Professional Documents
Culture Documents
Operator Manual IGS 520 530 540
Operator Manual IGS 520 530 540
THIS SYSTEM, THE OPERATOR SHOULD BE THOROUGHLY ACQUAINTED WITH THE INSTRUCTIONS FOR USE AND SAFETY
FAILURE TO FOLLOW THE INSTRUCTIONS FOR USE AND SAFETY RECOMMENDATIONS PROVIDED IN THIS MANUAL CAN CAUSE
5499294-1-1EN
Revision 2
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
2
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
Table of contents
Chapter 1 - Foreword 31
1 Regulatory Requirements 36
1-1 Electromagnetic Compatibility (EMC) 36
1-1-1 Electromagnetic Compatibility of the vascular system 36
1-1-2 Detailed Electromagnetic Compatibility of the vascular system 36
1-1-3 Electromagnetic Compatibility of the vascular system with Fluoro UPS (Innova option) 37
1-1-4 Detailed IEC60601–1–2 Electromagnetic Standard Compliance conditions 37
1-1-4-1 Electromagnetic Emission 38
1-1-4-2 Electromagnetic Immunity IEC 60601-1-2 39
1-1-4-3 Recommended Separation Distances for Portable and Mobile RF Communications 41
Equipment IEC 60601-1-2
1-1-4-4 Limitations Management 41
1-1-4-5 Use Limitation 42
1-1-4-6 Installations Requirements and Environment Control 42
1-2 Additional Regulatory Requirements 43
1-3 Specific Requirements for product with Open Monitor suspension (Option) 44
1-4 Recycling 44
1-5 Control of Pollution caused by Electronic Information Products (EIP's) 45
1-5-1 Explanation of Pollution Control Label 45
1-5-2 Name and Concentration of Hazardous Substances 45
1-6 Vermont labeling regulations (US) on Mercury 46
1-7 Patient Confidentiality 47
2 General Safety 47
2-1 Equipment Application Specification 48
3
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
4
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
5
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
6
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
4-2-4 IQ Standard 86
4-2-5 Receptor Dose Limited Standard 92
4-2-6 Receptor Dose Limited Standard: Countries limited to 5 R/min in Fluoroscopy 98
4-2-7 IQ Plus 99
4-2-8 Receptor Dose Limited Plus 104
4-3 Dose to Patient in Japan 109
4-3-1 Fluoroscopy/Roadmap/Blended Roadmap/Subtracted Fluoroscopy Air Kerma Levels 110
4-4 Dose to Patient in Australia/WA and Australia/ACT 110
4-4-1 Fluoroscopy/Roadmap/Blended Roadmap/Subtracted Fluoroscopy Air Kerma Levels 110
4-5 Dose to patient in Canada/Ontario 110
4-6 Dose to Operator (Isodoses) 110
4-6-1 Gantry in Vertical position - Dose at 1 meter from ground 110
4-6-2 Gantry in Vertical position - Dose at 1.5 meter from ground 111
4-6-3 Gantry in Lateral position - Dose at 1 meter from ground 112
4-6-4 Gantry in Lateral position - Dose at 1.5 meter from ground 113
4-7 Technical Factors Range for Innova IGS 520 114
4-8 Contribution of Filtration with 20 cm Detector 114
4-8-1 Contribution of Filtration along the Beam 114
4-8-2 Total filtration 115
4-9 Disable X-ray Whenever X-ray Are Not Necessary 115
4-10 Dosimetric Calibration 115
4-11 Application Mode and Service Mode are not exclusive 115
4-12 Leakage Radiation 116
4-13 Spectral Filtration 116
4-14 Dose Management 116
5 Protection regarding Ionizing Radiations Hazards for Innova IGS 530 117
5-1 Mode of Operation 118
5-1-1 Innova Digital 118
5-1-2 Reference Display 118
5-1-3 Live Display 119
5-1-4 Table Side Status Control (TSSC) 120
5-1-5 Innova Central touch screen 120
5-2 Dose to Patient 122
7
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
8
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
9
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
10
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
11
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
12
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
13
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
14
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
15
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
16
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
17
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
18
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
19
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
20
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
21
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
22
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
23
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
1 Introduction 577
1-1 What is QAP? 577
1-2 How to launch QAP? 577
1-3 Web Pages 577
1-4 Status 578
24
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
1 Introduction 581
2 Starting ILinq 581
3 ILinq Features 582
3-1 Contact GE 582
3-2 Settings 582
3-3 Messages 582
3-4 TiP Virtual Assist 582
3-5 iLinq Help 583
3-6 About iLinq 583
3-7 Close 583
4 iLinq Notifications 583
4-1 Message Notification 584
4-2 Chat Notification 584
Chapter 19 - Workstation Configuration 585
25
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
1 Introduction 610
26
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
27
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
1 Warning 637
2 Learning objective 637
3 Teaching points 637
4 Radiation dose is a natural consequence of x-ray imaging 637
5 Displayed Dose Parameters 637
5-1 Cumulated Air Kerma 638
5-2 Cumulated Dose Area Product (DAP) 638
6 Effects of ionizing radiation on humans 639
6-1 Deterministic effects 639
6-2 Stochastic effects 640
7 Dose reduction measures for the patient 640
7-1 List 640
28
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
1 Overview 651
2 Procedure for Single Model 651
2-1 PA measurement 651
2-1-1 Set up geometrical measurements 651
2-1-2 Set up feature configuration 653
2-1-3 Run the measurements 653
2-1-4 Analyse the results 654
2-2 Lateral Measurement 655
2-2-1 Set up geometrical measurements 655
2-2-2 Set up feature configuration 656
2-2-3 Run the measurements 657
2-2-4 Analyze the results 657
3 Procedure for Dual Model 658
3-1 PA measurement 658
29
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
30
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
Chapter 1 - Foreword
1 Using This Module
Throughout the text in this manual, certain type styles and symbols differentiate between one tool or graphic and
another. The conventions are as follows:
• Menu titles and window names appear in boldface: Application menu.
• Menu options appear in boldface: Done.
• Graphical mouse-selected or key-selected buttons (on-screen "soft keys") or field names (filled in by the user)
appear in boldface, within brackets: [View].
• Graphical touchkey buttons (Innova console on-screen buttons) appear in boldface, within braces: {TUBE
WARMUP}.
• On-screen prompts and messages appear in boldface typewriter font: Login:.
• User typed-in responses appear in regular typewriter font: root.
• Keyboard hardkeys, mouse buttons and Table Side User Interfaces (TSUI) hardkeys/buttons with text inside,
are within <> : <Enter>, <left>, <M>, <Fr>, <Low>, <Med>, <High>, <Pref>, <STOP>, <+>, <->.
• Icon with drawing/symbol inside (on-screen "soft keys") or TSUI buttons with drawing/symbol inside : the symbol
of the button appears in the text. Symbol examples: Stopwatch, Room Light, Rotation Brakes, Lat/Long Brakes.
Symbols may also be represented as the name of the symbol in boldface white on a gray background:
Stopwatch
• Functions for the TSUI joystick and panning handle are in boldface: Table Panning handle, Collimator joystick.
To continue this module, select item from menu list in panel on the left.
Foreword 31
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
The operator manual is available on the system itself (see section Innova Digital / Innova Digital Screen / Utilities
Key)
The Operator Manual is also available on the Internet at:
http://apps.gehealthcare.com/servlet/ClientServlet?REQ=Enter+Documentation+Library
• On the home page enter the part number of the Operator Manual in the search window and click on [Search] to
launch the search.
The part number is indicated in the table below. Choose the one matching your language.
OM part-no Language
5499294-1-1BG Bulgarian
5499294-1-1CS Czech
5499294-1-1DA Danish
5499294-1-1DE German
5499294-1-1EL Greek
5499294-1-1EN English
5499294-1-1ES Spanish
5499294-1-1ET Estonian
5499294-1-1FI Finnish
5499294-1-1FR French
5499294-1-1HR Croatian
5499294-1-1HU Hungarian
5499294-1-1ID Indonesian
5499294-1-1IT Italian
5499294-1-1JA Japanese
5499294-1-1KO Korean
32 Foreword
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
5499294-1-1LT Lithuanian
5499294-1-1LV Latvian
5499294-1-1NL Dutch
5499294-1-1NO Norwegian
5499294-1-1PL Polish
5499294-1-1PT-BR Portuguese (Brazil)
5499294-1-1PT-PT Portuguese (Portugal)
5499294-1-1RO Romanian
5499294-1-1RU Russian
5499294-1-1SK Slovakian
5499294-1-1SR Serbian
5499294-1-1SV Swedish
5499294-1-1TR Turkish
5499294-1-1ZH-CN Simplified Chinese
1 - Search window
This screen capture is given as an example and is not legally binding.
The result of the search will be the Operator Manual in pdf format in one language.
• Click on the underlined Filename.
• In the next window, click on [ACCEPT] to download the file.
A paper copy of Operator Manual can be ordered at no additional cost. Please, send a request to your Sales or
Service representative. They will transfer your request to CEMEURDIST@med.ge.com. In application of the EU
Commission Regulation on electronic instructions for use of medical devices, in European Union, your request
should be treated within 7 days.
4 Quebec
Foreword 33
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
34 Foreword
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
Chapter 2 - Introduction
Product Name: Medical X-ray angiography equipment.
The Innova IGS 520, Innova IGS 530, Innova IGS 540 is a fully integrated single plane angiographic x-ray imaging
system that meets a wide range of clinical needs for interventional and diagnostic imaging with excellent image
quality, extensive real-time processing, innovative dose management, ease of positioning, improved workflow and
image management for excellent clinical versatility without compromise.
Innova IGS 520 is an angiographic X-ray system equipped with a 21 cm Digital Detector.
Innova IGS 530 is an angiographic X-ray system equipped with a 31 cm Digital Detector.
Innova IGS 540 is an angiographic X-ray system equipped with a 41 cm Digital Detector.
The interventional configuration allows to perform diagnostic and interventional procedures. Additionally the OR
configuration (OR table) allows to perform surgery and x-ray guided surgical procedures (refer to OR Configuration
(Option) and Application Environment.)
Product Software Version: IGS5_2.0
Introduction 35
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
This symbol stands for the prescription legend statement “Caution: US federal law restricts this
device to sale by or on the order of a physician."
The following symbols are used in this document to highlight safety information:
1 Regulatory Requirements
1-1 Electromagnetic Compatibility (EMC)
The Innova IGS 520, Innova IGS 530 and Innova IGS 540 need special precautions regarding EMC. The Innova
IGS 520, Innova IGS 530 and Innova IGS 540 have to be installed and put into service according to the EMC
requirements joined in the Operator Manual and in the Pre-Installation Manual.
1-1-1 Electromagnetic Compatibility of the vascular system
This equipment complies with IEC60601-1-2 Edition 2.1 and Edition 3 EMC standard for medical devices.
This equipment generates, uses, and can radiate radio frequency energy. The equipment may cause radio
frequency interference to other medical and non-medical devices and radio communications nearby.
In order to provide a reasonable interference protection against radio communication services, this product
complies with the IEC60601-1-2 standard requirements at the premises of a typical user, and the CISPR11
standard emission limits (CISPR11 Group1 Class A).
This product is not likely to introduce significant disturbance in its vicinity, in the CISPR11 standard Group1
bandwidth (150 KHz - 1 GHz), assuming a separation distance can be set up and maintained versus sensitive
products.
The Innova system has no significant intentional radio emitter, which radiates electromagnetic energy and
electromagnetic fields (EMF) susceptible to induce known adverse health effects.
The level of these non-ionizing radiation was assessed under the required and recommended limits, addressing the
workers and general public located nearby the Innova system.
Compliance with these limits will ensure that workers and the general public exposed to the electromagnetic fields
are protected against all known adverse health effects in the 0 Hz to 300 GHz bandwidth.
1-1-2 Detailed Electromagnetic Compatibility of the vascular system
Detailed requirements and recommendations about the power supply distribution and installation are listed in the
next paragraph called "Detailed IEC60601-1-2 Electromagnetic Standard Compliance conditions", and in the PIM
Access to CPR position in less than IEC 60601-1 Clause 4.3 - From Risk CPR position is defined as:
15s (+1s in tilted conditions) Management tilt angle =0° and table top fully moved
towards feet along the longitudinal ax-
is.
Ability to position patient support for IEC 60601-1 Clause 4.3 - From Risk
imaging the zone of interest Management
Ability to position and orient the X-ray IEC 60601-1 Clause 4.3 - From Risk Includes the ability to position gantry
beam vs. patient Management isocenter vs. patient
Ability to distinguish anatomic struc- IEC 60601-2-54: Imaging perfor- Ability to distinguish images noise or
tures and interventional devices in mances artifacts from anatomic structures
images IEC 60601-2-54: Automatic Control and interventional devices.
System The automatic control system regu-
lates the acquisition parameters to
make visible anatomic structures of
interest and interventional devices.
Ability to guide interventional gesture IEC 60601-2-54: Imaging perfor-
with X-Ray imaging mances
Ability to export DICOM Radiation IEC60601-2-43: Radiation dose docu- Capability to export DICOM Radiation
Dose Structured Report mentation Dose Structured Report through the
network
System reset in less than 3 minutes IEC60601-2-43: Recovery manage-
ment
No unwanted movement of the table IEC60601-2-46 Clause 201.4.3 Applicable only to surgical configura-
in any SINGLE FAULT CONDITION tion and Tilting table
and combined fault condition with dor-
For EMC testing, the SINGLE FAULT
mant failure
CONDITION requirements of the gen-
eral standard do not apply. See §4.2
of IEC60601-1-2 Ed3
The Innova system is intended for use in the electromagnetic environment specified below.
The customer or the user of the Innova system should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment
Radio–Frequency Group1 The Innova system uses Radio Fre-
quency energy only for its internal
Emissions Class A
function. Therefore, its Radio Fre-
CISPR11 Limits quency emissions are very low and
are not likely to cause any interfer-
ence in nearby electronic equipment.
The Innova system is suitable for use
in all establishments other than do-
mestic and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions Not Applicable The Innova system is suitable for use
in all establishments other than do-
IEC 61000-3-2
mestic and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Voltage fluctuations / flicker emissions Not Applicable The Innova system is suitable for use
in all establishments other than do-
IEC 61000-3-3
mestic and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
The Innova system is intended for use in the electromagnetic environment specified below.
The customer or the user of the Innova system should assure that it is used in such an environment.
IMMUNITY Test IEC 60601–1–2 Compliance Level Electromagnetic Environment
Test Level
Electrostatic +/- 6 kV contact +/- 6 kV contact Floors are wood, concrete or
discharge (ESD) ceramic tile or floors are covered with
+/- 8 kV air +/- 8 kV air
synthetic material and the relative humidity is at
IEC 61000-4-2
least 30 %.
Electrical fast +/- 2 kV for power +/- 2 kV for power Mains power quality is that of a typical commer-
transient/burst supply lines supply lines cial or hospital environment.
IEC 61000-4-4 +/- 1 kV for input/ +/- 1 kV for input/
output lines output lines
Surge +/- 1 kV line(s) to +/- 1 kV line(s) to Mains power quality is that of a typical commer-
line(s) line(s) cial or hospital environment.
IEC 61000-4-5
+/- 2 kV line(s) to +/- 2 kV line(s) to
earth earth
Voltage dips, short 0% Un for 5 sec 0% Un for 5 sec Mains power quality is that of a typical commer-
interruptions and cial or hospital environment. If the user of the
voltage variations Innova system requires continued operation
on power supply during power mains interruptions, it is recom-
input lines mended that the Innova system be powered
from an uninterruptible power supply.
IEC 61000-4-11
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) is at levels characteristic of a typical location in
magnetic field a typical commercial or hospital environment.
IEC 61000-4-8
Note: Un is the a.c. mains voltage prior to application of the test level.
Table 3
The Innova system is intended for use in the electromagnetic environment specified below.
The customer or the user of the Innova system should assure that it is used in such an environment.
IMMUNITY Test IEC 60601-1-2 Test Compliance Level Electromagnetic Environment
Level
Conducted Radio Fre- 3 Vrms V1=3v Portable and mobile RF communications equip-
quency ment is used no closer to any part of the Innova
150 kHz to 80 MHz
IEC 61000-4-6 system, including cables, than the recommended
separation distance calculated from the equation
Radiated Radio Fre- 3 V/m E1=3v/m
applicable to the frequency of the transmitter.
quency
80 MHz to 2.5 GHz Recommended separation distance:
IEC 61000-4-3
Note: * Field strengths from fixed transmitters, such as base stations for cellular telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be performed. If the measured field strength exceeds the RF compliance level above, observe the
Innova system to verify normal operation in each use location. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Innova system.
** Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
THE INNOVA IGS SYSTEM HAD BEEN TESTED IN SITU FOR RADIATED RF IMMUNITY ONLY AT
SELECTED FREQUENCIES IN THE RANGE 80 MHz TO 2.5 GHz.
Frequency of Transmit- 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
ter
Equation
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the equation in the
corresponding column, where P is the power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note: These are guidelines. Actual conditions may vary.
1-1-4-4 Limitations Management
Adhering to the distance separation recommended in section "Recommended Separation Distances for Portable
and Mobile RF Communications Equipment IEC 60601-1-2" Table 4, between 150 KHz & 2.5 GHz, will reduce
disturbances recorded at the image level, but may not eliminate all disturbances. However, when installed and
operated as specified herein, the system will maintain its essential performance by continuing to acquire, display,
and store diagnostic quality images safely.
For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3 meters (see (*) in section
"Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC 60601-1-2"
Table 4) apart from the Innova system (in order to avoid images interferences risks).
1-1-4-5 Use Limitation
– Static field is specified less than <1 Gauss in Examination room, and in the Control Area.
– Static field is specified less than <3 Gauss in the Technical Room.
• Electrostatic discharges environment & recommendations:
– In order to reduce electrostatic discharge interference, install a charge dissipative floor material to avoid
electrostatic charge buildup.
– The relative humidity shall be at least 30 percent.
– The dissipative material shall be connected to the system ground reference, if applicable.
• EMI Filter Box for the Fluoro UPS (CE) Option: The Fluoro UPS (CE) option requires an EMI filter box placed
upstream the PDB (Power Distribution Panel). The EMI filter box will be delivered as a single product by GE
Consumer & Industrial.
The maximum cable length between the entrance EMI filter and the PDB is 3 m. (Implying the separation
distance between the EMI filter box and the PDB is less than 3 m).
Route separately EMI filter incoming power lines and outgoing power lines (refer to
Electrical Connections).
Manufacturer's name and address, address of the manufacturing site and European registered
place of business (the two addresses correspond to the same entity):
FOR CONTINUOUS COMPLIANCE WITH THE MEDICAL DEVICE DIRECTIVE 93/42/CEE (CE
MARKING), CONSULT YOUR GE REPRESENTATIVE BEFORE ADDING OPTIONS OR
ACCESSORIES OTHER THAN THOSE INDICATED IN THIS MANUAL. THE CE LABEL FOR
THE SYSTEM IS LOCATED ON THE C-ARC.
• Medical Device Good Manufacturing Practice Manual issued by the FDA (Food and Drug Administration,
Department of Health, USA).
• Underwriters' Laboratories, Inc. (UL), an independent testing laboratory.
• Canadian Standards Association (CSA).
• International Electro technical Commission, international standards organization, when applicable.
• EMC IEC-60601-1-2: Edition 2.1 (2004) and Edition 3 (2007).
• Innova IGS system has been designed, manufactured and tested to meet the applicable requirements of 21
CFR 1020.30, 1020.32. When the Innova IGS system is installed and tested according to the manufacturer
instructions, the system meets the applicable requirements. Each components falling under the performances
regulated in 21 CFR 1020.30 bears a certification label mentioning the following text:
"COMPLIES WITH DHHS RADIATION PERFORMANCE STANDARDS 21 CFR SUBCHAPTER J".
• CDRH certification label of the C1 cabinet and C2 cabinet can be found on the door of the cabinet.
• CDRH certification label of the Jedi generator can be found on the part into the C1 Cabinet.
• CDRH certification label for the Collimator, Tube and Detector (CDRH label = model number) are located on the
parts.
• CDRH certification label of VCIM (Vascular Control Interface Module) and the TSSC (Table Side Status Control)
are located on the parts.
• CDRH certification label of Tabletop can be found on the backside of the tabletop.
– A copy of the labels of the certified CDRH components that are not accessible to view is provided on the
L-Arm for collimator and X-ray tube; and on the Gear cover for X-ray flat panel detector.
• Regulatory markings for a specific country are only in the concerned local language. In general, these markings
are designated by the related country code as per ISO 3166-1.
• Eurasian Conformity (EAC)
This product passed all conformity assessment (approval) procedures that correspond to the requirements of
applicable technical regulations of the Customs Union.
This product complies with the regulatory requirements set in the regulation CU TR 020 / 2011 ‘electromagnetic
compatibility of technical products'.
1-3 Specific Requirements for product with Open Monitor suspension (Option)
For Innova IGS Systems with Open Monitor Suspension option, the overhead monitor suspension shall be installed
by strictly following the GEMS installation instructions. The manufacturer specifically disclaims any and all liability
arising out of or relating to the use or performance of the monitor suspension (including cables), including, without
limitation, any liability or claims relating to patient injury, death, or the reliability of such monitors suspension(s).
This option consists in the delivery of a Innova IGS System without the standard fully integrated Mavig overhead
suspension, its associated suspension rails and cables, to allow the customer (purchaser) to choose his (their) own
monitors suspension(s). Only GE monitors are permitted as primary image display.
The Innova IGS system with Open Monitor Suspension option is delivered with up to 6 exam-room 19” monitors,
and the display features are then the same as a Innova IGS System delivered with a standard Mavig monitors
suspension.
Note: This option does not modify the Control room monitor configuration delivered with the Innova IGS system.
Where an overhead monitor suspension(s) is supplied by the Purchaser of the Innova IGS System, the overhead
monitors suspension(s) shall comply with the applicable Regulation enforced in the country (e.g., when installed in
an European Community country, the associated monitors suspension(s) shall be CE marked under Directive
93/42/EEC).
The association of Innova IGS product delivered with Open Monitor Suspension option and the purchaser’s
(customer) monitors suspension(s), is not covered by Innova IGS product certification.
1-4 Recycling
• Instruments or accessories at end of life span:
Disposal of equipment and accessories must meet the applicable national regulations for waste treatment.
Equipment and accessories at the end of their life span must have all materials or components removed that
can be hazardous to the environment (such as accumulators, batteries, transformer oil, etc). Please consult
your local GEHC contact before disposing of these products.
• Packaging:
The materials used for our equipment packaging are recyclable. These materials must be collected and
processed according to the regulations applicable in the countries in which equipment or accessories are
unpacked.
This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted
municipal waste and must be collected separately. Please contact an authorized representative of the
manufacturer for information concerning the decommissioning of your equipment.
• Battery recycling (Battery 2006/66/EC):
The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be
recycled or disposed of in accordance with local or country laws. The letters below the separate collection
symbol indicate whether certain elements (Pb=Lead, Cd=Cadmium, Hg=Mercury) are contained in the battery.
To minimize potential effects on the environment and human health, it is important that all marked batteries that
you remove from the product are properly recycled or disposed.
For information on how the battery may be safely removed from the device, please consult the service manual
or equipment instructions. Information on the potential effects on the environment and human health of the
substances used in batteries is available at this url:
http://www.gehealthcare.com/euen/weee-recycling/index.html.
1-5 Control of Pollution caused by Electronic Information Products (EIP's)
The purpose of this chapter is to answer the position that GEHC has taken regarding the People's Republic of
China Electronic Industry Standard SJ/T11364-2006 "Marking for the Control of Pollution Caused by Electronic
Information Products" (EIP's).
1-5-1 Explanation of Pollution Control Label
This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese
standard SJ/T11363-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic
Information Products. The number in the symbol is the Environment-Friendly Use Period (EFUP), which indicates
the period during which the toxic or hazardous substances or elements contained in electronic information products
will not leak or mutate under normal operating conditions so that the use of such electronic information products will
not result in any severe environmental pollution, any bodily injury or damage to any assets. The unit of the period is
"Year"
In order to maintain the declared EFUP, the product shall be operated normally according to the instructions and
environmental conditions as defined in the product manual.
Consumables or certain parts may have their own label with an EFUP value less than the product.
This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled
properly after decommissioning.
1-5-2 Name and Concentration of Hazardous Substances
Innova Single Plane - Hazardous substances' name and concentration
This product consists of devices that may contain mercury, which must be recycled or disposed of in accordance
with local, state, or country laws. (Within this system, except 19" Monitor, the backlight lamps in some monitor
display contain mercury.)
Devices containing Mercury bear the following label:
2 General Safety
In case of a discrepancy between a generic and a specific instruction for use, the specific instruction will always
prevail.
This system contains operating safeguards to provide maximum safety. Before calling for service, be certain proper
operating procedures are being used.
Satisfactory equipment performance requires the use of service personnel specially trained on X-ray apparatus. GE
Healthcare is responsible for the effects on safety, reliability and performance only if the following conditions are
met:
• The electrical wiring of the relevant rooms complies with all national and local codes as well as the Regulations
for the electrical equipment of buildings published by the Institution of Electrical Engineers.
• All assembly operations, extensions, re-adjustments, modifications or repairs are carried out by GE Healthcare's
authorized service representatives.
• The equipment is used in accordance with the instructions for use.
ALWAYS BE ALERT TO SAFETY WHEN YOU OPERATE THIS EQUIPMENT. YOU MUST BE
FAMILIAR ENOUGH WITH THE EQUIPMENT TO RECOGNIZE ANY MALFUNCTIONS THAT
CAN BE A HAZARD. IF A MALFUNCTION OCCURS OR A SAFETY PROBLEM IS KNOWN TO
EXIST, DO NOT USE THIS EQUIPMENT UNTIL QUALIFIED PERSONNEL CORRECT THE
PROBLEM.
OF SAFETY.
2-1 Equipment Application Specification
2-1-1 Patient Population
• Age: newborn to geriatric.
• Weight: 204 kg maximum.
2-1-2 Intended Operator
• Knowledge: The usage of the system described in this Operator Manual requires specific technical and medical
knowledge and skills regarding, at a minimum, radiation protection, safety procedures and patient safety.
People using, moving, working with the system must have acquired such knowledge and skills during their
education curriculum.
• Language Understanding:
Operator Manual and System Screens are translated* in the following languages. Users must understand one of
these languages before using the system.
RU Russian English
SR Serbian English
ZH-CN Simplified Chinese English
SK Slovakian English
ES Spanish Spanish
SV Swedish Swedish
TR Turkish English
Note: * Depending on the region of sale, some translations may not be available.
• Equipment Training: Training is available from GE Healthcare. Some country regulations require that equipment
training be followed before any use of the system. The facility's management should assess the initial and
continuous training needs of qualified users, radiologists or other physicians responsible for the interpretation
and / or acquisition parameters. The facility should identify the individuals that should take part in the training
relative to dose-reduction features, and dose related acquisition parameters.
Contact GE Healthcare representative for training options.
• Training Options: From in person application training to self-training with the Online Operator Manual delivered
with the Innova IGS System and the safety instructions of the paper booklet, Operator Manual DVD. The
Operator Manual DVD shall be placed in a location easily accessible to the operator in case of emergency.
• Training Frequency Recommendation: The facility’s management should assess the initial and continuous
training needs of qualified users, radiologists or other physicians responsible for the interpretation and / or
acquisition parameters. The facility should identify the individuals that should take part in the training relative to
dose-reduction features, and dose related acquisition parameters.
For safe and effective use of the system, GE Healthcare recommends that each user of the equipment reviews
the procedures and safety precautions at least on annual basis and specially for dose management, GE
Healthcare recommends a semestrial training.
Contact GE Healthcare representative for refresher training options as part of the Radiological Technologists
continued education requirements for a licensure maintenance.
It is highly recommend that physicians be provided adequate time to participate in training and image quality
feedback.
2-1-3 Application Environment
• General
– Indoor use only.
– The Innova is suitable for Interventional X-ray procedures (angiograms, device placement, non vascular or
minimally invasive surgery) and can be used:
• either in an Interventional Room.
• or in a dedicated lab within an Operating Room Environment which is outside a location designated for
marked "AP" or "APG" Equipment. Category AP Equipment is an Equipment or Equipment part complying
with specified IEC60601-1 requirements on construction, marking and documentation in order to avoid
sources of ignition in a flammable anaesthetic mixture with Air. Category APG Equipment is an
Equipment or Equipment part complying with specified IEC60601-1 requirements on construction,
marking and documentation in order to avoid sources of ignition in a flammable anaesthetic mixture with
Oxygen or Nitrous Oxide.
– The Operating Room Environment is an environment which has environmental characteristics equivalent to
an Operating Room for infection control. Special attention is required to clean suspended equipment as per
device recommendation and Preventive Maintenance.
– The device is not to be used to do open surgical procedures except when equipped with the OR table and if
installed in a dedicated Operating Room.
• Conditions of Visibility
– Ambiant luminance: 150 lux maximum.
• Physical
– Temperature range: 15°C to 32°C.
– Relative humidity range: 30% to 70%, non condensing.
– Altitude range: sea level to 2000 m (15-32°C) .
Note: The chiller is able to dissipate maximum continuously power at 3000 meters in an ambient
temperature up to 20°C, for this altitude the technical room temperature shall not exceed 20°C.
– Atmospheric pressure range: 794 hPa to 1060 hPa.
Read and understand all the instructions in the Operator Manual before attempting to use the system and request
additional training from GE Healthcare if needed.
Periodically open the Operator Manual to review the procedures and safety precautions.
operating properly.
When covering the equipment with sterile drapes or other protective covering, be sure th e covering does not
interfere with the Motion Controls or Motion Enables.
Note: As per local regulations, record of these functional checks have to be kept with the equipment.
Motion Controls, Motion Enables and Table Top Brakes.
– Perform a daily check of the gantry pivot axis (RAO/LAO rotation) position by moving it to PA position by
using the Auto Positioner command.
Check that the gantry C arm is in vertical position and that "RAO/LAO 0 deg" +/- 1 deg is displayed on the
Geometry area, left on the live display.
Note: When using the InnovaIQ Table, position the table horizontally prior to moving the gantry to a PA
position.
Move the gantry pivot axis and check that the displayed angle is updated.
In case of mismatch detected, call your local Service Representative.
– For system using an Omega IV, Omega V or InnovaIQ Table, perform a daily check of the table Up/Down axis
by moving the table to its lowest position.
Check that " " is displayed on the Geometry area, left on the live display.
Move the table Up and check that the displayed value is updated.
In case of mismatch detected, call your local Service Representative.
2-2-1 Emergency power off
CHECK DAILY THE PRESENCE OF THE LIVE IMAGE AND THE REFERENCE DISPLAY ON
THE 2 BACKUP MONITORS THAT ARE LOCATED EITHER AT THE BACK OF THE
SUSPENSION OR ON AN ADDITIONAL SUSPENSION.
2-2-4 Innova Central touch screen.
• Check functionally the Innova Central touch screen mechanical fixing on the rail.
• Inspect visually the Innova Central touch screen housing and cable integrity.
• Inspect visually the Innova Central touch screen for chemical residuals or blinded spots.
• Check the operation of the touch-button selection, the joystick motion and pushbutton functions .
THE DIGITAL HEAD HOLDER MUST NOT BE USED TO RESTRAIN THE PATIENT'S HEAD
WHEN THE TABLE IS TILTED, ONLY WHEN IT IS HORIZONTAL. IT IS NOT DESIGNED
FOR THAT USAGE AND MAY CAUSE INJURY TO THE PATIENT.
– Horizontal Armboard (HB-1)
The horizontal armboard (HB-1) with a flat plate sliding under the mattress enables to support patient’s arm.
Its radiolucency allows imaging the arms.
– Quick Strap
Disposable Velcro-like strap helps immobilize patient.
Quick strap is not to be used when the table is tilted. The strap might exert mechanical
force which may harm the patient.
– Clear-Vu Arm Support
The Clear-Vu Arm Support provides comfortable means of supporting the patient’s arms in a prone or supine
position during an interventional procedure.
Attach only GE approved accessories to the table. Be sure to connect each accessory to the
proper connector, as labeled on the table. In particular, be sure to connect the grounding
cables.
On each screw of the table accessory rail, the maximum allowed torque is 64 N.m. All GEHC accessories
listed above are compliant with this specification. In case of non-GEHC accessory installation on the table
accessory rail, make sure the torque remains below 64 N.m.
2 - If needed, an optional rail can be installed at table foot end of the Omega V table for other accessories
installation. The total weight of all devices installed on this rail shall remain below 20 kg.
In any case, always make sure the weight of devices on the table side-rail remains below 40 kg, and below
20 kg on the optional table foot end rail. Also, make sure the torque on each screw is below 64 N.m,
according to the following formula:
Torque = 10 x weight (kg) x distance to the rail (m).
For example, a 5 kg load, located 10 cm away from the rail, has a torque of : 10 x 5 (kg) x 0.1 (m) = 5 N.m.
Check that each accessory is properly mounted and clamped if provided with clamps before
any use.
2-8 Periodic maintenance and inspection
Periodic maintenance is required for continued SAFE operation. Qualified personnel should
perform periodic maintenance as specified in the maintenance schedule of the Service
Manual and Operator Manual.
• Descriptions and scheduled frequency of the required periodic maintenance are provided in the Service
documentation supplied with the equipment. It is the owner's responsibility to supply or arrange for this service.
Refer to Maintenance Schedule Chapter for the list of the periodic maintenance requirements.
• Inspection intervals are based on average daily use of one eight-hour shift. More frequent inspection is
appropriate where equipment use is above average. Refer to Section Safety and Regulatory / General Safety /
Be Sure the Equipment is Functioning Before each Use in this chapter or refer to the table below for details.
Daily Checks
Actions Refer to
-
Shutdown at least once per day
Gantry pivot axis check Safety and Regulatory / General Safety / Be Sure the Equipment is
Functioning Before each Use
Table up/down check Safety and Regulatory / General Safety / Be Sure the Equipment is
Functioning Before each Use
Collision sensors check Safety and Regulatory / General Safety / Be Sure the Equipment is
Functioning Before each Use
Diamentor constancy check Diamentor M4 KDK (Option) / Performing a Measurement / Constan-
cy Check
Innova Central touch screen Safety and Regulatory / General Safety / Be Sure the Equipment is
Functioning Before each Use
Monthly checks
Actions Refer to
Wall and control panel emergency stop Safety and Regulatory / General Safety / Be Sure the Equipment is
Functioning Before each Use
Suspensions cleaning Safety and Regulatory / Protection against contamination / infection /
allergy / toxicity hazards / biological hazards / Suspensions Cleaning
QAP check Quality Assurance Program (QAP) / Parameter monitoring / Fre-
quency
Put storage media (tapes, CDs) in safe conditions: temperature (see the instructions on CD
packages).
2-10 Extra Iodine Injection
When used in automatic mode, the injector can be triggered by the system before or after X-ray (X-ray delay or
Inject delay).
When the injector is fired before X-ray, there is always an unpredictable residual risk that X-ray will not start, so, the
injection could be lost (our internal data give a very low occurance rate).
Always use the X-ray delay mode (injection first then X-ray) only when clinically appropriate.
2-11 Image Presentation
The image presentation does not automatically follow the patient position over the table top
(Prone, Supine, Head to Toes reversed…). It is the responsibility of the Operator to control
the patient orientation on the image. Failure to control may lead to misinterpretation of the
image displayed for diagnostic purposes.
In case of long procedures and/or when X-ray is focused on a single location. Use low
fluoro and record detail in order to reduce risk of patient injury due to ionizing radiation and
other recommendation listed in Safety and Regulatory / Protection against radiation / How
to Reduce Radiation Exposure Dose.
3-2 Dose Monitoring
• "Dose monitoring" is a feature of the Innova system, displaying air kerma rate, cumulated air kerma and DAP.
• Output parameters are displayed in the DL console and on the live & reference displays.
• The dose levels are determined with a measurement method based on the measured Generator X-ray
Technique parameters (kVp, mAs), on the Spectral Filtration position reported by the collimator, on measured
mR/mAs calibration values obtained with a radiation dosemeter, and on a transfer function (computation).
• The air kerma (referred as "Exposure Dose" / "Rate") and exam dose (referred as "Exposure Dose" /
"Cumulated") parameters are computed for the Interventional Reference Point.
• The DAP (Dose Area Product) is computed based on collimator opening (FOV and collimation).
• The accuracy of the displayed dose rate and cumulated dose is ±35% down to 100 mGy for dose and 6
mGy/min for dose rate. The accuracy of the DAP is ±30% down to 2.5 Gy.cm2 for DAP.
3-3 Concept of Interventional Reference Point (Patient Entrance Reference Point)
The INTERVENTIONAL REFERENCE POINT (IRP) or PATIENT ENTRANCE REFERENCE POINT, is intended to
be representative of the point of intersection of the X-ray beam axis with the patient. For the GE Healthcare
systems including an isocenter, the IRP is the point on the reference axis 15 cm from the isocenter towards the
focal spot.
This distance is assumed to represent a good approximation of the value of the actual focal spot to skin distance
during interventional procedures. If one considers currently available methods to estimate absorbed doses to
selected tissues for radioscopic and cine-angiographic examinations of the coronary arteries of adults, these
methods rely on the use of distinct operating conditions commonly used in radiological examinations of the heart.
These operating conditions are associated with a view, an arterial projection, and technique factors on the X-ray
equipment such as the X-ray tube voltage (kV), the half value layer (HVL), and the focal spot to skin distance, the
focal spot to image receptor distance and the entrance field size.
A review of the operating conditions derived from analyses of practice indicates that the defined interventional
reference point is, in fact, a fair approximation of the focal spot to skin distance for each field. Unfortunately, similar
data are not available for non-cardiac diagnostic and interventional procedures.
Isocenter
The Isocenter is the point at which the rotational axes of the L-Arm and the rotational axis of the C-Arm intersect.
Focal Spot to Interventional Reference Point Distance
For Frontal plane X-ray Tube: 57 cm (22.4").
3-4 Significant Zone of Occupancy
There is a significant zone surrounding the table that identifies proximity to the radiation source. Included in the
significant area is the table side area where the Physician stands to perform the procedure.
3-5 Skin Spacer Use
EXPOSURE RATE IS APPROXIMATELY 1.5 TIMES GREATER AT 305 mm FROM THE FOCAL
SPOT THAN AT 380 mm. USE THE SKIN SPACER TO ENSURE A MINIMUM DISTANCE OF
380 mm.
3-6 How to Reduce Radiation Exposure Dose
3-6-1 Radiation Time
In Fluoro: check and control the elapsed fluoroscopy time. When possible change the entry point of the beam in
order to reduce local dose.
In Record: try to keep the frame rate at the lowest value compatible with medical requirements; minimize the total
sequence.
3-6-2 Protection
When the tube is positioned under the table the patient helps protect the staff.
Additional lead aprons and protective devices such as lead gloves, glasses and thyroid shields are also effective in
limiting occupational exposure dose.
In case of obvious X-ray emission (audible signal and/or red light signal on) without image on the live display,
perform the following actions:
1. Release the footswitch.
2. If X-ray are still emitted, depress the "RESET" button on the system console until the system resets.
3. If X-ray are still emitted, switch off the Power Distribution Box (PDB) by turning to OFF the RED PDB switcher
on the system cabinet, in the technical area.
3-8 Protective devices and Accessories
Refer to the accessories catalog for a complete list of available radio-protection devices. (Rad shield, gloves,
apron, glasses, etc.)
3-9 Retake - Improper Use
Improper use of the system could lead to retakes of diagnostic images, thus requiring
additional patient and operator exposure to radiation. Therefore, observe operating
instructions.
3-10 Dose and Dose Rate Measuring Procedure
This procedure is intended to provide the details to be able to conduct measurement of radiation data consistently
with radiation data documented in the user instructions specific to each vascular system.
3-10-1 Test Equipment
3-10-1-1 Phantom
Material - Polymethyl-methacrylate (PMMA)
Rectangular blocks with sides equal to or exceeding 25 cm.
Nominal thickness: 20 cm. (the phantom may be fabricated from layers of material) area density of the nominal 20
cm.
Phantom: 23.5 gcm-2 +/- 5 %.
3-10-1-2 Dosimeter
The measuring probe must be small enough to cover not more than 80 % of the area of the X-ray beam in the
plane of measurement and the area of its surface perpendicular to the source-detector axis must not exceed 30
cm².
3-10-2 Measurement Procedure
3-10-2-1 Relevant parameters
The following are the factors that need to be referenced, when relevant to the equipment settings concerned:
• Equipment configuration:
1. Orientation of the X-ray BEAM (vertical unless otherwise specified);
2. PATIENT SUPPORT in or out;
3. ANTI-SCATTER GRID in or out;
4. Appropriate ENTRANCE FIELD SIZE selected (FOV).
• Operating settings (representative of normal use):
1. Technical details of parameters included in each mode of operation;
2. Frame rate;
3. Selectable added filters automatically applied;
4. Selectable added filters manually applied.
• Test geometry:
1. Focal spot to image receptor distance (SID);
2. Focal spot to detector distance;
3. X-ray field size at the measuring detector;
4. Positioning of phantom;
5. Positioning of measuring detector.
For each of these factors, a value is assigned and is indicated prior the radiation data stated (see hereafter in the
user instructions specific to each vascular system) they must absolutely be used in case of measurement of
radiation data.
3-10-2-2 Positioning of phantom and measuring chamber
The phantom is placed near the image receptor, leaving as much of the available distance as possible between the
X-ray tube assembly and the entrance surface of the phantom. (This is to minimize the effect of scattered radiation
on the measurements). Refer to radiation data in the user instructions specific to each vascular system.
Recommendation:
Position the measuring detector at a point that is either at the interventional reference point (see Safety and
Regulatory / Protection against radiation / Concept of Interventional Reference Point (Patient Entrance Reference
Point) (only if there is at least 20 cm distance between the measuring detector and the phantom) or half-way
between the focal spot and the entrance surface of the phantom. In that case, the readings are to be corrected to
the appropriate distance. Any other position that would minimize stray radiation may be used.
The surface of the phantom must be aligned perpendicular to the X-ray beam axis within ±2 degrees in all
directions.
3-10-2-3 Checking the test conditions
Before any dosimetric measurements are done, verify that the equipment settings under test and the associated
measuring arrangements are in compliance with Configurations of the equipment and test geometries defined
in radiation data in the user instructions specific to each vascular system.
Dosimetry is not to proceed, and therefore compliance with dosimetric statements cannot be determined, unless
the associated information is in itself compliant.
3-10-2-4 Basic measurements
Measure the dose rate (air kerma rate) for each of the radioscopic settings.
Measure the air kerma per image for each radiographic settings.
Measurements are made using the 20 cm phantom.
3-10-2-5 Operator selectable parameters
For settings in normal mode measured in previous paragraph "Basic measurements", measure the dose rate (air
kerma rate) using the nominal 20 cm PMMA phantom:
• for each operator selectable entrance field size (FOV); (as defined in radiation data).
• for each operator selectable added filter; (if applicable as defined in radiation data).
• for each operator selectable pulse repetition frequency; (if applicable).
3-10-3 Compliance Criteria
From the measured values obtained in the tests, calculate the corresponding values of reference air kerma (rate) at
Interventional Reference Point by using the inverse square law if necessary.
Compliance is verified if:
• no value stated in the instructions for use (hereafter in paragraph "radiation data") differs from the
AFTER POWERING OFF THE SYSTEM BY THE GENERAL BREAKER OR THE CONSOLE
ON/OFF, SOME PARTS CONTINUE TO BE ENERGIZED THROUGH A BATTERY FOR A
COUPLE OF MINUTES IN ORDER TO PROPERLY SHUTDOWN THE COMPUTERIZED
COMPONENTS. BEFORE INSPECTING OR SERVICING THE SYSTEM, ENSURE THAT ALL
PARTS ARE CORRECTLY DE-ENERGIZED. (REFER TO THE SAFETY PROCEDURE DEFINED
IN THE SERVICE MANUAL).
4-4 Laser
A laser product is used to connect external modalities to the Large Display Monitor.
This product implements an optical module used for data transmission, manufactured by OPTICIS CO., LTD., and
classified as laser Class 1.
Center wavelength minimum = 1300 nm and maximum = 1550 nm.
Invisible LED radiation not in excess of class 1 safety limits (CDRH and EN 60825).
If applicable:
The system anti collision device is based on a virtual volume representing the patient loaded on the tabletop in
"standard position".
This "standard position" is defined as head of the patient located at head end (gantry side) of the tabletop.
1 Patient Head
For any other position of the patient over the table top (use of table head extender, patient positioned lower on the
table top, patient head to toes reversed...), a special attention is required from users to watch any gantry motion
when approaching the patient, the virtual patient volume being in this case not in the same location than the patient
on the table top.
5-2 Use Extreme Care when Positioning the Equipment
The safety of everyone near the moving equipment depends on the operator at the controls. Although collision
sensors are provided on the image receptor and collimator, they are not intended to be a substitute for careful
operation. Be sure the equipment path is clear before starting a motion. When positioning the equipment, carefully
observe the moving parts, and stop the motion before making contact. The sensors are provided for added
protection in case of inadvertent contact, and are not to be relied on as a normal means for stopping the motion.
Improper equipment positioning could result in serious injury or equipment damage.
IN THE "FOLLOW THE GANTRY" MODE, USER’S ATTENTION IS MORE FOCUSED ON THE
AW IN-ROOM MONITOR THAN ON THE MOVING GANTRY. DURING SUCH OPERATION,
ALSO KEEP CONSTANT ATTENTION TO THE GANTRY TO AVOID COLLISION WITH THE
PATIENT AND PEOPLE WORKING CLOSE TO THE GANTRY. FAILURE TO FOLLOW THIS
RECOMMENDATION COULD RESULT IN SEVERE INJURIES TO THE PATIENT OR STAFF
WORKING IN THE ROOM.
Always use proper accessories to hold the patient and prevent risk of injury during the
motion of the equipment.
Never leave the patient unattended. An unattended patient could fall from the table, activate
a motion control, or encounter other problem, which could be hazardous. In addition, it is
recommended to use restraints, such as Velcro straps. The restraints limit, but do not
eliminate the risk for the Patient to fall; in no way they can be used to support the Patient
weight.
With a InnovaIQ Table, Velcro straps are not appropriate to immobilize the patient when the table is tilted,
patient could slightly slide over the mattress and be hurt by the Velcro straps. To prevent the risk for the
patient to fall, never leave the patient unattended while the table is tilted.
The Omega tabletop mattress is provided with three Velcro strips on each side. The InnovaIQ Table Top
mattress is provided with a set of Velcro strips on bottom side. While loading or unloading the patient, it is
recommended to grip the Velcro strips to avoid the mattress shifting.
Ensure the Velcro strips are well stuck on the tabletop prior to loading or unloading the patient.
DRIVING THE TABLE TOP INTO FIXED OBJECTS MAY SEVERELY INJURE PATIENT. BE
SURE TO CHECK FOR OBSTRUCTIONS ABOVE AND BELOW THE TABLE BEFORE
RAISING OR LOWERING THE TABLE TOP.
5-7 Cardiac Pulmonary Resuscitation (CPR)
THE GANTRY AND TABLE MOTIONS MAY BE SLOWED DOWN BECAUSE OF A NEAR
COLLISION RISK. IN THIS CASE, IT MAY BE FASTER TO USE MANUAL COMMANDS TO
MOVE THE PATIENT TO A CPR POSITION.
5-7-2 CPR with Omega Table
CPR position can be reached by manually moving all needed axes.
To reach CPR position, move the table top toward the foot end. CPR can be performed when the system is in this
position.
In addition to reaching the CPR position as described above, the table height may be lowered and Detector may be
moved to its highest position.
In the event of power loss the table may be manually pushed longitudinally to the foot end
and rotated. This operation may require more than one person.
With the Omega Table, the Positioner Enable/Desable button only enables/disables
commands from the Smart Box/Handle and Table Side Status Control. It does not
enable/disable table top brakes command from the Fluoro/Record footswitch or the Table
Panning handle. This caution is not applicable on InnovaIQ Table.
In case of table hang when tilted which cannot recover after a system reset, a tilting
bed/gurney should be used to securely transfer the patient from the tilted table to the tilted
bed/gurney. Also, hold the tabletop and securely support the patient as the transfer is made
from table to bed/gurney. It is mandatory to comply with these instructions in order to
minimize the risk of patient falling from table during unloading process.
System User Interface
7 Positioner Enable/Disable: activate this button to disable all table/gantry axis commands (button light ON).
Press again on this button to reactivate all table/gantry axis commands (button light OFF)
5-9 Clearance of the Patient in case of Emergency (Power Loss, Equipment Failures)
In the event of power loss with the patient positioned between the image receptor and table,
the patient may be released by manually moving the table longitudinally to the foot end and
then rotating it. In case of InnovaIQ Table, the table rotation only can be manually performed.
This operation may require more than one person.
IN CASE OF INNOVAIQ TABLE, THE TABLE ROTATION ONLY CAN BE MANUALLY PERFORMED. THIS
OPERATION MAY REQUIRE MORE THAN ONE PERSON.
5-12 3rd Party Imaging/Data Medical Device
For 3rd party imaging/data medical devices, refer to the safety recommendations in sections: Cathlab Frontiers
Solutions (Option) and Workstation Configuration / CA 1000 Workstation Configuration (Option).
In order to prevent any patient contamination and according to the local regulation, after
each procedure, thoroughly clean and disinfect any parts likely to become soiled during
use. This includes parts which were in patient vicinity or in contact with the patient and any
parts that could be contaminated by operators, in particular the waterproof mattress.
Improper cleaning methods or the use of certain cleaning and disinfecting agents can
damage the equipment, cause poor imaging performance or increase the risk of electric
shock.
To avoid possible injury or equipment damage:
• Do not use harsh detergents, abrasive cleaners, high alcohol concentration or Methanol at any concentration.
• Do not expose equipment parts to steam or high temperature sterilization.
• Never allow liquids to enter the internal parts of the equipment. Do not apply cleaning sprays or liquids directly
to the equipment; always use a clean cloth dampened with the spray or liquid. If you become aware of liquid
entry, disconnect the electrical supply and have the equipment checked by qualified service personnel before
returning it to use.
6-2 Equipment Cleaning Instructions
The entire system surfaces should be scrubbed as needed using a soft, damp gauze or cloth to remove all visible
residue. Scrubbing with a soft bristle brush (such as a toothbrush) may be necessary to reach corners or to remove
material that has dried onto the surface.
Subsequent disinfection may not be effective if the surfaces are not thoroughly clean.
Rinse all surfaces with clean water to remove visible soap residue, taking care to avoid liquid entry to internal
equipment parts. Dry surfaces with a soft cloth to remove any visible residue.
Due to the relationship between length and multiple widths of the mattress, it is absolutely recommended not to
remove the cover from the mattress for laundry. Otherwise, it would then be extremely difficult to reposition the
mattress in its cover.
Although the footswitches are compliant with the applicable standards, it is recommended
to avoid liquid (bodily fluids, contrast agents, cleaning agents...) ingress into the footswitch.
