions
(interruption of a network conductor at a time) - (normal tester automatism).
NOTE The limit of leakage current in earth conductor for permanently installed M.E. equipment is ≤ 5
mA (see notes in the manual of the instrument used to carry out this setting)

20 GUIDELINES - SAFETY TEST EN


Dental units M10 2019 - TECHNICAL MANUAL

• “Casing”: start the test; it must be carried out both in normal use and in single failure conditions (interruption of
a network conductor at a time and interruption of the earth conductor*, ref. IEC 60601) - (normal tester
automatism).
NOTE In single failure conditions, the interruption of the earth conductor does not apply to the
protective earth conductor of permanently installed M.E. Equipment, whose disconnection is considered
unlikely (EN 60601_8.1.b_4th)

• “Patient”: start the test; it must be carried out both in normal use and in single failure conditions (interruption of
a network conductor at a time and interruption of the earth conductor, ref. IEC 60601) - (normal tester
automatism).
NOTE To measure patient leakage current in FULL OPERATION for each electrical instrument both on
dentist side and on assistant side (micromotor and scaler), it is necessary to:
• individually take out the instrument to be checked;
• set the maximum speed;
• activate the spray;
• move the pedal lever (if the scaler is available, to allow tip vibration, it is necessary to disconnect the
gripper, activate the scaler, then reconnect the gripper to the vibrating tip);
• finally, start the tester for value measurement.

• “Patient Auxiliary”
NOTE This test must be carried out only if the syringe with heater is available (dentist side or assistant
side).
Start the test; it must be carried out both in normal use and in single failure conditions (interruption of a
network conductor at a time and interruption of the earth conductor, ref. IEC 60601) - (normal tester
automatism).
NOTE To measure patient auxiliary leakage current in FULL OPERATION, it is necessary that:
• the syringe heater is activated;
• take out the scaler or the micromotor (if the scaler is not present);
• move the pedal lever (if the scaler is available, to allow tip vibration, it is necessary to disconnect the
gripper, activate the scaler, then reconnect the gripper to the vibrating tip);
• finally, start the tester for value measurement.

C) Measurement of insulation resistance

Select the specific test on the tester:

• “between network and earth”: this test does not require special wiring, the tester will perform it automatically
after your authorisation to proceed. Let all the probes connected as in the previous points.
NOTE PHASE and NEUTRAL are short-circuited with each other by the tester.

NOTE The test must last one minute.

EN GUIDELINES - SAFETY TEST 21


Dental units M10 2019 - TECHNICAL MANUAL

• “between applied part and earth”: this test does not require special wiring, the tester will perform it
automatically after your authorisation to proceed. Let all the probes connected as in the previous points.
NOTE The test must last one minute.

The tester issues (where applicable) a paper report with test results.
a) Check that all the tests are passed.
b) Add the values to the conformity report only in the absence of the paper report (otherwise attach the report and
tick the tests in the report as performed).
c) If one or more tests have not been passed, see section 4 (“Indications for the solution of non-conformities
detected in the test”).

WARNING: If safety checks are not passed, the dental unit must NOT be used; before using it
again, the device must be repaired by an authorised technician

Now it is necessary to check that the dental unit is correctly operating by carrying out a complete check of dental unit
functions (see section 1.C - adding the result in the report).

File the report in the warranty and maintenance booklet of the dental unit.

6.4. INDICATIONS FOR THE SOLUTION OF NON-CONFORMITIES DETECTED IN THE


TEST
EARTH CONDUCTOR RESISTANCE

 Check the presence and continuity of the earth conductor between the measurement point and the main
earth node and in the intermediate sections
 Check correct tightening of the fasteners of earth conductors

INSULATION RESISTANCE

ANALYSIS ON DENTAL CHAIR


 Disconnect direct power supply to the dental chair and repeat the test
 If the test is passed, reconnect power supply to the dental chair and disconnect power supply of motors and
capacitors (if powered with network voltage) and repeat the test
 If the test is passed, reconnect motors and capacitors one by one and repeat the test until the non-compliant
component has been identified

ANALYSIS ON DENTAL UNIT


 Disconnect cup heater, if any, and repeat the test
 If the test is not passed, disconnect spray heater, if any, and repeat the test
 If the test is not passed, disconnect any syringes with heater (6F) and repeat the test
 If the test is not passed, replace the connections board to exclude a possible network filter failure

