Professional Documents
Culture Documents
MenAMU Oxytrace 00
MenAMU Oxytrace 00
MenAMU Oxytrace 00
Operator’s Manual
A-96.250.541 / 080117
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
ID: A-96.250.541
Revision Issue
00 January 2008
Table of Contents
1. Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.1 Legal Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.2 General Inspection and Maintenance Duties . . . . . . . . 5
1.3.3 Spare Parts and Disposables . . . . . . . . . . . . . . . . . . . 5
1.3.4 Modifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.5 Electrical shock hazard . . . . . . . . . . . . . . . . . . . . . . . . 5
3. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.1 Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2 Mounting of the Instrument . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 Electrical Connection Scheme . . . . . . . . . . . . . . . . . . . . . 18
3.4 Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5 Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.2 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6 Signal Outputs 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.7 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.8 Sensor Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.9.1 RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.9.2 Profibus (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
A-96.250.541 / 080117 1
AMU Oxytrace
Table of Contents
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1 Establish sample flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1 Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2 Stop Operation before Maintenance . . . . . . . . . . . . . . . . 24
5.3 Maintenance of Oxygen Sensor. . . . . . . . . . . . . . . . . . . . 24
5.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.4.1 Calibration Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.4.2 Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6. Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.2 Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
A. Programm Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
B. RS232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
B.1 Connection of the Communication Cable. . . . . . . . . . . . . 35
B.2 Configuration of the AMU Transmitter . . . . . . . . . . . . . . . 36
B.3 HyperTerminal Configuration on the PC . . . . . . . . . . . . . 37
B.4 Functions of the AMU RS232 Interface . . . . . . . . . . . . . . 41
B.4.1 Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
B.4.2 Log download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
B.4.3 Clear Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
B.5 Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
C. Profibus (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
C.2 Cyclic Data Telegram. . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
B.3 Customizing the Cyclic Data Telegram . . . . . . . . . . . . . . 49
B.4 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
B.5 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
B.5.1 Device Master Files (GSD). . . . . . . . . . . . . . . . . . . . . . 54
B.5.2 GSD File Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
B.6 Acyclic Data Communication . . . . . . . . . . . . . . . . . . . . . . 55
2 A-96.250.541 / 080117
AMU Oxytrace
Table of Contents
C. Modbus (option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
C.2. Data and Control Functions . . . . . . . . . . . . . . . . . . . . . . . 58
C.3 Function Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
G. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
H. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
A-96.250.541 / 080117 3
AMU Oxytrace
Safety Instructions
1. Safety Instructions
This chapter contains general information assuring safe operation
of the instrument. More safety instructions are given throughout this
manual, at the respective locations where observation is most
important.
Strictly follow all safety instructions in this publication.
1.1 General
The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work
environment.
Target audience
4 A-96.250.541 / 080117
AMU Oxytrace
Safety Instructions
WARNING Warning
Symbols
Generally, the triangular warning symbol indicates the possibility
of personal injury or even loss of life if instructions are not followed,
e.g.
caustic
Attention Attention
Symbols The attention symbol indicates the possibility of equipment
damage, malfunctions or incorrect process results if instructions
are not followed.
Warning
For safe instrument installation and operation you must read and
understand the instructions in this manual.
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.
A-96.250.541 / 080117 5
AMU Oxytrace
Safety Instructions
1.3.4 Modifications
Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept
responsibility for any claim resulting from unauthorized modification
or alteration.
6 A-96.250.541 / 080117
AMU Oxytrace
Description of the system
A-96.250.541 / 080117 7
AMU Oxytrace
Description of the system
RUN 10:56:03
AMU Oxytrace
8 A-96.250.541 / 080117
AMU Oxytrace
Description of the system
Safety No data loss after power failure, all data is saved in a non-volatile
features memory. Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
Alarm Relay One potential free contact for summary alarm indication for
programmable alarm values and instrument faults.
