LT Service Manual

Industrial Weighing Terminal

1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

Contents
1. GENERAL………………………………………….………….
1.1. Introduction……………………..…………………………….
1.2. Technical specifications………………………………………

2. INSTALLATION & INTERFACING……………


2.1. Front view of the indicator……………………………………
2.1.1. Display………………………………………………………..
2.1.2. Keypad………………………………………………………..
2.1.3. Status LEDs…………………………………………………..
2.2. Analog load cell connection..…………………….…………...
2.3. Communication interfacing…………………………………..
2.3.1. RS232 Interface……………………………………….………
2.3.2. RS485 Interface……………………………………….………

3. Calibration ………………………………………
3.1. Offset setting ……………..……………….………….
3.2. Calibration procedure……………..……………….………….

4. OPERATION………………………………………
4.2. Parameter configuration…..………………………..….………
4.2.1. Accumulation Function…………………………………….
4.2.2. Tolerance-Check Function………………………………

5. SETTINGS ……………………………………
5.2.1. Power-ON Zero Function………………………………….
5.2.2. Baud rate selection ………………………………….

6. ERRORS……………………………

1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

1. Instructions for the use of the user manual, update


service

1 Purpose and scope of the MANUAL


• The USER MANUAL contains instructions and information to enable the
service personnel to ensure proper maintenance and repair work of
BIZERBA products. Possession of the MANUAL is obligatory for all BIZERBA
aftersales service organizations.

Use of the MANUAL


• In order to facilitate quick and easy access to the required information or
portion of text, the manual is divided into sections. Each section contains
a list of contents and the index number is printed in the page margin.

Distribution of the MANUAL and update Service


• Page number and relevant amendment appear at the foot of every page.
The MANUAL is issued with every product supply and in special cases the
MANUAL may be obtained directly from the aftersales service in Bizerba.
Orders should quote the order number indicated on the cover page.
• Changes thereafter made to the MANUAL will automatically be forwarded
in soft format to Bizerba local office as a routine. In the event that
changes in documentation will be submitted, the holder is advised to
supplement or exchange such pages immediately. Each amendment will
be recorded in the list of contents in index 2.

Copyright
• All rights reserved. This USER MANUAL or any part of it may not be
reproduced, stored in a retrieval system, or transmitted, in any form, or
by any means, electronical, mechanical, photocopying, recording, or
otherwise, without the prior and explicit permission of BIZERBA. All
efforts have been made to ensure the accuracy of this MANUAL. Subject
to modifications.

Release of the USER MANUAL


• The release of the 1st edition of the MANUAL by the service management

1 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

is indicated below. Updates are released by the signature of the product


manager on each service information document.

1
Supplementary documents pertaining to the MANUAL
• Operating instructions
• Programming instructions

Symbols used and their meanings

• The following three symbols appearing in the service manual have


specific meanings which should be noted as follows:

Highlighting
This symbol separates important information from the other
text and gives it prominence.

Attention
This symbol indicates that the relevant instructions are to be
! strictly adhered to in order to prevent data loss or damage to
the unit.

Warning
This symbol means that noncompliance with the relevant
instructions may lead to injury to the user.

1.001.PLT.050506 2

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

1. GENERAL

1.1 Introduction
1 With the BIZERBA Weighing Indicator LT, you own a perfect combination of
the reliable hardware design and the sophisticated software technology. The
unit is developed using the MCS – 51 single chip micro processing unit and
high speed analog to digital converter technology. This unit is an optimum
combination of the hardware and software technology giving high speed
conversion rates for analog to digital updates and with extremely high
stability rate which is extensively required in static weighing applications.

1.2 Technical Specifications

Material Of Construction Stainless Steel Grade 304


Signal input range 0 mV ~ 15 mV
Net signal input range 3 mV ~ 15 mV
Non-linearity 0.01% F.S.
Full-scale temperature coefficient 8 PPM/ ºC
Bridge power supply for load cell DC: 5V, 350mA
Display 6 digit single-rowed 7-segment LED,0.8
inches in character height and 9-status
indicators
Display Colour Red / Blue
Division 1/2/5/10/20/50 (Selectable)
Keypad 6 keys membrane dome type.
Communication protocol RS232C communication port
(Optional RS485)
Power supply AC220V (-15%~+10%),50Hz(-2%~+2%)
Operating temperature and humidity 0~40 ºC ;?90% RH
Storage temperature -20~50 ºC
Fuse 0.5A
External dimension 290 X 175 X 90 (mm)
Self-weight Approximately 3.5 kg
Battery Back-Up Internal battery 6 V; 4.5 Ah which
provides 24 Hrs of backup

3 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

2. INSTALLATION & INTERFACING

2.1 Front view of the indicator


1

2.1.1 Display
Display consists of 6 digit 7-segment LED's (red / blue). On power ON the
indicator checks itself by an automatic display check. On completion of the
check the controller enters into weighing mode.