LEAKAGE CURRENT IN EARTH CONDUCTOR

ANALYSIS ON DENTAL CHAIR

22 GUIDELINES - SAFETY TEST EN


Dental units M10 2019 - TECHNICAL MANUAL

 Disconnect direct power supply to the dental chair and repeat the test (if the test is not passed, perform
ANALYSIS ON DENTAL UNIT by reconnecting the dental chair power supply)
 If the test is passed, reconnect power supply to the dental chair and disconnect power supply of motors and
capacitors (if powered with network voltage) and repeat the test
 If the test is passed, reconnect motors and capacitors one by one and repeat the test until the non-compliant
component has been identified

ANALYSIS ON DENTAL UNIT


 Disconnect cup heater, if any, and repeat the test
 If the test is not passed, disconnect spray heater, if any, and repeat the test
 If the test is not passed, disconnect any syringes with heater (6F) and repeat the test
 If the test is not passed, replace the connections board to exclude a possible network filter failure

PATIENT LEAKAGE CURRENT

INSTRUMENT’S TABLE ANALYSIS


 Disconnect power supply to the dental chair and repeat the test (if the test is not passed, perform
HYDROGROUP ANALYSIS by reconnecting the instrument’s table power supply)
 If the test is passed, reconnect instrument’s table power supply, disconnect all the instrument board power
supplies and repeat the test
 If the test is passed, connect in sequence power supplies of instrument boards (repeat the test upon each
connection) until leakage origin has been detected
 Disconnect in sequence solenoid valves, tubing, handpiece and repeat the test until leakage origin has been
detected

HYDROGROUP ANALYSIS
 Disconnect hydrogroup board power supply and repeat the test
 If the test is passed, reconnect power supply to the hydrogroup, disconnect in sequence power supplies of
instrument boards in assistant's module (repeat the test upon each connection) until leakage origin has been
detected

EN GUIDELINES - SAFETY TEST 23


Dental units M10 2019 - TECHNICAL MANUAL

7. FUNCTIONING
See the user manual of the concerned dental unit.

24 FUNCTIONING EN
Dental units M10 2019 - TECHNICAL MANUAL

8. KEYPADS AND DISPLAY SYMBOLS


See the user manual of the concerned dental unit.

EN KEYPADS AND DISPLAY SYMBOLS 25


Dental units M10 2019 - TECHNICAL MANUAL

9. SERVICE MENU

26 SERVICE MENU EN
Dental units M10 2019 - TECHNICAL MANUAL

SERVICE MENU

EN SERVICE MENU 27
Dental units M10 2019 - TECHNICAL MANUAL

INSTRUMENT MENU

28 SERVICE MENU EN
Dental units M10 2019 - TECHNICAL MANUAL

GENERAL SERVICE MENU

EN SERVICE MENU 29
Dental units M10 2019 - TECHNICAL MANUAL

CONFIGURATION MENU

(1 = Function enabled / 0 = Function disabled)

DENTAL UNIT CONFIGURATION

SIGNALS DESCRIPTION

GLASS_HEATER 1 = if the cup water heater is present.

MOTORIZED_BOWL 1 = if the motorised cuspidor is present.

GLASS_SENSOR 1 = if the cup sensor is present.

WATER_ CIRCULATION 1 = if anti-stagnation feature is present.

CANISTER_CATTANI 1 = if the Cattani separator is present.

BOWL 1 = if the basin is present.

SUCTION_WASH 1 = if suction washing is present.

CENTRAL_SUCT_VALVE 1 = if the centralized suction valve is present.

HOSE_SELECTION 1 = if the independent suction tube selection is present.

30 SERVICE MENU EN
Dental units M10 2019 - TECHNICAL MANUAL

SIGNALS DESCRIPTION

PERISTALTIC_PUMP 1 = if the peristaltic pump is present.

H20_SPRAY_HEATER 1 = if the spray water heater is present.

BRAKE_SENSOR 1 = if the arm brake electric sensor is present.

H20_BOTTLE 1 = if the bottle is present.

H2O2_DISINFECTION 1 = if the automatic intensive disinfection system is present.

H2O_HYGIEN_SYSTEM 1 = if the automatic continuous sanitising system is present.

WATER_BYPASS_EV 1 = if the bypass valve is present.

CHAIR_ROTATION_EV 1 = if the dental chair rotation is present.

THREEPHASE_CHAIR 1 = if the dental chair seat three-phase motor is present.

SLIDING 1 = if the dental chair sliding motor is present.