Maximum load: 100mA / 50V
A-96.250.541 / 080117 9
AMU Oxytrace
Description of the system
Sensor head
Sensor cap
10 A-96.250.541 / 080117
AMU Oxytrace
Description of the system
Sample
outlet
Flow
meter
Oxygen
sensor
Sample
inlet
A-96.250.541 / 080117 11
AMU Oxytrace
Description of the system
Sample
outlet
Sample
inlet
12 A-96.250.541 / 080117
AMU Oxytrace
Description of the system
2.4 Software
2.4.1 Display
RUN 10:56:03
11.3 ppb
AMU Oxytrace
error
blinking: fatal error
Status Time
R1 Process dissolved oxygen
xx.x ppb
R2
Values
sample sample
flow l/h °C temperature
Relay Status
upper / lower limit not yet reached
upper / lower limit reached
control upw. / dnw.: no action
control upw. / dnw.: active, dark bar indicates control intensity
motor valve: valve closed
motor valve: open, dark bar indicates approx. position
timer
timer: timing active (hand rotating)
A-96.250.541 / 080117 13
AMU Oxytrace
Description of the system
Program
Navigation RUN 14:10:45 Main Menu 1
R1 Enter Messages
R2 x ppb Diagnostics
Maintenance
Exit Operation
25.0°C Installation
14 A-96.250.541 / 080117
AMU Oxytrace
Description of the system
Main Menu 1
Messages 1.1 Menu Messages 1
Messages Pending Errors Reveals pending errors as
Diagnostics Message List well as an event history
Maintenance (time and state of events
Operation
that have occured at an
Installation
earlier point in time).
Menu Maintenance 3
Maintenance 3.1
For relay and signal output
Simulation simulation, to set the
Set Time 02.04.05 16:49:30 instrument time, and to
Transmitter check off
issue the transmitter
Fine adjust
check. It is used by the
service personnel.
Menu Operation 4
Operation 4.1
Sensors Subset of menu
Relay Contacts Installation 5, but process-
Logger related. User relevant
parameters that might
need to be modified during
daily routine. Normally
password protected and
used by the process-
operator.
Menu Installation 5
Installation 5.1
Sensors For initial instrument set up
Signal Outputs by SWAN authorized
Relay Contacts person, to set all
Miscellaneous instrument parameters.
Interface Normally password
protected.
A-96.250.541 / 080117 15
AMU Oxytrace
Description of the system
Press Exit .
Signal output 1 5.2.1
Yes highlighted, press Enter to
ParameterSave ? Hydrazine
save the new parameter. The
Current loop 0-20 mA
Yes system reboots, the new
Function No lin
Scaling parameter is set.
Changing
parameter Alarm Dis 5.3.1.1 Enter
Select parameter and press .
values Alarm High 1000 ppb
Alarm Low 0.00 ppb
Hysteresis 1.0 ppb to set required value.
Delay 5 Sec
Alarm Dis 5.3.1.1 Enter to confirm new value.
Alarm High 30.0 ppb
Alarm Low 0.00 ppb Exit to exit the menu.
Hysteresis 1.0 ppb
Delay 5 Sec
16 A-96.250.541 / 080117
AMU Oxytrace
Installation
3. Installation
A-96.250.541 / 080117 17
AMU Oxytrace
Installation
89.8
96
89.8
96
112
18 A-96.250.541 / 080117
AMU Oxytrace
Installation
Use only the terminals shown in this diagram, and only for
the mentioned purpose. Use of any other terminals will
cause short circuits with possible damage to material and
personnel.
A-96.250.541 / 080117 19
AMU Oxytrace
Installation
Rear view
SENSOR
LINE
RS232
OPTION
PROFIBUS or MODBUS
SHIELD
Neutral
Phase
20 A-96.250.541 / 080117
AMU Oxytrace
Installation
Terminal 5/6
3.7 Input
Use only potential-free (dry contacts).
Terminal: 11/10
A-96.250.541 / 080117 21
AMU Oxytrace
Installation
19 20 21 22
Swan Connect the Swan flow meter of the QV-Flow cell as given below.
flow meter
Black White Brown
7 8 9
3.9 Interfaces
3.9.1 RS232
22 A-96.250.541 / 080117
AMU Oxytrace
Installation
Please buy a PROFIBUS cable with free endsfrom your local suppli-
er.
NOTE: The color of the wires might be different. Before connecting
them to the transmitter, please check the corresponding data sheet.
In-and outcoming leads have to be bridged, as shown in the picture
above.
