Professional Documents
Culture Documents
SSB 800 07
SSB 800 07
SSB 800 07
SSB-800-07
AUTOMATIC
Page 1 of 2
12/31/2007
Automatic Switchover Medical Liquid x High Pressure Manifold with Automatic Reset
(LifeLine)
SPECIFICATION
Automatic Switchover Medical Liquid x High includes gas specific pigtail-to-header high-flow check
Pressure Manifold valves to permit changing of cryogenic container(s) (left
The BeaconMedaes Lifeline automatic switchover bank) without gas leakage. Thermoplastic hose
manifold accommodates a left bank of cryogenic assemblies, designed for cryogenic service, are provided
containers and a right bank of high pressure cylinders for each container gas connection. In addition, the
while providing an uninterrupted, reliable supply of gas. manifold includes high-pressure modular header
Left bank is designated as “Primary” source of gas and assemblies with gas specific pigtail-to-header check
right bank stands in reserve as “Secondary” source. valves to permit changing of cylinders (right bank)
The manifold system referred to as “liquid x high- without gas leakage. Stainless steel flexible pigtails are
pressure” or “LQ x HP”, consists of a manifold control provided for each cylinder gas connection, except for
panel with its wall mounting bracket and a high- oxygen, helium, and mixed gas applications which are
pressure reserve header assembly (separately ordered, provided with rigid copper pigtails. A separate power
see SSB-800-08). The manifold is cleaned, tested, and supply is furnished and converts 120 VAC to 24 VAC
prepared for the indicated gas service and constructed in output power. The power supply provides dry normally
accordance with requirements of the latest edition of closed contacts for (3) alarm conditions: secondary
NFPA 99 and CGA. supply, reserve in use, and reserve low. The power
supply is housed in a NEMA 3R enclosure with
Manifold Design electrical requirements of (0.125) amp at 120 VAC,
A bank regulator (one for each supply bank) is used to 60Hz, single phase. External LED visual indicators for
initially reduce the container / cylinder pressure to the “RESERVE IN USE” (red) and “RESERVE LOW”
two line regulators which control the final line pressure. (red) are included within the power supply assembly.
Both line pressure regulators are in service at all times The manifold is supplied with ¾” FNPT (manifold
to maximize flow rates. The manifold automatically outlet) and ½” FNPT (high-preserve reserve header
changes from the depleted primary supply bank to the inlet) o-ring sealed “zero clearance” unions. A ¾” full
secondary supply bank without fluctuation in line port, three piece, ball-type source shut-off valve with a
pressure utilizing dome-bias loading and unloading of 1/8” FNPT port is also included. Source valve has a ¾”
the bank regulators. When liquid container(s) are NPT attachment to the outlet union and a ¾” nominal
replenished, the manifold will automatically switch copper (type k) tube for brazing to main supply line.
back to the primary (left) supply bank. Manual
resetting of the control panel is not necessary. If the NOTE:
primary and secondary supply banks of container(s) / The flow capacity of a nitrous oxide and carbon
cylinders are both depleted, the manifold system will dioxide manifold depends upon environmental
automatically switch to the external high-pressure conditions at the installation site and the number of
reserve header assembly. A pressure switch is provided containers in service. Installing a nitrous oxide or a
to signal the master gas alarm system just before change carbon dioxide manifold in a location that exposes it
over from the secondary (right) supply bank to the high- to an ambient temperature below 32O F (0O C) is not
pressure reserve. The manifold includes a line pressure recommended.
gauge, two bank contents gauges (left-bank and right-
bank), and LED visual indicators for “IN USE” (green), Environmental Considerations
“READY” (yellow), and “EMPTY” (red) for each bank. Manifolds are to be installed in accordance with
The manifold has bank regulator, intermediate and line requirements stated by NFPA 99, CGA, and all
pressure relief valves internally connected to a common applicable local codes. Manifold components are
vent port, terminating into a ½” FNPT o-ring sealed designed to work best over a temperature range of 32O F
“zero clearance” union. Master shutoff valves (one for through 130O F. Wider temperature variations may
each bank) are located within the manifold cabinet and cause manifold malfunctions to occur. BeaconMedaes
are fabricated with metallic seating surfaces. The recommends the control cabinet be located with its
cabinet enclosure is easily removable by releasing draw power supply at an installation site protected from
latches for component accessibility. The enclosure may moisture, continuous exposure to direct sun rays, ice
be secured from unauthorized access by locking the and snow.
draw latches (locks provided by others). The manifold
BeaconMedæs 1800 Overview Drive Rock Hill, SC 29730 Phone: (803) 817-5600 Fax: (803) 817-5750
SSB-800-07
Page 1 of 2
12/31/2007
# of HP Cylinders on
4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Right Bank
Overall Length 33” 38” 43” 48” 53” 58” 63” 68” 73” 78” 83” 88” 93” 98” 103”
BeaconMedæs 1800 Overview Drive Rock Hill, SC 29730 Phone: (803) 817-5600 Fax: (803) 817-5750