OPTISONIC
Agenda
Measurement principle
Basic terms
Wave length
v [m/s]
Liquids
Methanol 1076 m/s
Naphta 1225 m/s
Kerosene 1324 m/s
Water 1497 m/s
Glycol 1658 m/s
Solids
PVC 2400 m/s
Titanium 3125 m/s
Carbon steel 3230 m/s
Ultrasonic flowmeter
Distance Transducer A
1 Transit time (t) =
Velocity
TB→A - TA→B ~ vm
vm
Transducer B
Distance Transducer A
1 Transit time (t) =
Velocity
TB→A - TA→B ~ vm
L v
Transit time of the signal from A to B D
α
L
2 TA →B =
CAB + v * cos a
Chemical industry
Conductive and non-conductive liquids
Addition of chemical products (urea, caprolactam, MDI or
liquid sulphur)
Auxiliary processes (demin water, nitrogen, air, steam)
Oil and gas Raw
products Inter-
Production mediates End
products
Loading and offloading
Pipelines and allocation metering
Petrochemicals / refineries
Process applications
Blending and allocation of crude and refined products
Auxiliary processes (demin water, nitrogen, offgas, steam)
Energy
Solar power plants
Thermal power plants
District heating
Water / wastewater
Wastewater treatment plants
Raw water intake
Water treatment and water
purification
Demineralised water
OPTISONIC
Agenda
Product portfolio
Ultrasonic process flowmeters
OPTISONIC 3400 OPTISONIC 7300 OPTISONIC 7300 Biogas OPTISONIC 6300, stationary
OPTISONIC
Agenda
OPTISONIC 3400
Product information
OPTISONIC 3400 C
DN350…2000 (14…80”)
OPTISONIC 3400 C
DN25...300 (1...12“)
OPTISONIC 3400 C
DN350…2000 (14…80”)
OPTISONIC 3400 C
DN350…3000 (14…120”)
Maintenance required
Output signal still valid
Up to 250°C (482°F)
Carbon steel flow sensor
Fully welded sensor housing
Field version only
Sensor: DN25…300 (1…12“)
Sensor: DN350…3000 (14…120“)
qi qp qs
Measurement accuracy
Class 1: ±1% (1 ± 0.01qp:q) with limit of 3.5%
Class 2: ±2% (2 ± 0.02qp:q) with limit of 5%
Class 3: ±3% (3 ± 0.03qp:q) with limit of 5%
OPTISONIC
Agenda
OPTISONIC 3400
Target application
Petrochemical / refining
Customer: Multiple refineries
Application: Hydrocarbons; high and low viscosities. Molten sulphur
Solution: Hydrocarbons: Full bore, unobstructed flow sensor.
No pressure loss. Compliant with hazardous area approvals.
No problem with local heat tracing. Ability to measure from
low to high viscosities. High turndown ratio. Molten sulphur:
Need for heating jacket to keep flow sensor at stable
temperature to avoid crystalisation of molten sulphur.