This bridging allows the replacement of an AMU transmitter without
interruption of the Profibus.
A-96.250.541 / 080117 23
AMU Oxytrace
Instrument Setup
4. Instrument Setup
5. Maintenance
Maintenance frequency depends strongly on the water quality.
24 A-96.250.541 / 080117
AMU Oxytrace
Maintenance
A-96.250.541 / 080117 25
AMU Oxytrace
Maintenance
Thread
Anode
Sensor cap
Cathode
Guard electrode
• Take the new cap of the replacement kit (A-87.290.050) and fill
it completely with fresh electrolyte.
• There is a groove in the electrode thread, where excessive air
and electrolyte can escape while mounting the sensor cap. Hold
the sensor in upright position, sensor tip pointing downwards,
with the groove on the upper side.
• Slowly fix the electrode head and the sensor cap to allow the
excess electrolyte to escape without building up too much
pressure inside the electrode. Fasten sensor cap finger-tight.
• Clean the electrode thoroughly and dry the sensing membrane
with a soft tissue.
• Switch on power. Let the sensor run-in at air for at least 30 min,
better 1 h. Afterwards perform an air calibration and install the
sensor into the flow cell again.
26 A-96.250.541 / 080117
AMU Oxytrace
Maintenance
5.4 Calibration
5.4.1 Calibration Procedure
Go into Maintenance, select Calibration.
If the instrument is in operation, turn off the sample flow by turning
the needle valve.
To take the electrode out of the flow cell, turn the fastening of the
sensor counter clockwise. Unscrew it completely. Open the sample
valve slightly for a moment. This will loosen the tight seal of the
sealing ring, and the electrode may be taken out of the flow cell
easily. Avoid pulling the electrode with force.
Dry the electrode's body and sensing membrane with a soft paper
tissue. This is very important!
Place the electrode slightly tilted into the flow cell, so that the
sensor cap of the electrode rests on the rim for the sealing ring.
A-96.250.541 / 080117 27
AMU Oxytrace
Maintenance
In the wet flow cell, the atmosphere will be saturated with water
vapour. This atmosphere will produce the most accurate calibration
results. The sensing part of the electrode must not be in direct
contact with water!
Choose menu 3 (maintenance), then calibration (point 3.1). Press
Enter.
The calibration is now done automatically. As soon as the reading
is stable, the microprocessor will store the calibration data in the
memory.
The required time for a calibration depends mainly on the
difference between temperature and oxygen content in the sample
and in the air. It can take 15 - 20 minutes. This is also the case, if
the electrolyte was exchanged. The end of the calibration is
indicated on the display.
Push Enter to confirm the calibration.
Insert the electrode back into the flow cell, push down completely.
Tighten the sensor fastening.
Turn the sample flow on and off until all air bubbles are flushed out.
28 A-96.250.541 / 080117
AMU Oxytrace
Maintenance
5.4.2 Zero-Verification
SWANSENSOR OXYTRACE G for the measurement of low
oxygen content (< 1 ppb).
Prepare a 5%-sodium sulfite solution with demineralized water.
If necessary, install the electrode according to the description in
the manual (Maintenance of Oxygen Sensor, p. 25), i. e. place a
measuring membrane into the head screw cap of the electrode
and fill it with electrolyte.
Calibrate according to the manual
(Calibration Procedure, p. 27).
Put the electrode into the sodium sulfite solution afterwards.
Assure that there are no air bubbles at the electrode.
The measured value should now be < 1 ppb.
Depending on the state of the electrode this process can take
several hours. In case of an electrode-refill, this can take days
until the measured value is lower than 1 ppb.