DOCTOR_TABLE 1 = if the instruments table board is present.

HYGIENE_BOARD 1 = if the hygiene board is present.

HYDRO_GROUP 1 = if the hydrogroup board is present.

ORTHO 1 = in case of Ortho dental unit.

SIDE_DELIVERY 1 = in case of Side Delivery ergonomics.

ASSISTANT_TABLE 1 = if the assistant's module is present.

CHAIR_EXT 1 = if the dental chair with split leg rest is present.

LAMP_IN_HYDRIC 1 = if the lamp is connected to the hydrogroup board.

LAMP_IN_CHAIR 1 = if the lamp is connected to the dental chair/connections board.

RECIPROCATING 1 = if the reciprocating mode is present.

FLUSHING 1 = if the Flushing is present.

PEDAL_TYPE Selection of the foot control type.

LAMP_TYPE Selection of the lamp type.

CONSOLE Selection of the console type.

MODEL Selection of the dental unit model.

ERGONOMICS Selection of the ergonomics type.

EN SERVICE MENU 31
Dental units M10 2019 - TECHNICAL MANUAL

SIGNALS DESCRIPTION

DOC_TABLE_TYPE Selection of the instruments table type.

DOC_TABLE_HOLDER Selection of the table arm type.

SYRINGE_POSITION Selection of the syringe position.

APEX YES = if the apex locator is present.

POWER RANGE +/- 10V automatic.

WIRELESS_PEDAL 1 = if the wireless foot control is present.

INSTRUMENT_A Selection of the instrument type in module A.

INSTRUMENT_B Selection of the instrument type in module B.

INSTRUMENT_C Selection of the instrument type in module C.

INSTRUMENT_D Selection of the instrument type in module D.

INSTRUMENT_E Selection of the instrument type in module E.

SYRINGE_TYPE Selection of the syringe type.

ASSIST_INSTRUMENT_1 Selection of the instrument type in assistant module 1.

ASSIST_INSTRUMENT_2 Selection of the instrument type in assistant module 2.

ASSIST_INSTRUMENT_3 Selection of the instrument type in assistant module 3.

DENTAL CHAIR CONFIGURATION

SIGNALS DESCRIPTION

BOWL_AND_ZEROPOSITION Bowl rotation with home position.

BOWL_AND_RINSEPOSITION Bowl rotation with rinsing position.

BOWL_AND_ABCD_POS Bowl rotation with ABCD programs.

0 = in case of dental chair interference, the bowl moves towards the lamp
post.
FAST_RETURN_BOWL
1 = in case of dental chair interference, the bowl moves from the fastest side.

ZEROPOS_BOTTLE_PRESS_OFF Bottle depressurisation with home position.

GLASS_AND_RINSEPOSITION Cup filling with rinsing position.

LAMPOFF_AND_
Lamp OFF with home position.
ZEROPOSITION

LAMPOFF_AND_RINSEPOSITION Lamp OFF with rinsing position.

32 SERVICE MENU EN
Dental units M10 2019 - TECHNICAL MANUAL

SIGNALS DESCRIPTION

BOWL_WATER_AND_RINSEPOSITIO
Bowl rinsing with rinsing position.
N

BOWLWATER_AND_LASTPOSITION Bowl rinsing with last dental chair position recall.

BOWLWATER_AND_ZEROPOSITION Bowl rinsing with home position.

PEDAL_LEVER_REVERSE Foot control lever MIN/MAX inversion.

JOYSTICK_EXTENDED_MODE Enabling of foot control joystick extended mode.

MONOSTABLE_BELL Monostable bell button.

1 = to use the lamp button that activates the bell relay (basic or CL version
EXTERNAL_LAMP
lamp).

POWERPEDALPROGRAM Dental chair programs available from Power foot control.

AUTOSTANDBY Automatic standby (enabled).

STDBYTIME Standby activation time (minutes).

HYDROGROUP CONFIGURATION

SIGNALS DESCRIPTION

BOWL_TIMER Bowl timer (Enabled / Disabled).

SUCTIONSTOP_DELAY Suction stop delay (Enabled / Disabled).

SUCTIONSTOP_BIG_HOSE Suction stop on big suction tube.

SUCTIONSTOP_SMALL_HOSE Suction stop on small suction tube.

WATER_BOWL_AND_GLASS Cup/bowl automatism.

GLASS_HEATER Cup heater presence.