In case of liquid ingress, ensure the footswitch is thoroughly rinsed and dried. In addition
we recommend to put a drape over the footswitch to avoid any bodily fluids, contrast
agents, cleaning agents ingress.
disinfected with a liquid chemical germicide rated for high-level disinfections. The same process used as that for
intermediate level disinfections is generally followed; however, the time of contact is usually much longer for
high-level disinfections.
6-6 Recommended Cleaning Agents
• Soap with lukewarm water.
6-7 Recommended Disinfecting Agents
The following legally marked products have been used on GE equipment without causing equipment damage.
• SURFANIOS (Supplier: ANIOS)
• CLEANSINALD (Supplier: TASKI) replacing the TASKI DS 5001 no more available
• MORNING MIST (Supplier: BUTCHERS)
6-8 Guidance for the Use of the Cleaning and Disinfecting Agents Listed
Note: For the Large Display Monitor, only Surfanios, Cleansinald and Morning Mist products have been tested.
Always follow the cleaning and disinfecting agents manufacturer's instructions and
precautions for mixing, storage, method of application, contact time, rinsing requirements,
protective clothing, shelf life and disposal to help assure effective and safe use of the
product.
6-9 Flammable Gasses
6-10 Latex
There is no presence of latex in the system and no part in contact with the patient contains latex.
FOR EVERY EQUIPMENT CONNECTED ON THE NETWORK AND USING DICOM, USERS
MUST VERIFY THE COMPATIBILITY THROUGH DICOM CONFORMANCE STATEMENT
2 Safety
2-1 Environmental Conditions
The conformity with the standard IEC 60601-1 series has been established for an ambient temperature up to 32°
Celsius.
2-2 Equipment Classifications
The following equipment classifications are applicable to the product:
The OR table mattress has antistatic properties. As it is connected to the ground and placed on a conductive
tabletop, this provides an antistatic leakage path for the surgical configuration: it is mandatory to use the OR table
mattress provided with the equipment.
* Classification Exemption:
The following accessible conductible parts are grounded.
The Innova systems are compliant to electromagnetic compatibility IEC 60601-1-2 Edition 2.1 (2004) and Edition 3
(2007) standards for medical devices.
For further information, see the pre-installation manual.
3 Electrical Hazards
3-1 Power-off the System
AFTER THE SYTEM IS POWERED-OFF BY THE INNOVA CONSOLE OR THE PDB (POWER
DISTRIBUTION BOX), THE FOLLOWING COMPONENTS ARE STILL ACTIVATED BY THE
UPS (UN-INTERRUPTIBLE POWER SUPPLY) BATTERY FOR ABOUT 5 MINUTES. IT
ALLOWS THE COMPUTERIZED COMPONENTS OF THE SYSTEM TO BE PROPERLY
SHUTDOWN:
• C1 CABINET
• INNOVA DIGITAL SCREEN AND REMOTE KVM POWER SUPPLY (MOUSE AND KEYBOARD POWER
SUPPLY BOX)
• THE LARGE DISPLAY CABINET IN THE TECHNICAL ROOM AND THE LARGE DISPLAY MONITOR IN
THE EXAM ROOM
An Emergency Power OFF/STOP Button is also located on the Power Distribution Box (PDB).
PDB EU PDB UL
1 - ON 1 - ON
2 - OFF 2 - OFF
– Wait for 30 s that the UPS "Power On Indicator" stays ON (no more blinking) (Item 2).
AutoExposure preferences and dose reduction strategy is set by Service via the Service User Interface (SUIF).
Dose for Low Detail mode is about 50% of Normal Detail mode for Fluoroscopy, Dynamic Record and
InnovaChaseTM (except for Fluoroscopy around high reference dose (87 mGy, 43 mGy, 21 mGy)), and < 60% of
Normal Detail dose for DSA, Bolus and 3DCT.
Depending on system configuration and available options, the Normal detail for unsubtracted Fluoroscopy is further
optimized for dynamic situations to visualize fast moving objects: High Contrast Fluoroscopy (HCF) option, for
Normal (HCF) detail.
Depending on system configuration and available options, several acquisition modes can be available.
• DSA mode: primary use for general angiography procedures.
• Innova 3D (3D) mode / Innova CT (3DCT) mode: primary use for 3D reconstruction and display of any region of
interest.
• Dynamic Record mode: primary use for general cardiac procedures.
• InnovaChaseTM mode: primary use for angiography procedures not requiring subtraction.
4-1 Mode of Operation
4-1-1 Innova Digital
1 Record parameters display: record mode, detail level, frame rate, acquisition duration and X-ray/inject delay
Record technique parameters display: kV, mA (peak for DSA and Chase, otherwise: average), ms
2 Fluoro parameters display: fluoro mode, detail level and frame rate
Fluoro technique parameters display: kV, mA
Record Screen
Fluoro Screen
Note 5: 3D acquisition can be used in unsubtracted and subtracted modes. In these cases the same loading
factors are used, that results in identical dose/ frame values. For this reason detailed data are provided only for
unsubtracted 3D. Since subtracted 3D requires 2 consecutive spins, thus the integrated dose in sub 3D is the
double of unsubtracted 3D acquisition.
Note 6: Unsubtracted fluoroscopy in 3.75 fps uses the same settings per frame as in 7.5 fps. This results in
identical kVp, pulse width, and spectral filters, while the time averaged mA and dose rate in 3.75 fps fluoroscopy
are half of the corresponding 7.5 fps values. Because of this direct relationship separate tables for 3.75 fps are not
provided.
Note 7: Loading factors and dose rate tables for High Contrast Fluoroscopy (HCF) option are provided for Normal
detail only, since this option is active in Normal fluoro detail only.
4-2-1 Skin Dose Levels
The skin dose levels in normal use may be high enough to cause deterministic effects. The availability of the
different settings (table below) has a considerable effect on the radiation quality, the delivered dose rate and the
image quality.
4-2-2 Available Settings
Selectable Without or Without or Without or Without or Without or with one Without or Without or
added filters with one of the with one of the with one of with one of of the below Cu with one of with one of
automatically below Cu below Cu the below the below thicknesses: the below the below Cu
applied thicknesses: thicknesses: Cu thick- Cu thick- 0.1 mm Cu, Cu thick- thicknesses:
0.1 mm Cu, 0.1 mm Cu, nesses: nesses: 0.2 mm Cu, nesses: 0.1 mm Cu,
0.2 mm Cu, 0.2 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.3 mm Cu 0.1 mm Cu, 0.2 mm Cu,
0.3 mm Cu, 0.3 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.3 mm Cu
0.6 mm Cu, 0.6 mm Cu, 0.3 mm Cu, 0.3 mm Cu, 0.3 mm Cu,
0.9 mm Cu 0.9 mm Cu 0.6 mm Cu, 0.6 mm Cu, 0.6 mm Cu,
0.9 mm Cu 0.9 mm Cu 0.9 mm Cu
Geometry
Source to im- 85 to 119 cm +/- 1 cm
age distance
(SID)
Modes of op- Low, Normal Low, Normal Low, Nor- Low, Nor- Low, Normal Low, Nor- Low, Normal
eration mal mal mal
Default detail Normal/Low Normal/Low Normal/Low Normal/Low Normal/Low Normal/Low Normal/Low
in Adult/
Pediatric
mode
Loading fac- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Normal: Low or Nor- Low or Nor-
tors range af- mal: mal: mal: mal: 60 - 120 kV mal: mal:
ter selection 60 - 120 kV 60 - 120 kV 50 - 125 kV 60 - 120 kV 60 - 120 kV 60 - 120 kV
of the mode
mA: mA: Normal mA: Normal mA: mA: 100 - 866 Normal mA: Normal mA:
Min: 0.0075 Min: 0.015 mA 41.7 - 1000 1 - 866 1 - 786 100 - 866
mA Average Average @
Low mA: Low mA: Low mA: Low mA:
@ 3.75 fps, 2 7.5 fps, 2 ms
41.7 - 500 1 - 681 1 - 786 100 - 786
ms pw pw
Max: 33.6 mA Max: 31.2 mA
Average @ 30 Average @ 30
fps, 8 ms pw fps, 8 ms pw
Highest refer- See tables at See tables at No dose No dose No dose limits ap- No dose No dose lim-
ence dose each Auto Ex- each Auto Ex- limits ap- limits ap- plied: limits ap- its applied:
(IRP = Inter- posure Prefer- posure Prefer- plied: plied: plied:
Normal setting Normal set-
ventional ence ence
Normal set- Normal set- gives approx. 2 Normal set- ting gives ap-
Reference
ting gives ting gives times more dose ting gives prox. 2 times
Point)
approx. 2 approx. 2 than Low setting approx. 2 more dose
times more times more times more than Low set-
dose than dose than dose than ting
Low setting Low setting Low setting
Typical set of values in cardiac procedures
Diagnostic Normal Normal Normal Normal Normal Normal Normal
Interventional Normal Normal Normal Normal Normal Normal Normal
Note: For 5 R receptor dose limited countries, refer to Safety and Regulatory Instructions for Use / Protection
regarding Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited Standard:
Countries limited to 5 R/min in Fluoroscopy.
4-2-3 Radiation Data
Procedure for measuring dose rate.
Refer to the section: Generic Safety and Regulatory instructions for use for vascular systems.
WHEN MEASURING DOSE, ALWAYS DO A FLUORO RUN PRIOR TO THE RUN IN STEP1
(ACQUIRING TECHNIQUE FACTORS), OTHERWISE SPECTRAL FILTER SELECTION, AND
THUS THE MEASURED DOSE MAY NOT BE REPRODUCIBLE.
4-2-4 IQ Standard
Dose rate with 20 cm PMMA Configuration and Operating Settings
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or 108 cm SID), during a 3D/3DCT spin, when the gantry
is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4:The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited
Standard.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors auto- 74.5 kVp, 25.0 mA avg 80.7 kVp, 25.0 mA avg 80.3 kVp, 25.5 mA avg
matically selected
Loading factors automatically selected 80.0 kVp, 15.4 mA avg 80.0 kVp, 23.5 mA avg
Dose rate at International Standardization 6.8 x 1.4 mGy/min 17.4 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automat- 76.5 kVp, 15.6 mA avg 80.3 kVp, 18.7 mA avg 82.2 kVp, 17.9 mA avg
ically selected
Loading factors automatically selected 80.0 kVp, 7.0 mA avg 80.0 kVp, 8.6 mA avg
Dose rate at International Standardization 5.3 x 1.4 mGy/min 12.8 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 15 fps are the 50% of the 30 fps measured
values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors auto- 77.8 kVp, 9.4 mA avg 85.3 kVp, 9.3 mA avg 84.3 kVp, 10.0 mA avg
matically selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 4.6 mA avg 80.0 kVp, 6.2 mA avg
Dose rate at International Standardization 3.5 x 1.4 mGy/min 9.2 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 7.5 fps are the 25% of the 30 fps measured
values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically select- 71.9 kVp, 72.9 mA avg (30 fps), 70.8 kVp, 130.7 mA avg (30 fps), 7.0
ed 5.0 ms ms
Dose rate at International Standard- 38.6 x 1.4 µGy/frame 96.3 x 1.4 µGy/frame
ization conditions (IEC 60601-1)
Loading factors automatically selected 85.5 kVp, 313.5 mA peak, 50.2 94.5 kVp, 327.4 mA peak, 61.7 ms
ms PW PW
Dose rate at International Standardiza- 0.95 x 1.4 mGy/frame 1.54 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 87.0 kVp, 324.1 mA peak, 44.2 98.9 kVp, 326.0 mA peak, 50.0 ms
ms PW PW
Dose rate at International Standardiza- 0.90 x 1.4 mGy/frame 1.40 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors automatically selected 74.8 kVp, 25.3 mA avg, 7.0 ms PW 79.9 kVp, 28.5 mA avg, 7.0 ms
PW
Dose rate at International Standardization 136 x 1.4 µGy/frame 288 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors automatically 74.0 kVp, 56.9 mA avg, 3.5 ms PW 73.0 kVp, 112.1 mA avg, 7.0 ms PW
selected
Dose rate at US regulation con- 123 x 0.73 µGy/frame 239 x 0.73 µGy/frame
ditions (21 Code Of Federal
Regulation)
Dose rate at International Stan- 123 x 1.6 µGy/frame 239 x 1.6 µGy/frame
dardization conditions (IEC
60601-1)
Loading factors automatically 88.8 kVp, 53.8 mA avg, 5.0 ms PW 84.7 kVp, 122.5 mA avg, 7.0 ms PW
selected
Dose rate at US regulation con- 55.0 x 0.73 µGy/frame 110 x 0.73 µGy/frame
ditions (21 Code Of Federal
Regulation)
Dose rate at International Stan- 55.0 x 1.6 µGy/frame 110 x 1.6 µGy/frame
dardization conditions (IEC
60601-1)
Field Of View 20 cm 17 cm 15 cm 12 cm
(nominal)
Field Of View 20 cm
Operating settings
Frame rate 30 fps, 15 fps, 30 fps, 15 fps, 7.5 fps, 3.75 30 fps, 15 fps 30 fps 30 fps
7.5 fps, 3.75 fps 7.5 fps fps, 1.875 fps,
1.0 fps, 0.5 fps
Selectable In Low 30 fps: In Low: 0.1 mm In Low: 0.3 mm In Low: 0 mm In Low: 0.3 mm
In Low 30 fps:
added filters 0.9 mm Cu Cu Cu Cu Cu
0.9 mm Cu
automatically In Normal 30
In Normal 30 In Normal: 0.1 In Normal: 0.3 In Normal: 0 mm In Normal: 0.3
applied fps:
fps: mm Cu mm Cu Cu mm Cu
0.6 mm Cu
0.6 mm Cu
In Normal
(HCF)30 fps: In Low 15 fps:
0.6 mm Cu 0.6 mm Cu
In Normal 15
In Low 15 fps:
fps:
0.9 mm Cu
0.3 mm Cu
In Normal 15
fps: In Low 7.5 fps:
0.6 mm Cu 0.6 mm Cu
In Normal In Normal 7.5
(HCF)15 fps: fps:
0.6 mm Cu 0.3 mm Cu
In Low 7.5 fps:
0.9 mm Cu
In Normal 7.5
fps:
0.6 mm Cu
In Normal
(HCF)7.5 fps:
0.6 mm Cu
Geometry
Source to im- 102.5 cm Max SID
age distance
(SID)
Distance of fo- 67.5 cm 72 cm
cal spot to mea-
suring detector
X-ray field size Square of: 174 cm² Square of: 145 Square of: 178
at the measur- cm² cm²
ing detector
Positioning of Distance between top of phantom and anticollision plate (airgap): Phantom centered in isocenter
phantom 17 cm
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or 108 cm SID), during a 3D/3DCT spin, when the gantry
is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatically 73.0 kVp, 12.6 mA avg 71.5 kVp, 19.3 mA avg 71.5 kVp, 19.2 mA avg
selected
Loading factors automatically selected 80.0 kVp, 6.9 mA avg 80.0 kVp, 9.9 mA avg
Dose rate at International Standardization 3.0 x 1.4 mGy/min 7.5 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automatically 74.0 kVp, 8.4 mA avg 74.1 kVp, 11.6 mA avg 73.6 kVp, 12.0 mA avg
selected
Loading factors automatically selected 80.0 kVp, 3.2 mA avg 80.0 kVp, 4.1 mA avg
Dose rate at International Standardization 2.4 x 1.4 mGy/min 6.1 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 15 fps are the 50% of the 30 fps measured
values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically 72.1 kVp, 7.0 mA avg 72.0 kVp, 9.4 mA avg 75.1 kVp, 7.8 mA avg
selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 2.1 mA avg 80.0 kVp, 2.7 mA avg
Dose rate at International Standardization 1.6 x 1.4 mGy/min 4.0 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 7.5 fps are the 25% of the 30 fps measured
values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically selected 67.1 kVp, 62.8 mA avg (30 fps), 67.8 kVp, 125.1 mA avg (30 fps),
5.0 ms 7.0 ms
Dose rate at International Standardization 25.4 x 1.4 µGy/frame 53.8 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors automatically selected 81.7 kVp, 221.1 mA peak, 42.5 85.7 kVp, 313.0 mA peak, 51.0 ms
ms PW PW
Dose rate at International Standardiza- 0.51 x 1.4 mGy/frame 0.98 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 82.0 kVp, 230.1 mA peak, 40.5 87.7 kVp, 321.8 mA peak, 44.1 ms
ms PW PW
Dose rate at International Standardiza- 0.50 x 1.4 mGy/frame 0.91 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors automatically 75.0 kVp, 32.0 mA avg, 3.3 ms PW 76.0 kVp, 55.4 mA avg, 5.6 ms PW
selected
Loading factors automatically 90.0 kVp, 29.8 mA avg, 3.3 ms PW 88.7 kVp, 53.7 mA avg, 5.0 ms PW
selected
Field Of View 20 cm 17 cm 15 cm 12 cm
(nominal)
Fluoro 30 fps Normal < 43.8 mGy/min < 97.4 mG/min < 43.8 mGy/min < 97.4 mGy/min
Frame- (5 R/min) (5 R/min)
Rate
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
15 fps Normal < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
7.5 fps Normal < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mG/min
(2.5 R/min) (2.5 R/min)
3.75 fps Normal < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7.mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
4-2-7 IQ Plus
For DSA, InnovaChaseTM and 3DCT Normal/Low, all frame rates, please refer to measurements in Dose to Patient
for IQ Standard.
Dose rate with 20 cm PMMA Configuration and Operating Settings
Selectable added Low 30 fps: 0.6 mm Cu Low 30 fps: 0.6 mm Cu In Low: 0.1 mm Cu In Low: 0 mm Cu
filters automati- Normal 30 fps: 0.3 mm Normal 30 fps: 0.6 mm In Normal: 0.1 mm Cu In Normal: 0 mm Cu
cally applied Cu Cu
Normal (HCF) 30 fps: Low 15 fps: 0.6 mm Cu
0.3 mm Cu Normal 15 fps: 0.3 mm
Cu
Low 15 fps: 0.6 mm Cu
Low 7.5 fps: 0.6 mm Cu
Normal 15 fps: 0.3 mm
Normal 7.5 fps: 0.3 mm
Cu
Cu
Normal (HCF) 15 fps:
0.2 mm Cu
Low 7.5 fps: 0.3 mm Cu
Normal 7.5 fps: 0.2 mm
Cu
Normal (HCF) 7.5 fps:
0.2 mm Cu
Geometry
Source to image 102.5 cm Max SID
distance (SID)
Distance of focal 67.5 cm 72 cm
spot to measur-
ing detector
X-ray field size at Square of: 174 cm² Square of: 145 cm²
the measuring
detector
Positioning of Distance between top of phantom and anticollision plate (airgap): 17 cm Phantom centered in
phantom isocenter
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or 108 cm SID), during a 3D/3DCT spin, when the gantry
is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited Standard, countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited
Standard.
For "Dose to Patient" and "Variation of factors" data with country settings Japan, for the Low Dose preference,
please refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding
Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / IQ Standard. The only exception is Max Dose
Reduction 7.5 fps Low Dose preference which is already limited to 21.9 mGy/min when using "IQ Plus".
For "Dose to Patient" and "Variation of factors" data with country settings Australia/WA, Australia/ACT, please
refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 520 / Dose to Patient / IQ Standard.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatically 69.4 kVp, 25.0 mA avg 70.4 kVp, 25.0 mA avg 70.4 kVp, 25.0 mA avg
selected
Loading factors automatically selected 80.0 kVp, 11.8 mA avg 83.6 kVp, 24.6 mA avg
Dose rate at International Standardization 9.0 x 1.4 mGy/min 22.1 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automatically 71.2 kVp, 15.6 mA avg 71.8 kVp, 18.2 mA avg 69.4 kVp, 14.4 mA avg
selected
Loading factors automatically selected 80.0 kVp, 8.6 mA avg 80.0 kVp, 11.2 mA avg
Dose rate at International Standardization 6.6 x 1.4 mGy/min 16.7 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 15 fps are the 50% of the 30 fps measured
values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically 73.6 kVp, 4.1 mA avg 72.6 kVp, 8.2 mA avg 73.9 kVp, 7.5 mA avg
selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.4 kVp, 5.8 mA avg 83.4 kVp, 6.4 mA avg
Dose rate at International Standardization 4.5 x 1.4 mGy/min 10.8 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 7.5 fps are the 25% of the 30 fps measured
values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / Receptor Dose Limited
Standard.
For "Dose to Patient" and "Variation of factors" data with country settings Italy, Spain, Portugal and Belgium,
please refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding
Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient / IQ Standard.
Loading factors automatically selected 69.2 kVp, 63.5 mA avg (30 fps), 68.8 kVp, 125.5 mA avg (30 fps),
5.0 ms 7.0 ms
Dose rate at International Standardization 67.7 x 1.4 µGy/frame 132.4 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors automatically 79.0 kVp, 40.1 mA avg, 3.5 ms PW 75.0 kVp, 97.0 mA avg, 7.0 ms PW
selected
Dose rate at US regulation con- 102 x 0.73 µGy/frame 220 x 0.73 µGy/frame
ditions (21 Code Of Federal
Regulation)
Dose rate at International Stan- 102 x 1.6 µGy/frame 220 x 1.6 µGy/frame
dardization conditions (IEC
60601-1)
Field Of View 20 cm 17 cm 15 cm 12 cm
(nominal)
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or 108 cm SID), during a 3D/3DCT spin, when the gantry
is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4:The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
Loading factors automatically 70.4 kVp, 10.1 mA avg 73.0 kVp, 9.8 mA avg 70.0 kVp, 11.5 mA avg
selected
Loading factors automatically selected 81.0 kVp, 4.2 mA avg 80.0 kVp, 10.9 mA avg
Dose rate at International Standardization 3.4 x 1.4 mGy/min 8.2 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automatically 74.4 kVp, 12.0 mA avg 72.7 kVp, 18.7 mA avg 75.1 kVp, 15.6 mA avg
selected
Loading factors automatically selected 83.0 kVp, 3.4 mA avg 80.0 kVp, 5.1 mA avg
Dose rate at International Standardization 3.0 x 1.4 mGy/min 7.7 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 15 fps are the 50% of the 30 fps measured
values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically 73.1 kVp, 9.4 mA avg 73.2 kVp, 5.8 mA avg 72.0 kVp, 6.1 mA avg
selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 3.7 mA avg 80.0 kVp, 4.7 mA avg
Dose rate at International Standardization 2.8 x 1.4 mGy/min 6.9 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro Low Frame-Rate Dose Reduction Strategy selection
(Balanced IQ/Dose). If Max Dose Reduction is used, the dose rates in 7.5 fps are the 25% of the 30 fps measured
values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Frame- 30 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 87.6 mGy/min < 194.7 mGy/min
Rate (10 R/min) (10 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 87.6 mGy/min < 194.7 mGy/min
(10 R/min) (10 R/min)
15 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
7.5 fps Normal < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
Low < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
3.75 fps Normal < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Loading factors automatically selected 67.5 kVp, 59.8 mA avg (30 fps), 66.7 kVp, 118.6 mA avg (30 fps),
5.0 ms 7.0 ms
Dose rate at International Standardization 37.0 x 1.4 µGy/frame 71.1 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors automatically 69.0 kVp, 31.9 mA avg, 3.3 ms PW 67.0 kVp, 60.2 mA avg, 6.2 ms PW
selected
Field Of View 20 cm 17 cm 15 cm 12 cm
(nominal)
Vertical 1 meter
Lateral 1 meter
X-ray should be disabled whenever they are not necessary to prevent the possibility of
irradiation being emitted through the inadvertent actuation of an irradiation switch.
Press on the X-ray disable button on the touchscreen or the digital screen (monitoring bar) to disable/re-enable
X-ray.
The button is lit when X-ray are disabled (refer to Innova Digital / Innova Digital Screen / Monitoring Bar / X-ray
Disable Button).
In case of obvious X-ray emission (audible signal and/or red light signal on) without image on the live display, the
user shall perform the following actions:
• First, to release the footswitch.
• Second, if X-ray are still emitted, to depress the "RESET" button on the Innova console until the system resets.
• Third, if X-ray are still emitted, to switch off the Power Distribution Box (PDB) by turning to OFF the RED PDB
switcher on the system cabinet, in the technical area.
4-10 Dosimetric Calibration
To maintain the calibrations of all dosimetric indications, the "mR/mAs" calibration procedure shall be strictly made
following the Periodic Maintenance (PM) interval for that calibration (6 months). (Refer to the Logbook).
4-11 Application Mode and Service Mode are not exclusive
NEVER ACQUIRE IMAGES WITH A PATIENT ON THE TABLE, IF THE REFERENCE DISPLAY
LISTS THE FOLLOWING:
"CALIBRATION" IN PLACE OF "PROTOCOL NAME"
"SERVICE" IN PLACE OF "EXAM TYPE"
THAT MEANS THAT THE SYSTEM IS NOT PROPERLY SET FOR APPLICATION AND THOSE ACQUISITION
PARAMETERS MIGHT BE INADEQUATE.
4-12 Leakage Radiation
The maximum operating factors (leakage technique factors) of the Innova IGS 520 are: Fluoro, 120 kV, 16 mA, SID
max.
In these conditions, the leakage radiation of the radiation source assembly measured is: 0.45 mGy (52
milliroentgens) in 1 hour (worst case 0.65mGy (75 milliroentgens)) at one meter in any direction from the diagnostic
source assembly composed of:
• X-ray Tube Performix 160A.
• Beam Limiting Device (Card Collimator AMP).
4-13 Spectral Filtration
Spectral filtration is added filtration that is inserted between the X-ray tube source and the patient in order to
harden the beam and reduce the soft radiation. The harder the beam, the better the penetration of the anatomy
allowing the X-ray photons to exit the patient and reach the digital detector. The spectral filters also absorb the
softer radiation so that it never reaches the patient and keeps the Air Kerma radiation and absorbed dose to a
minimum.
In addition, the beam limiting device features additional filters of: 0.1, 0.2, 0.3, 0.6 and 0.9 mm Cu which
corresponds to 6.5, 9.6, 12.6, 22 and 30 mm Alu equivalent at 70 kV.
The value of the additional spectral filtration is permanently displayed in the Fluoro and Record window located at
the bottom of the Innova Digital Screen.
The additional spectral filtration value is part of the Fluoro and Record techniques.
• Innova 3D (3D) mode / Innova CT (3DCT) mode: primary use for 3D reconstruction and display of any region of
interest.
• Dynamic Record mode: primary use for general cardiac procedures.
• InnovaChaseTM mode: primary use for angiography procedures not requiring subtraction.
• Bolus mode: primary use for bolus chasing procedures.
1 Record parameters display: record mode, detail level, frame rate, acquisition duration and X-ray/inject delay
Record technique parameters display: kV, mA (peak for DSA, Chase and Bolus, otherwise: average), ms
2 Fluoro parameters display: fluoro mode, detail level and frame rate
Fluoro technique parameters display: kV, mA
Record Screen
Fluoro Screen
Note 5: 3D acquisition can be used in unsubtracted and subtracted modes. In these cases the same loading
factors are used, that results in identical dose/ frame values. For this reason detailed data are provided only for
unsubtracted 3D. Since subtracted 3D requires 2 consecutive spins, thus the integrated dose in sub 3D is the
double of unsubtracted 3D acquisition.
Note 6: Unsubtracted fluoroscopy in 3.75 fps uses the same settings per frame as in 7.5 fps. This results in
identical kVp, pulse width, and spectral filters, while the time averaged mA and dose rate in 3.75 fps fluoroscopy
are half of the corresponding 7.5 fps values. Because of this direct relationship separate tables for 3.75 fps are not
provided.
Note 7: Loading factors and dose rate tables for High Contrast Fluoroscopy (HCF) option are provided for Normal
detail only, since this option is active in Normal fluoro detail only.
5-2-1 Skin Dose Levels
The skin dose levels in normal use may be high enough to cause deterministic effects. The availability of the
different settings (table below) has a considerable effect on the radiation quality, the delivered dose rate and the
image quality.
5-2-2 Available Settings
Frame rate 30 fps, 15 30 fps, 15 7.5 fps, 30 fps, 15 5 fps 1 fps, 2 fps 30 fps 30 fps
fps, 7.5 fps, fps, 7.5 fps 3.75 fps, fps
3.75 fps 1.875 fps,
1.0 fps, 0.5
fps
Selectable Without or Without or Without or Without or Without or 0.1 mm Cu Without or Without or
added fil- 0.1 mm Cu, 0.1 mm Cu,
0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu,
ters auto- 0.2 mm Cu, 0.2 mm Cu,
0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu,
matically or 0.3 mm or 0.3 mm
or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm
applied Cu Cu
Cu Cu Cu Cu Cu
Selectable Not applicable
added fil-
ters manu-
ally applied
Geometry
Source to 89 to 119 cm +/- 1 cm
image dis-
tance (SID)
Modes of Low, Nor- Low, Nor- Low, Nor- Low, Nor- Low, Normal Low, Normal Low, Normal Low, Normal
operation mal mal mal mal
Default de- Normal/LowNormal/LowNormal/LowNormal/Low Normal/Low Normal/Low Normal/Low Normal/Low
tail in Adult/
Pediatric
mode
Loading Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor-
factors mal: mal: mal: mal: mal: mal: mal: mal:
range after 60 - 120 kV 60 - 120 kV 50 - 125 kV 60 - 120 kV 60 - 120 kV 50 - 125 kV 60 - 120 kV 60 - 120 kV
selection of
mA: mA: Normal mA: Normal mA: mA: 100 - Normal mA: Normal mA: Normal mA:
the mode
Min: 0.0075 Min: 0.015 41.7 - 1000 1 - 866 866 41.7 - 866 1 - 786 100 - 866
mA Aver- mA Aver-
Low mA: Low mA: Low mA: Low mA: Low mA:
age @ 3.75 age @ 7.5
41.7 - 500 1 - 786 41.7 - 500 1 - 786 100 - 786
fps, fps, 2 ms
2 ms pw pw
Max: 33.6 Max: 25 mA
mA Aver- Average @
age @ 30 30 fps, 8
fps, 8 ms ms pw
pw
Highest ref- See tables See tables No dose No dose No dose lim- No dose lim- No dose lim- No dose lim-
erence at each Au- at each Au- limits ap- limits ap- its applied: its applied: its applied: its applied:
dose (IRP = to Exposure to Exposure plied: plied:
Normal set- Normal set- Normal set- Normal set-
Interven- Preference Preference
Normal set- Normal set- ting gives ting gives ting gives ting gives
tional Ref-
ting gives ting gives approx. 2 approx. 2 approx. 2 approx. 2
erence
approx. 2 approx. 2 times dose times dose times dose times dose
Point)
times dose times dose than Low than Low than Low than Low
than Low than Low setting setting setting setting
setting setting
Typical set of values in cardiac procedures
Diagnostic Normal Normal Normal Normal Normal Normal Normal Normal
Note: For 5 R receptor dose limited countries, refer to Safety and Regulatory Instructions for Use / Protection
regarding Ionizing Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited Standard:
Countries limited to 5 R/min in Fluoroscopy.
5-2-3 Radiation Data
Procedure for measuring dose rate.
Refer to the section: Generic Safety and Regulatory instructions for use for vascular systems.
WHEN MEASURING DOSE, ALWAYS DO A FLUORO RUN PRIOR TO THE RUN IN STEP1
(ACQUIRING TECHNIQUE FACTORS), OTHERWISE SPECTRAL FILTER SELECTION, AND
THUS THE MEASURED DOSE MAY NOT BE REPRODUCIBLE.
5-2-4 IQ Standard
Dose rate with 20 cm PMMA Configuration and Operating Settings
Selectable In Low: In Low: In Low: 0.1 In Low: 0.2 In Low: 0.1 In Low: 0.1 In Low: 0 In Low: 0.3
added filters 0.3 mm Cu 0.3 mm Cu mm Cu mm Cu mm Cu mm Cu mm Cu mm Cu
automatical- @ 30 fps @ 30 fps
In Normal: In Normal: In Normal: In Normal: In Normal: 0 In Normal:
ly applied 0.3 mm Cu 0.3 mm Cu
0.1 mm Cu 0.2 mm Cu 0.1 mm Cu 0.1 mm Cu mm Cu 0.3 mm Cu
@ 15 fps @ 15 fps
0.3 mm Cu 0.3 mm Cu
@ 7.5 fps @ 7.5 fps
In Normal: In Normal:
0.3 mm Cu 0.3 mm Cu
@ 30 fps @ 30 fps
0.3 mm Cu 0.3 mm Cu
@ 15 fps @ 15 fps
0.3 mm Cu 0.3 mm Cu
@ 7.5 fps @ 7.5 fps
In Normal
(HCF):
0.3 mm Cu
@ 30 fps
0.3 mm Cu
@ 15 fps
0.3 mm Cu
@ 7.5 fps
Geometry
Source to 102.5 cm Max SID 108 cm, Max
image dis- SID
tance (SID)
Distance of 67.5 cm 72 cm
focal spot to
measuring
detector
X-ray field Square of: 174 cm² Square of: Square of: Square of:
size at the 390 cm² 145 cm² 178 cm²
measuring
detector
Positioning Distance between top of phantom and anticollision plate (airgap): 17 cm Phantom centered in
of phantom isocenter
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or at 108 cm SID), during a 3D/3DCT spin, when the
gantry is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited
Standard.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatically 82.0 kVp, 5.0 mA avg 78.0 kVp, 14.2 mA avg 79.0 kVp, 13.7 mA avg
selected
Dose rate at International Stan- 10.3 x 1.4 mGy/min 24.4 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Loading factors automatically selected 82.0 kVp, 5.1 mA avg 80.0 kVp, 13.3 mA avg
Dose rate at International Standardization 10.7 x 1.4 mGy/min 25.8 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automatically se- 87.0 kVp, 2.9 mA avg 80.0 kVp, 9.4 mA avg 80.0 kVp, 9.9 mA avg
lected
Dose rate at International Stan- 7.3 x 1.4 mGy/min 18.7 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Loading factors automatically selected 80.8 kVp, 4.0 mA avg 80.0 kVp, 9.6 mA avg
Dose rate at International Standardization 8.1 x 1.4 mGy/min 18.7 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
Loading factors automatically 81.0 kVp, 2.6 mA avg 82.4 kVp, 6.3 mA avg 81.1 kVp, 6.7 mA avg
selected
Dose rate at International Stan- 5.1 x 1.4 mGy/min 13.0 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 2.7 mA avg 80.0 kVp, 6.4 mA avg
Dose rate at International Standardization 5.3 x 1.4 mGy/min 12.3 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically selected 68.6 kVp, 68.2 mA avg (30 fps), 68.0 kVp, 132.2 mA avg (30 fps),
5.0 ms 7.0 ms
Dose rate at International Standardiza- 53.3 x 1.4 µGy/frame 102.2 x 1.4 µGy/frame
tion conditions (IEC 60601-1)
Loading factors automatically selected 83.3 kVp, 312.4 mA peak, 45.0 90.5 kVp, 298.4 mA peak, 69.1 ms
ms PW PW
Dose rate at International Standardiza- 0.93 x 1.4 mGy/frame 1.70 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 84.2 kVp, 333.2 mA peak, 41.1 93.8 kVp, 299.9 mA peak, 50.0 ms
ms PW PW
Dose rate at International Standardiza- 0.93 x 1.4 mGy/frame 1.33 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors automatically selected 76.9 kVp, 14.7 mA avg, 7.0 ms 76.7 kVp, 29.7 mA avg, 7.0 ms PW
PW
Dose rate at International Standardization 160 x 1.4 µGy/frame 321 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors auto- 110.1 kVp, 161.3 mA 112.8 kVp, 296.6 mA 95.5 kVp, 473.4 mA 98.0 kVp, 686.6 mA
matically selected peak, 16.9 ms PW peak, 17.2 ms PW peak, 10.4 ms PW peak, 12.7 ms PW
Dose rate at Interna- 0.35 x 1.4 mGy/ 0.600 x 1.4 mGy/ 0.34 x 1.4 mGy/ 0.71 x 1.4 mGy/
tional Standardiza- frame frame frame frame
tion conditions (IEC
60601-1)
Loading factors automatically 76.0 kVp, 34.6 mA avg, 6.5 ms PW 74.0 kVp, 77.0 mA avg, 7.0 ms PW
selected
Loading factors automatically 88.1 kVp, 43.5 mA avg, 7.0 ms PW 87.1 kVp, 92.9 mA avg, 7.0 ms PW
selected
Loading factors automatically 86.8 kVp, 36.2 mA avg, 7.0 ms PW 86.2 kVp, 75.8 mA avg, 7.0 ms PW
selected
Field Of View 30 cm 20 cm 16 cm 12 cm
(nominal)
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or 108 cm SID), during a 3D/3DCT spin, when the gantry
is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatical- 88.7 kVp, 1.7 mA avg 80.6 kVp, 6.3 mA avg 80.7 kVp, 6.2 mA avg
ly selected
Loading factors automatically selected 88.7 kVp, 1.8 mA avg 80.0 kVp, 6.3 mA avg
Dose rate at International Standardiza- 5.2 x 1.4 mGy/min 12.2 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automatical- 85.6 kVp, 1.4 mA avg 77.4 kVp, 5.0 mA avg 80.4 kVp, 4.7 mA avg
ly selected
Loading factors automatically selected 86.0 kVp, 1.3 mA avg 80.9 kVp, 4.5 mA avg
Dose rate at International Standardiza- 3.5 x 1.4 mGy/min 9.1 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automati- 83.0 kVp, 1.1 mA avg 77.0 kVp, 3.5 mA avg 77.6 kVp, 3.6 mA avg
cally selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 84.5 kVp, 1.0 mA avg 80.0 kVp, 3.1 mA avg
Dose rate at International Standardiza- 2.4 x 1.4 mGy/min 6.1 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically select- 71.0 kVp, 40.7 mA avg (30 fps), 67.9 kVp, 107.3 mA avg (30 fps), 7.0
ed 5.0 ms ms
Dose rate at International Standard- 24.4 x 1.4 µGy/frame 55.0 x 1.4 µGy/frame
ization conditions (IEC 60601-1)
Loading factors automatically selected 83.2 kVp, 163.1 mA peak, 44.8 83.4 kVp, 314.3 mA peak, 45.1 ms
ms PW PW
Dose rate at International Standardiza- 0.49 x 1.4 mGy/frame 0.95 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 83.5 kVp, 160.0 mA peak, 45.2 83.5 kVp, 314.5 mA peak, 45.2 ms
ms PW PW
Dose rate at International Standardiza- 0.49 x 1.4 mGy/frame 0.95 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors auto- 103.6 kVp, 166.3 mA 104.9 kVp, 284.8 mA 91.5 kVp, 485.6 mA 93.3 kVp, 714.0 mA
matically selected peak, 14.6 ms PW peak, 16.0 ms PW peak, 8.8 ms PW peak, 10.8 ms PW
Dose rate at Interna- 0.35 x 1.4 mGy/ 0.58 x 1.4 mGy/ 0.33 x 1.4 mGy/ 0.73 x 1.4 mGy/
tional Standardiza- frame frame frame frame
tion conditions (IEC
60601-1)
Loading factors automatically 81.8 kVp, 24.6 mA avg, 4.2 ms PW 80.0 kVp, 46.4 mA avg, 7.0 ms PW
selected
Loading factors automatically 86.9 kVp, 28.7 mA avg, 5.2 ms PW 88.2 kVp, 43.4 mA avg, 7.0 ms PW
selected
Loading factors automatically 85.9 kVp, 23.5 mA avg, 5.1 ms PW 86.9 kVp, 35.8 mA avg, 7.0 ms PW
selected
Field Of View 30 cm 20 cm 16 cm 12 cm
(nominal)
5-2-7 IQ Plus
For DSA, Bolus, InnovaChaseTM and 3DCT Normal/Low, all frame rates, please refer to measurements in Dose to
Patient for IQ Standard.
Dose rate with 20 cm PMMA Configuration and Operating Settings
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or at 108 cm SID), during a 3D/3DCT spin, when the
gantry is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
Loading factors automati- 70.6 kVp, 9.4 mA avg 68.6 kVp, 27.4 mA avg 66.0 kVp, 23.0 mA avg
cally selected
Loading factors automatically selected 81.7 kVp, 5.5 mA avg 80.0 kVp, 14.0 mA avg
Dose rate at International Standardiza- 11.6 x 1.4 mGy/min 27.2 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automati- 64.8 kVp, 9.4 mA avg 66.9 kVp, 22.2 mA avg 67.3 kVp, 15.2 mA avg
cally selected
Loading factors automatically selected 80.8 kVp, 3.9 mA avg 80.0 kVp, 9.8 mA avg
Dose rate at International Standardiza- 7.8 x 1.4 mGy/min 18.8 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatical- 69.3 kVp, 4.7 mA avg 65.2 kVp, 11.2 mA avg 67.8 kVp, 6.1 mA avg
ly selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 2.7 mA avg 80.0 kVp, 6.6 mA avg
Dose rate at International Standardiza- 5.2 x 1.4 mGy/min 12.8 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Frame- 30 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 87.6 mGy/min < 194.7 mGy/min
Rate (10 R/min) (10 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 87.6mGy/min < 194.7 mGy/min
(10 R/min) (10 R/min)
15 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
7.5 fps Normal < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
Low < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
3.75 fps Normal < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Loading factors automatically se- 73.5 kVp, 77.1 mA avg (30 fps), 5.0 72.1 kVp, 151.9 mA avg (30 fps), 7.0
lected ms ms
Dose rate at International Stan- 75.2 x 1.4 µGy/frame 140.7 x 1.4 µGy/frame
dardization conditions (IEC
60601-1)
Loading factors automatically 80.0 kVp, 46.4 mA avg, 7.0 ms PW 74.0 kVp, 77.1 mA avg, 7.0 ms PW
selected
Field Of View 30 cm 20 cm 16 cm 12 cm
(nominal)
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID (or at 108 cm SID), during a 3D/3DCT spin, when the
gantry is in PA position. 2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited
Standard.
For "Dose to Patient" and "Variation of factors" data with country settings Japan, for the Low Dose preference,
please refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding
Ionizing Radiations Hazards for Innova IGS 530 / Dose to Patient / Receptor Dose Limited Standard. The only
exception is Max Dose Reduction 7.5 fps Low Dose preference which is already limited to 21.9 mGy/min when
using "Receptor Dose Limited Plus".
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatical- 61.0 kVp, 13.0 mA avg 65.3 kVp, 24.8 mA avg 64.3 kVp, 18.7 mA avg
ly selected
Loading factors automatically selected 83.0 kVp, 3.8 mA avg 80.0 kVp, 10.9 mA avg
Dose rate at International Standardiza- 8.5 x 1.4 mGy/min 20.9 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automatical- 61.0 kVp, 9.8 mA avg 63.9 kVp, 19.0 mA avg 63.7 kVp, 13.2 mA avg
ly selected
Loading factors automatically selected 83.0 kVp, 2.9 mA avg 81.3 kVp, 7.2 mA avg
Dose rate at International Standardiza- 6.5 x 1.4 mGy/min 14.7 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatical- 61.8 kVp, 6.2 mA avg 67.7 kVp, 10.1 mA avg 66.5 kVp, 7.8 mA avg
ly selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 2.2 mA avg 80.0 kVp, 5.1 mA avg
Dose rate at International Standardiza- 4.3 x 1.4 mGy/min 9.8 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically se- 68.2 kVp, 69.6 mA avg (30 fps), 5.0 67.9 kVp, 139.3 mA avg (30 fps), 7.0
lected ms ms
Field Of View 30 cm 20 cm 16 cm 12 cm
(nominal)
Vertical 1 meter
Lateral 1 meter
X-ray should be disabled whenever they are not necessary to prevent the possibility of
irradiation being emitted through the inadvertent actuation of an irradiation switch.
Press on the X-ray disable button on the touchscreen or the digital screen (monitoring bar) to disable/re-enable
X-ray.
The button is lit when X-ray are disabled (refer to Innova Digital / Innova Digital Screen / Monitoring Bar / X-ray
Disable Button).
In case of obvious X-ray emission (audible signal and/or red light signal on) without image on the live display, the
user shall perform the following actions:
• First, to release the footswitch.
• Second, if X-ray are still emitted, to depress the "RESET" button on the Innova console until the system resets.
• Third, if X-ray are still emitted, to switch off the Power Distribution Box (PDB) by turning to OFF the RED PDB
switcher on the system cabinet, in the technical area.
5-9 Dosimetric Calibration
To maintain the calibrations of all dosimetric indications, the "mR/mAs" calibration procedure shall be strictly made
following the Periodic Maintenance (PM) interval for that calibration (6 months). (Refer to the Logbook).
5-10 Application Mode and Service Mode are not exclusive
NEVER ACQUIRE IMAGES WITH A PATIENT ON THE TABLE, IF THE REFERENCE DISPLAY
LISTS THE FOLLOWING:
"CALIBRATION" IN PLACE OF "PROTOCOL NAME"
"SERVICE" IN PLACE OF "EXAM TYPE"
THAT MEANS THAT THE SYSTEM IS NOT PROPERLY SET FOR APPLICATION AND THOSE ACQUISITION
PARAMETERS MIGHT BE INADEQUATE.
5-11 Leakage Radiation
The maximum operating factors (leakage technique factors) of the Innova IGS 530 are: Fluoro, 120 kV, 16 mA, SID
max.
In these conditions, the leakage radiation of the radiation source assembly measured is: 0.45 mGy (52
milliroentgens) in 1 hour at one meter in any direction from the diagnostic source assembly composed of:
• X-ray Tube Performix 160A.
• Beam Limiting Device (Collimator) AF DSA 01.
5-12 Spectral Filtration
Spectral filtration is added filtration that is inserted between the X-ray tube source and the patient in order to
harden the beam and reduce the soft radiation. The harder the beam, the better the penetration of the anatomy
allowing the X-ray photons to exit the patient and reach the digital detector. The spectral filters also absorb the
softer radiation so that it never reaches the patient and keeps the Air Kerma radiation and absorbed dose to a
minimum.
In addition, the beam limiting device features three additional filters of: 0.1, 0.2 and 0.3 mm Cu which corresponds
to 6.5, 9.6, 12.6 mm Alu equivalent at 70 kV.
The value of the additional spectral filtration is permanently displayed in the Fluoro and Record window located at
the bottom of the Innova Digital Screen.
The additional spectral filtration value is part of the Fluoro and Record techniques.
interest.
• Dynamic Record mode: primary use for general cardiac procedures.
• InnovaChaseTM mode: primary use for angiography procedures not requiring subtraction.
• Bolus mode: primary use for bolus chasing procedures.
1 Record parameters display: record mode, detail level, frame rate, acquisition duration and X-ray/inject delay
Record technique parameters display: kV, mA (peak for DSA, Chase and Bolus, otherwise: average), ms
2 Fluoro parameters display: fluoro mode, detail level and frame rate
Fluoro technique parameters display: kV, mA
Record Screen
Fluoro Screen
Note 5: 3D acquisition can be used in unsubtracted and subtracted modes. In these cases the same loading
factors are used, that results in identical dose/ frame values. For this reason detailed data are provided only for
unsubtracted 3D. Since subtracted 3D requires 2 consecutive spins, thus the integrated dose in sub 3D is the
double of unsubtracted 3D acquisition.
Note 6: Unsubtracted fluoroscopy in 3.75 fps uses the same settings per frame as in 7.5 fps. This results in
identical kVp, pulse width, and spectral filters, while the time averaged mA and dose rate in 3.75 fps fluoroscopy
are half of the corresponding 7.5 fps values. Because of this direct relationship separate tables for 3.75 fps are not
provided.
Note 7: Loading factors and dose rate tables for High Contrast Fluoroscopy (HCF) option are provided for Normal
detail only, since this option is active in Normal fluoro detail only.
6-2-1 Skin Dose Levels
The skin dose levels in normal use may be high enough to cause deterministic effects. The availability of the
different settings (table below) has a considerable effect on the radiation quality, the delivered dose rate and the
image quality.
6-2-2 Available Settings
Frame rate 30 fps, 15 30 fps, 15 7.5 fps, 30 fps, 15 5 fps 1 fps, 2 fps 30 fps 30 fps
fps, 7.5 fps, fps, 7.5 fps 3.75 fps, fps
3.75 fps 1.875 fps,
1.0 fps, 0.5
fps
Selectable Without or Without or Without or Without or Without or 0.1 mm Cu Without or Without or
added fil- 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu, 0.1 mm Cu,
ters auto- 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu, 0.2 mm Cu,
matically or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm or 0.3 mm
applied Cu Cu Cu Cu Cu Cu Cu
Selectable Not applicable
added fil-
ters manu-
ally applied
Geometry
Source to 95 to 119 cm +/- 1 cm
image dis-
tance (SID)
Modes of Low, Nor- Low, Nor- Low, Nor- Low, Nor- Low, Normal Low, Normal Low, Normal Low, Normal
operation mal mal mal mal
Default de- Normal/LowNormal/LowNormal/LowNormal/Low Normal/Low Normal/Low Normal/Low Normal/Low
tail in Adult/
Pediatric
mode
Loading Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor- Low or Nor-
factors mal: mal: mal: mal: mal: mal: mal: mal:
range after 60 - 120 kV 60 - 120 kV 50 - 125 kV 60 - 120 kV 60 - 120 kV 50 - 125 kV 60 - 120 kV 60 - 120 kV
selection of
mA: mA: Normal mA: Normal mA: mA: 100 - Normal mA: Normal mA: Normal mA:
the mode
Min: Min: 41.7 - 1000 1 - 866 866 41.7 - 866 1 - 786 100 - 866
0.0075 mA 0.015 mA
Low mA: Low mA: Low mA: Low mA: Low mA:
Average @ Average @
41.7 - 500 1 - 786 41.7 - 500 1 - 786 100 - 786
3.75 fps, 7.5 fps, 2
2 ms pw ms pw
Max: Max:
33.6 mA 25 mA Av-
Average @ erage @ 30
30 fps, 8 fps, 8 ms
ms pw pw
Highest ref- See tables See tables No dose No dose No dose lim- No dose lim- No dose lim- No dose lim-
erence at each Au- at each Au- limits ap- limits ap- its applied: its applied: its applied: its applied:
dose (IRP = to Exposure to Exposure plied: plied:
Normal set- Normal set- Normal set- Normal set-
Interven- Preference Preference
Normal set- Normal set- ting gives ting gives ting gives ting gives
tional Ref-
ting gives ting gives approx. 2 approx. 2 approx. 2 approx. 2
erence
approx. 2 approx. 2 times dose times dose times dose times dose
Point)
times dose times dose than Low than Low than Low than Low
than Low than Low setting setting setting setting
setting setting
Typical set of values in cardiac procedures
Diagnostic Normal Normal Normal Normal Normal Normal Normal Normal
Note: For 5 R receptor dose limited countries, refer to Safety and Regulatory Instructions for Use / Protection
regarding Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited Standard:
Countries limited to 5 R/min in Fluoroscopy.