A-96.250.541 / 080117 29
AMU Oxytrace
Error list
6. Error list
30 A-96.250.541 / 080117
AMU Oxytrace
Error list
A-96.250.541 / 080117 31
AMU Oxytrace
Error list
32 A-96.250.541 / 080117
AMU Oxytrace
Programm Overview
A. Programm Overview
Instrument 2.1.3.1*
Identification 2.1* Factory Test 2.1.3* Motherboard
Front End
Years 2.1.4.1*
Operating Time 2.1.4* Days
Hours
:
Sample ID 2.3.1*
Temperature (°C)
Sample 2.3* Nt5k (Ohm)
Sample Flow (l/h)
Sample flow raw value (Hz)
Protocol 2.5.1*
Interface 2.5* (only with RS485
Baud rate 2.5.2* or RS232 interface)
* Menu numbers
A-96.250.541 / 080117 33
AMU Oxytrace
Programm Overview
Calibration 3.1*
Active 4.2.4.1*
Signal Outputs 4.2.4.2*
Input 4.2.4* Output/Control 4.2.4.3*
Fault 4.2.4.4*
Delay 4.2.4.5*
* Menu numbers
34 A-96.250.541 / 080117
AMU Oxytrace
Programm Overview
O2 Offset 5.1.2*
Parameter 5.2.X.1*
Signal Outputs Signal Output 1/2 Current Loop 5.2.X.2*
5.2* 5.2.1*/5.2.2* Function 5.2.X.3*
Scaling / Control Parameters 5.2.X.4X*
Active 5.3.4.1*
Signal Outputs 5.3.4.2*
Input 5.3.4* Output/Control 5.3.4.3*
Fault 5.3.4.4*
Delay 5.3.4.5*
Language 5.4.1*
Miscellaneous 5.4* Set defaults 5.4.2*
Load Firmware 5.4.3*
Password 5.4.4*
Sample ID 5.4.5*
Protocol 5.5.1*
Interface 5.5* (only with RS485
Baud rate 5.5.2* or RS232 interface)
* Menu numbers
A-96.250.541 / 080117 35
AMU Oxytrace
RS232 Interface
B. RS232 Interface
36 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
Interface 5.5.1
Protocol HyperTerminal Protocol
Baud Rate 115200 Baud Profibus
Modbus RTU
SWAN Bus
HyperTerminal
Webserver
2 Enter menu 5.5.2 and choose 115200 Baud for the Baud rate.
Interface 5.5.2
Protocol HyperTerminal
Baud Rate 115200 Baud Baud Rate
2400 Baud
4800 Baud
9600 Baud
19200 Baud
38400 Baud
57600 Baud
115200 Baud
A-96.250.541 / 080117 37
AMU Oxytrace
RS232 Interface
38 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
* Note:
If your computer has no COM port, you have to use an USB-COM-
Converter to connect the cable to the PC.
After connection find out via the Device Manager which port you
are using (Settings/Control Panel/System/Hardware/Device
Manager).
Clicking on Ports
(COM & LPT) shows
you which port is
used.
This address of the port should be used for the configuration of the
HyperTerminal.
A-96.250.541 / 080117 39
AMU Oxytrace
RS232 Interface
40 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
A-96.250.541 / 080117 41
AMU Oxytrace
RS232 Interface
Time 25 min 2h 25 h 5d 10 d 31 d 62 d
42 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
On the keyboard,
press Esc.
Browse to the
folder you have
created to archive
printouts.
A-96.250.541 / 080117 43
AMU Oxytrace
RS232 Interface
Changing the To change the folder where the files are saved,
folder for 1 press the “Receive”-button.
download
data 2 choose the directory and the folder
3 and the receiving protocol must be Zmodem.
44 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
A-96.250.541 / 080117 45
AMU Oxytrace
RS232 Interface
AMU Rescon
46 A-96.250.541 / 080117
AMU Oxytrace
RS232 Interface
A-96.250.541 / 080117 47
AMU Oxytrace
Profibus (option)
C. Profibus (option)
C.1 Introduction
The AMU family of transmitters provides a common base for a
number of analytical instruments with application specific sensor
interfaces. Each instrument has a basic structure with identical
functionality.