2 Ultrasonic flowmeter
6 Return line
OPTISONIC
Agenda
OPTISONIC 3400
Technical data
OPTISONIC 3400
Sensor types OPTISONIC 3000, versions: Standard, eXtended Temperature, Cryogenic, High Viscosity
Pressure range EN1092-1: DN25…80: PN40, DN100…150: PN16, DN200…1000: PN10, DN1200…3000: PN 6
ASME B16.5: 1…24 inch: 150…900 lb
Hazardous area approvals ATEX, IEC-Ex , cCSAus, NEPSI, EAC, DNV, PESO
I/Os Basic IO (Current Out, Pulse Out, Status Out, Control In); Ex-i IO, Modular IO
Diagnostics on I/Os and comm. protocols VOS, status and error messages acc. NAMUR NE107
OPTISONIC 3400
Medium properties for heating Water only: Compact version:, 0…90°C, Remote: 0…180°C
Converter type UFC 400 Basic I/O, Modular I/O with active pulse output *
I/Os and HART® communication Current Out, Pulse Out, Status Out, Control In & HART ® vs 7
OPTISONIC
Agenda
OPTISONIC 4400
Product information
OPTISONIC
Agenda
OPTISONIC 4400
Target applications
Piping for molten salts in a CSP plant (storage tank right on the back)
Well injection
OPTISONIC
Agenda
OPTISONIC 4400
Technical data
High temperature 1...40 inch (DN25...1000)* High pressure 1...8 inch (DN25...200)
Number of paths: 1“ to 2.5“; 1 path, 3“ and larger; 2 paths* Number of paths: 1“ to 2.5“; 1 path, 3“ and larger; 2 paths
Carbon steel ASTM A106 Gr B/A105N, other materials on request Stainless steel AISI 316(L), other materials on request
Max. process temperature: 600°C; 1112°F Max. process temperature: 180°C; 356°F
ASME (up to 2500 lbs) or butt-weld connection. EN connection on request ASME (up to 2500 lbs) or butt-weld connection. EN connection on request
OPTISONIC 4400
Hazardous area approvals ATEX, IEC-Ex , cQPSus, NEPSI, EAC, DNV, PESO
I/Os Basic IO (Current Out, Pulse Out, Status Out, Control In); Ex-i IO, Modular IO
Diagnostics on I/Os and comm. protocols VOS, status and error messages acc. NAMUR NE107
OPTISONIC
Agenda
OPTISONIC 6300 / 6300 P
Product information
Installation
Adjustment and guaranteed alignment of transducer in a
rail
Mounting of sensor rail with fixation units and metal straps
on pipe
Start-up within 15 minutes; no specialist, no training and
no special tools required
Converter’s installation wizard guides user step-by-step
− Program pipe and liquid parameters in converter
− Advised distance between transducers
Measure flow anywhere you need it, while processes
continue
Robustness
All-metal rail
Stainless steel fixation units and strap
Stationary OPTISONIC 6000 sensor includes:
− Cover to protect against dust, dirt, sunlight
− Double shielded industrial cabling
− Optional: stainless steel – eXtended temperature types
Portable OPTISONIC 6000 sensor:
− Transducer pair exchangeability
− High-quality coax connectors
Portable clamp-on
Trunk on wheels and pull bar for easy transport
Fits 4 rails and 2 transducer pairs
Backpack with space for converter and 2 sensor rails
Power adapter incl. plugs (EU, UK, US, AUS)
USB stick
Velcro strap for converter mounting on horizontal or vertical pole
Documentation (manual, short/quick start guide, calibration report)
Optional I/O connection box for pulse/current output and for energy
measurement
Diameter range and measuring mode DN15...100 | 0.5...4" DN50...400 | 2...16" DN200...4000 | 8…160"
< DN25 | 1": W-mode (4 traverses) V-mode (2 traverses) V-mode (2 traverses) up to DN2000 | 80"
≥ DN25 | 1": V-mode (2 traverses) Z-mode (1 traverse)
Optional sensor types Stainless steel, eXtended temperature (XT) Stainless steel, eXtended temperature (XT) -
2 traverses 1 traverse
2 traverses
4 traverses
OPTISONIC 6000 P sensor Small Medium - one rail Medium – two rails
Diameter range and measuring mode DN15...150 | 0.5...6" DN50...250 | 2...10" DN200…750 | 8…30"
< DN25 | 1": W-mode (4 traverses) V-mode (2 traverses) V-mode (2 traverses)
≥ DN25 | 1": V-mode (2 traverses) DN200…1500 | 8…60"
Z-mode (1 traverse)
Flexible solution
Metal and non-metal pipes
Internal and external coated pipes
Old and new pipes
Low-and high-pressurised pipes
Small and large diameter pipes
Process conditions
Stable process temperature