GLASS_SENSOR Cup sensor presence.

H2O_BOTTLE_STATUS Distilled water bottle presence.

ASSISTANT_PROGRAMENABLED Enabling of assistant dental chair programs.

LAMP_SENSOR Presence of operating light sensor.

LAMP_SOFT_START Lamp light soft start.

0 = switching from standard light to composite light via sensor.


LAMP_SENSOR_MODE
1 = ON/OFF lamp via sensor.

GLASS_TIME Cup water filling time via button.

EN SERVICE MENU 33
Dental units M10 2019 - TECHNICAL MANUAL

SIGNALS DESCRIPTION

GLASS_SENSOR_TIME Cup water filling time via sensor.

BOWL_TIME Bowl water time.

GLASS_HEATER_TEMP Cup water temperature.

BOWL_SPEED Bowl rotation speed.

GLASS_SENSOR_SENS Cup sensor sensitivity.

BOWL_VALVE_DELAY Bowl drain valve switching off delay.

INSTRUMENT'S TABLE CONFIGURATION

SIGNALS DESCRIPTION

SYRINGEFO_ON Syringe fiber optic enabling.

APEXSOUNDON Apex locator ON/OFF alarm sound.

NOWARNINGL1L2 No basic console error messages.

DOCTORPROGRAMENABLED Enabling of doctor’s console dental chair (ABDC) programs.

TLEDAUTO 1 = for automatic operating lamp light change upon T-LED activation.

SURGYAUTO ALWAYS 0.

HEATER Spray water heater presence.

AUTOCHIPAIR Automatic chip air ON/OFF.

AUTOCHIPAIR_DELAY Automatic chip air delay.

AUTOCHIPAIR_TIMING Automatic chip air time.

SYRINGEFO_LEVEL Syringe fiber optic level.

ALARMAPEXLEVEL Apex locator alarm level.

CLOUD CONFIGURATION

SIGNALS DESCRIPTION

STATION_SSID WiFi network name.

STATION_PWD WiFi network password.

34 SERVICE MENU EN
Dental units M10 2019 - TECHNICAL MANUAL

10. HOW TO QUIT SHUTDOWN / FAULT CONDITIONS DURING


DISINFECTING CYCLE

WARNING!
If H2O2 is present in the ducts, wash the cords manually with the relative foot control and flush the
water to cup duct by pressing the fill key several times.

EN HOW TO QUIT SHUTDOWN / FAULT CONDITIONS DURING DISINFECTING CYCLE 35


Dental units M10 2019 - TECHNICAL MANUAL

WARNING!
If H2O2 is present in the ducts, wash the cords manually with the relative foot control and flush the
water to cup duct by pressing the fill key several times.

36 HOW TO QUIT SHUTDOWN / FAULT CONDITIONS DURING DISINFECTING CYCLE EN


Dental units M10 2019 - TECHNICAL MANUAL

11. ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING


An up-to-date list of Alarms/Errors/Warnings/Troubleshooting can be found in the TECHNICAL MANUALS
section of the EXTRANET.

EN ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING 37


Dental units M10 2019 - TECHNICAL MANUAL

12. SCHEDULED MAINTENANCE AND TECHNICAL SERVICE


The Manufacturer requires that periodic scheduled maintenance operations are carried out at least once every 12
months in order to maintain all the certifications related to the device and the manufacturer’s responsibility for the CE
marking of the product.
To this end, special PERIODIC MAINTENANCE KITS, which include all components that need to be replaced due to
wear, are provided based on the concerned device.
The technician who carries out scheduled maintenance operations takes care of the correct and truthful filling in of
the GUIDE SHEET PROVIDED FOR PERIODIC PREVENTIVE MAINTENANCE OPERATIONS (ANNUAL).

The GUIDE SHEET must be handed over to the owner of the device and constitutes conclusive proof of the control.

Following periodic scheduled maintenance operations, the technician must ensure that the “Electrical safety tests”
specified in chapter 6 of this manual are performed by him/her or other authorised personnel.

It also resets the “reset & maintenance” counter to set the periodic maintenance counter to zero via the easyaccess
application.

In case of operations on the device on which an easycheck licence is activated, the technician is required to update
the “Maintenance Booklet” of the device on easycheck Cloud portal.

NOTE The manufacturer requires a compulsory general service of the turbines 3 years after installation by
returning the devices to the headquarters or authorised repair centres.
This service must then be repeated every three years.