48 A-96.250.541 / 080117
AMU Oxytrace
Profibus (option)
A-96.250.541 / 080117 49
AMU Oxytrace
Profibus (option)
Byte 1
Bit Description
0 Hold
1 Control Off
2..7 Not used
50 A-96.250.541 / 080117
AMU Oxytrace
Profibus (option)
A-96.250.541 / 080117 51
AMU Oxytrace
Profibus (option)
Figure 6: DP Diagnostics
Byte Description
0 Status 1 Standard DP Diagnostics
1 Status 2
2 Status 3
3 Master address
4 Ident number (high byte)
5 Ident number (low byte)
6 Header (block length incl. Header byte) = 14 Status coding according
7 Status Type (0x81) to DP/V1
8 Slot Number (0x00)
9 Status Specifier
10 ..13 Diagnosis
14 ..19 Diagnosis Extension
52 A-96.250.541 / 080117
AMU Oxytrace
Profibus (option)
A-96.250.541 / 080117 53
AMU Oxytrace
Profibus (option)
54 A-96.250.541 / 080117
AMU Oxytrace
Profibus (option)
B.5 Configuration
A-96.250.541 / 080117 55
AMU Oxytrace
Profibus (option)
56 A-96.250.541 / 080117
AMU Oxytrace
Modbus (option)
C. Modbus (option)
C.1 Introduction
The AMU Oxytrace is an electronic transmitter / controller for the
measurement of dissolved oxygen in high purity water. It can be
operated as an intelligent slave device on the Modbus/JBUS using
twisted pair RS485 lines.
Modbus/Jbus The Modbus / Jbus Protocol communicates using a master-slave
technique, in which only one device (the master) can initiate
transactions (queries). The other devices (the slaves) respond by
supplying the requested data to the master, or by taking the action
requested in the query. Typical master devices include host
processors, PLCs etc. Typical slaves can be PLCs or intelligent
transmitter/controller like the AMU family of the instruments.
The Query The function code in the query tells the addressed slave device
what kind of action to perform. The data bytes contain any
additional information the slave will need to function. For example,
function code 03 will query the slave to read holding registers and
respond with their contents. The data field must contain the
information telling the slave which register to start at and how many
registers to read.
The Re- If the slave makes a normal response, the function code in the
sponse response is an echo of the function code in the query.
The data bytes contain the data collected by the slave.
Serial Modbus supports two transmission modes: ASCII or RTU.
Transmission The AMU Instruments support Modbus RTU (Remote Terminal
modes Unit) mode only.
RTU Mode In RTU mode, each 8-bit byte in a message contains two 4-bit
hexadecimal characters. The main advantage of this mode is that
its greater character density allows better data throughput than
ASCII for the same baud rate. Each message must be transmitted
in a continuous stream. Users select the desired serial port
communication parameters (baud rate, parity mode, etc). The serial
parameters must be the same for all devices on a Modbus network.
A-96.250.541 / 080117 57
AMU Oxytrace
Modbus (option)
Coding System
Eight bit binary
1 start bit
8 data bits, least significant bit sent first
1 bit for even / odd parity, no bit for no parity
1 stop bit if parity is used, 2 bits if no parity
Cyclic Redundancy Check (CRC)
RTU framing: messages start with a silent interval of at least 3 ½
character times. A similar interval of at least 3 ½ character times
marks the end of the message.
Slave
Start Function Data CRC END
Address
T1-T2-T3-T4 8 Bits 8 Bits n x 8 Bits 16 Bits T1-T2-T3-T4
58 A-96.250.541 / 080117
AMU Oxytrace
Modbus (option)
A-96.250.541 / 080117 59
AMU Oxytrace
Modbus (option)
Discrete Inputs
Alarm &Status bits are packed as one input per bit of the data field.
60 A-96.250.541 / 080117
AMU Oxytrace
Modbus (option)
A-96.250.541 / 080117 61
AMU Oxytrace
Modbus (option)
2 Hold no yes
62 A-96.250.541 / 080117
AMU Oxytrace
Modbus (option)
4000.. 4126 2001.. 2064 Event_Rec Record EVENT_REC Get Event Record r
64 Records max.
5000.. 8070 2501.. 4036 Log_Rec Record LOG_REC Get Logger Record r
1536 Records max.
A-96.250.541 / 080117 63
AMU Oxytrace
Modbus (option)
64 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
2.1.1 View the Designation of instrument: Always AMU Oxytrace
2.1.2 Version: firmware of instrument, e.g. V4.11-11/07
2.1.3 Factory test of the instrument; date of mother board and front-end
QC test
2.1.5 Operating time: years, days, hours, minutes, seconds
A-96.250.541 / 080117 65
AMU Oxytrace
Program List and Explanations
2.2 Sensors
2.2.1 Oxytrace G
Current value: in ppb
Raw value tc: raw value in nA, temperature compensated.