6-2-3 Radiation Data
Procedure for measuring dose rate.
Refer to the section: Generic Safety and Regulatory instructions for use for vascular systems.
WHEN MEASURING DOSE, ALWAYS DO A FLUORO RUN PRIOR TO THE RUN IN STEP1
(ACQUIRING TECHNIQUE FACTORS), OTHERWISE SPECTRAL FILTER SELECTION, AND
THUS THE MEASURED DOSE MAY NOT BE REPRODUCIBLE.
6-2-4 IQ Standard
Dose rate with 20 cm PMMA Configuration and Operating Settings
Selectable In Low: In Low: In Low: 0.1 In Low: 0.2 In Low: 0.1 In Low: 0.1 In Low: 0 In Low: 0.3
added filters 0.3 mm Cu 0.3 mm Cu mm Cu mm Cu mm Cu mm Cu mm Cu mm Cu
automatical- @ 30 fps @ 30 fps
In Normal: In Normal: In Normal: In Normal: In Normal: 0 In Normal:
ly applied 0.3 mm Cu 0.3 mm Cu
0.1 mm Cu 0.2 mm Cu 0.1 mm Cu 0.1 mm Cu mm Cu 0.3 mm Cu
@ 15 fps @ 15 fps
0.3 mm Cu 0.3 mm Cu
@ 7.5 fps @ 7.5 fps
In Normal: In Normal:
0.3 mm Cu 0.3 mm Cu
@ 30 fps @ 30 fps
0.3 mm Cu 0.3 mm Cu
@ 15 fps @ 15 fps
0.3 mm Cu 0.3 mm Cu
@ 7.5 fps @ 7.5 fps
In Normal
(HCF):
0.3 mm Cu
@ 30 fps
0.3 mm Cu
@ 15 fps
0.3 mm Cu
@ 7.5 fps
Geometry
Source to 102.5 cm Max SID
image dis-
tance (SID)
Distance of 67.5 cm 72 cm
focal spot to
measuring
detector
X-ray field Square of: 174 cm² Square of: Square of: Square of:
size at the 694 cm² 145 cm² 178 cm²
measuring
detector
Positioning Distance between top of phantom and anticollision plate (airgap): 17 cm Phantom centered in
of phantom isocenter
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID, during a 3D/3DCT spin, when the gantry is in PA position.
2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited
Standard.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatically se- 82.5 kVp, 5.5 mA avg 78.0 kVp, 16.3 mA avg 79.0 kVp, 15.6 mA avg
lected
Dose rate at International Stan- 10.2 x 1.4 mGy/min 24.9 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Loading factors automatically selected 82.6 kVp, 5.5 mA avg 80.0 kVp, 15.3 mA avg
Dose rate at International Standardization 10.4 x 1.4 mGy/min 25.6 x 1.4 mGy/min
conditions (IEC 60601-1)
Loading factors automatically se- 87.0 kVp, 3.2 mA avg 80.0 kVp, 10.7 mA avg 80.6 kVp, 11.3 mA avg
lected
Dose rate at International Stan- 7.3 x 1.4 mGy/min 19.3 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Loading factors automatically selected 83.3 kVp, 4.1 mA avg 80.0 kVp, 11.4 mA avg
Dose rate at International Standardization 7.9 x 1.4 mGy/min 19.1 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
Loading factors automatically se- 81.0 kVp, 2.9 mA avg 82.5 kVp, 6.9 mA avg 82.3 kVp, 7.2 mA avg
lected
Dose rate at International Stan- 5.2 x 1.4 mGy/min 13.2 x 1.4 mGy/min
dardization conditions (IEC
60601-1)
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 3.1 mA avg 80.0 kVp, 7.6 mA avg
Dose rate at International Standardization 5.3 x 1.4 mGy/min 12.7 x 1.4 mGy/min
conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically selected 70.3 kVp, 66.5 mA avg (30 fps), 69.7 kVp, 126.0 mA avg (30 fps),
5.0 ms 7.0 ms
Loading factors automatically selected 85.2 kVp, 308.8 mA peak, 47.8 93.2 kVp, 290.3 mA peak, 67.8 ms
ms PW PW
Dose rate at International Standardiza- 0.96 x 1.4 mGy/frame 1.62 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 85.6 kVp, 328.1 mA peak, 43.7 98.1 kVp, 287.5 mA peak, 50.0 ms
ms PW PW
Dose rate at International Standardiza- 0.95 x 1.4 mGy/frame 1.33 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors automatically selected 80.0 kVp, 13.8 mA avg, 7.0 ms 80.0 kVp, 27.2 mA avg, 7.0 ms PW
PW
Dose rate at International Standardization 155 x 1.4 µGy/frame 303 x 1.4 µGy/frame
conditions (IEC 60601-1)
Loading factors auto- 105.3 kVp, 142.8 mA 106.8 kVp, 265.7 mA 89.8 kVp, 386.6 mA 92.4 kVp, 656.6 mA
matically selected peak, 14.6 ms PW peak, 14.9 ms PW peak, 8.8 ms PW peak, 9.5 ms PW
Dose rate at Interna- 0.22 x 1.4 mGy/ 0.44 x 1.4 mGy/ 0.25 x 1.4 mGy/ 0.49 x 1.4 mGy/
tional Standardiza- frame frame frame frame
tion conditions (IEC
60601-1)
Loading factors automatically 76.0 kVp, 38.9 mA avg, 7.0 ms PW 74.0 kVp, 84.0 mA avg, 7.0 ms PW
selected
Loading factors automatically 89.6 kVp, 43.4 mA avg, 7.0 ms PW 88.6 kVp, 89.6 mA avg, 7.0 ms PW
selected
Field Of View 40 cm 32 cm 20 cm 16 cm
(nominal)
Frame rate 30 fps, 15 fps, 30 fps, 15 fps, 7.5 fps, 30 fps, 15 1 fps, 2 fps 30 fps 30 fps
7.5 fps, 3.75 7.5 fps 3.75 fps, fps
fps 1.875 fps,
1.0 fps, 0.5
fps
Selectable added In Low 30 fps: In Low 30 fps: In Low: 0.1 In Low: 0.3 In Low: 0.1 In Low: 0.3 In Low: 0.3
filters automati- 0.3 mm Cu 0.3 mm Cu mm Cu mm Cu mm Cu mm Cu mm Cu
cally applied In Normal 30 In Normal 30
In Normal: In Normal: In Normal: In Normal: 0.1 In Normal:
fps: fps:
0.1 mm Cu 0.3 mm Cu 0.1 mm Cu mm Cu 0.3 mm Cu
0.3 mm Cu 0.3 mm Cu
In Normal
In Low 15 fps:
(HCF) 30 fps:
0.3 mm Cu
0.3 mm Cu
In Normal 15
In Low 15 fps: fps:
0.3 mm Cu 0.3 mm Cu
In Normal 15
In Low 7.5 fps:
fps:
0.3 mm Cu
0.3 mm Cu
In Normal 7.5
In Normal
fps:
(HCF) 15 fps:
0.3 mm Cu
0.3 mm Cu
In Low 7.5 fps:
0.3 mm Cu
In Normal 7.5
fps:
0.3 mm Cu
In Normal 7.5
fps:
0.3 mm Cu
Geometry
Source to image 102.5 cm Max SID
distance (SID)
Distance of focal 67.5 cm 72 cm
spot to measuring
detector
X-ray field size at Square of: 174 cm² Square of: Square of: 145 Square of:
the measuring de- 694 cm² cm² 178 cm²
tector
Positioning of Distance between top of phantom and anticollision plate (airgap): 17 Phantom centered in
phantom cm isocenter
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID, during a 3D/3DCT spin, when the gantry is in PA position.
2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. For all acquisition modes except 3D/3DCT, the dose values without the correction factor of
1.4 correspond to the dose rates at US regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
Loading factors automatical- 89.0 kVp, 1.9 mA avg 81.0 kVp, 7.0 mA avg 81.0 kVp, 6.9 mA avg
ly selected
Loading factors automatically selected 90.2 kVp, 1.9 mA avg 80.0 kVp, 6.9 mA avg
Dose rate at International Standardiza- 4.9 x 1.4 mGy/min 11.7 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automatical- 86.0 kVp, 1.6 mA avg 78.0 kVp, 5.6 mA avg 81.0 kVp, 5.1 mA avg
ly selected
Loading factors automatically selected 86.0 kVp, 1.5 mA avg 81.0 kVp, 5.1 mA avg
Dose rate at International Standardiza- 3.4 x 1.4 mGy/min 9.0 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Loading factors automati- 83.0 kVp, 1.3 mA avg 77.8 kVp, 3.8 mA avg 78.0 kVp, 3.9 mA avg
cally selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 87.3 kVp, 1.0 mA avg 80.0 kVp, 3.6 mA avg
Dose rate at International Standardiza- 2.3 x 1.4 mGy/min 6.0 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically select- 71.0 kVp, 45.4 mA avg (30 fps), 68.1 kVp, 125.0 mA avg (30 fps), 7.0
ed 5.0 ms ms
Dose rate at International Standard- 25.2 x 1.4 µGy/frame 59.8 x 1.4 µGy/frame
ization conditions (IEC 60601-1)
Loading factors automatically selected 85.5 kVp, 150.6 mA peak, 48.3 85.2 kVp, 311.1 mA peak, 47.8 ms
ms PW PW
Dose rate at International Standardiza- 0.49 x 1.4 mGy/frame 0.97 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Note: Data applicable to all frame rates lower than 7.5 fps.
Dose in DSA 7.5 fps (radiography) with 20 cm PMMA
Loading factors automatically selected 88.9 kVp, 143.7 mA peak, 42.2 85.7 kVp, 327.9 mA peak, 44.0 ms
ms PW PW
Dose rate at International Standardiza- 0.44 x 1.4 mGy/frame 0.95 x 1.4 mGy/frame
tion conditions (IEC 60601-1)
Loading factors auto- 100.0 kVp, 150.4 mA 100.9 kVp, 278.8 mA 88.9 kVp, 384.0 mA 90.0 kVp, 674.1 mA
matically selected peak, 12.5 ms PW peak, 12.8 ms PW peak, 8.5 ms PW peak, 8.9 ms PW
Dose rate at Interna- 0.18 x 1.4 mGy/ 0.35 x 1.4 mGy/ 0.23 x 1.4 mGy/ 0.44 x 1.4 mGy/
tional Standardiza- frame frame frame frame
tion conditions (IEC
60601-1)
Loading factors automatically 82.0 kVp, 26.0 mA avg, 5.1 ms PW 81.0 kVp, 39.7 mA avg, 7.0 ms PW
selected
Loading factors automatically 88.1 kVp, 27.3 mA avg, 5.5 ms PW 89.6 kVp, 43.4 mA avg, 7.0 ms PW
selected
Field Of View 40 cm 32 cm 20 cm 16 cm
(nominal)
6-2-7 IQ Plus
For DSA, Bolus, InnovaChaseTM and 3DCT Normal/Low, all frame rates, please refer to measurements in Dose to
Patient for IQ Standard.
Dose rate with 20 cm PMMA Configuration and Operating Settings
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID, during a 3D/3DCT spin, when the gantry is in PA position.
2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited
Standard.
For "Dose to Patient" and "Variation of factors" data with country settings Japan, for the Low Dose preference,
please refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding
Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / IQ Standard. The only exception is Max Dose
Reduction 7.5 fps Low Dose preference which is already limited to 21.9 mGy/min when using "IQ Plus".
For "Dose to Patient" and "Variation of factors" data with country settings Australia/WA, Australia/ACT, please
refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 540 / Dose to Patient / IQ Standard.
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatical- 71.0 kVp, 11.5 mA avg 71.0 kVp, 21.8 mA avg 67.2 kVp, 24.7 mA avg
ly selected
Loading factors automatically selected 83.1 kVp, 5.8 mA avg 80.0 kVp, 16.2 mA avg
Dose rate at International Standardiza- 11.2 x 1.4 mGy/min 26.9 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automatical- 65.6 kVp, 10.4 mA avg 68.3 kVp, 18.3 mA avg 70.4 kVp, 11.4 mA avg
ly selected
Loading factors automatically selected 82.5 kVp, 4.1 mA avg 80.0 kVp, 11.5 mA avg
Dose rate at International Standardiza- 7.7 x 1.4 mGy/min 19.3 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatical- 69.5 kVp, 5.7 mA avg 68.5 kVp, 8.0 mA avg 68.4 kVp, 7.7 mA avg
ly selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.0 kVp, 3.0 mA avg 80.1 kVp, 7.8 mA avg
Dose rate at International Standardiza- 5.1 x 1.4 mGy/min 13.1 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Frame- 30 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 87.6 mGy/min < 194.7 mGy/min
Rate (10 R/min) (10 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 87.6mGy/min < 194.7 mGy/min
(10 R/min) (10 R/min)
15 fps Normal < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
Low < 87.6 mGy/min < 194.7 mGy/min < 43.8 mGy/min < 97.4 mGy/min
(10 R/min) (5 R/min)
7.5 fps Normal < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
Low < 43.8 mGy/min < 97.4 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(5 R/min) (2.5 R/min)
3.75 fps Normal < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Low < 21.9 mGy/min < 48.7 mGy/min < 21.9 mGy/min < 48.7 mGy/min
(2.5 R/min) (2.5 R/min)
Loading factors automatically se- 75.6 kVp, 74.6 mA avg (30 fps), 5.1 73.8 kVp, 143.5 mA avg (30 fps), 7.0
lected ms ms
Dose rate at International Stan- 74.2 x 1.4 µGy/frame 133.2 x 1.4 µGy/frame
dardization conditions (IEC
60601-1)
Loading factors automatically 81.0 kVp, 39.8 mA avg, 7.0 ms PW 74.0 kVp, 83.9 mA avg, 7.0 ms PW
selected
Field Of View 40 cm 32 cm 20 cm 16 cm
(nominal)
Note 1: For all data in this section, measurements are done in 2-steps: 1st step (technique factors acquisition) in
automatic mode with PMMA and table in; 2nd step - Dose measurement in manual mode, repeating the technique
factors without PMMA and table.
Note 2: In 3D/3DCT the 2nd step of the measurement method is different form other (2D) modes. The 1st step
(technique factors acquisition) is done at maximum SID, during a 3D/3DCT spin, when the gantry is in PA position.
2nd step (dose measurement) is done with the dosemeter probe in isocenter on the table.
Note 3: IEC 60601-1 conditions for all dose rate data are derived by applying a geometrical correction factor using
the inverse square law. The values without the correction factor of 1.4 correspond to the dose rates at US
regulation conditions (21 Code Of Federal Regulation).
Note 4: The dose rate in normal detail is the same with or without the High Contrast Fluoroscopy (HCF) option,
although loading factors can be different.
Dose rate in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted fluoroscopy
(radioscopy) with 20 cm PMMA
For "Dose to Patient" and "Variation of factors" data with country settings Luxembourg, New Zealand, please refer
to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited Standard: countries limited to 5
R/min in Fluoroscopy.
For "Dose to Patient" and "Variation of factors" data with country settings Austria, Denmark, Germany,
Switzerland, please refer to the corresponding data in section Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited
Standard.
For "Dose to Patient" and "Variation of factors" data with country settings Japan, for the Low Dose preference,
please refer to the corresponding data in section Safety and Regulatory Instructions for Use / Protection regarding
Ionizing Radiations Hazards for Innova IGS 540 / Dose to Patient / Receptor Dose Limited Standard. The only
exception is Max Dose Reduction 7.5 fps Low Dose preference which is already limited to 21.9 mGy/min when
using "Receptor Dose Limited Plus".
Dose rate in fluoroscopy (radioscopy) at 30 fps with 20 cm PMMA
Loading factors automatical- 61.0 kVp, 14.4 mA avg 66.2 kVp, 24.8 mA avg 65.4 kVp, 18.7 mA avg
ly selected
Loading factors automatically selected 83.0 kVp, 4.3 mA avg 80.0 kVp, 12.0 mA avg
Dose rate at International Standardiza- 8.1 x 1.4 mGy/min 20.0 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Loading factors automatical- 61.0 kVp, 11.0 mA avg 63.2 kVp, 21.5 mA avg 62.9 kVp, 15.2 mA avg
ly selected
Loading factors automatically selected 83.0 kVp, 3.3 mA avg 82.0 kVp, 8.0 mA avg
Dose rate at International Standardiza- 6.3 x 1.4 mGy/min 14.6 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 15 fps are the 50% of the 30 fps measured values.
Dose rate in fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automati- 62.5 kVp, 6.6 mA avg 68.1 kVp, 10.9 mA avg 67.9 kVp, 7.8 mA avg
cally selected
Dose rate in roadmap/blended roadmap/subtracted fluoroscopy (radioscopy) at 7.5 fps with 20 cm PMMA
Loading factors automatically selected 80.7 kVp, 2.4 mA avg 80.0 kVp, 6.0 mA avg
Dose rate at International Standardiza- 4.2 x 1.4 mGy/min 10.0 x 1.4 mGy/min
tion conditions (IEC 60601-1)
Note: The dose data above correspond to the default Fluoro/Roadmap/Blended Roadmap/Subtracted Fluoroscopy
Low Frame-Rate Dose Reduction Strategy selection (Balanced IQ/Dose). If Max Dose Reduction is used, the dose
rates in 7.5 fps are the 25% of the 30 fps measured values.
Highest reference doses in fluoroscopy (radioscopy) and roadmap/blended roadmap/subtracted
fluoroscopy (radioscopy)
Loading factors automatically se- 68.4 kVp, 67.1 mA avg (30 fps), 5 ms 68.0 kVp, 134.5 mA avg (30 fps), 7
lected ms
Dose rate at International Stan- 34.7 x 1.4 µGy/frame 69.5 x 1.4 µGy/frame
dardization conditions (IEC
60601-1)
Field Of View 40 cm 32 cm 20 cm 16 cm
(nominal)
Vertical 1 meter
Lateral 1 meter
X-ray should be disabled whenever they are not necessary to prevent the possibility of
irradiation being emitted through the inadvertent actuation of an irradiation switch.
Press on the X-ray disable button on the touchscreen or the digital screen (monitoring bar) to disable/re-enable
X-ray.
The button is lit when X-ray are disabled (refer to Innova Digital / Innova Digital Screen / Monitoring Bar / X-ray
Disable Button).
In case of obvious X-ray emission (audible signal and/or red light signal on) without image on the live display, the
user shall perform the following actions:
• First, to release the footswitch.
• Second, if X-ray are still emitted, to depress the "RESET" button on the Innova console until the system resets.
• Third, if X-ray are still emitted, to switch off the Power Distribution Box (PDB) by turning to OFF the RED PDB
switcher on the system cabinet, in the technical area.
6-9 Dosimetric Calibration
To maintain the calibrations of all dosimetric indications, the "mR/mAs" calibration procedure shall be strictly made
following the Periodic Maintenance (PM) interval for that calibration (6 months). (Refer to the Logbook).
6-10 Application Mode and Service Mode are not exclusive
NEVER ACQUIRE IMAGES WITH A PATIENT ON THE TABLE, IF THE REFERENCE DISPLAY
LISTS THE FOLLOWING:
"CALIBRATION" IN PLACE OF "PROTOCOL NAME"
"SERVICE" IN PLACE OF "EXAM TYPE"
THAT MEANS THAT THE SYSTEM IS NOT PROPERLY SET FOR APPLICATION AND THOSE ACQUISITION
PARAMETERS MIGHT BE INADEQUATE.
6-11 Leakage Radiation
The maximum operating factors (leakage technique factors) of the Innova IGS 540 are: Fluoro, 120 kV, 16 mA, SID
max.
In these conditions, the leakage radiation of the radiation source assembly measured is: 0.45 mGy (52
milliroentgens) in 1 hour at one meter in any direction from the diagnostic source assembly composed of:
• X-ray Tube Performix 160A.
• Beam Limiting Device (Collimator) AF DSA 01.
6-12 Spectral Filtration
Spectral filtration is added filtration that is inserted between the X-ray tube source and the patient in order to
harden the beam and reduce the soft radiation. The harder the beam, the better the penetration of the anatomy
allowing the X-ray photons to exit the patient and reach the digital detector. The spectral filters also absorb the
softer radiation so that it never reaches the patient and keeps the Air Kerma radiation and absorbed dose to a
minimum.
In addition, the beam limiting device features three additional filters of: 0.1, 0.2 and 0.3 mm Cu which corresponds
to 6.5, 9.6, 12.6 mm Alu equivalent at 70 kV.
The value of the additional spectral filtration is permanently displayed in the Fluoro and Record window located at
the bottom of the Innova Digital Screen.
The additional spectral filtration value is part of the Fluoro and Record techniques.
THE INTERRUPTION OF THE MAIN POWER SUPPLY TO THE EQUIPMENT DURING ITS USE
COULD LEAD TO A HAZARDOUS SITUATION. THE EQUIPMENT MAIN POWER SUPPLY
MUST BE PROVIDED THROUGH AN UNINTERRUPTIBLE POWER SUPPLY (UPS) AT
FACILITY LEVEL OR USING THE FLUORO UPS OPTION, OR THROUGH A STAND-BY
GENERATOR. REFER TO SECTION APPENDIX - TECHNICAL PUBLICATIONS / SYSTEM
LINE VOLTAGE WITHOUT FLUORO UPS FOR THE CHARACTERISTICS OF THE INPUT LINE.
Functions maintained in Equip- Equipment with Fluo- Equipment with institu- Equipment with institu-
case of power supply ment ro UPS option tional stand-by genera- tional uninterruptible
main failure tor3 power supply3
Preservation of stored im- Yes Yes Yes Yes
ages
Availability of fluoroscopy No Yes1* After boot2 Yes
2
Availability of record No No After boot Yes
Availability of equipment No Yes1 After boot2 Yes
motion
Availability of equipment No Yes1 After boot2 Yes
motion for CPR
1
: Functions remain available until batteries are unloaded. At least 5 minutes of fluoroscopy are available if the
batteries are fully loaded.
2
: When the power supply is restored by the generator, the system needs to be powered-on following the procedure
available in the instruction for use.
3
: Refer to section Appendix - Technical Publications / System Line Voltage without Fluoro UPS for the
characteristics of the input line.
*
: On configurations with Large Display Monitor Option, the images display is guaranteed on the backup monitors
only.
7-4 System Recovery after a power Cut
• Overview
The Innova System provides an automatic shutdown for the power cut of more than 500 ms. An UPS
(Un-interruptible Power Supply) delivers power to the DL and the RTAC to make this shutdown clean.
• System restart
Refer to section System - How to Perform System Power-Up for the description of the procedure to restart the
system after a power cut.
7-5 Emergency Procedure for Image Recovery or System Lock Up with 19" monitors
There may still be a residual risk of:
• One system lock up every 1 000 patients.
• One X-ray exposure sequence interrupted every 10 000 records.
Handle Emergency or at risk procedures with precaution.
7-6 Emergency Procedure for Image Recovery or System Lock Up with Large Display
monitor
There may still be a residual risk of:
• One system lock up every 1 000 patients.
• One X-ray exposure sequence interrupted every 10 000 records.
• The Large Display option may experience power cut separately from the Interventional system.
Handle Emergency or at risk procedures with precaution.
7-7 Latency (Image Display Delay), Polymerization Time and Acquisition mode use
LATENCY OR IMAGE DISPLAY DELAY IS DEFINED AS THE TIME LAG BETWEEN THE
OPERATOR ACTION AND THE DISPLAY OF THAT ACTION ON THE MONITOR. THIS LAG
VARIES DEPENDING ON THE ACQUISITION FRAME RATE.
• ABOUT 200 MS FOR FRAME RATES OVER 15 FPS.
• ABOUT 850 MS FOR FRAME RATE OF 1.9 FPS.
THE OPERATOR SHOULD HAVE THIS LATENCY IN MIND WHEN MAKING A CHOICE OF EMBOLIC AGENTS
USED DURING INTERVENTIONS IN RELATIONSHIP TO POLYMERIZATION TIME.
There is a residual risk of a frozen Image being displayed on the live display if a failure occurs, this has never been
observed though. If X-ray has been interrupted during Fluoro or Record, a Frozen image may remain displayed on
the live display. The X-ray on indicator is active, "X-ray acquisition in progress message" is displayed, and the
X-ray on light remains on. Special precautions should then be taken for all interventional procedures and
manipulation.
7-8 Control of transfer of data to an archiving system
Innova System does not feature archiving but provides capability to connect an archiving system. So, not being an
archiving system, Innova ensures the control of transfer of the data to the archiving system.
The control of transfer is ensured by:
• the double confirmation before deletion of an item if this one is not successfully transferred,
• the network queue.
In the network queue items selected for archive are flagged using the following statuses:
• ACTIVE: The data transfer is in progress; only one item is flagged as "ACTIVE" in the queue.
• PENDING: The data transfer is waiting to be performed; multiple items can be flagged as "PENDING".
• FAILED: The data transfer was not completed due to an error ("FAILED"). The user shall re-transfer the item.
An item may be a patient, an exam, a sequence or a photo.
If the data was successfully transferred, the item is removed from the network queue. An empty queue means that
all items were successfully transferred.
In case of successful transfer, the DL guarantees that the transfer occurred with no data loss, but the DL does not
give any guarantee that the data has been successfully archived on the receiving system, it is recommended that
the user checks that the data is correct on the receiving system before deleting it from DL.
7-9 SID Failure - Degraded mode
In case of SID failure, the following message is displayed on the in-room monitor:
SID failure. Collimation activated. To raise detector, use pink button
If this message appears on the in-room monitor, user has to move the detector towards its highest position in using
the emergency back-out button (1) placed on the detector as shown above. X-ray are available even if the error
message is shown, applied x-ray parameters ensures minimal dose to the user and the patient.
7-10 Prevent Unnecessary Interruption of a Procedure Arising from a Collision
Note: Different measures have been implemented to prevent unnecessary interruption of a procedure arising from
a collision. At obstacle approach:
• the gantry slows down (half speed).
• warning icons are displayed on live display.
• an optional audible signal warns the operator of an imminent close collision.
7-11 Automatic Exposure Control
The procedure to verify the Automatic Exposure Control is available in the DVD SM "InnovaTM IGS 520, InnovaTM
IGS 530, InnovaTM IGS 540" (Functional Checks Tabs/CHK0024 - ABC Stabilization Point Checks). Apply this Job
Card.
7-12 Lead Shield Positioning
SPECIAL ATTENTION MUST BE TAKEN WHEN MOVING THE LEAD SHIELD CLOSE TO THE
PATIENT DURING A PROCEDURE. ALWAYS VERIFY FIRST THAT THE CATHETER CANNOT
BE ACCIDENTALY WITHDRAWN FROM THE PATIENT BEFORE MOVING THE LEAD SHIELD.
7-13 Risk of confusion between live Fluoro images and automatic replay of stored
Fluoro images
Secure all devices or equipment located on the table top prior to tilting the table to avoid
injury to patients and staff.
7-15 Degraded Fluoro image quality
Some parts of the comfort accessories are not radiolucent and may cause image quality
artifacts that require distinguishing them from anatomical structures.
7-17 Open Suspension Positioning
SPECIAL ATTENTION MUST BE GIVEN WHEN MOVING THE OPEN MONITOR SUSPENSION
CLOSE TO THE PATIENT DURING A PROCEDURE. BEFORE MOVING IT ALWAYS VERIFY
THAT THE CATHETER CANNOT BE ACCIDENTALY WITHDRAWN FROM THE PATIENT.
Involuntary mechanical shocks on the Smart Box joysticks may produce unwanted
positioner motion and cause injuries. In case the Smart Box needs to be removed from the
table, always leave it in an area where its joysticks are protected from shocks.
WHILE CLAMPING TSUI ON RAIL (ON THE TABLE SIDE, ON THE CART OR ON THE
REMOTE STAND), IT IS IMPORTANT TO DOUBLE CHECK THAT TSUI IS CORRECTLY
CLAMPED ON BOTH ITS LATERAL SIDES. THERE IS A RISK OF TSUI FALL, UNWANTED
TABLE AND GANTRY MOTION AND INJURY WHEN THE TSUI IS NOT CORRECTLY
CLAMPED.
Upper side
Lower side
Recommended Position
Position to avoid
Note: When a Smart Box or TSSC is installed near the foot end of the rail, it may be difficult to access and view the
controls when the table is tilted Trendelenburg. Move the tableside controls closer to the head end of the rail.
8-2-2 Tableside Cart (Option)
The Tableside Cart is provided to the customer fully installed and ready to use. Any mechanical configuration
modification must be performed by qualified and trained engineers only.
This product must be serviced with Manufacturer approved original replacement parts only. Failure to follow this
recommendation will void the Manufacturer’s warranty and will release the Manufacturer from liability and warranty
claims.
The Manufacturer cannot be held liable for unauthorized modifications made to the system or use of the system for
unintended purposes. Since improper modifications may impair proper operation, safety, or reliability of the system,
all modifications need to be authorized by the Manufacturer.
When positioning or moving the Tableside Cart, avoid collision with other equipment.
Failure to do so may lead to severe damage on both the Tableside Cart and other
equipment. In case of severe collision or equipment damage, immediately stop operation of
the system. Keep the area around the Tableside Cart clear of any equipment.
The Tableside Cart is not designed to hold other accessories than those installed at tableside of GE Healthcare
Cardiovascular Imaging systems.
Compatible Table Side User Interface (TSUI) allowed to be installed on the Tableside Cart:
• Smart Box
• Table Side Status Control (TSSC)
• Central Touch Screen
• In-room 3D Mouse
• Volcano Touch Pad Controller
The Tableside Cart is not designed to hold the injector head or injector control panel.
The maximum weight a rail can hold is limited to 13 kg, which is always more than any combination of 2 of the
above TSUI.
The maximum weight the Tableside Cart can hold is limited to 50 kg. Do not stand on wheels or Tableside Cart
base.
To prevent the Tableside Cart from falling over, never move or use it on a slope greater than
10°.
Do not open covers.
8-2-3 Mattress
To prevent any patient fall risk during a table tilt, the mattress is secured to the table top using a set of Velcro
strips.
DO NOT USE THE MATTRESS UPSIDE DOWN. THERE IS A RISK OF PATIENT FALL WHEN
THE TABLE IS TILTED BECAUSE VELCRO STRIPS ARE LOCATED UNDER THE MATTRESS
ONLY.
A set of Velcro strips is installed on the table top and under the mattress at mid length and at foot end.
In case the mattress needs to be removed from the table top, pull up the mattress at foot end, then at the mid
section to release the Velcro strips.
To install the mattress, start from head end.
1. Align the mattress head end to the table top head end.
2. Roll out the mattress aligned to the table top longitudinal axis.
3. Check that the Velcro strips are facing each other and press on the mattress to engage the strips at mid section
and foot end.
8-3 Precaution regarding Detector Anti-Collision Device
Bodily fluids may damage internal components if they are allowed inside the equipment.
Use drapes, if necessary, to protect equipment when performing procedures.
If you are performing a procedure where draping is necessary, you can choose among the following solutions. You
can contact your local GE representative for more information on how to order these solutions.
Image Receptor
Smart Handle
Smart Box
9-1-3 Accessories
In-room 3D Mouse
10 Symbols
Symbols used on the vascular systems and in its accompanying documents are shown and explained in this
section.
• Audio/Visual Indications
• Special Notices
• X-ray Tube
• Power ON and OFF
• Type B Applied Part
• Electrical Class
• Electrical Current
• Ground
• Collimator
• Cardiac Pulmonary Resuscitation (CPR)
• Maximum Patient Weight
• Maximum permissible table load
• Table end rail maximum load
• Footswitch and Table Panning Device (TPD)
• Smart Handle - Smart Box - Table Side Status Control
• Innova Console
• Application
• Grid Out
• Acquisition
• InnovaIQ Table
• Status Icons
• Remote Stand (Option)
Note: * The behavior may change depending on country selection. Refer to Safety and Regulatory Instructions for
Use / Protection regarding Ionizing Radiations Hazards for Innova IGS 520 / Dose to Patient in Japan or Safety and
Regulatory Instructions for Use / Protection regarding Ionizing Radiations Hazards for Innova IGS 530 / Dose to
Patient in Japan or Safety and Regulatory Instructions for Use / Protection regarding Ionizing Radiations Hazards
for Innova IGS 540 / Dose to Patient in Japan for more detail.
Buzzer Kit: Allows to propagate all audio indications in the exam room.
• The loudness of the loudspeaker was adjusted at installation. Contact your GE Service Representative for any
further modification.
• For cleaning, unplug the power source before cleaning. Do not use liquid cleaners or aerosol cleaners. Use a
damp cloth for cleaning.
• Do not attempt to service any part of the speaker yourself as opening or removing covers may expose.
Fluoroscopy and Radioscopy audible signals may be configured. Push Fluoro icon or Record icon in monitoring bar
to open the sound configuration window:
The following settings are available for Fluoroscopy and Radioscopy. By default at installation, the system is
configured with a short beep for Fluoroscopy and a continuous tone for Radioscopy.
Note for Japan: As described in Safety and Regulatory Instructions for Use / Protection regarding Ionizing
Radiations Hazards for Innova IGS 520 / Dose to Patient in Japan or Safety and Regulatory Instructions for Use /
Protection regarding Ionizing Radiations Hazards for Innova IGS 530 / Dose to Patient in Japan or Safety and
Regulatory Instructions for Use / Protection regarding Ionizing Radiations Hazards for Innova IGS 540 / Dose to
Patient in Japan, the usage of special high level Fluoroscopy mode is indicated permanently on the Reference in
room monitor by the annotation "Normal" and by a permanent audible signal. Therefore, Fluoro and Record beep
settings do not apply in this special mode.
Note: With the continuous tone setting, slower frame rates will sound like a beep signal.
Fluoroscopy and Radioscopy sound can be selected from the respective drop-down menu:
Extra caution should be taken if the Fluoro or Record sound is set to "None". Inadvertent
x-ray activation and unknowingly activating the footswitch will not be detectable by an
audible signal because the sound will be deactivated.
10-2 Special Notices
General Caution.
This symbol is used to highlight the fact that there are
specific warnings or precautions associated with the
device, which are not otherwise found on the label.
Class II Equipment. Protection against electrical shock does not rely on basic insulation
only, but in which additional safety precautions such as double or reinforced insulation
are provided. There is no provision for protective earthing or reliance upon insulation
conditions.
10-8 Ground
10-9 Collimator
Table Top Longitudinal and Lateral Brakes Release (on footswitch and TPD)
This sticker is located on the back side of each Table Side User Interface
(TSUI). Do not place TSUI on the floor. Unwanted table or gantry motion
could be activated.
10-17 Application
Displayed on left side of the live display to show that the displayed image is a
live fluoroscopic image.
Displayed on left side of the live display during a Fluoro performed in Auto Flu-
orostore mode.
Displayed on left side of the live display to show that the displayed image is a
recorded image.
Displayed on left side of the live display to show that the displayed image is the
frozen last fluoroscopic image.
Displayed on left side of the live display to show that the displayed image, pre-
viously recorded, is reviewed.
Displayed on left side of the live display to show that a mask is available for the
Roadmap fluoro.
Displayed on the left side of the live display to show the system is in Blended
Roadmap and the selected percentage of vessel visibility
30% Displayed on left side of the live display to show the selected level of landscape
for Subtracted or Roadmap fluoro.
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Forward nominal
speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Forward half speed
(review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Forward high speed
(review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Backward nominal
speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Backward half
speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Play mode Backward high
speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously recorded image reviewed in Pause mode (frozen image)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Forward
nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Forward
half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Forward
high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Backward
nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Backward
half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Play mode Backward
high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed im-
age is a previously stored fluoroscopic image reviewed in Pause mode (frozen
image)
Displayed on left side of the live display to show that an ECG signal is being
detected
Displayed on left side on the live display, in the Geometry area, to show that
the anti-scatter grid has been removed and is not in the X-ray beam.
10-19 Acquisition
Displayed at the bottom of the control room console to show that a fluoroscopic
or record acquisition is in progress
This sticker is located on the top right and top left of the
bellows surrounding the table base.
Do not leave hand/fingers on the bellows
There is a potential risk of hand/fingers pinch when the
table is moving down.
Do not stand between the table top head end and the
gantry pivot when the positioner is not disabled using the
Positioner Enable/Disable button located on the Smart
Box.
Risk of severe injury when the table top is moving to-
wards the gantry pivot.
Displayed on the right side of the right side of the reference display to show
that disk space is still available.
Displayed on the right side of the right side of the reference display to show
that disk starts to be full.
Blinking icon displayed on the right side of the reference display to show that
disk is near to be full.
Displayed on the right side of the reference display to show that disk is full.
Displayed on the right side of the reference display to show that at least 50% of
heat units are still available in the X-ray tube.
Displayed on the right side of the reference display to show that available heat
units in the X-ray tube are between 30% and 50%.
Blinking icon displayed on the right side of the reference display to show that
available heat units in the X-ray tube are below 30%.
Displayed on the right side of the reference display to show that the maximum
heat units are reached in the X-ray tube.
Displayed on the right side of the reference display to show that the Fluoro is
ready (enabled).
Displayed on the right side of the reference display to show that the Fluoro is
not available (disabled).
Displayed on the right side of the reference display to show that the Record is
ready (enabled).
Displayed on the right side of the reference display to show that the Record is
not available (disabled).
Sending Some network activity is in-progress Recommended not to perform a delete operation.
If deletion is still required, the system will provide warn-
ings and perform the deletion operation after user’s con-
firmation.
Archiving Archival is in-progress (Image or Dose This activity can take some time. Depends on the
Structured Report) Archiving station setup.
Recommended not to perform a delete operation.
If deletion is still required, the system will provide warn-
ings and perform the deletion operation after user’s con-
firmation.
Information Some network activity did not take Recommended to place the mouse over the icon to read
place (or) all the network activities and perform the action required.
completed but the patient exam is still
Recommended not to perform a delete operation.
active.
Completed All network activities completed. Can be safely deleted.
Failed Some network activity failed. Recommended to place the mouse over the icon to read
and perform the action required.
Recommended not to perform delete operation.
Symbol indicating that the Instruction For Use (Operator Manual) is sup-
plied in electronic format. It is located on the L-arm.
Label for maximum load for surgical and patient comfort accessories.
Note: Refer to the table in Surgical and Patient Comfort Accessories /
Accessory rail load consideration for the weight values.
Lock/Unlock symbol.
Side Up symbol.
Bolus button
Table Speed
X-ray ready
10-30 Filtration
Indicates the manufacturer's catalogue number so that the medical device can be
identified.
Indicates the manufacturer's serial number so that a specific medical device can be
identified.
2 Doses parameters
There are different types of dose depending on where and how the measurement is taken and if the measurement
is for the patient or the detector dose.
2-1 Incident dose
The incident dose is the dose measured in the middle of a radiation field on the surface of a body or a phantom.
However, it is only measured at this point if there is no body in the path of the X-ray beam. Thus, there is no scatter
radiation from the body during this measurement. When radiation strikes a substance, there is always a certain
scattering of radioactive particles. This is comparable to light striking a glass surface; a certain portion of the light is
always reflected.
The unit used to measure the incident dose is joules per kilogram, and is known as "Gray" where 1 Gray (Gy) = 1
J/kg. The former unit used to measure the incident dose was the "Rad", and using this unit, 1 Rad (rd) = 0.01 Gy,
or 1 Gy = 100 rd. But because today's doses are generally very small, they are usually described using the unit
measured dose
required time
The SI unit used to measure the dose rate is Gray per second: (Gy/s) or in usual unit (mGy/min)
2-6 Dose-area product
The dose-area product is a measurement of the amount of radiation that the patient absorbs. The dose-area
product is independent of the distance between the X-ray tube and the measuring device because the further away
from the X-ray tube this measurement is taken, the more the size of the device increases, and the dose itself
decreases (see diagram). The dose to the patient can be calculated from the dose-area product, the size of the
measuring device, and the distance to the X-ray tube and the patient.
Dose-area product = dose * surface area of the measuring device
The SI unit used to measure the dose-area product is the Gray * meter² (Gy*m²) or in usual unit cGy*cm² or
Gy*cm².
Dose-area product
The dose-area product at 50 cm from X-ray tube is just as great as dose-area for 100 cm or 200 cm, because the
size of the measuring device increases with greater distance to the X-ray tube. But the dose itself decreases with
greater distance to the tube. Thus the dose-area product is the same at each position if the size of the measuring
device enables it to detect all of the radiation.
2-7 Body dose and effective dose
The body dose is the comprehensive concept for the organ or partial-body dose equivalent and the effective dose.
In the practical application of radiation protection, however, local and individual doses are monitored, because body
doses cannot be measured directly. The Radiation Protection Regulations therefore use the concept of effective
dose, in which all the individual doses to the irradiated organs or parts of the body are multiplied by a factor and
then added together. The resulting value may not exceed the dose limit for the effective dose that a patient is
allowed to receive.
Body dose = sum of all organ or partial-body doses
Effective dose <= patient dose limit
The SI unit used to measure the body dose and the effective dose is the sievert, where 1 sievert = 1 Sv = 1
Joule/kilogram = 1 Gray
reference point is, in fact, a fair approximation of the focal spot to skin distance for each field.
In general, the dose amounts to 1/x². Therefore, if you double the detector-to-target distance, you will need four
times as much radiation to achieve the same image blackening.
A Skin Spacer may be used to assure a minimum distance of 380 mm to the patient's skin. If you positioned the
patient too close, this would lead to excessive skin dose radiation to the patient; thus, increasing the distance
between the X-ray tube and patient helps to reduce the amount of skin dose radiation to the patient. For example,
the exposure rate is approximately 1.5 times greater at 305 mm from the focal spot to the patient than 380 mm.
4-2 Distance between the patient and the detector
The detector should be kept as close to the patient as possible for all exposures. This helps the Innova system to
select the lowest optimal technique factors to visualize the anatomy properly. When the distance between the
patient and the detector is excessive, not only is the magnification increased but so are the technique parameters
which could lead to a poor image that is too gray and flat losing small vessel detail.
4-3 Collimation
Collimation brings about a genuine dose reduction and also produces better image quality. Collimation is performed
using collimators (multi-leaf collimators or iris diaphragms) that are attached directly in front of the X-ray tube.
Collimation at the target is the most effective radiation protection for the patient and personnel, because it narrows
the area that the radiation can strike.
4-4 Compression
Because radiation scatters in a body exposed to X-rays, compression of the body is another way to reduce the
radiation dose. Scattered radiation also produces an undesirable reduction in contrast in the X-ray image. With
compression, the thickness of the body is reduced, and so a lower dose is absorbed by the body. Additionally,
compression ensures that less scattered radiation occurs.
4-5 Anti scatter grid
The grid ratio is 13:1
Before mounting, check if the reference is the same on the anti-scatter grid and the following illustration:
Anti-scatter grid
Anti-scatter grid
Supplier Reference:
Anti-scatter grid 21 cm: 9896 010 67911
Anti-scatter grid 31 cm: 9896 010 67931
Anti-scatter grid 41 cm: 9896 010 67941
The anti-scatter grid is located between the patient and the Revolution Digital Detector. It is the most effective
method of reducing scattered radiation. The grid absorbs a portion of the scattered radiation in its lead plates.
This absorbed dose therefore does not reach the Revolution Digital Detector, even though it has already passed
through the patient. Thus, the use of an anti-scatter grid leads to an increase in the dose, because the amount of
radiation that reaches the Revolution Digital Detector, is not reduced until it has passed through the patient: if the
anti-scatter grid is used, the patient must be exposed to a higher dose of radiation in order for the minimum dose to
reach the Revolution Digital Detector. But, we are not talking about excessive amounts of increased dose and one
needs to understand the benefits as well as the trade offs of using a grid. In cardiac procedures, the chest area of
the patient , where the X-ray beam has to penetrate, is the largest portion of the patient as well as some patients
are over weight and present a challenge to imaging to begin with. Then add steep angles for the views of the heart,
and you have fairly high techniques as compared with the rest of the body. If there was no grid used, all the
secondary scatter exiting the patient would reach the detector as well as the primary beam. The scatter has no
usefulness for creating a good image so the grid helps to eliminate it before it can affect the overall image quality.
We can differentiate between individual anti-scatter grids using their grid ratios. This is the relationship between the
height of the plates to the distance from each other.
The greater the grid ratio, the greater the grid's effect. Thus, the required dose increases with the grid ratio. The
typical grid ratio is 13:1 for the Innova Systems.
4-6 Record and Fluoro mode selections
The Innova System provides the user with two Record modes and two Fluoroscopy modes. The Low selection for
each will use lower dose to the receptor while providing lower patient skin dose. The image quality of the Low mode
may be compromised because of the patient's size or the steepness of the angle. In that case, the Normal mode
should be used. Neither mode will increase the skin dose to the patient beyond the allowed limit of the country.
When possible, change the entry point of the beam in order to reduce the local dose to the patient. For long
interventional procedures, this will help to reduce any potential of skin burns to the patient.
4-7 Frame rates
In Record mode, try to keep the frame rate at the lowest value compatible with the medical requirement. Frame
rates are set depending on the blood flow of the vessel and or the movement of the anatomy. In addition, the
shorter the length of the sequence and the fewer sequences acquired will help to reduce the total exam dose to the
patient.
4-8 Field of View
There are four Field of Views available on the Innova Systems. Even though the detector doesn't need additional
dose to create an optimal image, there is a slight increase of dose to the receptor to minimize the noise seen in the
image. Even though this dose increase is minimal, using the largest Field of View with the least amount of
magnification will assure the least amount of dose used.
4-9 Radiation filtration / hardening
The quality of the X-ray also plays a great role in the size of the administered dose. X-ray radiation normally has
so-called "hard" and "soft" particles, that is, particles with a lot of energy and particles with little energy. Hard
particles are better for the patient, because they pass through the body. Soft particles, by contrast, get caught
inside the body because they are too weak to pass through and out of it. Therefore, it is primarily soft radiation that
creates unnecessary exposure to the patient. For this reason, copper and aluminum are used as filters in front of
the X-ray tube. The soft radiation is caught in the filter plates, and the remaining radiation emerging from the filter is
"harder". This additional filtration can also reduce the dose to the patient without diminishing image quality,
because in any case only the "hard" rays reach the Revolution Digital Detector.
radiation, the intensity of the radiation per unit mass is reduced by a factor of four.
For the greatest safety, stay as far from the source of radiation as possible. When the tube is positioned under the
table the patient helps to protect the staff. Radiation travels in a straight line. Don't stand where the tube is pointing
towards you, but behind the X-ray tube if possible. Also, any distance six feet (1.83 m) or more away from the X-ray
source and patient is considered safe.
There is a significant zone surrounding the table that identifies proximity to the radiation source. Included in the
significant area is the tableside area where the Physician stands to perform the procedure. All equipment within this
zone 1.5 m around the patient are medical components (table, positioner, monitors, injector, radiation protection
shield, etc.) that must be in compliance with medical device regulation.
5-1-2 Shielding
Any material placed between you and the source of X-ray radiation will absorb some of the radiation, thus reducing
its intensity. The more material with the greatest density will absorb more radiation. High atomic number materials
such as lead, are the most useful for absorbing X-rays. Depending on the energy of the X-rays, only 1 or 2 mm of
lead will reduce the radiation dose by a factor of more than 100.
Shielding can come in many forms. Lead aprons and protective devices such as lead gloves, glasses, thyroid
shields are also effective in limiting occupational exposure dose. When working near the X-ray equipment other
objects can be placed in between you and the radiation. An eye and thyroid lead window shield, a lead radiation
shield mounted on the table or equipment, or a mobile stand shield. Of course, the best shielding is standing
behind lead window panels and lead walls outside the angiographic room.
Proper shielding cannot eliminate X-ray, but it can reduce the quantity to an acceptable level.
5-2 General Precautions
1. Stay as far away as possible from the source of radiation. Know the path of the primary beam and work as far
from it as possible.
7 Smart Fluoro
Fluoroscopy is performed at low dose compared to record and inherently shows more noise. Quantum noise is a
direct function of the quantity of X-rays. At a given level of clinically acceptable noise, dose reduction can be
achieved either by at least two methods: Slowing the pulsing rate or advanced digital processing of successive
frames acquired with low dose pulsed fluoroscopy.
Fast (30 fps) pulsed fluoroscopy uses short, high intensity pulses to better freeze motion in the image. It is also
possible to skip pulses, filling the "gap" with previous image. This reduces the quantity of X-rays used and results in
less dose per image. When you reduce the pulse rate, the inherent temporal averaging performed by the human
visual system is less effective and images may look a bit noisier at lower frame rate. This can be compensated by
increasing dose per frame appropriately, yet achieving dose saving in reduced frame rate mode to the extent
possible. The images may appear slightly jerky at lower frame rates. Innova systems offer reduced pulse rate
option during fluoroscopy.
An alternative method of lowering dose is to digitally process images, by averaging several successive images into
one, or by performing averaging within a given frame. Temporal averaging allows the use of the same X-rays more
than once. This is equivalent to more photons to the image, for the same patient dose. Using image processing, the
temporal averaging effect of the eye is thus further enhanced. During digital processing, moving objects have to be
treated differently to avoid blurring. GE has a patented technique called Smart Fluoro for effective temporal
averaging of frames. The algorithm effectively discriminates moving object edges from the background to ensure
that noise is reduced while moving objects are preserved. In addition GE offers an advanced spatial denoising
algorithm that combines wavelet-based multi-resolution decomposition with edge-preserving spatial filters. This
algorithm reduces noise level without degrading the clarity of edges and fine details.
7-1 Benefits of LIH
Last Image Hold (LIH) provides an image in between the live fluoro images to use as a reference image and
eliminates the need for extended fluoro exposures
7-2 Spectral Filtration
Spectral filtration is added filtration that is inserted between the X-ray tube source and the patient in order to
harden the beam and reduce the soft radiation. The harder the beam, the better the penetration of the anatomy
allowing the X-ray photons to exit the patient and reach the digital detector. The spectral filters also absorb the
softer radiation so that it never reaches the patient and keeps the Air Kerma radiation and absorbed dose to a
minimum. The Innova Systems have automatic spectral filtration that appropriately inserts either 0.1 mm, 0.2 mm,
0.3 mm, 0.6 mm or 0.9 mm of copper (depending on system) into the fluoroscopy or record exposure if needed.
7-3 Tabletop
The material from which the tabletop is constructed is also significant for the required dose, because the tabletop is
penetrated by the radiation and weakens it before it reaches the image intensifier. Therefore, if at all possible the
tabletop should not contain any material that strongly weakens the radiation or absorbs it well, such as lead or
metals in general. Carbon fiber has proven to be the best material for X-ray system tabletops because its radiation
absorption is minimal and the tabletop can take a great amount of stress; today a tabletop is expected to be able to
8 Special Case
8-1 Pediatrics
Because children have a greater sensitivity to radiation than adults, special conditions apply to pediatric radiology.
When possible, an attempt should be made to avoid X-ray examinations altogether and to use alternative
procedures. But when X-rays can not be avoided, there are a few basic practices that can be used to minimize the
dose to a child.