38 SCHEDULED MAINTENANCE AND TECHNICAL SERVICE EN


Dental units M10 2019 - TECHNICAL MANUAL

13. SYSTEM RESET

Use easyaccess App.

EN SYSTEM RESET 39
Dental units M10 2019 - TECHNICAL MANUAL

14. ELECTRONIC BOARDS


14.1. DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD
97661871 DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD

40 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD FUSES

EN ELECTRONIC BOARDS 41
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD LED

42 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD LED

EN ELECTRONIC BOARDS 43
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR-CONNECTIONS BOX BASIC BOARD LED

44 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR MOVEMENT LIMIT SWITCH SETTING (AUTOMATIC SETTING)

EN ELECTRONIC BOARDS 45
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR MOVEMENT LIMIT SWITCH SETTING (AUTOMATIC SETTING)

46 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661871 DENTAL CHAIR MOVEMENT LIMIT SWITCH SETTING (AUTOMATIC SETTING)

EN ELECTRONIC BOARDS 47
Dental units M10 2019 - TECHNICAL MANUAL

14.2. BASIC HYDROGROUP BOARD


97661873 BASIC HYDROGROUP BOARD

48 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661873 BASIC HYDROGROUP BOARD FUSES

EN ELECTRONIC BOARDS 49
Dental units M10 2019 - TECHNICAL MANUAL

97661873 BASIC HYDROGROUP BOARD LEDS

50 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661873 BASIC HYDROGROUP BOARD LEDS

EN ELECTRONIC BOARDS 51
Dental units M10 2019 - TECHNICAL MANUAL

97661873 SETTING BASIN ROTATION LIMIT SWITCH

52 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

14.3. BASIC INSTRUMENT’S TABLE BOARD


97661874 INSTRUMENT'S TABLE BASIC BOARD

EN ELECTRONIC BOARDS 53
Dental units M10 2019 - TECHNICAL MANUAL

97661874 INSTRUMENT’S TABLE BASIC BOARD FUSES

54 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661874 INSTRUMENT’S TABLE BASIC BOARD LEDS

EN ELECTRONIC BOARDS 55
Dental units M10 2019 - TECHNICAL MANUAL

97661874 INSTRUMENT’S TABLE BASIC BOARD LEDS

56 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661874 INSTRUMENT’S TABLE BASIC BOARD LEDS

EN ELECTRONIC BOARDS 57
Dental units M10 2019 - TECHNICAL MANUAL

14.4. HYGIENE BOARD


97661903 HYGIENE BOARD

58 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661903 HYGIENE BOARD LEDS

EN ELECTRONIC BOARDS 59
Dental units M10 2019 - TECHNICAL MANUAL

97661903 HYGIENE BOARD LEDS

60 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

14.5. ASSISTANT’S MODULE BOARD


97661907 ASSISTANT’S MODULE BOARD

EN ELECTRONIC BOARDS 61
Dental units M10 2019 - TECHNICAL MANUAL

97661907 ASSISTANT’S MODULE BOARD LEDS

62 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

14.6. MULTI-FUNCTION FOOT CONTROL BOARD


97661933 MULTIFUNCTION FOOT CONTROL BOARD

EN ELECTRONIC BOARDS 63
Dental units M10 2019 - TECHNICAL MANUAL

97661933 MULTIFUNCTION FOOT CONTROL BOARD LED

64 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661933 SETTING PROCEDURE

EN ELECTRONIC BOARDS 65
Dental units M10 2019 - TECHNICAL MANUAL

14.7. POWER FOOT CONTROL BOARD


97661934 POWER FOOT CONTROL BOARD

66 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661934 POWER FOOT CONTROL BOARD

EN ELECTRONIC BOARDS 67
Dental units M10 2019 - TECHNICAL MANUAL

14.8. VENUS PLUS L MCT LAMP CONTROL BOARD


97661955 VENUS PLUS L MCT LAMP CONTROL BOARD

68 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661955 VENUS PLUS L MCT LAMP CONTROL BOARD LEDS

EN ELECTRONIC BOARDS 69
Dental units M10 2019 - TECHNICAL MANUAL

97661955 POTENTIOMETER SETTING PROCEDURE

70 ELECTRONIC BOARDS EN
Dental units M10 2019 - TECHNICAL MANUAL

97661955 NO-TOUCH SENSOR SENSITIVITY SETTING

EN ELECTRONIC BOARDS 71
www.cefla.com

You might also like