Raw value: raw value in nA, uncompensated.
Saturation: Current saturation in %
2.2.1.5 Cal history:
Review the diagnostic values of the last calibration of the oxygen
sensor. Only for diagnostic purpose.Max. 65 data records are
memorized.
2.2.2 Miscellaneous
Case Temp.: Actual temperature in °C inside the transmitter.
Air pressure: Displaying of the current air pressure in hPa
2.3 Sample
2.3.1 Sample ID
Sample ID: Review the programmed code. The code is defined by
the user to identify the sample point in the plant.
Temperature
Current sample temperature in °C and additionally the resistance of
the Nt5k temperature probe in Ohm
2.5 Interface
Review programmed communication settings.
66 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
3 Maintenance
3.1 Calibration
3.1.5 Start a calibration and follow the instructions on the screen.
Displayed values are saturation in % and the saturation current in
nA. The indication bar shows the progress.
3.2 Simulation
3.2.1-3.2.6 Select alarm relay, relay 1 /2 or signal output 1/2 to be tested with
the <Enter> key and press the up/down key to change the value.
After confirming the setting with the Enter key, the value is
simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min.
If you quit the menu, all simulated values will be reseted.
4 Operation
4.1 Sensors
4.1.1 Filter time constant:
Used to damp noisy signals. The higher the filter time constant, the
slower the system reacts to changes of the measured value.
Range: 5-300 sec
4.1.2 Hold after cal:
To allow the instrument to stabilize again after calibration. During
cal. and hold time the signal outputs are frozen, alarms and limits
are not active.
Range: 0 - 6000 sec
A-96.250.541 / 080117 67
AMU Oxytrace
Program List and Explanations
68 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
4.3 Logger
The instrument is equipped with an internal logger. The data can be
downloaded to a PC by Hyperterminal interface (see Log
download, p. 43 for more details)
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, measured value
uncompensated, temperature, flow.
4.3.1 Log Interval: Select a convenient log interval. Consult the table
below to estimate the max logging time. When the loggin buffer is
full, the oldest data record is erased to make room for the newest
one. (circular buffer)
Time 25 min 2h 25 h 5d 10 d 31 d 62 d
4.3.2 Clear logger: If confirmed with yes, all data is erased and a new
data series is started.
A-96.250.541 / 080117 69
AMU Oxytrace
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1 Flow:
If a flow cell without flow measurement (B-Flow) is used, choose
none. With flow measurement select Q-Flow
5.1.2 O2 Offset:
Manual, small correction of the offset. Range 0-3 ppb
70 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
As process values:
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below
mA
20
e ar
bilin
0-20 mA: 10
4-20 mA: 12
e ar
lin
0/4
0.0 0.1 0.2 0.3 0.4 0.5 0.6 unit
meas. value
mA
20
0-20 mA:10
4-20 mA: 12
0/4
1 2 3 6 30
10 100 1000 10000 value
0 1 2 3 4 log
A-96.250.541 / 080117 71
AMU Oxytrace
Program List and Explanations
As control output:
Signal outputs can be used for driving control units.
We distinguish different kinds of controls:
P-controller:
The controller action is proportional to the deviation from the
setpoint. The controller is characterized by the P-Band. In the
steady-state, the setpoint will never be reached. The deviation is
called steady-state error.
Parameters: setpoint, P-Band
PI-controller:
The combination of a P-controller with an I-controller will minimize
the steady-state error. If the reset time is set to zero, the I-controller
is switched off.
Parameters: setpoint, P-Band, reset time
PD-controller:
The combination of a P-controller with a D-controller will minimize
the response time to a fast change of the process value. If the
derivative time is set to zero, the D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller:
The combination of a P-, an I- and a D-controller allows a proper
control of the process.
parameters: setpoint, P-Band, reset time, derivative time
response on maximum
control output
a time
The point of intersection of the tangent with the respective axis will
result in the parameters a and L .
Consult the manual of the control unit for connecting and
programming details. Choose control upwards or downwards.