For infants up to 1 year old the anti scatter grid can be removed from acquisitions performed at Field Of View
smaller than 20 cm.
Collimation to the specific area of interest is still the first choice of reducing dose to a child.
And when possible, it is especially important to use a gonad shield.
For access to specific dose settings for pediatrics in Innova Systems, select a protocol on the DL console where
the name contains ‘peds’ or ‘pediatrics’ (see Innova Digital / How to select an Acquisition Protocol).
For details about how the system optimizes the exposure techniques and how it complies with regulation about
dose settings for pediatrics, (SV-Richtlinie, point about Tabelle I 1, Kinder KB25 < 0.2 µGy/s and the alternative
application note E21), please refer to Appendix “Dose settings for pediatrics in Innova systems” or the following
article:
Barry Belanger, John Boudry: Management of pediatric radiation dose using GE fluoroscopic equipment. Pediatr
Radiol (2006) 36 (Suppl 2): 204–211.
9 Legal Provisions
In many countries or states, by means of rules, guidelines or regulations, lawmakers have contributed to improving
radiation protection for patients and medical personnel; after all, medical exposure to radiation is the single largest
source of radiation exposure among the general population. On an international level, guidelines are laid down by
the International Commission on Radiological Protection (ICRP). Many of the rules, guidelines or regulations are
governed by the ALARA concept (As Low As Reasonably Achievable), meaning the production of a diagnostically
relevant image at minimum possible dose.
5 Highest table working height • Use the highest table working height acceptable.
• Maximize X-ray tube to patient distance.
6 Collimate • Always collimate to the anatomical area of interest.
• Use the largest acceptable image detector Field of View, with col-
limation, in lieu of image magnification.
• If image magnification is required, use the lowest acceptable
dose rate.
7 Monitor cumulative patient dose • Set thresholds of cumulative patient dose (Yellow/Orange/Red
levels).
• Keep track of cumulative dose display on monitor relative to these
thresholds during the case.
Image Quality Trajectories • Options are available to reduce dose up to 75%.
• Review with GEHC Applications or Service the available IQ Tra-
jectories.
• Ask them to customize the IQ protocols (IQ+, IQ Standard, RDL+,
RDL).
• Dose strategy need to be adjusted according to the user and clini-
cal procedure needs.
The DFP replaces the analog image intensifier (II), camera optics, pickup tube or CCD camera, and analog to
digital converters. It therefore provides the first full digital imaging chain. Because it replaces the components that
degrade image quality with a DFP detector, it can capture information with minimal loss. The result is improved
image quality.
1 Imaging Chain
Conventional Cardiac Imaging Chain with Image Intensifier
• The digital detector receives the X-rays transmitted through the patient
The Revolutionary Digital Detector replaces everything but the X-ray tube and patient.
As you can see, there is a reduction of the total number of parts that make up the digital versus the analog imaging
chain. Image quality can be affected by each component in the chain. Not only are there fewer parts that can fail,
but fewer to optimize. An image chain is only as good as it's weakest link. Note that at each stage in this process,
the X-ray signal is degraded to some extent, even if the individual components are optimized for the application. As
a result, typically less than 40% of the original image information is available for use in image production.
Because of its high Detective Quantum Efficiency (DQE), it has the potential to capture over 80% of the original
image information. And it equips the user with a wide range of post-processing tools to further improve that signal -
including many that can be applied automatically.
The operation:
intensities, as well as very high contrast resolution to permit the display of thousands of shades of gray. In fact,
that's one of the digital X-ray's great advantages over conventional imaging systems. Digital X-ray detectors can
successfully image areas that might be under or over exposed.
When testing very low dose acquisition modes, it is recommended to use 0.1mm Pb thick phantoms such as:
phantom 07-541 (type 41) of thickness 0.1 mm Pb, and groups 0.6 to 3.4 lp/mm.
The detector spatial resolution performances of the Innova IGS Systems also answer to country regulations.
Detector spatial resolution acceptance criteria per modes for Germany
Detector spatial resolution acceptance criteria in fluoroscopy for Italy and Spain
Detector spatial resolution acceptance criteria for Australia VIC and QLD
7 Radiation Dose
Reducing radiation dose is another potential advantage of digital detectors. A detector with high DQE has the
potential to deliver significant object detectability improvement at the equivalent dose, or to permit object
detectability comparable to film's at a reduced dose. Or to say it another way, high DQE allows acquisition of the
same image quality at a lower dose, or better image quality at the same dose. Now the cardiologist can choose
between superior image quality or dose savings to the patient or staff. For example, the doctor may not need the
same image quality for a diagnostic study and can choose a setting that will allow for lower dose. But if the
procedure develops into an interventional procedure, the doctor can then select a different setting allowing for
superior image quality for placing the balloon or stent.
In addition, with the development of AutoEx, the optimal exposure technique for the patient size and the level of
contrast and sharpness desired can be selected. This allows for the entire procedure to be viewed with the same
image quality, no matter the angle or field of view (magnification). In addition, GE uses adaptive spatial and
temporal denoising algorithms with dynamically adjusted parameters depending on the size of the patient and the
level of noise seen in the image. On small to average size patients, the system is rarely pushed to the maximum
technique or dose limit. But for large patients and steep angles, that is a different story. In the US, the skin Air
Kerma radiation that the patient can receive during fluoroscopy is 10R/min. But many countries are reducing that
limit to 5R/min. As we have already discussed, the less Signal (dose) you have the more noise in the image you
have. To counteract that, the dynamic Image Processing will adjust to provide for a smoother more contrasty
fluoroscopy image when the dose is limited or at the maximum amount allowed.
1 Gantry
2 Table
3 Smart Box
4 Table Side Status Control (TSSC)
5 X-ray Tube
6 Digital Detector
2 System Components
Availability of configurations may vary according to the country regulatory registration
2-1 Basic Components
• C-arm unit (including C-arm, control cabinet and table-side control boxes)
• Angio Table: Omega V or InnovaIQ Table; Cardiac table: Omega IV (all with mattress included)
• Table Side User Interface (TSUI): Smart Box or Smart Handle (not available with InnovaIQ Table), Table Side
Status Control (TSSC)
• X-ray high-frequency and high-voltage generator: JEDI 100
Peripheral Device connections for Omega Table Peripheral Device connections for InnovaIQ Table
In addition, the Omega Table allows the connection of the following devices:
1 Gantry
2 Status control
3 Debug screen
4 Footswitch
5 TPD
6 Gantry
7 Status control
8 Injector
9 3D mouse
10 Central Touch Screen
In addition, the InnovaIQ Table allows the connection of the following devices:
More information can be found in the service manual provided with the system.
When the cover is opened, do not touch the accessible contacts of connectors and the patient simultaneously.
3 Innova Console
4 General Specifications
Specifications may be subject to change without modification in the behavior of the system.
4-1 Generator
Specifications apply to generator independently from tube association and system applications.
Nominal output power: 100 kW
Max kW of generator's output when loading time is 0.1s, 100kW.
• Peak power
Maximum peak power: 100 kW (may be limited, depending on the mode or application)
• Average power
Average power is intended to be the average power computed in a wide time frame, greater than one hour.
Maximum average power: 3.2 kW (may be limited, depending on the mode or application)
• Range of high frequency used by the high-voltage generator: 20 kHz - 60 kHz
• kV-mA range
kV from 40 kV to 125 kV
Maximum mA: 1000 mA
• Large focus
Maximum mA: 1000 mA
Maximum power: 100 kW
• Small focus
Maximum mA: 400 mA
Maximum power: 48 kW
• Compressed focus
Maximum mA: 200 mA
Maximum power: 16 kW
4-2 Gantry
• L-arm rotates on its vertical axis +/- 100° (+/- 95° motorized limit set).
• Offset C-arm (pivot) permits -117°/+105° RAO/LAO rotation.
• The C-arm permits 50° cranial and 45° caudal angulation of the imaging system.
• The combination of movements of the C-arm and L-arm permits +/- 55° cranial and caudal angulations.
• The offset arm throat depth of 107 cm (42") with L-arm at 0° provides femoral coverage on most patients.
• Rotation speed of Offset-C-arm (Pivot) and C-arm; 0-15° per second (0-20° per second for Innova IGS 520 and
Innova IGS 530 with the InnovaSense option).
• Rotation speed of L-arm: 10° per second.
• SID Fully motorized (8.9 cm/s - 3.5 in/s).
• SID range of 85 cm to 119 cm (33.5" to 46.8") for Innova IGS 520. Travel distance of 34 cm (13.4").
• SID range of 89 cm to 119 cm (35" to 46.8") for Innova IGS 530. Travel distance of 30 cm (11.8").
• SID range of 95 cm to 119 cm (37.4" to 46.8") for Innova IGS 540. Travel distance of 24 cm (9.5").
• X-ray tube Focal Spot to Isocenter 72 cm (28.3") .
• Isocenter to the floor of 107 cm (42").
• X-ray tube reference axis is tilted by 3° versus the X-ray beam axis in the anode direction on the Innova IGS
520.
• X-ray tube reference axis is tilted by 1° versus the X-ray beam axis in the cathode direction on the Innova IGS
540.
• A green LED is located on Pivot inner cover to indicate the X-ray status. This LED is on when the system is
ready to perform X-rays, otherwise the LED is off.
4-3 Table
4-3-1 Omega Tables
The Omega table can operate while engaging the motor drive, table height up and down while fully extended,
supporting a maximum table weight of 304 kg: maximum of 204 kg for patient weight, maximum weight of 40 kg of
accessories can be installed on each table accessory rail and maximum of 20 kg on the table end rail.
InnovaIQ Table
Table Top filtration Less than 0.85 mm aluminium equivalent
Mattress 5 cm (2") Mattress : less than 1.20 mm aluminium equivalent
Maximum Patient Weigth 204 kg (450 lb)
Maximum Table Load 320 kg (705 lb) including accessories
Maximum Tabletop Load 251 Kg (553 lb) including surgical and patient comfort acces-
sories
Tabletop Length 333 cm (131")
Tabletop Width 46 cm (18") in Patient Trunk Area - 67 cm (26.4") max
Tabletop Width 53 cm (21") in Patient Trunk Area (for Wide Table Top only)
Horizontal Float Movement 8-way
Horizontal Motorized Movement 8-way
Longitudinal Speed 0 to 25 cm/s
Longitudinal Travel 170 cm (67")
Equivalent Fluoroscopic Coverage Innova IGS 520: 186 cm (73")
Innova IGS 530: 194 cm (76")
Innova IGS 540: 198 cm (78")
Transverse Travel Motorized +/-13 cm (+/-5.1")
Transverse Speed 0 to 10 cm/s
Vertical Travel 57 cm (22.4”)
Vertical Travel above Floor From 80 cm (31.5") to 137 cm (53.9")
Vertical Speed 0 to 2.5 cm/s
Table Base 73 cm x 55 cm (29" by 21.7")
Tilt Travel From 20° Head Down to 12° Head Up
Tilt Speed 0 to 2°/s
Table Rotation +/- 180°
Patient Support Type Cantilevered
On the DL image disk, it is possible to store up to 4000 sequences and photos altogether as long as disk space is
still available.
4-5 Innova Imaging
4-5-1 Innova IGS 520
1 X-ray Indicator
2 Audible Signals
3 System Reset
4 ON/OFF
5 X-ray Timer Reset
6 Auto inject ON/OFF
Note: In case a system power-Up is required right after a system power-Down, wait until the control room console
is fully blackened prior to initiate the system power-Up. The control room console then stays blackened for about 1
minute before the new display appears, but the power-Up action is taken into account. There is no need to initiate
additional power-Up during this 1 minute of blackened screen.
Whenever the Innova System Main power has been removed with the wall main circuit breaker, turning ON the
Innova will require additional steps.
• Verify the wall emergency button is unlocked.
• Switch ON the wall main circuit breaker if needed.
• Go to the technical room where the Power Distribution Box (PDB) and the UPS stand.
• Verify the PDB emergency button is unlocked.
• Verify the UPS test switch is on "normal" position.
For PDB (CE):
• Close the disconnecting switch DM1 (item 1) on EMI filter ( if EMI filter is present) and switch S1 (item 2) on
PDB, if needed:
• Wait for 20 seconds then ensure the pilot lamps PL1 (item 7) and PL2 (item 8) are lit on.
• Push the green PDB START button (item 9) on the PDB front panel.
If the Large Display Monitor option is installed:
1 X-ray ON Indicator
2 Audible Signals
- X-ray timer
- X-ray on
3 System RESET
4 System ON-OFF
5 X-ray Timer Reset
6 Auto Inject ON/OFF
No selection is allowed during Record or Fluoro acquisition. Only Monitoring Bar can be opened by clicking on one
of the monitoring icons.
2-1 Monitoring Bar
The monitoring bar consists of the seven icons, which provide information about the state and availability of the
following functions: ECG Signal, Disk Space, Heat Units, X-ray Timer, Fluoro, Record, X-ray disable button.
ECG Signal Disk Space Heat Units Sta- X-ray Timer Fluoro Ready Record Ready X-ray Disable
Status Button Status Button tus Button Status Button Icon Icon Button
Note: Always check that enough disk space is available before starting a procedure.
2-1-3 Heat Units Status
Click on the Heat Units Status button to display Heat Units Information.
Click on the X-ray Timer Status button to display the elapsed X-ray time since last timer reset.
The X-ray button allows disabling or enabling X-rays, outside an acquisition, by pressing the button.
The X-ray button has three states:
• INACTIVE state: during an acquisition, the button is inactive:
iLinq: Access to direct interactive communication with GEHC via In Site Connection.
Service: Access to a panel of menus such as Quality Assurance Program (QAP), InnovaSpin Parameters,
configuration and Service...
When selecting Service, a pop up window opens:
Fill in the patient information. Patient Name and Identification are required fields. If they are left
blank, an error message appears.
information that has been entered for a procedure. Also, it allows the modification of this information.
• New and editable fields will appear and can be modified while the original patient information is still shown on
the left side.
Note: If the patient information was retrieved from a worklist, it cannot be modified and the patient and identification
fields will remain grayed out. Only information manually typed in can be modified.
• Click on [Apply] to accept the changes.
• Click on [Cancel] to exit with no changes.
• Click on [Start Exam] to begin the procedure.
The protocols are classified inside categories. The list of protocols is organized in a 2 levels hierarchy.
A folder list displays and selects protocols.
5-1 List of Categories
Find below the protocols available:
Abdomen
Cardiac & Combo
Chest
EP DoseSense
Head
Lower Limbs
Pediatrics
Surgery (only with the surgical system configuration)
Upper Limbs
Vascular
X-ray Quality Test
Pelvis
Popliteal
Pediatrics Pedia Abdomen
Pedia Cardiac
Pedia Cardiac Neonate
Pedia Cerebral
Pedia Chest
Pedia Lower Limbs
Pedia Upper Limbs
Surgery 1 Surgey dose limited
2 Single shot
Abdomen
Chest
Extremities
Head/Neck
Upper Limbs 1 Upper Limbs dose limited
Elbow-Forearm
Hand
Shoulder-Humerus
Vascular DSA 0.3 Focal Spot
REC 0.6 Focal Spot
REC 1.0 Focal Spot
REC 2 fps-20 s, Fl 15 fps
X-ray Quality Test Quality Assessment Cardiac
Quality Assessment DSA
Use this protocol only in case a navigation system using electromagnetic waves is required
during a cardiac procedure. Otherwise Fluoro image quality might not be optimum.
Due to the large size of the 40cm detector, we recommend to avoid performing cardiac
procedure requiring the use of a navigation system using electromagnetic waves on the
Innova IGS 540. In such case, the artifacts created by the navigation system are not
compensated which can significantly lower the Fluoro image quality.
• On the Innova IGS 530, to optimize the visibility of soft tissues in the brain during intra cranial procedures, it is
recommended to select/use the "Soft Tissues" protocol located in the "Head" anatomy when a 3DCT acquisition
at 10°/s is required using the 30 cm field of view. The SID being reduced for such 3DCT acquisition, particular
attention must be brought to well center the patient's head in lateral to avoid risk of collision between gantry and
table top.
• To adjust the edge enhancement filter level used by default in a protocol, refer to section "Edge Enhancement
Filter Selection".
All acquisition protocols are preloaded with parameters to deliver a standard image quality. It is always possible to
modify the preloaded protocol parameters in order to customize/optimize the image processing parameters in
Record, Fluoro NoSub, Fluoro Sub, Fluoro Roadmap and Blended Roadmap. The strength of temporal and spatial
denoising, the brightness, contrast and sharpness of the images can be fine-tuned according to individual
preferences. Blended Roadmap also allows modifying the wire contrast over the vessel and the default vessel
density. If needed, contact your GE Healthcare Representative for any acquisition protocol
modification/optimization.
Any parameters, protocols and/or application changes to GEHC Equipment shall be Customer's sole responsibility
as well as any decision having regard to such clinical responsibilities and duty of care owed to patients.
Accordingly, the Customer will be solely liable for the decision to modify the protocol and the consequences
thereof. GEHC shall in no event be held liable for any damages resulting directly or indirectly from such
modification by the Customer of the protocol, as such modification will be made on Customer's instructions only.
5-2-1 Dose limited protocols
The system is delivered with one dose limited protocol for each of the following protocol categories: abdomen,
cardiac, chest, head, low limbs, surgery and upper limbs.
A dose limited protocol is set with following parameters:
• Fluoro: Receptor Dose Limited Plus autoexposure; 15 fps; Low detail.
• Dynamic record: Receptor Dose Limited Plus autoexposure; 15 fps; Low detail.
• DSA record : Receptor Dose Limited Plus autoexposure.
The predefined dose limited protocol is at the top of the list in each protocol category and have "dose limited" in its
name. It is selected by default upon system boot or reset, but this function can be changed.
It is recommended not to rename or modify these protocols.
If a dose limited protocol is modified, there is a risk of delivering higher dose than how the protocol is to be used.
Depending on local regulation, the large footswitch cover may be mandatory. It is recommended to install the large
cover on the footswitch located in remote position.
1 Fluoro
2 Record
3 Table Top Longitudinal and Lateral Brakes Release.
Note: With the InnovaIQ Table, the activation of the switch will only release the lateral table top brake when the
table is tilted.
Do not use the footswitch cover as a footrest or stand on it. The cover may break and lead
to uncontrolled X-ray.
A green led located on the gantry (ready for X-ray LED below) indicates when X-rays are available.
- THE BATTERY IS FULLY CHARGED. THE BATTERY CHARGE DURATION TIME IS ABOUT 8 HOURS.
- A REGULAR MAINTENANCE CHECK IS ENSURED BY PERIODIC PREVENTIVE MAINTENANCE (REFER
TO YOUR SERVICE CONTRACT)
In case of main power failure, the Innova system will automatically switch to UPS mode within 2 s maximum.
The message "Powered with UPS" displayed at bottom of the live display will inform users that the Innova system
is working under UPS.
Note: In case of main power failure during a Record acquisition, the acquisition will be aborted but already acquired
images will be saved on the image disk.
While Innova system is running under UPS mode:
• When less than 3 minutes of UPS power are available, the message "!!! 3 min before X-ray inhibit. UPS low.
Secure patient (1.9)" is displayed on the live display for 1 min.
• When less than 2 minutes of UPS power are available, the message "!!! 2 min before X-ray inhibit. UPS low.
Secure patient (1.9)" is displayed on the live display for 1 min.
• When less than 1 minute of UPS power are available, the new message "!!! 1 min before X-ray inhibit. UPS low.
Secure patient (1.9)" is displayed on the live display for 40 s.
If the main power is back before the UPS battery is discharged, the Innova system automatically switches off from
UPS mode within 2 s maximum.
In case of main power recovery during a Fluoro acquisition, the message "Power restored. Release fluoro for full
functionality" is displayed. The Innova system will switch off from UPS mode only after the end of the Fluoro in
progress.
Notes:
Regular Preventive maintenances are required to ensure the availability of the fluoro UPS option. In addition, the
UPS device has the ability to detect cases of undercharged batteries and some failure conditions of the UPS. In
both cases, the message "!! Exam interruption risk if power is lost. UPS failure. Call Service (2.15)" will be
displayed.
This message will be cleared if the batteries are recharged, or if the UPS failure condition disappears. Note that if
the operator presses the Mute button on the UPS, this message will also disappear while the error condition
remains and will remain unnoticed until the next UPS automatic test. It is therefore important to contact GE service
when pressing the Mute button on the fluoro UPS.
• If the Innova system does not recover full acquisition functionality after the main power is restored, reselect the
protocol in use and resume.
• All Maintenance and Service of the UPS must be performed by qualified personnel only.
• In case of UPS storage for more than 3 months, recharge UPS battery every 3 months to prevent any battery
damage.
is displayed at the bottom of the control room console to show that a fluoroscopic acquisition is in
progress.
After each Fluoro exposure, a Last Image Hold (LIH) is displayed on the live display.
7-1 Field of View (FOV)
There are four available FOV settings:
• 20 cm, 17 cm, 15 cm and 12 cm for Innova IGS 520.
• 30 cm, 20 cm, 16 cm and 12 cm for Innova IGS 530.
• 40 cm, 32 cm, 20 cm, and 16 cm for Innova IGS 540.
The FOV can be changed on the Innova Digital screen (see Illustration FOV - Innova Digital Screen) or on the
tableside control (TSSC).
For Innova IGS 520: 20 cm = 0 Mag, 17 cm = 1 Mag, 15 cm = 2 Mag and 12 cm = 3 Mag.
For Innova IGS 530: 30 cm = 0 Mag, 20 cm = 1 Mag, 16 cm = 2 Mag and 12 cm = 3 Mag.
For Innova IGS 540: 40 cm = 0 Mag, 32 cm = 1 Mag, 20 cm = 2 Mag and 16 cm = 3 Mag.
There are + and - symbols to allow for ease of up/down selection on the tableside control (see Illustration FOV -
TSSC, Item 4).
FOV - TSSC
The FOV can be changed dynamically during fluoroscopy without releasing the pedal. This provides the clinician
improved flexibility and workflow on the Innova IGS 520, Innova IGS 530, Innova IGS 540. The default FOV is 20
cm on Innova IGS 520, 30 cm on Innova IGS 530 and 40 cm on Innova IGS 540.
7-2 Fluoro Detail
There are two settings Low and Normal that are selected at tableside on the TSSC. See Illustration Fluoro Details -
TSSC (Item 13) or on the Innova Digital screen see Illustration Fluoro Detail - Innova Digital Screen.
Note: The level of fluoro detail selected is always displayed with the Fluoro parameters on the reference display
while in Exam.
Fluorostore - TSSC
Depending on the Fluoro frame rate used, the Fluorostore function allows to store up to:
• 15 s of Fluoro when performed at 30 fps.
• 30 s of Fluoro when performed at 15 fps.
• 60 s of Fluoro when performed at 7.5 fps.
• 120 s of Fluoro when performed at 3.75 fps.
The stored images are added into the DL database as a new Fluoro sequence.
The looping display starts automatically after the sequence is stored and enables better visualization of low
contrast objects thanks to repeated display without repeated exposures to the patient.
The stored Fluoro sequence has the same basic review capabilities (autoreview, Play/Pause...) as typically used
for Record sequence.
The Fluoro sequence is shown in the sequence browser with a specific icon (see Illustration - Fluoro Sequence
icon) and has limited processing capabilities (measurements and quantifications are not allowed on such fluoro
images).
Fluorostore mode.
– The autostorage selection will automatically be deactivated at End Exam.
7-5 X-ray Timer
The X-ray TIME RESET button (Item 17 lower left corner) will blink and reset the X-ray timer.
This timer counts the cumulated time while the Fluoro foot switch is depressed.
Note: For US only (Compliance to 21 CFR 10.20.32), this timer counts the total cumulated Fluoro and Record time
since last timer reset.
When the timer reaches 5 minutes, a warning tone is audible during Fluoro.
Press the button to reset the X-ray timer after the elapsed duration, or at any time, depending on local regulation.
The timer can be reset either at the tableside control or at the Innova Console.
In case of continuous fluoro without interruption, X-ray will be terminated after 10 min and the error message
"Fluoro interrupted: reset Fluoro timer to continue" will be displayed.
Note: In Italy, after 10 min of cumulated X-rays, X-rays are disabled until the X-ray timer is manually reset.
7-6 Cumulated Time
It is reset to zero at end Exam. If the exam is resumed from an existing exam, the value is set to the total X-ray time
(Fluoro and Record) recorded in this exam.
IN CASE OF SYSTEM FAILURE, THE ROTATION OF THE IMAGE DISPLAYED ON THE LIVE
DISPLAY MIGHT NOT BE AUTOMATICALLY CORRECTED TO KEEP THE "HEAD UP"
DISPLAY WHEN THE L ARM IS OUT OF 0° OR 90°. IN CASE SUCH CONDITION OCCURS,
AVOID TO START PROCEDURE WHICH REQUIRES A PERMANENT PATIENT "HEAD UP"
DISPLAY.
By default the [Image Rotation] key is active. Deselect the [Image Rotation] to prevent the image to automatically
rotate 90° when the L-Arm is moved more than +/- 45°. In such case, the image will be displayed "head up" only
when the L-Arm will be at 0°.
Patient Position will automatically default to Head First Supine when End Exam is selected.
Patient Position in the exam room
The current selected Patient Position is displayed in the exam room on the live monitor.
If the displayed patient position on the live monitor does not match the actual patient position in the exam room,
correct the patient position on the DL screen.
If the table is 180° rotated, an icon “180°” indicating the table rotation state will appear on the live monitor.
The patient position relative to the table top does not change when the table is rotated so there is no need to
reselect a patient position on the DL screen.
However, the patient position relative to the imaging system is the opposite of the display (patient position display
shows his head under detector, while his feet are actually under the detector).
Note: the patient orientation markers (L, I, R) are described in the "Patient Orientation markers" section below.
Patient Position at start exam
When clicking on Start Exam, a display of the current selected patient position on the DL screen will alternate with
a question mark reminding the user to verify the correct patient position has been selected. This effect will stop
upon selecting a patient position from the list or after 30 seconds.
1 - Patient's anterior (A) is the dominant direction and the patient's right side (R) is also viewed.
2 - Patient's superior (S) is the dominant direction
3 - Patient's left side (L) is the second dominant direction
4 - Patient's anterior (A) direction is also viewed but it is the least important.
Superior/Inferior or Head/Feet display
By default, the Superior/Inferior patient direction is indicated by letters “S” and “I”. However, there is a configuration
to display this direction with the letters “H” (Head) and “F” (Feet) instead. Contact your GE representative to change
this setting.
In the Preferences Tab on the DL screen, “Patient Orientation Preferences” allows the display or not of the patient
orientation markers. This configuration is accessible at any time except during an acquisition.
When the configuration is set to “Yes”, the patient orientation markers on acquired sequences or photos will be
displayed and transferred to workstations.
When the configuration is set to “No”, the patient orientation markers will not be displayed or transferred. If the
configuration is “No” and you need to retrieve the patient orientation of your acquired sequences, click on “Yes” and
re-launch the sequence or photo review. The markers will appear. If the sequence or photo without patient
orientation was already transferred to a workstation, they will need to be re-pushed.
Note: the change of display preference will not affect the images that are in the queue for transfer.
Modification of Patient Orientation
If the patient position on the table was not selected properly, the patient orientation markers will be incorrect. In this
case, it is possible to manually annotate the images through the Modify Patient Orientation interface.
Select the sequence or photo to modify. To select multiple sequences or photos, press and hold the shift key while
selecting.
Right click on your selection then left click on the “Modify Patient Orientation” dialog box.
Click on the box that matches the correct patient orientation for the selected images or the Clear Orientation box.
Click on Apply. The change will apply to all frames in a sequence or photo, even if the patient orientation changes
during the acquisition.
Any sequence or photo that is in the queue for transfer cannot be modified.
If the sequence or photo containing incorrect patient orientation markers was already transferred, they will need to
be re-pushed.
7-9 Image Flip
Image Flip will be automatically removed at each Start Exam (default image flip settings at start exam = no or
flip selected).
The icon will be also displayed in the Geometry area (Item 1) of each live display and over any image
reviewed on each live or reference display (Item 1).
- Perform a fluoroscopy depressing the fluoro footswitch; this is the mask acquisition phase.
The Fluoro icon is displayed on the left side of the live display during Fluoro to show that
acquired.
- Depress again the fluoro footswitch. The displayed fluoro images are subtracted. The vessel roadmap should
be visible in white color.
- The subtraction level can be adjusted using the LANDSCAPE function. Select an appropriate "Landscape"
level in the Fluoro window or use the "LANDSCAPE" button on the Table Side Status Control (TSSC).
DURING THE ROADMAP FLUOROSCOPY, DO NOT MOVE THE TABLETOP AND / OR THE
GANTRY NOR CHANGE THE FIELD OF VIEW, THE FLUORO DETAIL, THE IMAGE FLIP
SETTINGS, INSERT OR REMOVE CONTOUR FILTERS. IF DONE THE IMAGE QUALITY WILL
BE SERIOUSLY DEGRADED.
Note: The mask is aborted when Roadmap mode is exited, or when "END EXAM" is activated or when another
protocol is selected.
Note: When the roadmap fluoro is deselected, the landscape function is disabled.
DL Console - Roadmap
TSSC
When Blended Roadmap is selected, the label Blended Rdmp is displayed on the reference display.
Reference display
Note: When trying to enter Blended Roadmap, an error message may pop up in the following cases:
In this case, first select a good vessel image / change the settings, then re-enter Blended Roadmap.
9-3 Reposition Gantry/Table
The vessel information blended with the live fluoro is relevant only if the live fluoro is taken from the same
angulations as the vessel photo. After activating Blended Roadmap, the vessel image position and the current
system positions are compared and if needed, tools are provided to help repositioning of the gantry and the table
back to where the vessel image was acquired.
9-3-1 Systems equipped with an Omega table
• If the gantry position has changed, the system automatically triggers the Send Angle function. Activate the Auto
Positioner joystick (top or bottom) to drive the gantry back to where the vessel mask was acquired.
• If the table position has changed, the system automatically displays the table repositioning tool. Manually adjust
the axis of the table until a check mark is displayed next to all axis.
Note: A check mark is displayed when within a range of 5 mm for longitudinal, lateral and vertical movement
and 0.3 degrees for rotation.
• If neither the gantry nor the table was moved, no specific action is required. Perform a fluoro acquisition (see
next step).
See illustrations below.
Need to reposition the Gantry Need to reposition the Table No repositioning needed
Need to reposition the System Need to reposition the Rotation No repositioning needed
Note: Blended Roadmap Send Angles may fail or be cancelled, and an error message may pop up in the following
cases:
fluoroscopic image.
AFTER ENOUGH FRAMES HAVE BEEN ACQUIRED FOR 1 OR 2 SECONDS, THE SYSTEM
AUTOMATICALLY STOPS THE ACQUISITION.
Release the fluoro footswitch. A "Vessel" icon is displayed on the left hand side of the live display, in
order to show that the system is ready for a Blended Roadmap acquisition, and the percent of vessel density.
LIH shows the subtraction between the fluoro mask and the vessel photo to verify the registration of the subtracted
roadmap.
See illustration below:
If neither gantry or table was repositioned, and the patient did not move, registration should not be necessary.
because misregistration artifacts may occur (subtraction of images of different sizes). If FOV change is needed, exit
Blended Roadmap, change FOV, re-enter Blended Roadmap and acquire fluoro mask in the new FOV.
See the illustration below.
If a vessel photo is taken with a smaller FOV than live fluoro, then the available vessel information does not extend
to the whole fluoro image.
See illustrations below
Note: In that case, the registration image looks like the following; the borders of the fluoro image appear
unsubtracted.
and, using the mouse, click on the Record acquisition mode needed to perform the next acquisition.
All Record parameters are automatically set and the Innova system is ready to acquire using the selected mode.
• Innova 3D (3D) mode / Innova CT (3DCT) mode: primary use for 3D reconstruction and display of any region of
interest.
The Record icon is displayed on the left side of the live display during Record acquisition to show that the
displayed image is a live image.
The radiation icon is displayed at the bottom of the control room console to show that a record
acquisition is in progress.
13-1 DSA Mode (Option depending on system configuration)
• Select among "Abdomen, Chest, EP DoseSense, Head, Lower Limbs, Upper Limbs, Vascular" from the
"Protocols". Double click on the anatomy name to open it.
Then click on one procedure among the proposed choice.
The DSA acquisition is divided into 3 segments. A different frame rate and duration can be set for each of the 3
segments.
During acquisition, a dedicated button located on the side of the Record hand switch allows to commute from the
current segment to the next one before the total duration of the current segment.
• In the Record window DSA tab, select the frame rate needed for each of the three segments using the frame
rate pull down list between 0.5; 1; 2; 4 or 7.5 fps.
• Select the duration of each segment using the duration pull down list between:
– 1; 3; 5; 10; 15; 20; 30 s for the first segment.
– 0; 1; 3; 5; 10; 15; 20; 30 s for the second segment.
For the third segment, the duration list is dynamic and the maximum allowed duration is
When Auto Inject is selected (injector automatically triggered by the Innova system), the
X-ray or Injection Delay must be set (or kept) at Zero on the injector unit. In order to prevent
any conflict between the injector unit and the Innova system, X-ray or Injection Delay must
only be selected from the Control Room Innova Console. Setting (or keeping) some X-ray or
Injection Delay on the injector unit itself could create some acquisition abort.
In case of Automatic injection selection, the injector is automatically triggered by the system.
When Auto Injection is selected, it is possible to preset an injection delay (X-ray then Injection start) or X-ray delay
(Injection then X-ray start) by opening the selected pull down list.
Click on the +/- field to open the Injection/X-ray delay edit window and add/remove new/existing values.
To add a new value in the list, type the value and click on .
To delete an existing value, open the pull down list, select the value to remove and click on .
From this window, it is also possible to enter a new delay to use for the next acquisition. Just enter the value and
click OK to close the edit window and apply the new value. In this case, this new value will not be saved and added
to the list.
Note: New Injection or X-ray delay must be entered by step of 0.5 s only (i.e. 1 s, 1.5 s, 2 s...).
• Perform DSA acquisition
Depress and hold the hand/foot switch to initiate the acquisition.
Acquisition parameters are computed from the previous fluoro. In case of table/gantry motion, FOV change,
geometry modification... since the previous fluoro, the Record acquisition could be inhibited and a new Fluoro
required.
If such situation occurs, perform a short fluoro (2 s minimum) then depress the acquisition switch again.
Note: Always check the image quality at the beginning of the run before launching the injection (images shall not
be dark nor contain excessive saturation).
Note: If the expose trigger of the hand/foot switch is released during acquisition, X-ray and injection will stop.
In case of Manual injection, "Inject" will be displayed on the left side of the live display to inform when injection can
start.
The acquisition starts using parameters set for segment 1, then those set for segment 2, then those set for
segment 3.
In case it is required to commute from one segment to the next one before the complete duration set for that
segment, pressing on the Skip Segment button located on the side of the record hand switch will make the
acquisition switch to parameters set for the next segment.
- The Heat Units Information displayed are only estimations made by the system depending
on acquisition parameters computation. - The Heat Units Information are available from the
control room Innova Digital screen Heat Units window and from the reference display. At the
beginning of a DSA sequence, the system will automatically check the possibility of the
requested run. If the entire run duration is not possible because of thermal limitation (tube
too hot), then a 5-seconds long popup is displayed on the frontal live display, informing the
user of the maximum possible run duration. If the user does not release the pedal, the run
starts and will end when reaching that duration. Before a Record acquisition is performed, it
is the responsibility of the user to check the Heat Units informations. Depending on the
information displayed, the next Record acquisition may or may not be completed as
requested by the user.
Depending on acquisition parameters computation and available heat units, the system will display a Warning pop
up on the frontal live display in case the requested run cannot be completed.
In this pop up, the computed available run time is displayed and a 5 seconds count down starts to let the user take
the decision to perform or not the acquisition, knowing that the run will be aborted before its completion.
Keep the hand/foot switch pressed to allow the run to start.
Release the hand/foot switch within 5 seconds (before count down reaches zero) to abort the acquisition.
Change acquisition parameters or wait for the X-ray tube to cool down to succeed in acquisition completion.
Note: Reaching X-ray tube thermal limit will interrupt the acquisition and require to wait for X-ray tube to cool down
or to modify acquisition parameters. Fluoro is always available. After reaching the thermal limits, the subsequent
DSA acquisition will have a 10 seconds access time.
13-2 Dynamic Mode (Option depending on system configuration)
Note: Before performing the first cardiac procedure after an angio procedure, check the Heat Units Status window
for heat units availability. It should be at least 50% for a diagnostic cardiac case or 70% for an interventional
cardiac case. The subsequent cardiac procedures can be performed without any thermal limitation. Fluoroscopy is
always fully available.
• Select "Cardiac & Combo" from the "Protocols". Double click on the anatomy name to open it.
Then click on one procedure among the proposed choice.
In addition to specific protocols such as Coronaries, EP, Needle Placement, Pedia Cardiac, Pedia Cardiac Neonate
and Ventriculogram, two protocols labeled Cardiac 1 and Cardiac 2 can be modified at installation to better fit local
needs during cardiac procedures. Two other protocols labeled Combo 1 and Combo 2 can also be modified at
installation to access other angio acquisition modes needed for angiography and peripheral procedures.
Automatically the Acquisition tab confirms the selected acquisition mode.
• In the Record window Dynamic tab, select the Frame Rate needed using the Frame rate pull down list between
30 or 15 fps.
Select Auto injection to automatically trigger the injector from the Innova system (for more detail, see Innova Digital
/ How to perform a Record Acquisition / DSA Mode).
• Perform Cardiac acquisition
13-3 InnovaChaseTM Mode (Option)
• After selection of an angio protocol, select the Chase acquisition mode from the Current Application window
located on the left side of the Innova Digital screen. If the selected protocol does not offer this choice, select
another protocol from the "Protocols".
• In the Record window, select the level of detail and injection parameters.
For more detail about injection parameters, refer to Innova Digital / How to perform a Record Acquisition / DSA
Mode).
• Perform InnovaChaseTM acquisition
InnovaChaseTM Acquisition is available using only:
5 fps acquisition frame rate.
1.0 large focal spot.
During the bolus acquisition, the table top will move between the Start and End positions at
high speed. Always check that all tubings (injector, pressure...) connected to the Patient will
be long enough and will have an extra length of at least 15 cm when the table will reach the
Start or End position. Failure to this recommendation could cause severe injury to the
Patient at punction location.
13-4-2-3 Anticollision Precaution
Make sure to clear the area before launching the procedure to avoid any activation of the
anticollision system that would abort the X-ray acquisition and injection.
13-4-2-4 On the Innova Console
• After selection of an angio protocol, select the Bolus acquisition mode from the Current Application window
located on the left side of the Innova Digital screen. If the selected protocol does not offer this choice, select
another protocol from the "Protocols".
• If the table top is in the allowed range, the S/E button of the Table Control Panel lits.
Any protocol selection performed after the Start/End positions were memorized will lose
these positions.
• Select the acquisition parameters:
- Available frame rates at rest (when the table top does not move) are: 1 fps and 2 fps.
- Step length is fixed to 5 cm.
- Auto/Manual injection with X-ray/Injection Delay.
- Normal or Low Record Detail.
Note: Press once on the S/E button to store the Start position.
Press a second time on the S/E button to store the End position (there must be at least 30 cm between Start and
End positions).
Press a third time on the S/E button to reset (clear) the Start and End positions.
Note: A 15 minutes time out is allowed between:
- Memorization of Start and memorization of End position.
- Memorization of End position and beginning of bolus acquisition.
13-4-3 Making an Acquisition in Bolus Mode
13-4-3-1 On the Innova Console
• Select the Bolus Chase protocol and the appropriate acquisition parameters (refer to Innova Digital / How to
perform a Record Acquisition / InnovaBreezeTM Mode (Option) / Before You Start / On the Innova Console).
13-4-3-2 In the Examination Room
• In order to always get the maximum field coverage on the patient even using a small field of view, move the
table top up to the end and keep the Digital Detector close to the patient.
• Move the table top to center the patient over the abdomen. The S/E button of the Control Panel must be ON to
allow to store the table position. Press the S/E button to memorize the table Start position. The S/E button goes
OFF.
• Correct centering is checked using fluoroscopy, from the abdomen to the feet.
Longitudinal movement is the only movement available. When the table top is moved more than 30 cm from the
Start position and is in the allowed range, the S/E button blinks.
• After centering over the feet, press the S/E button to memorize the table End position. The S/E button goes
OFF.
Note: From the time the Start position is memorized to the time the exposure switch is pressed all gantry axis are
locked except:
- Table Up/Down.
- Digital Detector Up/Down.
- Longitudinal table top.
Collimation, Contour Filters and Field of View selection are available until the exposure switch is pressed.
Fluoro is always available until the acquisition starts.
13-4-3-3 At the Console
1 Table Speed
2 Bolus Button
• When the table speed button is released, the table stops on the next Mask position. Images are then acquired at
the frame rate at rest.
• To optimize the last acquisition, the table will stop at 4 cm maximum further than the memorized End position.
• To stop the X-ray Run:
- Release the prep/expose Innova handswitch button at any time or,
- Press then release the Bolus button on the Bolus Chasing handswitch after the table has reached its stop
position.
Note: Until the Bolus Chase protocol is exited or the S/E button is depressed to clear the Start and End memorized
positions, the system is still ready to perform another identical bolus acquisition.
A 15 minutes time out will automatically clear both memorized table positions if no action is perform on the system
after the completion of the bolus.
13-5 Innova 3D, Innova Subtracted 3D and Innova CT Modes (Options)
13-5-1 Application Description
Depending on system configuration, three 3D Spin modes can be available.
Innova CT is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide
images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up.
Innova CT is intended for imaging bone and soft tissues as well as other internal body structures.
Innova 3D is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to assist the
physician in diagnosis, surgical planning, interventional procedures and treatment follow-up.
Innova Subtracted 3D reconstructs 3D volumes from 2 consecutive rotational acquisitions to provide images that
help clinicians to quickly visualize vessels without the need to remove surrounding bone, tissue and implanted
devices during procedures using angiography for aneurysm coiling, stent insertion, vessel embolization, etc.
Innova CT is intended for imaging bone and soft tissues as well as other internal body structures.
Innova 3D / Innova Subtracted 3D / Innova CT images should not be used on their own for
diagnosis or treatment but always in complement of 2D images (live or recent recorded
images).
The Innova 3D and Innova CT reconstruction is carried out on the Advantage Workstation and allows
reconstructions from a single 200° (approximate) spin acquired at a frame rate of 30 fps.
The Subtracted 3D reconstruction is carried out on the Advantage Workstation VS4 release or above, and allows
reconstructions from two consecutive 200° (approximate) spins of approximately 150 images each acquired at a
frame rate of 30 fps and a rotation speed of 40° per second.
All reconstructed 3D models can have 5123 resolution. A 2563 resolution mode is also available when faster
reconstructions are needed.
3D models produced can be visualized on the Advantage Workstation using the Volume Viewer Plus software.
The Volume Viewer software offers the ability to:
• Manipulate the 3D model around all three axes.
• Perform distance, and volume measurements.
• Navigate inside vessels (with Navigator option).
• Display MPVR (Multi Planar Volume Reconstruction) cross-sections: axial, sagittal, coronal and oblique
cross-sections.
• Detect vessels and remove un-wanted structures from the image.
13-5-2 3D Spin Acquisition
The 3D Spin Acquisition procedure is performed in two steps:
• 3D calibration acquisition using the helix phantom for definition of volume reference points (to be performed
once every six months).
Two levels of information message are displayed:
– 6 months-15 days: "3D Calibration near to expire. Call Service to perform 3D Calibration".
– After 6 months: "3D Calibration out of date. Call Service to perform 3D Calibration".
3D acquisition and reconstruction are still allowed.
A small "*" will be displayed left to the reconstruction date of the 3D model on the Advantage Workstation.
• Acquisition of patient images using a specific 200° (approximate) rotation protocol around the patient.
13-5-2-1 3D Spin Images Acquisition
• In Innova 3D mode, the rotation is performed at 40°/s, at a frame rate of 30 fps, providing approximately 150
images in a 5 seconds acquisition.
• In Innova Subtracted 3D mode, two rotations, "Mask" and "Contrast" are performed at 40°/s, at a frame rate of
30 fps, and provide approximately 300 images in approximately 20 seconds.
• In Innova CT mode, 3 rotation speeds are available: 40°/s, 20°/s and 10°/s, also at a frame rate of 30 fps,
providing approximately 150 images, 300 images or 600 images respectively. Using more views allows to
improve the visualization of bones and soft-tissues.
Note: As the source images are acquired during the 3D spin, several detector and processing corrections are
applied and tuned to optimize resulting 3D image consistency during reconstruction. This process may result in
the presence of certain anomalies visible while reviewing the source images. The potential anomalies may
include a horizontal line or contrast change across the middle of an image and/or image-to-image contrast
oscillations. In general these anomalies are generated by rapid corrections taking place to account for the
presence of raw radiation within the field of view during the spin. The degree of visibility of these anomalies will
depend on several factors related to the patient, system and geometry of the spin. These anomalies appear
solely in the 2D spin image and allow for an improved display of the 3D model.
13-5-2-2 Patient Positioning
Using Fluoro in PA and Lateral projections, properly center the region of interest (ROI) at isocenter.
Special attention must be paid to the patient immobilization during the 3D Spin acquisition
sequence. Any patient motion during the fast rotation of the gantry could injure the patient.
It is advised, particularly if the patient is not anaesthetized, to use appropriate restraining
devices to completely immobilize the patient (head, arms, legs) in order to keep the patient,
and so the region of interest, perfectly still throughout the acquisition sequence. Any
patient motion will also reduce the quality of the images and the 3D reconstruction. Explain
the procedure to the patient and ask him not to move prior to performing a 3D spin images
acquisition.
Use appropriate collimation to limit patient radiation and cut raw radiation.
Note: Because contour filter blades can create artifacts on 3D models, blades are automatically parked as soon as
a 3D acquisition mode is selected.
13-5-2-3 On the Innova Digital Screen
For an Innova 3D acquisition, select the 3D acquisition mode in the Acquisition Mode tabs area. If the selected
protocol does not offer this choice, select another angio protocol from the Protocols.
For an Innova Subtracted 3D acquisition, select the 3D acquisition mode in the Acquisition Mode tabs area. If the
selected protocol does not offer this choice, select another angio protocol from the Protocols.
For an Innova CT acquisition, select the 3DCT acquisition mode in the Acquisition Mode tabs area.
Use appropriate injection parameters (volume, rate, delay) in order that vessels are properly
opacified during the entire image acquisition phase.
On the Innova IGS 530, when using the "Soft Tissues" protocol, the SID is reduced to 108
cm (42.5") for 3DCT acquisition performed at 10°/s to optimize the visibility of soft tissues in
the brain. - Make sure that the table top is centered in lateral to avoid collision during the
gantry test rotation. - Use such protocols with reduced SID for intra cranial 3DCT
acquisition at 10°/s only.
Make sure to clear the area before launching the procedure to avoid any activation of the
anticollision system that would abort the X-ray acquisition and injection.
Special attention must be paid to select the appropriate 3D spin mode for the current exam
(Innova 3D mode versus Innova Subtracted 3D mode) to avoid: - Unnecessary retake,
causing extra X-ray dose and injection to patient (when Innova 3D mode is selected instead
of Innova Subtracted 3D), - Or mechanical hazard to clinical staff arising from unexpected
gantry movement (when Innova Subtracted 3D mode is selected instead of Innova 3D).
At the end of the test rotation, the Test button goes Off and the message "Ready. Clear rotation area" is displayed
on the live display.
Note: From the beginning of the Test Rotation, all table and gantry axes are locked. Pressing on the Test button
after the completion of the Test Rotation will abort the 3D Spin set up and free all table and gantry axes. If needed,
the patient can be repositioned. A new Test Rotation is then required.
Press and hold the prep/expose Innova handswitch button:
Note: Except in case of emergency, this button must NOT be released during the entire 3D Spin
acquisition
• The first X-ray run, or Mask, starts at a speed of 40 degree/second. A message "Spin Acquisition in Progress" is
displayed in the bottom left corner of the live display.
• The gantry automatically goes back to the start position at a speed of 30 degree/second. A message "Returning
to Start Position" is displayed.
• The injection starts.
• The second X-ray run, Contrast, starts after the programmed x-ray delay. A message "Spin Acquisition in
Progress" is displayed in the bottom left corner of the live display.
• The review automatically starts subtracted at 30 fps. Mask, Contrast and Subtracted spins can be reviewed
separately.
Note: Stenosis Analysis is available on the subtracted images.
Note: The 2 spin of approximate 150 images ("Mask" and "Contrast") are concatenated into a single approximate
300 images sequence.
Note: Typical duration from the press of the exposure handswitch to the end of X-ray, is approximate 20 seconds
for an X-ray delay of 0 seconds.
When the gantry reaches the End position, X-ray are stopped and the Test button starts to blink again; the system
is ready for a new 3D Spin acquisition (new Test Rotation must be performed first).
All calibration data needed for the 3D reconstruction are saved with the images in the same sequence.
This will allow to reconstruct the 3D Spin acquisition anytime.
13-5-3 Image Network
After the 3D Spin acquisition is completed, acquired images are automatically transferred to the Advantage
Workstation.
Note: 3D images will always be sent to a predefined specific remote host (AW workstation).
Non 3D images will be sent to the remote host selected in the Host Selection screen.
3D images can be sent to 2 different hosts: the specific 3D host (AW workstation) and the host selected in the Host
Selection screen. In such case, the status of the 3D sequence in the DL Sequence Browser will change to sent
only after successful completion of the 3D sequence data transfer to both hosts.
13-5-4 3D Image Reconstruction
As soon as all acquired images are received at the Advantage Workstation, the 3D reconstruction is started
automatically.
Note: For detailed information on the use of the reconstruction software, refer to the Innova 3D XR Operator
Manual.
13-5-5 3D Model Display
The 3D model of vessels and/or interventional devices can be analyzed using a range of 3D analysis tools
including:
• MIP (maximum intensity projection).
• Volume Rendering.
• MPVR (multi-planar volume reconstruction).
• Axial, sagittal, coronal and oblique cross-sections.
• Electronic scalpel and 3D shutter.
• Distance and volume measurements.
• Navigator (endovascular-like).
• Customized screen layouts for Innova 3D and Innova CT.
13-5-6 3D Image Archiving
The 3D model is saved as a DICOM object, and as such can be archived on a CD-ROM. This model can later be
reviewed on any Advantage Workstation with the Volume Viewer Plus software.
Additionally, the user can archive the spin acquisition on PACS, then at any time later retrieve and reconstruct a 3D
model from it.
Archiving of "screen capture" images that are created from the 3D model is also possible on a PACS.
13-5-7 3D Quick Guide
13-5-7-1 3D Acquisition
• Isocenter the region to be imaged using lateral and frontal fluoro.
• When using the 30/32 cm or 40 cm FOV, use appropriate collimation to avoid raw radiation, which can degrade
the 3D image.