2.1.2 Keypad

Used to enter parameter configuration modes if


MENU pressed during weighing mode

TARE key is used to tare a specific amount of weight eg.


TARE Empty container weight so as to display net weight
which is indicated by the NET status LED on the display.
To clear the Tare value, press this key again & the
indicator will again show gross weight.
On pressing ZERO key the weight will be zeroed & the
ZERO zero status LED will be ON only if the current weight
is in valid zero range which is 4% of the full scale capacity.

1)Used to add a LSB digit on display during configuration


or calibration.
2)Used to hold the value of the weight reading
on the display.

1.001.PLT.050506 4

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

1)Used to increment the LSB from 0 to 9 and back to 0


during the parameter setting or calibration.
2)Used for Accumulation function.

1)Used to confirm a particular amended parameter in


ENTER configuration mode.
2)Used to enter into the settings of the parameter

2 currently shown on the display if pressed after MENU key.

2.1.3 Status LED Description

LED Description when ON


Mains To Indicate that the MAINS supply is present.
Hold To indicate that the displayed value is made to hold.
Accu To indicate that the reading shown is accumulated weight
Stable To indicate that the displayed reading is stable.
Net To indicate that the displayed weight is net weight i.e. some
weight is tared by pressing TARE key.
Zero Glows when the indicated value on the display is zero.
Under To indicate that the measured weight is less than acceptable
limits if Tolerance-Check Mode is enabled.
Accept To indicate that the measured weight is within the
acceptable limits if Tolerance-Check Mode is enabled.
Over To indicate that the measured weight is more than
acceptable limits if Tolerance-Check Mode is enabled.

2.1 Analog load cell connection


5- Pin Female RF connector is used for connecting load cell with the
indicator. Pin configuration is as shown below.

Loadcell interface Pin Nos Colour Code


-EXC (E-) 1 Yellow
+EXC (E+) 6 Blue
+SIG (S+) 9 Red
-SIG (S-) 8 Orange
SHIELD (SH) 5 Green

5 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

2.1 Communication interfacing


The serial ports can transmit data to a printer, computer, remote display etc.
The serial port also has RS485 which is optional port. The maximum
recommended cable length for RS-232 communications is 50 feet ( 15
meters), for RS-485 communications is 1000 feet ( 330 meters).
Connection details of communication ports are as shown below.

2.1.1 RS232 Interface 2


9- Pin Female connector is used for connecting RS232 cable to the indicator,
pin configuration of which is as shown below.

RS232
interface 9-pin RF Colour
marks Connerctor Pin Nos

GND 5 Green
TXD 3 Red
RXD 2 Black

2.1.1 RS485 Interface

9- Pin Female connector is used for connecting RS485 cable to the indicator,
pin configuration of which is as shown below.

RS485
interface 9-pin RF Colour
marks Connerctor Pin Nos
GND 5 Green
- 3 Red
+ 2 Black

1.001.PLT.050506 6

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

3. CALIBRATION
Connect load cell as per connection diagram & do adc offset setting the
calibration procedure described bellow:

3.1 ADC Offset Setting:


Switch on the indicator & follow the sequence of steps bellow.

Sequence of Operation Display

ZERO
Press & hold key

7 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

Raw counts will be displayed


on the e.g.16105

To increase the counts

displayed press key

To decrease counts display


3
press key

Adjust the no load counts so


that counts should be
between
20000 to 45000

ENTER
Press

3.2 Calibration Procedure

Sequence of Operation Display

MENU
Press & hold during Power
ON till CAL message appears on
the display

A raw count will be displayed


to verify whether count changes
against the change in the weight.
e.g. If raw counts are 36105

ENTER
Press

1.001.PLT.050506 8

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

Set the display format for


weight taking the decimal
places into consideration.
e.g. To set 2 decimal places,
key in [0.00 ]

Press ENTER

3 Key in the full scale capacity in


the format set at [dec pt ] field.

Press ENTER
ENTER

Enter least count in the format


set at [dec pt ] field

using key.

E.g. 1/2/5/10/20/50

Press ENTER

Put known weight of value


within the full scale range on
the platform and key in the
same value in the format set
at [dec pt ] field.

Sequence of Operation Display

Press ENTER A raw count for

[CAL Pt1 ] will be displayed


and STABLE LED will glow when
weight becomes stable.
e.g. Counts for CAL Pt1 are 36105

9 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

Press ENTER

Put known weight of some


other value within the full scale
range on the platform and key
in the same value in the
format set at [dec pt ] field

Press ENTER A raw count for

[CAL Pt2 ] will be displayed


and STABLE LED will glow when 3
weight becomes stable.e.g.
Counts for cal pt2 are 182883

Press ENTER

Now, the indicator will


automatically restart and show
the weight information in the
format set at [dec pt ] field.