72 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
Control upwards/downwards:
5.2.x.43.10 Setpoint: User-defined process value (Oxygen, Temperature,
Sample flow or Saturation).
5.2.x.43.20 P-Band: Range below (upwards control) or above (downwards
control) the set-point, within the dosing intensity is reduced from
100% to 0% to reach the set-point without overshooting.
5.2.X.43.3 Reset time: The reset time is the time till the step response of a
single I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0-9000 sec.
5.2.X.43.4 Derivative time: The derivative time is the time till the ramp
response of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0 - 9000 sec.
5.2.X.43.5 Control timeout: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped to safety reasons. Range: 0 - 720 min.
A-96.250.541 / 080117 73
AMU Oxytrace
Program List and Explanations
74 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
Limit upper/lower:
A-96.250.541 / 080117 75
AMU Oxytrace
Program List and Explanations
5.3.X.32 Settings
Choose the respective actuator:
Time proportional
Examples of metering devices that are driven time proportional are
solenoid valves, peristaltic pumps.
Dosing is controlled by the operating time
Parameters:
Cycle time: duration of one control cycle (on/off change).
Range: 10-600 sec.
Response time: Minimal time the metering device needs to react
Range: 0-240 sec.
Frequency
Examples of metering devices that are puls frequency driven are
the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots
Parameters:
Pulse frequency: Max. pulses per minute the device is able to
respond to. Range: 20-300/min.
Motor valve
Dosing is controlled by the position of a motor driven mixing valve.
Parameters:
Run time: Time needed to open a completly closed valve
Range: 5-300 sec.
Neutral zone: Minimal response time in % of the of the runtime. If
the requested dosing output is smaller than the response time, no
change will take place.
Range: 0-240 sec.
76 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
Timer
The relay will be closed repetitively depending on the programmed
time scheme.
This function is intended mainly to run Swan's cleaning modules or
to program an automatic filter flush.
5.3.X.24 Operation mode: Select the operation mode (interval, daily, weekly)
of the signal outputs.
5.3.X.340 Interval: Set interval time. range 1-1440 min.
5.3.X.44 Run Time: Time the relay stays closed. Range: 5-6000 sec.
5.3.X.54 Delay: during run time plus the delay time the signal and
control outputs are held as programmed below .
5.3.X.6 Signal Outputs:
Select the behavior of the signal outputs when the relay closes
Continuous: Signal output continues normally.
Hold: Signal output continues on the last valid value.
Off: Signal output is switched off and goes to 0/4 mA
4.2.2.6 Output / Control:
Select the behavior of the control outputs when the relay closes
Continuous: Controller continues normally.
Hold: Controller continues on the last valid value.
Off: Controller is switched off.
A-96.250.541 / 080117 77
AMU Oxytrace
Program List and Explanations
Fieldbus
The relay will be switched via the Profibus input. No further
parameters are needed.
See section Profibus (option), p. 48 for more details
5.3.4 Input
78 A-96.250.541 / 080117
AMU Oxytrace
Program List and Explanations
5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Available settings: Deutsch / English / Français / Español
5.4.2 Set default values: Reset the instrument to factory default values
in three different ways:
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All other
values
are set back to default values.
Completely: Sets back all values including communication
parameters.
5.4.3 Load firmware: Firmware updates should be done by instructed
service personnel only.
5.4.4 Password: Select a password different from 0000 to prevent
unauthorized access to the menus “Messages”, “Maintenance”,
“Operation” and “Installation”.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN
representative.
5.4.5 Sample ID: Identify the process value with any meaning full text,
such as KKS number.
5.5 Interface
5.5.1 Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Profibus: Device Address: 0 - 126
ID Number: Analyzer, Manufacturer, Multivariable
Local operation: enable, disable
see Profibus (option), p. 48 for details.
Modbus RTU: Device Address: 0 - 126
Baud Rate: 2400 - 115200
Parity: none, even, odd
see Modbus (option), p. 57 for details
HyperTerminal: Baud Rate: 2400 - 115200
see RS232 Interface, p. 36
Webserver: Device Address: 0-126
see separate Webserver Manual.