• Use collimator blades as required, but as the 3D image might be slightly smaller than 2D image, make sure to
keep an appropriate margin by NOT collimating too tightly.
• For optimum vessel contrast, injection should be as selective as possible.
• Vessels should be completely opacified from the beginning of the injected spin until the end.
– X-ray delay: If a DSA series is available, use the "scale" button on the DL keypad to display the time line and
measure the delay between injection start and vessel completely filled. Enter this delay as X-ray delay.
– Injection parameters: select volume and flow rate such that the injection duration is equal to the
above-computed X-Ray delay plus 5 seconds rotation duration.
– As a guideline, injection parameters should be equal or inferior to the one used for an unsubtracted 3D.
Subtracted 3D enables automatic bone removal, one can expect same vessel segmentation and visibility
with volume of contrast product inferior to the one used for a no sub 3D.
• Any patient motion during either gantry rotations (mask and / or contrast spin) or in between those two rotations,
will degrade the 3D image quality causing artifacts such as streaks or blurring. Informing the patient about the
gantry rotation and instructing the patient not to move, and if possible not to breathe (approximately 20
seconds), during the entire 3D acquisition will help. In case of cerebral procedures, ask the patient to close
his/her eyes during the spin/s. If possible with general anesthesia, stop the respirator.
• For intracranial vascular studies, it is recommended to use the 20 cm FOV to avoid the presence of skull in the
3D image. This simplifies the 3D processing on the AW and may require the skull to be manually removed.
13-5-7-2 3D reconstruction
• 3 modes are available to reconstruct 3D Spin images:
– To get optimum spatial resolution, reconstruct the 3D Spin using the "512" 3D volume size.
– To speed up the reconstruction keeping a very good spatial resolution, use the "Fast 512" 3D volume size to
reconstruct the 3D Spin. Reconstruction time is about 50% compared to "512" with a spatial resolution
slightly lower.
– In case of large amount of images to reconstruct or to get the shortest reconstruction time, reconstrut the 3D
Spin using the "256" 3D volume size. Reconstruction time is about 12% (8 time less) compared to "512" with
the lowest spatial resolution.
• Reconstruction filters
– Vascular: primary choice is None, Bone filter enhances 3D vessel sharpness but increases noise in
cross-section.
13-5-7-3 3D display
• When looking at a 3D image, keep in mind that it is a reconstructed image which may also contain
reconstruction artifacts. Always compare a 3D image with other patient data in case unexpected structures are
discovered.
• When using scalpel, select only the 3D images you are working on, otherwise black areas will be created in
cross-sections.
• A fast way to isolate and display a ROI is to use the Sphere Shutter (adjust shutter diameter as required) or the
Box Shutter.
• On cross-section images, a large slice thickness (between approximately 5 mm and a few centimeters)
improves vessel visibility.
• Influence of "VR Opacity" on the 3D image.
A Higher VR Opacity will provide smoother vessels, but will also increase the visibility of undesired structures
(bone, etc).
• When performing measurements, it is better to drop points on a cross-section of minimal thickness than on a 3D
view. For measuring vessel diameter, the cross-section must be orthogonal to the vessel.
• How to create a cross-section orthogonal to a vessel:
– The wrong way to do it:
Oblique cross-section
which is not necessarily
orthogonal to the vessel
axis (because in 3D the
vessel is not necessarily
parallel to the plane of
the screen).
MIP or VR image
MIP or VR image
1st Oblique cross- 2nd Oblique cross- Oblique cross-
section parallel to section perpendicu- section orthogonal
the vessel axis lar to the vessel ax- to the vessel axis
is
– When "projection (2D)" is selected, points are dropped in a plane and result refers to the distance between
these points in the same plane.
– When "from volume (3D)" is selected, points are dropped in the space along a virtual line. Points are
dropped at the intersection of these virtual lines with any anatomy found. Result refers to the distance
between these 2 points which can be located in different planes.
– The same visual distance between both points can display two completely different results.
- "2D" displays the length of the segment visualized on the image
- "3D" displays the length of the segment drawn in the space. In fact, the 3D segment is going from the small
vessel located on the front of the image to the large vessel located in the background of the image. This
oblique segment is much longer than it seems.
"Non-Sub 3D" after PVE Glue only (from mask) Vessels only (from "Sub 3D")
"Sub 3D merge"
"Drag & Drop" merge showing glue
and vessels
Exam Browser
If there is more than one Exam available for the patient, both of them can be reviewed from the Sequence Browser.
To activate this function, highlight the desired exams to review by holding down the Control key and clicking on
them with the mouse.
• Select the exam you want to review and double click on it. This brings up the SEQUENCE BROWSER window
and automatically launches the review of first sequence.
Sequence Browser
• In the SEQUENCE/PHOTO BROWSER window, double click on any sequence/photo that you wish to review.
A Sequence name is displayed on the image in the upper right corner during review.
The Review icon is displayed on the left side of the live display as soon as the review is launched to
show that the displayed image in review was previously recorded.
, , , are also displayed at the bottom right of the Live display to show that the displayed
image is either a previously recorded or stored fluoroscopic image reviewed in Play or Paused mode (review in
loop or frozen image).
If there are no Sequences acquired at this time, the Browser will be blank. As each Record Sequence is stored to
the digital disk, a new icon will appear with the number of the sequence.
A single click on a Sequence icon only activates it for other processing. The icon will be highlighted to indicate that
it has been selected. Selection can be for the following:
1. Name - Each Sequence can be named up to 8 characters. A menu window will open displaying a predefined
list. Adding or deleting from this list is allowed.
Sequence naming can be performed during Fluoro or Record acquisition.
2. Delete - this will delete the Sequence from the digital disk. It can not be retrieve after the OK for deletion has
been accepted. Multiple sequences can be deleted at a time by holding down the Control key and clicking on
each icon with the mouse. If a Sequence is deleted from a list, the numbering of the other ones is not changed.
3. Network - this will start the transfer of the Sequence to the Network. Multiple sequences can be networked at a
time by holding down the Control key and clicking on each icon with the mouse.
Sequence network can be initiated during Fluoro or Record acquisition.
If the Sequence icon has a Yellow box around it, this indicates that this is the Sequence that is presently being
reviewed.
Review of the Sequence can be accomplished by using the Keypad, Remote control or the Digital console and
mouse. None of these are active during an exposure.
Note: Reviewing images at frame rates higher than acquisition frame rates will cause some frames been skipped
during display.
Photo Browser
• In the PHOTO BROWSER window, double click on any photo that you wish to review.
A Photo name is displayed on the image in the upper right corner during review.
If there are no Photos stored at this time, the Browser will be blank. As each Photo is stored to the digital disk, a
new icon will appear with the number of the sequence. During an exposure, a message will appear "Acquisition in
Progress" on the Digital console. A single click on a Photo icon only activates it for other processing. The icon will
be highlighted to indicate that it has been selected. Selection can be for the following:
1. Name - Each Photo can be named up to 8 characters. A menu window will open displaying a predefined list.
Adding or deleting from this list is allowed.
Photo naming can be performed during Fluoro or Record acquisition.
2. Delete - this will delete the Photo from the digital disk. It can not be retrieve after the OK for deletion has been
accepted. Multiple Photos can be deleted at a time by holding down the Control key and clicking on each icon
with the mouse. If a Photo is deleted from a list, the numbering of the other ones is not changed.
3. Network - this will start the transfer of the Photo to the Network. Multiple photos can be networked at a time by
holding down the Control key and clicking on each icon with the mouse.
Photo network can be initiated during Fluoro or Record acquisition.
If the Photo icon has a Yellow box around it, this indicates that this is the Photo that is presently being reviewed.
Review of the Photo can be accomplished by using the Keypad, Remote control or the Digital console and mouse.
None of these are active during an exposure.
The empty catheter must be in the vessel to analyse and at the same level as the lesion to
quantify, else the measurement results displayed will be wrong. See Innova Digital /
Stenosis Analysis (Option) / Calibrating on a Catheter Diameter.
Note: Before starting acquisition, exit the Stenosis Analysis (SA) application to have access to the display of the
X-ray technique factors on the INNOVA Digital Screen.
17-1 Analysis Limitations
The software will not allow analysis under the following conditions:
• Image type is not the original acquired image (i.e. photos cannot be used to perform an analysis).
• Some functionalities only work in the following conditions:
– Stenosis quantification can be used on Dynamic, subtracted DSA, subtracted Bolus and subtracted 3D
images only (not available on InnovaChaseTM, 3D/3DCT images and Cardiac images acquired using a Field
of View larger than 20 cm).
Multi-segments and Distance Measurement functionalities are always available.
Note: Stenosis quantification and Distance measurement cannot be performed on Fluoro image: Fluoro LIH and
Fluoro sequence created using the Fluorostore function. In such case, a pop up will be displayed and the
application will fail.
1 Autocalibration
2 Catheter Calibration
3 Stenosis Quantification
4 Show/Hide Results
5 Hide Contours
6 Show Real Contours
7 Show Real and Ideal Contours
8 Multi Segments
9 Measure Distance
10 Zoom Image
11 Store Photo
12 Exit
(+/- 1.5 cm) as the catheter and for all the images acquired with the same geometric parameters of the X-ray
system.
The catheter calibration can be performed on cardiac images or non-subtracted digital angiographic images. If in
subtracted mode, the application will switch automatically to non-subtracted mode when clicking on the catheter
calibration button. Automatic catheter detection may not work on subtracted frames.
The imprecision displayed below the calibration factor is the estimated confidence interval around the result
returned by the application (in %).
For example, if you get a calibration factor of 0.1 mm/pixel with an imprecision of 10%, it means that the real factor
is between 0.09 mm/pixel and 0.11 mm/pixel.
17-6 How to achieve an accurate catheter calibration
• Select a frame where the catheter is completely empty: no contrast media and no guide-wire or any kind of tool.
Guide Line:
Always calibrate on a completely empty catheter
Watch the imprecision percentage displayed in the calibration result box: if higher than 20%, select another
frame and redo the calibration in order to get the lowest imprecision as possible.
• Enter the exact size of the catheter used.
The catheter must be at the same distance from the detector than the vessel to be measured.
A difference in distance of 5 cm may cause a 5% error in the vessel dimension. The catheter
must be as close as possible to the vessel to be measured in the image.
Reject those, which do not fit the external edges of the catheter.
• Select a catheter segment long enough (at least 2 cm on the screen) in the straight section of the catheter.
• Examine carefully the traced contours.
• Steps to calibrate a catheter
• Click on the icon for Catheter Calibration (2) on the Stenosis Analysis Toolbar.
The Input the Catheter Size dialog box appears.
Stenosis Analysis application shall only be used with catheter between 3 and 9 French (1-3
mm). Also, depending of the manufacturer, the given product catheter size may not be
accurate, else the measurement results displayed will be wrong. In such case, you may
possibly want to customize your catheter sizes. In all cases, it is recommended to check the
displayed catheter size before accepting the calibration results.
• Place two points within the catheter, as close as possible to the end but not on a curve, to define the section for
diameter detection.
• The system automatically finds and displays the catheter edges. The "Confirm the catheter detection" message
appears.
THE AUTOMATIC EDGE DETECTION PROGRAM WILL SHOW THE DETECTED BORDER OF
THE CATHETER. YOU SHALL CHECK THAT THE EDGES BEING DISPLAYED ARE
MATCHING THE CATHETER BORDER BEFORE ACCEPTING THE CALIBRATION RESULTS.
IF THE MATCHING IS NOT ACCURATE, PLEASE CANCEL AND RESTART A NEW
CALIBRATION.
17-7 Stenosis Quantification (available for Dynamic, subtracted DSA, Bolus and
subtracted 3D Modes only)
The basic principles of this functionality are the automatic detection of the vessel edges in a selected portion of the
artery, then the quantitative measurement of the vessel length and diameters along the selected segment.
The application is able to quantify vessels up to 5 mm on Cardiac images acquired on Dynamic mode with FOV
less than or equal to 20 cm and vessels up to 50 mm on subtracted DSA, Bolus and subtracted 3D images.
The contrast media used to differentiate the vessels from the patient background must be radio-opaque (iodine) so
that vessels appear as dark structures with respect to the background. Quantification is not applicable on
• Place two or more points within the vessel from the proximal to the distal part of the stenosis to define the
section for automatic detection. Double click the last point to start the calculation and get the results. To move a
point, drag it with left mouse button. To delete a selected point, press the Delete key.
Note: The ability to accurately measure a vessel can be hindered by:
• The presence of a stent or guiding catheter in the vessel.
• A calibration or quantification performed on edge or corner of the image (conic deformation of X-ray) can
degrade the accuracy by up to 7.5%.
17-7-1 In case of Auto-calibration
• If the image is not calibrated, the Auto-calibration dialog box opens asking you to enter the "object to table top
distance".
The goal of auto-calibration is to allow the software to compute the pixel size in a given image without using any
kind of reference object. Auto-calibration is entirely based on the known geometry of the imaging system. The
only parameter that the user has to specify is the Object to Tabletop distance defined as being the shortest
distance between the center of the object to be measured and the top of the table.
As opposed to catheter calibration, auto-calibration is available only after a region of interest has been defined
in the image. It means that auto-calibration can be launched only after vessel contours or segments are drawn
in the image.
Calibration results are automatically saved at the sequence level. If you launch the application on a previously
calibrated sequence, the calibration information will also be loaded as well as the sequence itself.
Enter the object (vessel) to table top distance or use the default value (15 cm for Cardiac image; 8 cm for Angio
image), then click OK to get results in metric units.
THE OBJECT (VESSEL) TO TABLE TOP DISTANCE IS DEFINED FROM THE TOP OF THE
TABLE TOP AND NOT FROM THE TOP OF THE MATTRESS. IF THE DISTANCE ENTERED
DIFFERS FROM THE ACTUAL DISTANCE, THE RESULTING MEASUREMENTS WILL BE
INACCURATE OF ABOUT 1% PER CENTIMETER OF DIFFERENCE.
In case a new auto-calibration is needed, after points were dropped, it is always possible to launch it manually
pressing on the Auto-calibration icon (1) on the Stenosis Analysis toolbar.
Upon completion of auto-calibration, Stenosis Analysis report is displayed with the computed results.
Notes: To accurately measure a vessel, note the followings:
• Input the correct distance the anatomy is from the table top.
• The greater the gantry angulation towards a lateral position, the higher the imprecision result will be. Avoid
using Auto-calibration for lateral images; use catheter calibration instead.
• Always take into account the imprecision displayed with each measurement.
17-7-2 In case of Catheter Calibration
The Results Box displays results of stenosis quantification and their imprecision.
The imprecision displayed after each measurement is the estimated confidence interval around the result returned
by the application (in mm).
For example, if you get a Reference Diameter of 2 mm ± 0.2 mm, it means that the real reference diameter is
ENSURE THAT THE MEASURED OBJECT IS IN THE SAME PLANE AS THE CATHETER,
PARALLEL TO THE DETECTOR AND TAKE INTO ACCOUNT THE IMPRECISION FOUND
AROUND THE CALIBRATION FACTOR.
Note: Only one Multi-segments can be displayed at a time.
17-11 Innova OneTouch QA (option available on the optional Innova Central touch
screen)
Using the Review Tab of the Innova Central touch screen located at table side, it is also possible to launch the
Stenosis Analysis application.
From the Review tab, select Stenosis Analysis (SA) to enter the application.
The most useful functions are available to quantify a vessel or measure a distance.
Note: Only Autocalibration is available at table side. In case of catheter calibration required, perform the calibration
from the control room console.
Launch the transfer of the image displayed on the live display in the Innova Central touch screen.
After the completion of the transfer, the central part of the image appears on the Innova Central touch screen.
And the displayed Region Of Interest (ROI) is visualized on the live display.
Use the Innova Central touch screen joystick to roam the image:
• Tilt the joystick left/right to see further left/right (the viewport moves left/right, the image moves toward right/left).
• Tilt the joystick up/down to see further up/down (the viewport moves up/down, the image moves down/up).
The location of the ROI visualized on the live display always matches the image displayed on the Innova Central
touch screen.
Use the Zoom/Roam functions to better visualize and center the part of vessel to quantify.
The Zoom button on the Innova Central touch screen switches zoom state on both images, Innova Central touch
screen and live display.
Drop points apart of the stenosis to quantify from proximal to distal by just touching the vessel (at least 2 points
must be dropped).
All dropped points can be deleted by using the Delete All Points button.
One single point can be selected by just touching it, moved by just dragging it or deleted by pressing on the Delete
Selected Point button.
After all points are dropped, press on the OK button to launch the quantification.
After completion of the quantification, results are displayed on the live display.
The image can be zoomed using the Zoom button and roamed using the Innova Central touch screen joystick.
The Calibration Result window can be hidden by pressing on the Show/Hide Calibration button.
Note: The Stenosis Analysis Results and Vessel Profile windows cannot be hidden.
Vessel diameters along the vessel can be moved by using the Move Diameter button.
Select the diameter which needs to be moved by selecting either "Reference", "At Stenosis", "Lesion Start" or
"Lesion End".
The selected diameter is highlighted on the live display. Tilt the Innova Central touch screen joystick right/left to
move the selected diameter. When the optimum position is found, select another diameter or press the Move
Diameter button again to end diameter movement.
As for Stenosis Quantification, launch the transfer of the image displayed on the live display in the
Select the Contours key and then one of the predefined displays from the pull down list. This function
When possible, allows to toggle between Sub and NoSub image display.
During a vessel quantification, allows to move the "Reference" or "At Stenosis" or "Lesion Start" or
stenosis quantification, the image displayed on the Touch Screen can be roamed by using the Innova Central touch
screen joystick.
Allows to exit from the in-room Stenosis Analysis application (in-room SA).
After In-room Stenosis Analysis is launched, it is always possible to launch the review of the selected sequence or
of another one from the control room console. In such case, the pop up below is displayed:
will fail.
Note: Before starting acquisition, exit the Ventricular Analysis (VA) application to have access to the display of the
X-ray technique factors on the INNOVA Digital Screen.
Contrast media
For ventricular analysis to be successful, the contrast media injected in the left ventricle should have enough
volume to fill the entire ventricle during multiple cardiac cycles and to visualize the end diastolic and end systolic
phases of the cycles.
Quantification
Based on the end-diastolic and end-systolic contour of the left ventricle defined manually by the user, Ventricular
Analysis provides a means to perform Wall Motion (WM) and Global Ejection Fraction (GEF) calculations. Wall
Motion analysis is built upon the Sheehan's centerline method. Global Ejection Fraction analysis provides results
calculated with both the Simpson's rule method and the Dodge-Sandler Area-Length method.
18-1 Ventricular Analysis at a glance
Using the Ventricular Analysis application involves the five activities summarized below.
1. Select study image.
Review patient LV-gram.
2. Launch Ventricular Analysis application.
The VA application can be launched while in review or paused.
3. Select two frames and manually draw LV contour:
The one that most clearly shows the end of Diastolic contour.
The one that most clearly shows the end of Systolic contour.
4. Generate the Global Ejection Fraction and Wall Motion report.
On application request, enter the object to tabletop distance (or use default value) to perform the automatic
calibration. The distance entered shall be determined from the center of the measured object to the top plane of
the table without mattress.
5. Save the GEF and WM report.
When the report is saved, it is added in the Photo Browser as a standard photo.
18-2 Performing Ventricular Analysis
1. Once need for analysis has been established, select the desired sequence from the SEQUENCE BROWSER
and begin playback.
2. Review the images that most clearly show the end of Diastolic and end of Systolic contours.
3. Click the Ventricular Analysis button on the SEQUENCE BROWSER to start the analysis.
The Ventricular Analysis button can be selected while in a review or paused mode.
Following the instructions in the hint window, enter the points of the diastolic contour by clicking with the left mouse
button. The LV diastolic contour shall contain at least 3 points.
Each point entered is represented by a yellow cross.
The application builds up the LV diastolic contour by connecting the user-specified LV edge points using a bicubic
spline interpolation. The diastolic contour is represented by a red curve.
To edit a point already entered, move over the point, click on it, and drag it to the correct position with the left
mouse button kept pressed.
To insert a new point in the contour, click on the contour with the left mouse button.
To delete a selected point, press the Delete key on the keyboard.
If you do not get an acceptable curve using the editing functions, the Delete Diastole (3) allows you to
It is possible to change the frame during the systolic contour edition. The systolic frame will be the one on which the
contour edition is finished.
Dodge's method
The left ventricle is approximated by a prolate ellipsoid with its long axis in the 30° RAO projection. The calculation
of the volume of a prolate ellipsoid only requires knowledge of the length of the longest axis and the area of the
ellipse. The axis of revolution length is the long axis of the LV. The computation of the ellipse area is based on the
mechanical tracing of the LV projection border.
After both end of Diastolic and Systolic contours are drawn, click on Generate GEF report (7) on the
The goal of auto-calibration is to allow the software to compute the pixel size in a given image without using any
kind of reference object. Auto-calibration is entirely based on the known geometry of the imaging system. The only
parameter that the user has to specify is the Object to Tabletop distance defined as being the shortest distance
between the center of the object to be measured and the top of the table.
As opposed to catheter calibration, auto-calibration is available only after a region of interest has been defined in
the image. It means that auto-calibration can be launched only after LV contours or segments are drawn in the
image.
Calibration results are automatically saved at the sequence level. If you launch the application on a previously
calibrated sequence, the calibration information will also be loaded as well as the sequence itself.
Enter the object (heart) to table top distance or use the 15 cm default value, then click OK to get results in metric
units.
THE OBJECT (HEART) TO TABLE TOP DISTANCE IS DEFINED FROM THE TOP OF THE
TABLE TOP AND NOT FROM THE TOP OF THE MATTRESS. IF THE DISTANCE ENTERED
DIFFERS FROM THE ACTUAL DISTANCE, THE RESULTING VOLUMES WILL BE
INACCURATE OF ABOUT 3% PER CENTIMETER OF DIFFERENCE.
In case a new auto-calibration is needed, it is always possible to launch it manually pressing on the Auto-calibration
icon (1) on the Ventricular Analysis toolbar.
Upon completion of auto-calibration, centerline computation and GEF quantification, the GEF report is displayed
with the computed results.
The GEF analysis results report includes the report header, patient and exam information, the Diastolic and
Systolic image, and the results table.
The results table provides the end diastolic, end systolic and stroke volumes of the left ventricle computed in two
ways (Dodge's method and Simpson's method). These volumes are provided in both ml and ml/m² (normalized by
the body surface area. Patient weight and height must be entered). Also, the table contains the Global Ejection
Fraction value expressed in %.
If the sequence has not been calibrated for Ventricular Analysis, then the resulting units of ml or ml/ m² are not
displayed. In this case, only percentage values appear.
IF THE SEQUENCE HAS NOT BEEN CALIBRATED FOR VENTRICULAR ANALYSIS, THIS
CORRECTION IS NOT APPLIED, WHICH LEADS TO A LESS ACCURATE VALUE OF THE
GEF.
If you hide the GEF report by pressing again the Generate GEF report button, you will see both Systolic and
Diastolic contours in the image connected with the 100 equidistant chords.
In case you are not satisfied by results delivered in the GEF report:
1. Hide the current GEF report by pressing again the Generate GEF report button.
2. Modify the position of the existing points on the Diastolic or Systolic contour, add new points, or delete the
contours and draw new ones.
3. Press the Generate GEF Report button. (If requested, perform auto-calibration.)
The new GEF report appears in the screen.
Before exiting the GEF report, click on Store Photo (10) to save the report in the Photo Browser as a
standard photo.
18-7 Wall Motion Analysis
Wall Motion Analysis analyzes the wall motion dynamics of the left ventricle using the Sheehan's centerline
method. This method computes the lengths and locations of 100 chords equally spaced between the user-defined
heart contours.
According to the Sheehan's centerline method, each chord is drawn perpendicular to the centerline which is
constructed midway between the End- Diastolic (ED) and the End-Systolic (ES) contour. The length of the chord
represents the wall motion at a given location of the ventricle contour. The value will be negative if, at a given point
of the contour, the ES contour is located outside the ED contour.
Since the heart size varies from patient to patient, the absolute extent of motion will vary even in normal hearts of
different size. To normalize for heart size, each chord length is divided by the length of the ED perimeter. The
normalized chord data can then be standardized.
This is done by subtracting from each chord the average of a group of normal patients (called the normal mean),
and dividing the result by the standard deviation from the normal mean for that chord.
Note: This method is validated for LV acquired in RAO 30° or LAO 60° angulations. So, if the sequence was not
acquired in RAO 30° +/- 10° or LAO 60° +/- 10°, or with a CRA or CAU angulation greater than 10°, or with more
than 5° difference between the end of Diastolic and end of Systolic frame, a warning message about the loss of
result accuracy will be displayed.
After both end of Diastolic and Systolic contours are drawn, click on Generate Wall Motion report (8) on
The Wall Motion analysis report includes a report header, patient and exam information, a heart graph, a standard
deviation graph, and a results table.
The Heart graph field (top right of the WM report) contains the Systolic and Diastolic contours previously drawn by
the user, together with the corresponding RAO/LAO values. Also, the 100 chords representing the current wall
motion of the left ventricle are displayed (current patient chords).
To gather precise and relevant information on hypokinesia and hyperkinesia in different ROIs, study the Standard
deviation graph and the Results table.
The standard deviation graph (bottom left of the WM report) shows the normalized wall motion as a function of
chord number, based on Wall Motion computation.
In the case of hypokinesia, the normalized wall motion is a negative value. For hyperkinesia, the normalized wall
motion is positive.
Left ventricle with normalized motion between -2 and +2 can be considered as normal.
The Results table field (bottom right of the WM report) contains a table with the results of Wall Motion Regional
analysis.
The results provide information about the impact on the wall motion of the three main coronary arteries (LAD, RCA,
and CFX) affected with single or multiple diseases. Also, the chord interval is specified for each vessel ROI in the
'Chord #' column. The chord interval associated with a main coronary territory is different in the case of RAO 30°
+/- 10° and LAO 60° +/- 10° acquisition. The table below shows the chord intervals for each ROI in both cases.
Region of Interest (ROI) Chord interval for RAO 30° Chord interval for LAO 60°
CFX single 10 - 80 19 - 67
CFX multiple 10 - 80 19 - 67
LAD single 10 - 66 50 - 67
LAD multiple 10 - 58 50 - 67
RCA single 51 - 80 38 - 74
RCA multiple 59 - 80 38 - 74
The 'Hypokin. SD' column displays the average of hypokinesia severity in standard deviation unit in a certain ROI. If
the average of the hypokinesia value is lower than -2, then it is displayed in bold. Negative values in this column
refer to hypokinesia. Wall motion is considered normal, if it is between -2 and +2.
For each ROI, the '% Total LV contour' column displays what percentage of the entire length of the left ventricle
contour can be characterized with hypokinesia outside the [-2;+2] range.
For each ROI, the 'Hyperkin. op. ROI SD' column displays the possible hyperkinesia compensation in the opposite
ROI. It corresponds to a significant hyperkinesia if it is positive and outside the range [-2; +2].
If you hide the WM report by pressing again on the Generate WM report button, you will see both Systolic and
Diastolic contours in the image connected with the 100 equidistant chords.
In case you are not satisfied by results delivered in the WM report:
1. Hide the current WM report by pressing again the Generate WM report button.
2. Modify the position of the existing points on the Diastolic or Systolic contour, add new points, or delete the
contours and draw new ones.
3. Press the Generate WM Report button. (If requested, perform auto-calibration.)
The new WM report appears in the screen.
Before exiting the WM report, click on Store Photo (10) to save the report in the Photo Browser as a
standard photo.
18-8 Multi Segments
Click on the icon Multi-segments (9) on the Ventricular Analysis tool menu to draw a link made of multiple
segments (multi-segments) on a single frame of a sequence, and to estimate the length of the segments.
Note: Only one Multi-segments can be displayed at a time.
Following the instructions in the Hint window, enter the points of the multi-segment by clicking with the left mouse
button.
Before you place the first point, you can perform frame navigation using the Play/Pause or Next/Previous frame
functions.
The adjacent points of the multi-segments can be as close as you want from each other. However, remember that
the smaller the distance between the points is, the higher the imprecision will be.
While creating the multi-segments, you can edit any of the points already entered. Move over a point, the pointer
will change to a crosshair cursor, and then click and drag with the left mouse button to the intended position.
Also, you can insert points between two points already entered. Move over the intended position on the segment
line (the pointer will change to a crosshair cursor), and click with the left mouse button.
Double-click the last or any existing point to indicate that the multi-segment is finished.
To draw a new multi-segment on the selected frame, repeat the above procedure.
Before exiting, click on Store Photo (10) to save the image with measurements in the Photo Browser as a
standard photo.
After all analysis are performed and all images/graphics/reports saved, click on the Exit icon
19 StentViz (Option)
The StentViz application is designed to enhance the visibility of a stent placed during an interventional Cardiac
procedure. Additionally, StentViz may help the cardiologist assess the correct deployment of the stent in the vessel.
If a stent was already placed, it may also be used to verify the positioning of a new balloon before the deployment
of the second adjacent stent, or before re-dilatation of the stent.
The result is shown on the static image below detailing the enhanced image image quality and contrast of the stent.
There are two StentViz images displayed on the reference display. One where the full guidewire is displayed and a
second where the portion of the guidewire between the balloon marker-balls has been subtracted. The subtracted
guidewire is intended to ease the visualization of the stent.
There are four different displays the image will be divided depending on the original direction the vessel was
imaged.
Note: This application is restricted to heart anatomy only and can be launched only on images acquired using the
automatic StentViz workflow described below.
The success of the StentViz application might be altered by the poor quality of the images (high noise level…), the
type of balloon used (low marker radio-opacity), the radio-opacity of the guidewire, and or high radio-opacity
structures in the vicinity of the markers.
Note: The StentViz resulting photo is a recomputed image. Always associate it with the original recorded sequence
for interpretation.
19-1 Automatic StentViz Workflow
To launch the StentViz application acquisition to image a deployed stent the following is required:
1. Deflate the balloon and leave it in position inside the stent. StentViz will detect and focus on both markers of the
balloon and on the guidewire to perform the image processing.
2. A cardiac related protocol needs to be selected. On DL console or on the Innova Central touch screen (option),
select Dynamic and then StentViz to launch the application. An icon will be displayed on Live monitor to indicate
the StentViz acquisition mode is set for the next single dynamic acquisition.
The message “READY FOR STENT VIZ” is displayed at the bottom left of the Live monitor in the Status area.
3. Perform a short Record acquisition centered on the display stent and deflated balloon. A "Stop X-Ray" blinking
message will be displayed on the Live monitor when enough frames are acquired. You can release the pedal at
this stage.
The acquired sequence is stored in the DL Sequence browser with a specific default "Pre-StentViz" label.
4. At the end of acquisition, StentViz processing is automatically launched. The message "STENT VIZ IN
PROGRESS" is displayed at the bottom left of the Live monitor in the Status area. The typical processing time is
less than 10 seconds. Because image processing is performed in the background, fluoro, record, review and
image processing are always available.
5. At the end of StentViz processing time, the message "STENT VIZ IN PROGRESS" is replaced by "STENT VIZ
PHOTO READY" at the bottom left of the Live monitor and the resulting image is automatically displayed on the
reference display as a photo showing the stent with enhanced image quality and contrast. As for other photos,
this photo is stored in the DL Photo browser with a specific default "StentViz" label.
Note: In case of fluoro or record acquisition, or image review in progress at the completion of the StentViz
processing, the resulting photo is displayed only at the end of the fluoro or when the review switches to Pause
mode.
Note: The next Dynamic acquisition will be performed without StentViz processing. In case a new StentViz
acquisition is required, reselect the StentViz button on the DL Console or Innova Central touch screen (option).
19-2 Manual StentViz Workflow
If the StentViz processing is not successful, StentViz can be manually launched on previously acquired StentViz
sequence:
1. Select the StentViz native sequence (labeled “Pre-StentViz”) and launch the review. Zoom the image and center
it on the stent. This will increase the sucess rate and decrease the processing time.
2. Launch StentViz through the “StentRefine” button from the DL sequence browser on the control room console,
the DL remote control using the Menu key, or from the Innova Central touch screen (Option) using Review
menu or the SmartNav menu.
Note: StentRefine is used indifferently for StentViz or StentVesselViz post-processing. The selection of StentViz
or StentVesselViz post-processing is automatically done by the system depending of the acquisition setting of
the native sequence.
DL sequence browser
SmartNav menu
Live and Refer- StentViz already running. Restart it The previous StentRefine or StentViz post-processing is
ence monitor manually later still running
This is not an eligible sequence for The sequence selected for the StentRefine post-
StentRefine processing has not been acquired using a StentViz
workflow
StentViz not available. Not enough The acquire sequence or selected sequence does not
frames in sequence have enough frames for StentViz post-processing.
Redo and wait until "X-Ray Stop" is displayed.
20 StentVesselViz (Option)
The StentVesselViz application is designed to enhance the visibility of a stent placed during an interventional
Cardiac procedure. Additionally, StentVesselViz may help the cardiologist assess the correct deployment of the
stent in relation with the lumen of the artery.
If a stent was already placed, it may also be used to verify the positioning of a new balloon before the deployment
of the second adjacent stent
The result is shown on:
1. The static image below detailing the enhanced image quality and contrast of the stent (StentViz).
2. A dynamic zoomed image sequence which displays a progressive fading between the enhanced stent image
and an injected vessel image.
Note: The StentVesselViz application includes the StentViz processing.
20-1 Output Description
20-1-1 StentViz Images
There are two StentViz images displayed on the reference display. One where the full guidewire is displayed and a
second where the portion of the guidewire between the balloon marker-balls has been subtracted. The subtracted
guidewire is intended to ease the visualization of the stent.
There are four different displays the image will be divided depending on the original orientation the vessel was
imaged.
Note:The StentViz resulting photo is a recomputed image. Always associate it with the original recorded sequence
for interpretation.
20-1-2 StentVesselViz Sequence
The StentVesselViz outputs a sequence of images that displays a video fading between a stent enhanced image
(StentViz image) and the lumen of the vessel it is deployed in. The sequence starts by displaying the injected
vessel. Then the stent enhanced image (with subtracted guidewire) progressively appears over the vessel. The
stent enhanced image is registered so that the balloon marker-balls are at the same location as in the vessel
image. In a third phase, the vessel disappears to leave the full stent enhanced image. Finally the guidewire is
displayed over the stent image.
Note: This application is restricted to heart anatomy and can be launched only on images acquired using the
Automatic StentVesselViz workflow described below.
The success of the StentVesselViz application might be altered by poor quality of the images (high noise level…),
the type of balloon used (low marker radio-opacity), the radio-opacity of the guidewire, and or high radio-opacity
structures in the vicinity of the markers.
20-2 Automatic StentVesselViz Workflow
To launch the StentVesselViz application acquisition to image a deployed stent the following is required:
1. Deflate the balloon and leave it in position. StentVesselViz will detect and focus on both markers of the balloon
and on the guidewire to perfom the image processing.
2. A cardiac related protocol needs to be selected. On DL console or on the Innova Central touch screen (option),
select Dynamic and then StentVesselViz to launch the application. An icon will be displayed on Live monitor to
indicate the StentVesselViz acquisition mode is set for the next single Dynamic acquisition.
The message “READY FOR STENT VESSEL VIZ” is displayed at the bottom left of the Live monitor in the
Status area.
3. Perform a single and continuous Record acquisition centered over the deployed stent and delated balloon, first
part without injection of contrast media, second part with injection of contrast media:
– Non-injected frames: an "inject" blinking message will be displayed on the Live monitor when enough non
injected frames are acquired.
– Injected frames: proceed as usual for an injected acquisition - standard injection-.
The acquired sequence is stored in the DL Sequence browser with a specific default "Pre-StentVesselViz" label.
6. At the end of StentVesselViz processing time, the message "STENT VESSEL VIZ IN PROGRESS" is replaced
by "STENT VESSEL VIZ READY" at the bottom left of the Live monitor. The typical processing time is less than
30 seconds. The resulting sequence is automatically displayed on the Live monitor as a sequence showing the
stent with enhanced image quality and contrast and the vessel filled with contrast media.
The sequence will be stored in the DL sequence browser with a specific default "Post-StentVesselViz" label.
Note: If the "Pre-StentVesselViz" sequence (or native sequence) is not in review or paused at the end of the
StentVesselViz processing, the review of the StentVesselViz post-sequence is to be launched manually. Launch
"Post- StentVesselViz" review through the "Play/Pause" or "Next/Prev" button from the DL remote control or from
the DL keypad, or through the "Recall sequence" button on the Innova Central touch screen (Option).
Note: The next Dynamic acquisition will be performed without StentVesselViz processing. In case a new
StentVesselViz acquisition is required, reselect StentVesselViz button on DL Console or Innova Central touch
screen (option).
DL sequence browser
DL remote control
SmartNav menu
– Avoid large breathing motion during the injected phase of the sequence as much as possible.
20-5 StentVesselViz User Messages
Live and Refer- StentVesselViz already running. The previous StentRefine or StentVesselViz post-
ence monitor Restart it manually later processing is still running
This is not an eligible sequence for The sequence selected for the StentRefine post-
StentRefine processing has not been acquired using a StentVes-
selViz workflow
StentVesselViz not available. Not The acquire sequence or selected sequence does not
enough frames in sequence have enough frames for StentVesselViz post-
processing. Redo and wait until "Inject" in displayed be-
fore injecting.
21 Dose Map
THE CUMULATED AIR KERMA (CAK) REMAINS THE REFERENCE FOR DOSE
MANAGEMENT.
During the exam, the Dose Map will be displayed on the Reference display:
• Automatically, if the amount of dose received locally has exceeded a pre-defined threshold
• On request, using one of the following ways:
– [Dose Map] button on the DL Sequence Browser
– {Dose Map} button on the “Dose” tab on the Innova Central touch screen
– <Dose Map> menu button on the remote control or {Dose Map} button on Smart Nav
Note: The same buttons can be used to close the Dose Map on Innova Central touch screen and Browser.
Additionally, the next pedal press for fluoroscopy or record will automatically close the Dose Map. Any photo that
was hidden by the Dose map will be redisplayed on the Reference display.
Note: Dose Map cannot be displayed during an acquisition.
For systems equipped with the Innova Central touch screen, the "Dose" tab will be automatically displayed when
the system determines that a threshold of local dose is reached and lower dose settings would be beneficial.
If predefined, the final Dose Map will be automatically saved as a photo at the end of the exam.
Note: Dose Map cannot be saved using the [Store Photo] command.
21-2 How to set Dose Map?
21-2-1 Dose Map preferences
Dose Map predefined preferences must be set before the start of an exam in the Preferences tab on the DL
screen.
to end distance field; which allows the system to use this automatically for each patient.
Note: Head to end distance or Default head to end distance are used by the system to set dual
model used for exams done in the head area. There is no need to enter a value for Head to end
distance or Default head to end distance if the exam is done outside the head area (thorax, pelvis,
etc…).
• Dose Map automatically stored as photo:
– If [Yes] is selected, Dose Map will be automatically saved as photo at the end of the exam. It will then be
managed as any other exam photo taken during the same exam.
– If [No] is selected, no Dose Map will be saved and no photo will not be accessible at a later time.
• Mattress type: Dose Map internal computation uses the mattress type to have a better precision. It can be:
– 1" / 2.5 cm Omega standard mattress with slicker
– 1" / 2.5 cm Omega anti sore mattress
– 2" / 5 cm Omega anti sore mattress
– 2" / 5 cm InnovaIQ anti sore mattress
Note: This list only includes mattresses delivered with GE IGS systems. The GE service engineer
or applications specialist can help select the appropriate mattress.
In case you don't find the used mattress in the previous list, you have two possibilities:
1. Select the 1" / 2.5 cm Omega anti sore mattress
2. If the Al equivalent filtration of the mattress is known at 100kV, select the GE mattress with the
closest lower equivalency of Al filtration.
• Preferences tab: Default head to end distance (refer to Dose Map Preferences above)
Note: Head to end distance is not a parameter filled using Worklist. It had to be manually entered prior to the
exams.
21-3 Dose Map limitations and recommendations
Dose Map models (single or dual) are only based on thorax estimation, not including the arms.
No Dose Map is provided when the decubitus patient position is used.
Each time the following parameters are changed during an exam, the current Dose Map is closed and a new map
is set:
• Patient data change: birth date, age, weight, height, head to end distance
• Protocol change driven by an anatomical change: from all except head area to head area
• Patient position change
• [End exam] followed by a Resume Exam
The following are recommendations:
• Enter patient data before starting the exam, including head to end distance. Avoid changing data during the
exam.
• Place patients at the same Head to end distance which allows the use of the Default head to end distance to be
set on the system. If not, then the Head to end distance will have to be measured prior to each exam.
• If it is anticipated to perform an exam on the head or neck region, select a dedicated protocol at the beginning of
the exam.
• Avoid changing patient position during an exam.
• If there is an interruption in the exam, it is recommended to activate X-ray lock versus [End Exam] then [Start
Exam]. This will avoid the system from creating a new dose map.
Additional recommendations:
• When using the Large Display Monitor, the Reference display must be included in the selected layout.
• In case of multiple dose maps created during the same exam, the total estimated local dose amount is the sum
of the estimated local dose in each map.
21-4 Information provided by Dose Map
21-4-1 Dose Map representation - On Reference Display
Local cumulated dose is represented as a 2D view of a 3D model.
The length of the model is the length of the table, with a pixel resolution of 1 cm2. Cumulated local dose levels are
represented using a grey scale. The grey scale is directly linked to the local dose threshold chosen in the
Preferences tab. The least amount of radiation is represented from black to white, the highest amount of radiation.
The Dose Map is displayed on the Reference display’s image area. The Dose Map is composed of three areas:
1. Full view of the cumulated local dose on the model
2. Zoomed view of the cumulated local dose on the model
3. Information area
Indicates the usage of multiple Dose Map after a second map was created dur-
ing the same procedure.
Estimation of local cumulated dose has to be done adding the local value on
the two maps.
Displayed when more than 1 Gy of the sum of Air Kerma on the map is missing
dose data during the Dose Map calculation.
In case of multiple Dose Maps created during the same procedure, this icon will be includ-
ed on the last photo.
In this case, two maps have been created and stored during the procedure.
Note: If the mattress used has no GE part number or the GE part number doesn’t figure in the mattress list defined
in Dose Map / How to set Dose Map? / Dose Map preferences, or in case of accessory usage in the beam: Dose
Map accuracy is not guarantee.
Refer to Appendix - Dose Map Accuracy procedure.
22 Remote Control
12 Next Image 24 Slower Review / Contrast - / Pixel Shift Down / Roam / IRB
Sequence thumbnails are displayed first with their original sequence number; photo thumbnails are then displayed
with a dedicated photo icon and their original photo number.
A thick horizontal white line separates sequences and photos.
Note: In case of multiple exams performed on a Patient, sequences of each exam are displayed first (a thin
horizontal white line separates each exam), then, after the thick white line, photos of each exam are displayed (a
thin white line also separates each exam).
Once the IRB is displayed on the reference display, use the Exam Loop key to switch focus either on the IRB or on
the live display. Refer to the highlighted label in side panel of both IRB or live displays to visualize where the focus
is located.
Focus on IRB
Note: The IRB will stay displayed on the reference display even when the focus is on the live display.
The IRB will be removed only in case of Start/End Exam or Store/Recall photo.
It will not be possible to activate or take focus on the IRB during a Stenosis quantification of acquisition.
When the focus is on the IRB, use one of the eight Remote Control arrows to select the sequence or photo to
review.
A dedicated icon shows the selected thumbnail:
After selection, press on the Remote Control Enter key to launch the review.
A dedicated icon shows the sequence in review:
Note: The thumbnail selection can also be performed using the DL keypad joystick. Press also on the DL keypad
Enter key to launch the review.
A small arrow is displayed in the bottom right corner of the IRB to indicate that more than 16 thumbnails are
available (see 1).
Use the Remote arrow keys to scroll and see the other available thumbnails.
While focus is on the IRB, the 6 blue keys of the Remote Control (or equivalent functions on the DL keypad) can be
used without exiting focus from the IRB.
Any other processing function activation (Zoom, Pixel Shift...) will automatically switch focus on the live display.
In case of photo thumbnail selection and Enter key pressed, the selected photo will replace the IRB on the
reference display and the system will react as if a Recall photo was activated.
Press on Exam Loop again to re-display the IRB when the photo is not needed anymore.
23 DL Keypad
Note: The monitor pattern is not available while in review or under Ready for exam status (exam in progress).
• To display the pattern on both live and reference displays, press the blank (19) button.
• To remove the pattern:
1. Creating a new patient and starting the exam will blacken both screens.
2. Starting a review on any sequence will remove the pattern from the live display.
3. Recalling a photo will remove the pattern from the reference display.
23-1 Edge Enhancement Filter Selection
Edge Enhancement can be used to sharpen or smooth the image content. A range of filter values is available and
permits to obtain different image looks. By default, the edge filter used during acquisition is applied during DL
review. This default edge enhancement value can be changed.
The following procedure should be followed to adjust the desired level of edge enhancement for Cardiac/DSA no
Sub/Bolus/Chase/DSA/Innova 3D/Innova 3DCT acquisitions:
• Load the sequence to be reviewed.
• Press the "Edge Enhancement Filter Selection" and select the preferred filter setting for the type of acquisition
reviewed.
• Apply the same number as the filter selected to the "Customer Sharpness Sensitivity" parameter in the
acquisition protocol (ask Applications Specialist).
In case of system failure requiring a reset during a procedure, dose related to fluoro
acquisitions performed after the last record acquisition is lost.
If a reset is needed during a procedure requiring fluoro only, all patient dose information are entirely lost. A way to
estimate the cumulated patient skin dose (worst case) in Gy is obtained by multiplying the estimated performed
fluoro time in minutes by 0.1.
e.g. for a performed fluoro time of 60 min, the estimated cumulated patient Air Kerma is: 60 x 0.1 = 6 Gy.
A way to estimate the cumulated DAP (Dose Area Product) in Gray.cm² (worst case) is by multiplying the estimated
performed fluoro time in minutes by 0.1 and further multiplied by 100.
e.g. for a performed fluoro time of 60 min, the estimated DAP is: 60 x 0.1 x 100 = 600 Gy x cm²
Always end any exam by using the "End Exam" key located on the top of the DL flat panel in
order to save all dose information related to the exam.
At "End exam", the actual dose information are recorded and displayed on the appropriate fields in the exam
browser.
Doing a start exam on an already performed exam will update the dose information on the reference display with
the values previously recorded at exam level.
The icon is used as a visual update of the status of the ECG signal. Waveform & detected peaks are sent in real
time during all kinds of acquisition modes to the AW. When the ECG is enabled and well received (i.e. status is
Online), the ECG data will be automatically stored in the DICOM header of the corresponding sequence for the
following record modes: DSA, Dynamic mode, unsubtracted Innova 3D and Innova CT.
Click on the ECG icon on top of the Innova DL screen to display the beats per minute and to reset the R-wave peak
to peak detection
There are 4 possible ECG status modes, which will be displayed on the DL and the live display (Offline, Stabilizing,
Online and No Trigger).
25-1 ECG icon indication on the DL by color and on the live display
Stabilizing (yellow):
A signal is received, and the system is "learning" the ECG.
The algorithm detecting the peaks needs a period of 10 seconds to analyze the waveform (polarity, amplitude, etc.)
before being able to trigger.
An hour glass will be displayed during the stabilization phase.
Online (green):
A signal is received and the system is delivering triggers.
No Trigger (orange):
A signal is received but the system is not capable of delivering a trigger.
The system is receiving the signal, the current BPM are displayed.
25-2-2 Status "Stabilizing"
The "- -" will be displayed if there is no R – Peak detection or there is noise in the signal.
Note: Always refer to the hemodynamic or electrophysiology system as the prime reference for the status of the
patient’s ECG, not the Innova system.
25-2-4 Status "Offline"
No ECG signal is received, the ECG line is black, there is no ECG indication
If the ECG signal is lost:
In most situations, check the hemodynamic monitor verifying there is no electrode failure. Once that is confirmed,
perform a reset of the R-peak detection to try to recover the signal.
There are some situations that the ECG waveform analyzer will be 'lost'.
Examples:
• Noise cannot be differentiated from a regular signal and will try to detect peaks.
• When peaks are detected at a frequency greater than 300 BPM, the display of the BPM value will be
">300BPM" until the frequency is less.
• If the polarity of the signal is changing, it is probable that no peak will be detected.
• If the signal is too noisy: false peaks may be detected.
• If the signal amplitude is increasing: false peaks may be detected.
• If the signal amplitude is decreasing: no peak may be detected.
28 How to Delete
Note: Before to delete any data from the Innova Digital, always check that all data was successfully pushed and
available on the AW Workstation.
28-1 How to delete a Patient
Patient Browser
Delete Patient
• Click on the [Patients] tab at top left of DL screen to select the PATIENT BROWSER. By a simple click; select
the patient(s) you want to delete.
Note: You can select multiple patients:
If not all of the sequences of a patient have been pushed/archived, another pop-up "Patient
'Last Name' was not sent to network/archived yet. Confirm delete patient?" will appear. If the
Innova system is connected to a Patient Monitoring/Scheduling/Reporting system and at
least one MPPS message was not sent yet, a similar pop-up will be displayed indicating
“MPPS data is not yet sent” and asking to confirm the deletion. If DICOM Dose Structured
Report is configured and atleast one DICOM Dose Structured Report message was not sent
yet, a confirmation popup shall be displayed to user indicating “Dose Report is not yet
sent” and asking to confirm the deletion.
• After a delete is done, the PATIENT BROWSER will be updated and redisplayed.
28-2 How to delete an Exam
• Click on the [Patients] tab at top left of DL screen to select the PATIENT BROWSER. Select a single patient by
a simple click.
– Click on the [Exams] tab to select the EXAM BROWSER. The Exam Browser window appears.
– Select on the exam(s) to be deleted.
Note: You can select multiple exams:
• Using <Shift> + left mouse button for contiguous exams.
• Using <Control> + left mouse button for non-contiguous exams.
• While holding the [Control] key, click on the [Delete] key.
• The pop-up "Confirm Delete Exam" will appear.
• Select [Cancel] key to cancel deletion.
• Select [OK] or press the <Enter> key to confirm deletion.
If not all of the sequences of an exam have been pushed/archived, another pop-up "Exam
from date 'DD-MMM-YYYY' was not sent to network/archived yet. Confirm delete exam?" will
appear. If the Innova system is connected to a Patient Monitoring/Scheduling/Reporting
system and at least one MPPS message was not sent yet, a similar pop-up will be displayed
asking to confirm the deletion. If DICOM Dose Structured Report is configured and atleast
one DICOM Dose Structured Report message was not sent yet, a confirmation popup shall
be displayed to user indicating “Dose Report is not yet sent” and asking to confirm the
deletion.
• After a delete is done, the EXAM BROWSER will be updated and redisplayed.
If one of the sequences has not been pushed/archived, another pop-up "Sequence was not
sent to network/archived yet. Confirm delete sequence?" will appear.