Stable flow profile
Factory calibration
Standard wet dynamic flow calibration for every
OPTISONIC 6300 / 6300 P leaving the factory
On a dedicated calibration circuit
No additional costs
Calibration report enclosed as standard
Zero flow calibration at factory
− i.e. no zero flow calibration needed in the field
OPTISONIC
Agenda
OPTISONIC 6300
Target applications
• Wall thickness: 10 mm
Solution: OPTISONIC 6300 - large
• Cost-effective solution
Chemical - troubleshooting
Customer: Catalyst manufacturer
Application: Temporary replacement of inline flowmeter
Measurement Portable clamp-on meter speeds up troubleshooting
requirement Flow measurements can continue while inline EMF meters
DN40…50 are repaired
Data logging
OPTISONIC
Agenda
OPTISONIC 6300 / 6300 P
Technical data
Sensor type OPTISONIC 6000 standard, stainless steel (SS), OPTISONIC 6000 portable
eXtended temperature (XT),
Diameters & sensors Small DN15…100 | 0.5…4 inch: standard, XT, SS Small DN15… 150 | 0.5…6 inch
Medium DN50…400 | 2…16 inch: standard, XT, SS Medium DN50…1500 | 2…60 inch
Large DN200…4000 | 8…160 inch: standard
Measuring accuracy ±1% of measured value for DN≥50 mm / 2" ±1% of measured value for DN≥50 mm / 2"
±3% of measured value for DN<50 mm / 2" ±3% of measured value for DN<50 mm / 2"
Flow velocity 0.5…20 m/s; 1.6…66 ft/s bi-directional 0.5…20 m/s; 1.6…66 ft/s bi-directional
Medium properties Max. 2% gas, 5% solids, max. viscosity 100 cSt Max. 2% gas, 5% solids, max. viscosity 100 cSt
Hazardous area approvals ATEX, FM, CSA, NEPSI, EAC, DNV Non-Ex
Power Mains powered, voltage 100…230 VAC, 24 V AC/DC Standalone, battery powered (+ power adapter)
I/Os Basic IO (Current Out, Pulse Out, Status Out, Control In); Ex-i IO, Modular Option: current, pulse output, 2x current inputs
IO
Human machine interface Optical buttons Push buttons on full numeric keypad; full colour graphics display
Diagnostics Empty pipe, signal strength, parameter programming Empty pipe, signal quality, parameter programming
Data logger Via PactWare + DTM, xFC300 Toolbox Built-in data logger + USB port
OPTISONIC
Agenda
OPTISONIC 7300
Product information
Technology
Revolutionary patented transducer design
Acoustic signal thas goes back through the transducer towards the pipe
wall is dampened, killing the crosstalk
Use of titanium and high-power signal maximises the level of acoustic
energy in the gas
Acoustic dampers
Digital signal processing techniques guarantee signal reception and
detection under all circumstances
Titanium OPTISONIC 7300 transducer
Accuracy
4”…24” (DN100..DN600): 2 path
Dry calibration (standard): ± 2%
Air calibration: ± 1%
2”, 2.5”, 3” (DN50, 65, 80): 1 path
Dry calibration (standard) ±3%
Air calibration: ± 1.5%
Applies to an absolute flow speed: 1...30 m/s; 3.3...98 ft/s
Calibration with air under atmospheric conditions
Calibration at VSL, Delft, air calibration lab, “large facility”, using a master meter
as a reference.
Medium: air at atmospheric conditions
RVA accredited, NR K999, Calibration and Measurement Capability
(CMC): 0.15 %
Five point calibration procedure at 10, 25, 40, 70, 100% of the full scale with a
maximum flow rate of 15,000 m3/h (depending on diameter)
K-factor will be set for optimum accuracy
Flow curve correction may be set for linearization purposes Calibration at VSL, with open air outlet
indicating atmospheric calibration conditions
KROHNE calibration certificate, VSL calibration certificate on request
Converter
GFC 300, converter incl. all standard options, GDC compliant
Compact or remote converter
Aluminum or stainless steel housing
Remote version requires one cable per acoustic path
W-housing not available
Basic, Ex-i and Modular I/O
4…20 mA inputs available for P and T input
Communication options: GFC 300
Modbus
Hart 5
FF
Converter functions
Optional conversion to standard conditions (functional
converter option)
Using temperature and pressure input Actual: 100°C, 100 bara
OPTISONIC
Agenda
OPTISONIC 7300
Target applications
Process applications
Oil and gas production
− Gas from separator
− Gas re-injection