A-96.250.541 / 080117 79
AMU Oxytrace
MSDS Filling electrolyte
Transport :
IMDG-code 8/II
UN No. 1814
ICAO/IATA Class 8
8 UN 1814 PAX 809 CAO 813
First Aid :
- Inhalation : --
- Eyes : Wash with lot of water, summon eye specialist
immediately.
- Contact with skin : Wash with lot of water
Information on toxicology : -
Information on ecology : Less than 1% caustic potassium solution (KOH)
80 A-96.250.541 / 080117
AMU Oxytrace
Default Values
G. Default Values
Operation:
Sensors: Filter time constant: 30s
Hold after cal: 300 s
Relay 1, same as in Installation
Relay 2, Input:
Log. Interval: 30 min
Clear Logger no
Installation:
Sensors: Flow: None
O2 Offset: 0.0 ppb
Signal Outputs: Parameter Output 1: Oxygen
Parameter Output 2: Temperature
Current loop: 4-20 mA
Function: linear
Scaling:
Oxygen range low: 0.0 ppb
Oxygen range high: 10.00 ppm
Temperature range low: 0.0 °C
Temperature range high: 50.0 °C
Flow range low: 0 .0 l/h
Flow range high: 50.0 l/h
Saturation range low: 0.0 %
Saturation range high: 100 %
Alarm Relay: Alarm Oxygen high: 10 ppm
Alarm Oxygen low: 0.0 ppb
Hysteresis: 100 ppb
Delay: 5 s
Flow alarm: yes
Flow alarm high: 20.0 l/h
Flow alarm low: 5 l/h
Sample temp. alarm high: 100 °C
Sample temp. alarm low: 0 °C
Saturation alarm high: 120 %
Saturation alarm low: 0.0 %
Hysteresis: 2 %
Delay: 5 s
Case temp. high: 65 °C
Case temp. low: 0 °C
A-96.250.541 / 080117 81
AMU Oxytrace
Default Values
82 A-96.250.541 / 080117
AMU Oxytrace
Index
H. Index
A
Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 21, 69
Attention Symbols . . . . . . . . . . . . . . . . . . . . . . . . . 5
B
B-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . 12
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Communication Cable . . . . . . . . . . . . . . . . . . . . . . 36
Control Function . . . . . . . . . . . . . . . . . . . . . . . . . . 9
D
Data Communication, acyclic . . . . . . . . . . . . . . . . 57
Data Communication, cyclic . . . . . . . . . . . . . . . . . 50
Data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Description of the system . . . . . . . . . . . . . . . . . . . 7
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
E
Electrical Connection Scheme. . . . . . . . . . . . . . . . 19
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
F
Filter time constant . . . . . . . . . . . . . . . . . . . . . . . . 68
Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . 46
G
GSD File Types . . . . . . . . . . . . . . . . . . . . . . . . . . 56
H
HyperTerminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Hysteresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
I
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 21, 69
Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 22
L
Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . 42, 43
Logger, clear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
A-96.250.541 / 080117 83
AMU Oxytrace
Index
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Maintenance of Oxygen Sensor . . . . . . . . . . . . . . 25
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . 24
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 7
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . 8, 10
Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
MSDS Filling electrolyte. . . . . . . . . . . . . . . . . . . . . 81
O
On-line operation . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
P
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Profibus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 49
Program List and Explanations . . . . . . . . . . . . . . . 66
Programm Overview . . . . . . . . . . . . . . . . . . . . . . . 33
Q
QV-Flow PMMA OTG . . . . . . . . . . . . . . . . . . . . . . 11
R
Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 21, 69
RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22, 36
S
Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . 22
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 21, 71, 74
Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . 15
Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . 10
T
Technical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Temperature Input . . . . . . . . . . . . . . . . . . . . . . . . . 9
Temperature monitoring . . . . . . . . . . . . . . . . . . . . 9
W
Warning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . 5
Z
Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . 29
84 A-96.250.541 / 080117
AMU Oxytrace
A-96.250.541 / 080117 85
AMU Oxytrace
SWAN
is represented worldwide by subsidiary companies
and distributors.
SWAN Products
Analytical Instruments for:
Potable Water
Cooling Water
Made in Switzerland.
86 A-96.250.541 / 080117