• After selection has been done, the SEQUENCE BROWSER will be updated and redisplayed.
28-4 How to delete a Photo
• Click on the [Patients] tab at top left of DL screen to select the PATIENT BROWSER. Select a single patient by
a simple click.
• Click on the [Exams] tab to select the EXAM BROWSER Select a single exam by a simple click.
• Click on the [Photos] tab. The PHOTO BROWSER window appears.
• Click on the photo to be deleted.
Note: In case of a biplane photo, both Frontal and Lateral photo will be deleted at the same time.
Note: You can select multiple photos:
• Using <Shift> + left mouse button for contiguous icons.
• Using <Control> + left mouse button for non-contiguous icons.
• While holding the [Control] key, click on [Delete] key. The message "Confirm Delete Selected Photo(s)"
appears.
• Select [Cancel] key to cancel deletion.
• Select [OK] or press the <Enter> key to confirm deletion.
If one of the photos has not been pushed/archived, another pop-up "Photo was not sent to
network/archived yet. Confirm delete photo?" will appear. Select [OK] (or press the <Enter>
key) to confirm deletion. Select [Cancel] key to cancel deletion.
• After a selection is made, the PHOTO BROWSER will be updated and redisplayed.
Alternatively, it is possible to preset the Auto Archive to ON or OFF by default in the Browser. By default, the Auto
Archive is set to ON. The update will require a system restart.
Note: Transfer will pause when there is a fluoro, record or review in progress.
29-2 How to configure and select a Host
• Click on the [Net] button at the top of the DL screen. The NETWORK QUEUE appears:
• Click on the [Select Host] button. This opens the host configuration screen. The Worklist and MPPS tabs are
displayed if Worklist option and MPPS are enabled.
[Verify] button allows checking if the network connection is alive between the Innova system and the remote host.
By pressing [Save] button and if all fields are filled in, the new host will be added to the list in the left window.
User can select the 3D reconstruction host (if 3D option is selected).
User can set the auto push option. 'Yes' is selected by default. If automatic push behavior is not required, this
parameter is to be selected as ‘No’. It requires a system restart to update changes made to this setting.
From the available host(s) list in the left window, the user can select one or several hosts by clicking on its/their
name with the left mouse button.
In case of several hosts selected, use the Up and Down buttons to prioritize the transfer. The hosts may also be
dragged up or down to change their priority. Transfer is performed to the host from first line of the list to last bottom
line of the list.
At the bottom of each Patients/Exams/Sequences/Photos Browser, the selected host(s) is (are) displayed in priority
order from left to right.
Host set in the system for 3D acquisition network is labeled with "3D" written after the host name).
Note: Whatever be the priority of the 3D host in the host list, a 3D acquisition is always networked first. Host set as
archive station is labeled "host name*" (small star written after the host name).
CASE 2: If the Network status at Patient Browser is INFORMATION, then place the mouse over the
INFORMATION status.
CASE 2.1: If the tooltip indicates "There is no image host configured" then, configure and select an Image push
host. Refer How to configure and select a Host
CASE 2.2: If the tooltip indicates "There are images that have not been transfered. A manual push is required"
then, go to Exams tab.
In the Exams tab, click on the network status for the exam with "Information" status.
If the tooltip indicates “There are unsent images” then, go to Sequences and Photos tab.
Manually push the unsent Images.
CASE 2.3: If Tooltip indicates "This patient is under exam", it implies that the exam is ongoing.
CASE 3: If the Network status at Patient Browser is SENDING, then place the mouse over the SENDING status.
If the tool-tip indicates: "Sending Image(s) is in progress" then, image push is underway. Need to wait for the
network operation to end.
CASE 4: If the Network status at Patient Browser is ARCHIVING, then place the mouse over the ARCHIVING
status.
If the tool-tip indicates: "Archiving of Image(s) is in progress" then, image archival is underway. Need to wait for the
network operation to end.
CASE 5: If the Network status at Patient Browser is COMPLETED, then place the mouse over the COMPLETED
status.
If the tooltip indicates "Images have been transferred successfully" then, it implies successful transfer of all images
for the selected patient.
• Delete Selected: Remove the selected item from the queue. This function is available only on items with
PENDING or FAILED status.
• Clear Pending: Remove all PENDING items from the queue, after completion of the ACTIVE item transfer.
• Retry Failed: Return all items with FAILED status to the active part of the queue; their status is changed to
PENDING.
• Retry Selected: Allow to resend selected items with PENDING status (neither successfully pushed nor failed) in
case of communication problem.
• Abort: Immediately abort the current ACTIVE task.
• Select Host: Give access to configuring different host(s).
User can modify the host configuration details of the existing host.
User can DELETE the host by clicking the [Clear] and [Save] button in the given order.
User can set the automatic retrieve of Worklist at boot. 'Yes' is selected by default. Changing this parameter would
require a system restart.
Pressing the [Refresh] button on the patient Browser retrieves the worklist.
While the query is in progress, the system remains fully available for other operations.
During the query, the [Refresh] button toggles to [Cancel Query], allowing canceling the query in progress.
If the query fails, the last successful query result remains displayed.
The worklist can be sorted by patient name or by scheduled start date and time by clicking on the header row.
The system will retrieve items that satisfy the criteria defined in all three boxes. In the "Date" area, only one item
can be selected.
The "Modality Worklist" criteria and the "Date" criteria are stored persistently on the system. The patient criterion is
stored as long as the system is not re-booted.
MODALITY WORKLIST
One or both or none of the checkboxes may be selected
• This room: The scheduled system's DICOM AE-Title matches the local dicom AE-Title.
• This modality: The modality of the scheduled procedure step is "XA", X-ray angiography.
• Leaving both checkboxes unchecked: Selecting this will have the effect of bypassing the scheduled system
selection criteria, since it will always match.
• Both checkboxes checked: The scheduled system's DICOM AE-Title and the modality of the scheduled
procedure step will be used for filtering.
Note: This selection must be used when a GE Mac-Lab is used to retrieve the worklist.
Note: To see the local system's AE-Title, select Service from the Browser. In Service User Interface, choose
configuration application, then DICOM.
DATE
Any date: Selecting this will have the effect of bypassing the date selection criteria, since it will always match.
From ... to: Enter the selected dates in the following format: 01-Jan-2001.
PATIENT SEARCH
Multiple matching criteria can be defined in this area.
For the last name and first name, the query will match if the name in the worklist provider contains the given name.
E.g.: querying for "John" will return "John", "Johnson" ...
The tab key navigates between the fields.
[Refresh now] button
This has the same effect as [Apply] + [Refresh].
[Apply] button
The window is closed. The settings definition is saved.
[Cancel] button
All edits are discarded. The window is closed.
Case 2: The patient has more than one scheduled procedure step:
• To create one exam from each procedure step, select and import each procedure step individually.
• Selecting and importing the patient line or a group of scheduled procedure steps will import the first procedure
step.
Importing items from the worklist will open the patient and exam info screen, pre-filled with the information from the
worklist provider.
To ease the patient identification in Review Stations and Archive Stations, the patient demographic information is
not editable if it was created from the worklist (patient last and first name, patient ID, date of birth, other patient
identification and sex). If editing is required, it should be done on the worklist provider, then the worklist should be
refreshed and the patient re-imported.
This opens the window to configure the hosts. The tabs MPPS and Worklist is displayed if MPPS and Worklist
option is enabled.
Selecting the MPPS tab of Host Configuration screen, the user can configure the MPPS host.
Only one MPPS host can be configured.
All fields will have to be filled in to create a new host.
By pressing [Save] button if all fields are filled in, the new host will be saved.
A [Verify] button allows checking if the network connection is alive between the Innova system and the remote
host.
User can modify the host configuration details of the existing host.
User can DELETE the host by clicking the [Clear] and [Save] buttons in the given order.
If any Exam displays Network status as "FAILED", place the mouse over the FAILED status.
Follow the tool-tip instructions to recover the error.
CASE 2: If the Network status at Patient Browser is COMPLETED, then place the mouse over the COMPLETED
status.
If the tooltip indicates "MPPS data has been transferred successfully" then, it implies successful transfer of all
MPPS for the selected patient.
THE EXAM STARTED ON INNOVA SYSTEM MUST ALWAYS BE THE EXAM THAT IS ACTIVE
ON THE PATIENT MONITORING/SCHEDULING/REPORTING SYSTEM. STARTING ANOTHER
EXAM ON INNOVA SYSTEM WILL LEAD TO FAILING THE EXPORT OF THE EXAM DOSE
AND DATA.
MAKE SURE THAT THE EXAM IS PROPERLY CLOSED BY PRESSING THE [END EXAM or
ABORT EXAM] BUTTON ON THE INNOVA SYSTEM BROWSER SCREEN. PRESSING THE
[END EXAM or ABORT EXAM] BUTTON WILL ACTIVATE THE AUTOMATIC TRANSFER OF
THE EXAM DATA. IF THE INNOVA SYSTEM IS RE-STARTED WITHOUT PROPERLY
CLOSING THE EXAM, THE EXAM DATA WILL NOT BE AUTOMATICALLY TRANSFERRED.
Each time an exam is ended on Innova System, the X-Ray information of the recently performed exam is sent to
the patient monitoring/scheduling/reporting system.
If the patient information is not populated on the Patient Monitoring/Scheduling/Reporting system, or the study is
not activated on the Patient Monitoring/Scheduling/Reporting system, it will be impossible to retrieve on Innova the
patient demographics via the Worklist. Therefore, a new patient and exam must be created manually on Innova.
At the end of this unscheduled exam, the X-ray exam information will be integrated in the patient record of
Mac-Lab/CardioLab and Carddas.
31-4-1 Known limitations with Mac-Lab/CardioLab system
The 'Cine Dose', 'Fluoro Dose', 'Total Dose' and 'Cine Frames' fields is not displayed in the Dose Report of
Mac-Lab/CardioLab. The user may get these values from the information displayed on the Innova screens. The
user can also find these values in the DICOM Dose Structured Report document that will be available in the Dose
Report Receiving station.
This opens the window to configure different hosts. The tabs MPPS and Worklist is displayed if MPPS and Worklist
option is enabled.
Select the Dose SR tab of Host Configuration screen to create or modify Dose SR hosts.
If the user clicks on the 'Add' button, empty host parameter fields will be displayed for the new host.
All fields will have to be filled in to create a new host.
If Storage Commitment workflow is not enabled, the ‘Archive Station’ radio buttons will be disabled.
If Storage Commitment workflow is enabled, the ‘Archive Station’ radio buttons will be enabled.
Initial state of radio buttons: For Archive Station "No" is selected by default.
If the user checks ‘yes’ for ‘Archive Station’ option, the ‘Storage Commitment’ edit fields will be enabled and, if they
are empty, they will be filled in automatically with the same values as the storage host.
[Verify] button allows checking if the network connection is alive between the Innova system and the remote host.
By pressing [Save] button and if all fields are filled in, the new host will be added to the list in the left window.
From the available host(s) list in the left window, user can select one or several host by clicking on its/their name
with the left mouse button.
In case of several hosts selected, use the Up and Down buttons to prioritize the transfer. The hosts may also be
dragged up or down to change their priority. Transfer is performed to host from first line of the list to last bottom line
of the list.
CASE 2: If the Network status at Patient Browser is COMPLETED, then place the mouse over the COMPLETED
status.
If the tooltip indicates "Dose Report has been transferred successfully" then, it implies successful transfer of all
Dose Reports for the selected patient.
3. Test the connection to AW system. In DL Browser, go to the NET tab, click the [Select Host] button.
Click on Dose SR tab.
a. Highlight the remote host in the left side window.
b. Click on the [Verify] button. A message box is displayed indicating the successful Echo to AW host.
32-4-1 Known limitations with AW Reporting Tool display
1. It could be possible that for DAP value’s <10-3 or >107, values displayed in the report are inconsistent with
Innova value displayed with AW Reporting Tool 2.8 and before. Reporting Tool 2.9 to be installed to overcome
the problem.
2. In the Reporting Tool, click on [Options] button In Display options window, set Decimal places to "4".
3. Restart AW.
Live Display
Example:
• in case of 8 degree Trendelenburg (head down) tilt, the tilt icon and 8 degrees will be dis-
played.
• in case of 8 degree Reverse Trendelenburg (head up) tilt, the tilt icon and 8 degrees will
be displayed.
• in case of table horizontal icon will be displayed.
When the geometry angles are displayed using anatomical labels (CRA/CAU, LAO/RAO), the displayed angles
take into account the position of the patient on the table, the rotation of the table and the tilt of the table.
Examples:
• If the table is tilted 5 deg Trendelenburg and the gantry is 10 deg towards head, the display will show CRA 5.
• If the gantry is in the lateral position and the table is rotated 5 deg clockwise, the display will show LAO 90, CRA
5.
When the geometry angles are displayed with mechanical labels, plain gantry angles are used.
33-1-2 Omega Table
When the geometry angles are displayed using anatomical labels (CRA/CAU, LAO/RAO), the displayed angles
take into account the position of the patient on the table and the rotation of the table.
33-2 Status and Image Review
Displayed on left side of the live display to show that the displayed image
is a live fluoroscopic image.
Displayed on left side of the live display to show that the displayed image
is a recorded image.
Displayed on left side of the live display during a Fluoro performed in Au-
to Fluorostore mode.
Displayed on left side of the live display to show that the displayed image
is the frozen last fluoroscopic image.
Displayed on left side of the live display to show that the displayed image,
previously recorded, is reviewed.
Displayed on left side of the live display to show that a mask is available
for the Roadmap fluoro.
Displayed on the left side of the live display to show the system is in
Blended Roadmap and the selected percentage of vessel visibility
Displayed on left side of the live display to show the selected level of 30%
landscape for Subtracted or Roadmap fluoro.
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Forward
nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Forward
half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Forward
high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Backward
nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Backward
half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Play mode Backward
high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously recorded image reviewed in Pause mode (frozen
image)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Forward nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Forward half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Forward high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Backward nominal speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Backward half speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Play mode
Backward high speed (review in loop)
Displayed at the bottom right of the live display to show that the displayed
image is a previously stored fluoroscopic image reviewed in Pause mode
(frozen image)
Displayed on left side of the live display to show that an ECG signal is be-
ing detected
When reviewing a sequence or photo acquired when the table is tilted, a specific "Tilted Table" icon: or
Reference display
Displayed on the right side of the right side of the reference display to
show that disk space is still available.
Displayed on the right side of the right side of the reference display to
show that disk starts to be full.
Blinking icon displayed on the right side of the reference display to show
that disk is near to be full.
Displayed on the right side of the reference display to show that disk is full.
Displayed on the right side of the reference display to show that at least
50% of heat units are still available in the X-ray tube.
Displayed on the right side of the reference display to show that available
heat units in the X-ray tube are between 30% and 50%.
Blinking icon displayed on the right side of the reference display to show
that available heat units in the X-ray tube are below 30%.
Displayed on the right side of the reference display to show that the maxi-
mum heat units are reached in the X-ray tube.
Displayed on the right side of the reference display to show that the Fluoro
is ready (enabled).
Displayed on the right side of the reference display to show that the Fluoro
is not available (disabled).
Displayed on the right side of the reference display to show that the
Record is ready (enabled).
Displayed on the right side of the reference display to show that the
Record is not available (disabled).
Live and reference displays are managed independently, so, in some cases, it is possible to display images from
two different Exams of the same patient.
There is a delay between Photo storage and Photo display. In case of X-ray started before the Photo is displayed,
use Recall Photo to display the last stored Photo.
33-3 Technique factors display accuracy
The relative variation between actual and displayed high voltage (in kV unit) is within [0%; 0.07%].
The relative variation between actual and displayed tube current (in mA unit) is within [+/- 5%].
33-4 Dose Estimation
Error on dose estimation is within +/- 35 % interval in accordance with the international safety standard for Dose
Rate and Total Dose and which is valid down to 100 mGy for dose and 6 mGy/min for dose rate.
Error on Dose Area Product (DAP) estimation is within +/- 30 % interval in accordance with Switzerland legislation
which is valid down 2.5 Gy.cm² for DAP.
Compliance is tested with acquisition or duration longer than 3 seconds.
The estimation of the total dose (cumulated Air Kerma) (mGy) delivered to the patient is continuously displayed and
updated on the reference display. On the live display, the ratio (in percentage) of the actual Total Dose of the Dose
Threshold value is set within the protocol. The Dose Threshold value and a gauge icon visualizing the Dose
percentage of the Threshold are displayed during acquisition. After the Dose Threshold value is reached, the value
is displayed in reverse video. Once the Dose Threshold reaches twice the set Dose Threshold value, a warning
icon is displayed on the corresponding live display next to gauge icon.
Display prior to the predefined Dose Threshold value being reached:
Display after the predefined Dose Threshold value has been exceeded (above 100%):
Display on live display after double of the predefined Dose Threshold value has been reached:
Live Display
Possible states of the gauge icon depending on the actual Total Dose ratio to the Dose Threshold value.
100% of Threshold
Dose Threshold values are set in each protocol. The default Dose Threshold values are:
• 1 Gy Threshold for Pediatric protocols
• 2 Gy Threshold for protocols in head category
• 3 Gy Threshold for others
The Dose Threshold value of each protocol can be configured to any of the following values:
• 0.5 GY
• 1 GY
• 2 Gy
• 3 Gy
• 4 Gy
• 5 Gy
If you want to change Dose Threshold level for any of the protocols, contact your GE Healthcare Representative.
33-5 Monitor Pattern
On the DL keypad, a dedicated key (19) can be used to check the in-room/control room monitors.
Note: The monitor pattern is not available while in review or under Ready for exam status (exam in progress).
• To display the pattern on both live and reference displays, press the blank (19) button.
• To remove the pattern:
1. Create a new patient and start the exam will blacken both screens.
2. Start a review of any sequence will remove the pattern from the live display.
3. Recall a photo will remove the pattern from the reference display.
33-5-1 Monitor Contrast Adjustment Check
The Contrast Adjustment Tool is used to set monitor brightness and contrast levels. They are correctly adjusted
when the 5% square inside of the black or white squares is visible. If this is not visible, brightness and contrast may
need to be readjusted. Monitor changes are to be performed only by service.
33-6 Stopwatch
The Stopwatch when selected on the Innova Central touch screen has the capabilities of Start, Pause, Restart the
timer and Exit the function. Exit will abort the stopwatch and remove the display from the reference display's lower
left corner. Reclicking on Stopwatch button will close the window but leave the stopwatch displayed in the mode it
was last using.
The electronic shutter is always displayed few millimeters larger than the image area to keep the collimator blades
always visible during Fluoro or Record acquisition.
Note: In case collimator blades are moved during an acquisition, the electronic shutter displayed in review and
pushed will correspond to the largest position of the blades.
Using the Shutter button located on the DL Keypad it is possible to activate or not the electronic shutter in review.
In review, the electronic shutter is applied as close as possible to the image area to hide the collimator blades
completely.
Note: Only images acquired with electronic shutter can be reviewed with or without electronic shutter. For images
acquired without electronic shutter the Keypad’s Shutter button is inactive.
TM
Chapter 9 - InnovaSpin (Option)
The InnovaSpin option allows to perform an acquisition while the gantry rotates around its horizontal axis (pivot).
Rotation from RAO to LAO can be performed at up to 40°/s.
Rotation from LAO to RAO can be performed at up to 30°/s.
To prevent any collision during the rotation a test rotation at 10°/s must be performed first.
Time for gantry to speed up and slow down is included in the rotation trajectory.
1 Application Description
InnovaSpin is a software option that permits fast spin rotational angiography.
2 Safety Recommendations
1. Secure the patient on the table using appropriate restraints to prevent any motion during the gantry rotation.
2. Explain the procedure to the patient and ask him not to move during the gantry rotation
3. Clear the rotation area to prevent any collision during the rotation
4. In case of any problem during the rotation, releasing the exposure foot switch will activate the emergency break
of the gantry.
– At the end of the rotation, the gantry motion is stopped and X-ray can still be performed until the exposure
footswitch is released
• Releasing the exposure footswitch will exit from the InnovaSpin mode. "Dynamic" or "Chase" is displayed back
on the reference display.
Note: If you choose to keep shooting X-ray once the InnovaSpin is finished, some gantry vibrations might occur
and be visible on the image depending on the trajectory performed.
4 Trajectories Settings
Spin trajectories are preloaded in the Innova system.
However, gantry Start position, End position and rotation speed can be modified for each trajectory.
If needed, contact your GE Healthcare Representative for any trajectory modification.
5 User Messages
Information and error messages are displayed in the status and Error Message areas located at the bottom of the
live display.
"SELECTION NOT ALLOWED, ACQUISITION IS IN • You tried to enter spin mode with an X-ray switch-
PROGRESS" appears in error area. pressed. This in not allowed.
or
• You tried to reselect a trajectory with an X-ray switch
pressed. This is not allowed.
"SELECTION NOT ALLOWED, RELEASE TEST BUT- • You tried to enter spin mode with Test button-
TON" appears in error area. pressed. This is not allowed.
or
• You tried to reselect a trajectory with Test button-
pressed. This is not allowed.
"SELECTION NOT ALLOWED, TABLE IS ROTATED" You tried to enter spin mode with Table rotated. This is
appears in error area. not allowed.
"RECORD ACQUISITION NOT ALLOWED, SPIN TEST You tried to perform a record during test phase. This is
IS IN PROGRESS" appears in error area. not allowed.
"RELEASE FLUORO SWITCH FOR SPIN ACQUISI- You tried to select a new spin position during a Fluoro
TION" appears in error area. acquisition. This is not allowed.
"RELEASE TEST BUTTON FOR SPIN ACQUISITION" After test phase completed, you tried to start acquisition
appears in error area. with test button already pressed. This is not allowed.
"SPIN SETUP ABORTED, RESELECT TRAJECTORY" You got out of spin mode before performing spin acqui-
appears in error area. sition.
This can happen if:
• There was a user gantry or a table motion between
end and start position during test phase.
or
• There was a collision between end and start position
during test phase.
or
• You exited operator "F".
or
• You selected a new protocol.
or
• 5 minutes has been elapsed after a spin test.
"SPIN ACQUISITION ABORTED, RESELECT TRA- You stopped acquisition during Spin acquisition.
JECTORY" appears in error area. There is no error or status message and you cannot se-
lect the F operator: You tried to select InnovaSpin mode
but other than a Cardiac or the InnovaChaseTM protocol
is selected. This is not allowed.
"READY FOR SPIN TEST" appears in status field. You just entered Spin mode.
"SPIN TEST IN PROGRESS" appears in status field. Spin safe test is running.
"READY. CLEAR ROTATION AREA" appears in status You are ready to start acquisition then launch Spin.
field.
# Symbol Name
1 Table Panning: press down first, then
pan.
The Positioner Enable/Disable button only enables/disables commands from the Smart
Box/Handle and Table Side Status Control. It does not enable/disable table top brakes
commands from the Fluoro/Record footswitch or the Table Panning handle.
Each joystick is equipped with a double activation (Enable and Command) system to avoid unintended activation of
the motion control commands. To activate a joystick, press down first to activate the Enable, and then apply the
desired Command.
Note: In case of collision between two mechanical parts or with the patient, with activation of a collision sensor (see
Innova Positioner / Collision Sensors), the anti-collision software stops the gantry. Release the enable trigger and
press it again. Gantry motion will automatically reverse to clear the collision. After the collision was cleared, call
Service in case the gantry moves permanently at low speed.
In case of collision between two mechanical parts or with the patient, without activation of a collision sensor, the
anti-collision software progressively slows down the gantry motion. An audible signal is activated and the visual
gantry display located at the bottom left of the live display is highlighted. The command of the gantry or table must
be released to prevent any collision.
Note:The Smart Box is equipped with protection barriers to prevent as much as possible the risk of unintended
motions. Do not remove the protection barrier.
1-2 InnovaIQ Table Smart Box
# Symbol Name
1 Table Panning Assist Joystick: press
down first, then pan.
Note: In case of collision between two mechanical parts or with the patient, with activation of a collision sensor (see
Innova Positioner / Collision Sensor), the anti-collision software stops the gantry. Release the enable trigger and
press it again. Gantry motion will automatically reverse to clear the collision. After the collision was cleared, call
Service in case the gantry moves permanently at low speed.
In case of collision between two mechanical parts or with the patient, without activation of a collision sensor, the
anti-collision software progressively slows down the gantry motion. An audible signal is activated and the visual
gantry display located at the bottom left of the live display is highlighted. The command of the gantry or table must
be released to prevent any collision.
1-3 Table Panning Assist Joystick (item 1)
1 - The table automatically moves Up or Down and the longitudinal table top automatically moves toward
Head and Foot to keep the ROI centered when tilting the table.
2 - The gantry pivot and C-arm automatically move to keep the X-ray beam angulation related to the patient
constant when tilting the table.
3 - If the user pans and tilts the table while in the Anatomical mode, the Isocenter Tracking mode will
deselect and only keep the Incidence keeping mode active to allow the tilt of the gantry to follow the
panning and tilt of the table simultaneously.
Whenever tilting the table more than 5 degrees, explain to the patient that it is normal for
him/her to feel slight movement on the table. This will assure the patient and prevent any
unwanted movement during imaging.
When the Machine mode is activated (button 5 light ON), all table and gantry axes work separately.
In such mode, the ROI is not kept centered and the gantry angulation is not preserved when the table is tilted.
Moving the joystick Right/Left will only tilt the table.
1-5 Emergency Stop (item 11)
In case of emergency, press the Emergency Stop button.
THE STOP BUTTON (11) LIGHT PERMANENTLY OFF INDICATES A FAILURE IN THE
EMERGENCY STOP BUTTON MECHANISM. HANDLE THE SYSTEM WITH CARE AND
CONTACT YOUR GEHC SERVICE REPRESENTATIVE AS SOON AS POSSIBLE.
1-6 Omega Smart Handle (Not available with InnovaIQ Table)
# Symbol Name
1 Table Control Up/Dow, Motorized
Longitudinal Tabletop: press down
first, then tilt.
2 Detector Elevator Control Up/Down:
press down first, then tilt Up/Down.
Move the joystick up or down to raise
or lower the detector.
3 C-Arm Gantry Motion: Press Gantry
Enable first, then press button right or
left.
4 Table Panning/Pivot Gantry Motion:
press respective Enable first, then ro-
tate.
5 Anatomic/Machine Mode Control.
Pressing this button allows to toggle
between Anatomical and Mechanical
modes. The selection affects the be-
havior of the positioning joysticks.
In Mechanical mode, each axis of the
gantry and table are controlled inde-
pendently.
In Anatomical mode, the motion of the
axes is combined in order to allow
motion with respect to the patient.
6 L-Arm Control: squeeze first, then tilt
to position the L-Arm.
Note: In case of collision between two mechanical parts or with the patient, with activation of a collision sensor (see
Innova Positioner / Collision Sensor), the anti-collision software stops the gantry. Release the enable trigger and
press it again. Gantry motion will automatically reverse to clear the collision. After the collision was cleared, call
Service in case the gantry moves permanently at low speed.
In case of collision between two mechanical parts or with the patient, without activation of a collision sensor, the
anti-collision software progressively slows down the gantry motion. An audible signal is activated and the visual
gantry display located at the bottom left of the live display is highlighted. The command of the gantry or table must
be released to prevent any collision.
1-7 Manual L-Arm Positioning
Always use the L-Arm brake release button (1) to manually position the L-Arm. The brake release is located on a
handle at the upper end of the L-Arm. Press one or both side buttons on the handle to release the brakes and
manually position the L-Arm.
You can freely move the L-Arm to its limits in either direction. A built-in hard stop limits travel.
Note: In case of failure of the L-Arm brake, it is still possible to overcome the L-Arm brake by pushing with
sufficient strength so clear the patient as needed in emergency situations.
In normal situation avoid pushing the L-Arm without using the brake button, as it can create
a small drift of the "zero" position of the L-Arm, which can impact the Innova Vision
applications.
Note: With the InnovaIQ Table, the activation of the switch will only release the lateral table top brake when the
table is tilted.
1-9 Omega Table Side Status Control (TSSC)
Note: In case a second TSSC is used in control room, all commands driving any gantry motorized axis are disabled
(i.e.: gantry Test rotation).
In case of Emergency, press the Emergency Button (10). All Gantry and table motions will stop.
To reset, press the RESET Button at TSSC underside.
Note: When a second TSSC is used in the control room, the Lateral/Longitudinal brakes (button 16) and/or the
table Rotation brakes (button 18) are disabled.
Note: In case the Lateral/Longitudinal brakes (button 16) and or the table Rotation brakes (button 18) are activated
to release the brakes, these brakes will be automatically reengaged in case no motion is detected within the next 3
minutes.
In case of emergency, press the Emergency button (10). All Gantry and table motions will stop.
To reset, press the RESET button at TSSC underside.
Note: When the Stop button (10) is activated, the button is fully highlighted.
When the Stop button (10) is not activated, the button light is dimmed and fluctuates to show that the button is alive
and in working condition.
THE STOP BUTTON (10) LIGHT PERMANENTLY OFF INDICATES A FAILURE IN THE
EMERGENCY STOP BUTTON MECHANISM. HANDLE THE SYSTEM WITH CARE AND
CONTACT YOUR GEHC SERVICE REPRESENTATIVE AS SOON AS POSSIBLE.
1-11 Table brakes button management
Table brakes behavior versus button's state:
• Omega Table
If you press on the Table longitudinal/lateral button or Table rotation button, the table release brake function, will
be immediately activated: the user will be able to move the table manually.
• InnovaIQ Table
If you press on the Table Brake longitudinal/lateral button or Table rotation button, you will have to release the
button first to activate the selected function.
If you don't release it and continue to press, NO Table Release Brake function will be activated: the user will not
be able to move the table manually, the table will still be locked.
1-12 Tableside Cart (Option)
1-12-1 General Description
The Tableside Cart is delivered with two accessory rails, each designated to hold up to two GE Healthcare Table
Side User Interface (TSUI). Depending on the configuration, the Tableside Cart can be installed with one or two
accessory rails.
There are four configurations of the Tableside Cart with one accessory rail. The approximate height from the floor
to the rail for these four configurations is the following:
• 75.5 cm (29.7 in)
• 82 cm (32.3 in)
• 98 cm (38.6 in)
• 104.5 cm (41.2 in)
There are two configurations of the Tableside Cart with two accessory rails. The approximate height from the floor
to the rail for these two configurations is the following:
• Bottom rail: 75.5 cm (29.7 in) Top rail: 98 cm (38.6 in)
• Bottom rail: 82 cm (32.3 in) Top rail: 104.5 cm (41.2 in)
Contact your GE Service Representative for any mechanical modification.
Handles located on the front and on each side allow moving the Tableside Cart in any direction.
Two brakes located on the front side of both front wheels, can be used to immobilize the Tableside Cart when
needed.
1-12-2 Environmental working conditions
Because the TSUI’s cables are connected to the table, the Tableside Cart must remain near the InnovaIQ table at
a distance of 2 or 3 m around the table base, depending on the cable lengths.
If one of the TSUIs mounted on the Tableside Cart has a cable length of 3 m (118 in), the Tableside Cart must
remain at a maximum distance of 2 m around the table base.
If all the TSUIs mounted on the Tableside Cart have a cable length of 4 m (157.5 in), the Tableside Cart must
remain at a maximum distance of 3 m around the table base.
1-12-3 How to move the Tableside Cart
Release first both front wheel brakes by lifting up the brake command. Then use the frontal handle or handles on
each side of each accessory rail to move the Tableside Cart in any direction.
When the Tableside Cart is positioned at the proper location, push down the brake command located in front of
1 Brake released
2 Brake engaged
3 Handles used to move
BEFORE USING THE TSUIs ON THE TABLESIDE CART, BE SURE THEY ARE WELL
CLAMPED ON THE RAILS. FAILURE TO DO SO MAY LEAD TO STAFF INJURY DUE TO
TSUIs FALLING.
When using the Smart Box on the Tableside Cart to move the gantry or the table, keep the
Tableside Cart accessory rails parallel to the table accessory rails or to the Tabletop end.
1-12-5 Cable Routing
The cable routing on the Tableside Cart is designed to avoid TSUIs cables hanging down from the rails or lying on
the floor.
Two cable strips are used for tying TSUI cables on the Tableside Cart and four cable winders are used for winding
the TSUI cables to minimize the cables lying on the floor.
1 Cable winder
2 Cable tie
ALWAYS ENSURE THE TSUI CABLES FROM THE TABLESIDE CART ARE NEVER IN A
LOCATION WHERE PATIENT OR STAFF COULD TRIP. USE THE FOUR CABLE WINDERS TO
REDUCE THE MINIMUM LENGTH OF THE TSUI CABLES LYING ON THE FLOOR.
1-12-6 Safety mechanism
The shape and size of the front and side handles is designed to avoid unexpected table and gantry motion due to
an unwanted activation of the TSUI in case of collision between the Tableside Cart and other equipment.
Wheels are designed to prevent any unexpected X-ray pedal activation and to roll over any TSUI cables.
THIS TYPE OF DYSFUNCTION MAY CAUSE UNEXPECTED GANTRY MOTION AND RESULT
IN SEVERE INJURY TO THE PATIENT OR PEOPLE AROUND THE GANTRY.
AS SOON AS THE ERROR CODE IS DISPLAYED ON THE CONTROL PANEL (TSSC), YOU MAY TERMINATE
YOUR EXAM WITH CAUTION AND MUST REPORT TO YOUR GE SERVICE REPRESENTATIVE
IMMEDIATELY.
1-14 Remote Stand (Option)
The Remote Stand is installed in the control room by your GE Service Engineer. It is used to mount the Smart Box
and TSSC in the control room allowing movement of the gantry and table remotely. The Remote Stand can be
configured to have the main column either to the left or right side of the rail arms. The direction will be selected
based on the position it is mounted on the counter to make the controls easily accessible to the user but not
extending beyond the counter surface. It is important that the stand and controls are always in a position that the
patient and equipment are clearly visible and the view unobstructed to the operator. Keep the surrounding area free
of objects that could fall on the controls and cause an uncontrolled motion.
The Remote Stand is equipped with a knob to lock or unlock the rotation of the two arms:
• Turning the knob counter clockwise will unlock the arms and allow rotation around the axis of the stand. Adjust
the position of the arms as needed.
• Turning the knob clockwise will lock the rotation of the arms and hold them in the desired position.
Note: Do not use the joysticks of the Smart Box or TSSC to move the stand.
ALWAYS ENSURE THAT THE ROTATION OF THE REMOTE STAND IS LOCKED BY PULLING
ON ITS ARM BEFORE USING THE EQUIPMENT. FAILURE TO DO SO MAY RESULT IN
SEVERE PATIENT OR STAFF INJURY DUE TO AN UNCONTROLLED MOTION.
1-15 InnovaIQ Cable Holder
The cable holder for the InnovaIQ Table encloses and guides the Table Side User Interface (TSUI) cables during
the table movements and thereby protects the cables from entangling and getting cut during table movements.
• The sheath can be moved from one side of the table to the other without opening it.
• It can be opened to selectively add or remove cables.
• It can be removed for cleaning.
THE CABLE HOLDER WILL BE EFFECTIVE ONLY WHEN THE SHEATH IS PROPERLY
INSTALLED AS PER THE OM INSTRUCTIONS.
DO NOT USE THE GUIDE ROD TO ATTACH DEVICES OTHER THAT THE CABLE SHEATH.
DO NOT USE THE GUIDE ROD TO ATTACH CABLES WITHOUT USING THE SHEATH.
MAKE SURE NO OBJECT LIES UNDER THE TABLE, WHICH THE CABLES COULD GET
CAUGHT INTO.
The following TSUI cables shall always be routed through the Cable Holder when installed on the table rail.
• Table Side Status Control (TSSC)
• Smart Box
• Table Panning Device (TPD)
WHEN USING THE TABLESIDE CART, ALWAYS RELEASE THE CABLE LOOPS OF THOSE
CABLES ON THE TABLESIDE CARTOR THE CABLES WILL BE TOO SHORT TO LAY ON
THE FLOOR.
IF, AFTER USING THE TABLESIDE CART, THE DEVICES ARE BROUGHT BACK TO THE
TABLE RAIL, THE EXTRA CABLE LENGTH MUST BE LOOPED AGAIN TO AVOID CABLES
GETTING CAUGHT.
• If all the devices are moved to the TableSide Cart, the sheath can be left on the cables.
• If only some devices are moved to the cart, the cables of those devices should be taken out of the sheath.
1-15-4 How to Remove and Install the Sheath
Note:
While fixing the sheath zipper on each side, ensure all the cable markers (see illustration
Cable Markers, item 1) are aligned over the end of the zipper.
Cable Markers
Note: If in case the two sheaths are separated, connect them with the two flap buttons.
THE CABLE HOLDER WILL BE EFFECTIVE ONLY WHEN THE SHEATH IS PROPERLY
INSTALLED AS PER THE OM INSTRUCTIONS.
DO NOT LEAVE ANY TSUI OR GROUNDING CABLES OUTSIDE THE SHEATH WHEN CMD IS
IN USE.
MAKE SURE NO OBJECT LIES UNDER THE TABLE, WHICH THE CABLES COULD GET
CAUGHT INTO.
The following TSUI cables shall always be routed through the Cable Holder when installed on the table rail.
• Table Side Status Control (TSSC)
• Smart Box
• Table Panning Device (TPD)
• Innova Central touch screen
• In-room 3D Mouse
• Ground wire from table
• Injector (Mark V Table Mounted)
1-16-1 How to swap TSUIs
When side sheath option is used, follow below instructions to swap accessories:
• Unhook the accessory cable from rail hook.
• Remove the side sheath zipper; take only the desired cable out from the internal strap and button it back, then
close the zipper.
1-16-2 How to Rotate the Table beyond 90 deg. when Cable Holder is in use
• Remove all ferrules from both side Cable Holder (with Cable Holder option); or unhook from the rails the two
hooks of the side sheath (with side sheath option).
• When the table is rotated back to its home position, make sure that all ferrules and hooks are placed back into
their original positions.
1-16-3 How to Use with TableSide Cart
WHEN USING THE TABLESIDE CART, ALWAYS RELEASE THE CABLE LOOPS OF THOSE
CABLES ON THE TABLESIDE CARTOR THE CABLES WILL BE TOO SHORT TO LAY ON
THE FLOOR.
IF, AFTER USING THE TABLESIDE CART, THE DEVICES ARE BROUGHT BACK TO THE
TABLE RAIL, THE EXTRA CABLE LENGTH MUST BE LOOPED AGAIN TO AVOID CABLES
GETTING CAUGHT.
• If all the devices are moved to the TableSide Cart, the sheath can be left on the cables.
• If only some devices are moved to the cart, the cables of those devices should be taken out of the sheath.
1-16-4 How to Remove and Install the Sheath
1-16-4-1 Detaching the Sheath
• Remove the side sheath zipper, unbutton the internal straps and take the cables out.
• Unhook the side sheath from the rail
• Remove the center sheath and middle sheath zippers to remove the center sheath.
• Unhook the sheath metal holder from the two supports.
• Put the cables inside the internal straps, and close the side sheath zipper.
• Mount the side sheath hooks on the channel block of rail.
1-16-5 How to Loop the Extra Cable Length
Once the center sheath is installed, the extra length of cables coming out of it has to be managed through cable tie
and holder.
• Use the cable tie to group cables on rest point bracket. Provide extra length of ~120mm from cable tie to the
center sheath end.
Note: In case of shortage of length of cables, adjust the cables in the center sheath to accommodate for extra
length requirement
• Use cable tie and holder to carry the loop of cables inside the rest point board.
• Confirm all TSUI connections are good; Close the rest point board with cover.
2 Collimator Control
The collimator is controlled automatically when operator selects a Field Of View on the TSSC.
The collimator may also be operated manually at the TSSC.
There is a permanent safety mechanism which ensures automatically for the operator that the X-ray beam will
never expose patient body areas which are not displayed on the monitor (SID and FOV tracking).
1 Not Used
2 Not Used
3 Lateral Blades Closed
4 Lateral Blades Open
5 Longitudinal Blades Open
6 Longitudinal Blades Closed
7 Auto-Collimation
Auto-Collimation: Provides the means to quickly open collimator blades to maximum exposure area for the
selected FOV. When in auto-collimation, LEDs on both iris blades control buttons are ON. As soon as the collimator
blades are manually entered in the FOV, LEDs go OFF.
There are two ways to visualize the location of the collimator blades:
• While performing Fluoro:
Move the collimator joystick to visualize and position real time the collimator blades over the selected anatomy.
• Out of Fluoro:
Move the collimator joystick to visualize and position virtual lines displayed over the Last Image Hold (LIH) or
over any image in review on the live display.
Note: Virtual lines displayed represent the current location of the collimator blades and may not match, in case
of gantry or table motion, with the anatomy of an image previously acquired.
These virtual lines are displayed just few seconds over the image each time the collimator blades are moved
either manually using the collimator joystick or either automatically when changing the gantry SID or detector
FOV.
Note: There is always a short delay between the time the collimator blades move and the time the position of
the virtual collimation lines is updated on the image. If X-ray is performed before the position of the virtual
collimation lines stabilizes, the X-ray beam might not perfectly match the area shown by the virtual collimation
lines.
Recommendation: wait for the virtual collimation lines to stabilize before performing fluoro or record
acquisition.
Gantry informations linked to the acquired image and displayed in the bottom right corner of the live display are
removed during the few seconds the virtual collimator lines are displayed.
Note: Collimator blades can be moved during DSA, Chase and Dynamic Record acquisitions. Collimator blades
controls are disabled during Bolus and 3D Record acquisitions to allow further successful image reconstruction.
2-1 Contour Filter Control
8 Operator Selector
To store a position:
• Move the gantry in the position to store.
• Using the Operator Selector (button 8), select an operator number.
• Using the Auto Positioner Selector (button 3), select a position number. The button light goes On.
• Press the store icon (button 6) to store the gantry position, the light of the position number goes Off then On
again to show that the storage is performed.
Note: If no storage action is performed within 30 seconds after the position number is selected, the light of the
position number will automatically go Off. In such case, reselect the position number again.
To recall a position:
• Using the Operator Selector (button 8), select an operator number.
• Using the Auto Positioner Selector (button 3), select a position number. The button light goes On and the gantry
position previously stored is displayed in reversed video on the live display.
• Using the Auto Positioner joystick, press down first then tilt the joystick Up or Down to move the gantry to the
previously stored position. At the end of the motion, the information "Position Reached" appears in the lower left
corner of the live display.
3-3 2D Send Angles
The 2D Send Angles function allows to position the gantry and the SID back to where an image was previously
acquired.
First, select the image of interest. The image you can send the angles from can be:
• A photo under review on the reference display.
• The image of a sequence under review on the live display.
• The image of a sequence or a photo under review on the In-room Browser.
Note: Particular photos, such as SMPTE photos, are not supported.
Note: Sequences under review need to be paused prior to sending the angles.
Note: When sending the angles from a sequence of the In-room Browser, the angles of the 1st frame of the
sequence are sent.
Then, send the angles of the selected image to the Auto Positioner, by:
• Pressing the Store Angle button on the Remote Control (see Remote Control illustration).
• Pressing the Send Angles button on the Review tab of the Innova Central touch screen (see Innova Central
touch screen illustration).
The Auto Positioner Selector displays the "0" Operator, and the Position lights turn On.
The angles sent are displayed and high-lighted on the live display. They are displayed in the same display mode
(anatomical / mechanical) as the current selection. See illustrations below.
Note: The angles sent remain high-lighted until a gantry or table movement is initiated.
Note: The Send Angles may be cancelled in the following situations:
• 30 s time-out (no action during 30 s after the Send Angles request).
• Protocol or fluoro mode change.
• Patient position change.
• Auto Positioner Operator new selection.
In that case, a message “POSITION NOT REACHED” appears in the lower left corner of the live display, the angles
come back to normal display, and the Auto Positioner Selector comes back to the previous setting.
Note: The Send Angles request may fail in the following situations:
• System is configured in Bolus mode, and the start position has been selected.
• System is configured in 3D or Spin mode, and the test phase has started.
• Positioner is unable to reach the final position.
• Positioner is in degraded mode.
In that case, a message “POSITION NOT REACHED” appears in the lower left corner of the live display.
Finally, press the Auto Positioner joystick down and tilt it upward or downward to move the gantry to the
previously stored position. When the gantry reaches the final position, the message "POSITION REACHED"
appears in the lower left corner of the live display.
Note: In case of collision, the gantry motion is stopped, Send Angles is cancelled, and the message "POSITION
NOT REACHED" appears in the lower left corner of the live display.
Note: In order to avoid collision, gantry motion can be stopped by releasing the Auto Positioner joystick. Move the
table and re-start the motion by pressing the Auto Positioner joystick again.
To store a position:
• Move the gantry and/or table in the position to store.
• Using the Operator Selector (button 8), select an operator number.
• Using the Auto Positioner Selector (button 3), select a position number. The button light goes On.
• Press the gantry icon (button 7) to only store a gantry position, the light of the position number goes Off then On
again to show that the storage is performed.
• Press the gantry + Table icon (button 6) to store gantry and table position, the light of the position number goes
Off then On again to show that the storage is performed.
• Press on the Table icon (button 8) to only store the table position, the light of the position number goes Off then
On again to show that the storage is performed.
Note: If no storage action is performed within 30 s after the position number was selected, the light of the position
number will automatically goes Off. In such case, reselect the position number will automatically go Off.
To recall a position:
• Using the Operator Selector (button 8), select an operator number.
• Using the Auto Positioner Selector (button 3), select a position number. The button light goes On and the
gantry, or table, or gantry + table position previously stored is displayed in reversed video on the live display.
• Using the Auto Positioner joystick, press down first then tilt the joystick Up or Down to move the gantry, or table,
or gantry + table to the previously stored position. At the end of the motion, the information "Position Reached"
appears in the lower left corner of the live display.
Note: The safety of everyone near the moving equipment depends on the operator at the controls. When
positioning the equipment using the Auto Positioning function, continuously observe all moving equipment, and
stop the motion before a collision can occur. The sensors are provided for added protection in case of an
inadvertent contact and are not to be relied on as a normal means for stopping the motion. Improper use of the
equipment while using the Auto Positioning function, could result in serious injury or equipment damage.
4-3 2D Send Angles (Gantry, Table, Gantry+Table)
The 2D Send Angles function allows to position the gantry (including SID), the table, or both the gantry and the
table back to where an image was previously acquired.
First, select the image of interest. The image you can send the angles from can be:
• A photo under review on the reference display.
• The image of a sequence under review on the live display.
• The image of a sequence or a photo under review on the In-room Browser.
Note: Particular photos, such as SMPTE photos, are not supported.
Note: Sequences under review need to be paused prior to sending the angles.
Note: When sending the angles from a sequence of the In-room Browser, the angles of the 1st frame of the
sequence are sent.
Then, send the angles of the selected image to the Auto Positioner, by:
• Pressing the Store Angle button on the Remote Control (see Remote Control illustration).
• Pressing the Send Angles button on the Review tab of the Innova Central touch screen (see Innova Central
touch screen illustration).
The Auto Positioner Selector displays the "0" Operator, and the Position lights turn On.
The angles sent are displayed and high-lighted on the live display. They are displayed in the same display mode
(anatomical / mechanical) as the current selection. See illustrations below.
Note: The angles sent remain high-lighted until a gantry or table movement is initiated.
Note: The Send Angles may be cancelled in the following situations:
• 30 s time-out (no action during 30 s after the Send Angles request).
• Protocol or fluoro mode change.
• Patient position change or table tilt change.
• Auto Positioner Operator new selection.
In that case, a message “POSITION NOT REACHED” appears in the lower left corner of the live display, the angles
come back to normal display, and the Auto Positioner Selector comes back to the previous setting.
Note: The Send Angles request may fail in the following situations:
• System is configured in Bolus mode, and the start position has been selected.
• System is configured in 3D or Spin mode, and the test phase has started.
• Positioner is unable to reach the final position.
• Positioner is in degraded mode.
In that case, a message “POSITION NOT REACHED” appears in the lower left corner of the live display.
In order to reposition:
• The gantry, press the "Gantry" Store button of the Auto Positioner Selector (refer to button 7 of the Auto
Positioner Selector picture above). The "Gantry" Store button turns On, and the other Store buttons turn
blinking. The angles get high-lighted on the live display.
• The table, press the “Table” Store button of the Auto Positioner Selector (refer to button 8 of the Auto
Positioner Selector picture above). The "Table" Store button turns On, and the other Store buttons turn blinking.
The table information gets high-lighted on the live display.
• Both the gantry and the table, press the "Gantry+Table" Store button of the Auto Positioner Selector (refer to
button 6 of the Auto Positioner Selector picture above). The "Gantry+Table" Store button turns On, and the
other Store buttons turn blinking. Both the angles and table information get high-lighted on the live display.
By default, the Send Angles is configured as Gantry Send Angles.
"Gantry" Send Angles is active "Table" Send Angles is active "Gantry+Table" Send Angles is active
Note: Only table height and tilt axis are displayed high-lighted. Other axis are not shown.
Note: Gantry Send Angles repositions all axis of the Gantry, including the SID. It considers the current table tilt
position to reach the same incidence as the reviewed image.
Note: Table Send Angles repositions the Longitudinal, Lateral, Height and Tilt axis of the table. SID automatically
goes to max SID in order to avoid collision. In case patient contouring is active, SID is adjusted to the patient
position.
Note: Gantry+Table Send Angles repositions all axis of the Gantry, the SID and the Longitudinal, Lateral, Height
and Tilt axis of the table.
Finally, press the Auto Positioner joystick down and tilt it upward or downward to move the gantry, table, or
gantry+table respectively to the previously stored position. When the gantry, table, or gantry+table reach the final
position, the message "POSITION REACHED" appears in the lower left corner of the live display.
Note: In case of collision, the motion is stopped, Send Angles is cancelled, and the message "POSITION NOT
REACHED" appears in the lower left corner of the live display.
Note: In order to avoid collision, motion can be stopped by releasing the Auto Positioner joystick. Move the system
and re-start the motion by pressing the Auto Positioner joystick again.
Raise (only) the Digital Detector by pressing one of the Emergency Back-Out buttons located on the Digital
Detector (See Illustration).
Note: This control must be used only:
1. To clear a collision Level 2 (see Innova Positioner / Collision Sensors) by allowing a 2.5 cm additional SID
course.
2. To move away the Digital Detector from the patient in case of gantry failure.
3. In case of SID failure. Use this button to bring the Digital Detector to its highest position to still allow X-ray.
5-3 InnovaTraceTM (on Innova IGS 520 and Innova IGS 530 only)
The Digital Detector housing is surrounded by a capacitive sensor ring (1) to prevent as much as possible collision
with the patient.