1.001.PLT.050506 10

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

4. OPERATION

4.1 Parameter Configuration

Before we can start using LT indicator for any application, it has to be


configured i.e. there are some parameters which has to be set in order to
have efficient operation and to get desired output out of the indicator.
Following are the steps to be followed for this purpose.

Discription Display

During the weighing mode if the

4 MENU
key is pressed, display
shows following information
in sequence.

To view the accumulated weight


and number of accumulations.

To amend tolerance parameters.

To activate or deactivate
Tolerance- Check function.

To amend a particular paramete

press ENTER when that

parameter name is displayed.

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

4.1.1 Accumulation Function

Accumulation function allows us to add up the weight values shown on the


display if the ACCU key is pressed after every weight value to be included in
the accumulation is displayed. Maximum 255 additions are possible using
this indicator. To view the accumulated weight and the number of times the
accumulation is done, we follow the following procedure.

Sequence of Operation Display

MENU
Press in weighing mode.

4 Press Value shown is the


ENTER

accumulated weight.

Press ZERO to clear

the accumulator.

Otherwise to retain the


accumulated value and to make
it available for the successive
MENU
additions press Now

display will show the number


of times the accumulation
is done.

MENU
Press to return to

weighing mode.

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

4.1.2 Tolerance-Check Function

Using this indicator we can set upper and lower tolerance limits for check-
weighing application. To set these limits and to activate and de-activate this
tolerance-check function, we follow the following procedure.

Sequence of Operation Display

MENU
Press in weighing mode.

Press ENTER

4 Key in the nominal weight value


in the format set at [dec pt ] field.

Press ENTER

Key in the upper limit for the


tolerance band in the format
set at [dec pt ] field.

Press ENTER

Key in the lower limit for the


tolerance band in the format set
at [dec pt ] field.

Press ENTER to return to

weighing mode.

To activate Tolerance –Check

function, press MENU

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

Press ENTER Now display will

show the weight information in


the format set at [dec pt ]
field with status LEDs indicating
the tolerance check.

To de-activate Tolerance –Check

function, press MENU

ENTER
Press Now display will

show the weight information in

4 the format set at [dec pt ]


field with tolerance-check
LEDs turned OFF.

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

5. SETTINGS

5.1 Power-ON Zero Setting


This feature allows us to nullify the weight applied on the platform before
the indicator is switched ON. To activate or deactivate this feature, we
follow the following procedure.

Sequence of Operation Display


ENTER
Press & hold during

Power ON till P on 0 message


appears on the display

5 Press ENTER

Press ENTER to activate Power-

ON Zero function. Now display


shows the weight information in
the format set at [dec pt ] field.

To deactivate the Power-


ON Zero function, press & hold

TARE during Power ON till P on

0 message appears on the display

Press ENTER

Press

Press ENTER Now display shows

the weight information in the


format set at [dec pt ] field.

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

5.2 Baud Rate setting:


Switch OFF the indicator the switch it ON & follow the sequence of steps
bellow.

5.2.1 To enable the rs232 option:

Sequence of Operation Display

Press & hold “HOLD” key

Indicator will display message


If rs232 option is to be enabled

5 press ENTER key

If weight data at rs232 port


required in continuous mode

press ENTER key Other wise

if data required on user

request press

press ENTER Key

Press & hold key

Baud rate “9600” will be


displayed on the display. Press

key to increase the

baud rate Different baud rates


that can be selected are 9600
19200

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

ENTER
press key

5.2.1 To disable rs232 option:


Sequence of Operation Display

Press & hold “HOLD” key

Indicator will display message


If rs232 option is to be enabled

press ENTER key

5 If this option not required press

key to increase the

baud rate

press ENTER

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com


LT User Manual

6. ERRORS
6.1 Error Indication
Display Description

ERROR0 The weight to be zeroed is not within the zero limit.

ERROR1 Value entered at CAL PT 1 field is greater than the set capacity.

ERROR2 Value entered at CAL PT 2 field is greater than the set capacity.

Value entered at CAL PT 2 field is greater than the value entered


ERROR3
at CAL PT 1 field but the internal counts for CAL PT 2 weight is
less than that of CAL PT 1 weight.

ERROR4 Value entered at CAL PT 2 field is less than the value entered at

6
CAL PT 1 field but the internal counts for CAL PT 2 weight is
greater than that of CAL PT 1 weight.

11 1.001.PLT.050506

PDF created with pdfFactory trial version www.pdffactory.com