− Natural gas storage and transportation
Petrochemical industry and refineries
− Hydrocarbon gasses
− Hydrogen
− Nitrogen
Chemical industry
− Furnace residue gasses
− Process gasses
− Air separation
Other
− Compressed air 2x 10” at DOW Terneuzen, NL
− Natural gas distribution and consumption
Akor ehf. - www.akor.is - email: akor@akor.is - Sími: +5179100
OPTISONIC 7300
Target applications
OPTISONIC
Agenda
OPTISONIC 7300
Technical data
Measured value
Volume flow, corrected volume flow, mass flow, molar mass, flow speed, flow direction, speed of sound, gain,
Secondary measured values
signal-to-noise ratio, reliability of flow measurement, quality of acoustic signal
Design
Features 1 or 2 path fully welded flow sensor with O-ring fitted titanium transducers
Measurement range 0.3...30 m/s; 1...98.4 ft/s, bi-directional, turndown ratio: 1:100
Measuring accuracy
Theoretical calibration (standard) DN100...600 / 4…24": < ± 2% of actual measured flow rate, for 1...30 m/s
DN50...80 / 2…3": < ± 3% of actual measured flow rate, for 1...30 m/s
Gas calibration DN100...600 / 4…24": < ± 1 % of actual measured flow rate, for 1...30 m/s
DN50...80 / 2…3": < ± 1.5% of actual measured flow rate, for 1...30 m/s
Operating conditions
Process pressure All standard sensor designs are fully rated. Higher pressure ratings up to 2500 lbs / PN420 on request
Signal converter
Inputs / outputs Current (incl. HART®), pulse, frequency and/or status output, limit switch and/or control input (depending on the I/O version)
Counters 2 internal counters with a max. of 8 counter places (e.g. for counting volume and/or mass units)
Approvals
OPTISONIC
Agenda
OPTISONIC 7300 Biogas
Product information
Converter
GFC 300 converter
Compact or remote converter
Aluminum housing
Remote version requires one cable per acoustic path
W-housing not available
Standard Ex and Modular I/O:
− I/O selection, current output and pulse output
− Module 1, 2: Active current input
Communication: GFC 300
− HART 5 (standard)
− Modbus (optional)
Temperature sensor
Temperature sensor, PT100
4…20 mA transmitter: Range: 0…100°C
10 mm thermowell, length 150 mm, replaceable insert
Fitting on flowmeter: Compression fitting in G1/2 nozzle
Ex: ATEX Ex-d
Integrated functionality
Methane content calculation with standard temperature
sensor
Optional conversion to standard conditions using optional Actual: 100°C, 100 bara
pressure sensor
Calculation to standard conditions according to the
universal gas law for ideal gases
OPTISONIC
Agenda
OPTISONIC 7300 Biogas
Target applications
Biogas contains:
50…70 vol % of methane depending on the bio feedstock and process
The remaining part is CO2 (30…50 vol %)
Small amounts of H2S, O2 and N2
Water vapour (when not dried; 100% saturated)
(water condensates on the pipe wall as the gas cools down in the pipeline
The condensate also flows through the flowmeter on to the bottom of the
pipeline.)
A CHP unit (gas engine with generator and heat
recovery) is a common way to use biogas
Temperature: +20…50°C
Pressure: 50-60 mBar(g)
OPTISONIC
Agenda
OPTISONIC 7300 Biogas
Technical data
Measured value
Volume flow, corrected volume flow, mass flow, molar mass, flow speed, flow direction, speed of sound, gain,
Secondary measured values
signal-to-noise ratio, reliability of flow measurement, quality of acoustic signal, methane content
Design
Features 1 or 2 path fully welded flow sensor with O-ring fitted titanium transducers
Modular construction The measurement system consists of a measuring sensor and a signal converter
Materials
Sensor
Measuring accuracy
Gas calibration DN50, 80 / 2 , 3": < ± 1.5% of actual measured flow rate, for 1...30 m/s; < ± 1.5 cm/s for 0.3...1 m/s
DN100, 150, 200 / 4, 6, 8": < ± 1 % of actual measured flow rate, for 1...30 m/s; < ±1 cm/s for 0.3...1 m/s
Operating conditions
Operating conditions
EN 1092-2 / DIN 2642F DN50…200: PN10, lap joint flange, pressed plate
Ambient temperature
Signal converter
Current (incl. HART®), pulse, frequency and/or status output, limit switch and/or control input (depending on the I/O version).