WITH SYSTEMS USING THE OMEGA TABLE, THE REAR OF THE DETECTOR LIFT IS NOT
PROTECTED AS WELL AS THE OTHER PARTS OF THE GANTRY. SPECIAL ATTENTION
MUST BE PAID TO THE PATIENT WHEN WORKING WITH THE L-ARM POSITIONED
AROUND 90° FROM THE TABLE AND A DEEP CAUDAL GANTRY ANGULATION IS USED. IN
SUCH POSITION, IT MAY HAPPEN THAT CAPACITIVE SENSORS WILL NOT DETECT SOON
ENOUGH PATIENT ANATOMY. ALWAYS WATCH PATIENT AND GANTRY WHEN L-ARM IS
AROUND 90° FROM THE TABLE AND GANTRY MOVED IN DEEP CAUDAL ANGULATION.
NARROW AND SMALL HUMAN OBJECTS (I.E. NOSE) ARE MORE DIFFICULT TO DETECT
BY THE SENSORS. SO IT IS RECOMMENDED TO TURN PATIENT'S HEAD TO THE LEFT OR
RIGHT AND TO REDUCE GANTRY SPEED WHEN IN THE VICINITY OF THE PATIENT'S
HEAD.
5-4 InnovaSense patient contouring (option on Innova IGS 520 and Innova IGS 530 only)
InnovaSense patient contouring incorporates implementation of advanced capacitive sensiting technology.
The system in this mode is aware of location of the patient and can dynamically adapt the receptor distance to
position receptor safely in proximity to the patient.
The Digital Detector housing is surrounded by a capacitive sensor ring (1) to prevent as much as possible collision
with the patient.
NARROW AND SMALL HUMAN OBJECTS (I.E. NOSE) ARE MORE DIFFICULT TO DETECT
BY THE SENSORS. SO IT IS RECOMMENDED TO TURN PATIENT'S HEAD TO THE LEFT OR
RIGHT AND TO REDUCE GANTRY SPEED WHEN IN THE VICINITY OF THE PATIENT'S
HEAD.
When InnovaSense patient contouring is activated (Item 8 on Smart Box/Smart Handle), the Digital Detector
always stays at a constant safe distance from the patient while the gantry is moved in any direction. At the end of
the motion, keeping the gantry enable pressed, will move the detector down for even closer positioning (stops at
few cm from patient).
The Digital Detector will also move Up/Down in response to the table movement providing a smart system
response.
With InnovaSense patient contouring activated, bringing the detector at its highest position will prevent it to
automatically move down during a table Up/Down motion and /or longitudinal tabletop motorization.
When using the Auto Positioner, the Digital Detector will automatically move Up/Down during gantry motion to stay
at safe distance from the Patient. There is a button provided to activate or disable this function (Item 8).
BE DISABLED (ITEM 8) PRIOR MOVING THE GANTRY EITHER MANUALLY OR WITH THE
AUTO POSITIONER; THESE OBJECTS/DEVICES BEING TOO SMALL AND/OR NOT
DETECTED BY THE CAPACITIVE SENSORS. USE THE "NEEDLE PLACEMENT" PROTOCOL
LOCATED EITHER IN THE ABDOMEN OR CARDIAC & COMBO ANATOMY WHICH
AUTOMATICALLY DISABLES THE INNOVASENSE PATIENT CONTOURING FUNCTION.
FAILURE TO THIS MAY CAUSE SEVERE PATIENT INJURY IN CASE THE DETECTOR HITS
THE INVASIVE OBJECT/DEVICE.
A "Needle Placement" protocol, located in the "Cardiac & Combo" and "Abdomen" anatomy automatically disables
the InnovaSense patient contouring function and forbids its activation.
Such protocol must be selected prior to start any procedure that involve the use of small invasive objects/devices
which project out from the patient skin surface.
Note: When using an Angio protocol (located out of the "Cardiac & Combo" anatomy), the InovaSense patient
contouring function is automatically disabled at each new protocol selection or end of exam activation.
Note: Manual control of the Digital Detector is available at any time.
THE RECEPTOR WILL NOT AUTOMATICALLY MOVE OUT OF THE WAY DURING MANUAL
PANNING UNLESS GANTRY ENABLE IS PRESSED.
6 Collision Sensors
The sensor pads surrounding the X-ray tube and Digital Detector contain pressure-sensitive collision sensors.
There is one level of collision detection for the X-ray tube sensor (see Innova Positioner / Collision Sensors / Level
1), and there are two levels for the Digital Detector sensor (see Innova Positioner / Collision Sensors / Level 1 and
Innova Positioner / Collision Sensors / Level 2).
With the InnovaIQ Table, additional sensors are located at the rear of the detector lift.
6-1 Level 1
During normal operation, Level 1 sensors activate immediately upon sensor detection of collision.
Note: Certain environmental pressure changes or operations such as cleaning may cause the system to detect a
Level 1 collision proactively under some circumstances. In most cases, the Level 1 indication would automatically
clear after a few minutes. If the Level 1 persists, the collision system can be reinitialized by pressing the detector
bumper with both hands for a few seconds and then releasing it. If both these methods fail, perform a system reset
to clear the Level 1 indication.
Activation:
When Level 1 collision sensors are activated, corresponding parts of the visual gantry display located on the left
side of the live display are highlighted and flashing. An audible signal is activated. Motions and drives for the
positioner and the table stop.
Clearing Procedure:
After a Level 1 collision sensor activates, remove the obstruction to allow the sensor to clear itself automatically. If
you are unable to remove the obstruction and the system must be driven to clear the sensor, perform the following
procedure:
1. Release the Smart Handle or Smart Box motion enable button. Press the button again, and the system
automatically backs out of the collision.
a. Collimator collision clearing. If you are clearing a collimator collision, then during collimator backout, the
system retraces the motion that preceded the collision.
b. Digital Detector collision clearing. If you are clearing a Digital Detector collision, the system first attempts to
raise the Digital Detector along its motion axis. If this is not successful, the system retraces the motion that
preceded the collision along the remaining axes.
2. If the collision remains after the backward motion, you can resume the positioner movement by releasing and
pressing the Motion Enable Trigger.
Note: In case the table top is under collision and the collision cannot be cleared as described above, such collision
can be cleared by moving manually the table top in longitudinal or lateral direction or by manually rotating it. With
the InnovaIQ Table, the table top can only be moved in lateral direction when the table is tilted (longitudinal direction
is locked).
Note: In case of collision sensor failure, messages such as "...Press Tube front plate to reinit" or "...Press Detector
front plate to reinit" will be permanently displayed on the Innova console at system boot. In such case, press once
on the specified sensor to reinitialize it.
Note: With the InnovaIQ Table, when the level 1 collision sensor is activated, all table axes are available in the
opposite way to better clear the collision.
Auto-Backout (not available with the InnovaIQ Table):
This feature is configured by your Service Engineer to either be ON or OFF. If the Auto-Backout feature is
configured ON, a collision with the Digital Detector while driving the major axes of the gantry (Pivot, C-Arm or
L-Arm) with the Smart Handle or Smart Box will cause the system to raise the Digital Detector along its motion axis,
without stopping the positioner movement. If the Digital Detector is at maximum SID, gantry movement will stop as
a normal Level 1 collision.
6-2 Level 2 (Digital Detector only)
Level 2 detection is a back-up for Level 1 detection. The Level 2 collision sensor on the Digital Detector is activated
if sensor detection continues beyond the normal and expected machine stopping distance.
Activation:
When the Level 2 collision sensor is activated, the Status Display indicator lits and an audible signal is activated. All
motions and drives for the gantry and the table are disabled.
Clearing Procedure:
1. Level 2 prohibits powered motions; therefore, you must clear the sensor manually or press the Digital Detector
emergency backout button (see Innova Positioner / Digital Detector Housing Controls).
Note: With the InnovaIQ Table, when the level 2 collision sensor is activated, all table axes are available in the
opposite way to better clear the collision.
2. Once you remove the obstruction or clear Level 2 using the backout button, all motions and drives are restored.
7 Skin Spacer
Depending on country regulation (e.g: as required in USA and New Zealand), the Skin Spacer must be installed
over the X-ray tube cover.
Innova IGS 520 with Skin Spacer Innova IGS 520 without Skin Spacer
Innova IGS 530 and Innova IGS 540 with Skin Spacer Innova IGS 530 and Innova IGS 540 without Skin
Spacer
When the Skin Spacer is installed over the X-ray tube cover, particular attention must be taken when moving the
table down, especially during patient loading/unloading phase.
With the Skin Spacer installed, do not stop the table too close to the spacer (keep about 5
cm in between) when loading the patient on the table.
Doing so will prevent the table top to collide with the Skin Spacer due to its flexibility.
The purpose of the Innova Central touch screen is to bring an easy and direct access from table side to most
commonly used functionalities related to the acquisition, image review, electrophysiology and Intra Vascular
Ultrasound.
A joystick is used to navigate in some of the functionalities.
A reset button located under the Innova Central touch screen (left side) allows to reinitialize it in case it hangs or
does not respond properly.
8-1 Innova Central touch screen Installation
The primary location of the Innova Central touch screen, connected at table base, is at table side. It can also be
installed on a cart, on an articulated arm or just placed on a table.
To install the Innova Central touch screen at table side, pull the handle located underside, position it over the table
accessories rail and release the handle.
Note: Before using the Innova Central touch screen, make sure it is well secured on the table accessories rail.
8-2 Innova Central touch screen menus
At the bottom of the Innova Central touch screen, permanent buttons are always available whatever menu is
selected.
Depending on configuration, on the top of the Innova Central touch screen, tabs give access to Innova fluoro,
record and image review, AW Vision, electrophysiology and intra vascular ultrasound dedicated screens.
When a small joystick is displayed on a button, it indicates that the Innova Central touch screen joystick needs to
be used to interact in the selected function. For example, when Recall Sequence is selected, move the joystick
right/left to visualize the next/previous sequence.
When a green frame is displayed around a button, it indicates that the given function is enabled.
For example:
When Auto Injection is switched on, it is displayed as follows:
Note: This status feedback is working only for the Innova and AW related functions. Do not expect visual feedback
on the Maclab and IVUS tabs.
• Innova Record screen:
This screen gives access to the most used Record acquisition parameters (frame rates, detail
levels, injection parameters...) as well as protocols and image flip selection..
Label Description
Protocols Allows to select an acquisition protocol within an anatomy.
Image Flip Allows to flip the image horizontally and / or vertically.
X-Ray Lock Interrupts the system from taking exposures when Fluoro or Record is no longer re-
quired.
End Exam Allows to close the current exam.
Record Mode Allows to select one of the available acquisition mode.
Detail Allows to select among Normal or Low Record detail.
3D Spin Speed Depending on system configuration, allows to select among three 3D gantry rotation
speed.
X-ray Delay Allows to set the delay between Injection and X-ray.
Subtraction Activates the Sub or NoSub display of the live image during the acquisition.
Frame Rate/Duration Allows to select a Record acquisition frame rate and duration.
Injection Delay Allows to set the delay between X-ray and injection.
Auto Injection (permanent Activates or deactivates the Auto Injection function.
area)
Stopwatch (permanent Allows to display/hide a stopwatch.
area)
This screen gives access to the most used Fluoro acquisition parameters (frame rates, detail
levels,...) as well as protocols and image flip selection.
Label Description
Protocols Allows to select an acquisition protocol within an anatomy.
Image Flip Allows to flip the image horizontally and / or vertically.
X-Ray Lock Interrupts the system from taking exposures when Fluoro or Record is no longer re-
quired.
End Exam Allows to close the current exam.
Fluoro Mode Allows to select NoSub or Sub Fluoro, Roadmap or Blended Roadmap. When in Sub or
Roadmap Fluoro, allows to activate the Sub and NoSub simultaneous display.
Frame Rate Depending on the selected acquisition mode, allows to select among the available Flu-
oro frame rates.
Simultaneous Display When 2 live displays are configured, allows for the display of a subtracted fluoro image
simultaneously with the unsub fluoro image.
Detail Allows to select among two levels of Fluoro detail.
All basic review and display functions are directly available from this screen. When the In-room
Browser option is activated, all images acquired during the procedure can easily be reviewed using
the Innova Central touch screen joystick. For angio processing, Mask select and Pixel Shift are
directly accessible.
Label Description
Send Angles Activates the 2D send angles
Time Scale Displays/Hide a time line under the image in review on
the live display.
Stenosis Analysis Allows to launch the Stenosis Analysis application.
Stent Viz Activate the Stent visualization application
This Dose screen summarizes the parameters that can be adjusted to directly affect dose.
Label Description
Dose Map Displays or hides the local dose map
Dose reduction Selection of this button automatically reduces the dose
(refer to Innova Positioner / Innova Central touch screen
(option) / Dose Reduction on Innova Central touch
screen)
Protocol dose default Reselect the default dose settings of the currently select-
ed protocol
Fluoro Detail Allows to select between Normal or Low detail
Fluoro Frame Rate Depending on the selected acquisition mode, allows to
select among the available Fluoro frame rates
Record Detail Allows to select between Normal or Low detail
Record Frame Rate/ Duration Allows to select a Record acquisition frame rate and du-
ration
A green/ yellow color is displayed when a dose reduction can be achieved by modifying that
parameter.
A yellow color is displayed when a parameter is not utilizing any dose reduction and any
modification will decrease the dose.
• AW screens: refer to Innova Vision applications User Guide - 5392954-x.
• Mac-Lab/CardiaoLab screens (Options):
Depending on system configuration, when a Mac-Lab and/or CardioLab system is/are connected to the Innova
system, the most commun physio functions such as probe selection, zero setting, scale selection,
measurements, gain selection... are directly accessible.
Refer to the Mac-Lab/CardioLab specific Operator Manual for detailed description of all these functions.
Label Description
Study > Emer- Begins a new study, bypassing the patient demographic input. The patient information is set as
gency follows:
• Last Name: Timestamp
• First Name: No Name
• MRN: Timestamp
Study > Close Closes the current Mac-Lab study in progress.
• Yes/No confirmation on the Innova Central touch screen.
• No confirmation from the Mac-Lab application.
Labels Labels a selected blood pressure channel ART, PA, VEN, PV, LA, RA, AO, RV, VC, SP, LV, or
PCW.
12 Lead Prints a Real-Time 12 lead. Also adds 12-Lead event to the log.
Scale Scales all pressures to 25, 50, 100, or 200 mmHg as selected.
P1 Selects blood pressure channel 1 in the Real-Time window.
P2 Selects blood pressure channel 2 in the Real-Time window.
Measure Measures the current active pressure. The pressure measurement is displayed in the Review and
Log windows.
Measure All Measures all displayed pressures. The pressure measurements are displayed in the Review and
Log windows.
Zero Sets the selected pressure to zero for calibration.
Zero All Sets all displayed pressures to zero for calibration.
Sweep Up Increases the scroll/sweep speed in the active window to the next scroll/sweep speed value up in
the list.
Sweep Down Decreases the scroll/sweep speed in the active window to the next scroll/sweep speed value
down in the list.
Update Saves about 10 seconds of signal data in the Review window (the actual time of data saved is de-
pendant on the number of channels displayed). A corresponding entry appears in the Log (green
time-stamp).
Switch Video Toggles the video signal sent to the monitor in the procedure room.
Stopwatch • Open: Opens a new stopwatch on the Real-Time window.
• Close: Closes the stopwatch opened above.
• Reset: Resets the time on the stopwatch to 0:00.
Start/Stop Stop- Starts or stops the stopwatch.
watch
Emergency Immediately starts saving signals in the Real-Time window to the optical disk. Also saves as
Save much data as possible from memory.
(permanent
area)
Stat Vitals Performs a Stat Vitals measurement. The current HR, SPO2, RR, temperature, Inspired CO2, and
(permanent Expired CO2 are recorded as-is from the Status area into the Log window.
area)
Label Description
Close Study Closes the current Mac-Lab study in progress.
• Yes/No confirmation on the Innova Central touch screen.
• No confirmation from the CardioLab application.
Mark Creates a new note at the current time with the Mark text in the Event Log.
12 Lead Prints a Real-Time 12 lead. Also adds 12-Lead event to the log.
Pace Switches the pacing settings to the highlighted channel poles.
Page Up Changes the display of the Real-Time window up one page.
Page Down Changes the display of the Real-Time window down one page.
Gain Up Increases the gain for the selected ECG, intracardiac, or high-level input channels.
Gain down Decreases the gain for the selected ECG, intracardiac, or high-level input channels.
Auto Save Enables/Disables the automatic saving option.
Manual Save Starts/Stops saving the signals in the Real-Time window.
Sweep Up Increments the scroll/sweep speed in the active window to the next scroll/sweep speed value up in
the list.
Sweep Down Decreases the scroll/sweep speed in the active window to the next scroll/sweep speed value down
in the list.
Update Saves about 10 seconds of signal data in the Review window (the actual time of data saved is de-
pendant on the number of channels displayed). A corresponding entry appears in the Log (green
time-stamp).
Switch Video Toggles the video signal sent to the monitor in the procedure room.
Image Freeze Freezes an image in the CardioImage window.
Emergency Immediately starts saving signals in the Real-Time window to the optical disk. Also saves as much
Save data as possible from memory.
(permanent
area)
Stat Vitals Performs a Stat Vitals measurement. The current HR, SPO2, RR, temperature, Inspired CO2, and
(permanent Expired CO2 are recorded as-is from the Status area into the Log window.
area)
Label Description
Home / Live IVUS: If the system is not on the Home screen, go there.
If the system is on the Home screen and not in Live
mode, initiate Live mode, If the system is already on the
home screen and in Live mode, do nothing.
FFR: If an archived file is not loaded then go to the Home
screen and run.
Permanent Buttons:
Label Description
Auto Injection
In Room-Browser Activates/Desactivates the Collage/In-Room Browser
function:
• Move joystick up/down, right/left to select the se-
quence/photo to review.
• Press on top of joystick to launch the review of the se-
lected sequence/photo.
Recall Sequence Brings focus back on live display:
• Press on top of joystick to Play/Pause the review.
• Move joystick right/left to review next/previous se-
quence when in Play; next/previous frame when in
Pause.
Recall Photo Displays on the reference display the last stored photo.
Store Photo Saves on disk the image displayed on live display and
displays this image on reference display.
Emergency Save Immediately starts saving signals in the Real-Time win-
dow to the optical disk. Also saves as much data as pos-
sible from memory.
Stat Vitals Performs a Stat Vitals measurement. The current HR,
SPO2, RR, temperature, Inspired CO2, and Expired CO2
are recorded as-is from the Status area into the Log win-
dow.
Stopwatch
Innova Central touch screen customization enables to collect the frequently used functionalities to one single
surface based on the user preferences or the type of intervention. It means that you are able to copy the frequently
used Innova Central touch screen buttons to one single location, and use only this surface during the examination
without navigating through the different pages of the Innova Central touch screen.
The following section describes step by step how you can customize Innova Central touch screen. Please be aware
that the customization mode is available only out of exam, for safety reasons. During the examination you can use
the created favorite pages.
8-3-1 Location of the customized pages (Favorites tab)
Innova Central touch screen has a specific tab page called Favorites tab, which is indicated with a star icon. You
can collect your preferred functionalities under different favorite sub-tabs. After the system installation, Favorites
tab has one single sub-tab, which contains a short description about the usage of customization feature.
2. Select the Add new tab item from the displayed list: a virtual keyboard is opened.
3. Enter the name of the new favorites sub-tabs, and press the [OK] button. The new favorites sub-tab with the
given name will be created.
2. Select the name of the Favorite sub-tab where you want to copy the button with the help of the pager arrows,
and press the [OK] button. The selected control functionality will be copied to the selected Favorite sub-tab.
3. Repeat the previous actions until you collect all functionalities you need for your examination.
Please be aware that a single functionality can be placed only once in a favorites sub-tab (functionalities cannot be
duplicated inside a single favorites sub-tab). The functionalities coming from different modalities can be simply
distinguished with the help of the colored stripe and caption of the buttons. Each modality has one specific color
that can also be found on the bottom left corner (colored squares) of the corresponding Tab:
• Purple: Innova / Record
• Green: Innova / Review
• Beige: Innova / Fluoro
• Yellow: AW
• Cyan: MLab
• Red: Volcano s5i
By selecting an allowed location (displayed with green or blue arrows) you can put the button to an empty position
or you can swap two buttons. (Red icon means a position which is not allowed for the given button.)
8-3-4-2 Arrange buttons between tabs
You can arrange buttons between favorites sub-tabs by copying or moving them to another tab. To do this, you
have to select the Copy to or Move to item from the menu.
When you select Copy to or Move to item, a dialog window is displayed to enable selecting the new location of the
button.
Select the name of the sub-tab where you want to copy or move the button with the help of the pager arrows, and
press the [OK] button. The selected button will be copied or moved to the first free position of the selected tab.
8-3-4-3 Delete a button from the current tab
You can remove a button from the favorites sub-tab by selecting the Delete button item from menu. Before the
button is removed a confirmation dialog is displayed. Answer the question of the dialog according to your needs.
The following additional functionalities are available in the extended sub-tab management menu:
• Move tab: change the order of tabs
• Delete all buttons: delete the content of the selected sub-tab
• Delete current tab: delete the selected sub-tab
8-3-5-1 How to change the tab order
Select the Move tab item of the menu: a dialog window will be displayed where you can chose where you want to
insert the selected sub-tab page (before another tab, or at the end of the tab page by selecting at the end).
The approval of content deletion will result in an empty favorites sub-tab, which you can populate again according
to your preferences.
8-3-5-3 How to delete a Favorites sub-tab
Select the Delete current tab item of the menu. A confirmation dialog will be displayed where you can approve the
deletion of the selected sub-tab.
8-4 Dose Reduction on Innova Central touch screen
A {Dose Reduction} button is available on the Dose tab of the Central touch screen located at table side.
This button reduces the dose during the exam by modifying the following parameters (called dose optimization
settings) in one action:
• Fluoroscopy and Dynamic record modes: frame rate, detail, Autoexposure preference
• Angiographic record modes: Autoexposure preference
The {Dose Reduction} button applies the minimum between the current and the configured value of each dose
optimization setting. The current values are defined as the effective ones during the exam, whereas the configured
values are set in the Service User Interface (SUIF) and can be customized by a Service engineer or Application
specialist.
The default configured values are the same as in the Dose Limited protocols (refer to Innova Digital / How to select
an Acquisition Protocol / List of Protocols / Dose limited protocols).
• Not available: means that all current values of dose optimization settings are lower or equal to the
corresponding configured ones.
The {Protocol dose default} button sets the default values of the dose optimization settings from the current
protocol.
8-5 Smart Nav
“Smart Nav” is a graphical interface present on the system if the Innova Central touch screen option is available on
the system. Smart Nav provides an easy access to system functionalities (applications and actions to perform) from
a menu displayed on user request, on the Reference display in 2 monitors configuration or on the Live 2 display in
the 3 monitors configuration.
Smart Nav can be opened In Room by pressing the [Menu] button on either the Innova Central touch screen or the
DL remote control, or in the Control Room by pressing the [Menu] button on the DL keypad. The user can open
Smart Nav with a long press of the Innova Central touch screen joystick button.
If Smart Nav is displayed, the user can hide this menu by pressing the Innova Central touch screen [Menu] button,
using the remote control or DL keypad, or with a long press of the Innova Central touch screen joystick button.
Functions accessible through Smart Nav are organized in a contextual tree. If for example "Smart Nav” is
requested and the system focus is on a review, the menu will be displayed in “the review context” meaning the
review sub-menu will be pre-selected by the system on user display request.
8-5-1 Functions and application accessible through Smart Nav
The Smart Nav tree organization is presented below: sub menus are presented with a blue background and
applications or functions are presented with a white background.
Note that items related to optional applications or functionalities are present in the Smart Nav tree only if the option
is activated on the system:
• StentViz is available with the StentViz option
• InRoomBrowser is available with the In Room Browser option
• Pixel Shift, Mask Select and Substract are available with DSA, Bolus or Sub 3D options.
The context states recognized by the system are:
• Review: state in which the user requests the Menu display when reviewing fluoro or recorded sequences on the
Live 1 (Review has the focus).
• Photo: state in which the user requests the Menu display when reviewing photos on the Reference monitor
(Photo has the focus).
This logo indicates that the selected item triggers an application or functionality.
In order to navigate, select an application or function, and control a functionality in Smart Nav, you can either use
the Innova touch screen joystick or the DL remote control arrows or the DL keypad joystick.
• On the Innova Central touch screen:
– Perform a Left, Up or Down movement of the joystick to navigated in the menu.
– Perform either a right movement of the joystick, or a short press on the joystick button to access a sub menu.
– Press the joystick button to select a function or application (a short press for IRB and Play/Pause).
• On the DL remote control:
– Use the Left, Up or Down arrows to navigated in the menu.
– Either use the right arrow or a press [Enter] to access a sub menu.
– Press [Enter] to select a function or application.
• On the DL keypad:
– Perform a Left, Up or Down movement of the joystick to navigated in the menu.
– Perform either a right movement of the joystick , or press [Enter] to access a sub menu.
– Press [Enter] to select a function or application.
8-5-3 Smart Nav sub menu and functions descriptions
The Review sub menu group functionalities are either useful in a review context or applicable on a fluoro or
recorded sequences.
• Select the Proc. Tools sub menu to reach processing functionalities.
• Select [Send Angles] to perform a smart angle on the last displayed recorded sequence.
• Select [IRB] to display the IRB (Smart Nav will be hidden).
• Select [Store Photo] to store a photo of the displayed frame of the sequence (Smart Nav will be hidden).
• Select [Review Photo] to display the last photo displayed (Smart Nav will be hidden).
• Select [Review Seq.] to play the last recorded sequence (Smart Nav will be hidden).
In Proc. Tools sub menu:
• Select the Display Tools sub menu to reach display functionalities.
• Select [Filter] in order to choose a filter to apply. Use the touch screen joystick, the remote control arrows or the
Dl keypad joystick to select the filter to apply on the sequence. Press on the touch screen joystick button or on
enter on either the remote control or the Dl keypad to exit the function.
The state of the Smart Nav will change. Select [Validate] to keep the performed modification.
Select [Reset] to reset the sequence to its initial state.
Use the touch screen joystick, the remote control arrows or the Dl keypad joystick to adjust the
brightness and contrast. Press on the touch screen joystick button or on enter on either the remote
control or the Dl keypad to exit the function. The state of Smart Nav will change.
Select [Validate] to keep the performed modification. Select [Reset] to reset the sequence to its
initial state.
• Select [Zoom] to apply a x2 zoom on the sequence. Use the touch screen joystick, the remote control arrows or
the Dl keypad joystick to adjust the roam the sequence. Press on the touch screen joystick button or on enter on
either the remote control or the Dl keypad to exit the function. The state of Smart Nav will change. Select
[Validate] to keep the performed modification. Select [Reset] to reset the sequence to its initial state.
• Select [Pixel Shift] to apply the pixels shift of a subtracted acquisition. Use the touch screen joystick, the
remote control arrows or the Dl keypad joystick to adjust the shift of the sequence. Press on the touch screen
joystick button or on enter on either the remote control or the Dl keypad to exit the function. The state of Smart
Nav will change. Select [Validate] to keep the performed modification. Select [Reset] to reset the sequence to
its initial state.
• Select [Mask Select] to select the mask of a DSA sequence or to switch between mask and sequence review
for bolus or subtracted 3D acquisitions. Use the touch screen joystick, the remote control arrows or the DL
keypad joystick. Press on the touch screen joystick button or on enter on either the remote control or the DL
keypad to exit the function. The state of Smart Nav will change. Select [Validate] to keep the performed
modification. Select [Reset] to reset the sequence to its initial state.
• Select [Subtract] to apply or remove subtraction on the sequence. If subtraction is applied, a check is displayed
on the subtraction icon.
9 Table accessories
9-1 Shoulder Rests
A set of shoulder rests is delivered with the InnovaIQ Table.
This set is made of three parts: one central module and two identical shoulder rests.
1 Shoulder Rests
2 Central Module
In the case of Trendelenburg tilt during a procedure, it is recommended to install the shoulder rests prior to starting
the procedure to increase the patient confidence and prevent movement while the table is tilting.
The shoulder rests must be installed as shown in the picture.
THE DISTANCE BETWEEN THE TWO SUPPORTING ARMS OF THE SHOULDER REST MAY
NOT SUPPORT SMALL CHILDREN AND ADDITIONAL SUPPORT WILL BE REQUIRED.
One shoulder rest pole can hold a 125 kg max patient weight. Do not use the shoulder rest
assembly with patients heavier than 125 kg in case one of the shoulder rest needs to be
removed during the procedure while the table is tilted or needs to be tilted afterward.
1- Clean and dry carefully the table top before installing the central module.
2- Install the central module from table top head end as shown on the picture and secure it to the table top by
rotating the knob clockwise (3) until the torq limitor is activated.
3- Insert the shoulder rest perpendicular to the central module, push UP (4) and rotate it (5) toward foot side until it
is locked.
One shoulder rest pole can hold a 125 kg max patient weight. Do not use the shoulder rest
assembly with patients heavier than 125 kg in case the weight will be applied on one point
only on the footrest when the table will be tilted.
1- Install the central module at the head end as shown in the picture and secure it to the table top by rotating the
knob clockwise (3) until the torq limitor is activated.
2- Insert the shoulder rest perpendicular to the central module, push UP (4) and rotate it (5) toward foot side until it
is locked.
4- Slide the footrest onto the shoulder rest's poles and tighten each nut of the footrest onto the poles.
Table head extender is not compatible with digital headholder, head widener, shoulder rest
and foot rest.
9-4 IV pole
The IV pole provides support for any IV bag when intravenous therapy needs to be administered to the patient.
The IV pole clamps with through hole are also available for IV pole mounting. These clamps are mounted on the
removable rails (sleeve) or the rail extender.
The IV pole clamps are clamped either on the sleeve or the rail extender.
The alarm index displayed at the end of the error message refers to the alarm description in this section of the
Operator Manual and in the Alarm Messages window available from the Utilities key located on top of the Innova
Digital screen.
X-ray / Auto-injection 1s 3s
Alarm system not functional 3s 5s
2.2 !! Collision risk. Detector lift Reset system to recover. Detector lift brake failure de-
brake failure. Reset system tected at system boot or re-
set. Risk of uncontrolled de-
tector lift down motion. De-
tector lift commands are dis-
abled after five consecutive
detections.
2.3 !! Collision risk. Table position Continue on going procedure with care. Unknown table position.
failure. Move with care Risk of uncontrolled colli-
sion. All gantry axes move
at half speed.
2.4 !! Extra dose risk. Acquisition Release exposure switch. The frame rate used during
frame rate error. Reselect proto- Reselect the protocol and restart the acqui- acquisition does not match
col sition. the requested frame rate.
Patient Air Kerma may be
higher than expected (up to
four time to nominal).
2.5 !! Extra dose risk. Spectral filtra- Stop acquisition and resume. Error detected in spectral fil-
tion failure. Restart acquisition If problem persists, terminate promptly on tration selection.
going procedure. Image quality may not be
optimal and patient skin
dose may be higher (up to
four times nominal).
2.7 !! Patient fall risk. Gantry/table Verify that Gantry and Table motion are dis- One of the Gantry or Table
lock failed. Handle patient care- abled. brake has not been activat-
fully Handle patient with care. ed after a Gantry/Table dis-
able command. One of the
gantry axis is not locked.
Table longitudinal or lateral
axis is not locked. Risk of
patient fall in case of acci-
dental table or gantry com-
mand during patient load /
unload.
2.8 !! Patient fall risk. Gantry/Table Stop Gantry/Table motion then resume. Gantry/Table disable com-
lock failed. Repeat lock com- mand not available during
mand an ongoing motion, a 3D, a
Spin Cardiac or a Bolus ac-
quisition. Risk of patient fall
in case of accidental table or
gantry command during pa-
tient load / unload.
2.9 !! Patient fall risk. Tabletop Handle patient with care. Longitudinal/lateral brakes
brakes failure. Handle patient cannot be activated any-
carefully more. Risk of patient fall
during patient load / unload.
2.10 !! System lock up risk. Long time Reset system before the end of the day. System was not reset for 30
without reset. Reset system consecutive days. A reset of
the system is needed to
eliminate a growing risk of
system lock up.
2.11 !! Extra dose risk. Detector fail- Terminate promptly ongoing procedure to Error detected in detector
ure. Secure patient secure patient. X-rays will be inhibited after power supply. Image quality
the end of the current procedure. may not be optimal and pa-
tient Air Kerma may be high-
er (up to four times nomi-
nal).
2.12 !! Patient fall risk. Immediately Move back immedialety the table in horizon- One of the table brakes has
move table back to horizontal tal position. Handle the patient with care. been detected as defective.
Risk of patient fall due to a
table brake failure and un-
controlled motion of the ta-
ble top.
2.13 ! Hidden anatomy risk. Deselect Stop on going Fluoro or Record. Deselect The electronic shutter posi-
auto-shutter and resume electronic shutter using the Shutter button tion is not accurate and clini-
on the Innova Digital Screen and resume cal data may be hidden.
procedure.
2.14 !! Collision risk. Collision predic- Continue on going procedure with care. Unknown gantry or table po-
tion failure. Move with care sition. Risk of uncontrolled
collision. All gantry and table
axis move at reduced
speed. Auto-Positioning and
automatic CPR are dis-
abled.
2.15 !! Exam interruption risk if power Contact immediately GE service for repair of Undercharged UPS batter-
is lost. UPS failure. Call Service the UPS. ies or UPS failure. Risk that
the exam is interrupted and
the Innova system suddenly
unavailable in case of main
power failure.
1.1 !! Collision risk. Collision detec- Continue on going procedure with care. Gantry emergency stop fail-
tion failure. Move with care ure or bumper sensor fail-
ure. Risk of patient collision.
All gantry axes move at half
speed.
1.2 !! Collision risk. L-Arm clutch Continue on going procedure with care. Gantry clutch failure detect-
failure. Move with care ed. Risk of staff injury in
case of collision with L-arm
movement. All gantry axes
move at half speed.
1.3 ! Extra dose risk. Cumulated Cumulated dose displayed is not accurate; The cumulated dose dis-
Dose not accurate. Reset system do not use this information in patient played might be less than
records. the cumulated dose deliv-
Reset System. ered to the patient.
1.4 ! Extra dose risk. Spectral filtra- Stop acquisition and resume. Error in spectral filtration se-
tion failure. Restart acquisition If problem persists, terminate on going pro- lection. Image quality may
cedure. not be optimal and patient
Air Kerma may be higher
(up to twice nominal).
1.5 ! Over injection risk. Injection Reset Innova system and reinitialize injector. Automatic injection failure
stop failure. Reset system If problem persists, use manual injection. detected. Injector did not
stop the injection as expect-
ed.
1.6 !! Unintended motion risk. Mo- Continue with caution. Smart Handle / Smart Box /
tion command failure. Move Secure the Load/Unload patient procedure. TSSC not fully operational.
carefully Risk of unintended motion
due to loss of command pro-
tection.
1.7 ! Unintended motion risk. Check Verify on footswitch/console hand switch are Record foot / hand switch
exposure switches are released released. permanent activation is de-
Correct and reset system to recover. tected by the system con-
trol. Risk of unintended mo-
tion during 3D, Spin Cardiac
or Bolus acquisition.
1.8 ! Unused X-ray risk. Injector un- Stop acquisition if started. Automatic Injection failure
available. Use manual injection Reinitialize injector and resume or select detected. Injector got dis-
manual injection. armed before automatic in-
jection start.
1.9 !!! 3 min before X-Ray inhibit. Terminate promptly current operation to se- System will be shut downed
UPS low. Secure patient cure patient. 3 min after the first message
is displayed. Less than 20 s
!!! 2 min before X-Ray inhibit.
prior system shutdown after
UPS low. Secure patient
last message is displayed.
!!! 1 min before X-Ray inhibit.
UPS low. Secure patient
!!! System near to be shut down.
UPS low. Exit exam
1.10 !!! 3 min before X-Ray inhibit. Terminate promptly current operation to se- The X-ray tube is not cooled
Tube cooling failure. Secure pa- cure patient. anymore. X-rays are
tient stopped 3 min after the first
message is displayed.
!!! 2 min before X-Ray inhibit.
Tube cooling failure. Secure pa-
tient
!!! 1 min before X-Ray inhibit.
Tube cooling failure. Secure pa-
tient
2 System Components
2-1 Basic Components
• A large display cabinet in the technical room.
• A large display monitor in the exam room.
• A suspension in the exam room.
• A set of video source devices either in the exam or control room.
• A table side interface for controlling the display (Innova Central touch screen).
2-2 Option
• A large display monitor to replicate images from the exam room.
• A set of parts to have HD 2MP and HD 8MP output.
• A set of video source devices either in the exam or control room.
3 Safety recommendations
THE WORKING POSITION WITH THE LARGE DISPLAY MONITOR OPTION IS TWO METERS
MAXIMUM FROM THE MONITOR.
For Innova signals, the images can be reduced to 75% (960 x 768 pixels) of the native resolution (1280 x 1024
pixels). It is estimated that the image is still readable at 75% for a working distance of two meters or less.
IMAGE RESIZING: THE LARGE DISPLAY MONITOR ALLOWS RESIZING OF THE IMAGES
(ZOOM, SHRINK). THE USER SHOULD PAY ATTENTION WHEN DOING A DIAGNOSTIC OR A
TREATMENT ON OVER ESTIMATED VESSELS/STENOSIS.
MAKE SURE THAT ALL THE IMAGES ON THE LARGE DISPLAY MONITOR OPTION DISPLAY
THE SAME PATIENT INFORMATION TO AVOID DATA MISMATCH.
• Vivid E9
FOR THE OTHER MODALITIES, THE LARGE DISPLAY MONITOR ONLY PROVIDES A
CONNECTION BUT DOES NOT ENSURE AN EQUIVALENT IMAGE QUALITY AND A
FREQUENCY MATCH.
If there is a degradation of the signal on the Large Display Monitor that prevents the use of the modality image, use
the original monitor of the modality.
If the frequency of the signal does not match the frequency of the allocated input port of the Large Display Monitor,
use the original monitor of the modality.
Note: Five groups can be displayed at the same time in the group list. The user may navigate using up and down
arrow in order to access to the whole group list.
5-1-1 How to Create a New Group
Press ‘Manage Groups’, then ‘Create new group’.
After tapping on the 'Assign new layout' box, a new page will appear with all the video inputs that are available.
Tap each video input to be displayed in the layout and the corresponding check mark will be displayed on the
layout button.
Selecting too many inputs at a time could cause the layout filter to return 0 matches. Reduce the number of
video inputs to widen the search.
Once the selection is done, press on "Apply Filter".
All layouts that respect the filtering criteria will be displayed.
Each window on the Innova Central touch screen shows 4 layouts..
Use the right and left arrow in order to navigate among pages of proposed layouts. An indicator shows the number
of pages and which one is currently displayed.
After creating a group, tap on the “Save Group” button and return to the main screen of the available groups.
5-1-2 How to Modify a selected Group
Press the ‘Groups’ button on the bottom of the Innova Central touch screen.
Press ‘Manage Groups’.
Select the group to modify.
Select ‘Edit selected group’ option.
The selected group's layouts will appear allowing the user to add, move or delete any layouts.
Type in the new name for that group with the soft keyboard.
Tap 'OK’.
5-1-4 How to Delete a selected Group
Press the ‘Groups’ button on the bottom of the Innova Central touch screen.
Press ‘Manage Groups’.
Select the group to delete.
Select ‘Delete selected group’ option.
• The Frontal Live 2 image, and its possible duplicate, is defined by a medium blue horizontal line at the top of the
image.
• The Frontal Reference image is defined by a light blue horizontal line on the top.
• The Frontal Reference duplicate image is defined by a light blue full border.
• The Lateral Live 1 image, and its possible duplicate, is defined by a medium orange horizontal line at the top of
the image.
• The Lateral Live 2 image, and its possible duplicate, is defined by a medium orange horizontal line at the top of
the image.
• The Lateral Reference image is defined by a light blue horizontal line on the top.
• The Lateral Reference duplicate image is defined by a light blue full border.
On the Innova Central touch screen, the rectangles representing the Innova inputs could be identified by the
specific labeling:
• The Frontal Live 1 image, and its possible duplicate, is represented by a medium blue rectangle.
• The Frontal Live 2 image, and its possible duplicate, is represented by a medium blue rectangle.
• The Frontal Reference image, and its possible duplicate, is represented by a light blue rectangle.
• The Lateral Live 1 image, and its possible duplicate, is represented by a medium orange rectangle.
• The Lateral Live 2 image, and its possible duplicate, is represented by a medium orange rectangle.
• The Lateral Reference image, and its possible duplicate, is represented by a light orange rectangle.
The Innova IGS 520, Innova IGS 530, Innova IGS 540 can be delivered with 3, 4 or 6 LCD monitors suspension.
The LCD Monitor Suspension comes fully installed and pre-cabled, suspended by a Suspension Arm. For your
operational needs, no additional features need to be considered.
1-2 Large Display Monitor Suspension
1-2-1 Standard use
The Innova IGS 520, Innova IGS 530, Innova IGS 540 can be delivered with a Large Display Monitor suspension.
The Large Display Monitor suspension is delivered with one Large Display Monitor and two 19" monitors mounted
to the back of the suspension to use in case of a degraded mode of the Large Display Monitor.
The Large Display Monitor suspension comes fully installed and pre-cabled, suspended by a suspension Arm. For
your operational needs, no additional features need to be considered.
The rotation of the suspension in case of LCD failure can be done in one direction only depending on the room
configuration. The rotation angle is 180°, if it cannot be reached, rotate the suspension in the opposite direction.
The label fixed on the handles indicates to the user the presence of backup monitors located at the back of the
suspension.
IN A CASE WHERE THE LARGE DISPLAY MONITOR IS ACCIDENTALLY TURNED OFF, THE
BACK-UP MONITORS WILL STILL BE ON AND SHOULD BE USED.
1-3 Precautions
The installation, connection, removal and repair of all components must be performed by qualified, licensed
technicians only.
Ceiling mounted systems must be installed properly. Failure to follow the instructions
provided may lead to a potentially hazardous and unstable condition of the system.
These products must be serviced with manufacturer-approved, original replacement parts only. Failure to follow this
recommendation will void the manufacturer's warranty and will release the manufacturer from liability and warranty
claims!
When installing and adjusting the LCD Monitor Suspension System and/or associated components, or performing
any work on the system, be sure to remove all electrical power to the entire system. This is a hazardous
condition!
The suspension arm is under constant spring tension. When installing or removing components to or from the
system, always move the suspension arm slowly to its upper stop (upper maximum hub)! Otherwise, when
disconnecting components from the system, the suspension arm may jump upwards, causing damage or injury.
The Manufacturer can not be held liable for unauthorized modifications made to the system, or use of the system
for unintended purposes. Since improper modifications may impair proper operation, safety, or reliability of the
system, all modifications need to be authorized by the Manufacturer in writing.
Avoid all Collisions with other Equipment or Walls, Ceiling, etc. Severe Damage may occur!
In case of Collision or Damage, immediately stop Operation of System.
Disconnect the Power Supply immediately!!
Do not position anyone or walk underneath the system!
Inform Manufacturer immediately!
Do not use System in humid surroundings.
Do not open Covers, Cable Channels, etc.! Hazardous!
1-4 Operation
Note: The suspension arm contains GAS SPRINGS. The gas inside of the springs reacts to temperature changes
by expanding or contracting, thus slightly changing their spring range.
Therefore, extensive temperature differences may affect the Suspension Arm performance. Please let the gas
springs adjust to your room climate, before you have the balance of your Suspension Arm adjusted. Even 10°
Celsius will have a (small) effect. Allow the suspension arm to acclimate to the room temperature for
approximately 12 hours.
Swing System in all desired directions. Avoid harsh movements. Do not turn more than 350° - System is blocked at
360° to avoid Cable torque.
Upward movement is easier if you swing System outwards (away from point of Suspension Arm fixation) before
you lift the System.
Downward movement is easier if you swing System inwards (towards point of Suspension Arm fixation) before you
lower the System.
1-5 Periodic Maintenance
On a regular basis (min. annually), have a technician check screws and cables of System for security.
Please refer to Service Manual for LCD Monitor Suspension for all necessary Measurements and Specification.
Note: System is designed for a life time of 10 (ten) years. After 10 years, exchange of the System is strongly
recommended.
INTERFACES (LIKE DICOM), ALL ASSOCIATIONS OF THE GE ANGIOGRAPHIC X-RAY SYSTEM WITH A 3RD
PARTY IMAGING/DATA MEDICAL DEVICE MUST BE APPROVED FOR USE BY THE MANUFACTURER
GEHC.
Cathlab Frontiers Solutions is the common name for the following system options:
• The Integrated Innova - s5i system option (Volcano s5iTM Imaging System: PN 400-0100.02, Software version
3.2.2)
• The Integrated Vivid E9 solution (Vivid E9 System: BT' 12 version).
The s5i screen can be displayed on GE angiographic system’s exam room display solution: either on a color
monitor from the LCD monitor suspension or on the Large Display Monitor.
In the LCD monitor suspension case a video switch may be used in the control room to share the s5i display and
the AW display on the same exam room color monitor.
In the Large Display Monitor case a dedicated "IVUS" linkset is used to route the s5i video signal to the Large
Display Monitor. Using the Innova Central touch screen, you can select from the predefined "IVUS" layouts.
However, it is possible to further optimize the layouts to fulfill all the needs of the clinical procedure (fully
customized input types, sizes, and positions). Contact the GE Service Office for your country to ask support for
further layout customization.
2-6 Use of the s5i system in OR configuration
To maintain the compliance with the operating table safety standard IEC 60601-2-46, not all s5i controls and
Patient Interface Modules (PIMs) are allowed to be mounted on the table rails or the table itself. Only the following
s5i controls and PIMs are allowed:
• s5i Trackball Controller (Control Console II
• Joystick
• s5i IVUS PIM
• s5i Rotational PIM (only in a sterile bag)
• FFR PIM
Note: The traditional s5i Control Console (Control Console I - IPX0 rating only) needs to be placed in the tableside
cart.
Note: The Volcano Touch Pad is applicable only to s5i GE Rev1, Rev2 and Rev2+ systems. It is not used any
more with the s5iTM Imaging System PN: 400-0100.02, Software Version 3.2.2.
To maintain the integrity of the sterile field the bedside mounted s5i system equipment should be located under a
sterile drape when configured for normal use. Refer to the Volcano operator manuals delivered with your system for
the use of sterile drapes/covers on the s5i system components.
For a complete description of the Vivid E9 systems and the 6VT-D (4D TEE) probe refer to the following User
Manuals:
• Vivid E9 User Manual - GA092907.
• TEE probes User Manual - KZ192871.
3-1 System Setup
During the clinical procedures the Vivid E9 system with the 6VT-D (4D TEE) probe will be positioned either on the
left or on the right side of the patient head area, at a distance which permits comfortable access of the TEE probe
towards the patient (transesophageal access). The Vivid E9 operator is standing between the equipment and the
patient head. The patient head area positions are shared with the anesthesiology equipment and operator, and the
interventional cardiologist and/or surgeons.
During the X-Ray procedure, the Innova IGS gantry movements are also limiting the Vivid E9 positions around the
head area, and a dedicated back out path has to be cleared for Vivid E9 park position.
The Vivid E9 has 3 types of connection cables:
• Power
• Network
• Large Display Monitor image link called Vivid E9 linkset cable
The power and the network connection cables belong to the Vivid E9. They shall be disconnected from the wall
access sockets and roll up to Vivid E9 cable hooks before moving the equipment out of the exam room.
The LDM image link cable should be disconnected from the back of the Vivid E9 and needs to be protected when
not in use.
The Vivid E9 linkset cable has the following end (including an EMC foam that needs to be used all the time):
ACCIDENTAL WITHDRAWAL OF A TEE PROBE MAY CAUSE DEEP SOFT TISSUE INJURY.
THE USER SHOULD MINIMIZE THE RISK OF THE INJURY BY: - LOCKING THE WHEELS OF
THE VIVID E9 SYSTEM WHILE IT IS IN USE - LOCKING THE INNOVA GANTRY AND TABLE
MOTIONS WHEN NO MOVEMENT IS NEEDED - TAKING SPECIFIC CARE WHEN MOVING
THE GANTRY AND/OR TABLE TO PREVENT THE ACCIDENTAL WITHDRAWAL OF THE
6VT-D (4D TEE) PROBE.
Since the Vivid E9 system is used in the vicinity of the patient, it may accidentally collide with Innova IGS Gantry
and/or Table.
1 General Description
The OR configuration of the product is designed in accordance with the operating table standard (IEC 60601-2-46),
it corresponds to the Innova IGS 520, Innova IGS 530 or Innova IGS 540 product equipped with:
• OR table (InnovaIQ table identified with an "OR" labeling on the Covers, refer to the picture below).
• IPX4 Table side user interface.
• Mandatory UPS.
• Removable table rails - Available in US (9.52 x 28.5 mm) and European (10 x 25 mm) standard sizes.
• Dedicated surgery protocols.
To maintain the compliance with the operating table safety standard, not all controls are allowed to be mounted on
the table rails. The following controls only are allowed:
• IPX4 Table Side Status Control (TSSC).
• IPX4 Smart Box (Smart Box).
• IPX4 Table Panning Device (TPD).
• IPX4 Innova Central touch screen.
• Innova IVUS PIM and TouchPad.
A tableside cart is provided to place any other control (3D mouse, Innova IVUS console and joystick).
Table mounted injectors are not allowed with the OR configuration.
PRODUCT DIFFERENTIATION:
The OR table corresponds to the OR configuration of the Innova x-ray angiographic system.The symbol on the
table with "OR" corresponds to OR Table and OR configuration of system.
OR Table
THE SYSTEM CAN BE USED FOR OPEN SURGERY OR IMAGE GUIDED SURGERY ONLY
WITH THE OR CONFIGURATION OPTION AND IF INSTALLED IN A DEDICATED OPERATING
ROOM.
THE TABLE DOES NOT HAVE BREAKABLE TABLE TOP (ARTICULATED JOINTS). IT IS NOT
RECOMMENDED TO PERFORM PROCEDURES REQUIRING INTRAOPERATIVE
ARTICULATIONS WITH THIS DEVICE.
THE TABLE DOES NOT HAVE LATERAL TILT (CRADLING). IT IS NOT RECOMMENDED TO
PERFORM PROCEDURES REQUIRING INTRAOPERATIVE LATERAL TILT (CRADLING) WITH
THIS DEVICE.
TABLE MOUNTED INJECTORS ARE NOT ALLOWED TO MOUNT ON THE TABLE AS THEY
ARE NOT COMPATIBLE IN THE OR CONFIGURATION OF THE INNOVA SYSTEM.
3D MOUSE AND INNOVA IVUS (EXCEPT PIM & TOUCHPAD) SHALL BE MOUNTED ON THE
WHEELED STAND IN THE OR CONFIGURATION OF THE INNOVA SYSTEM AS THEY ARE
NOT COMPATIBLE WITH THE OPERATING TABLE.
Bodily fluids may damage internal components if they are allowed inside the equipment.
Use drapes, if necessary, to protect equipment when performing procedures.
You can choose among the following solutions, that must be ISO 10993 certified. You can contact your local GE
representative for more information on how to order these solutions.
Full drape kit:
For a safe and easy mounting of the drapes, 2 people are necessary.
