BDC Medical Equipment Supplies and Services

San Juan Kalayaan, Laguna

CALIBRATION REPORT
 

Customer: City of Gen.Trias Doctors Medical Center Date: November 24, 2021
Address:  Gen.Trias,Cavite  
DESCRIPTIVE MODEL NO. BRAND SERIAL NO. CODE NO.
1. CONTROL PANEL AMX G.E. 720625  
300MA
2. HIGH TENSION TRANSFORMER   G.E 5843  
3. X-RAY TUBE ASSEMBLY   G.E. 922901  
TAPPING CALIBRATION
CP H.T.T X.T.A TIME MA KVP
  sec pulse Sec/pulse setting actual setting HTT In HTT Out
      .1 .12   50ma 5mas 60kv 100v 60.3
      .1 .12   50ma 5mas 70kv  120v 70.1
      .1 .12   50ma 5mas 80kv 140v 80.2
      .1 .12   100ma 10mas 60kv 120v 61.5
      .1.       .12   100ma 10mas 70kv 135v 70.4
      .1 .12   100ma 100mas 80kv 150v 80.2
      .1 .12   200ma 20mas 60kv 135v 61.2
      .1 .12   200ma 20mas 70kv 150v 70.2
      .1 .12   200ma 20mas 80kv 160v 80.4
      .1 .12   300ma 30mas 60kv 140v 60.2
      .1 .12   300ma 30mas 70kv 165v 70.3
      .1 .12   300ma 30mas 80kv 175v 80.4
 
Performed by:
 
Dante Navarro
BIOMEDICAL Tech.
 