Inputs / outputs
Analogue input for temperature/pressure signals
Counters 2 internal counters with a max. of 8 counter places (e.g. for counting volume and/or mass units)
Approvals
OPTISONIC
Agenda
OPTISONIC 8300
Product information
OPTISONIC 8300
Accuracy:
− DN100 / 4": < ± 1.5% of actual measured volume flow rate
− DN150...600; 6…24": < ± 1% of actual measured volume
flow rate
1…60m/s; 3…197 ft/s, bi-directional
Measures from zero, turndown ratio: 60:1
Temperature range: -40 to 600°C; -40 to 1112°F
Pressure: up to 200 bar; 2900 psi
OPTISONIC 8300
Standard
Diameter range: DN100…DN600 (4…24”)
Connections:
− Flangeless (butt welding into pipeline)
− ASME 150 or 300 lbs flanges, equivalent EN flange on request
Standard flow sensor material
− Carbon steel (ASTM A105/106)
Economic alternative for differential pressure (DP) measurement with
primary elements:
Low investment (capex)
Low operational cost (opex)
EX:
− Non-Ex
− ATEX
Design:
− Remote only, with 5 to 30 meters of cable (1 cable per acoustic path)
Calibration:
− With atmospheric air at VSL (same procedure as OPTISONIC 7300)
Finish:
− Uncoated
Options
Low alloy carbon steel materials for higher pressure and
temperature
High-pressure natural gas calibration (accuracy +/-1% on
mass flow)
High-temperature coating
OPTISONIC 8300
GFC 300
Temperature
Pressure
Energy flow
Volumetric flow
Mass flow
OPTISONIC 8300 as volume flowmeter Mass flow calculation in customer system Energy flow
Volumetric flow
Mass flow
Temperature
Pressure
OPTISONIC 8300 as volume flowmeter Mass flow calculation in flow computer Energy flow
Volumetric flow
Mass flow
Temperature
Pressure
Optional calibration:
Medium: pressurised natural gas
Flow rate and pressure are adjusted to cover
Reynolds range of operation as indicated by the
sizing tool
Recommended for custody transfer of steam
Standard procedure:
6 flow rates, 3 measurements per flow rate
Verification on one flow rate after adjustment of K-factor
One general K-factor will be established. No (Reynolds)
correction curve (polynomial correction) will be
implemented, neither in GFC 300 F nor in a flow computer
Can be implemented at Euroloop. Provides an Euroloop
calibration certificate
OPTISONIC
Agenda
OPTISONIC 8300
Target applications
Superheated steam
Low-, medium- and high-pressure steam in process
plants
Accurate measurement of steam from boiler/power plant
to user (process)
Allocation of steam in process plants
Steam flow in HRSG
(heat recovery steam generator) in power plants
Steam flow to steam turbine in power plants OPTISONIC 8000
OPTISONIC 8000
Contractor: Mitsubishi
1311 MW (3 units)
OPTISONIC
Agenda
OPTISONIC 8300
Technical data
Measured value
Secondary measured values Volume flow, enthalpy flow, mass flow, flow speed, flow direction, speed of sound, gain, signal-to-noise ratio,
reliability of flow measurement, quality of acoustic signal
Design
Remote version In field (F) mount version: OPTISONIC 8000 with GFC 300 F signal converter
Materials
Sensor
Measuring accuracy
Air calibration (standard) DN100 / 4": < ± 1.5% of actual measured volume flow rate
Repeatability
< ± 0.2%
Measuring accuracy
Repeatibility
< ± 0.2%
Operating conditions
Process temperature
Process pressure
Standard version (high pressure) 100 bar / 1450 psi (200 bar / 2900 psi)
Ambient temperature
Signal converter
Current (incl. HART®), pulse, frequency and/or status output, limit switch and/or control input (depending on the I/O version)
Inputs / outputs
Analogue input for temperature/pressure signals
Counters 2 internal counters with a max. of 8 counter places (e.g. for counting volume and/or mass units)
Approvals