• Break the “Tear Here” stickers (4 pieces) at the perforation lines and slide the drape to expand over the top and
bottom sections of the C-arc.
Note: Particular attention should be paid to avoid any contact between the sterile operator and the detector.
For a safe and easy mounting of the drapes, 2 people are necessary.
In order to get proper clearance and access, you can position the gantry and the tabletop away from one another
by rotating the tabletop or by moving the gantry in backout or parking.
• Remove the image detector drape from the kit.
• Lower the detector to the lowest position. Lay the drape on a table, unfold the left and right and fan folds
keeping the label “this side up” exposed.
• Insert hands (left and right) at the bottom of the drape in the layer where hand labels are indicated.
• Lift drape up with hands inside (the inside of the drape will be exposed).
• Start positioning the drape over the detector.
Note: Particular attention should be paid to avoid any contact between non sterile body parts of the operator
(like head) and the drape.
At this stage of the draping procedure 2 people are necessary.
• Uncurl the drape completely.
• Match the drape’s trim sealed blue corners to the detector’s front corners.
• Raise the detector to its maximum height and fasten the first strap at its landing point area (if this strap is not
placed in the right spot, the drape will sag and not fit tightly to the detector.
• Lower the detector to the lowest position and fasten the 2nd strap. Fasten the strap such that no C-arc clip is
entrapped.
• Apply the drape. The drape should overlay the C-arc drape, but should not entrap any C-arc drape clip.
• Gather the excess material at the top of the C then place the 2nd rubber band around the drape.
• Using the adhesive bands at the drape’s corner, it is possible to improve the fitting of the drape.
If yes, re-clip and try to be smoother when sliding the drape over the C-arc the next time. Be sure to break the
"Tear Here" stickers before sliding the drape.
• No C-arc clip have been entrapped in the Image detector drape or tube drape, as shown on the following
pictures. If so, there is a risk for the clips to be ripped out during extreme angulations of the C.
1 - Clip 1 - Clip
2 - Poly tie badly positioned, below the last clip 2 - Poly tie well positioned, above the last clip
Image detector drape is not correctly positioned, the Image detector drape is correctly positioned, the poly
poly tie is attached too low on the C-arc tie is attached above the last clip.
Tube drape not correctly positioned, last clip is inside Tube drape is correctly positioned, last clip is outside
the bag the bag
Clip should be outside the tube drape. Clip is outside the tube drape.
Half drape:
This drape is covering the entire Image receptor assembly, and the upper half of the C-arc. The mounting
instructions are identical to the Image detector drape above 2. There is an additional poly tie to attach around the
C-arc, at the end of the mounting.
This drape is to be used for procedures not requiring any C-arc movement. Pivot, swivel
and detector movements are allowed.
L Arm Access Plate & Access Holes:
This portion of L-arm need to be draped.
1 Access plate
2 Access holes
Below is the weight and load capacity of accessories that get mounted on removable table rails. This table
to be considered for loading of accessories to ensure not to exceed maximum permissible load limits:
SOME PARTS OF THESE ACCESSORIES ARE NOT RADIOLUCENT AND MAY CAUSE
IMAGE QUALITY ARTIFACTS, THAT MAY BE DIFFICULT TO DISTINGUISH FROM
ANATOMICAL STRUCTURES.
Check that each accessory is properly mounted and clamped if provided with clamps before
any use. Failure to this recommendation may lead to patient injury and equipment damage
These accessories are delivered with their own instructions for use. Refer to them before any use.
3-1 Removable Table Rails
It is recommended that the table be rotated away from gantry during removable table rails insertion on/removal
from the table.
3-1-1 Installation
Before mounting the removable table rails, locate and secure all tubing, wires or catheters to avoid pinching, pulling
or otherwise interfering with safe and proper patient care.
1. Slide the removable table rails on to the tabletop until it hits the table Top.
2. Rotate the knob upward into the locked position to secure it.
3. A potential equalization terminal is available on the rails to bring the various parts of equipment or systems to
the same potential when connected together.
3-1-2 Removal
1. Before removing the removable table rails, check that tubing, wires or catheters are not entangled.
2. Remove all accessories from it including any potential equalization plugs.
3. Unlock the removable table rails by rotating the lock knob downward.
4. Grip the removable table rails by the rails and pull it towards you.
3-1-3 Equipment care
Inspect the removable table rails and rails for damage, dents, or cracks before each use. If they seem damaged or
broken, contact your GE service representative for repair or replacement.
3-2 Other accessories
Once the removable table rails is mounted on the tabletop, it is used to mount other accessories like armboard, rail
extender, adaptor rail, universal clamp, round post clamp, patient restraint strap, anesthesia screen holder and
table width extender.
3-2-1 Armboard with thick pad (armrest)
The armboard with thick pad is a non-invasive medical device that can be attached to the removable table rails to
support patient’s arm(s) during imaging procedures or for administering anesthesia with improved patient comfort
(lateral, prone or supine).
3-2-2 Rail Extender
The rail extender is a non-invasive medical device designed to clip on the front end of the removable table rails
mouted on the table to extend the rails. Mounting of anesthesia screen holder and the addition of an IV Pole are
common applications.
3-2-3 Adaptor rail
The adaptor rail is a non-invasive medical device intended to mount table side control units on the trunk section of
the table top. This feature improves operator comfort for procedures requiring close work to the head of the table,
such as pediatric procedures.
3-2-4 Head Widener
The head widener is a non-invasive medical device that can be mounted at the head of the table in order to provide
additional support for the patient’s head.
3-2-5 Table Width Extender
The table width extender is a non-invasive medical device that can be mounted on the removable table rails to
increase the support surface for large or obese patients or to improve peri-operative comfort by supporting the
arms of any patient.
3-2-6 Round Post Clamp
The round post clamp is a non-invasive medical device designed to be clipped on the removable table rails in order
to hold other accessories with a round post connection. The clamp can hold the accessory, such as an anesthesia
screen holder, and rotate to offer a variety of alignment options.
3-2-7 Universal Clamp
The universal clamp is a non-invasive medical device designed to be clipped or slided on the removable table rails
in order to hold other accessories with a round post or blade connection. The clamp can hold the accessory, such
as an anesthesia screen holder, perpendicular to the rail.
3-2-8 Anesthesia Screen Holder
The flexible anesthesia screen holder is a non-invasive medical device designed to be mounted into two surgical
rail clamps (using a 5/8" OD round post) and support a typical screen between the sterile field and the patient’s
head. It can be flexed into most practical, clinical positions and securely hold the weight of the screen.
3-2-9 Patient Restraint Strap
The patient restraint strap is a non-invasive medical device designed to be attached to the removable table rails in
order to immobilize the patient.
3-2-10 Accessory Cart
The accessory cart is a non-medical device intended to store various medical accessories. It can be transported
between locations utilizing rolling casters and be secured in place.
Start Page
Acquisition Page
1-4 Status
When all images have been acquired and processing is finished, the application displays a status:
Some of the standard physics measurements are performed using metallic (copper and/or aluminum) phantoms for
simulating patient attenuation.
These are selected in such a way that the total X-ray absorption of the phantom is similar to an average patient.
However, since the atomic numbers of the metal phantoms are much higher than the ones of the elements in a real
patient, the response they give to tube voltage variations is very different. In practice it means that metallic
phantoms are more sensitive to voltage adjustments than the real anatomies. As a consequence, when the
exposure control loop tries to regulate for a metal phantom, it may overreact and produce oscillating techniques
(the displayed kV and mA values may periodically increase and decrease). Such oscillations are normal in the
presence of metal phantoms, and they do not occur when real anatomy is in the field. In order to avoid the
oscillations we suggest the usage of water, plexiglas or anatomical phantoms, which are much closer to real
tissues in X-ray absorption properties.
2 Parameter monitoring
2-1 The parameters that are monitored are
• Electronic Noise
• Row Correlated Noise
• Brightness Uniformity
• Signal to Noise Ratio uniformity
• Bad Pixels
• Noise Power Spectrum
• Conversion Factor
• Modulation Transfer Function
• Resolution uniformity
• Dynamic Range Linearity
• Dynamic Range Accuracy
• Contrast to noise ratio
• Contrast
2-2 Frequency
QAP shall be run on a periodic timeline going from every week to at least every month.
Logs will be automatically retrieved by OnLine-Center.
3 Quality criteria
Comparing the current values to typical values monitored with optimal image quality makes the Pass/Fail status.
This is made using the 6-sigma methodology.
4 Material
In order to perform QAP, you need the QAP Suitcase, which contains:
• The flat-field phantom (20 mm Al):
2 Starting ILinq
To access the iLinq features, click on the iLinq button on the DL Browser screen of your GE equipment. This is
located under the "Utilities" icon.
The iLinq main screen will appear.
If your GE equipment does not have the iLinq licenses installed, contact your local GE Service representative for
details on how to get it. iLinq is activated remotely, via the InSite connection, and does not require any manual step
or access to the equipment.
It is from this main screen that you select which iLinq feature to use.
1 Contact GE
2 Settings
3 Messages
4 TiP Virtual Assist
5 iLinq Help
6 About iLinq
7 Close
8 GE Healthcare= 'GE Santé' in Province of Quebec - Canada
3 ILinq Features
3-1 Contact GE
Contact GE page is displayed as the homepage of the iLinq application.
Use this to ask questions to applications specialists or to send a problem report directly to the Online Center.
You can send requests for the same system or for some other system in the same facility using their System ID.
3-2 Settings
Clicking on the Settings link (2) opens a new page.
You can use this page to change the language settings for the iLinq application and to add the list of user names,
phone number and other System IDs which can be used later while submitting the Contact GE request.
3-3 Messages
Clicking on the Messages link (3) opens a new page.
A list of your sent and received messages appears, the most recent messages listed first. You may need to scroll
down to see all the messages.
Once connected, the remote User will not be dependent on the Customer for any activity.
This feature does not impact regular clinical applications, and the Customer can disconnect the remote User at any
time from the console.
Notice: Depending on local regulation, patient consent might be required to autorize any GE Healthcare
support to access his/her demographic information prior the launch of the TVA program.
Note: The performances of the system may be degraded during a TVA session. The announced performances do
not apply in this mode.
3-5 iLinq Help
Clicking on iLinq Help link (5) opens a new page.
This page provides the user with a detailed tutorial on how to use the iLinq application and all its functionality.
3-6 About iLinq
Clicking on About iLinq link (6) opens a new page.
This page provides the version details of the iLinq application.
3-7 Close
Clicking on Close link (7) closes the iLinq application and returns you to the clinical browser.
4 iLinq Notifications
This feature allows the user to be notified when there is an incoming message or Chat request from the GE Online
Center.
4-1 Message Notification
User shall be notified with the below image (toggling between them) on the clinical application User Interface (DL
Browser) if there is an incoming iLinq message from the GE Online Center.
This notification shall be stopped once the user opens the iLinq application and access the Messages page, or if
the system is reset.
4-2 Chat Notification
User shall be notified with the below image (toggling between them) on the clinical application User Interface (DL
Browser) if there is Chat initiated from the GE Online Center.
Once you open the iLinq application, you will see a small window (as shown below) on the right lower corner of the
iLinq page. You can accept to continue to chat with a GE Representative, or reject if you do not wish at that
moment.
This notification shall be stopped once the user opens the iLinq application and accept or decline the Chat
invitation, or if the system is reset.
- The Advantage Workstation (AW) delivered for the Innova system is loaded with a specific
software to allow Innova image review and processing. - Subtracted images acquired on the
Innova system can be visualized in Sub mode or reprocessed only on the AW delivered for
the Innova system. - Innova images networked to any other workstation/PACS... can be
reviewed only in No Sub mode.
1-1-1 Helpful Information
- To start AW:
Default login: sdc
Default Password: adw4.x, where 4.x is the AW core version number, as shown in table below:
Note: When a new AW is installed, it is loaded with the "sdc" default login and "adw4.x" default password. It is then
the responsability of users to ask their GE Service Representative for any login and password change. In case of
AW software upgrade, previous login and passwords are preserved. Also ask your GE Service Representative for
any change if needed.
- In AW you will be able to know the names of icons locating the mouse on the icon and waiting a green help
appears.
- The Browser of AW displays the patient structure like DICOM: One STUDY contains SERIES, one SERIES
contains images. The DL sequences are DICOM images, thus displayed in the IMAGE window. All the
sequences coming from Innova system are contained in one SERIES. All the photos coming from Innova
system are contained in a second SERIES.
1-1-2 Compliance to Standards
DICOM V3.0 implementation in AW is described in its Conformance Statement (2340652-100). It can be used to
verify compatibility with other DICOM devices.
1-2 Patient List Browser
Go into patient list browser using "Patient list" tab at left of the screen.
1-2-1 Tools
Click on "Tools" key.
1-2-2 Patient List preference
Select "Patient list Preference"
Set preference for:
• Layout
In "Select Profile":
• Set "General Purpose" to burn CDs non compressed.
• Set "Compress XA 512 cardiac" to burn CDs with 512 cardiac images compressed using the jpeg 14 lossless
compression ratio (1024 cardiac images must be downscanned first to 512 before being compressed).
• Do not set "Portable Disk" to keep the multi CD capability.
In "Embedded applications":
• Set "Add Centricity DICOM Viewer" to burn the DICOM Viewer with the images.
In "Other Options":
• Do not set "Downscan XA cardiac images to 512" to burn cardiac images in their original matrix size.
The purpose of downscan option is to convert a monochrome 1280 x 1024 @ 72 Hz video signal into a Low
Line Rate 625l @50Hz or 525l @60Hz video signal, to be connected to standard peripheral devices such as a
printer, a recorder or a monitor.
IMAGES CONVERTED THROUGH DOWNSCAN FEATURE ARE NOT VALIDATED FOR
DIAGNOSIS AND/OR INTERVENTIONNAL PURPOSES.
• Set "Check full media" to check all burnt data after CD is written.
Click on "Ok" before to exit.
1-3 Viewer
Go into patient list browser using "Viewer" tab at left of the screen.
1-3-1 Display Tools
1-3-2 Display
1-3-5 Edit
• Screen
– Screen annotation -> full
– Mag glass size -> normal
– Scale -> OFF
– Hospital, patient_name, patient_id, esi, system, date, geometry, acquisition_params, ras_coordinates,
frame_info, Windowing, graphic results -> all ON
• Film
– -> full
– Hospital, patient_name, patient_id, esi, system, date, geometry, acquisition_params, ras_coordinates,
frame_info, Windowing, graphic results -> all ON
• Autocine
– Select Autocine "On" to automatically launch the cine mode when opening the Viewer on a sequence. If "Off"
is selected, the cine mode will need to be manually started.
1-3-9 Layout Manager
Click on the Layout Manager key on top left of the Viewer to open the Layout Manager window.
At the bottom right of this window, click on the Protocol selection arrow to open the preloaded display protocol list.
In case the optional Adv Paste 2.0 application is loaded on the AW station, a set of 5 dedicated display protocols is
available (protocol names start with "Past..."). Use these display protocols on any "Pasted Series" for specific
display of a pasted image.
A specific protocol so called "Divided Screen" can also be used to divide the viewer/image in two parts and so allow
1 Video ON/OFF
Note: With the In-room 3D application, it is recommended to deactivate the screen saver time out on the AW
workstation.
1-4-2 Main "In-room 3D" Control Panel
Multimodality In-room 3D offers the capability to manipulate 3D images and to activate gantry synchronization
functions from the exam room.
Features are accessible through a dedicated software interface or from buttons located on the In-room 3D Mouse.
In-room 3D functions are applicable to CT, MR and XA images.
1-4-3 Operation
The user can visualize and control 3D models and their related reformatted views (Axial, Coronal, Sagittal, Oblique,
Curved, cross-section, and Lumen views) from the exam room. In-room 3D functions are applicable to CT, MR and
X-ray-Angiography 3D models. Typical controls are rotation of the 3D image rotation, adjustment of the volume
rendering, selection and rotation around a vessel of interest. The Gantry synchronization features, Send Angulation
and Follow the Gantry, can be controlled from the table-side. Features are accessible through a dedicated software
interface or from buttons located on the In-room 3D Mouse.
Exam room can take user interface control just by pressing any button on the In-room 3D Mouse (except for the
reset button).
Control room can take user interface control just by pressing any key on the AW control room keyboard.
1-4-4 User Interface
This section describes the In-room 3D features.
Refer to AW Operator Manuals for detailed explanations of "standard" 3D application and basic use of the AW.
Main path
1 Start Exam Room User interface from the Control Room
using the left control panel.
Open the In-room 3D tool located on the Display tools of
Volume Viewer and select In Room Control.
2 Start Exam Room User interface from the view contextual
menu.
In room interface can only be activated from views in left
screen monitor.
3 Select a 3D View and Activate the ‘Follow the Gantry’
mode
In-room 3D interface provides 2 ways of activating the
‘Follow the Gantry’ mode:
1 Click on the Follow the Gantry button on the interface
. by using the In-room 3D mouse cursor mode.
2 Push the toggle button "G/M" in the In-room 3D
. Mouse to toggle between Follow the Gantry and Man-
ual modes (cf. Workstation Configuration / Advantage
Workstation (AW) Configuration (Option) / In-room 3D
(Option) / In-room 3D Interface Launch at Table-Side /
In-room 3D Mouse Interface Description).
9 Set 3D image rotation to default orientation 21 Adjust window width and window level adjustment for
(Anterior-Posterior-Left-Right-Superior-Inferior) reformatted view and 3D MIP view
Adjust Volume Rendering threshold for 3D VR view
10 Set mode "Rotation/Paging" on the 3D joystick: 22 Save secondary capture of the selected view
• 3D Rotation on 3D view
• Image paging on Axial, Coronal, Sagittal,
Curved and Lumen views.
11 Set mode "Move Cursor position" on the 3D joy- 23 Select VR preset for 3D VR view
stick
12 Set mode "Adjust shutter diameter" on the 3D
joystick
The 3D joystick controls are dependent on user interface mode and are also contextual to the current view type
(3D, Curved, Lumen….). See the table below for joystick control description.
Position your hand on the In-room 3D Mouse and move gently to position the cursor on the desired function.
1-4-5-4 In-room 3D Mouse Interface Description
Note: The 3D Mouse box is active only when the In-room 3D application is launched and the "Exam Room" is
displayed on top left of the main menu.
Pressing this button allows to reset the In-room 3D Mouse device in case of abnormal motion of the image
without touching the mouse.
Note: Do not touch the In-room 3D Mouse during the reset (about 10 s). Wait until the initialization mes-
sage disappears. A total of 5 beeps can be heard during the In-room 3D Mouse reinitialization.
Pressing this button allows to toggle between the multiple views and the full screen view image display.m,
...).
Pressing this button allows to send the current 3D image angulations to the gantry.
Note:Make sure to select a 3D view on the Volume Viewer views layout.
Note: Make sure the proper patient position was selected on the Innova Digital Screen (prone, supine,
head to toes reversed…) prior to sending the current 3D image angulations to the gantry.
To reach the desired angulations, move the Auto Positioner joystick UP to move the frontal gantry to the
selected angles.
Pressing this button allows to move to the next view in the AW left screen view layout
Note: The function may be applied when the view is enlarged.
Press one of the associated buttons to store the new orientation. The previous orientation already
mapped to this button will be replaced.
Note: If the "Save Store Angle as Secondary Capture" is selected in the In-room 3D Mouse Preferences
Orientation screen, each stored 3D image is automatically archived in the AW database and can be re-
viewed as Secondary Capture from the Image Browser.
Note: All stored 3D images are saved until Volume Viewer is exited from the current patient.
Pressing this button allows to recall previously stored 3D image angulations (if any).
In case of more than one angulations have been stored, the angulations button map picture is displayed
on screen to recall the selected orientation.
Pressing this button allows to switch between the "Follow the Gantry" and "Manual" modes.
Note: Make sure to select a 3D view on the Volume Viewer views layout.
When in "Manual" mode, 3D image can be manipulated at any time independently from the gantry posi-
tion.
When in "Follow the gantry" mode, 3D image orientation is linked to the frontal gantry position.
Moving the frontal gantry will make 3D image(s) rotate(s) in synchrony with the gantry.
Note: The Image rotation/flip and the table tilt/rotation are not taken into account in "Follow the Gantry"
mode.
Note: Make sure the proper patient position was selected on the Innova Digital Screen (prone, supine,
head to toes reversed…) prior to performing "Follow the Gantry".
The function of this button depends on the type of active view (the view contoured in red):
1 The active view is a 3D view. Pressing this button allows to successively display the 3D image in Su-
. perior, Inferior, Anterior, Posterior, Left or Right orientation.
Note: The sequence of orientations can be configured in the Preferences panel from the In-room 3D
assistant screen. (i.e. A-R-L-P can be preset).
2 The active view is any of the other views (Axial, Coronal, Curved, Lumen….). Pressing this button
. allows to page forward through a list of vessels of interest outlined in the QuickAVA or Advanced Ves-
sel analysis applications. If no vessels have been defined, this button is idle.
Caution: Make sure when paging through vessels of interest to visualize the correct vessel. This function
is not allowed in Full screen and Oblique view layouts.
Pressing this button allows to quickly switch between "Rotation" and "Roam and Zoom" modes.
The sensitivity of the mouse when used in Rotation or Translation can be customized using the Prefer-
ences panel of the In-room 3D assistant screen.
Note: When the ownership is in the Control room, pressing on any button (except Reset) allows to get
control to the Examination Room.
Note: Preset n°1 is always used as default VR preset when VR rendering mode is selected on the In-room 3D
Main Panel.
• Click on key to select up to 6 different VR display from the VR Presets library.
– Select "Save Store Angle as Secondary Capture" to automatically archive in the AW database each image
stored using the Store Angle button either on the In-Room 3 panel or In-room 3D Mouse box.
These images can be reviewed as any other Secondary Capture from the image browser.
To record the Dose, DAP, Dose Rate, and DAP Rate, a double ionization chamber is fixed at the output of the
X-ray tube collimator. There is a cable connecting the chamber to the main unit.
One or two secondary displays connected to the main control unit and fixed on the monitor boom display the dose
information in the X-ray Room.
One ionization chamber can be connected to the main unit to record the Dose, DAP, Dose Rate, and DAP Rate of
one X-ray tubes.
An optional printer can be connected to the main unit to enable dose report printing on labels.
When the CAK reaches 10 Gy, the Diamentor M4 KDK displays OL (Overload). To avoid
information loss, before the CAK reaches 10 Gy (e.g. 5 Gy), write down (or print) the CAK
value and reset the Diamentor M4 KDK. Repeat this operation as many times as needed to
avoid the 10 Gy limit. At the end of the patient exam, add all written (or printed) values to
estimate the exam CAK value.
When an optional Diamentor M4 KDK is installed on the system, the values displayed on the DL console, live or
reference monitor may be different from the Diamentor M4 KDK values since they are based on the internal
dosimetric indications of the system.
1-2 Dose Measurement at Interventional Reference Point
A Diamentor is a device that displays dose values measured at a fixed distance of the focal point.
The interventional reference point is defined as the point on the reference axis, 15 cm from the isocenter, in the
direction of the focal point.
The following table presents the different possible Interventional Reference Point distances:
Chamber/Patient Distance (CPD) and Source/Chamber Distance (SCD) must be set at the following values
depending on the detector size :
System Flat Panel SOD (cm) Reference point dis- CPD SCD
(Source Object Dis- tance from source (in (in cm) (in mm)
tance) cm)
1. Digital display.
2. <Reset> key to reset to zero the dose area product and entrance dose readings as well as the time and to quit
the setup menu without saving the modified settings. Secondary displays connected are also reset.
Depending on the setup menu configuration, <Reset> will also initiate a printout; refer to Diamentor M4 KDK
(Option) / System Configuration (Setup Menu). In addition, the key is used to terminate entry of the
chamber-patient distance without saving the new value.
3. key: cursor control key used in the setup menu (up) and to enter the Chamber-Patient Distance.
4. <Enter> key to access submenus of the setup menu and to enable the Chamber-Patient Distance entry mode.
5. key: cursor control key used in the setup menu (down), to enter the Chamber-Patient Distance and to display
the calibration menu.
6. Contrast adjustment .
7. <Print> key to initiate a printout and to quit the setup menu, saving the modified settings. In addition, the key
is used to terminate entry of the Chamber-Patient Distance, saving the new value.
2-2 Diamentor M4 KDK, top view
1. Chamber connection B.
2. High-voltage fuse for ionization chamber (20 mA, slow-blow).
3. Chamber connection A.
4. Power switch.
5. Power input fuse.
6. Power input.
7. Nameplate (on rear panel).
8. <Test> button to perform a Diamentor M4 KDK (Option) / Performing a Measurement / Constancy Check of the
device.
9. Serial interface for connection of a printer or PC.
10.2 connectors for secondary displays.
Symbols as marked on the equipment
| power on
power off
contrast adjustment
brightness adjustment (secondary display)
up
down
AC voltage
3 Performing a Measurement
3-1 Constancy Check
Check the device performance every day, before putting the Diamentor M4 KDK into operation.
1. Turn on the device and wait for the display to illuminate.
2. Press the <Test> button 2.
The test screen will be displayed (Illustration Test screen). When no problem is found, the message "Test passed"
appears briefly afterwards. The test screen clears automatically after approx. 3 seconds.
Test screen
Note: This test is meant to check the stability of the electrical calibration. For this purpose a test charge is injected
to the amplifier input (deviation +/-1 %) and the resulting gain is compared with the factory calibration.
The percentage deviation is indicated for each channel. Deviations of +/-5 % are admissible (message: Test
passed). When the deviation exceeds +/ 5 %, the message "Test failed" appears (confirm message with <Print>).
In the second case, repeat the test. Should the "Test failed" message be displayed again, the Diamentor M4 KDK
must be repaired before it can be used again.
3-2 Measurement
When the device has passed the test, the measured values will be shown in the selected format as soon as
radiation is detected.
A filter or tubus placed between chamber and patient will reduce the dose area product and
entrance dose relevant for the patient. In this situation the Diamentor M4 KDK will display
dose area product and entrance dose values which are too high. According to national
regulations in Germany (1), however, it is sufficient to record the measuring value and the
filter or tubus in use for reconstruction of the radiation exposure level. Should the system
be permanently used with the filter or tubus, it is recommended to recalibrate the Diamentor
M4 KDK.
(1) Ask your competent authority about the regulatory situation.
3-2-1 Note on warm-up time
The device requires a minimum warm-up period of 15 minutes before a measurement can be performed, because
the leakage is very high during this time. As the leakage changes in the beginning, it would not make sense to
perform the zero adjustment within the first 15 minutes. The leakage has a strong effect only on measurements
with small measuring values and long measuring times. Therefore, measurements with high values and short times
Display (Submenu):
• Selection of the display format (independently for the Diamentor M4 KDK and the secondary display).
• Adjustment of the bar graph limits.
• Selection of the filter depth.
• Activation of the display test.
• Enabling/disabling the time display.
• Selection of the display resolution.
Date & Time (Submenu):
• Entry of date and time of day. Correction (Submenu):
• Initiating the zero adjustment.
• Entry of the chamber temperature.
• Entry of the chamber ambient pressure.
• Entry of the correction factor.
• Entry of the focus chamber distance.
Printer (Submenu):
• Entry of the printer control commands.
Language:
• English, German, Spanish, Italian, French.
Dose area product Unit:
• Selecting the unit for dose area product measurement: cGy.cm², µGy.m², Gy.cm².
RS232:
• Printer, PC or unassigned.
Print Mode:
• AUTO ON/OFF (initiate an automatic printout with <Reset>: yes/no).
4-2 Factory Settings
The device is delivered with the following factory settings:
Display format (1-4-5):
[D-d] [E-e] [D-d] [E-e] [D-d] [E-e]
Bar graph limits (1-4-6):
low = 0 and high = 3600 cGy.cm²/s
Filter depth (1-4-7):
3
Time (1-4-8):
ON
Display test (1-4-10):
N/A
Date & Time (1-4-11):
Is set to the date & time in Germany
Zero adjustment (1-4-12):
N/A
Chamber temperature (1-4-13):
20 °C
Air pressure (1-4-13):
1013 hPa
Correction factor (1-4-13):
1
Init code (1-4-15):
<000> <000> <000> <000>
Paper feed (1-4-15):
<009> <010> <000> <000>
Note: This is the printer setup for the STAR printer DP8340S (PTW ordering number L991096). This is also the
default setting even if the Diamentor M4 KDK is delivered without a printer.
Language (1-4-4):
English
Dose area product unit (1-4-17):
µGy.m²
RS232 (1-4-18):
Computer
Note: If the Diamentor M4 KDK is delivered with a printer, this option is set to "Printer".
Print mode (1-4-19):
AUTO OFF
4-3 Displaying the Setup Menu
• Press or and, keeping either key depressed, push <Enter> to display the setup menu. (If you press the
<Enter> key first, the Diamentor M4 KDK changes to the mode for entry of the Chamber-Patient Distance).
Setup menu
For option [EFD], [efd] and [E-e], the measuring value for the entrance dose rate, respectively, is displayed in the
display area to the left. In the display area to the right you will see the current chamber-patient distance. The line is
labeled with a "E:" in the left margin.
Using the cursor key , select "Display" and press <Enter>.
The Display submenu will be shown (Illustration Display menu).
Display menu
Example
1 - Measuring unit for the first line (display area to the left)
2 - Measuring unit for the second line (display area to the left)
3 - Measuring unit for the first line (display area to the right)
4 - Measuring unit for the second line (display area to the right)
5 - Fluoroscopy time
6 - Chamber-Patient Distance
7 - Bar graph for the dose area product rate
8 - Entrance dose
9 - Dose area product
4-6 Bar Graph Limits (Display Submenu)
The length of the bar graph corresponds to the dose area product rate. If the measured dose area product rate is
below the low limit, the bar graph will not deflect. If the measured value reaches the high limit, the bar graph
deflects to full scale. The bar graph starts flashing when the dose area product rate exceeds the high limit.
Note: The fact that the reading exceeds the high limit is indicated only by the bar graph flashing.
The bar graph scale is linear. Limits can be adjusted in the range between 0.000 cGycm²/s and 30,000 cGycm²/s.
The following limitations must be observed: (max. - min.) > 1 cGycm²/s; min. > 0.8 max.
• Call up the Display menu.
• Using the cursor keys, select "Bargraph limits" and press <Enter>.
• Using the cursor keys, adjust the first digit of the low limit, or press <Enter> to access the second digit.
4-7 Filter Depth (Display Submenu)
The Diamentor M4 KDK calculates the dose area product rate on the basis of the change of the dose area product
over a period of time T between approx. 0.6 and 6 seconds. The time T automatically adapts to the radiation pulse
frequency of the X-ray device, allowing the dose area product rate to be indicated even in the presence of pulsed
fluoroscopy.
The chain of dose area product rate values determined in this manner are sent through a digital filter with
adjustable filter depth for further smoothing before being displayed.
Major signal changes exceeding 12.5 % pass this filter undelayed. Minor changes build up in dependance of the
selected filter depth:
5 = filter off residual noise for pulsed fluoroscopy approx. 9 % / 5 setting time to 90 %
approx. 5 x 2.3 x T
9 = filter off residual noise for pulsed fluoroscopy approx. 9 % / 9 setting time to 90 %
approx. 9 x 2.3 x T
• Adjust the other digits in the same manner. The internal clock does not stop while the menu is being displayed.
The Diamentor adopts the new time as soon as you quit the menu with <Print>.
If the system displays the message "Off-set current out of range. Confirm with PRINT" at the
end of the procedure, the system could not be zeroed.
Corrections menu
4-13 Chamber Temperature, Air Pressure and Correction Factors (Corrections Menu)
Before putting the Diamentor into operation, the chamber temperature and air pressure for air density correction
must be specified. If there are radiation absorbing parts between the patient and the chamber (such as the patient
couch), the corresponding correction factor must be entered as well.
• Display the Corrections menu.
• Using the cursor keys, select one of the menu items and press <Enter>.
• Using the cursor keys again, adjust the first digit and confirm it with <Enter>.
• Proceed in the same manner to adjust all other digits.
Adjustment ranges:
temperature (0 ... 70) °C
air pressure (500 ... 1500) hPa
correction factor 0.000 ... 9.999
The air density correction factor is calculated by the following formula:
PO = 1013 hPa
tO = 20 °C
t = actual temperature
P = actual air pressure
You can set only one correction factor for the Diamentor M4 KDK. As both measuring chambers form a unit, the
installation is the same for both chambers (under-table tube or over-table tube). Therefore, it would not make sense
to set different correction factors for the two chambers.
4-14 Source-Chamber Distance
The Diamentor M4 KDK calculates the value of the entrance dose in the patient plane from the value measured in
the chamber plane.
The values are calculated according to the following formula:
Printer menu
STAR DP8340S <02> <000> <000> <000> <000> <009> <010> <000> <000>
COSYS LP2042 <06> <027> <081> <063> <000> <001> <012> <010> <000>
ELTRON COMPANION PLUS <06> <094> <080> <094> <080> <001> <094> <080> <000>
5 Maintenance
The measures provided by the M4 KDK are determined by Calibration factors.
A maintenance of the M4 KDK is necessary every two years to ensure calibration factors are valid (i.e. measures
are accurate).
If your equipment is maintained by GE Healthcare, this maintenance will be performed during a periodic
maintenance.
If your equipment is maintained by a service provider other than GE Healthcare, a request should be made to your
service provider to perform this maintenance or, alternatively, you may request a preventive maintenance from your
local GE Healthcare service provider.
6 Restrictions
The values recorded by the Diamentor M4 KDK are accurate as required by the International Standards in the
following working conditions:
• Temperature within 10 to 40 °C.
• Field of views not smaller than the smallest Field of view available with collimator blades wide opened.
the Gantry
and table for
integrity, and
rework/redo
as necessary
Table Check the ta- CHK0196 x
ble moving
protective
earth cables
C1 Cabinet DL/RTAC fan PM0060 x
cleaning
DL battery re- PM0058 x
placement
(every 2
years)
RTAC battery PM0059 x
replacement
(every 2
years)
System back- SW0250, see x
up sections 3.1 &
3.2
PDU Battery PM0032 x
replacement
(for system
without Fluoro
UPS: every
year)
PDU Battery PM0032 x
Replacement
(for system
with Fluoro
UPS: every 2
years)
Jedi fan diag- PM0037 x
nostic
Jedi RAM KV PM0029 x
board re-
placement
(every 5
years)
KV accuracy PM0038 x
Filters Clean- PM0033 x
ing and Fan
Check
C2 Cabinet Filters Clean- PM0033 x
ing and Fan
Check
Technique factors that constitute the maximum line current condition is 100 kV, 1000 mA.
Hospital feeder line protection:
• 480V 3~ or 3N 60Hz PDB configuration: circuit breaker, rating 150 A.
• 380/400/415V 3~ or 3N 50/60Hz PDB configuration: circuit breaker, rating 80 A.
Note: 0.1s exposure time is in the range of DSA record exposure time.
kV mA max
Focal Spot 1 50 290
70 630
125 640
Focal Spot 0.6 50 169
70 313
125 267
2 Learning objective
Increase your knowledge and help you develop safe work habits. Quick overview of X-ray radiation and dose
through a simple explanation of the equipment, technology and to discuss the different techniques available to
reduce dose.
3 Teaching points
• Radiation dose is a natural consequence of x-ray imaging.
• Displayed Dose Parameters.
• Effects of ionizing radiation on humans.
• Dose reduction measures for the patient.
• Dose reduction measures for the staff.
• Pediatrics Case.
• Legal Provisions.
1: Focal Spot
2: Reference Point
3: Isocenter
A: 57 cm on Innova monoplane units.
B: 15 cm
1: Focal Spot
S: Surface of Area
AK: Air Kerma
DAP: Dose (Air Kerma) Area Product
0 to 2 • None Expected
2 to 5 • Transient erythema/epilation
For full table, refer to S. Balter et al. Fluoroscopically Guided Interventional Procedures: A
Review of Radiation Effects on Patients' Skin and Hair Radiology 254(2) 2010.
• Sequence duration:
– Minimize Radiation time, use the X-ray timer indications.
– The X-ray time flashes and emits an audible signal after 5 minutes of X-ray time.
• Use of LIH (Last Image Hold):
– LIH provides an image in between the live fluoro images to use as a reference image and eliminates the
need for extended fluoro exposures.
– Further reducing the need for additional fluoro exposure can be facilitated by storing the last frames (up to
450) of the fluoro sequence with the fluoro store option .
Example from the "Dose to patient" section, for Innova IGS 520, RDL+, Balanced Dose/IQ, 7.5 fps, Normal, at 20
cm patient size (PMMA cm). Refer to this Operator Manual section to find the data corresponding to your system
configuration.
: Low
: Normal
Note: Depending on country regulations, some of the choice may not be available (e.g. only RDLx choices in
Germany)
Lower dose rate selectable within a large range.
Choice of AutoEx preference to be preset in protocol by Service or Application specialist.
7-8 Frame Rate
Lower dose rate:
• Use as low Frame rate as possible (at tableside or console) compatible with medical requirements for the image
refresh rate (versus anatomy and/or tool(s) movements or changes).
• Have protocols preselected with lower dose setup (fluoro dose strategy).
Pulsed fluoroscopy with a choice of frame rate: 30 fps, 15 fps, 7.5 fps, 3.75 fps**.
** Option
Two Dose Strategies (Max Dose Reduction and Balanced IQ/Dose) are provided on systems (to be preset within
protocols by Service or Application Specialist).
Balanced IQ/Dose: Maintains static object detectability, noise appearance. Based on perception experi-
ment (Case Western University, OH, from the D.L. Wilson group).
Highest reference dose (mGy/min) for For Autoexposure preference: IQ For Autoexposure preference: IQ
Fluoro settings Plus, RDL Plus Standard, RDL Standard
Balanced IQ/Dose Max Dose Reduc- Balanced IQ/Dose Max Dose Reduc-
tion tion
30 fps Normal < 87.6 < 87.6 < 87.6 < 87.6
15 fps Normal < 87.6 < 43.8 < 87.6 < 43.8
7.5 fps Normal < 43.8 < 21.9 < 43.8 < 21.9
3.75 fps Normal < 21.9 < 21.9 < 21.9 < 21.9
30 fps Low < 87.6 < 87.6 < 43.8 < 43.8
15 fps Low < 87.6 < 43.8 < 43.8 < 21.9
7.5 fps Low < 43.8 < 21.9 < 21.9 < 21.9
3.75 fps Low < 21.9 < 21.9 < 21.9 < 21.9
• Dose Limit in Fluoro can be lowered when reducing frame rate (up to 4x) and using a different fluoro frame rate
strategy (up to 2x).
• Two main different ways – main difference is on the Low detail management for the standard AutoEx
preferences which are set lower than the regular ones (up to 2x).
7-10 Threshold
Display prior to the predefined Dose Threshold value being reached: • Watch the dose display and apply
dose reduction means as appropri-
Live display Reference display
ate.
– The display changes above the
predefined threshold.
• Other recommendations exist for
procedural dose warning.
– eg Values above 5 Gy indicates
that the patient should have
clinical follow-up for determinis-
tic radiation-induced injury
(SIR-CIRSE recommenda-
tions).
Display after the predefined Dose Threshold value has been exceeded – eg. Use 500 Gy.cm2 as signifi-
(above 100%): cant radiation threshold
Live display Reference display (SIR-CIRSE).
– eg. Use 3000 mGy as first noti-
fication, then each further addi-
tional 1000 mGy (SIR-CIRSE).
– eg. Use first notification 3000
mGy, then second one at 6000
mGy (ACR).
SIR-CIRSE recommendation: Stecker
MS, Balter S, Towbin RB, et al.
Guidelines for patient radiation dose
management. J Vasc Interv Radiol
2009 20(7 supp) S263-S273.
Display on live display after double of the predefined Dose Threshold ACR recommendation: Mahdevappa,
value has been reached: M, Detorie, N. How to prepare for the
joint commission's sentinel event poli-
cy pertaining to prolonged fluo-
roscopy. American College of Radiol-
ogy. The Medical Physics Consultant
2008 601-603.
9 Pediatrics Case
The Innova IGS system is specifically designed to minimize radiation exposure while providing good image quality
for the intended pediatric and adult patients. Exposure to ionizing radiation is of particular concern in pediatric
patients as younger patients are more radiosensitive than adults and have a longer potential lifetime for the effects
of radiation exposure to manifest as cancer. Thus, the use of equipment and exposure settings designed for
pediatric patients are different from the ones used for adult patients.
• Children have a greater sensitivity to radiation than adults. When possible, avoid X-ray examinations and use
alternative procedures.
• Further to the presented means to minimize dose, also consider the application of:
– For infants up to 1 year old, remove the anti scatter grid from acquisitions performed at Field Of View smaller
than 20 cm.
– Use collimation to the specific area of interest.
– When possible, use a gonad shield.
• Use pediatrics protocols. Pediatrics protocols are set with lower dose setup than for adults.
– Select a protocol on the system console where the name contains "peds" or "pediatrics".
Note: Dose rate for pediatrics are automatically reduced as the system senses a small patient thickness from the
dose signal to the detector. See section Appendix - Radiation Safety Tutorial / Dose settings for pediatrics in
Innova systems for more details.
9-1 Dose-related Parameters Default Values
Dose-related parameters default values of adult vs. pediatric protocols
Dose Threshold 2 or 3 Gy 1 Gy 1 Gy 1 Gy
* Few adult protocols have Fluoro or Dynamic detail default value at Low (a non-exhaustive example is Dose
Limited protocols).
** Few adult protocols have Fluoro or Dynamic FPS default value different from 15.
***Most adult protocols take the AutoExposure set in the Service User Interface (SUIF), but few adult protocols
have a defined AutoExposure (a non-exhaustive example is Dose Limited protocols defined at RDL+).
In addition to the dose-related default values, the system is able to optimize temporal resolution for pediatric
patients. The system also adapts a smaller ROI in dynamic fluoroscopy and in dynamic record, based on the exam
type (Angio or Cardiac) and the patient type (Adult or Pediatric). These correspond to attributes of a patient
protocol, and are set according to its usage. Further, there is a minimal spectral filtration of 0.1 mm for pediatric
protocols due to a higher radiosensitivity.
9-2 List of Pediatric Protocols
For pediatrics imaging, select a protocol on the system console under the “Pediatrics” category, or where the name
contains "Pedia".
Pedia Cerebral
Pedia Chest
In case of adult protocol selected while the patient’s age is below the “Pediatrics limit age threshold”, the following
Normal
Air Kerma rates at the interventional reference point (IRP) in case of fluoroscopy with 30 pulses per second, normal
and low detail Receptor Dose Limited (RDL) settings in FOV20, measured using phantoms with different thickness
on Innova IGS 530.
12 Legal Provisions
• In many countries or states, by means of rules, guidelines or regulations, lawmakers have contributed to
improving radiation protection for patients and medical personnel.
• On an international level, guidelines are laid down by the International Commission on Radiological Protection
(ICRP). Many of the rules, guidelines or regulations are governed by the ALARA concept (As Low As
Reasonably Achievable), meaning the production of a diagnostically relevant image at minimum possible dose.
• The material presented in this tutorial contains ways to improve radiation safety on Innova’s but is not deemed
to be exhaustive, read the Radiation Safety section for more details. Local legal provisions also prevails over the
ways presented in this document when applicable.
5. Set the PMMA elevator kit layers on the mattress on each side of the dosimeter ( or equivalent if not available).
6. Set the 15 cm of PMMA phantom above the dosimeter.
7. Set the table height as the function of the mattress:
Mattress 1'' Mattress 2''
Table height (as displayed 11 cm 13 cm
on Live display)
8. Use Fluoro to center the ionization chamber in the X-ray beam. Center the ionization chamber in the image as
shown in the Note. During this step do not move the table height.
Note: The ionization chamber position is very critical for the accuracy measurements. To
determine the center of the display area, plot on the monitor its two diagonals as shown on the
illustration below or use a ruler. The center of the display area and the center of the ionization
chamber must merge.
10.End exam.
2-1-4 Analyse the results
1. Compute "Dose3" in mGy by applying the following formula:
Dose3 (mGy) = Dose2 (%) x Adult local dose threshold (Gy) x 1000
Note: The Adult local dose threshold is set in the Preferences tab on the DL browser (refer to
Innova Digital / Dose Map / How to set Dose Map? / Dose Map preferences).
For example with Adult local dose threshold at 2 Gy and Dose2 at 37.5%:
Dose3 (mGy) = 37.5/100 x 2 x 1000 = 750 mGy
2. Compute the accuracy of Dose Map by applying the following formula:
Inaccuracy (%) = abs (Dose1 - Dose3) / Dose1 x 100
For example with Dose1 at 940 mGy and Dose3 at 750 mGy:
Inaccuracy (%) = abs (940 - 750) / 940 x 100 = 20%
2-2 Lateral Measurement
2-2-1 Set up geometrical measurements
Set the geometrical parameters as follow:
1. Center the table laterally:
1. Set an object on the mattress, centered on the lateral axis of the table.
2. Use Fluoro to center the object in the X-ray beam by moving the table laterally.
2. Move the Gantry into Lateral position.
3. Set the PMMA elevator kit layers on the mattress (or equivalent if not available):
4. Put the 15 cm PMMA phantom on the top of the elevator kit centered on the table.
5. Set the table height as the function of the mattress (do not move the table until the end of the measurement).
Mattress 1'' Mattress 2''
Table height (as displayed on Live 11 cm 13 cm
display)
Note: The table height is required in Lateral configuration in order to measure the dose at the right point
compared to the theoretical patient model computed by Dose Map. If the table height is disregarded, the dose
map accuracy is not guarantee.
6. Connect the ionization chamber to dosimeter controller.
7. Set the dosimeter controller in dose mode (cumulated dose).
8. Stick the dosimeter probe to the PMMA by using scotch tape (or any equivalent mean).
9. Use Fluoro to center the ionization chamber in the X-ray beam. Center the ionization chamber in the image as
shown in the Note. During this step do not move the table height and the table lateral axis.
Note: The ionization chamber position is very critical for the accuracy measurements. To
determine the center of the display area, plot on the monitor its two diagonals as shown on the
illustration below or use a ruler. The center of the display area and the center of the ionization
chamber must merge.
10.End exam.
8. Use Fluoro to center the ionization chamber in the X-ray beam. Center the ionization chamber in the image as
shown in the Note. During this step do not move the table height.
Note: The ionization chamber position is very critical for the accuracy measurements. To
determine the center of the display area, plot on the monitor its two diagonals as shown on the
illustration below or use a ruler. The center of the display area and the center of the ionization
chamber must merge.
Note: The table height is required in Lateral configuration in order to measure the dose at the right point
compared to the theoretical patient model computed by Dose Map. If the table height is disregarded, the dose
map accuracy is not guarantee.
6. Connect the ionization chamber to dosimeter controller.
7. Set the dosimeter controller in dose mode (cumulated dose).
8. Stick the dosimeter probe to the PMMA by using scotch tape (or any equivalent mean).
9. Use Fluoro to center the ionization chamber in the X-ray beam. Center the ionization chamber in the image as
shown in the Note. During this step do not move the table height and the table lateral axis.
Note: The ionization chamber position is very critical for the accuracy measurements. To
determine the center of the display area, plot on the monitor its two diagonals as shown on the
illustration below or use a ruler. The center of the display area and the center of the ionization
chamber must merge.
10.End exam.
3-2-4 Analyze the results
1. Compute "Dose3" in mGy by applying the following formula:
Dose3 (mGy) = Dose2 (%) x Adult local dose threshold (Gy) x 1000
Note: The Adult local dose threshold is set in the Preferences tab on the DL browser (refer to
Innova Digital / Dose Map / How to set Dose Map? / Dose Map preferences).
For example with Adult local dose threshold at 2 Gy and Dose2 at 37.5%:
Dose3 (mGy) = 37.5/100 x 2 x 1000 = 750 mGy
2. Compute the accuracy of Dose Map by applying the following formula:
Inaccuracy (%) = abs (Dose1 - Dose3) / Dose1 x 100
For example with Dose1 at 940 mGy and Dose3 at 750 mGy:
Inaccuracy (%) = abs (940 - 750) / 940 x 100 = 20%
Chapter 25 - Glossary
1 Acronyms in use
Acronym Defi- Comment Generic Standards Unit Medical appli- System com-
nition cation ponent
2D 2 Dimensional X
Application
3D 3 Dimensional X
Application
ABC Automatic X
Brightness Con-
trol
AE-Title Application Enti- X
ty Title
ALARA As Low As Rea- X
sonably Achiev-
able
AW Advantage X
Workstation
CAU Caudal X
CCD Coupled X
Charged Device
CDRH Center for De- X
vices and Radi-
ological Health
CFR Code of Federal X
Regulations
CFX Circumflex X
artery
CISPR Comité Interna- X
tional Special
des Perturba-
tions
Radioélectrique
s
cm centimeter X
CPD Chamber-Pa- X
tient Distance
CPR Cardiac Pul- X
monary Resus-
citation
CRA Cranial X
CSA Canadian Stan- X
dards Associa-
tion
664 Glossary
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
CT Computed To- X
mography (or
equivalent ap-
plication)
DAP Dose Area X
Product
DFP Digital Flat Pan- X
el
DICOM Digital Imaging X
and COmmuni-
cations in Medi-
cine
DL Digital Leader X
DQE Detective Quan- X
tum Efficiency
DSA Digital Subtract- X
ed Angiography
DVD+R DVD Record- X
able
DVD+RW DVD X
ReWritable
ECG ElectroCardioG- X
raphy
ED End-Diastolic X
EFUP Environment-Friend- X
ly Use Period
EIP Electronic Infor- X
mation Products
EMC Electro Magnet- X
ic Compatibility
EP Electro Physiol- X
ogy
ES End-Systolic X
F french X
FDA Food and Drug X
Administration
FNR Fluoro Noise X
Reduction
FOV Field of View X
fps Frame Per Sec- X
ond
GE General Electric X
GEF Global Ejection X
Fraction
Glossary 665
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
666 Glossary
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
Glossary 667
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
PM Periodic Mainte- X
nance
PMMA PolyMethyl- X
MethAcrylate
PMS Periodic Mainte- X
nance Schedule
QAP Quality Assur- X
ance Program
RAO Right Anterio X
Oblique
RCA Right Coronary X
Artery
RDL Receptor Dose X
Limited
REC Record X
RF Radio Frequen- X
cy
ROI Region Of Inter- X
est
RTAC Real Time Ac- X
quisition Con-
troller
s second X
SA Stenosis Analy- X
sis
SCD Source- X
Chamber Dis-
tance
SCPT Secondary Cap- X
ture
SID Source to Im- X
age Distance
SNR Signal to Noise X
Ratio
SUB Subtraction X
SUIF Service User In- X
terface
TFT Thin-Film Tran- X
sistors
TSSC Table Side Sta- X
tus Control
TSUI Table Side User X
Interface
668 Glossary
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
Glossary 669
InnovaTM IGS 520, InnovaTM IGS 530, InnovaTM IGS 540
Direction 5499294-1-1EN , Revision 2
www.gehealthcare.com